(120 days)
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Adult Cephalic; Trans-rectal (TR); Trans-vaginal (TV); Musculo-skeletal (Conventional); Musculo-skeletal (Superculo-skeletal (Superficial); Cardiac Adult; Cardiac Pediatric; Peripheral Vascular (PV); and Urology (including prostate).
E-CUBE 11 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.
The provided text describes the ALPINION MEDICAL SYSTEMS Co., Ltd. E-CUBE 11 ultrasound system, a Class II medical device, and its substantial equivalence to predicate devices, focusing on technical characteristics and safety standards rather than human performance studies or detailed acceptance criteria with reported device performance.
Therefore, the following information cannot be fully extracted from the provided text:
- A table of acceptance criteria and the reported device performance.
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth for the test set and their qualifications.
- Adjudication method for the test set.
- Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human reader improvement with AI.
- Whether a standalone (algorithm only) performance study was done.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
- The sample size for the training set.
- How the ground truth for the training set was established.
The document states: "The subject of this premarket submission, E-CUBE 11, did not require clinical studies to support substantial equivalence." This indicates that no new clinical efficacy studies evaluating device performance against acceptance criteria were conducted or presented in this 510(k) submission. Instead, the submission relies on demonstrating substantial equivalence to previously cleared predicate devices by comparing technical specifications, imaging modes, and adherence to safety standards.
However, I can provide the acceptance criteria in terms of the standards and features the device conforms to, as reported in the "Summary of Non-Clinical Tests" and "Determination of Substantial Equivalence" sections.
1. Table of Acceptance Criteria (as per conformance to standards and features) and the Reported Device Performance (as stated in the summary of non-clinical tests and comparison to predicates):
| Acceptance Criteria (Conformance to Standards/Features) | Reported Device Performance (Stated Conformance/Equivalence) |
|---|---|
| Safety and Performance Standards | |
| IEC 60601-1 (General Electrical Safety) | Conforms to IEC 60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety |
| IEC 60601-1-2 (EMC Requirements) | Conforms to IEC 60601-1-2, Medical Electrical Equipment – Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests |
| IEC 60601-2-37 (Ultrasonic Medical Equipment Safety) | Conforms to IEC 60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment |
| ISO 10993-1 (Biocompatibility) | Conforms to ISO10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing; Transducer materials and other patient contact materials are biocompatible. |
| ISO 14971 (Risk Management) | Conforms to ISO14971, Application of risk management to medical devices |
| AIUM Medical Ultrasound Safety | Conforms to AIUM Medical Ultrasound Safety |
| NEMA UD 2 (Acoustic Output Measurement) | Conforms to NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment |
| NEMA UD 3 (Real-Time Acoustic Output Display) | Conforms to NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic ultrasound Equipment |
| Quality Management System Measures | |
| Medical Device Risk Management | Applied |
| Requirements Reviews | Applied |
| Design Reviews | Applied |
| Component Verification | Applied |
| Integration Review (System Verification) | Applied |
| Performance Testing (System Verification) | Applied |
| Safety Testing (Compliance Test) | Applied |
| Design Validation | Applied |
| Imaging Modes | (All listed modes are present, aligning with various predicate devices) |
| 2D(B) mode | Supported |
| Harmonic mode (HAR) | Supported |
| M mode | Supported |
| Color M mode | Supported (similar to E-CUBE i7, new compared to E-CUBE 15 & 7) |
| Anatomical M mode | Supported (similar to E-CUBE i7, new compared to E-CUBE 15 & 7) |
| Color Flow Doppler (CF) Mode | Supported |
| Power Doppler (PD) Mode | Supported |
