K Number
K161509

Validate with FDA (Live)

Device Name
3D OCT-1 Maestro
Manufacturer
Date Cleared
2016-07-28

(57 days)

Product Code
Regulation Number
886.1570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Topcon 3D OCT-1 Maestro is a non-contact, high resolution tomographic and biomicroscopic imaging device that incorporates a digital camera for photographing, displaying and storing the data of the retina and surrounding parts of the eye to be examined under Mydriatic and non-Mydriatic conditions.

The 3D OCT-1 Maestro is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging and measurement of posterior ocular structures, including retinal nerve fiber layer, macula and optic disc as well as imaging of anterior ocular structures.

It also includes a Reference Database for posterior ocular measurements which provide for the quantitative comparison of retinal nerve fiber layer, optic nerve head, and the human retina to a database of known normal subjects. The 3D OCT-1 Maestro is indicated for use as a diagnostic device to aid in the diagnosis, documentation and management of ocular health and diseases in the adult population.

Device Description

The Maestro is a non-contact, high-resolution, tomographic and bio-microscopic imaging system that merges OCT and fundus cameras into a single device. The technological characteristics of the OCT employed are similar to those of already 510(k)-cleared OCT products, such as Topcon's 3D OCT-2000 (K092470), in that it employs conventional spectral domain OCT with widely-used 840 nm light source. The technological characteristics of the fundus camera employed are also similar to those of already cleared fundus cameras, such as Topcon's TRC NW300 (K123460), in terms of field of view (FOV) and camera sensor resolution.

The Maestro captures an OCT image and a color fundus image sequentially. It can take anterior OCT images in addition to fundus OCT images. It also includes a reference database for fundus OCT. Captured images are transferred from the device to an off-the-shelf personal computer (PC) via LAN cable, where the dedicated software for this device is installed. The transferred data is then automatically processed with analysis functions such as the automatic retinal layers segmentation, the automatic thickness calculation with several grids, the optic disc analysis and comparison with a reference database of eyes free of ocular pathology, and is finally automatically saved to the PC.

Two software programs for installation on an off-the-shelf PC are provided with the device. The first PC software program, called "FastMap", captures the images from the device, analyzes them and enables viewing of the data. The second PC software program, called "OCT Viewer", can only analyze and view the data.

Accessories include the following: power cord; chin-rest paper sheet; LAN cable; chin-rest paper pins; external fixation target; dust cover; spare parts case; and stylus pen. An optional Anterior Segment Kit allows the user to activate the anterior segment imaging functionality of the Maestro device.

AI/ML Overview

The Topcon 3D OCT-1 Maestro is a non-contact, high-resolution tomographic and biomicroscopic imaging device. The provided text outlines its performance data, primarily focusing on repeatability and reproducibility measurements for various ocular structures in different patient populations.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria for the repeatability and reproducibility of the measurements. Instead, it presents the calculated repeatability and reproducibility measurements (SD, Limit, CV%) for the Maestro device across different parameters and patient groups. The "acceptance criteria" appear to be implied by the presentation of these results, suggesting that the device's performance, as measured, is considered acceptable for demonstrating substantial equivalence to predicate devices.

However, based on the provided tables, here's a summary of the reported device performance:

Measurement TypePopulationScan PatternTypical Repeatability CV% (Range)Typical Reproducibility CV% (Range)
Full Retinal ThicknessNormal Eyes12x9 3D Wide0.286% - 1.115%0.526% - 1.461%
6x6 3D Macula0.305% - 0.684%0.498% - 1.025%
Retinal Disease Eyes12x9 3D Wide0.378% - 1.478%0.595% - 1.897%
6x6 3D Macula0.376% - 1.090%0.660% - 1.336%
Glaucoma Eyes12x9 3D Wide0.493% - 1.199%0.661% - 1.639%
6x6 3D Macula0.332% - 1.288%0.719% - 1.239%
Ganglion Cell + IPLNormal Eyes12x9 3D Wide0.404% - 0.950%0.508% - 1.162%
6x6 3D Macula0.364% - 1.044%0.557% - 1.148%
Retinal Disease Eyes12x9 3D Wide1.041% - 2.673%1.101% - 3.604%
6x6 3D Macula0.690% - 1.452%0.984% - 1.824%
Glaucoma Eyes12x9 3D Wide0.628% - 1.563%0.716% - 1.784%
6x6 3D Macula0.593% - 1.288%0.736% - 1.451%
Ganglion Cell Complex ThicknessNormal Eyes12x9 3D Wide0.470% - 0.821%0.645% - 1.056%
6x6 3D Macula0.498% - 1.400%0.729% - 1.607%
Retinal Disease Eyes12x9 3D Wide1.112% - 3.213%1.112% - 3.232%
6x6 3D Macula0.485% - 1.093%0.601% - 1.093%
Glaucoma Eyes12x9 3D Wide0.638% - 1.189%0.687% - 1.240%
6x6 3D Macula0.508% - 1.131%0.678% - 1.265%
Retinal Nerve Fiber Layer (RNFL) - AverageNormal Eyes12x9 3D Wide1.318%1.517%
6x6 3D Disc0.933%1.099%
Retinal Nerve Fiber Layer (RNFL) - SectoralNormal Eyes12x9 3D Wide2.461% - 16.711%3.040% - 18.538%
6x6 3D Disc3.738% - 13.898%4.405% - 14.407%
Retinal Disease Eyes12x9 3D Wide1.594% - 8.143%1.888% - 8.675%
6x6 3D Disc1.084% - 5.725%1.480% - 7.387%
Glaucoma Eyes12x9 3D Wide1.970% - 8.261%2.097% - 8.299%
6x6 3D Disc1.929% - 6.480%1.933% - 7.074%
Optic DiscNormal Eyes12x9 3D Wide3.520% - 6.600%4.233% - 7.967%
6x6 3D Disc3.313% - 6.359%4.074% - 8.139%
Retinal Disease Eyes12x9 3D Wide3.858% - 8.404%4.981% - 20.586%
6x6 3D Disc2.855% - 5.627%3.438% - 11.024%
Glaucoma Eyes12x9 3D Wide3.179% - 14.274%3.811% - 17.103%
6x6 3D Disc1.852% - 5.813%1.959% - 7.201%

