K Number
K161414

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-01-19

(241 days)

Product Code
Regulation Number
888.3560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ANATOMIC® Total Knee System is indicated for use in total knee arthroplasty as a result of:

  • Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis;
  • Post-traumatic loss of knee joint configuration and function;
  • Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
  • Revision of previous unsuccessful knee replacement or other procedure.

The ANATOMIC® Total Knee System is for single use only and is intended for implantation with bone cement only.

Device Description

The ANATOMIC® Total Knee System belongs to the category of sliding posterior-stabilized total knee prostheses (PS, posterior-cruciate substituting).

The ANATOMIC® Total Knee System is intended for use as a semi-constrained replacement system. It consists of a femoral implant, a tibial implant and a patellar implant. The tibial implant consists of two components: tibial insert and tibial baseplate. The patellofemoral joint may be preserved or resurfaced with the patellar resurfacing implant.

The femoral implant comes in right and left versions and in various sizes. The tibial baseplate and tibial insert are symmetrical and can be used both for right and left sides. The tibial baseplate comes in various sizes. The tibial insert is available in various sizes and thicknesses.

The nominal size of the insert includes the thickness of the tibial baseplate. The standard keel (supplied in the tibial baseplate packaging) must be mounted and blocked on the implant by the surgeon before implantation.

The components are for cemented used only.

The selection of the appropriate implants can be made by using the recommendations of the surgical technique and by using the x-ray templates and trial implants supplied with the instrumentation.

AI/ML Overview

Please note: The provided document describes the ANATOMIC® Total Knee System, which is a medical device (knee prosthesis). The request asks for information relevant to "AI/ML powered device" or "software" performance criteria. Since this document is for a mechanical implant and not an AI/ML powered device, many of the requested fields (like "AI vs without AI assistance," "number of experts for ground truth," "adjudication method," "sample size for training set," etc.) are not applicable and thus couldn't be extracted from the provided text.

Based on the provided document, here's an analysis of the acceptance criteria and study information for the ANATOMIC® Total Knee System:

1. Table of acceptance criteria and the reported device performance

The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance results in the format typically used for software or AI/ML devices. Instead, it refers to a "substantial equivalence" conclusion based on a comparison to predicate devices and adherence to relevant standards through non-clinical testing.

The comparison table (Table 1 on pages 4-6) lists numerous characteristics and states "YES" under the "Substantial equivalence?" column for each feature, indicating that the device's characteristics are considered equivalent to the predicate devices.

Key areas covered by the non-clinical testing for substantial equivalence:

Acceptance Criteria Category (Derived from Testing Type)Reported Device Performance (as implied by "Substantial Equivalence")
Tibial Baseplate FatigueMet required fatigue strength per FDA guidance.
Insert (post) FatigueMet required fatigue strength per FDA guidance.
WearDemonstrated acceptable wear characteristics.
Range of MotionAchieved intended range of motion (High flexion = 130° max).
Constraints Femoro-TibialDemonstrated appropriate femoro-tibial constraints.
Constraints Patello-FemoralDemonstrated appropriate patello-femoral constraints.
Contact Area/Stress Femoro-TibialExhibited acceptable contact area and stress distribution.
Contact Area/Stress Patello-FemoralExhibited acceptable contact area and stress distribution.
Interlock Strength insert/BaseplateDemonstrated sufficient interlock strength.
UHMWPE Material PropertiesConformed to test standards (ISO 5834-2, ASTM F648 for predicates).
Galvanic Effect on CorrosionAssessed and demonstrated conformity to relevant standard.

2. Sample size used for the test set and the data provenance

The document describes non-clinical laboratory testing, not human patient data. Therefore, concepts like "test set sample size" (in terms of patients) and "data provenance" (country of origin, retrospective/prospective) are not applicable in the context of this device. The tests are performed on physical samples of the device components.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth, in the context of this document, refers to physical measurements and engineering standards, not expert evaluation of clinical data.

