K161332

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-04-05

(328 days)

Product Code
Regulation Number
888.3010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NILE Alternative Fixation and NILE Proximal Fixation Spinal Systems are temporary implants for use in orthopedic surgery. The systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

  1. Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring techniques;

  2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, and spondylolisthesis:

  3. Spinal degenerative surgery, as an adjunct to spinal fusions.

The NILE Alternative Fixation and NILE Proximal Fixation Spinal Systems may also be used in conjunction with other medical implants made of similar metals whenever 'wiring' may help secure the attachment of other implants.

Device Description

The NILE Proximal Fixation Spinal System is a multiple component spinal fixation system intended for application to the thoracic, lumbar and sacral spine. The band is manufactured from polyethylene terephthalate (PET) and the connectors are made from titanium alloy in accordance with ASTM F136.

Function: The NILE Proximal Fixation Spinal System is designed to be used in conjunction with spinal rod constructs for attachment to the posterior vertebral structures.

The purpose of this submission is to add bands and connectors to the Nile system.

AI/ML Overview

This document (K161332) is a 510(k) premarket notification for the NILE Proximal Fixation Spinal System. It asserts substantial equivalence to a predicate device, the K2M NILE Alternative Fixation Spinal System (K143350). The FDA has reviewed and determined the device is substantially equivalent for its stated indications for use.

Here's an analysis of the provided text to answer your questions regarding acceptance criteria and performance studies:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative "acceptance criteria" for the NILE Proximal Fixation Spinal System in the traditional sense (e.g., minimum accuracy, sensitivity, or specificity values). Instead, the performance evaluation is based on a comparative study against predicate devices using recognized ASTM standards. The reported performance is a qualitative statement of equivalence or superiority.

Acceptance Criteria (Implied)Reported Device Performance
Mechanical performance in static and dynamic compressionPerformed equally to or better than predicate devices
Mechanical performance in static torsionPerformed equally to or better than predicate devices
Mechanical performance in static and dynamic tensionPerformed equally to or better than predicate devices
Mechanical performance in axial slipPerformed equally to or better than predicate devices
Bacterial endotoxin limit<20 EU/Device maintained (from ANSI/AAMI ST72:2011)

2. Sample sized used for the test set and the data provenance

The document does not specify the sample sizes (number of devices or test repetitions) used for the mechanical performance testing conducted in accordance with ASTM F1717 and ASTM F1798. It also does not mention the data provenance (e.g., country of origin, retrospective or prospective) as the tests are non-clinical, mechanical device performance assessments.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to the provided document. The "ground truth" for mechanical testing is established by the specified ASTM test methods and objective measurements of force, displacement, etc., not by expert clinical opinion or interpretation.

4. Adjudication method for the test set

This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used for clinical studies involving expert review of medical images or patient outcomes, not for non-clinical mechanical performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study described. This document pertains to a spinal fixation system, which is a physical implant, not an AI-assisted diagnostic or interpretative device. Therefore, the concept of "human readers improve with AI" is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a mechanical implant, not an algorithm, so standalone AI performance is not relevant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical performance evaluation, the "ground truth" is based on objective, quantifiable physical measurements as defined by the ASTM standards (e.g., load-displacement curves, failure loads, pull-out forces). There is no "expert consensus," "pathology," or "outcomes data" ground truth utilized in this context.

8. The sample size for the training set

This information is not applicable. As the device is a mechanical implant and not an AI/machine learning system, there is no "training set."

9. How the ground truth for the training set was established

This information is not applicable due to the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 5, 2017

K2M, Inc. Ms. Nancy Giezen Manager, Regulatory Affairs 600 Hope Parkway Southeast Leesburg, Virginia 20175

Re: K161332

Trade/Device Name: NILE Proximal Fixation Spinal System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: OWI Dated: March 6, 2017 Received: March 7, 2017

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161332

Device Name NILE Proximal Fixation Spinal System

Indications for Use (Describe)

The NILE Alternative Fixation and NILE Proximal Fixation Spinal Systems are temporary implants for use in orthopedic surgery. The systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

  1. Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring techniques;

  2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, and spondylolisthesis:

  3. Spinal degenerative surgery, as an adjunct to spinal fusions.

The NILE Alternative Fixation and NILE Proximal Fixation Spinal Systems may also be used in conjunction with other medical implants made of similar metals whenever 'wiring' may help secure the attachment of other implants.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY NILE Proximal Fixation Spinal System

Submitter

K2M, Inc. 600 Hope Parkway SE Leesburg, VA 20175

Contact Person: Nancy Giezen Telephone: 571-919-2000 Date Prepared: 4/3/2017

Classification

Trade Name:NILE Proximal Fixation Spinal System
Common Name:Bone Fixation Cerclage
Regulatory Class:Class II

Classification Name(s): Bone Fixation Cerclage (21 CFR 888.3010, Product Code OWI)

Predicate Device(s)

Primary Predicate: K2M NILE Alternative Fixation Spinal System (K143350) Reference Devices: K2M Range/ Denali/ Mesa Spinal System (K153031) K2M Everest Spinal System (K151727)

Device Description

The NILE Proximal Fixation Spinal System is a multiple component spinal fixation system intended for application to the thoracic, lumbar and sacral spine. The band is manufactured from polyethylene terephthalate (PET) and the connectors are made from titanium alloy in accordance with ASTM F136.

Function: The NILE Proximal Fixation Spinal System is designed to be used in conjunction with spinal rod constructs for attachment to the posterior vertebral structures.

The purpose of this submission is to add bands and connectors to the Nile system.

Indications For Use

The NILE Alternative Fixation and NILE Proximal Fixation Spinal Systems are temporary implants for use in orthopedic surgery. The systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

{4}------------------------------------------------

  1. Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring techniques:

  2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, kyphosis, and spondylolisthesis;

  3. Spinal degenerative surgery, as an adjunct to spinal fusions.

The NILE Alternative Fixation and NILE Proximal Fixation Spinal Systems may also be used in conjunction with other medical implants made of similar metals whenever 'wiring' may help secure the attachment of other implants.

Technological Comparison to Predicate(s)

The NILE Proximal Fixation Spinal System implants were compared to predicate devices and the design features, materials and intended uses were found to be substantially equivalent to these systems.

Non-clinical Performance Evaluation

The NILE Proximal Fixation Spinal System implants were tested in accordance with ASTM F1717 and compared to predicate devices. The NILE Proximal Fixation system implants performed equally to or better than these systems in static and dynamic compression, static torsion, static and dynamic tension and axial slip per ASTM F1798. Bacterial endotoxin testing (BET) as specified in ANSI/AAMI ST72:2011 was also performed and will be routinely repeated to ensure that the Endotoxin limit of <20EU/Device is maintained.

Conclusion

There are no significant differences between the NILE Proximal Fixation Spinal System implants and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and performance.

§ 888.3010 Bone fixation cerclage.

(a)
Identification. A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.(b)
Classification. Class II.