| Directional PD mode | Supported (similar to E-CUBE i7, new compared to E-CUBE 15 & 7) |
| Pulsed Wave Doppler (PWD) Mode | Supported |
| Continuous Wave Doppler (CWD) Mode | Supported |
| High PRF Doppler mode | Supported (similar to E-CUBE i7, new compared to E-CUBE 15 & 7) |
| Tissue Doppler Imaging (TDI) Mode | Supported (similar to E-CUBE i7, new compared to E-CUBE 15 & 7) |
| 3D/4D mode | Supported (similar to E-CUBE 15 & 7, new compared to E-CUBE i7) |
| Imaging Functions (Advanced Features) | (All listed functions are present, aligning with various predicate devices) |
| Xpeed™ | Supported |
| Full SRI™ | Supported |
| Spatial Compounding Image (SCI) | Supported |
| Frequency Compounding image (FCI) | Supported (similar to E-CUBE 15 & 7, new compared to E-CUBE i7) |
| Panoramic | Supported |
| Stress Echo | Supported (similar to E-CUBE 15, new compared to E-CUBE i7 & 7) |
| Cube Strain™ | Supported (similar to E-CUBE 15, new compared to E-CUBE i7 & 7) |
| Live HQ™ | Supported (similar to E-CUBE 15, new compared to E-CUBE i7 & 7) |
| Needle Vision™ / Needle Vision™ Plus | Supported (similar to E-CUBE 15 & i7, new compared to E-CUBE 7) |
| Elastography | Supported (similar to E-CUBE 15, new compared to E-CUBE i7 & 7) |
| Cube view™ | Supported (similar to E-CUBE i7, new compared to E-CUBE 15 & 7) |
| Intended Use / Clinical Applications | (Indications for Use are listed as identical to predicate) |
| Fetal | Supported |
| Abdominal (renal & GYN/pelvic) | Supported |
| Pediatric | Supported |
| Small Organ (breast, testes, thyroid) | Supported |
| Adult Cephalic | Supported |
| Trans-rectal (TR) | Supported |
| Trans-vaginal (TV) | Supported |
| Musculo-skeletal (Conventional) | Supported |
| Musculo-skeletal (Superficial) | Supported |
| Cardiac Adult | Supported |
| Cardiac Pediatric | Supported |
| Peripheral Vessel (PV) | Supported |
| Urology (including prostate) | Supported |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document explicitly states: "The subject of this premarket submission, E-CUBE 11, did not require clinical studies to support substantial equivalence." Therefore, no specific test set sample size or data provenance from clinical studies is provided. The evaluation was based on non-clinical tests ensuring safety and performance compliance with standards and comparison to predicate devices' technical specifications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no clinical studies with expert-established ground truth were conducted for this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no clinical studies were conducted for this submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool, and no MRMC studies were conducted.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable, as the device is a diagnostic ultrasound system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no clinical studies with ground truth establishment were conducted for this submission.
8. The sample size for the training set
Not applicable, as the device is a diagnostic ultrasound system and no specific training set for an AI model is mentioned or relevant to this 510(k) submission.
9. How the ground truth for the training set was established
Not applicable, as no training set for an AI model is mentioned or relevant to this 510(k) submission.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ALPINION MEDICAL SYSTEMS Co., Ltd. % Mr. Donghwan Kim Quality Management Representative (QMR) 1FL and 6FL Verdi Tower, 72, Digital-ro(St) 26-gil(Rd), Guro-gu Seoul 08393 REPUBLIC OF KOREA
September 22, 2016
Re: K161439 Trade/Device Name: E-CUBE 11 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 19, 2016 Received: August 22, 2016
Dear Mr. Kim:
This letter corrects our substantially equivalent letter of September 21, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
Jeff Bollyer
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name E-CUBE 11
Indications for Use (Describe)
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Adult Cephalic; Trans-rectal (TR); Trans-vaginal (TV); Musculo-skeletal (Conventional); Musculo-skeletal (Superficial); Cardiac Adult; Cardiac Pediatric; Peripheral Vascular (PV); and Urology (including prostate).
