K Number
K161527

Validate with FDA (Live)

Date Cleared
2017-07-21

(414 days)

Product Code
Regulation Number
862.1225
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Creatinine Enzymatic Reagent Kit is a device which is intended for measurement of creatinine level in human serum, in vitro diagnostic use only. Test results may provide information regarding the status of kidney function and the diagnosis of renal diseases, and also serve as a component of several calculations for determination of creatinine clearance or glomerular filtration rate (GFR).

Device Description

Creatinine Enzymatic Reagent Kit is a dual reagent one contains Good's buffer, creatine amidinohydrolase, sarcosine oxidase and ESPMT (3-(N-Ethyl-3methylanilino) propanesulfonic acid sodium salt). Reagent two contains Good's buffer, creatinine amidohydrolase, Peroxidase and 4-aminoantipyrine.

AI/ML Overview

The provided text describes the performance characteristics and acceptance criteria for the Teco Creatinine Enzymatic Reagent Kit, but it does not detail a study involving human readers or AI assistance. The device is a diagnostic test kit, not an AI-powered diagnostic system requiring human interpretation.

Therefore, questions related to human reader performance, AI assistance, multi-reader multi-case studies, and human interpretation of ground truth are not applicable to this document. The information provided focuses solely on the analytical performance of the diagnostic reagent kit.

Here's the breakdown of the information that is applicable from your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly list acceptance criteria in a dedicated table, but it presents performance data against established guidelines (CLSI standards) and compares it to a predicate device. The "acceptance criteria" are implied by meeting CLSI standards and demonstrating comparable performance to the predicate.

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Teco Creatinine Enzymatic Reagent Kit)
Intended UseQuantitative determination of creatinine in human serumQuantitative determination of creatinine in human serum
SpecimenSerumSerum
MethodologyEnzymaticEnzymatic
Linearity (Measuring Range)Comparable to predicate (0.04 to 5.1 mg/dL for serum)0.37 - 5.06 mg/dL
Correlation (R²)High correlation with predicate (R closer to 1.0)R² = 0.9986 (compared to predicate)
Storage2-8 °C2-8 °C
Intra-assay RepeatabilityLow CV% (e.g., < 2%)Patient Pool 1: 0.93% CV; Patient Pool 2: 0.87% CV; Patient Pool 3: 0.90% CV
Within-Laboratory PrecisionLow CV% (e.g., < 3%)Patient Pool 1: 1.98% CV; Patient Pool 2: 2.18% CV; Patient Pool 3: 1.08% CV
Limit Of Blank (LoB)Low (ideally 0.00 mg/dL)0.00 mg/dL
Limit Of Detection (LoD)Low (e.g., < 0.05 mg/dL)0.01 mg/dL
Limit Of Quantitation (LoQ)Low (e.g., < 0.1 mg/dL) with CV% < 20%0.10 mg/dL (with accuracy goal of CV < 20%)
Analytical SpecificityMinimal interference from common substancesMost substances (<10% interference); list of interfering substances provided (e.g., conjugated bilirubin at 24.95 mg/dL, triglycerides at 3000 mg/dL)
Reference IntervalVerified against established ranges (Larsen K., 1972)Men: 0.9-1.5 mg/dL; Women: 0.7-1.4 mg/dL