4. Adjudication method for the test set

Not applicable. This concept applies to human reader studies or clinical trials, not non-clinical laboratory testing of a mechanical implant.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a mechanical implant, not an AI/ML powered device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a mechanical implant, not an AI/ML powered device.

7. The type of ground truth used

The ground truth for the non-clinical tests is based on established engineering standards and regulatory guidance (e.g., FDA Guidance Class II/Special Controls Guidance Document Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, issued January 16, 2003, and test standards for materials like ISO 5832-4, ASTM F75, ISO 5834-2, ASTM F648).

8. The sample size for the training set

Not applicable. This is a mechanical implant; there is no "training set" in the AI/ML sense. The "training" in manufacturing would refer to design, material selection, and process validation, which are not typically quantified by a "sample size" in this context.

9. How the ground truth for the training set was established
Not applicable, for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 19, 2017

Amplitude Mireille Lemery Director, Quality and Regulatory Affairs 11 Cours Jacques Offenbach Valence, FR 26000 France

Re: K161414 Trade/Device Name: Anatomic Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: December 16, 2016 Received: December 19, 2016

Dear Mireille Lemery:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161414

Device Name ANATOMIC® Total Knee System

Indications for Use (Describe)

The ANATOMIC® Total Knee System is indicated for use in total knee arthroplasty as a result of:

  • Painful, disabling joint disease of the knee resulting from degenerative arthritis,

rheumatoid arthritis or post-traumatic arthritis;

  • Post-traumatic loss of knee joint configuration and function;

  • Moderate varus, valgus or flexion deformity in which the ligamentous structures can

be returned to adequate function and stability;

  • Revision of previous unsuccessful knee replacement or other procedure.

The ANATOMIC® Total Knee System is for single use only and is intended for implantation with bone cement only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Amplitude. The logo features a stylized letter "A" in blue, with a small circle attached to the upper left side of the "A". Below the "A" is the word "AMPLITUDE" in black, with a small degree symbol to the right of the word.

510(k) Summary of Safety and Effectiveness Information Traditional 510(k) Premarket - ANATOMIC® Total Knee System

Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92

Date prepared: December 16th, 2016

Assigned 510(k) number:K161414
Company:AMPLITUDE11, cours Jacques Offenbach, Zone Mozart 226000 Valence, FRANCEPhone: +33 4 75 81 34 95Fax: +33 4 75 41 41 78
Contact:Mireille LEMERY
Trade/Proprietary Name:ANATOMIC® Total Knee System
Classification Name:Knee Prosthesis
Classification/Product Code:21 CFR 888.3560, Class II, Product Code JWH
Predicate Devices:Scorpio® NRG® Knee, Howmedica Osteonics CorpK030978, K042343, K071991Freedom® Total Knee System, Maxx OrthopedicsInc. K082019, K090411

Service Client France

Porte du Grand Lyon 01700 Neyron - France Tél : +33 (0)4 37 85 19 19 Fax : +33 (0)4 37 85 19 18

1. Device Description

The ANATOMIC® Total Knee System belongs to the category of sliding posterior-stabilized total knee prostheses (PS, posterior-cruciate substituting).

The ANATOMIC® Total Knee System is intended for use as a semi-constrained replacement system. It consists of a femoral implant, a tibial implant and a patellar implant. The tibial implant consists of two components: tibial insert and tibial baseplate. The patellofemoral joint may be preserved or resurfaced with the patellar resurfacing implant.

Export Customer Service

11, cours Jacques Offenbach Zone Mozart 2 26000 Valence - France Tél : +33 (0)4 75 41 87 41 Fax : +33 (0)4 75 41 87 42

E-mail : amplitude@amplitude-ortho.com

{4}------------------------------------------------

The femoral implant comes in right and left versions and in various sizes. The tibial baseplate and tibial insert are symmetrical and can be used both for right and left sides. The tibial baseplate comes in various sizes. The tibial insert is available in various sizes and thicknesses.

The nominal size of the insert includes the thickness of the tibial baseplate. The standard keel (supplied in the tibial baseplate packaging) must be mounted and blocked on the implant by the surgeon before implantation.

