K Number
K161220

Validate with FDA (Live)

Date Cleared
2016-08-02

(95 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARIES® Flu A/B & RSV Assay is a polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) nucleic acid in nasopharyngeal swab (NPS) specimens from patients with signs and symptoms of respiratory tract infection in conjunction with clinical and laboratory findings. The test is intended for use as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV in humans and is not intended to detect Influenza C.

Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, X-ray findings) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.

Performance characteristics for influenza A were established during the 2015-2016 influenza seasons when influenza A/H3N2 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

The ARIES® Flu A/B & RSV Assay is indicated for use with the ARIES® Systems.

Device Description

The ARIES® Flu A/B & RSV Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system that will consist of an ARIES® System with its included software, an assay-specific cassette, and an assay-specific protocol file. The ARIES® Flu A/B & RSV Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® Flu A/B & RSV Assay cassette directly detects and differentiates influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) nucleic acid in nasopharyngeal swabs (NPS) specimens from patients with signs and symptoms of respiratory tract infection in conjunction with clinical and laboratory findings. Specifically, the ARIES® Flu A/B & RSV Assay cassette detects the matrix protein genes of influenza A and influenza B viruses, and the fusion gene of RSV and a RNA Sample Processing Control.

The specimen is lysed and nucleic acid is extracted using an ARIES® System. An extractable sample processing control (SPC) target is present in the ARIES® Flu A/B & RSV assay cassette and is processed with the specimen. The SPC controls for specimen lysis, for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify proper specimen lysis and nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument. The Tm value of the SPC is used as a reference for determining the target Tm.

The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® Flu A/B & RSV Assay lyophilized PCR reagents in the PCR tube that contain primer pairs specific to influenza A, influenza B, RSV, and the SPC sequence. The specific primer pairs are labeled with distinct fluorophore labels. PCR amplification is performed and assay fluorescence is monitored on an ARIES® System. Incorporation of the quencherlabeled nucleotide causes a decrease in assay fluorescence. Following amplification, the reaction is slowly heated and fluorescence is monitored. The strands of the amplification products will separate at a specific melting temperature (Tm) that is determined by an increase in fluorescence as the strands are separated. The instrument fluorescence output is analyzed and test results are determined using the ARIES® Flu A/B & RSV Assay protocol file. A printed results report is generated.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ARIES® Flu A/B & RSV Assay, based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

Clinical Performance Acceptance Criteria:

AnalyteAcceptance Criteria (PPA)Acceptance Criteria (NPA)
Influenza A90% with a lower bound 95% CI of at least 80%90% with a lower bound 95% CI of at least 90%
Influenza B90% with a lower bound 95% CI of at least 80%90% with a lower bound 95% CI of at least 90%
RSV90% with a lower bound 95% CI of at least 80%90% with a lower bound 95% CI of at least 90%

Reported Device Performance (Clinical Study):

AnalytePositive Percent Agreement (PPA)95% CI (PPA)Negative Percent Agreement (NPA)95% CI (NPA)
Influenza A95.8%93.0% - 97.8%98.4%97.8% - 98.9%
Influenza B93.8%82.8% - 98.7%99.4%99.0% - 99.7%
RSV97.1%94.4% - 98.7%98.4%97.7% - 98.9%

Additional Influenza B Performance (Supplemented Study):

AnalytePositive Percent Agreement (PPA)95% CI (PPA)
Influenza B100%91.2% - 100%

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Prospective Clinical Study: 2479 eligible unique nasopharyngeal swab (NPS) specimens.
  • Sample Size for Supplemented Influenza B Study: 40 pre-selected Influenza B positive specimens + 40 unique negative clinical specimens (total 80).
  • Data Provenance: Retrospective and prospective.
    • Prospective Data: Specimens collected during the 2014/2015 (Jan 18, 2015 to Mar 20, 2015) and 2015/2016 (Nov 02, 2015 to Feb 29, 2016) flu seasons.
    • Geographical Origin: Collected at 3 clinical sites (2014/2015) and 4 clinical sites (2015/2016) in the United States and Canada.
    • Retrospective/Pre-selected Data: 40 banked (pre-selected) Influenza B positive specimens collected at a single clinical laboratory (collection site) in Canada for the supplemental study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number or qualifications of "experts" used to establish ground truth in the traditional sense (e.g., radiologists interpreting images).

Instead, the ground truth for the clinical study was established using:

  • An FDA-cleared molecular comparator assay (primary reference standard).
  • Bi-directional sequencing for discordant results.

This approach means the "ground truth" relies on the performance of another validated diagnostic device and a molecular sequencing technique, rather than human expert consensus on clinical presentation or imaging.


