K Number
K161506

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-09-26

(117 days)

Product Code
Regulation Number
870.5150
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Penumbra Aspiration Catheters and Separators: As part of the Penumbra Aspiration System, the Penumbra Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.
Penumbra Aspiration Tubing: As part of the Penumbra Aspiration System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Aspiration Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The Penumbra Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the coronary and peripheral vasculature. The Aspiration Catheter and Separator are available in multiple configurations. The devices are provided sterile, non-pyrogenic, and intended for single use only. Intended users for this device are physicians who have received appropriate training in interventional radiology. The Penumbra Aspiration System is designed to remove thrombus from the coronary and peripheral vasculature using continuous aspiration. The Aspiration Catheter targets aspiration from the pump directly to the thrombus. The Separator may be used to clear the lumen of the Aspiration Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using an Aspiration Catheter with an I.D. of 0.054 in or larger. The Aspiration Catheter is introduced through a guide catheter or long introducer sheath and into the coronary or peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The Aspiration Catheter is used with the Penumbra Pump MAX to aspirate thrombus from an occluded vessel. As needed, the Separator may be deployed from the Aspiration Catheter to assist with thrombus removal. The Separator is advanced and retracted through the Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Aspiration Catheter tip. For the aspiration source, the Aspiration Catheter is used in conjunction with the Penumbra Pump MAX, which is connected using the Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Separator is provided with an introducer and torque device. The Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and a peelable sheath. The Separator is provided with an introducer and torque device. The Aspiration Catheter and Separator are visible under fluoroscopy. The Penumbra Pump MAX is the aspiration source for the Penumbra Aspiration System. The Penumbra Pump MAX operates using AC power and is designed to be portable if needed. The Penumbra Pump MAX provides vacuum of up to 29 inHg. The pump is available in both 110Vac and 230Vac versions. The front face of the pump has a display panel with a vacuum gauge, vacuum regulator dial, and power switch. The pump connects to the canister reservoir with a tubing assembly (Penumbra Pump/Canister Tubing), which is provided as an accessory. The Penumbra Pump/Canister Tubing consists of a short tubing segment with an inline filter and connectors on each end to facilitate attachment to the pump's vacuum port. The tubing is provided pre-attached to the canister reservoir lid. The Penumbra Pump/Canister Tubing is provided non-sterile and is used outside the sterile field.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Penumbra Aspiration System. It details the device's indications for use, its classification, and a comparison to predicate and reference devices. However, the document primarily focuses on demonstrating substantial equivalence through non-clinical (benchtop) and animal study data, largely leveraging existing data from previously cleared similar devices.

Crucially, this document does NOT contain information about specific "acceptance criteria" or "device performance" in terms of clinical outcomes, nor does it describe a study involving human subjects or experts for establishing ground truth for AI/algorithm performance.

Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, along with the specified sub-points (table of acceptance criteria/performance, sample size, experts, adjudication, MRMC study, standalone performance, type of ground truth, training set details) because this information is not present in the provided text.

The document discusses:

  • Leveraged Non-Clinical Data (Section 1.10): This section lists benchtop tests (Friction, Visual & Dimensional, Pouch Seal Strength, Flow Rate, Tensile, Elongation, Bond Strength, Hub Air Aspiration, Torsion, Burst, Simulated Use, Particulate, Flexibility, Packaging, Corrosion). It states that "all established requirements and acceptance criteria were met," but it does not specify what those acceptance criteria were in measurable terms (e.g., "flow rate X L/min," "tensile strength Y N"). It also does not define "device performance" in the way requested (e.g., diagnostic accuracy, sensitivity, specificity).
  • Leveraged Animal Studies (Section 1.11): This section summarizes findings from design validation in GLP animal testing, noting "No vessel injury," "No abnormal gross or histology findings," and "no significant vascular response." This is performance data in an animal model, but not clinical performance in humans, nor does it relate to acceptance criteria for an AI/algorithm-driven device.

