K Number
K161483

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-10-31

(153 days)

Product Code
Regulation Number
N/A
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Unger Quad Injector is intended to be used as the needle infusion set component of a tissue expander fill kit for breast tissue expanders designed to be filled with sterile saline using 21-gauge needles. Specifically, the Unger Quad Injector is indicated for assisting the clinician in delivery of sterile saline into the surgically-placed, sub-dermal, temporary, removable tissue expander.

Device Description

The Unger Quad Injector is a 'needle infusion set' component to a breast tissue expander fill kit. The device is comprised of four 21 gauge hypodermic lumen needles, which are connected to a needle hub, a luer adapter and tubing.

AI/ML Overview

This document is a 510(k) premarket notification for the Unger Quad Injector, indicating it is a device intended for assisting in the delivery of sterile saline into breast tissue expanders. However, the provided text does not contain acceptance criteria or a study that evaluates the device against specific performance metrics for the purpose of demonstrating clinical accuracy or diagnostic performance, which is typically found in documentation for diagnostic or AI-enabled devices.

Instead, this document focuses on demonstrating substantial equivalence to existing predicate devices based on technological characteristics and non-clinical bench testing. The "acceptance criteria" discussed here are primarily related to conforming to established standards and matching the performance of predicate devices in specific non-clinical tests.

Here's a breakdown of the information that can be extracted or inferred from the provided text, adjusted to the context of this device:


Acceptance Criteria and Device Performance (Non-Clinical Bench Testing)

Since this is a non-AI, non-diagnostic device, the "acceptance criteria" are related to manufacturing standards, material properties, and functional performance compared to predicate devices, rather than accuracy metrics like sensitivity or specificity.

Acceptance Criteria (Standard / Comparator)Reported Device Performance (Unger Quad Injector)
Hub / Needle Bond Strength: Minimum 44N (Per ISO 7864)Meets Criteria: 21G – Minimum 44N (Per ISO 7864)
Tip Configuration: Per ISO 7864Meets Criteria: Per ISO 7864
Needle Color: Green (for 21G) (Following ISO 6009)Meets Criteria: 21G Needle Color: Green
Flow Rate Performance: Comparable to B. Braun infusion setDemonstrated Comparability: Flow rate comparison study conducted. (Specific results not detailed)
Material Biocompatibility: Identical to equivalent components in 510(k) cleared devicesDemonstrated Comparability: Materials are identical or comparable to predicate/reference devices.
Sterilization: In accordance with ANSI/AAMI/ISO 11135-2014, ISO 11135-1 & -2Validated: Product is EO sterilized, validation in accordance with standards.
Shelf Life/Expiration Dating: Established through appropriate aging testingUnder Establishment/Validated: Aging testing conducted; shelf life will be established.
Bioburden/Endotoxin: Pass specific testsPassed: Bioburden Test, Endotoxin Test, Bioburden Validation Test conducted.
Filling, Sealing, Leakage Prevention: Comparable to predicate devicesDemonstrated Comparability: Testing results demonstrate capability. (Specific results not detailed)
Conformance: ISO 7864 and ISO 9626Demonstrated Conformance: Bench testing confirmed conformance.

1. A table of acceptance criteria and the reported device performance:
See table above. The "reported device performance" indicates whether the device met the stated standard or showed comparability to the predicate/reference device, as described in the "Non Clinical Tests performed" section and the "Summary of the Technological Characteristics" table.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document does not detail specific sample sizes for each non-clinical bench test. The studies were non-clinical bench tests, not involving human subjects or real-world data provenance in the sense of patient data. The tests were conducted to verify equivalence.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This is not applicable as this device is not an AI/diagnostic device that requires expert-established ground truth for a test set. The evaluation is based on engineering and material standards and comparison to predicate devices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. The tests are bench tests against established standards or direct comparisons, not requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is not an AI-enabled diagnostic or assistance tool. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This device does not involve an algorithm. The "standalone" performance here refers to the device's functional characteristics in isolation, which were assessed through the bench tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for this type of device is defined by:

  • Established Industry Standards: Such as ISO 7864, ISO 9626, ISO 6009, ANSI/AAMI/ISO 11135.
  • Performance of Legally Marketed Predicate Devices: The McGhan Tissue Expander Fill System (K870754) and other reference devices served as comparators for technological characteristics, materials, and functional performance.