4. Adjudication Method for the Test Set

  • Primary Comparison: All 2479 eligible clinical specimens were initially tested by an FDA-cleared molecular comparator and the ARIES® Flu A/B & RSV Assay.
  • Discordant Analysis: For specimens where the ARIES® Flu A/B & RSV Assay results differed from the comparator assay result, they were further assessed by bi-directional sequencing using analytically validated primers that targeted genomic regions distinct from the ARIES® Flu A/B & RSV Assay.
  • Reporting: Results from discordant testing analysis were not included in the initial calculation of Positive Percent Agreement and Negative Percent Agreement for each target. These discordant results were noted as footnotes in the performance evaluation tables (e.g., "False Negative by bi-directional sequencing").

This indicates a form of secondary adjudication using an independent, highly specific molecular method to resolve discrepancies between the new device and the predicate.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, this was not a Multi-Reader Multi-Case (MRMC) comparative effectiveness study involving human readers. This study evaluated the standalone performance of a molecular diagnostic assay against a comparator assay and sequencing, not the improvement of human readers with AI assistance.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was conducted. The ARIES® Flu A/B & RSV Assay is an automated molecular diagnostic test system. The clinical performance evaluation directly tested the device's output against a molecular comparator and sequencing, without human interpretation of the assay's raw output determining the final diagnostic call. The assay itself provides a qualitative "Positive" or "Negative" result.


7. The Type of Ground Truth Used

The ground truth for the test set was an combination of:

  • Molecular Comparator Assay Results: An FDA-cleared molecular test used as the primary reference standard.
  • Bi-directional Sequencing: Used as a confirmatory "tie-breaker" or definitive reference for specimens yielding discordant results between the ARIES® assay and the primary comparator.

8. The Sample Size for the Training Set

The document does not directly specify a "training set" sample size in the context of machine learning model development. For in vitro diagnostic devices like the ARIES® Flu A/B & RSV Assay, "training" typically refers to the process of developing and optimizing the assay's chemical reagents, amplification protocols, and the algorithms for interpreting raw signal (e.g., Ct values, Tm values) into a qualitative result. This development process uses numerous laboratory experiments with contrived samples and potentially some early clinical samples, but these are generally not referred to as a "training set" in the same way as in AI/ML literature.

The document describes extensive analytical performance studies (e.g., Limit of Detection, Inclusivity, Specificity, Precision, Stability) that are crucial for defining the assay's operating characteristics and establishing the cut-offs and parameters that lead to accurate results. These analytical studies utilize:

  • Precision/Reproducibility: 675 replicates (initial precision) and 1260 replicates (site-to-site reproducibility) using contrived samples (viral cultures at specific concentrations diluted in simulated nasal matrix).
  • Limit of Detection: At least 20 replicates for each of 7 viral strains.
  • Inclusivity: Triplicate testing for each of 34 different viral strains (some required additional testing).
  • Interfering Substances: Triplicate testing for viral targets with 18 different interfering substances.
  • Cross-Reactivity: Triplicate testing for 32 microorganisms (some required additional testing).
  • Microbial Interference/Co-infection: Triplicate testing for various pathogen combinations.

These extensive studies with known concentrations and types of pathogens and interferents essentially "train" and validate the design and interpretation rules of the assay.


9. How the Ground Truth for the Training Set was Established

Again, the concept of a "training set" with ground truth in the AI/ML sense is not explicitly detailed. However, for the analytical studies (which inform the assay's design and operating parameters), the "ground truth" was established based on:

  • Known Concentrations of Characterized Viral Stocks/Strains: Viral cultures were quantified and diluted to precise concentrations (e.g., TCID50/mL) to create samples with a known positive status and concentration.
  • Known Negative Samples: Simulated nasal matrix (SNM) or Universal Transport Medium (UTM) were used as known negative samples.
  • Known Interfering/Cross-Reactive Agents: Specific microorganisms or substances were added at defined concentrations to evaluate their impact.

Essentially, the ground truth for these analytical studies was derived from laboratory-controlled conditions where the presence, identity, and concentration of analytes were precisely known.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol with three human profiles facing right, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 2, 2016

Luminex Corporation Wendy Ricker Senior Regulatory Affairs Associate 12212 Technology Blvd. Austin TX 78727

Re: K161220 Trade/Device Name: ARIES® Flu A/B & RSV Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: II Product Code: OCC, OOI, OZE Dated: April 29, 2016 Received: April 29, 2016

Dear Ms. Ricker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tamara V. Feldblyum -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161220

Device Name

ARIES® Flu A/B & RSV Assay

Indications for Use (Describe)

The ARIES® Flu A/B & RSV Assay is a polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) nucleic acid in nasopharyngeal swab (NPS) specimens from patients with signs and symptoms of respiratory tract infection in conjunction with clinical and laboratory findings. The test is intended for use as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV in humans and is not intended to detect Influenza C.

Neqative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, X-ray findings) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.