The "Limit" values in the tables are calculated as 2.8 x SD, representing a range within which 95% of repeated measurements are expected to fall. The "CV%" is the Coefficient of Variation, indicating precision relative to the mean.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Test Set (Clinical Studies for Repeatability and Reproducibility):
    • Normal Subjects: 25 subjects for macula and optic disc measurements (full retinal thickness, ganglion cell + IPL, ganglion cell complex thickness, retinal nerve fiber layer, optic disc parameters). Explicitly stated in the tables (N=25).
    • Subjects with Retinal Disease: 26 subjects for macula and optic disc measurements (full retinal thickness, ganglion cell + IPL, ganglion cell complex thickness, retinal nerve fiber layer, optic disc parameters). Explicitly stated in the tables (N=26).
    • Subjects with Glaucoma: 25 subjects for macula and optic disc measurements (full retinal thickness, ganglion cell + IPL, ganglion cell complex thickness, retinal nerve fiber layer, optic disc parameters). Explicitly stated in the tables (N=25).
  • Data Provenance: The document does not explicitly state the country of origin. The study is referred to as "clinical studies," but it's not specified if they were prospective or retrospective. Given that the manufacturer is Topcon Corporation of Japan, but the contact person is in the US, the studies could have been conducted in either or both regions.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish a "ground truth" for the test set in the context of the repeatability and reproducibility studies. The clinical studies were conducted to determine the agreement, repeatability, and reproducibility of measurement data, not for diagnostic accuracy against a ground truth.

However, it mentions: "Consistent with the labeling for the test and control devices, the clinical site was permitted to make manual adjustments to automated segmentation based on the clinician's judgment." This indicates that clinicians (likely ophthalmologists or optometrists) were involved in reviewing and potentially adjusting automated segmentation, but it doesn't define them as "ground truth" experts in the context of defining disease states or specific measurements for the purpose of validating the AI's accuracy against a known truth.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method in the context of establishing ground truth for the test set. The clinical studies focused on repeatability and reproducibility of quantitative measurements, rather than classification or diagnosis that would typically require an adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study evaluating human reader improvement with or without AI assistance was not reported in this document. The clinical studies conducted were focused on the device's measurement precision (repeatability and reproducibility) and agreement with predicate devices rather than human-AI collaboration for diagnostic accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The clinical studies described primarily assessed the precision of the device's measurements (which involve algorithms for segmentation and thickness calculation) rather than the standalone diagnostic performance of an AI algorithm. The device performs automatic retinal layer segmentation, automatic thickness calculation, and optic disc analysis. The phrase "the clinical site was permitted to make manual adjustments to automated segmentation based on the clinician's judgment" (page 6) suggests that the device's algorithms operate with potential human oversight, implying it's not strictly a standalone AI performance evaluation for diagnostic purposes. The data presented are for the device's ability to consistently provide these measurements.

7. The Type of Ground Truth Used

For the repeatability and reproducibility studies, the "ground truth" is not a diagnostic label (e.g., pathology, outcomes data). Instead, the studies assess the consistency of the device's quantitative measurements of ocular structures (e.g., retinal thickness, RNFL thickness) by comparing multiple measurements taken under similar or varied conditions. The reference database uses "known normal subjects" but this is for comparative analysis against a normal population rather than for establishing a "ground truth" for disease diagnosis in the test set.

8. The Sample Size for the Training Set

The document specifies a "Reference Database" was compiled using "399 subject eyes from normal study subjects." This database is for "quantitative comparison... to a database of known normal subjects," which functions as a normative reference rather than a training set for an AI/algorithm in the conventional sense (e.g., for classification tasks).

If parts of the device's functionality (like automatic segmentation) involve machine learning, the training set size for those specific algorithms is not provided in this document. The 399 normal subjects form a reference database, not explicitly an algorithm training set.

9. How the Ground Truth for the Training Set Was Established

For the "Reference Database" of 399 normal subjects:

  • How it was established: The study collected measurements of various ocular structures from these normal eyes. The "normal" status of these subjects would have been established through clinical evaluation to ensure they were free of ocular pathology.
  • Type of Ground Truth: The ground truth for this reference database is the consensus clinical determination that the subjects have "normal eyes" and the quantitative measurements derived from these normal eyes form the expected range for a healthy population. The document states it provides "quantitative comparison... to a database of known normal subjects." It also mentions "a reference database of eyes free of ocular pathology."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines that resemble a ribbon or banner. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 28, 2016

Topcon Corporation % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004

Re: K161509

Trade/Device Name: 3D OCT-1 Maestro Regulation Number: 21 CFR §886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO, HKI Dated: June 1, 2016 Received: June 1, 2016

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K161509

Device Name

3D OCT-1 Maestro

Indications for Use (Describe)

The Topcon 3D OCT-1 Maestro is a non-contact, high resolution tomographic and biomicroscopic imaging device that incorporates a digital camera for photographing, displaying and storing the data of the retina and surrounding parts of the eye to be examined under Mydriatic and non-Mydriatic conditions.

The 3D OCT-1 Maestro is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging and measurement of posterior ocular structures, including retinal nerve fiber layer, macula and optic disc as well as imaging of anterior ocular structures.

It also includes a Reference Database for posterior ocular measurements which provide for the quantitative comparison of retinal nerve fiber layer, optic nerve head, and the human retina to a database of known normal subjects. The 3D OCT-1 Maestro is indicated for use as a diagnostic device to aid in the diagnosis, documentation and management of ocular health and diseases in the adult population.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

{3}------------------------------------------------

510(k) SUMMARY – K161509

Topcon Corporation's 3D OCT-1 Maestro

Submitter Information:

Applicant:Topcon Corporation75-1 Hasunuma-cho, Itabashi-kuTokyo, Japan 174-8580
Phone:(201) 599-5533
Fax:(201) 599-5248
Contact Person:Charles ReismanSenior Manager, Global R&DTopcon Medical Systems, Inc.111 Bauer DriveOakland, NJ 074361 (201) 599-5283
Date Prepared:July 25, 2016
Subject Device:
Device Name:3D OCT-1 Maestro
Common Name:Optical Coherence Tomography
Classification Name:Ophthalmoscope
Regulation:Class II, 21 C.F.R. § 886.1570
Product Code(s):OBO, HKI
Predicate Device(s):

Primary Predicate: Optovue's iVue with Normative Database (K121739) Topcon's TRC-NW300 Non-Mydriatic Retinal Camera (K123460) Secondary Predicate:

{4}------------------------------------------------

Indications for Use:

The Topcon 3D OCT-1 Maestro is a non-contact, high resolution tomographic and biomicroscopic imaging device that incorporates a digital camera for photographing, displaying and storing the data of the retina and surrounding parts of the eye to be examined under Mydriatic and non-Mydriatic conditions.