8. The sample size for the training set:
Not applicable. This device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established:
Not applicable. There is no training set for this type of device.


Summary of the Study:

The document describes a series of non-clinical bench tests rather than a clinical study. These tests were performed to demonstrate that the Unger Quad Injector is substantially equivalent to legally marketed predicate devices, primarily the McGhan Tissue Expander Fill System (K870754).

The specific non-clinical tests conducted include:

  • Pre-clinical Investigational Evaluation of Tissue Expander Inter-operative Filling Device
  • Quad Injector Flow Rate Comparison
  • Unger Quad Injector Bioburden Test
  • Unger Quad Injector Endotoxin Test
  • Unger Quad Injector Bioburden Validation Test
  • Shelf Life / Expiration Dating – Aging Testing
  • ISO 7864 Hub Needle Bond Strength
  • ISO 7864 and ISO 9626 Conformance

Purpose of the Studies: The main purpose was to verify the equivalence of the Unger Quad Injector's performance, safety, and specifications with the predicate devices, and to ensure conformance with relevant ISO standards.

Conclusion from the Studies: The data from these tests "supports a finding of substantial equivalence because the data demonstrates that the Unger Quad Injector is safely and effectively indicated for use as the 'needle infusion set' component... Testing results demonstrate the filling, sealing capability and leakage prevention as compared to lawfully marketed predicate devices. Performance testing results also demonstrate safety and efficacy in terms of biocompatibility, sterilization, shelf life, and general device performance."

No clinical tests were conducted. The device's safety and effectiveness were established through comparison of its technological characteristics and performance in bench tests to those of existing, legally marketed devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 31, 2016

Stradis Healthcare % Mr. Carl H. Jenkins The Wood Burditt Group 10 East Scranton Avenue, Suite 201 Lake Bluff, Illinois 60044

Re: K161483

Trade/Device Name: Unger Quad Injector Regulatory Class: Unclassified Product Code: LCJ Dated: September 8, 2016 Received: September 12, 2016

Dear Mr. Jenkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

David-Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161483

Device Name Unger Quad Injector

Indications for Use (Describe)

The Unger Quad Injector is intended to be used as the needle infusion set component of a tissue expander fill kit for breast tissue expanders designed to be filled with sterile saline using 21-gauge needles. Specifically, the Unger Quad Injector is indicated for assisting the clinician in delivery of sterile saline into the surgically-placed, sub-dermal, temporary, removable tissue expander.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date prepared: October 28, 2015

Submitter / Contact PersonH. Carl JenkinsThe Wood Burditt Group10 E. Scranton Ave, Suite 201Lake Bluff, IL 60044(ph) 847-234-7500 x 205(fax) 847-578-0728(email) hcjenkins@woodburditt.com
Applicant / SpecificationsDeveloper / PackagerStradis Healthcare805 Marathon Pkwy, Ste 100Lawrenceville, GA 30046(ph) 847-596-7716FDA Establishment Registration Number: 1055754FEI Number: 1000111025
ManufacturerSurgistar, Inc2310 La Mirada Dr
2310 La Mirada Dr.Vista, CA 92081(ph) 760-598-2480
FDA Establishment Registration Number: 2028661FEI Number: 3005765014
Device
Trade NameUnger Quad Injector
Proprietary NameUnger Quad Injector
Common NameNeedle Infusion Set - Breast Tissue Expander Accessory
Classification NameBreast Tissue Expander - Accessory
Advisory PanelGeneral and Plastic Surgery Advisory Panel
RegulationUnclassified
Product CodeLCJ
ClassificationUnclassified

{4}------------------------------------------------

Reason for 510(k) Submission

The applicant device is a new device.

Device Description

The Unger Quad Injector is a 'needle infusion set' component to a breast tissue expander fill kit. The device is comprised of four 21 gauge hypodermic lumen needles, which are connected to a needle hub, a luer adapter and tubing.

The Unger Quad Injector was developed with the intent to improve the infusion of saline solution into a tissue expander used in breast augmentation procedures.