Performance characteristics for influenza A were established during the 2015-2016 influenza seasons when influenza A/H3N2 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

The ARIES® Flu A/B & RSV Assay is indicated for use with the ARIES® Systems.

Type of Use (Select one or both, as applicable)

☑Prescription Use (Part 21 CFR 801 Subpart D)
☐Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is slightly slanted to the right. There is a small circle with an R inside of it to the right of the "x".

11.0 Executive Summary 510(k)

This Executive Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

A. 510(k) Number:

K161220

B. Purpose for Submission:

Traditional 510(k), New Device

C. Measurand:

Targets RNA sequences for the conserved regions of the matrix protein genes of influenza A and influenza B viruses, and the fusion gene of Respiratory Syncytial Virus (RSV).

D. Type of Test:

Qualitative Real Time Polymerase Chain Reaction (PCR)

E. Applicant:

Luminex Corporation

F. Proprietary and Established Names:

ARIES® Flu A/B & RSV Assay

G. Regulatory Information:

ProductCodeClassificationRegulation SectionPanel
OCCII21 CFR 866.3980—Respiratory viral panelmultiplex nucleic acid assayMicrobiology(83)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, dark gray circle above the "i" in "Luminex". A small registration mark is present to the right of the "x".

H. Intended Use:

1. Intended use(s):

The ARIES® Flu A/B & RSV Assay is a polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) nucleic acid in nasopharyngeal swabs (NPS) specimens from patients with signs and symptoms of respiratory tract infection in conjunction with clinical and laboratory findings. The test is intended for use as an aid in the differential diagnosis of Influenza B, and RSV in humans and is not intended to detect Influenza C.

Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Conversely, positive results do not rule-out bacterial infection with other viruses.

The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, X-ray findings) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.

Performance characteristics for influenza A were established during the 2014-2015 and the 2015-2016 influenza seasons when influenza A/H3N2 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

The ARIES® Flu A/B & RSV Assay is indicated for use with the ARIES® Systems.

    1. Indication(s) for use:
      Same as intended use.
    1. Special conditions for use statement(s):
      For prescription use only.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is slightly blurred, giving it a soft appearance. The overall impression is of a clean and modern logo.

4. Special instrument requirements:

For use with the ARIES® System or ARIES® M1 System.

l. Device Description:

The ARIES® Flu A/B & RSV Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system that will consist of an ARIES® System with its included software, an assay-specific cassette, and an assay-specific protocol file. The ARIES® Flu A/B & RSV Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® Flu A/B & RSV Assay cassette directly detects and differentiates influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) nucleic acid in nasopharyngeal swabs (NPS) specimens from patients with signs and symptoms of respiratory tract infection in conjunction with clinical and laboratory findings. Specifically, the ARIES® Flu A/B & RSV Assay cassette detects the matrix protein genes of influenza A and influenza B viruses, and the fusion gene of RSV and a RNA Sample Processing Control.

The specimen is lysed and nucleic acid is extracted using an ARIES® System. An extractable sample processing control (SPC) target is present in the ARIES® Flu A/B & RSV assay cassette and is processed with the specimen. The SPC controls for specimen lysis, for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify proper specimen lysis and nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument. The Tm value of the SPC is used as a reference for determining the target Tm.

The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® Flu A/B & RSV Assay lyophilized PCR reagents in the PCR tube that contain primer pairs specific to influenza A, influenza B, RSV, and the SPC sequence. The specific primer pairs are labeled with distinct fluorophore labels. PCR amplification is performed and assay fluorescence is monitored on an ARIES® System. Incorporation of the quencherlabeled nucleotide causes a decrease in assay fluorescence. Following amplification, the reaction is slowly heated and fluorescence is monitored. The strands of the amplification products will separate at a specific melting temperature (Tm) that is determined by an increase in fluorescence as the strands are separated. The instrument fluorescence output is analyzed and test results are determined using the ARIES® Flu A/B & RSV Assay protocol file. A printed results report is generated.

Substantial Equivalence Information: J.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small circle above the "i" in "Luminex". The word is slightly blurred, and the overall impression is of a company logo.

1. Predicate device name(s):

Simplexa Flu A/B & RSV Direct, Simplexa Flu A/B & RSV Positive Control Pack (manufactured by Focus Diagnostics, Inc.)

2. Predicate 510(k) number(s):

K120413

3. Comparison with predicate:

The following table compares the ARIES® Flu A/B & RSV Assay to Focus' Simplexa Flu A/B & RSV Direct, Simplexa Flu A/B & RSV Positive Control Pack (K120413). Table 11.1 shows similarities between the new device and the predicate, while Table 11.2 shows the differences.