The 3D OCT-1 Maestro is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging and measurement of posterior ocular structures, including retinal nerve fiber layer, macula and optic disc as well as imaging of anterior ocular structures.

It also includes a Reference Database for posterior ocular measurements which provide for the quantitative comparison of retinal nerve fiber layer, optic nerve head, and the human retina to a database of known normal subjects. The 3D OCT-1 Maestro is indicated for use as a diagnostic device to aid in the diagnosis, documentation and management of ocular health and diseases in the adult population.

Device Description:

The Maestro is a non-contact, high-resolution, tomographic and bio-microscopic imaging system that merges OCT and fundus cameras into a single device. The technological characteristics of the OCT employed are similar to those of already 510(k)-cleared OCT products, such as Topcon's 3D OCT-2000 (K092470), in that it employs conventional spectral domain OCT with widely-used 840 nm light source. The technological characteristics of the fundus camera employed are also similar to those of already cleared fundus cameras, such as Topcon's TRC NW300 (K123460), in terms of field of view (FOV) and camera sensor resolution.

The Maestro captures an OCT image and a color fundus image sequentially. It can take anterior OCT images in addition to fundus OCT images. It also includes a reference database for fundus OCT. Captured images are transferred from the device to an off-the-shelf personal computer (PC) via LAN cable, where the dedicated software for this device is installed. The transferred data is then automatically processed with analysis functions such as the automatic retinal layers segmentation, the automatic thickness calculation with several grids, the optic disc analysis and comparison with a reference database of eyes free of ocular pathology, and is finally automatically saved to the PC.

Two software programs for installation on an off-the-shelf PC are provided with the device. The first PC software program, called "FastMap", captures the images from the device, analyzes them and enables viewing of the data. The second PC software program, called "OCT Viewer", can only analyze and view the data.

Accessories include the following: power cord; chin-rest paper sheet; LAN cable; chin-rest paper pins; external fixation target; dust cover; spare parts case; and stylus pen. An optional Anterior Segment Kit allows the user to activate the anterior segment imaging functionality of the Maestro device.

Substantial Equivalence

The Maestro is substantially equivalent to the primary predicate device Optovue's iVue with Normative Database (iVue) cleared under K121739. The Maestro and the iVue with Normative Database have very similar indication for use statements with respect to the reference database and its ability to provide reference analyses. The Maestro and iVue are both OCT devices and each provide cross-sectional images of the fundus and measurement of posterior ocular structures. Both devices provide analysis functions for both retinal disease and glaucoma. Also, both are intended for use as a diagnostic device to aid in the detection and management of various ocular diseases.

{5}------------------------------------------------

The Maestro and the iVue have similar technological characteristics with respect to the kind of OCT images they capture, the processing methods, and presentation of OCT information. The light source used by both devices is a spectral domain OCT (SD-OCT) using a superluminescent diode (SLD) with a center wavelength of 840 nm. The Maestro has a faster scan speed of 50,000 A-scans per second compared to 26,000 A-scans per second for the iVue. In addition, the Maestro has deeper A-scan depth of 2.6 mm (in tissue) at axial resolution of 6 micron (in tissue) and transverse resolution of 20 micron (in tissue) compared to A-scan depth of 2.0 to 2.3 mm at axial resolution of 5 micron and unspecified transverse resolution for the iVue. Both devices use the 3D Scan pattern and other similar scan patterns for posterior and Radial and Line-like scans for anterior. For the anterior segment feature, the Maestro provides only a visualization while the iVue also provides automated measurement of central corneal thickness.

In addition to the primary predicate device, Topcon's TRC-NW300 Non-Mydriatic Retinal Camera (NW300) (K123460) as a secondary predicate device for functionality. This secondary predicate device serves as a comparator device to the subject device since the scientific methodology and the device performance for the individual components have already been established. Comparative data between the subject and secondary predicate device will be provided in the device labeling.

Based on the intended use/indications for use and technological characteristics described above, the subject device is substantially equivalent to the primary predicate device.

Performance Data

Performance bench testing has been conducted on the Maestro to demonstrate substantial equivalence to the predicate devices. A list of testing conducted is as follows:

  • Axial and Horizontal Resolutions
  • . Trueness and Precision for Axial and Horizontal Directions
  • . Testing to Consensus Standards for Ophthalmic Devices
    • ISO 15004-1:2006 Ophthalmic instruments Fundamental requirements and test methods-Part 1: o General requirements applicable to all ophthalmic instruments
    • ISO 10940: 2009 Ophthalmic instruments Fundus cameras o
    • ISO 15004-2:2007 Ophthalmic Instruments Fundamental requirements and test methods Part 2: O Light hazard protection
  • Electrical Safety ●
    • o AAMI ANSI ES 60601-1:2005/A2:2010/(R)2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (Third Edition)
  • Electromagnetic Compatibility
    • IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety O and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests (Third Edition)

Biocompatibility of materials used in the patient contacting components were already evaluated and established in company's predicate device submissions.

{6}------------------------------------------------

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could result in an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

Clinical Data

Topcon conducted clinical studies to determine agreement, repeatability and reproducibility of the measurement data between the Maestro device and the iVue with normative database. A comparative study between images of the Maestro and the TRC NW300 Non Mydriatic Retinal Camera was also conducted. These studies included data from 25 normal subjects with retinal disease, and 25 subjects with glaucoma. The studies included five main analysis, 2) an agreement analysis, 3) an agreement analysis, 3) a fundus photograph evaluation, 4) an anterior segment image evaluation, and 5) a B scan image quality evaluation.

A reference database study of the Maestro was also conducted.

Consistent with the labeling for the test and control devices, the clinical site was permitted to make manual adjustments to automated segmentation based on the clinician's judgment. The study device manual adjustment rates for macula scans ranged from approximately 0-5% for 6x6 macula and 10-15% for 12x9 wide scans. For RNFL thicknesses and optic disc parameters, the overall manual adjustment rates for the Maestro were 11-32% and 20-40% for 6x6 disc and 12x9 wide scans, respectively.

