K Number
K161457

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-11-21

(179 days)

Product Code
Regulation Number
886.3920
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The XEN Glaucoma Treatment System is indicated for the management of refractory glaucomas, including cases where previous surgical treatment has failed, cases of primary open angle glaucoma, and pseudoexfoliative or pigmentary glaucoma with open angles that are unresponsive to maximum tolerated medical therapy.

Device Description

The XEN Glaucoma Treatment System consists of the XEN45 Gel Stent preloaded into the XEN Injector. The XEN45 Gel Stent is composed of a gelatin derived from porcine dermis, formed into a tube, and then cross-linked with glutaraldehyde. The inside diameter of the tube is approximately 45 um, its outside diameter is approximately 150 um, and it has a length of approximately 6 mm. The XEN45 Gel Stent creates a permanent channel through the sclera allowing aqueous flow from the anterior chamber to the subconjunctival space.

The XEN45 Gel Stent is preloaded into an injector designed to place the XEN45 Gel Stent in the intended position. The injector with the preloaded implant is sterilized via gamma irradiation and is provided sterile. The injector is discarded after a one-time use.

AI/ML Overview

The document provided is a 510(k) summary for a medical device (XEN Glaucoma Treatment System) and details the performance data used to demonstrate substantial equivalence to predicate devices, rather than formally outlining acceptance criteria for AI/ML-based device performance and a study proving it meets those.

Therefore, many of the requested elements (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth types for test/training, training set build) are not applicable or cannot be extracted from this document, as they relate to AI/ML device validation, which is not the subject of this 510(k) summary.

However, I can extract information related to the clinical effectiveness and safety evaluation of the XEN Glaucoma Treatment System, which serves as its "performance data" in this context.

Here's an attempt to structure the available information relevant to "acceptance criteria" and "device performance" as per the document provided, interpreting "acceptance criteria" as the measures of effectiveness and safety used to demonstrate equivalence.


Device: XEN Glaucoma Treatment System

Study Type: K161457 510(k) Premarket Notification - demonstrating substantial equivalence to predicate devices.

Interpretation of "Acceptance Criteria" for this device: For this type of medical device (an aqueous shunt), acceptance criteria are primarily demonstrated through clinical effectiveness in lowering intraocular pressure (IOP) and an acceptable safety profile, comparable to legally marketed predicate devices. The "performance data" presented is the evidence that these clinical objectives were met.

1. Table of "Acceptance Criteria" (Effectiveness and Safety Measures) and Reported Device Performance

"Acceptance Criteria" Aspect (as inferred from effectiveness/safety goals)Reported Device Performance (XEN Glaucoma Treatment System)
Effectiveness:
IOP Reduction (> 20% from baseline)76.3% of subjects achieved a 12-month mean diurnal IOP reduction of > 20% from baseline on the same or fewer IOP-lowering medications.
Mean Diurnal IOP Reduction (at 12 months)-6.4 ± 1.1 (SE) mmHg mean diurnal IOP reduction from baseline at the 12-month visit.
Mean Diurnal IOP at 12 Months15.9 (± 5.2) mmHg (for 52 subjects with observed data) from a baseline of 25.1 (± 3.7) mmHg.
Mean Number of IOP-Lowering Medications at 12 Months1.7 (± 1.5) medications (from a baseline of 3.5 ± 1.0 medications).
Safety:
Adverse Event Profile Comparison to PredicatesThe adverse event profile of the XEN45 Gel Stent is comparable to publicly available data on legally marketed predicate devices.
Most Common Adverse EventsAdditional glaucoma surgery (with/without device explant), hypotony (< 6 mmHg), IOP increase ≥ 10 mmHg from baseline, needling procedures.
Overall Safety ConclusionAs safe as the predicates and does not raise any different or unanticipated safety questions.