Similarities
AttributeNew DevicePredicate Device (K120413)
Intended UseThe ARIES® Flu A/B & RSV Assay is apolymerase chain reaction (PCR) basedqualitative in vitro diagnostic test for thedirect detection and differentiation ofinfluenza A virus, influenza B virus, andrespiratory syncytial virus (RSV) nucleicacid in nasopharyngeal swabs (NPS)specimens from patients with signs andsymptoms of respiratory tract infectionin conjunction with clinical andlaboratory findings. The test is intendedfor use as an aid in the differentialdiagnosis of Influenza A, Influenza B, andRSV in humans and is not intended todetect Influenza C.Negative results do not precludeinfluenza virus or RSV infection andshould not be used as the sole basis fordiagnosis, treatment or othermanagement decisions. Conversely,positive results do not rule-out bacterialinfection or co-infection with otherviruses.The agent detected may not be thedefinite cause of disease. The use ofadditional laboratory testing (e.g.bacterial culture immunofluorescenceThe Focus Diagnostics Simplexa™ Flu A/B& RSV Direct assay is intended for use onthe 3M Integrated Cycler instrument forthe in vitro qualitative detection anddifferentiation of influenza A virus,influenza B virus, and respiratorysyncytial virus (RSV) RNA innasopharyngeal swabs (NPS) from humanpatients with signs and symptoms ofrespiratory tract infection in conjunctionwith clinical and epidemiological riskfactors. This test is intended for use as anaid in the differential diagnosis ofinfluenza A, influenza B, and RSV viralinfections in humans and is not intendedto detect Influenza C.Negative results do not precludeinfluenza virus or RSV infection andshould not be used as the sole basis fortreatment or other patient managementdecisions.Performance characteristics for influenzaA were established with clinicalspecimens collected during the2010/2011 influenza season when 2009

Table 11.1: Similarities between New Device and Predicate

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is in black and the background is white. There is a small circle with an R in it to the right of the x.

Assay Targets
X-ray findings) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.Performance characteristics for influenza A were established during the 2014-2015 and the 2015-2016 influenza seasons when influenza A/H3N2 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.The ARIES® Flu A/B & RSV Assay is indicated for use with the ARIES® Systems.H1N1 influenza and H3N2 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Assay TargetsRNA from Influenza A, Influenza B, and RSVRNA from Influenza A, Influenza B, and RSV
Influenza A TargetMatrix geneMatrix gene
Influenza B TargetMatrix geneMatrix gene
Sample typeNasopharyngeal swabs (NPS)Nasopharyngeal swabs (NPS)
Assay formatReal-time PCRReal-time PCR
Assay resultsQualitativeQualitative

Table 11.2: Differences between New Device and Predicate

Differences
AttributeNew DevicePredicate Device (K120413)
Extraction MethodAutomated by an ARIES® SystemNo extraction
RSV TargetFusion geneM gene
ControlsSample processing controlRNA Internal Control, Positive ControlsSold Separately
InstrumentARIES® System or ARIES® M1 System3M™ Integrated Cycler

K. Standards/Guidance Documents Referenced:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, dark gray circle above the "i" in "Luminex". The word is in black and the background is white. There is a small registered trademark symbol to the right of the "x".

    1. Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay, October 9, 2009
    1. Guidance for Industry and FDA Staff Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection or Detection and Differentiation of Influenza Viruses, July 15, 2011
    1. Guidance for Industry and FDA Staff, In Vitro Diagnostic Devices to Detect Influenza A Viruses: Labeling and Regulatory Path, May 1, 2007.

L. Test Principle:

The ARIES® Flu A/B & RSV Assay chemistry is based on an expanded genetic alphabet technology, consisting of synthetic DNA base pair 2'-deoxy-5-methyl-isocytidine (iC): 2'deoxyisoguanosine (iG). The isobases (iC and iG) pair specifically with each other and not with natural nucleotides. In addition, isobases are efficiently incorporated during PCR. During PCR amplification, a quencher-modified iGTP is incorporated by the polymerase opposite an iC and a fluorophore reporter attached to a PCR primer. If the target is present and is amplified, assay fluorescence decreases with every cycle as amplification product accumulates. The decrease in assay fluorescence is monitored in real time using an ARIES® System. Following PCR, the amplification products are thermally denatured and assay fluorescence is monitored. The strands of the amplification products are separated and assay fluorescence increases, thus enabling determination of the melting temperature (Tm) of the amplicon.

M. Performance Characteristics:

1. Analytical performance:

  • a. Precision/Reproducibility:

Precision

Within Laboratory Precision of the ARIES® Flu A/B & RSV Assay was evaluated by two operators performing testing with four ARIES "Systems using three lots of ARIES® Flu A/B & RSV Assay cassettes. Testing was performed on 12 nonconsecutive days and included a total of 675 replicates of a representative reproducibility panel. The reproducibility panel contained influenza A, influenza B, RSV-A, and RSV-B and representative viral cultures were diluted to three concentrations: moderate positive (10x LoD), low positive (3x LoD), and high negative (0.125x LoD for influenza A, and 0.25x LoD for influenza B, RSV-A and RSV-B). All dilutions were prepared in a simulated nasal matrix (SNM); negative samples consisted of only SNM. The overall

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Luminex. The logo is in black and white, with the word "Luminex" in a bold, sans-serif font. There is a small, gray circle above the "i" in Luminex. The logo is simple and modern.

invalid percentage for this study was 1.3% (9/675). The results of the reproducibility study are summarized in Table 11-3.