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
E-CUBE 11 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | N | N | N | N | N | N | N | N | |
| Abdominal | N | N | N | N | N | N | N | N | |
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | N | N | N | |
| Small Organ(breast, testes, thyroid) | N | N | N | N | N | N | N | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | N | N | N | N | N | N | N | ||
| Trans-rectal | N | N | N | N | N | N | N | N | |
| Trans-vaginal | N | N | N | N | N | N | N | N | |
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | N | ||
| Musculo-skeletal(Superficial) | N | N | N | N | N | N | N | ||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | N | N | N | N | N | N | N | N | |
| Cardiac Pediatric | N | N | N | N | N | N | N | N | |
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | N | N | N | N | N | N | N | ||
| Urology (including prostate) | N | N | N | N | N | N | N | N |
N = new indication; P = previously cleared by FDA ; E = added under appendix * Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{4}------------------------------------------------
E-CUBE 11 with SC1-4HS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | Color Doppler | Power Doppler | Tissue Harmonic Imaging | Combined* (Specify) | Other** (Specify) | |||
| Ophthalmic | |||||||||||
| Fetal | P | P | P | P | P | P | P | P | |||
| Abdominal | P | P | P | P | P | P | P | P | |||
| Intra-operative (Specify) | |||||||||||
| Intra-operative (Neuro) | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | |||
| Small Organ(breast, testes, thyroid) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Trans-rectal | |||||||||||
| Trans-vaginal | |||||||||||
| Trans-urethral | |||||||||||
| Trans-esoph. (non-Card.) | |||||||||||
| Musculo-skeletal(Conventional) | |||||||||||
| Musculo-skeletal(Superficial) | |||||||||||
| Intravascular | |||||||||||
| Other (Specify) | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Intravascular (Cardiac) | |||||||||||
| Trans-esoph. (Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Other (Specify) | |||||||||||
| Peripheral vessel | |||||||||||
| Urology (including prostate) | P | P | P | P | P | P | P | P |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{5}------------------------------------------------
E-CUBE 11 with SC1-6H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | ||
| Ophthalmic | ||||||||||
| Fetal | P | P | P | P | P | P | P | |||
| Abdominal | P | P | P | P | P | P | P | |||
| Intra-operative (Specify) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | P | |||
| Small Organ(breast, testes, thyroid) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | N | |||
| Musculo-skeletal(Superficial) | N | N | N | N | N | N | N | |||
| Intravascular | ||||||||||
| Other (Specify) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular (Cardiac) | ||||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Other (Specify) | ||||||||||
| Peripheral vessel | ||||||||||
| Urology (including prostate) | P | P | P | P | P | P | P |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{6}------------------------------------------------
E-CUBE 11 with SC1-4H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | ||
| Ophthalmic | ||||||||||
| Fetal | P | P | P | P | P | P | P | |||
| Abdominal | P | P | P | P | P | P | P | |||
| Intra-operative (Specify) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | P | P | |||
| Small Organ(breast, testes, thyroid) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal(Conventional) | ||||||||||
| Musculo-skeletal(Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Specify) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular (Cardiac) | ||||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Other (Specify) | ||||||||||
| Peripheral vessel | ||||||||||
| Urology (including prostate) | P | P | P | P | P | P | P | P |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{7}------------------------------------------------
E-CUBE 11 L3-12H™ with Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | Color Doppler | Power Doppler | Tissue Harmonic Imaging | Combined* (Specify) | Other** (Specify) | |
| Ophthalmic | |||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ(breast, testes, thyroid) | P | P | P | P | P | N | P | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | N | P | ||
| Musculo-skeletal(Superficial) | P | P | P | P | P | N | P | ||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | N | N | N | N | N | N | N | ||
| Urology (including prostate) |
N = new indication; P = previously cleared by FDA K132687; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{8}------------------------------------------------
E-CUBE 11 with L3-12H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ(breast, testes, thyroid) | P | P | P | P | P | P | P | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | P | P | ||
| Musculo-skeletal(Superficial) | P | P | P | P | P | P | P | ||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | P | P | P | P | P | P | P | ||
| Urology (including prostate) |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{9}------------------------------------------------
E-CUBE 11 with SP1-5 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | |||||||||
| Abdominal | P | P | P | P | P | P | P | ||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | P | P | P | P | P | P | P | ||
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | N | N | N | N | N | N | N | ||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | P | P | P | P | P | P | P | P | |
| Cardiac Pediatric | N | N | N | N | N | N | N | N | |
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) |
N = new indication; P = previously cleared by FDA K132687; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{10}------------------------------------------------