2. Sample Size and Data Provenance

  • Clinical Study Sample Size: 65 subjects were implanted with the XEN45 Gel Stent.
  • Data Provenance:
    • Clinical Study: Prospective, multi-center, single-arm, open-label clinical trial conducted at 12 sites in the U.S.
    • Additional Data Considered: OUS (Outside U.S.) data, including an abstract (Grover et al, 2016 AGS abstract "Describing the Development of a Minimally Invasive Collagen Stent for Treating Glaucoma: First 975 Eyes Treated with the XEN Gel Stent," p. 86) and post-market surveillance data from the European Union and Canada.

3. Number of Experts and Qualifications for Ground Truth

  • Not applicable to this device submission. This is a medical device for implantation, not an AI/ML diagnostic or image analysis tool requiring expert radiologists for ground truth establishment. The "ground truth" is the clinical outcome (IOP change, adverse events observed in patients).

4. Adjudication Method for the Test Set

  • Not applicable for a clinical trial evaluating device performance. Clinical trial data collection follows predefined protocols and statistical analysis plans.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not applicable. This is a clinical trial for an implantable medical device, not a comparative study of AI assistance for human readers.

6. Standalone Performance (Algorithm Only)

  • Not applicable. This is a physical medical device (implant), not an algorithm or AI. Its performance is inherent to its physical function within the patient.

7. Type of Ground Truth Used

  • Clinical Outcomes Data: For effectiveness, the "ground truth" was objective measurements of intraocular pressure (IOP) and number of IOP-lowering medications over time, as well as observed adverse events for safety. This is direct patient outcome data from a clinical trial.

8. Sample Size for the Training Set

  • Not applicable. This is not an AI/ML device that undergoes a "training" phase with data. The clinical trial serves as the primary dataset for demonstrating safety and effectiveness for regulatory approval.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable (as above). The "ground truth" (clinical outcomes) was established through direct observation and measurement in the prospective clinical trial.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular emblem that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 21, 2016

Allergan, Inc. Ms. Barbara Niksch Executive Director, Clinical Affairs 26970 Aliso Vieio Parkwav, Suite 200 Aliso Viejo, CA 92656

Re: K161457

Trade/Device Name: Xen Glaucoma Treatment System Regulation Number: 21 CFR 886.3920 Regulation Name: Aqueous Shunt Regulatory Class: Class II Product Code: KYF Dated: October 7, 2016 Received: October 11, 2016

Dear Ms. Niksch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161457

Device Name Xen Glaucoma Treatment System

Indications for Use (Describe)

The XEN Glaucoma Treatment Systemis indicated for the management of refractory glaucomas, including cases where previous surgical treatment has failed, cases of primary open angle glaucoma, and pseudoexfoliative or pigmentary glaucoma with open angles that are unresponsive to maximum tolerated medical therapy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

I. SUBMITTERAllergan, Inc.2525 Dupont DriveMail Code T2-6LIrvine, CA 92612
Contact Person:Barbara A. NikschExecutive Director, Clinical AffairsAllergan, Inc.(714) 246-6914
Date Summary Prepared:November 16, 2016
II. DEVICE
Trade Name:XEN Glaucoma Treatment System
Common Name:Glaucoma Implant

Classification Name: Aqueous shunt (21 CFR 886.3920)

Class II (special controls)

Device Product Code: KYF ("Implant, Eye Valve")

III. PREDICATE DEVICES

Device Class:

  • Ahmed Glaucoma Valve Implant, Model S2 (K925636) .
  • EX-PRESS Glaucoma Filtration Device (formerly referred to as the Ex-Press Mini Glaucoma . Implant (K012852))

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The XEN Glaucoma Treatment System consists of the XEN45 Gel Stent preloaded into the XEN Injector. The XEN45 Gel Stent is composed of a gelatin derived from porcine dermis, formed into a tube, and then cross-linked with glutaraldehyde. The inside diameter of the tube is approximately 45 um, its outside diameter is approximately 150 um, and it has a length of approximately 6 mm. The XEN45 Gel Stent creates a permanent channel through the sclera allowing aqueous flow from the anterior chamber to the subconjunctival space.