2. Sample size used for the test set and the data provenance

  • Precision/Reproducibility:
    • Sample Size: Three serum pools. For intra-assay precision, each of the three pools was assayed for 20 operating days, twice per day, in triplicates (3 pools * 20 days * 2 runs/day * 3 replicates/run = 360 measurements per pool per reagent lot, for two reagent lots).
    • Data Provenance: "Patient serum pools" are mentioned. The specific country of origin is not stated, but the company is US-based (Anaheim, CA), suggesting data may be from the US or procured globally for clinical lab use. The studies are prospective in nature, as they involve actively "assaying samples."
  • Linearity/Assay Reportable Range:
    • Sample Size: Eleven samples, assayed in duplicates for each reagent lot.
    • Data Provenance: "Patient serum pools." Prospective.
  • Detection Limit (LoB, LoD, LoQ):
    • LoB: Four diluted real pooled serum samples. Fifteen replicates for each sample, using two reagent lots (4 samples * 15 replicates * 2 lots = 120 measurements).
    • LoD: Four diluted serum samples (100x, 50x, 33x, 25x). Fifteen times for each sample (4 samples * 15 replicates = 60 measurements).
    • LoQ: Four low concentration samples from native serum sample pools. Fifteen times each (4 samples * 15 replicates = 60 measurements).
    • Data Provenance: "real pooled serum samples," "native serum sample pools." Prospective.
  • Analytical Specificity:
    • Sample Size: Samples with increasing amounts of potential interferents were tested in triplicate. Two different clinically relevant concentrations of creatinine (1.50 and 5.00 mg/dL) were used. The number of interferents tested is listed as 15. So, for each interferent, at least (2 creatinine conc * X interferent conc * 3 replicates) measurements were done.
    • Data Provenance: Not specified, but likely controlled laboratory samples. Prospective.
  • Method Comparison:
    • Sample Size: 98 serum samples.
    • Data Provenance: "Serum samples." Not explicitly stated if patient samples or contrived. Given the context of comparing to a predicate, historical patient samples are likely. The study is presented as a comparison, which implies a retrospective analysis against existing methods.
  • Expected Values/Reference Interval:
    • Sample Size: 20 healthy male donors and 20 healthy female donors (total 40 donors).
    • Data Provenance: Human serum samples from healthy donors (age 20-65). Prospective study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable as this is a diagnostic reagent kit. "Ground truth" is established by the analytical method itself (e.g., reference method for method comparison studies, or a gold standard for traceability) and the physiological state of the samples (e.g., healthy donors, patient pools). There are no human experts interpreting images or signals to establish ground truth in the way described for AI/human-in-the-loop systems.

The creatinine calibrator values are traceable to NIST reference material (SRM 967), which serves as a metrological ground truth.

4. Adjudication method for the test set

Not applicable. This device does not involve human interpretation or adjudication for its results.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No human readers or AI assistance are involved with this diagnostic kit.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, in a sense, the entire performance study is a "standalone" assessment of the reagent kit's analytical performance (analogous to an "algorithm only" performance for an AI system), as it measures the kit's ability to accurately quantify creatinine without human interpretative input influencing the result. The device is purely an in vitro diagnostic reagent kit.

7. The type of ground truth used

  • Reference method/Predicate device: For method comparison, the "Stanbio Creatinine LiquiColor test system" serves as the comparative reference.
  • NIST Reference Material (SRM 967): For traceability of calibrator values.
  • CLSI guidelines: For various analytical performance characteristics (precision, linearity, detection limits, specificity).
  • Established biological ranges/clinical literature: For expected values/reference intervals (e.g., "Larsen K. Clin Chem Acta 41:209, 1972").
  • Known concentrations: For studies like linearity, detection limits, and analytical specificity, samples are prepared with known concentrations of analyte or interferents.

8. The sample size for the training set

Not applicable. This is a chemical reagent kit, not a machine learning model, so there is no concept of a "training set" in the context of data used for algorithm development. Its development would involve chemical optimization and formulation rather than data-driven training.

9. How the ground truth for the training set was established

Not applicable (no training set for this type of device).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus or a staff with a serpent entwined around it, but in this case, it is a more abstract, flowing design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 21, 2017

TECO DIAGNOSTICS, INC. XIAOYAN HU RESEARCHER 1268 NORTH LAKEVIEW AVENUE ANAHEIM CA 92807

Re: K161527

Trade/Device Name: Teco Creatinine Enzymatic Reagent Kit Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: II Product Code: JFY Dated: July 13, 2017 Received: July 14, 2017

Dear Xiaoyan Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161527

Device Name Teco Creatinine Enzymatic Reagent Kit

Indications for Use (Describe)

Creatinine Enzymatic Reagent Kit is a device which is intended for measurement of creatinine level in human serum, in vitro diagnostic use only. Test results may provide information regarding the status of kidney function and the diagnosis of renal diseases, and also serve as a component of several calculations for determination of creatinine clearance or glomerular filtration rate (GFR).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a green hexagon with the letters "TC" in white. The letters are in a sans-serif font and are slightly offset from each other. The hexagon is a solid green color and has a slight gradient effect, making it appear slightly darker at the bottom. The image is simple and clean, with a focus on the logo itself.