Table 1: Full Retinal Thickness - Normal Eyes Repeatability and Reproducibility
RepeatabilityReproducibility
Device
MeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Central Fovea25233.992.6087.3021.1153.4199.5731.461
Superior Parafoveal25309.121.6474.6110.5332.4356.8190.788
Nasal Parafoveal25309.721.4704.1160.4752.5957.2660.838
Inferior Parafoveal25307.291.2813.5880.4172.2356.2590.727
Temporal Parafoveal25296.651.4173.9680.4782.2946.4240.773
Superior Perifoveal25268.841.4424.0390.5362.1405.9910.796
Nasal Perifoveal25286.140.8182.2890.2861.6814.7070.587
Inferior Perifoveal25257.541.0592.9650.4111.3563.7960.526
Temporal Perifoveal25253.970.9482.6550.3731.5194.2520.598
Maestro 6x6 3D Macula
Central Fovea25231.181.5814.4280.6842.3706.6371.025
Superior Parafoveal25310.611.6494.6190.5312.6167.3240.842
Nasal Parafoveal25309.711.3283.7170.4292.4686.9110.797
Inferior Parafoveal25307.781.2443.4840.4042.1125.9130.686
Temporal Parafoveal25295.581.2173.4090.4122.0665.7850.699
Superior Perifoveal25268.731.4684.1110.5462.0175.6480.751

Precision Assessment:

Table 1: Full Retinal Thicknoss - Normal Eves Popoatability and Roproducibili

\DC - 026811/000013 - 8588672 v1

{7}------------------------------------------------

RepeatabilityReproducibility
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Nasal Perifoveal25285.910.8732.4450.3051.9125.3540.669
Inferior Perifoveal25258.271.2853.5990.4982.1436.0020.830
Temporal Perifoveal25251.831.0883.0450.4321.9635.4960.779

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{8}------------------------------------------------

Table 2: Full Retinal Thickness - Retinal Disease Eyes Repeatability and Reproducibility

RepeatabilityReproducibility
Device
MeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Central Fovea26250.153.69710.3511.4784.74513.2851.897
Superior Parafoveal26317.621.5614.3710.4922.3716.6390.747
Nasal Parafoveal26315.721.5124.2330.4792.7377.6650.867
Inferior Parafoveal26313.432.0605.7690.6572.9838.3510.952
Temporal Parafoveal26303.672.0425.7180.6723.1648.8601.042
Superior Perifoveal26276.821.3563.7970.4901.7584.9220.635
Nasal Perifoveal26288.871.0933.0590.3781.7194.8120.595
Inferior Perifoveal26263.371.8005.0390.6832.3196.4950.881
Temporal Perifoveal26261.821.9505.4590.7452.2796.3820.871
Maestro 6x6 3D Macula
Central Fovea26247.022.6937.5411.0903.3019.2421.336
Superior Parafoveal26317.541.7544.9100.5522.7147.5990.855
Nasal Parafoveal26315.981.5154.2410.4792.5757.2100.815
Inferior Parafoveal26312.701.5554.3540.4972.4886.9660.796
Temporal Parafoveal26301.801.4844.1560.4922.7047.5700.896
Superior Perifoveal26276.601.5314.2860.5532.3656.6210.855
Nasal Perifoveal26288.211.0833.0330.3761.9025.3260.660
Inferior Perifoveal26263.981.6974.7510.6432.1776.0950.825
Temporal Perifoveal26258.611.1363.1820.4391.9945.5830.771

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{9}------------------------------------------------

Table 3: Full Retinal Thickness - Glaucoma Eyes Repeatability and Reproducibility
Table 3: Full Retinal Thickness - Glaucoma Eyes Repeatability and Reproducibility
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Central Fovea25243.092.9148.1591.1993.98511.1581.639
Superior Parafoveal25304.591.9215.3800.6312.7277.6350.895
Nasal Parafoveal25307.971.5204.2550.4932.6767.4920.869
Inferior Parafoveal25299.652.0125.6330.6712.7237.6240.909
Temporal Parafoveal25293.571.5874.4440.5412.6877.5240.915
Superior Perifoveal25261.201.7314.8460.6632.3086.4630.884
Nasal Perifoveal25276.951.3933.9010.5031.8315.1270.661
Inferior Perifoveal25246.041.7264.8320.7012.2256.2300.904
Temporal Perifoveal25247.981.6854.7180.6792.2296.2410.899
Maestro 6x6 3D Macula
Central Fovea25241.051.6774.6950.6962.9868.3611.239
Superior Parafoveal25305.201.4624.0950.4792.5557.1540.837
Nasal Parafoveal25307.461.2453.4870.4052.7917.8150.908
Inferior Parafoveal25299.201.2703.5560.4242.6447.4040.884
Temporal Parafoveal25292.571.2223.4210.4182.6367.3800.901
Superior Perifoveal25261.201.3793.8600.5282.1425.9980.820
Nasal Perifoveal25276.370.9192.5730.3321.9875.5630.719
Inferior Perifoveal25246.181.4924.1770.6062.2986.4330.933
Temporal Perifoveal25245.711.0302.8840.4192.3666.6240.963

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance Repeatability limit = 2.8 x Repeatability SD

Repeatability limit = 2.8 x Repeatability SD
Reproducibility limit = 2.8 x Reproducibility SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{10}------------------------------------------------

Table 4: Ganglion Cell + IPL - Normal Eves Repeatability and Reproducibility

RepeatabilityReproducibility
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D WideSuperior2570.370.6411.7950.9110.7152.0011.015
Superior Nasal2574.660.6711.8780.8980.7071.9800.947
Superior Temporal2571.460.5901.6510.8250.8312.3261.162
Inferior2568.110.6341.7760.9310.6631.8560.973
Inferior Nasal2573.590.6991.9570.9500.7682.1501.043
Inferior Temporal2573.120.5531.5500.7570.7081.9830.968
Average2571.890.2910.8140.4040.3651.0230.508
Maestro 6x6 3D MaculaSuperior2570.480.7362.0601.0440.8092.2661.148
Superior Nasal2574.620.3941.1020.5280.5481.5350.735
Superior Temporal2571.930.6111.7120.8500.7121.9930.989
Inferior2568.330.4531.2680.6630.6571.8410.962
Inferior Nasal2573.580.4361.2200.5920.5951.6670.809
Inferior Temporal2573.460.5621.5730.7650.6111.7120.832
Average2572.080.2630.7350.3640.4011.1240.557

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{11}------------------------------------------------

