K Number
K161492

Validate with FDA (Live)

Device Name
Juno VPAP ST-A
Manufacturer
Date Cleared
2017-01-19

(232 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).

The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).

The Juno VPAP ST-A is intended for home and hospital use.

The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.

Device Description

The subject device Juno VPAP ST-A retains all the same hardware, technologies and manufacturing characteristics as previously cleared in K153061. The device is used in a wider breathing circuit which typically comprises patient tubing and a mask (patient interface) to deliver a prescribed positive airway pressure treatment to patients.

Some of the kev features of the device include an in-line power supply: fully integrated humidifier: alarms module; heater controller; colour LCD and simple controls for ease of use. The device also allows data transfer/connectivity via an integrated wireless module (When used in a hospital setting, remote changes may not be appropriate for certain patients, as these setting changes may not be communicated to all hospital personnel treating the patient. Hospital staff should liaise with the patient's regular care provider such that the desired therapy outcome is achieved).

Juno VPAP ST-A utilizes a Micro-processor controlled blower system that generates positive airway pressure (CPAP) between 4-20 cmHJO as required to maintain an "air splint" for effective treatment of OSA and (Bilevel) pressures between 3-30 cmH2O for the treatment respiratory insufficiency.

The therapy modes available in the Juno VPAP ST-A include CPAP, Spontaneous, Spontaneous/Timed, Timed, PAC and iVAPS. In this subject device now features an optional "AutoEPAP" function on iVAPS mode. AutoEPAP automatically adjusts EPAP pressure (within set values) in response to flow limitations of the upper airway.

Juno VPAP ST-A is intended to be used under the conditions and purposes indicated in the labelling provided with the product.

It is a prescription device, supplied non-sterile.

AI/ML Overview

The document describes the Resmed Juno VPAP ST-A device (K161492), which is an update to a previously cleared device (K153061). The key change in the subject device is the addition of an optional "AutoEPAP" function to the iVAPS therapy mode.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets in the provided text. Instead, suitability for market is based on demonstrating substantial equivalence to the predicate device (Juno VPAP ST-A, K153061) and demonstrating that the new AutoEPAP algorithm provides equivalent therapy to the fixed EPAP setting of the predicate.

The reported device performance is that the AutoEPAP iVAPS therapy mode was as efficacious as iVAPS with fixed EPAP in terms of several clinical metrics.

Acceptance Criterion (Inferred from Substantial Equivalence Claim)Reported Device Performance (from Clinical Data)
Equivalence in maintaining upper airway patencyAutoEPAP iVAPS was as efficacious as iVAPS with fixed EPAP.
Equivalence in Apnea-Hypopnea Index (AHI)Equivalent
Equivalence in Oxygen Desaturation Index (ODI)Equivalent
Equivalence in Blood Oxygen Saturation (SPO2)Equivalent
Equivalence in Transcutaneous Carbon Dioxide (PtcCO2)Equivalent
Equivalence in Sleep QualityEquivalent
No serious adverse events or complicationsNo serious adverse events or complications related to the study device were recorded.
Compliance with relevant standards and specificationsThe subject device meets all requirements of the Juno VPAP ST-A System Specification and complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, FDA Draft Reviewer Guidance for Ventilators, and FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample size for the test set: Not explicitly stated. The document mentions "patients with respiratory insufficiency" but does not provide the number of participants.
  • Data provenance: Not explicitly stated, but the study was a "clinical trial" which implies prospective data collection. The country of origin is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. As this is a study evaluating the clinical efficacy of a ventilator mode, the "ground truth" would be established by direct physiological measurements and clinical assessments rather than expert review of images or data.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC study was not done. This device is a ventilator, not an AI-assisted diagnostic tool that would typically involve human readers interpreting data or images. The study focuses on the direct efficacy of the ventilator's automated function.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the clinical trial described appears to be a standalone performance evaluation of the AutoEPAP iVAPS algorithm. It directly compares the efficacy of the new automated algorithm with the fixed EPAP mode, which also functions automatically (without real-time human intervention during therapy delivery).

7. The Type of Ground Truth Used

The ground truth was established using physiological measurements and clinical outcomes data such as:

  • Apnea-Hypopnea Index (AHI)
  • Oxygen Desaturation Index (ODI)
  • Blood Oxygen Saturation (SPO2)
  • Transcutaneous Carbon Dioxide (PtcCO2)
  • Sleep Quality (presumably based on polysomnography or other clinical assessments)

8. The Sample Size for the Training Set

The document describes a clinical trial (double-blinded, randomized, crossover study) which evaluates the performance of the implemented algorithm. It does not mention a separate "training set" in the context of an AI/machine learning model development lifecycle. The AutoEPAP algorithm would have been developed and internally validated by the manufacturer, but the specific details of its training data (if any for a machine learning component) are not provided. The clinical trial serves as the validation or test set for the device's efficacy.