This Summary of 510(k) Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR § 807.92.

Owner's Name: Teco Diagnostics, Inc.

Address and Contact information:

1268 North Lakeview Avenue Anaheim, CA 92807 Phone: 714-463-1111 Fax: 714-463-1169

Contact:

Xiaoyan Hu

Date Prepared:

July 20, 2017

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image features a green hexagon with the letters 'TC' in silver. The letters are stylized with a slight 3D effect, giving them a rounded appearance. The green background is a solid color, and the hexagon shape is well-defined. The overall design is simple and modern.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiagnostics.com

510(k) SUMMARY

A. 510(k) Number:

  • K161527 B. Analytes:
    Creatinine

  • C. Type of Test: Enzymatic, quantitative

  • D. Applicant: Teco Diagnostics. Inc. E. Trade Name:

Teco Creatinine Enzymatic Reagent Kit

  • F. Common Name: Enzymatic Method, Creatinine
    G. Regulatory Information:

    1. Regulation Classification section: Class II: 21 CFR § 862.1225-Creatinine test system
    1. Product Code: JFY
    1. Panel:

75, Clinical Chemistry

H. Intended Use:

    1. Indication(s) for use:
      Creatinine Enzymatic Reagent Kit is a device which is intended for measurement of creatinine level in human serum, in vitro diagnostic use only. Test results may provide information regarding the status of kidney function and the diagnosis of renal diseases, and also serve as a component of several calculations for determination or estimation of creatinine clearance or glomerular filtration rate (GFR).
    1. Special conditions for use statement(s):
      This is an in vitro diagnostic test for prescription use only.
    1. Special instrument requirements: TC-MATRIX analyzer

I. Device Description

Creatinine Enzymatic Reagent Kit is a dual reagent one contains Good's buffer, creatine amidinohydrolase, sarcosine oxidase and ESPMT (3-(N-Ethyl-3methylanilino) propanesulfonic acid sodium salt). Reagent two contains Good's buffer, creatinine amidohydrolase, Peroxidase and 4-aminoantipyrine.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a green hexagon with the letters "TC" in silver. The letters are stacked on top of each other, with the "T" on top and the "C" on the bottom. The hexagon has a slight gradient, with the top being slightly lighter than the bottom. The overall impression is that of a logo or icon.

J. Substantial Equivalence Information:

    1. Predicate device name(s): Stanbio Creatinine LiquiColor test system
    1. Predicate K number(s): K050283
    1. Compare with predicate:

Table 5.1 Creatinine Enzymatic Reagent Kit vs Stanbio Creatinine LiquiColor test system

Teco Creatinine EnzymaticReagent KitStanbio CreatinineLiquiColor test system
Intended UseFor the quantitative determinationof creatinine.Same
Intended SpecimenSerumSerum and urine
Test MethodologyEnzymaticSame
LinearitySerum: 0.37- 5.06 mg/dLSerum: 0.04 to 5.1 mg/dLUrine: 0.04 to 200 mg/dL
WavelengthPrimary: 546 nmSecondary: 670 nm546 nm
Accuracy/Correlation(Serum)y = 1.004x+ 0.0379R2 = 0.9986y = 0.9993x+0.0018R = 0.9999
Storage2- 8 °CSame

K. Guideline Document Referred (if applicable):

  1. CLSI EP5-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline Third Edition.

  2. CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline.

  3. CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline Second edition.

  4. CLSI EP9-A3, Method Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline Third edition.

  5. CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approve Guideline Second Edition.

  6. CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline.

  7. CLSI EP28-A3C, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory: Approved Guideline Third Edition.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a green hexagon with the letters "TC" in white. The letters are in a sans-serif font and are slightly offset from each other. The hexagon has a subtle gradient, with the top being slightly lighter than the bottom. The overall impression is clean and modern.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiagnostics.com

510(k) SUMMARY

L. Test Principle:

Creatinase in Reagent 1 hydrolyzes existing creatine in a sample to sarcosine and urea. Sarcosine is oxidized by sarcosine oxidase to glycine and formaldehyde and hydrogen peroxide. This hydrogen peroxide is reduced to water soon. Ceatininase in Reagent 2 hydrolyzes creatinine in sample to creatine. By the same reaction steps above, the hydrogen peroxide from creatinine reacts with 4-aminoantipyrine and ESPMT in the presence of peroxidase to yield a quinoneimine dye. The change in absorbance at 546 nm is proportional to the creatinine concentration in sample.