Table 5: Ganglion Cell + IPL - Retinal Disease Eyes Repeatability and Reproducibility
RepeatabilityReproducibility
Device
MeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Superior2669.571.2343.4561.7741.7794.9822.558
Superior Nasal2673.780.7682.1501.0410.8122.2741.101
Superior Temporal2670.050.9632.6961.3751.7344.8542.475
Inferior2665.841.7604.9282.6731.8265.1132.774
Inferior Nasal2670.861.8345.1362.5891.8445.1642.603
Inferior Temporal2670.391.4273.9952.0272.5377.1033.604
Average2670.080.7622.1341.0880.8192.2941.169
Maestro 6x6 3D Macula
Superior2668.870.6871.9230.9970.6951.9451.008
Superior Nasal2674.040.8412.3551.1360.9262.5931.251
Superior Temporal2670.491.0232.8661.4521.2863.6011.824
Inferior2666.770.8102.2681.2130.9402.6331.408
Inferior Nasal2671.070.4911.3740.6900.6991.9580.984
Inferior Temporal2671.370.9932.7821.3921.0783.0171.510

0.375

Mean = Intercept of the ANOVA model

Average

Repeatability SD = Square root of the residual variance

26

70.45

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

1.049

0.532

0.465

1.302

0.660

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{12}------------------------------------------------

Table 6: Ganglion Cell + IPL - Glaucoma Eves Repeatability and Reproducibility

Table 6: Ganglion Cell + IPL - Glaucoma Eyes Repeatability and Reproducibility
DeviceMeasurementNMeanRepeatabilityReproducibility
SDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Superior2563.550.8842.4761.3921.0112.8321.591
Superior Nasal2568.770.9962.7901.4491.1693.2741.700
Superior Temporal2563.930.8832.4721.3810.9082.5421.420
Inferior2559.610.9322.6081.5631.0632.9771.784
Inferior Nasal2566.070.8682.4321.3150.9802.7431.483
Inferior Temporal2563.340.9022.5261.4240.9352.6181.476
Average2564.220.4031.1290.6280.4601.2880.716
Maestro 6x6 3D Macula
Superior2564.270.5921.6580.9210.7081.9831.102
Superior Nasal2569.540.4801.3450.6910.5591.5660.804
Superior Temporal2566.210.7662.1451.1570.9242.5861.395
Inferior2560.860.7842.1941.2880.8832.4721.451
Inferior Nasal2566.110.5031.4070.7600.6891.9281.042
Inferior Temporal2566.000.7652.1421.1590.8912.4941.350
Average2565.530.3881.0870.5930.4831.3510.736

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{13}------------------------------------------------

Table 7: Ganglion Cell Complex Thickness - Normal Eyes Repeatability and Reproducibility

Table 7: Ganglion Cell Complex Thickness - Normal Eyes Repeatability and Reproducibility
DeviceMeasurementNMeanRepeatabilityReproducibility
SDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Superior25106.360.7312.0460.6870.8722.4410.819
Superior Nasal25118.010.9692.7130.8211.0702.9950.906
Superior Temporal2594.110.5951.6650.6320.7552.1150.803
Inferior25106.270.7462.0870.7021.0072.8200.948
Inferior Nasal25120.430.8872.4830.7361.2723.5621.056
Inferior Temporal2597.480.6401.7910.6560.8212.2990.842
Average25107.110.5041.4110.4700.6911.9360.645
Superior Average25106.160.6081.7020.5730.7292.0420.687
Inferior Average25108.060.5791.6210.5360.8452.3670.782
Maestro 6x6 3D Macula
Superior25106.951.4974.1921.4001.7194.8131.607
Superior Nasal25118.020.6341.7760.5370.9422.6370.798
Superior Temporal2593.780.7111.9920.7590.9242.5870.985
Inferior25107.290.6811.9080.6351.1373.1841.060
Inferior Nasal25121.270.6871.9250.5670.9012.5220.743
Inferior Temporal2597.090.6821.9090.7020.9792.7411.008
Average25107.430.5351.4970.4980.7832.1930.729
Superior Average25106.250.7071.9810.6660.8952.5070.843
Inferior Average25108.550.5411.5150.4980.8452.3650.778

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{14}------------------------------------------------

Table 8: Ganglion Cell Complex Thickness - Retinal Disease Eyes Repeatability and Reproducibility

Table 8: Ganglion Cell Complex Thickness - Retinal Disease Eyes Repeatability and Reproducibility
RepeatabilityReproducibility
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Superior26108.981.4624.0951.3422.2956.4252.106
Superior Nasal26120.000.9262.5930.7720.9852.7580.821
Superior Temporal2694.961.0242.8671.0781.2623.5341.329
Inferior26105.202.2366.2592.1252.2366.2592.125
Inferior Nasal26116.093.72910.4423.2133.75210.5063.232
Inferior Temporal2697.862.2316.2482.2802.4206.7752.473
Average26107.201.1923.3371.1121.1923.3371.112
Superior Average26107.980.7562.1180.7001.1573.2401.072
Inferior Average26106.392.2806.3832.1432.2886.4072.151
Maestro 6x6 3D Macula
Superior26107.960.8992.5180.8331.0813.0271.001
Superior Nasal26120.100.7932.2210.6600.9522.6650.792
Superior Temporal2693.870.7412.0760.7900.9722.7211.035
Inferior26106.131.1603.2481.0931.1603.2481.093
Inferior Nasal26117.240.7212.0190.6150.8622.4140.736
Inferior Temporal2695.850.7802.1840.8140.9012.5240.940
Average26106.880.5181.4520.4850.6431.7990.601
Superior Average26107.310.5991.6770.5580.8022.2460.747
Inferior Average26106.410.6121.7120.5750.6861.9200.644

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{15}------------------------------------------------

Table 9: Ganglion Cell Complex Thickness - Glaucoma Eyes Repeatability and Reproducibility