9. How the Ground Truth for the Training Set Was Established

As explained above, a "training set" in the machine learning sense is not explicitly discussed. For the clinical trial (validation data), the ground truth was established through direct physiological measurements and clinical assessments as listed in point 7.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, arranged in a way that resembles a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 19, 2017

Resmed Ltd % Larissa D'Andrea Director, Government & Regulatory Affairs Resmed Corp. 9001 Spectrum Center Boulevard San Diego, California 92123

Re: K161492

Trade/Device Name: Juno VPAP ST-A Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: MNS Dated: December 16, 2016 Received: December 19, 2016

Dear Larissa D'Andrea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Tina Kiang, Ph. D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161492

Device Name Juno VPAP ST-A

Indications for Use (Describe)

The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).

The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).

The Juno VPAP ST-A is intended for home and hospital use.

The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for ResMed. The logo consists of a series of blue and red dots arranged in a curved line, resembling a sine wave. Below the dots, the word "ResMed" is written in a bold, sans-serif font.

510(k) SUMMARY [As required by 21 CFR 807.92(c)]

    1. Date prepared January 13, 2017

2. Applicant information

Company Name/OwnerResMed Ltd1 Elizabeth Macarthur DriveBella Vista NSW 2153 Australia
Submitter NameMr. Jean-Nicolas BoudaudRegulatory Affairs Manager+61 2 88841000 (phone)+61 2 88842004 (fax)Jean.boudaud@resmed.com.au
Correspondent Details/Official ContactMs. Larissa D'AndreaDirector, Government & Regulatory Affairs9001 Spectrum Center BlvdSan Diego CA 92123 USA(858) 836 6837 (phone)Larissa.D'Andrea@resmed.comMr. Jean-Nicolas BoudaudRegulatory Affairs Manager+61 2 88841000 (phone)+61 2 88842004 (fax)Jean.boudaud@resmed.com.au

3. Device details and substantial equivalence claim [807.92(a)(3)]

Trade/Device NamesJuno VPAP ST-A
Device Common NameVentilator, continuous, non-life supporting
Regulation Number21 CFR §868.5895
Regulation NameAnesthesiology devices, Continuous Ventilator
Regulatory ClassClass II
Product CodeMNS
Predicate DeviceResMed Juno VPAP ST-A (K153061)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for ResMed. The logo consists of the word "ResMed" in bold, sans-serif font. Above the word is a curved line of dots that transition in color from blue to red. The dots are arranged in a wave-like pattern, starting with larger blue dots and gradually decreasing in size as they transition to red.

4. Device description

The subject device Juno VPAP ST-A retains all the same hardware, technologies and manufacturing characteristics as previously cleared in K153061. The device is used in a wider breathing circuit which typically comprises patient tubing and a mask (patient interface) to deliver a prescribed positive airway pressure treatment to patients.

Some of the kev features of the device include an in-line power supply: fully integrated humidifier: alarms module; heater controller; colour LCD and simple controls for ease of use. The device also allows data transfer/connectivity via an integrated wireless module (When used in a hospital setting, remote changes may not be appropriate for certain patients, as these setting changes may not be communicated to all hospital personnel treating the patient. Hospital staff should liaise with the patient's regular care provider such that the desired therapy outcome is achieved).

Juno VPAP ST-A utilizes a Micro-processor controlled blower system that generates positive airway pressure (CPAP) between 4-20 cmHJO as required to maintain an "air splint" for effective treatment of OSA and (Bilevel) pressures between 3-30 cmH2O for the treatment respiratory insufficiency.

The therapy modes available in the Juno VPAP ST-A include CPAP, Spontaneous, Spontaneous/Timed, Timed, PAC and iVAPS. In this subject device now features an optional "AutoEPAP" function on iVAPS mode. AutoEPAP automatically adjusts EPAP pressure (within set values) in response to flow limitations of the upper airway.

Juno VPAP ST-A is intended to be used under the conditions and purposes indicated in the labelling provided with the product.

It is a prescription device, supplied non-sterile.

5. Indications For Use

The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).

The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).

The Juno VPAP ST-A is intended for home and hospital use.

The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for ResMed. The logo features a curved line made up of a series of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

Predicate comparison 6.