M. Performance Characteristics:

1. Analytical performance:

  • a. Precision/Reproducibility:
    Precision studies were performed based on upon the guidelines set forth in CLSI EP5-A. Intra-assay Precision was performed by assaying three serum pools at three concentrations over 20 operating days, on a single instrument at a single site and with two lots of reagents. Assay runs were performed twice per day and each sample was tested in triplicates. Results of the precision studies were summarized below:
Sample DescriptionMean (mg/dL)RepeatabilityWithin-Laboratory Precision
SDCV%SDCV%
Patient Pool 10.890.0083530.930.017611.98
Patient Pool 21.520.013230.870.033162.18
Patient Pool 33.090.028000.900.033611.08

I OT· 080201

LOT: 080202

SampleDescriptionMean(mg/dL)RepeatabilityWithin- LaboratoryPrecision
SDCV%SDCV%
Patient Pool 10.880.009221.050.020092.28
Patient Pool 21.510.010390.690.034022.25
Patient Pool 33.040.027360.900.036821.21

b. Linearity/assav reportable range:

Linearity studies were performed based on upon the guidelines set forth in CLSI EP6-A. Linearity was performed by preparing samples from patient serum pools at a low creatinine concentration and a high concentration. Eleven samples in all were assayed in duplicates for each reagent lot. The expected values were plotted against

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is a logo with a green hexagon as the background. Inside the hexagon are the letters 'TC' in white. The letters are slightly shadowed, giving them a three-dimensional appearance. The logo is simple and clean, with a focus on the letters 'TC' against the green background.

the observed values and a linear regression line from one of the reagent lots fitted with the following regression equation:

Expected Value(mg/dL)Observed Value(mg/dL)
0.370.374
0.940.946
1.501.518
2.062.090
2.712.663
3.313.235
3.833.807
4.384.379
4.904.951
5.495.523
6.126.095

Considering these testing concentrations from Linearity study and Method Comparison study, and the assay range of predicate device (0.04 - 5.1 mg/dL), the claimed measuring range for Teco Creatinine Enzymatic Reagent Kit is 0.37- 5.06 mg/dL. The regression is y = 0.9992x + 0.0007, R2 = 0.9983.

  • c. Traceability, Stability, Expected values:
    The stability claim was verified using an accelerated stability protocol. Real-time testing was conducted. The study results are summarized below:
Storage ConditionsClaimed stability
Closed-vial$37 \pm 2$ °C20 days
Open-vial2 - 8°C, 20-30% humidity60 days
Real-time, closed-vial2 - 8°C, 20-30% humidityPending

Traceability

The Teco chemistry calibrator values were obtained by multi determinations performed by Teco Diagnostics. The assigned value is the median of all values obtained. The assigned value of the calibrator is traceable to NIST reference material (SRM 967).

  • d. Detection limit:
    Limit Of Blank (LoB):

The study was determined following CLSI EP17-A2 as a guideline. Four diluted real pooled serum samples were tested in the limit of blank study. Fifteen replicates

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a green hexagon with the letters "TC" in silver. The letters are stacked on top of each other, with the "T" on top and the "C" on the bottom. The letters are in a sans-serif font and have a slight shadow effect. The hexagon is a solid green color and has a slight gradient effect.

of each sample were assayed using TC-MATRIX instrument and two reagent lots, with the total number of measurements being 60 for each lot.

Based upon the studies conducted for LoB, the data were assessed using the nonparametric data analysis option, and the following results were obtained: 0.00 mg/dL.

Limit Of Detection (LoD):

The LoD study was determined following CLSI EP17-A2 as a guideline. In the Limit Of Detection the serum was determined by using two reagent lots. Four diluted serum samples of 100 times, 50 times, 33 times and 25 times were assayed 15 times for each sample (N = 60) on the TC-MATRIX instrument. Based upon the studies conducted for LoD and the reported LoB determined above, this gave result of 0.01 mg/dL.