Indications for Use:

The Unger Quad Injector is intended to be used as the needle infusion set component of a tissue expander fill kit for breast tissue expanders designed to be filled with sterile saline using 21gauge needles. Specifically, the Unger Quad Injector is indicated for assisting the clinician in delivery of sterile saline into the surgically-placed, sub-dermal, temporary, removable tissue expander.

Predicate and Reference Device Summary Table

Based on the comparison of the device features, materials, intended use and performance the Unger Quad Injector was shown to be substantially equivalent to the lawfully marketed predicate devices indicated in the table below.

DeviceApplicant510(k) #Featured Element(s) ofSubstantial Equivalence
McGhan TissueExpander Fill System(Predicate Device)McGhanK870754Intended Use / Indicationsfor Use / TechnologicalCharacteristics
McGhan TissueExpander Fill Kit (theneedles for which arealso marketed as the B.McGhanK853014Intended Use / Indicationsfor Use; Needle Gauge;
Braun Infusion Set –Universal Fill Kit Component(Reference Device)Labeling / TechnologicalCharacteristics
Terumo Needle(Reference Device)TerumoK121607Needle Length; DeviceColor
IMC HypodermicNeedle(Reference Device)IMCK102584Tip Configuration
K-Pack II Needle(Reference Device)K-PackK150263Hub / Needle BondStrength
Sterilab Needles(Reference Device)SterilabK961524Biocompatibility;Materials
Surgistar Blades(Reference Device)SurgistarK923689Biocompatibility;Materials
Winged Infusion Set(Reference Device)Distronic SterileProductsK955709Biocompatibility;Materials
IntravascularAdministration Set(Reference Device)Distronic SterileProductsK945184Biocompatibility;Materials
Core-Resistant HuberInfusion Set(Reference Device)Distronic SterileProductsK950597Biocompatibility;Materials

{5}------------------------------------------------

{6}------------------------------------------------

Technological characteristics and comparison

The Unger Quad Injector is substantially equivalent to the McGhan Tissue Expander Fill System (K870754) in terms of technological characteristics and indications for use. The predicate device is comprised, in part, of a 21 G butterfly needle for filling the injection reservoir of the tissue expander.

In addition, several other Reference Devices demonstrate the safety and efficacy of the Unger Quad Injector in terms of supporting scientific methodology and/or standard reference values.

Summary of the Technological Characteristics of Subject Device and Predicate/Reference Devices (21 CFR 807.92(a)(6))

FeatureUnger Quad InjectorPredicate and Reference Devices
DesignNeedle infusion set component ofa tissue expander fill kit for breasttissue expanders; lumen needles atone end of the device; femaleconnector (hub) designed to matewith a male connector (nozzle) of asyringe at the other end of thedevice.Predicate Device:McGhan Tissue Expander FillSystem (K870754): "A 21G butterflyneedle for filling the injectionreservoir of the tissue expander"(p. 7, Attachment 12A).Reference Device:McGhan Tissue Expander Fill Kit(K853014): Needle infusion setcomponent "for use with AllerganFill Kit" (lumen needles at one endof the device; hub designed tomate with nozzle of a syringe at theother end of the device).
Gauge21GPredicate Device:McGhan Tissue Expander FillSystem (K870754): "A 21G butterflyneedle for filling the injectionreservoir of the tissue expander"
Reference Device:
McGhan Tissue Expander Fill Kit(marketed as the B. Braun InfusionSet - Universal Fill Kit Component)(K853014): "21 Ga. Needle InfusionSet".
Needle Length38mmReference Device:Terumo Needle (K121607): 21GNeedle Length: 38mm
Tip ConfigurationPer ISO 7864Reference Device:IMC Hypodermic Needle(K102584): Per ISO 7864
ColorHub color coded following ISO6009. 21G Needle Color: GreenReference Device:Terumo Needle (K121607): 21GNeedle Color: Green
Hub / Needle Bond Strength21G – Minimum 44NPer ISO 7864Reference Device:K-Pack II Needle (K150263): 21G –Minimum 44N
MaterialsNeedle: 304 Stainless SteelHypodermic TubingSame Needle Material (304Stainless Steel Hypodermic Tubing)used in:- K853014 - B. Braun InfusionSet (Predicate Device)- K955709 – Winged andSubcutaneous Infusion Set(Reference Device)- K961524 – Sterilab SterileDisposable Instruments –Needles - Identical 304Stainless Steel HypodermicTubing. (Reference Device)
Needle Cover/Sheath: 640I LDPE(Low Density Polyethylene)Same Cover/Sheath Material usedin:- K945184 - IntravascularAdministration Set (ReferenceDevice)- K950597 – Core-ResistantHuber Infusion Set (ReferenceDevice)
ABS Cap and Base: ABS PlasticSame ABS plastic used in:- K923689 – Surgistar Blades –Handles made of same ABSPlastic (Reference Device)
Tubing: Non-DEHP PVCSame material used for tubing in:- K955709 Winged &Subcutaneous Infusion Set(Reference Device)- K945184 - IntravascularAdministration Set (ReferenceDevice)
Luer Lock: White PolypropyleneSame material used in luer lock in:- K961524 – Sterilab Needleswith Polypropylene Hub(Reference Device)
Chevron Pouch – Mylar/TyvekSame Mylar/Tyvek pouches in:- K955709 Winged &Subcutaneous Infusion Set(Reference Device)