The XEN45 Gel Stent is preloaded into an injector designed to place the XEN45 Gel Stent in the intended position. The injector with the preloaded implant is sterilized via gamma irradiation and is provided sterile. The injector is discarded after a one-time use.

V. INTENDED USE

The XEN Glaucoma Treatment System has the same intended use as the predicates and all other devices regulated within the generic type of device known as aqueous shunts in accordance with 21 CFR 886.3920. The device is a prescription (Rx) device that is intended to be permanently implanted to reduce intraocular pressure for the management of glaucoma. The XEN Glaucoma Treatment System will bear the following indications for use statement:

The XEN Glaucoma Treatment System is indicated for the management of refractory glaucomas, including cases where previous surgical treatment has failed, cases of primary open angle glaucoma, and pseudoexfoliative or pigmentary glaucoma with open angles that are unresponsive to maximum tolerated medical therapy.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICES

The technological characteristics of the XEN Glaucoma Treatment System, the Ahmed Glaucoma Valve and the EX-PRESS Glaucoma Filtration Device are similar. Furthermore, the differences in technological characteristics that exist between the XEN Glaucoma Treatment System and both legally marketed predicate devices do not raise different questions of safety and effectiveness.

{5}------------------------------------------------

Comparison of the XEN Glaucoma Treatment System and the Predicate Devices

XEN GlaucomaTreatment SystemAhmed Glaucoma ValveEX-PRESS GlaucomaFiltration Device
Intended UseTo be permanentlyimplanted to reduceintraocular pressure for themanagement of glaucomaTo be permanentlyimplanted to reduceintraocular pressure for themanagement of glaucomaTo be permanentlyimplanted to reduceintraocular pressure for themanagement of glaucoma
Indications for UseThe XEN GlaucomaTreatment System isindicated for themanagement of refractoryglaucomas, including caseswhere previous surgicaltreatment has failed, casesof primary open angleglaucoma, andpseudoexfoliative orpigmentary glaucoma withopen angles that areunresponsive to maximumtolerated medical therapy.The Ahmed GlaucomaValve is indicated for themanagement of refractoryglaucomas, where previoussurgical treatment hasfailed, or by experience isknown not to providesatisfactory results. Suchrefractory glaucomas caninclude neovascularglaucoma, primary openangle glaucomaunresponsive to medication,congenital or infantileglaucoma and refractoryglaucomas resulting fromaphakia or uveitis.The EX-PRESS GlaucomaFiltration Device isintended to reduceintraocular pressure inglaucoma patients wheremedical and conventionalsurgical treatments havefailed.
Rx or OTCRxRxRx
Permanent ImplantYesYesYes
Mechanism of ActionCreates an outflow pathwayfrom the anterior chamberto the subconjunctivalspace through whichaqueous humor can flowresulting in a conjunctivalbleb.Creates a channel via a tubefor aqueous flow from theanterior chamber to thesubconjunctival spaceresulting in a bleb formedwithin the encapsulatedpolypropylene plate.Creates an outflow pathwayfrom the anterior chamberto the subconjunctivalspace through whichaqueous humor can flowresulting in a conjunctivalbleb.
Method of InsertionVia a pre-loaded XENInjectorManual insertion withClass I surgical instrumentsVia a pre-loadedEX-PRESS DeliverySystem (EDS)
DesignSingle piece tubeDrainage tube and avalve/reservoir bodySingle piece tube
MaterialPorcine collagen/gelatinSilicone drainage tubePolypropylenevalve/reservoir bodySilicone elastomer valvemembraneStainless steel
XEN GlaucomaTreatment SystemAhmed Glaucoma ValveEX-PRESS GlaucomaFiltration Device
Size(nominal dimensions)XEN45 Gel Stent drydimensions:6 mm length45 microns inner diameter150 microns outer diameterImplant:13 mm width, 16 mmlength, 1.9 mm heightTube: 0.317 mm innerdiameter (317 microns) and0.635 mm outer diameter(635 microns)Single piece tube: 2-3 mmlength and 0.4 mm (400microns) outer diameterModels R30 and R50 wereoriginally cleared and havedifferent flowcharacteristics created bydifferent wire diameterswelded into the device.The higher the number, thelower the resistance. The"30" and "50" designatorsrefer to the size of the innerdiameter in microns.
Sterile, Single-UseYesYesYes
Sterilization MethodGammaGammaGamma