Limit Of Quantitation (LoQ):

The LoQ study was determined following CLSI EP17-A2 as a guideline. For LoQ determination, four low concentration samples were prepared by native serum sample pools and assayed fifteen times each (N = 60). The CV% and the SD at each tested concentration were calculated. The LoQ accuracy goal is within 20% CV.

Results of Summary:

LoBLoDLoQ
Value (mg/dL)0.000.010.10

e. Analytical specificity:

Testing for interfering substances was based on CLSI EP7-A2. Testing was performed with two different clinically relevant concentrations of creatinine (1.50 and 5.00 mg/dL). Samples with increasing amounts of potential interferents were tested in triplicate and compared to a control sample without the interferent. The concentrations of substances that demonstrated no significant interference (≤10%) are listed below:

InterferenceSubstanceHighestconcentration testedwith no significantinterference (≤10%)
Unconjugated Bilirubin20 mg/dL
Conjugated Bilirubin16.7 mg/dL
Hemoglobin500 mg/dL
Triglycerides2500 mg/dL

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a green hexagon with the letters "TC" in white. The letters are stacked on top of each other, with "T" on top and "C" on the bottom. The hexagon has a gradient effect, with the top being slightly lighter than the bottom. The letters also have a slight shadow effect, making them appear to be raised from the surface of the hexagon.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiagnostics.com

510(k) SUMMARY

Ascorbic Acid20 mg/dL
Creatine6.0 mg/dL
Glucose1000 mg/dL
Acetaminophen20 mg/dL
Nitrofurantoin400 ug/dL
Uric acid25 mg/dL
Dopamine90.0 ug/dL
Methyldopa0.10 mg/dL
L- Dopa0.06 mg/dL
Calcium dobesilate1 mg/dL

The following table shows the substances which did interfere with Teco Creatinine Enzymatic Reagent Kit. The concentrations of interfering substance that demonstrated a significant interference are listed below:

Interference SubstanceConcentration ofSubstance at whichInterference wasobserved
Conjugated Bilirubin24.95 mg/dL
Triglycerides3000 mg/dL
Creatine8.0 mg/dL
Methyldopa0.45 mg/dL
L- Dopa0.11 mg/dL
Calcium dobesilate3 mg/dL

A summary of the interferents with effects is found in the labeling.

2. Comparison studies:

Method comparison

The performance of the Creatinine Enzymatic Reagent Kit was compared with the Stanbio Creatinine LiquiColor test system on TC-MATRIX analyzer. Study performed according to CLSI EP9-A3 using 98 serum samples covering the measuring range with values from predicate and the result from the linearity study in M 1.b. The linear regression is summarized as y=1.004x+0.0379, R2=0.9986.

Based on the detection limits determined in this study, the result from the method comparison study in M 2.a. and the result from the linearity study in M 1.b., the labeling claims a measuring range of 0.37 to 5.06 mg/dL.

3. Expected values/Reference interval:

The expected reference interval is derived from: Larsen K. Clin Chem Acta 41:209, 1972. The reference interval for Creatinine was verified using CLSI guideline EP28-

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows a green hexagon with the letters "TC" in white. The letters are in a sans-serif font and are slightly offset from the center of the hexagon. The hexagon has a subtle gradient, with the top being slightly lighter than the bottom. The overall impression is that of a logo or icon.

A3 by conducting a small study using human serum samples from 20 healthy male donors and 20 healthy female donors, from 20 to 65 years old. The samples were assayed on TC-MATRIX analyzer. All results from the total 40 healthy donors were found to be within the reported reference range. The Reference Interval for the study for Teco Creatinine Enzymatic Reagent Kit shows:

MenWomen
Serum (mg/dL)0.9 - 1.50.7 - 1.4

N. Conclusion:

The performance characteristics of Teco Creatinine Enzymatic Reagent Kit were verified by comparison, precision, linearity, detection limit, specificity and interference, shelf life, and stress studies. Testing results indicate that Teco Creatinine Enzymatic Reagent Kit performs satisfactorily when used appropriately, as outlined in the package insert. The result of performance studies demonstrates a substantial equivalency on performance between Teco Creatinine Enzymatic Reagent Kit and the predicate device, Stanbio Creatinine LiquiColor test system.

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.