The components are for cemented used only.

The selection of the appropriate implants can be made by using the recommendations of the surgical technique and by using the x-ray templates and trial implants supplied with the instrumentation.

2. Intended Use

The ANATOMIC® Total Knee System is intended for replacement of the knee joint to reduce pain and restore knee function in comparison with preoperative status.

3. Indications for use

The ANATOMIC® Total Knee System is indicated for use in total knee arthroplasty as a result of:

  • -Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis;
  • -Post-traumatic loss of knee joint configuration and function;
  • Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
  • Revision of previous unsuccessful knee replacement or other procedure. -

The ANATOMIC® Total Knee System is for single use only and is intended for implantation with bone cement only.

Summary of technological characteristics 4.

Table 1: Main features comparison
---------------------------------------
Service Client FrancePorte du Grand Lyon01700 Neyron - FranceTél: +33 (0)4 37 85 19 19Fax: +33 (0)4 37 85 19 18MATERIALSFemoralimplantANATOMIC® TotalKnee SystemScorpio® NRG®KneeFreedom® TotalKnee SystemSubstantialequivalence?
Export Customer Service11, cours Jacques OffenbachZone Mozart 226000 Valence - FranceTél: +33 (0)4 75 41 87 41Fax: +33 (0)4 75 41 87 42TibialbaseplateCoCrISO 5832-4CoCrASTM F75CoCrASTM F75YES
TibialinsertCoCrISO 5832-4CoCrASTM F75CoCrASTM F75YES
TibialinsertUHMW-PEISO 5834-2UHMW-PEASTM F648UHMW-PEASTM F648YES
PatellarimplantUHMW-PEISO 5834-2UHMW-PEASTM F648UHMW-PEASTM F648YES
EmployedtechnologyPosterior StabilizedPosterior StabilizedPosterior StabilizedYES
Anchoring systemCementedCementedCementedYES
Intended UseTotal KneeReplacementTotal KneeReplacementTotal KneeReplacementYES
SterilizationmethodGamma sterilizationGamma sterilizationGamma sterilizationYES
Maximum flexionHigh flexion = 130°maxHigh flexion = 130°maxHigh flexion = 155°maxYES
Femoral component features
Model9 sizes9 sizes8 sizesYES
DimensionsAP from 49.83 to72.86 mmML 56 from to 76.7mmAP from 51 to 75mmML from 57 to 82mmAP from 51 to 74mmML 54 from to 78mmYES
Anchoring systemCage's surfaceshave negativemacrostructuregrooves type andcones with a depthof 0.5 mm as wellas positivemacrostructurespyramid type with aheight of 0.5 mmThe surface incontact with bonecement is shot-blastedWafle structures witha depth of 0.6 mmThe surface incontact with bonecement is shot-blastedGroove depth 0.8mmThe surface incontact with bonecement is shot-blastedYES
Radius ofcurvatureSingle radius ofcurvature from 0°to 100° flexionSingle radius ofcurvature from 0° to95° flexionSingle radius ofcurvature from 15°to 120°YES
Insert component features
Model9 sizes x 6thickness5 sizes x 7 thickness8 sizesYES
ThicknessThickness: 10, 12,14, 16, 18 and 20mm with Tibial TrayMinimumPolyethyleneThickness: 6.8without Tibial TrayThickness: 8, 10, 12,15, 18, 21, 24 mmwith Tibial TrayMinimumPolyethyleneThickness: 6 withoutTibial TrayN/KYES
Insert minimumthickness (mm)6.9 +/-0.1 mm6 mm, toleranceunknownN/KYES
Service Client FrancePorte du Grand Lyon01700 Neyron - FranceTél : +33 (0)4 37 85 19 19Fax : +33 (0)4 37 85 19 18Export Customer Service11, cours Jacques OffenbachZone Mozart 226000 Valence - FranceTél : +33 (0)4 75 41 87 41Fax : +33 (0)4 75 41 87 42The anchoringsystem of theinsert onto thetibial baseplatePeripheral locks atsix places (twoanterior clips,posteromedial,posterolateral,lateral and medialdovetails)2 posteriorundercuts1 anterior wirePeripheral lipPeripheral locks atfive places (anterior,posteromedial,posterolateral,lateral and mediallocking)YES
Tibial component features
Model9 sizes8 sizes8 sizesYES
DimensionsAP from 39.09 to57.31 mmAP from 40 to 58mmAP from 40 to 54mmYES