{6}------------------------------------------------

VII. PERFORMANCE DATA

Bench Testing A.

The nonclinical bench testing conducted on the XEN Glaucoma Treatment System included design verification and functional product testing, sterilization validation, packaging and shelf life testing, and biocompatibility testing. Results of the nonclinical testing demonstrate that the XEN Glaucoma Treatment System meets the defined specifications.

Design Verification and Functional Product Testing:

The XEN45 Gel Stent and the XEN injector were evaluated to verify that the design output met the original design input and intent. This testing was based on tests described in ANSI Z80.27, Implantable Glaucoma Devices. The results establish that the XEN Glaucoma Treatment System meets the design intent and complies with the applicable requirements.

Sterilization Validation:

The gamma irradiation sterilization method was validated using the VDmas 26 method described in BS EN ISO 11137-1:2015 and BS EN ISO 11137-2:2015. Validation results demonstrate that a minimum exposure dose of 25 kGy has been substantiated for the routine sterilization of the XEN Glaucoma Treatment System to provide a 10-6 sterility assurance level (SAL).

{7}------------------------------------------------

Packaging and Shelf Life Testing:

The XEN Glaucoma Treatment System is labeled with an expiration date of 3 years. The shelf life study evaluated the functional performance of the XEN45 Gel Stent and the XEN injector, as well as the packaging integrity of the tray sealed with the Tyvek lid. Additional testing was completed to evaluate the impact of environmental conditioning and distribution factors. Test results confirm that the XEN45 Gel Stent and the XEN injector meet their functional requirements and the sterile barrier (package integrity) remains intact after simulated distribution and aging. This testing provides the justification for the 3-year shelf life and the maintenance of the sterile barrier.

Biocompatibility Testing

Biocompatibility of the final finished XEN Glaucoma Treatment System was demonstrated through testing in accordance with BS EN ISO 10993-1 "Biological evaluation of medical devices, Part 1: Evaluation and testing within a risk management process" (refer to the table below for a listing of all tests performed). Testing was performed for the Gel Stent material and for the injector components that have contact with the patient or the Gel Stent. All testing demonstrated that the device materials have an acceptable biocompatibility profile.

TestResults
Gel Stent Material Testing
Cytotoxicity (MEM Elution)Non-cytotoxic
Guinea Pig Maximization SensitizationNon-irritant compared to negative control
Acute Systemic Toxicity (Mouse)Non-toxic
Intraocular Irritation (Rabbit)No evidence of irritation
Bacterial Mutagenicity – Ames AssayNon-mutagenic
4 Week Intramuscular Implant (Rabbit)Non-irritant compared to USP negativecontrol
Direct Contact CytotoxicityNon-cytotoxic
Immunogenicity (Mouse)Non-immunogenic
Injector Component Testing
Cytotoxicity (MEM Elution)Non-cytotoxic
Guinea Pig Maximization SensitizationNon-irritant compared to negative control
Intraocular Irritation (Rabbit)No evidence of irritation

Listing of All Biocompatibility Tests Performed on All Patient Contacting Materials

B. Clinical Evidence Supporting Substantial Equivalence

A prospective, multi-center, single arm, open-label, clinical trial was conducted at 12 sites in the U.S. to evaluate the safety and effectiveness of the XEN45 Gel Stent in refractory glaucoma

{8}------------------------------------------------

subjects where previous filtering or cilioablative procedures failed or IOP was unresponsive to maximally tolerated medical therapy . Sixty-five subjects were implanted with the XEN45 Gel Stent and 18-month data were collected for safety.