StrainTarget ConcentrationExpected PositivityPositivity95%ConfidenceInterval
Influenza A/HongKong/8/68Moderate Positive100%100%(48/48)93% - 100%
Low PositiveApproximately 95%100%(48/48)93% - 100%
High Negative20% - 80%79%(38/48)65% - 90%
InfluenzaB/Florida/04/06Moderate Positive100%100%(48/48)93% - 100%
Low PositiveApproximately 95%100%(48/48)93% - 100%
High Negative20% – 80%81%(39/48)a67% - 91%
RSV A2Moderate Positive100%100%(48/48)93% - 100%
Low PositiveApproximately 95%96%(46/48)86% - 99%
High Negative20% - 80%48%(23/48)33% - 63%
RSV B WV/14617/85Moderate Positive100%100%(48/48)93% - 100%
Low PositiveApproximately 95%100%(48/48)93% - 100%
High Negative20% – 80%90%(43/48)a77% - 97%
Flu A/B & RSVNegativeNegative0%2.1%(2/96)b0% - 7%
Table 11-3: ARIES® Flu A/B & RSV Assav Within Laboratory Precision Determination Results
----------------------------------------------------------------------------------------------

° RSV B WV/14617/45 and influenza B/Florida/04/06 High Negative samples generated positivity that expected positivity results of 20-80%.

' Two Flu A/B & RSV Negative replicates tested by different test dates, on different instruments, with different cassette lots generated late Ct false influenza B Positive results. Overal percentage of negative results for all three assay targets is 98%.

Reproducibility

Reproducibility of the ARIES® Flu A/B & RSV Assay was evaluated by testing one lot of ARIES® Flu A/B & RSV Assay Cassettes on two ARIES "Systems by two operators at each of three clinical laboratory sites on five non-consecutive days. A reproducibility panel was prepared containing a moderate positive (10x LoD), low positive (3x LoD), and high negative (0.2x LoD) independently for influenza A, influenza B, RSV-A, and RSV-B, as well as a negative sample. The reproducibility panels were created by an independent operator and blinded to the testing sites. Each panel member was

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is in black and the background is white. There is a small circle with a "c" in it to the right of the "x".

tested in triplicate by each operator on each day of testing. The results of the reproducibility study are shown in Tables 11-4 and 11-5.

Table 11-4: ARIES c Flu A/B & RSV Assay Site-to-Site Reproducibility Results
Site 1Site 2Site 3
PositivityPositivityPositivity
High Negative27/3090.0%30/30100.0%26/3086.7%
Influenza ALow Positive30/30100.0%30/30100.0%30/30100.0%
Moderate Positive30/30100.0%30/30100.0%30/30100.0%
Negative1/303.3%0/300.0%0/300.0%
High Negative9/3030.0%7/3023.3%15/3050.0%
Influenza BLow Positive29/3096.7%30/30100.0%30/30100.0%
Moderate Positive30/30100.0%30/30100.0%30/30100.0%
Negative0/300.0%0/300.0%0/300.0%
High Negative39/6065.0%50/6083.3%43/6071.7%
RSVcLow Positive60/60100.0%60/60100.0%60/60100.0%
Moderate Positive60/60100.0%60/60100.0%59/6098.3%
Negative0/300.0%0/300.0%0/300.0%
Table 11-4: ARIES® Flu A/B & RSV Assay Site-to-Site Reproducibility Results 30
------------------------------------------------------------------------------------------

a An overall invalid rate of 0.9% (11/1260) was observed in the target replicates.

° The expected result for: a moderate positive, low positive target was approximately 95% positive, high negative was 20 – 80% positive, and negative was 0% positive.

^ RSV-A and RSV-B are not differentiated by the ARES Flu A/B & RSV Assay and, therefore, are combined and represented as RSV only.

{11}------------------------------------------------

Positivity95% C.I. lower limitupper limit
High Negative83/9092.2%84.6%96.8%
Influenza ALow Positive90/90100.0%95.9%100.0%
Moderate Positive90/90100.0%95.9%100.0%
Negative1/901.1%1.1%6.0%
High Negative31/9034.4%24.7%45.2%
Influenza BLow Positive89/9098.9%94.0%100.0%
Moderate Positive90/90100.0%95.9%100.0%
Negative0/900.0%0.0%4.0%
RSVcHigh Negative132/18073.3%66.2%79.6%
Low Positive180/180100.0%95.9%100.0%
Moderate Positive179/18099.4%96.9%100.0%
Negative0/900.0%0.0%4.0%

Table 11-5: Reproducibility Panel Total Results 3.6

a An overall invalid rate of 0.9% (11/1260) was observed in the target replicates.