E-CUBE 11 with MP3-8 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | |||||||||
| Abdominal | N | N | N | N | N | N | N | ||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | N | N | N | N | N | N | N | ||
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | N | N | N | N | N | N | N | ||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal( Conventional ) | |||||||||
| Musculo-skeletal( Superficial ) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | N | N | N | N | N | N | N | N | |
| Cardiac Pediatric | N | N | N | N | N | N | N | N | |
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) | |||||||||
| Clinical Application | Mode of Operation | ||||||||
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | P | P | P | P | P | P | P | P | P |
| Abdominal | P | P | P | P | P | P | P | P | P |
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | P | P | P | P | P | P | P | P | P |
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) | P | P | P | P | P | P | P | P | N |
N = new indication; P = previously cleared by FDA; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{11}------------------------------------------------
E-CUBE 11 with SVC1-6H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{12}------------------------------------------------
E-CUBE 11 with EV3-10H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | Color | Power | Tissue | Combined* | Other** | ||||
| Doppler | Doppler | Harmonic | (Specify) | (Specify) | ||||||||
| Imaging | ||||||||||||
| Ophthalmic | ||||||||||||
| Fetal | N | N | N | N | N | N | N | N | ||||
| Abdominal | ||||||||||||
| Intra-operative (Specify) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ | ||||||||||||
| (breast, testes, thyroid) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | P | P | P | P | ||||
| Trans-vaginal | P | P | P | P | P | P | P | P | ||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal | ||||||||||||
| (Conventional) | ||||||||||||
| Musculo-skeletal | ||||||||||||
| (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Urology (including prostate) | P | P | P | P | P | P | P | P |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{13}------------------------------------------------
E-CUBE 11 with EC3-10H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | N | N | N | N | N | N | N | ||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | P | P | P | P | P | P | P | P | |
| Trans-vaginal | P | P | P | P | P | P | P | P | |
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) | P | P | P | P | P | P | P | P |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{14}------------------------------------------------
E-CUBE 11 with VE3-10H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | N | N | N | N | N | N | N | N | |
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | P | P | P | P | P | P | P | P | P |
| Trans-vaginal | P | P | P | P | P | P | P | P | P |
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal( Conventional ) | |||||||||
| Musculo-skeletal( Superficial ) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac Pediatric | |||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) | P | P | P | P | P | P | P | P |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{15}------------------------------------------------
E-CUBE 11 with CW2.0 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal(Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | P | ||||||||
| Cardiac Pediatric | P | ||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{16}------------------------------------------------
E-CUBE 11 with CW5.0 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | PowerDoppler | TissueHarmonicImaging | Combined*(Specify) | Other**(Specify) | |
| Ophthalmic | |||||||||
| Fetal | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ(breast, testes, thyroid) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Card.) | |||||||||
| Musculo-skeletal( Conventional ) | |||||||||
| Musculo-skeletal( Superficial ) | |||||||||
| Intravascular | |||||||||
| Other (Specify) | |||||||||
| Cardiac Adult | P | ||||||||
| Cardiac Pediatric | P | ||||||||
| Intravascular (Cardiac) | |||||||||
| Trans-esoph. (Cardiac) | |||||||||
| Intra-cardiac | |||||||||
| Other (Specify) | |||||||||
| Peripheral vessel | |||||||||
| Urology (including prostate) |
N = new indication; P = previously cleared by FDA K150773; E = added under appendix
- Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
{17}------------------------------------------------
510(k) Summary
In accordance with 21CFR807.92, the following summary of information is provided;
May 17th, 2016 Date
- ALPINION MEDICAL SYSTEMS Co., Ltd. Submitter: Address: 1FL and 6FL, Verdi Tower, 72, Digital-ro(St) 26-gil(Rd), Guro-gu, Seoul, 08393, Republic of Korea
- Primary Contact Donghwan Kim Person Quality Management Representative(QMR) Address: 1FL and 6FL, Verdi Tower, 72, Digital-ro(St) 26-gil(Rd), Guro-gu, Seoul, 08393, Republic of Korea Phone: +82 70 7465 2068 Fax: +82 2 6220 2162 Email: qa ra@alpinion.com
Secondary JULIAN LEE Address: 21312 30th Dr SE Ste 100 Bothell, WA 98021, United States Contact Person Phone: 425 949 1059 Fax: 425 949 4910 Email: julian.lee@alpinionusa.com
Device Trade E-CUBE 11 Name:
Common/ Ultrasonic Pulsed Doppler Imaging System Usual Name:
- Classification System, Imaging, Pulsed Doppler Ultrasonic Names
- Product Code: Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX
K150773 E-CUBE 15 Diagnostic Ultrasound System Primary Predicate Device
- K153424 E-CUBE i7 Diagnostic Ultrasound System Predicate K132687 E-CUBE 7 Diagnostic Ultrasound System Device
E-CUBE 11 product is an ultrasound imaging system for medical diagnosis. Device This innovative system platform provides optimal patient diagnosis workflow Description: with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.