In summary, the provided text describes the regulatory clearance process for a medical device (an aspiration system) based on substantial equivalence to existing devices, supported by benchtop testing and animal studies. It does not contain the kind of information typically found in a study proving an AI/algorithm-driven device meets specific clinical acceptance criteria, which would involve human data, expert review, and metrics like accuracy, sensitivity, or specificity.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 26, 2016

Penumbra, Inc. Richard Kimura Regulatory Affairs Specialist 1 Penumbra Place Alameda, California 94502

Re: K161506

Trade/Device Name: Penumbra Aspiration System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: August 23, 2016 Received: August 24, 2016

Dear Mr. Kimura:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161506

Device Name Penumbra Aspiration System

Indications for Use (Describe)

Penumbra Aspiration Catheters and Separators

As part of the Penumbra Aspiration System, the Penumbra Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

Penumbra Aspiration Tubing

As part of the Penumbra Aspiration System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Yes
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Penumbra Aspiration System.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Richard Kimura Regulatory Affairs Specialist Phone: (510) 995-2034 FAX: (510) 217-6414 Email: rkimura@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

May 31, 2016

1.4 Device Trade or Proprietary Name

Penumbra Aspiration System

Device Classification 1.5

Regulatory Class: II Classification Panel: Cardiovascular Classification Name: Catheter, Embolectomy Regulation Number: 21 CFR §870.5150 Product Code: DXE

1.6 Predicate and Reference Devices

510(k) Number/Clearance DateName of DeviceName ofManufacturer
Predicate Device
K103405[22Dec2010]Pronto V4 Extraction CatheterVascular Solutions, Inc.
Reference Devices
K072718 [28Dec2007]Penumbra System [026, 032, 041]Penumbra, Inc.
K090752 [21Sep2009]Penumbra System [054]Penumbra, Inc.
K100769 [21May2010]Penumbra System Separator Flex [026, 032,Penumbra, Inc.

{4}------------------------------------------------

510(k) Number/Clearance DateName of DeviceName ofManufacturer
[041, 054]
K113163 [23Nov2011]Penumbra System MAXPenumbra, Inc.
K151623 [06Aug2015]Penumbra System 110 Aspiration TubingPenumbra, Inc.
K160449 [25May2016]Penumbra SystemPenumbra, Inc.
K121917 [19Sep2012]Penumbra Embolectomy Aspiration System(INDIGO Aspiration System)Penumbra, Inc.
K142870 [26May2015]Penumbra Embolectomy Aspiration System(INDIGO Aspiration System)Penumbra, Inc.
K160533 [24May2016]Penumbra Embolectomy Aspiration System(INDIGO Aspiration System)Penumbra, Inc.
K122756 [02Oct2012]Penumbra Pump MAXPenumbra, Inc.

{5}------------------------------------------------

K161506

Predicate and Reference Device Comparisor 1.7

Trade NamePredicate Device510(k) No.ClassificationIndication for UseReference DevicesSubject Device
Pronto V4 ExtractionCatheterK103405Class II, DXEThe Pronto Catheter isindicated for the removal offresh, soft emboli andthrombi from vessels in thecoronary and peripheralvasculature.
INDIGO Aspiration SystemK072718, K090752, K100769,K113163, K151623, K122756, andK160449K121917, K142870, and K160533Penumbra Aspiration System
SAMESAMETo be determined
Penumbra Reperfusion Cathetersand SeparatorsAs part of the Penumbra System, theReperfusion Catheters andSeparators are indicated for the usein the revascularization of patientswith acute ischemic strokesecondary to intracranial largevessel occlusive disease (within theinternal carotid, middle cerebral –M1 and M2 segments, basilar, andvertebral arteries) within 8 hours ofsymptom onset.INDIGO Aspiration Catheters andSeparatorsAs part of the PenumbraEmbolectomy Aspiration System(INDIGOTM Aspiration System), theINDIGO Aspiration Catheters andSeparators are indicated for theremoval of fresh, soft emboli, andthrombi from vessels of theperipheral arterial and venoussystems. Not for use in the coronariesor the neurovasculature.Penumbra Aspiration Cathetersand SeparatorsAs part of the PenumbraAspiration System, the PenumbraAspiration Catheters andSeparators are indicated for theremoval of fresh, soft emboli andthrombi from vessels in thecoronary and peripheralvasculature.
Penumbra Aspiration TubingAs part of the Penumbra System, thePenumbra Sterile Aspiration Tubingis indicated to connect the PenumbraReperfusion Catheters to thePenumbra Pump MAX.INDIGO Aspiration TubingAs part of the PenumbraEmbolectomy Aspiration System(INDIGOTM Aspiration System), theINDIGO Sterile Aspiration Tubing isindicated to connect the INDIGOAspiration Catheters to the PenumbraPump MAX.Penumbra Aspiration TubingAs part of the PenumbraAspiration System, the PenumbraSterile Aspiration Tubing isindicated to connect thePenumbra Aspiration Cathetersto the Penumbra AspirationPump.
Penumbra Pump MAXThe Penumbra Pump MAX isindicated as a vacuum source forPenumbra Aspiration Systems.Penumbra Pump MAXThe Penumbra Pump MAX isindicated as a vacuum source forPenumbra Aspiration Systems.Penumbra Aspiration PumpThe Penumbra Aspiration Pumpis indicated as a vacuum sourcefor Penumbra AspirationSystems.