ٌ The expected result for: a moderate positive, low positive target was approximately 95% positive, high negative was 20 – 80% positive, and negative was 0% positive.

် SV-A and RSV-B are not differentiated by the ARIES Flu A/B & RSV Assay and, therefore, are combined and represented as RSV only.

b. Linearity/assay reportable range:

Not applicable. The ARIES® Flu A/B & RSV Assay is a qualitative assay.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Stability:

Specimen Stability (Fresh vs. Frozen)

{12}------------------------------------------------

Fresh vs. frozen specimen stability was determined with contrived fresh and frozen specimens stored at either 2-8°C (fresh) or -65 to -95°C (frozen) and tested with the ARIES® Flu A/B & RSV Assay. This was assessed by testing 60 replicates for a single target pathogen (influenza A, influenza B, or RSV) prepared at 3 test concentrations (30 low positive, 15 moderate positive, and 15 high positive) across 9 different time points extending out to 12 months. In addition to the 60 replicates for each target organism, 6 replicates of an influenza A/B and RSV negative specimen in simulated nasal matrix (SNM) were tested for each time point extending out to 12 months. Data up to 3 months was collected with all targets yielding the expected result. Overall, high positive specimens were 100% positive, moderate positive specimens were 100% positive, and low positive specimens were positive approximately 95% of the time for all three evaluated targets. Negative specimens were negative 100% of the time. Influenza A, influenza B, and RSV specimens are stable for up to 7 days when stored at 2-8°C and 3 months when stored at -65 to -95°C.

Shelf-Life Stability

Real time stability study was performed to evaluate the shelf life of ARIES® Flu A/B & RSV Assay cassettes. Stability was established by testing 6 replicates of ARIES® Flu A/B & RSV Assay Extractable Control, 100X LOD blend (PN 10188) and negative targets (UTM) on three different lots of ARIES® Flu A/B & RSV Assay cassettes stored at 2 different temperatures: 4°C (2 = 8°C) and room temperature (15 = 30°C) at 10 different time points extending out to 19 months. Acceptance criteria for stability at each time point and temperature was established as 100% positivity for all influenza A/B and RSV replicates, and 100% negativity for all negative replicates. Data collected up to 5 months gave expected results indicating stability of the of ARIES® Flu A/B & RSV Assay cassettes up to 5 months.

Open Box Stability

An Open Box Stability study was performed in order to evaluate performance of ARIES® Flu A/B & RSV Assay Cassettes after they were removed from their individual pouches. Cassettes were removed from their pouches and placed on a laboratory bench where they were exposed to ambient temperatures, humidity, and light for up to ten (10) hours. Data was collected for contrived influenza A, influenza B, and RSV positive and influenza A, influenza B, and RSV negative (Copan Universal Transport Media) samples at five time points throughout a ten (10) hour duration. Three lots of cassettes were used to assess Open Box Stability. At the end of 10 hours, all three lots of cassettes produced expected results showing that ARIES® Flu A/B and RSV Assay Cassettes are stable in ambient laboratory conditions for up to ten (10) hours after they have been removed from the storage pouch.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is likely a logo or brand name. The overall image has a clean and professional appearance.

Controls:

Process Control

Each ARIES® assay cassette contains a Sample Process Control (SPC), which is processed with the sample and analyzed during the amplification reaction. The SPC verifies sample lysis, nucleic acid extraction, and proper reagent, cassette, ARIES® System, and assay protocol performance. The SPC has a known melting temperature (Tm) range and Ct range. Each time an assay is run, the system measures the temperature and fluorescence intensity of the SPC control to ensure the thermal and optical subsystems have remained in calibration.

External Controls

External controls should be tested according to guidelines or requirements of local, provincial and/or federal regulations or accreditation organizations. For example, reference influenza A, influenza B & RSV strains or well characterized influenza A, influenza B, and RSV clinical isolates may be used as positive controls. The ARIES® Flu A/B & RSV Assay Cassette Kit does not include external positive and negative controls.

d. Detection Limit:

Limit of Detection (LoD) was established for the ARIES® Flu A/B & RSV Assay using three influenza A, two influenza B, and two respiratory syncytial viral strains (1 RSV-A, and 1 RSV-B) diluted in a simulated nasal matrix (SNM) containing negative pooled human clinical matrix and Universal Transport media (UTM). The LoD for each viral strain was determined as the lowest concentration (TCIDso/mL) that had a positivity rate of ≥ 95%. Serial dilutions of each quantified viral strain in SNM were initially tested in a range finding study where the preliminary LoD TCIDso/mL concentrations were determined. The preliminary LoD concentrations were then confirmed by testing twenty (20) replicates of each strain. The final LoD concentrations for the seven viral strains are shown in Table 11-6.