RepeatabilityReproducibility
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Superior2595.851.0532.9471.0981.1883.3271.240
Superior Nasal25108.780.8502.3790.7810.8972.5130.825
Superior Temporal2585.281.0142.8401.1891.0332.8931.211
Inferior2591.591.0372.9031.1321.0712.9981.169
Inferior Nasal25105.430.7772.1750.7370.9742.7280.924
Inferior Temporal2585.080.7562.1160.8880.8182.2890.961
Average2595.340.6081.7030.6380.6551.8330.687
Superior Average2596.640.8052.2550.8330.8572.4000.887
Inferior Average2594.030.6511.8230.6920.7242.0280.770
Maestro 6x6 3D Macula
Superior2596.850.7001.9600.7230.9082.5420.937
Superior Nasal25108.810.6911.9350.6350.8922.4980.820
Superior Temporal2584.930.5971.6710.7030.8092.2640.952
Inferior2592.151.0422.9171.1311.1663.2641.265
Inferior Nasal25105.990.7061.9770.6660.9142.5580.862
Inferior Temporal2584.510.7422.0790.8780.9392.6291.111
Average2595.590.4861.3600.5080.6481.8140.678
Superior Average2596.860.5091.4250.5260.6761.8930.698
Inferior Average2594.220.6421.7980.6820.7822.1890.830
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Average25105.301.3883.8871.3181.5984.4741.517
Superior25127.893.75410.5122.9363.89310.9013.044
Nasal2579.372.7657.7433.4843.0448.5223.835
Inferior25135.233.3759.4492.4954.09411.4643.028
Temporal2578.681.3563.7961.7232.0875.8432.652
12-hour - T2565.261.6064.4962.4611.9845.5543.040
12-hour - TS2593.362.6367.3812.8243.0248.4673.239
12-hour - ST25138.214.53312.6913.2804.93313.8123.569
12-hour - S25125.155.96816.7114.7696.35517.7955.078
12-hour - SN25120.315.28214.7894.3905.53415.4964.600
12-hour - NS2598.295.16914.4745.2595.48315.3535.579
12-hour - N2562.752.4906.9723.9683.5149.8405.600
12-hour - NI2577.114.04211.3185.2424.07211.4025.281
12-hour - IN25113.675.01314.0354.4105.88016.4635.173
12-hour - I25145.755.96416.7004.0926.62118.5384.543
12-hour - IT25146.364.05711.3592.7724.90413.7323.351
12-hour - TI2577.412.2956.4262.9653.4909.7714.508
Maestro 6x6 3D Disc
Average25104.320.9732.7240.9331.1473.2111.099
Superior25128.412.6277.3562.0462.9178.1672.272
Nasal2578.142.0075.6192.5692.3026.4452.946
Inferior25135.512.0035.6101.4782.5037.0081.847
Temporal2575.221.3353.7381.7751.8595.2062.471
12-hour - T2561.651.3963.9082.2641.8165.0862.946
12-hour - TS2590.012.1616.0502.4012.7927.8173.101
12-hour - ST25139.483.2719.1572.3454.34412.1633.114
12-hour - S25125.524.96413.8983.9545.14514.4074.099
12-hour - SN25120.204.16011.6473.4614.29412.0243.573
12-hour - NS2597.393.62210.1433.7204.29512.0254.410
12-hour - N2561.111.5264.2732.4971.7594.9262.879
12-hour - NI2575.872.9958.3853.9473.4239.5854.511
12-hour - IN25113.573.2989.2342.9043.59010.0523.161
12-hour - I25147.833.63310.1722.4574.38812.2872.968
12-hour - IT25145.023.98011.1442.7454.36612.2243.010
12-hour - TI2573.982.2126.1942.9903.2599.1254.405

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{16}------------------------------------------------

Table 10: Retinal Nerve Fiber Layer - Normal Eyes Repeatability and Reproducibility

{17}------------------------------------------------

RepeatabilityReproducibility
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
T=Temporal; S=Superior; N=Nasal; I=InferiorAll statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interactionbetween operator/device and eye.
N = Number of subjects
Mean = Intercept of the ANOVA model
Repeatability SD = Square root of the residual variance
Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual variance
Repeatability limit = 2.8 x Repeatability SD
Reproducibility limit = 2.8 x Reproducibility SD
Repeatability CV% = (Repeatability SD)/Intercept x 100%
Reproducibility CV% = (Reproducibility SD)/Intercept x 100%

{18}------------------------------------------------

Table 11: Retinal Nerve Fiber Layer - Retinal Disease Eyes Repeatability and Reproducibility

DeviceMeasurementNMeanSDRepeatabilityReproducibility
SDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
Average26101.461.6184.5291.5941.9165.3641.888
Superior26123.564.09911.4763.3174.51712.6483.656
Nasal2676.763.3469.3704.3593.94811.0555.143
Inferior26128.813.3439.3602.5955.42715.1944.213
Temporal2676.731.7084.7832.2262.1666.0642.823
12-hour - T2667.021.9425.4382.8982.5107.0283.745
12-hour - TS2691.792.6917.5352.9323.1438.7993.424
12-hour - ST26133.094.60812.9013.4625.51515.4434.144
12-hour - S26122.776.63318.5725.4037.19120.1355.857
12-hour - SN26114.846.72018.8165.8526.72018.8165.852
12-hour - NS2691.746.55918.3657.1506.83919.1497.455
12-hour - N2664.723.1338.7724.8403.5129.8335.426
12-hour - NI2673.866.01416.8398.1436.40717.9398.675
12-hour - IN26111.635.62515.7495.0399.00425.2108.066
12-hour - I26146.305.48315.3533.7486.99419.5834.781
12-hour - IT26128.525.21214.5934.0555.93216.6104.616
12-hour - TI2671.392.5497.1363.5702.7917.8153.910
Maestro 6x6 3D Disc
Average26101.721.1023.0861.0841.5054.2141.480
Superior26125.252.5127.0342.0062.6097.3052.083
Nasal2676.242.1756.0902.8533.5239.8634.620
Inferior26131.152.2936.4191.7482.8307.9242.158
Temporal2674.251.2873.6031.7331.4464.0481.947
12-hour - T2664.471.6004.4802.4821.9165.3652.972
12-hour - TS2689.312.2186.2102.4832.6447.4042.961
12-hour - ST26134.934.22011.8153.1275.01514.0423.717
12-hour - S26123.865.35014.9804.3205.86516.4234.736
12-hour - SN26116.903.69410.3433.1604.23411.8543.622
12-hour - NS2691.834.17711.6944.5486.02616.8736.562
12-hour - N2663.163.2379.0655.1253.58310.0315.672
12-hour - NI2673.704.21911.8145.7255.44415.2427.387
12-hour - IN26114.304.23911.8693.7095.46915.3144.785
12-hour - I26150.514.35912.2062.8964.87413.6483.239
12-hour - IT26128.613.77010.5562.9324.83413.5353.759
12-hour - TI2668.952.0205.6562.9302.4736.9253.587