E-mail : amplitude@amplitude-ortho.com

{5}------------------------------------------------

E-mail : amplitude@amplitude-ortho.com

Internet : www.amplitude-ortho.com

{6}------------------------------------------------

ML from 60 to 88mmML from 61 to 88mmML from 59 to 78mm
Tibial traythickness (mm)3 mm2.5 mm3.3 mmYES
Stem length41 mm30 to 40 mmN/KYES
Surface aspectArea in contact withbone cement: shot-blastedSurface in contactwith tibial insert:mirror polishedShot-blastedShot-blastedYES
Patellar component features
Model4 diameters6 models4 diametersYES
Dimension30, 33, 36 and 39mm30, 32, 34, 36 and38 mm25, 28, 31 and 34mmYES
Thickness8 mm8 to 10 mmUnknownYES
Patella fixationResurfacing patellacemented withthree pegsResurfacing patellaare cemented withthree pegs andgrooves for cement.All-poly resurfacingpatella cementedwith three pegsYES
Patella shapeSpherical domeshape forresurfacing patellaSpherical domeshape for UniversalPatellaSpherical domeshape forresurfacing patellaYES
ManufacturerAmplitudeHowmedicaOsteonics Corp.Maxx Orthopedics,Inc.

The product subject of this premarket notification is substantially equivalent in design and functionality to the Scorpio® NRG® Knee (K030978) and Freedom® Total Knee System (K082019).

5. Non clinical testing

Functional testing was conducted in compliance with the FDA Guidance Class II/Special Controls Guidance Document Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, issued January 16, 2003.

Specifically, the following non-clinical laboratory tests were performed to determine substantial equivalence:

  • Tibial Baseplate fatique test -

  • Insert (post) fatigue test -

  • Wear test -

  • Range of Motion -

  • Constraints Femoro-Tibial -

  • Constraints Patello-Femoral -

  • Contact Area/ Stress Femoro-Tibial -

  • Contact Area/ Stress Patello-Femoral -

    • Interlock Strength insert/ Baseplate

E-mail :

Service Client France Porte du Grand Lyon

01700 Neyron - France Tél : +33 (0)4 37 85 19 19 Fax : +33 (0)4 37 85 19 18

Export Customer Service

11, cours Jacques Offenbach Zone Mozart 2 26000 Valence - France Tél : +33 (0)4 75 41 87 41 Fax : +33 (0)4 75 41 87 42

{7}------------------------------------------------

The UHMWPE material properties have been characterized according to test standards to demonstrate the conformity of the material.

The galvanic effect on the corrosion of the assembly between the tibial baseplate and the standard keel has been assessed according to relevant standard to demonstrate the conformity of the assembly.

MRI compatibility has not been evaluated for the ANATOMIC® Total Knee System.

The results of these tests demonstrate the equivalence between the ANATOMIC® Total Knee System and the Scorpio® NRG® Knee (K030978) and Freedom® Total Knee System (K082019).

6. Substantial Equivalence Conclusion

The proposed ANATOMIC® Total Knee System is believed to be substantially equivalent to the predicate devices with the same intended use and the same technological characteristics.

Service Client France

Porte du Grand Lyon 01700 Neyron - France Tél : +33 (0)4 37 85 19 19 Fax : +33 (0)4 37 85 19 18

Export Customer Service

11, cours Jacques Offenbach Zone Mozart 2 26000 Valence - France Tél : +33 (0)4 75 41 87 41 Fax : +33 (0)4 75 41 87 42

E-mail : amplitude@amplitude-ortho.com

Internet : www.amplitude-ortho.com

N/A