{6}------------------------------------------------

રેણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દ
l
9
lX
Predicate DeviceReference DevicesSubject Device
Trade NamePronto V4 ExtractionCatheterPenumbra SystemPenumbra Aspiration System
510(k) No.K103405K072718, K090752, K100769,K113163, K151623, K122756, andK160449K121917, K142870, and K160533To be determined
ClassificationClass II, DXESAMESAMESAME
Aspiration Catheter
MaterialsBiocompatible, commonlyutilized for interventionaldevicesSAMESAMESAME
CoatingHydrophilicSAMESAMESAME
MarkerbandRadiopaqueSAMESAMESAME
GuidewireCompatibleYesSAMESAMESAME
DimensionsOD [Maximum]0.0645" - 0.0865"(1.64 – 2.20 mm)0.037" - 0.083"(0.94mm – 2.1mm)0.037" - 0.112"(0.94mm – 2.84mm)SAME AS THE INDIGOSYSTEM
Working LengthApproximately 138 cm118m – 153cm30cm – 153cmSAME AS THE INDIGOSYSTEM
Separator
MaterialsN/ABiocompatible, commonly utilizedfor interventional devicesBiocompatible, commonly utilizedfor interventional devicesBiocompatible, commonlyutilized for interventional devices
Dimensions
Distal ODN/A0.022" – 0.045"(0.56mm – 1.1mm)0.022" – 0.068"(0.56mm – 1.7mm)SAME AS THE INDIGOSYSTEM
Working LengthN/A175cm – 200cm150cm – 200cmSAME AS THE INDIGOSYSTEM
Aspiration Tubing
MaterialsBiocompatible, commonlyutilized for interventionaldevicesSAMESAMESAME
Dimensions

{7}------------------------------------------------

રેણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દ
l
9
l
X
Predicate DeviceReference DevicesSubject Device
Trade NamePronto V4 ExtractionCatheterPenumbra SystemPenumbra Aspiration System
510(k) No.K103405K072718, K090752, K100769,K113163, K151623, K122756, andK160449K121917, K142870, and K160533
ClassificationClass II, DXESAMETo be determined
IDunknown0.071" – 0.110 "(1.8mm – 2.79mm)0.088" and 0.110 "(2.2mm and 2.79mm)
Lengthunknown112.0" (284.5cm)112.0" (284.5cm)
SterilizationEOSAMESAME
Shelf-Life36 monthsSAMESAME
Aspiration SourceAspiration SyringeAspiration PumpAspiration Pump
EC 60601-1ComplianceN/AYesYes
EC 60601-1-2ComplianceN/AYesYes
VoltageN/A100-115 Vac/230 Vac100-115 Vac/230 Vac
FrequencyN/A50 Hz/60 Hz50 Hz/60 Hz
SterilizationunknownNon sterileNon sterile
Shelf LifeunknownN/AN/A

{8}------------------------------------------------

Device Description 1.8

The Penumbra Aspiration System

The Penumbra Aspiration System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the coronary and peripheral vasculature. The Aspiration Catheter and Separator are available in multiple configurations. The devices are provided sterile, non-pyrogenic, and intended for single use only. Intended users for this device are physicians who have received appropriate training in interventional radiology.

The Penumbra Aspiration System is designed to remove thrombus from the coronary and peripheral vasculature using continuous aspiration. The Aspiration Catheter targets aspiration from the pump directly to the thrombus. The Separator may be used to clear the lumen of the Aspiration Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using an Aspiration Catheter with an I.D. of 0.054 in or larger. The Aspiration Catheter is introduced through a guide catheter or long introducer sheath and into the coronary or peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The Aspiration Catheter is used with the Penumbra Pump MAX to aspirate thrombus from an occluded vessel. As needed, the Separator may be deployed from the Aspiration Catheter to assist with thrombus removal. The Separator is advanced and retracted through the Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Aspiration Catheter tip. For the aspiration source, the Aspiration Catheter is used in conjunction with the Penumbra Pump MAX, which is connected using the Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Separator is provided with an introducer and torque device. The Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and a peelable sheath. The Separator is provided with an introducer and torque device. The Aspiration Catheter and Separator are visible under fluoroscopy.

The Penumbra Pump MAX

The Penumbra Pump MAX is the aspiration source for the Penumbra Aspiration System. The Penumbra Pump MAX operates using AC power and is designed to be portable if needed. The Penumbra Pump MAX provides vacuum of up to 29 inHg. The pump is available in both 110Vac and 230Vac versions.

{9}------------------------------------------------

The front face of the pump has a display panel with a vacuum gauge, vacuum regulator dial, and power switch. The pump connects to the canister reservoir with a tubing assembly (Penumbra Pump/Canister Tubing), which is provided as an accessory. The Penumbra Pump/Canister Tubing consists of a short tubing segment with an inline filter and connectors on each end to facilitate attachment to the pump's vacuum port. The tubing is provided pre-attached to the canister reservoir lid. The Penumbra Pump/Canister Tubing is provided non-sterile and is used outside the sterile field.

1.9 Indications for Use

Penumbra Aspiration Catheters and Separators

As part of the Penumbra Aspiration System, the Penumbra Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the coronary and peripheral vasculature.

Penumbra Aspiration Tubing

As part of the Penumbra Aspiration System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.10 Leveraged Non-Clinical Data

The subject Penumbra Aspiration System devices are identical to the reference Penumbra System and INDIGO System devices. All bench top testing conducted on the reference Penumbra System and INDIGO System devices to evaluate the physical and mechanical properties of the system are therefore applicable to the subject Penumbra Aspiration System. The following bench top tests were performed and all established requirements and acceptance criteria were met:

  • Friction Test . Visual & Dimensional ●
  • Pouch Seal Strength .
  • Flow Rate Test .
  • Tensile Test Elongation Test o .

{10}------------------------------------------------

  • Bond Strength ●
  • . Hub Air Aspiration . Torsion Test
  • Burst Test Simulated Use Test ● .
  • . Particulate Test . Flexibility Test
  • . Packaging Test

In addition, the Penumbra Aspiration System devices are manufactured using identical materials, and utilize the same manufacturing, packaging, and sterilization methods as the reference Penumbra System and INDIGO System devices. Therefore, biocompatibility data, sterilization data, and stability data from previous pre-market notifications listed in Section 1.6 are directly applicable and no additional testing was required or was performed to support the Penumbra Aspiration System.

Corrosion Test

1.11 Leveraged Animal Studies

The subject Penumbra Aspiration System is identical to the reference Penumbra System and reference INDIGO System. Design Validation (GLP Animal Testing) was conducted on the reference Penumbra System and INDIGO System, the results are which are also applicable to the subject Penumbra Aspiration System. Therefore, Animal Testing data from previous pre-market notifications listed in Section 1.6 are directly applicable and no further Animal Testing was required or was performed to support the Penumbra Aspiration System.

The studies concluded that:

  • No vessel injury was noted on the final angiograms following the vessel response o procedure.
  • No abnormal gross or histology findings were noted in test vessel segments. .
  • The use of the devices resulted in no significant vascular response in these o experimental conditions.

{11}------------------------------------------------

1.12 Summary of Substantial Equivalence

The subject Penumbra Aspiration System is substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, shelflife, packaging and sterilization processes.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).