{19}------------------------------------------------

RepeatabilityReproducibility
DeviceMeasurementNMeanSDSDLimitCV%SDLimitCV%
T=Temporal; S=Superior; N=Nasal; I=InferiorAll statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interactionbetween operator/device and eye.
N = Number of subjectsMean = Intercept of the ANOVA modelRepeatability SD = Square root of the residual varianceReproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual varianceRepeatability limit = 2.8 x Repeatability SDReproducibility limit = 2.8 x Reproducibility SDRepeatability CV% = (Repeatability SD)/Intercept x 100%Reproducibility CV% = (Reproducibility SD)/Intercept x 100%
Table 12: Retinal Nerve Fiber Layer - Glaucoma Eyes Repeatability and Reproducibility
DeviceMeasurementNMeanSDRepeatabilityReproducibility
LimitCV%SDLimit (Ratio)CV%
Maestro 12x9 3D Wide
Average2583.551.6464.6091.9701.7524.9062.097
Superior2597.323.81410.6793.9193.94811.0554.057
Nasal2570.633.80510.6535.3874.03611.3015.715
Inferior25102.903.2719.1603.1793.4029.5263.306
Temporal2563.291.3703.8372.1651.9415.4343.066
12-hour - T2555.911.7794.9803.1812.5347.0944.532
12-hour - TS2573.532.4716.9183.3602.8828.0713.920
12-hour - ST25101.844.69113.1344.6064.73113.2464.645
12-hour - S2599.856.17617.2946.1857.64821.4157.659
12-hour - SN2590.185.72516.0316.3495.82816.3186.463
12-hour - NS2580.966.68818.7278.2616.71918.8138.299
12-hour - N2558.294.13811.5867.0994.43612.4227.611
12-hour - NI2572.634.57812.8196.3044.90313.7276.750
12-hour - IN2594.444.45312.4684.7155.45715.2805.778
12-hour - I25117.006.21417.4005.3116.21417.4005.311
12-hour - IT2597.274.22211.8224.3414.68413.1144.815
12-hour - TI2560.462.2436.2803.7092.5307.0834.184
Maestro 6x6 3D Disc
Average2583.701.6154.5211.9291.6184.5301.933
Superior25100.603.58410.0353.5633.73210.4503.710
Nasal2570.122.2646.3383.2282.3646.6193.371
Inferior25103.113.4359.6173.3313.61910.1323.510
Temporal2561.032.1045.8923.4482.1265.9543.484
12-hour - T2553.802.6427.3974.9112.7407.6715.092
12-hour - TS2572.043.96111.0905.4983.96111.0905.498
12-hour - ST25103.274.06711.3863.9385.55315.5485.377
12-hour - S25103.836.08517.0395.8617.04619.7306.786
12-hour - SN2594.795.67815.8985.9905.72716.0366.042
12-hour - NS2582.984.30712.0615.1914.46912.5125.386
12-hour - N2555.652.6707.4774.7982.7347.6564.913
12-hour - NI2571.733.96411.0995.5274.03511.2975.625
12-hour - IN2594.946.15217.2266.4806.71618.8057.074
12-hour - I25119.035.62015.7354.7215.74016.0724.822
12-hour - IT2595.374.60712.8994.8314.69113.1354.919
12-hour - TI2557.262.2786.3793.9782.5727.2004.491

{20}------------------------------------------------

Table 12: Retinal Nerve Fiber Layer - Glaucoma Eyes Repeatability and Reproducibility

{21}------------------------------------------------

RepeatabilityReproducibility
Device
MeasurementNMeanSDLimitCV%SDLimit (Ratio)CV%
T=Temporal; S=Superior; N=Nasal; I=Inferior
All statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interaction
between operator/device and eye.
N = Number of subjects
Mean = Intercept of the ANOVA model
Repeatability SD = Square root of the residual variance
Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual variance
Repeatability limit = 2.8 x Repeatability SD
Reproducibility limit = 2.8 x Reproducibility SD
Repeatability CV% = (Repeatability SD)/Intercept x 100%
Reproducibility CV% = (Reproducibility SD)/Intercept x 100%

{22}------------------------------------------------

Table 13: Optic Disc - Normal Eyes Repeatability and Reproducibility
Table 13: Optic Disc - Normal Eyes Repeatability and Reproducibility
RepeatabilityReproducibility
Device
MeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
C/D Vertical250.460.0300.0856.6000.0310.0886.863
C/D Area250.250.0130.0375.1720.0150.0436.023
Disc Area252.100.0740.2073.5200.0890.2494.233
Cup Area250.570.0220.0623.8900.0250.0714.474
Rim Area251.540.0720.2034.7110.0870.2435.640
Cup Volume250.100.0060.0165.3220.0070.0207.014
Rim Volume250.280.0180.0516.5710.0220.0627.967
Maestro 6x6 3D Disc
C/D Vertical250.460.0210.0594.5490.0230.0665.089
C/D Area250.260.0130.0364.9370.0140.0405.406
Disc Area251.920.0640.1783.3130.0780.2194.074
Cup Area250.540.0200.0553.6660.0250.0714.730
Rim Area251.390.0650.1834.7080.0750.2115.443
Cup Volume250.100.0050.0144.7470.0070.0206.759
Rim Volume250.240.0150.0436.3590.0200.0558.139

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{23}------------------------------------------------

Table 14: Optic Disc - Retinal Disease Eves Repeatability and Reproducibility

Table 14: Optic Disc - Retinal Disease Eyes Repeatability and ReproducibilityReproducibility
Repeatability
DeviceMeasurementNMeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D WideC/D Vertical260.470.0250.0715.3770.0300.0856.430
C/D Area260.260.0190.0547.3870.0200.0577.789
Disc Area262.250.0870.2433.8580.1120.3144.981
Cup Area260.610.0410.1156.7280.0430.1217.080
Rim Area261.640.0950.2655.7620.1160.3267.088
Cup Volume260.110.0090.0258.4040.0220.06120.586
Rim Volume260.280.0200.0567.1860.0270.0769.775
Maestro 6x6 3D DiscC/D Vertical260.460.0190.053 (0.439)4.1170.0210.058 (0.472)4.519
C/D Area260.250.0090.024 (0.469)3.4000.0100.029 (0.533)4.079
Disc Area262.070.0650.183 (0.853)3.1630.0790.220 (0.754)3.803
Cup Area260.550.0160.044 (0.441)2.8550.0190.053 (0.513)3.438
Rim Area261.520.0620.172 (0.818)4.0570.0730.205 (0.732)4.817
Cup Volume260.100.0050.014 (0.192)4.8270.0120.032 (0.420)11.024
Rim Volume260.260.0150.041 (0.926)5.6270.0170.046 (0.949)6.419