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Component materials and specifications are presented in the table below:

DescriptionMaterialDimensionsAdditional Specs.
21G HypodermicNeedle304 StainlessHypodermicTubing2.087" X .0323" OD X.0233" IDTri-Facet Grind Bevel
Needle CoverSheath640I LDPE2.5" X .506" OD X.372" IDNatural Color
ABS Cap and BaseABS Plastic1" X .394" OD X .125"IDISO 6009 -- 21G Green
TubingNon-DEHP PVC12" X .184" OD X.125" IDClear, 70 Durometer
Luer LockWhitePolypropylene1" X .234" OD Barb X.117" IDWhite Plastic
Chevron Pouch48GA PET/.002LDPE7.75" X 3.75"Clear/White

The Hub/Needle Bond Strength is: Minimum 44N, Per ISO 7864.

Clinical test:

No clinical tests were conducted in support of this 510(k) submission.

{10}------------------------------------------------

Non Clinical Tests performed:

Bench testing related to performance, safety, effectiveness and specifications of the proposed device was conducted in order to verify the equivalence of performances with the predicate devices.

The following bench testing was conducted on the Unger Quad Injector in support of this 510(k) submission:

Bench TestPurpose
Pre-clinical Investigational Evaluation ofTissue Expander Inter-operative FillingDevice: Stradis Unger Quad InjectorCompare performance of the B. Braun infusionset with the Unger Quad Injector as they performas part of Tissue Expander Fill Kits
Quad Injector Flow Rate ComparisonCompare performance of the B. Braun infusionset with the Unger Quad Injector as they performas part of Tissue Expander Fill Kits
Unger Quad Injector Bioburden TestBioburden Test
Unger Quad Injector Endotoxin TestEndotoxin Test
Unger Quad Injector Bioburden ValidationTestBioburden Validation Test
Shelf Life / Expiration Dating – AgingTestingDevice Functionality and Packaging Integrity
ISO 7864 Hub Needle Bond StrengthTo Determine Hub Needle Bond Strength inconformance with ISO 7864
ISO 7864 and ISO 9626Conformance with ISO 7864 and ISO 9626

{11}------------------------------------------------

The data from the performance testing supports a finding of substantial equivalence because the data demonstrates that the Unger Quad Injector is safely and effectively indicated for use as the 'needle infusion set' component of a tissue expander fill kit for breast tissue expanders designed to be filled with sterile saline using 21 gauge needles. Testing results demonstrate the filling, sealing capability and leakage prevention as compared to lawfully marketed predicate devices. Performance testing results also demonstrate safety and efficacy in terms of biocompatibility, sterilization, shelf life, and general device performance.

Sterilization:

Product is EO sterilized. Sterilization validation is in accordance with ANSI/AAMI/ISO 11135-2014, ISO 11135-1 and 11135-2.

The shelf life of the product will be established in accordance with appropriate and valid aging testing results.

Biocompatibility:

The Unger Quad Injector components are made with materials identical to equivalent components in other 510(k) cleared devices.

Conclusion

The applicant device is substantially equivalent in its intended use, technology / principal of operation, materials, and performance to the predicate devices identified in this 510(k) submission. There is no significant difference that raises any issues of safety or effectiveness.

N/A