K Number
K161499

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-07-29

(58 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VertiForm Posterior Fixation System is intended for posterior, non-cervical fixation in skeletally mature patients for the following conditions: degenerative disc disease (DDD), DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, spondylolisthesis, trauma (fracture or dislocation), spinal stenosis, curvatures (scoliosis, lordosis), tumor, pseudarthrosis (previous failed fusion). When used as posterior non-cervical pedicle screw fixation in pediatric patients, the VertiForm Posterior Fixation System is intended to treat adolescent idiopathic scoliosis. The device is intended to be used with autograft to facilitate fusion.

Device Description

The VertiForm Posterior Fixation System is a non-cervical spinal fixation system consisting of pedicle screws, set screws, rods and cross connectors. The pedicle screws are offered in a Polyaxial. Fixed or Uniplanar pedicle screw configuration and are available in a variety of geometries and sizes to accommodate patient anatomy. They are offered non-sterile and sterile. The VertiForm Posterior Fixation System is manufactured from titanium alloy (ASTM F136), medical-grade commercially pure titanium (ASTM F67) and medical-grade cobalt chrome (ASTM F1537). This Special 510(k) Premarket Notification seeks clearance to add variable cross connectors to the currently marketed VertiForm Posterior Fixation System.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called the "VertiForm Posterior Fixation System." This type of document is for regulatory clearance and focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific performance criteria through a clinical study or detailed acceptance criteria for an AI/algorithm.

Therefore, the information required to fill out the table and answer the study-related questions (points 1-9) about acceptance criteria, study design, sample sizes, ground truth, experts, and MRMC studies is not present in this document.

Here's why and what information is available:

  • Device Type: This is a physical spinal implant system (pedicle screws, rods, cross connectors), not an AI/algorithm-based diagnostic or therapeutic device.
  • Regulatory Pathway: A 510(k) submission primarily relies on demonstrating that a new device is "substantially equivalent" to a legally marketed predicate device. This is often achieved through comparison of indications for use, technological characteristics, and sometimes non-clinical performance testing (e.g., mechanical tests for implants).
  • Focus of the Document: The document explicitly states that it seeks clearance to add "variable cross connectors" to an already marketed VertiForm Posterior Fixation System. The main comparison is between the modified device and the predicate device (the previous version of the VertiForm system, K141291).

What is present in the document is a summary of non-clinical (mechanical) testing performed to demonstrate equivalence, not AI performance metrics:

  • Non-Clinical Test Summary (Section 5):
    • Dynamic compression testing: Performed in accordance with ASTM F1717, "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model, 2004."
    • Bacterial endotoxin testing: Performed in accordance with ANSI/AAMI ST72.
    • Results: "Testing results demonstrated that the VertiForm Posterior Fixation System with variable cross connectors performs comparable with the predicate device."

In summary, the request pertains to information typically found in submissions for AI/algorithm-enabled medical devices, which is not applicable to this 510(k) for a spinal implant system.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 29, 2016

Next Orthosurgical Ms. Ellen Yarnall Director. Regulatory Affairs/Ouality Assurance 3270 Corporate View, Suite A Vista, California 92081

Re: K161499

Trade/Device Name: VertiForm Posterior Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MKB, MNI, MNH Dated: July 1, 2016 Received: July 5, 2016

Dear Ms. Yarnall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

VincentJ. Devlin -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161499

Device Name VertiForm Posterior Fixation System

Indications for Use (Describe)

The VertiForm Posterior Fixation System is intended for posterior, non-cervical fixation in skeletally mature patients for the following conditions: degenerative disc disease (DDD), DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, spondylolisthesis, trauma (fracture or dislocation), spinal stenosis, curvatures (scoliosis, lordosis), tumor, pseudarthrosis (previous failed fusion). When used as posterior non-cervical pedicle screw fixation in pediatric patients, the VertiForm Posterior Fixation System is intended to treat adolescent idiopathic scoliosis. The device is intended to be used with autograft to facilitate fusion.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

NEXT ORTHOSURGICAL

510(k) SUMMARY VertiForm Posterior Fixation System

1. GENERAL INFORMATION

Submitted by:Next Orthosurgical3270 Corporate ViewSuite AVista, CA 92081(760) 295-3600 (Tel)(760) 295-3610 (Fax)
Contact:Ellen Yarnall, Director of RA/QA
Proprietary Name:VertiForm Posterior Fixation System
Classification Name:Pedicle screw spinal system
Class:III
Product Code:NKB, OSH, MNI, MNH
CFR Section:21 CFR 888.3070
Device Panel:Orthopedic
Legally Marketed PredicateDevice:VertiForm Posterior Fixation System, K141291
Date Prepared:July 1, 2016

2. DEVICE DESCRIPTION

The VertiForm Posterior Fixation System is a non-cervical spinal fixation system consisting of pedicle screws, set screws, rods and cross connectors. The pedicle screws are offered in a Polyaxial. Fixed or Uniplanar pedicle screw configuration and are available in a variety of geometries and sizes to accommodate patient anatomy. They are offered non-sterile and sterile. The VertiForm Posterior Fixation System is manufactured from titanium alloy (ASTM F136), medical-grade commercially pure titanium (ASTM F67) and medical-grade cobalt chrome (ASTM F1537). This Special 510(k) Premarket Notification seeks clearance to add variable cross connectors to the currently marketed VertiForm Posterior Fixation System.

3. INDICATIONS FOR USE

The VertiForm Posterior Fixation System is intended for posterior, non-cervical fixation in skeletally mature patients for the following conditions: degenerative disc disease (DDD), DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, spondylolisthesis, trauma (fracture or dislocation), spinal stenosis, curvatures (scoliosis, lordosis), tumor,

{4}------------------------------------------------

pseudarthrosis (previous failed fusion). When used as posterior non-cervical pedicle screw fixation in pediatric patients, the VertiForm Fixation System is intended to treat adolescent idiopathic scoliosis. The device is intended to be used with autograft and/or allograft to facilitate fusion.

4. TECHNOLOGICAL CHARACTERISTICS SUMMARY

The VertiForm Posterior Fixation System with Variable Cross Connectors has the same indications for use, materials and technological characteristics as the predicate device. Both systems include rod-to-rod connectors used to add stability to posterior noncervical pedicle screw spinal systems. The subject device differs from the predicate in that the variable cross connectors allow angular and medial-lateral variability during implantation. Testing performed in accordance with guidance document; "Guidance for Industry and FDA Staff: Spinal Systems 510(k)s" resulted in a substantial equivalence determination.

5. NON-CLINCIAL TEST SUMMARY

Dynamic compression testing was performed in accordance with ASTM F1717, Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model, 2004. Bacterial endotoxin testing was performed in accordance with ANSI/AAMI ST72. Testing results demonstrated that the VertiForm Posterior Fixation System with variable cross connectors performs comparable with the predicate device.

6. CONCLUSIONS

Based upon evidence accompanying this Special 510(k), and the fact that the Indications for Use, fundamental scientific technology, materials and performance characteristics are the same, it has been determined that the VertiForm Posterior Fixation System is substantially equivalent to the previously cleared device.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.