{14}------------------------------------------------

Assay TargetStrainConcentration(TCID50/mL or CEID50/mL)Positivity95% ConfidenceInterval (Lower Limit /Upper Limit)
Influenza APR/8/341 x 10 $$100%83.2% - 100%
Hong Kong/8/681 x 10 $$100%83.2% - 100%
Mexico/4108/2009(H1N1)pdm091 x 10 $$95%75.1% - 99.9%
Influenza BFlorida/04/061 x 10 $$100%83.2% - 100%
Malaysia/2506/041 x 10 $$100%83.2% - 100%
RSVA2 (VR-1540)1 x 10 $$95%75.1% - 99.9%
WV/14617/85 (VR-1400)1 x 10 $$100%83.2% - 100%

Table 11-6: Limit of Detection of the ARIES® Flu A/B & RSV Assay

e. Analytical Reactivity (Inclusivity)

The analytical reactivity/inclusivity of the ARIES® Flu A/B and RSV Assay was evaluated against twenty-four influenza A strains, seven influenza B strains, and three respiratory syncytial virus (RSV) strains, which differ from those strains included in the Limit of Detection (LoD) study. Each strain was diluted in simulated nasal matrix (SNM) to a concentration near the LoD (1x10+ TCID50/mL or CEID50/mL) and tested in triplicate; SNM consists of a 1:1 mixture of Universal Transport Medium (UTM) and 100% negative clinical specimen pool. For five strains, additional testing at 1x10 TCID50/mL or CEID50/mL was required to achieve 100% positivity. For two strains, additional testing at 1x103 TCID50/mL or CEID50/mL was required to achieve 100% positivity. The study results are shown in Table 11-7.

Table 11-7: ARIES® Flu A/B & RSV Assay Analytical Reactivity/Inclusivity Results

Inclusivity Strains100% PositivityConcentrationResult
Influenza A /Perth/16/2009 (H3N2)-like$1x10^1$Influenza A Detected
Influenza A/Brisbane/10/07 H3$1x10^1$Influenza A Detected
Influenza A/Brisbane/59/07 H1$1x10^1$Influenza A Detected
Influenza A/Port Chalmers/1/73 H3N2$1x10^1$Influenza A Detected
Influenza A/Solomon Island/03/06 H1$1x10^1$Influenza A Detected
Influenza A/Swine H1N1/Iowa/15/1930$1x10^1$Influenza A Detected
Influenza A/Taiwan/42/06 H1N1$1x10^1$Influenza A Detected
Influenza A/Wisconsin/67/05 H3$1x10^1$Influenza A Detected
Influenza A/California/7/2009-like (pH1N1)$1x10^1$Influenza A Detected

{15}------------------------------------------------

Influenza A/Hong Kong/33982/2009 H9N2 x PR8-1x101Influenza A Detected
Influenza A/Indiana/08/2011 (H3N2)v1x101Influenza A Detected
Influenza A/Texas/50/2012 H3N21x101Influenza A Detected
Influenza A/WS/33 H1N11x101Influenza A Detected
Influenza A/New Caledonia/20/99 H1N11x102Influenza A Detected
Influenza A/Swine H1N1/USA/1976/19311x102Influenza A Detected
Influenza A/California/07/2009 NYMC x-179A1x102Influenza A Detected
Influenza A/Victoria/361/2011-like (H3N2)1x102Influenza A Detected
Influenza A/Minnesota/11/2010 (H3N2)v1x103Influenza A Detected
Influenza A/Ohio/02/2012 (H3N2)1x103Influenza A Detected
A/Anhui/01/2005 (H5N1)a1x101Influenza A Detected
A/Anhui/1/2013 (H7N9)a1x101Influenza A Detected
A/Egypt/321/2007 (H5N1)a1x101Influenza A Detected
A/Shanghai/1/2013 (H7N9)a1x101Influenza A Detected
A/Vietnam/1194/2004 (H5N1)a1x101Influenza A Detected
Influenza B/Massachusetts/2/2012-like1x101Influenza B Detected
Influenza B/Wisconsin/1/2010-like1x101Influenza B Detected
Influenza B/Florida/02/2006 (Victoria)1x101Influenza B Detected
Influenza B/Lee/401x101Influenza B Detected
Influenza B/Panama/45/90 (Yamagata)1x101Influenza B Detected
Influenza B/Brisbane/60/20081x101Influenza B Detected
Influenza B/Florida/07/04 (Yamagata)1x102Influenza B Detected
RSV A/Long1x101RSV Detected
RSV B/93201x101RSV Detected
RSV B/Wash/18537/621x101RSV Detected

a BPL-inactivated viral culture fluid provided from IRR with no unit/titer information. Dilution factors based on prior preliminary LoD testing of these strains with the ARIES® Flu A/B & RSV assay to approximate limit of detection.