CharacteristicsJuno VPAP ST-A(Subject device)Juno VPAP ST-A(Predicate device - K153061)Comments
Intended Use
Indications for useThe Juno VPAP ST-A isindicated to providenoninvasive ventilation forpatients weighing more than30lbs (13 kg) with respiratoryinsufficiency or obstructivesleep apnoea (OSA).The Juno VPAP ST-A isindicated to provide noninvasiveventilation for patients weighingmore than 30lbs (13 kg) withrespiratory insufficiency orobstructive sleep apnoea(OSA).Same intended use - Equivalent
The iVAPS mode is indicatedfor patients weighing more than66lbs (30 kg).The iVAPS mode is indicatedfor patients weighing more than66lbs (30 kg).
The Juno VPAP ST-A isintended for home and hospitaluse.The Juno VPAP ST-A isintended for home and hospitaluse.
The Humidifier is intended forsingle patient use in the homeenvironment and re-use in ahospital/institutionalenvironmentThe Humidifier is intended forsingle patient use in the homeenvironment and re-use in ahospital/institutionalenvironment
Environment of useHospital/HomeHospital/HomeSame environment of use - Equivalent
Prescription statusPrescription onlyPrescription onlySame prescription status - Equivalent
Therapies
Modes available• CPAP• S, ST, T• PAC• iVAPS with optionalAutoEPAP• CPAP• S, ST, T• PAC• iVAPS with fixed EPAP onlyThe subject device includes the same therapy modesas the predicate. An optional "AutoEPAP" function isadded on the iVAPS mode. The role of AutoEPAP is toautomatically adjust EPAP pressure (within a set min-max range) in order to maintain upper airway patency.Data is provided to demonstrate that the new algorithm
provides equivalent therapy to set fixed EPAP.
Equivalent
Pressure ranges• 4-20 cmH2O (CPAP)• 3-30 cmH2O (bi-level)o AutoEPAP: 3-25 cmH2O• 4-20 cmH2O (CPAP)• 3-30 cmH2O (bi-level)o Fixed EPAP: 3-25 cmH2OThe subject device operates within the same pressureranges as the predicate in all modes.
On the iVAPS (Bi-level mode), the range of adjustmentfor the new AutoEPAP feature is the same as with fixedEPAP, hence the same as the predicate.
Equivalent
Ramp settings• User selected as “Off” to 45minutes in 5 minuteincrements• Max Ramp time set atclinician's discretion• User selected as “Off” to 45minutes in 5 minuteincrements• Max Ramp time set atclinician's discretionSame ramp settings - Equivalent
Features
Alarms moduleYesYesSame integrated alarms module - Equivalent
HumidifierYesYesSame integrated humidifier - Equivalent
Motor typeBrush-less low voltage DCBrush-less low voltage DCSame motor - Equivalent
Operating systemMicrochip STM32F405ZGmicro-controller with ARM32-bitCortex™-M4 CPUMicrochip STM32F405ZGmicro-controller with ARM32-bitCortex™-M4 CPUSame operating system - Equivalent
Supplemental oxygenLabeled for use withSupplemental OxygenLabeled for use withSupplemental OxygenSame use with supplemental oxygen - Equivalent
Data transfer medium• SD Card• Fully integrated wirelessmodule• SD Card• Fully integrated wirelessmoduleSame data transfer and connectivity mediums -Equivalent
Hardware platform• Gen 10• Gen 10Same hardware - Equivalent

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for ResMed. The logo features a curved line made up of a series of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the ResMed logo. The logo features a curved line made of connected circles that transition in color from blue to red. Below the curved line, the word "ResMed" is written in a bold, sans-serif font.

7. Non Clinical data submitted

Bench test data is presented to demonstrate that the subject device meets all requirements of the Juno VPAP ST-A System Specification. For the new AutoEPAP algorithm, this includes characterization of the ventilator's response to a breathing machine that simulates patients flow limitations and apnoeas.

The Juno VPAP ST-A has also been tested to the relevant FDA consensus standards and other applicable requirements passing all test protocols. The Juno VPAP ST-A with and without the integrated heated humidifier (HumidAir) was designed and tested according to:

  • . IEC 60601-1:2005+AMD1:2012, Medical electrical equipment - Part 1: General requirements for safety Medical electrical equipment - General requirements for basic safety and essential performance
  • . IEC 60601-1-2:2014. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • . IEC 60601-1-8:2006, Medical electrical equipment -- Part 1-8: General requirements for basic safety and essential performance -- Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • IEC 60601-1-11:2010, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

The device complies with the applicable requirements referenced in the FDA quidance documents:

  • FDA Draft Reviewer Guidance for Ventilators (July 1995) .
  • FDA Guidance for the Content of Premarket Submissions for Software Contained in ● Medical Devices (May 11, 2005)

8. Clinical data

Clinical trial data is provided to demonstrate that the added AutoEPAP iVAPS algorithm performed as expected in maintaining upper airway patency in patients with respiratory insufficiency. The data relates to a double-blinded, randomised, crossover study comparing the efficacy of the AutoEPAP iVAPS therapy mode with the fixed EPAP iVAPS mode (as previously used on the predicate).

In patients with respiratory insufficiency, the trial demonstrated that therapy with AutoEPAP iVAPS was as efficacious as iVAPS with fixed EPAP in terms of AHI, ODI, SPO2, PtcCO2 and sleep quality.

No serious adverse events or complications related to the study device were recorded.

9. Substantial Equivalence Conclusion

This submission demonstrates that the subject device is substantially equivalent to the predicate Juno VPAP ST-A (K153061).

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).