{18}------------------------------------------------
1. Signal Mode:
2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler(CF) Mode, Power Doppler(PD) Mode, Directional PD mode, Pulsed Wave Doppler(PWD) Mode, Continuous Wave Doppler(CWD) Mode, High PRF Doppler mode, Tissue Doppler Imaging(TDI) Mode, 3D/4D mode
2. Combination Mode:
B/Color Doppler, B/PWD, B/Color Doppler/PWD
Acoustic output track:
Track 3
| SC1-4HS | SC1-4H | SC1-6H | L3-12H | SVC1-6H | |
|---|---|---|---|---|---|
| Transducertype | Convex | Convex | Convex | Linear | VolumeConvex |
| Applicablefrequency | 1~4MHz | 1~4MHz | 1~6MHz | 3~12MHz | 1~6MHz |
| Intendedusage | Fetal,Abdominal,Pediatric,Urology(includingprostate) | Fetal,Abdominal,Pediatric,Urology(includingprostate) | Fetal,Abdominal,Pediatric,Musculo-skeletal(Conventional)Musculo-skeletal(Superficial),Urology(includingprostate) | Small Organ(breast,testes, thyroid),Musculo-skeletal(Conventional)Musculo-skeletal(Superficial),Peripheralvessel | Fetal,Abdominal,Pediatric,Urology(includingprostate) |
| Applicablemode | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined/Other |
| ScanningDepth (mm) | 300 | 300 | 300 | 100 | 300 |
| Total numberof element | 166 | 192 | 192 | 192 | 192 |
| Predicate | Previouslycleared(K150773) | Previouslycleared(K150773) | Previouslycleared(K150773) | Previouslycleared(K150773) | Previouslycleared(K150773) |
Types of transducers compatible with the device :
{19}------------------------------------------------
| EV3-10H | VE3-10H | EC3-10H | CW2.0 | CW5.0 | |
|---|---|---|---|---|---|
| Transducertype | Endocavity | VolumeEndocavity | Endocavity | PencilDoppler | PencilDoppler |
| Applicablefrequency | 3~10MHz | 3~10MHz | 3~10MHz | 2.0MHz | 5.0MHz |
| Intendedusage | Fetal,Trans-rectal,Trans-vaginal,Urology(includingprostate) | Fetal,Trans-rectal,Trans-vaginal,Urology(includingprostate) | Fetal,Trans-rectal,Trans-vaginal,Urology(includingprostate) | Cardiac Adult,CardiacPediatric | Cardiac Adult,CardiacPediatric |
| Applicablemode | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined/Other | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | CWD | CWD |
| ScanningDepth (mm) | 140 | 140 | 140 | N/A | N/A |
| Total numberof element | 192 | 192 | 192 | 2(TX1, RX1) | 2(TX1, RX1) |
| Predicate | Previouslycleared(K150773) | Previouslycleared(K150773) | Previouslycleared(K150773) | Previouslycleared(K150773) | Previouslycleared(K150773) |
| L3-12HWD | SP1-5 | MP3-8 | |
|---|---|---|---|
| Transducertype | Linear | SectorPhased | Sector Phased |
| Applicablefrequency | 3~12MHz | 1~5MHz | 3~8MHz |
| Intendedusage | Small Organ(breast,testes, thyroid),Musculo-skeletal(Conventional)Musculo-skeletal(Superficial),Peripheralvessel | Abdominal,Pediatric,Adult Cephalic,Cardiac Adult,CardiacPediatric | Abdominal,Pediatric,Adult Cephalic,Cardiac Adult,Cardiac Pediatric |
| Applicablemode | B/M/PWD/ColorDoppler/PowerDoppler/TissueHarmonicImaging/Combined | B/M/PWD/CWD/Color Doppler/Power Doppler/TissueHarmonicImaging/Combined | B/M/PWD/CWD/Color Doppler/Power Doppler/Tissue HarmonicImaging/Combined |
| ScanningDepth (mm) | 100 | 300 | 300 |
| Total numberof element | 192 | 64 | 64 |
| Predicate | Previouslycleared(K132687) | Previouslycleared(K132687) | SP3-8/SP3-8T(K132687/K153424) |
{20}------------------------------------------------
Substantial Equivalence:
- The device is intended for use by a qualified physician for the evaluation of Indications soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & For Use: GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Adult Cephalic; Trans-rectal (TR); Trans-vaginal (TV); Musculo-skeletal (Conventional); Musculo-skeletal (Superficial); Cardiac Adult; Cardiac Pediatric; Peripheral Vessel (PV); and Urology (including prostate).