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{24}------------------------------------------------

Table 15: Optic Disc - Glaucoma Eyes Repeatability and Reproducibility
Table 15: Optic Disc - Glaucoma Eyes Repeatability and Reproducibility
DeviceMeasurementNRepeatabilityReproducibility
MeanSDLimitCV%SDLimitCV%
Maestro 12x9 3D Wide
C/D Vertical250.770.0310.0864.0300.0320.0904.185
C/D Area250.590.0200.0573.4690.0240.0684.130
Disc Area252.350.0750.2093.1790.0890.2513.811
Cup Area251.400.0480.1343.4300.0560.1574.017
Rim Area250.950.0640.1796.7220.0800.2258.449
Cup Volume250.410.0580.16414.2740.0700.19617.103
Rim Volume250.120.0110.0308.9530.0140.03811.152
Maestro 6x6 3D Disc
C/D Vertical250.760.0200.0552.5750.0210.0582.743
C/D Area250.590.0110.0311.8520.0120.0321.959
Disc Area252.150.0650.1823.0130.0710.1983.285
Cup Area251.280.0450.1253.4850.0480.1343.755
Rim Area250.870.0380.1074.3790.0410.1154.701
Cup Volume250.410.0240.0675.8130.0240.0685.859
Rim Volume250.110.0070.0206.6820.0080.0227.201

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{25}------------------------------------------------

Reference Database:

The reference database study included the assessment of 399 subject eyes from normal study subjects across a predefined range of ages and ethnicities. The study collected measurements of normal eyes for full retinal thickness, retinal nerve fiber layer thickness, ganglion cell layer thickness, ganglion cell complex and the optic disc in addition to small and large super pixel grid and TSNIT circle profile measurements. The reference limits at the 1°, 5°, 95° and 99°' percentile points were determined. The database age range is 18 – 88 years.

Total(N=399)
Age (years)
n399
Mean (SD)46.3 (16.3)
Median47.0
Min, Max18, 88
Age group, n (%)
18-30 years89 (22)
31-40 years76 (19)
41-50 years66 (17)
51-60 years78 (20)
61-70 years55 (14)
70+ years35 (9)
Gender, n (%)
Male173 (43)
Female226 (57)
Ethnicity, n (%)
Hispanic or Latino71 (18)
Not Hispanic and not Latino328 (82)
Race, n (%)
White234 (59)
Black/African American81 (20)
American Indian/AlaskanNative4 (1)
Asian51 (13)
Native American/PacificIslander7 (2)
Other22 (6)

Table 16: Reference Database Demographic Data

{26}------------------------------------------------

Substantial Equivalence Comparison Table

Subject Device:Primary Predicate:Secondary Predicate Device:
Topcon 3D OCT-1 MaestroOptovue iVue (K121739)Topcon TRC-NW300 (K123460)
Indications forUseThe Topcon 3D OCT-1 Maestro is anon-contact, high resolutiontomographic and biomicroscopicimaging device that incorporates adigital camera for photographing,displaying and storing the data ofthe retina and surrounding parts ofthe eye to be examined underMydriatic and non-Mydriaticconditions.The 3D OCT-1 Maestro is indicatedfor in vivo viewing, axial crosssectional, and three-dimensionalimaging and measurement ofposterior ocular structures, includingretina, retinal nerve fiber layer,macula and optic disc as well asimaging of anterior ocular structures.It also includes a ReferenceDatabase for posterior ocularmeasurements which provide for thequantitative comparison of retinalnerve fiber layer, optic nerve head,and the macula in the human retinato a database of known normalsubjects. The 3D OCT-1 Maestro isindicated for use as a diagnosticdevice to aid in the diagnosis,documentation and management ofocular health and diseases in theadult population.The iVue with Normative Databaseis an optical coherence tomographysystem intended for in vivo imaging,axial cross-sectional, three-dimensional imaging andmeasurement of anterior andposterior ocular structures. TheiVue is a non-contact, highresolution tomographic imagingdevice. It is intended for in vivoimaging, axial cross-sectional, andthree-dimensional imaging andmeasurement of anterior andposterior ocular structures, includingretina, retinal nerve fiber layer,ganglion cell complex (GCC), opticdisc, cornea, and anterior chamberof the eye. The iVue with NormativeDatabase is a quantitative tool forthe comparison of retina, retinalnerve fiber layer, ganglion cellcomplex, and optic discmeasurements to a database ofknown normal subjects. The iVuewith Normative Database isindicated for use as a device to aidin the diagnosis, documentation, andmanagement of ocular health anddiseases in the adult population.The TRC-NW300 intended foruse in capturing images of theretina and the anterior segmentof the eye and presenting thedata to the eye care professional,without the use of a mydriatic.
Product CodeOBO, HKIHLIHKI
TechnologicalCharacteristicsOCT:Optical coherence tomography isemployed to obtain cross-sectionalimages of the fundus and theanterior segment of the eye.Fundus Camera:A Xenon flash and a CMOS cameraare employed to obtainfundus/anterior Color/digital Red-free images.OCT:Optical coherence tomography isemployed to obtain cross-sectionalimages of the fundus and theanterior segment of the eye.Fundus Camera:A Xenon flash and a CCDcamera are employed to obtainfundus/anterior Color images.
OCT
Light SourceSpectral domain OCT (SD-OCT)using a superluminescent diode(SLD) with center wavelength 840nm.Spectral domain OCT (SD-OCT)using a superluminescent diode(SLD) with center wavelength 840nm.N/A
Subject Device:Topcon 3D OCT-1 MaestroPrimary Predicate:Optovue iVue (K121739)Secondary Predicate Device:Topcon TRC-NW300 (K123460)
AnalysisFor posterior:- Retinal layer segmentation- Thickness calculation- Optic disc analysisFor anterior:NAFor posterior:- Retinal layer segmentation- Thickness calculation- Optic disc analysisFor anterior:- Pachymetry measurement- Pachymetry map- Angle measurementN/A
Fundus Camera
ObservationLight source- IR LEDCamera- CMOS cameraN/ALight source- Halogen lampCamera- CCD camera

{27}------------------------------------------------

Conclusion

As described in this 510(k) Summary, the performance testing deemed necessary was conducted and the resultant data support the substantial equivalence of the 3D OCT-1 Maestro to the primary predicate device Optovue's iVue with Normative Database (iVue) (K121739).

§ 886.1570 Ophthalmoscope.

(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.