Analytical specificity: f.

Interfering Substances:

The potential inhibitory effect of non-microbial substances expected to be found in nasopharyngeal swab specimens was evaluated by the testing with the ARIES® FLU A/B & RSV Assay. Three (3) replicates each of influenza A, influenza B, RSV-A, and RSV-B were tested at concentrations near the assay LoD with a clinically relevant

{16}------------------------------------------------

concentration of each potentially interfering substance spiked into the contrived sample; additionally, negative Copan Universal Transport Medium (UTM) was spiked with the same concentration of each substance and tested for assay interference.

The results of the study demonstrate that all influenza A, influenza B, and RSV samples were 100% positive in the presence of a non-microbial substance at concentrations shown below; all negative samples containing only the non-microbial substance were 100% negative, with the exception of FluMist®.

Positive influenza results obtained in a patient who received FluMist prior to sample collection may be due to detection of the live attenuated influenza vaccine virus and may mask a true positive result caused by an influenza infection; additionally, FluMist may interfere with RSV detection due to high concentration of vaccine virus nucleic acid, causing a possible RSV false negative result. This statement has been added to the Analytical Performance section of the ARIES® FLU A/B & RSV Product Insert.

Interfering SubstanceTest Concentration
Benzocaine2.5% w/v
Budesonide25 mg/mL
Dexamethasone3 mg/mL
FluMist°0.5% v/v
Flunisolide55 mg/mL
Menthol1.7 mg/mL
Mometasone2.5 mg/mL
Phenylephrine0.5% w/v
Afrin° (Oxymetazoline)15% v/v
Tobramycin4 µG/mL
Mupirocin6.6 mg/mL
Beconase AQ° (Beclomethasone)5% v/v
Flonase° (Fluticasone)5% v/v
Zanamivir3.3 mg/mL
Tamiflu°1 μΜ
Triamcinolone5.5 mg/mL
Sodium chloride0.65% v/v
Human Whole Blood2% v/v
Mucin Protein60 μG/mL
ZICAM° (Galphimia glauca, Histaminum hydrochloricum, Luffa operculata, Sulfur)5% v/v

Table 11-8: Interfering Substances Tested

Cross-Reactivity:

{17}------------------------------------------------

The analytical specificity for the ARIES® Flu A/B & RSV Assay was evaluated by testing the potential cross-reactivity of 32 microorganisms listed in Table 11-9. The microorganisms tested consisted of 14 viral and 18 bacterial strains representing common respiratory pathogens, or those potentially encountered in the human nasopharynx region. The potential cross reactive organisms were spiked into simulated nasal matrix (SNM) that was negative for influenza A, influenza B, and RSV and tested with the ARIES® Flu A/B & RSV Assay in triplicate. Bacterial organisms were tested at concentrations ≥10° CFU/mL and viral organisms at ≥10° TCID50/mL, or the highest available concentration for both types of potential cross-reactive microorganism. All replicates of 31 microorganisms that were evaluated yielded negative results following initial testing. One replicate of three tested for Parainfluenza type 1 generated a late Ct indicating a weak influenza B positive result (Ct = 40) during the initial testing. An additional five replicates of the same Parainfluenza type 1 strain was tested with all 5 replicates yielding negative target results. Parainfluenza type 1 was considered to be non-reactive with the ARIES® Flu A/B & RSV Assay after the repeat testing.

Microorganism
Adenovirus type 1Haemophilus influenzaeParainfluenza type 1
Adenovirus 7aMycoplasma pneumonia M129Parainfluenza type 2
Bordetella pertussis (A639)Mycobacterium tuberculosisParainfluenza type 3
Chlamydia pneumoniaeLactobacillus plantarum (17-5)Pseudomonas aeruginosa
Coronavirus 229ELegionella longbeachaeRhinovirus type 1A
Coronavirus OC43MeaslesStaphylococcus aureus (COL)
Corynebacterium diphtheriaeMetapneumovirusStaphylococcus epidermidis
Cytomegalovirus (CMV)Moraxella catarrhalis Ne 11Streptococcus pneumoniae
Enterovirus 71MumpsStreptococcus pyogenes
Epstein Barr virusNeisseria elongataStreptococcus salivarius
Escherichia coli O157Neisseria meningitidis

Table 11-9: Microorganism Information

Microbial Interference:

Microbial interference for the ARIES® Flu A/B & RSV Assay was assessed with the 32 potential cross reactive microorganisms evaluated in the Cross-Reactivity Study, and identified in Table 11-9 above. