K Number
K161362

Validate with FDA (Live)

Date Cleared
2016-06-14

(29 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature. This device should not be used in coronary vasculature or neurovasculature.

Device Description

The ASAHI Corsair Armet consists of a distal tip, a shaft tube that is inserted into vasculature, a protector, and a connector for catheter control and infusion of contrast media. The device has a hydrophilic coating on the outer surface of distal tip and the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair Armet has a tapered shape. PTFE is applied to the inner lumen of the catheter for the purposes of a smooth transition and exchange of guidewires. The primary change in this Special 510(k) involves a minor change in the shape of the distal tip.

AI/ML Overview

The provided document is a 510(k) Summary for the ASAHI Corsair Armet. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving efficacy or safety through large-scale clinical trials. Therefore, the information typically requested in your prompt regarding acceptance criteria, study details, and expert involvement for diagnostic algorithm performance is not present in this document.

The document describes non-clinical (bench) testing performed to confirm the device's performance against pre-defined acceptance criteria, primarily related to its physical characteristics and functionality.

Here's a breakdown of the available information based on your request, with an emphasis on what is not provided due to the nature of this regulatory submission:

1. A table of acceptance criteria and the reported device performance

The document lists the types of non-clinical tests performed, but does not provide a specific table detailing the quantitative acceptance criteria for each test or the exact numerical performance values. It generally states that the device "met all acceptance criteria" and "performed similarly to the predicate devices."

Test CategoryAcceptance Criteria (Not explicitly quantified in document)Reported Device Performance (Summary statement)
Appearance/DimensionsImplied conformance to specificationsMet all acceptance criteria
Corrosion ResistanceImplied resistance to corrosionMet all acceptance criteria
Force at BreakImplied strength/durabilityMet all acceptance criteria
Liquid Leakage under PressureImplied absence of leakageMet all acceptance criteria
Air LeakageImplied absence of leakageMet all acceptance criteria
Leak and Damage under High Static PressureImplied resistance to damage/leakageMet all acceptance criteria
Radio-DetectabilityImplied sufficient visibility under imagingMet all acceptance criteria
Torque TransmissionImplied efficient torque transferMet all acceptance criteria
Slide DurabilityImplied resistance to wear during slidingMet all acceptance criteria
Kink ResistanceImplied resistance to kinkingMet all acceptance criteria
Torque DurabilityImplied maintained torque performance over timeMet all acceptance criteria

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for the non-clinical tests. These are typically bench tests conducted on a sufficient number of units to demonstrate consistency, but not usually reported with statistical sample sizes in a 510(k) summary.
  • Data Provenance: The tests were "non clinical laboratory testing." No country of origin for the data is specified, but the applicant is Asahi Intecc Co., Ltd. from Japan, and the testing likely occurred internally or at a contracted lab.
  • Retrospective or Prospective: These were bench tests, so the concept of retrospective or prospective (which applies to clinical data) is not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable to this type of non-clinical device testing. "Ground truth" established by human experts (like radiologists) pertains to diagnostic algorithms, not to the physical performance of a catheter. The ground truth for these tests is based on engineering specifications and measurement standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as it relates to expert consensus on clinical or diagnostic data. For bench testing, results are typically analyzed against predefined, objective, and measurable acceptance criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This device is a percutaneous catheter, not an AI-powered diagnostic tool. Therefore, MRMC studies and the effect of AI assistance on human readers are irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This device is hardware (a catheter), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For non-clinical testing, the "ground truth" implicitly refers to the engineering specifications and performance standards. For example, for "Force at Break," the ground truth would be a defined minimum force the device must withstand, determined by engineering principles and safety margins. For "Dimensions," the ground truth is the specified dimensional tolerances.

8. The sample size for the training set

This information is not applicable. There is no "training set" as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

This information is not applicable, as there is no training set for this type of medical device submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a human figure, with three profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June14, 2016

Asahi Intecc Co., Ltd. % Candace Cederman Senior Regulatory Affairs Consultant Cardiomed Device Consultants, LLC 5523 Research Park Drive, Suite 205 Baltimore, MD 21228

Re: K161362

Trade/Device Name: Asahi Corsair Armet Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: May 13, 2016 Received: May 16, 2016

Dear Ms. Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161362

Device Name ASAHI Corsair Armet

Indications for Use (Describe)

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature. This device should not be used in coronary vasculature or neurovasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

(as required by 21 CFR 807.92)

HI INTECC CO.LI 1703 Wakita-cho, Moriyama-ku, Nagoya, Aichi 463-0024 Japan Tel. +81-52-768-1211 Fax. +81-52-768-1221 Branch offices: Tokyo, Nagoya, Osaka, Hong Kong, Amsterdam, Singapore, Beijing Research Facilities and Factories: Osaka, Seto, Thailand, Hanoi

ASAHI Corsair Armet

510(k) _______________________________________________________________________________________________________________________________________________________________________

DATE PREPARED:May 13, 2016
APPLICANTAsahi Intecc Co., Ltd.1703 Wakita-cho, Moriyama-kuNagoya, Aichi 463-0024, Japan
CONTACTCarroll Councilman, Sr. RA ManagerAsahi Intecc USA, Inc.2500 Red Hill Avenue, Suite 210Santa Ana, CA 92705Tel: (949) 756-8252, FAX: (949) 756-8165e-mail: ASAHI.ra-fda@ASAHI-intecc.com
TRADE NAME:ASAHI Corsair Armet
DEVICE CLASSIFICATION:Class 2 per 21 CFR §870.1250
CLASSIFICATION NAME:Percutaneous Catheter
PRODUCT CODEDQY – Catheter, Percutaneous
PREDICATE DEVICES:ASAHI Corsair Armet (K152249)

Intended Use/Indications for Use

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of quide wires in the peripheral vasculature, and can be used to exchange one quide wire for another.

The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.

This device should not be used in coronary vasculature or neurovasculature.

Device Description:

The ASAHI Corsair Armet consists of a distal tip, a shaft tube that is inserted into vasculature, a protector, and a connector for catheter control and infusion of contrast media. The device has a hydrophilic coating on the

{4}------------------------------------------------

outer surface of distal tip and the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair Armet has a tapered shape. PTFE is applied to the inner lumen of the catheter for the purposes of a smooth transition and exchange of guidewires.

The primary change in this Special 510(k) involves a minor change in the shape of the distal tip.

Comparison with Predicate Devices:

Comparisons of the modified ASAHI Corsair Armet to its predicate device show that the technological characteristics of the Subject device such as the product performance, intended use/indications, components, materials, sterilization method, shelf life, manufacturing process, and operating principle are identical to the currently marketed predicate devices. There are only minor dimensional variations between the Subject and predicate device.

Name of DeviceASAHI Corsair ArmetASAHI Corsair Armet
510(k)Current ApplicationK152249
Intended Use and IndicationsThe ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another.The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.This device should not be used in coronary vasculature or neurovasculature.
Target Body LocationPeripheral
Hydrophilic coatingYes
Effective Length600-1500 cm
Nominal Outer DiameterDistal : 0.75mmProximal : 0.83 mm
Catheter Shaft MaterialPolyamide-elastomer
Distal Tip Length1.2mm2.0mm
Single UseYes
SterilizationProvided sterile via Ethylene Oxide to SAL10-6
Shelf Life3 Years

{5}------------------------------------------------

Non Clinical testing / Performance Data:

Confirmatory non clinical laboratory testing was performed on the modified ASAHI Corsair Armet to determine substantial equivalence.

The following testing/assessments were performed:

  • Appearance/Dimensions
  • Corrosion Resistance ●
  • Force at Break
  • Liquid Leakage under Pressure ●
  • Air Leakage ●
  • Leak and Damage under High Static Pressure ●
  • Radio-Detectability ●
  • Torque Transmission ●
  • Slide Durability ●
  • Kink Resistance ●
  • . Torque Durability

The in vitro bench tests demonstrated that the modified ASAHI Corsair Armet met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended, and is substantially equivalent to the predicate devices.

BIOCOMPATIBILITY:

The modified ASAHI Corsair Armet was compared to the predicate device. Based on the identical materials and manufacturing process used in the subject device to its predicate, the biocompatibility of the modified ASAHI Corsair Armet was verified to be the same as those of the predicate.

CONCLUSION:

The modified ASAHI Corsair Armet has identical intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended.

Therefore, the modified ASAHI Corsair Armet is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).