| Determination | Comparison table with Predicate devices: |
|---|---|
| of |
| Feature | ProposedE-CUBE 11 | PredicateE-CUBE 15K150773 | PredicateE-CUBE i7K153424 | PredicateE-CUBE 7K132687 |
|---|---|---|---|---|
| Indications for Use | ||||
| - Fetal | √ | √ | √ | √ |
| - Abdominal(renal&GYN/pelvic) | √ | √ | √ | √ |
| - Pediatric | √ | √ | √ | √ |
| - Small Organ(breast, testes, thyroid) | √ | √ | √ | √ |
| - Adult Cephalic | √ | √ | √ | √ |
| - Trans-rectal | √ | √ | √ | √ |
| - Trans-vaginal | √ | √ | √ | √ |
| - Musculo-skeletal(Conventional) | √ | √ | √ | √ |
| - Musculto skeletal(Superficial) | √ | √ | √ | √ |
| - Cardiac (Adult) | √ | √ | √ | √ |
| - Cardiac (Pediatric) | √ | √ | √ | √ |
| - Peripheral Vessel | √ | √ | √ | √ |
| - Urology(including prostate) | √ | √ | √ | √ |
| Imaging modes | ||||
| - 2D(B) mode | √ | √ | √ | √ |
| - Harmonic mode(HAR) | √ | √ | √ | √ |
| - M mode | √ | √ | √ | √ |
| - Color M mode | √ | √ | ||
| - Anatomical M mode | √ | √ | ||
| - Color Flow Doppler(CF) mode | √ | √ | √ | √ |
| - Power Doppler(PD) mode | √ | √ | √ | √ |
| - Directional PD mode | √ | √ | ||
| - Pulsed waveDoppler (PWD)mode | √ | √ | √ | √ |
| - Continuous waveDoppler (CWD)mode | √ | √ | √ | √ |
| - High PRF Dopplermode | √ | √ | ||
| - Tissue Dopplerimaging (TDI) mode | √ | √ | ||
| - 3D/4D mode | √ | √ | √ | |
| Imaging Functions | ||||
| - Xpeed™ | √ | √ | √ | √ |
| - Full SRI™ | √ | √ | √ | √ |
| -Spatial CompoundingImage (SCI) | √ | √ | √ | √ |
| - Frequency | √ | √ | √ | |
| Compounding image(FCI) | ||||
| - Panoramic | √ | √ | √ | √ |
| - Stress Echo | √ | √ | ||
| - Cube StrainTM | √ | √ | ||
| - Live HQTM | √ | √ | ||
| - Needle VisionTM/ Needle VisionTMPlus | √ | √ | ||
| - Elastography | √ | √ | ||
| - Cube viewTM | √ | √ | ||
| Thermal, mechanical and electrical safety | ||||
| - NEMA UD2, UD3 | √ | √ | √ | √ |
| - AIUM MedicalUltrasound Safety | √ | √ | √ | √ |
| - IEC 60601-1 | √ | √ | √ | √ |
| - IEC 60601-1-2 | √ | √ | √ | √ |
| - IEC 60601-2-37 | √ | √ | √ | √ |
ALPINION MEDICAL SYSTEMS Co., Ltd.
{21}------------------------------------------------
Summary of Non-Clinical Tests:
E-CUBE 11 has been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. E-CUBE 11 and its application comply with voluntary standards as detailed in this premarket submission.
- � IEC60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety
- � IEC60601-1-2, Medical Electrical Equipment – Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
- � IEC60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- � ISO10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
- � ISO14971, Application of risk management to medical devices
- � AIUM Medical Ultrasound Safety
- � NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- � NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic ultrasound Equipment
The following quality management system measures were applied to the development of E-CUBE 11:
- Medical Device Risk Management �
- � Requirements Reviews
- � Design Reviews
- � Component Verification
- � Integration Review (System Verification)
- � Performance Testing (System Verification)
- � Safety Testing (Compliance Test)
- � Design Validation
Transducer materials and other patient contact materials are biocompatible.
{22}------------------------------------------------
Summary of Clinical Tests:
The subject of this premarket submission, E-CUBE 11, did not require clinical studies to support substantial equivalence.
Discussion:
E-CUBE 11 was compared with the predicate devices. The subject device is in conformance with applicable safety standards.
Therefore, the differences between E-CUBE 11 and the predicate devices would not affect the safety, effectiveness and essential performance.
- Conclusion: ALPINION MEDICAL SYSTEMS Co., Ltd. considers E-CUBE 11 to be as safe, as effective. Performance, technology and software are substantially equivalent to the predicate devices.
ALPINION MEDICAL SYSTEMS Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA or the requirements will be published in guidance documents.
N/A