The safety and effectiveness data from this clinical study establish, in the context of the publicly available data for the predicate devices, that there is no difference with respect to substantial equivalence in safety and effectiveness outcomes between the XEN45 Gel Stent and legally marketed predicate devices, confirming that the XEN45 Gel Stent is as safe and effective as the predicate devices.

Effectiveness

The results for the specified effectiveness outcomes (based on multiple imputations) are as follows:

  • 76.3% of subjects were able to obtain a 12-month mean diurnal IOP reduction of . > 20% from baseline on the same or fewer IOP-lowering medications.
  • The mean diurnal IOP reduction from baseline at the 12-month visit was -6.4 ± 1.1 (SE). .

The results of additional analyses are as follows:

  • The mean diurnal IOP at baseline for all subjects (n=65) was 25.1 (+ 3.7) mmHg and the . mean diurnal IOP at 12 months (using only observed data) for the 52 subjects who were seen at the 12-month visit was 15.9 (± 5.2) mmHg.
  • The mean number of IOP-lowering medications at baseline was 3.5 (± 1.0) compared to . 1.7 (± 1.5) medications at the 12-month visit (based on observed data)

The effectiveness data, when compared to the publicly available data for the predicate devices, demonstrate that the XEN Glaucoma Treatment System lowers IOP to similar levels as the predicate devices without raising different questions regarding effectiveness.

Safety

The most common adverse events reported prior to and including the 12-month visit were: additional glaucoma surgery with or without device explant, hypotony defined as < 6 mmHg at any time point regardless of whether there were any associated complications or sequelae related to the low pressure, IOP increase ≥ 10 mmHg from baseline, and needling procedures.

In addition to the U.S. Clinical Trial, OUS data (e.g., Grover et al, 2016 AGS abstract "Describing the Development of a Minimally Invasive Collagen Stent for Treating Glaucoma: First 975 Eyes Treated with the XEN Gel Stent," p. 86 and post-market surveillance data from the European Union and Canada) were considered for thoroughness.

{9}------------------------------------------------

When comparing the adverse event profile of the XEN45 Gel Stent to publicly available data on the legally marketed predicate devices, the XEN45 Gel Stent is as safe as the predicates and does not raise any different or unanticipated safety questions.

VIII. CONCLUSIONS

The XEN Glaucoma Treatment System has the same intended use as the legally marketed predicate devices identified in this 510(k) submission and all other aqueous shunts regulated by FDA under 21 CFR § 886.3920. The indications for use statement differs from those for the predicate devices, however, the differences do not alter the therapeutic effect of the device type and are based on clinical experience with the device. The XEN Glaucoma Treatment System's technological characteristics differ from the predicate devices, however, the differences do not raise new or different questions of safety or effectiveness. Results of the nonclinical testing demonstrate that the XEN Glaucoma Treatment System meets the defined specifications. Clinical data establish that the device effectively lowers intraocular pressure in refractory glaucoma patients demonstrating that the XEN Glaucoma Treatment System is as safe and effective as the predicate devices. In summary, the XEN Glaucoma Treatment System is substantially equivalent to the Ahmed Glaucoma Valve and the EX-PRESS Glaucoma Filtration Device when used in the management of refractory glaucoma.

§ 886.3920 Aqueous shunt.

(a)
Identification. An aqueous shunt is an implantable device intended to reduce intraocular pressure in the anterior chamber of the eye in patients with neovascular glaucoma or with glaucoma when medical and conventional surgical treatments have failed.(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) “Aqueous Shunts—510(k) Submissions.”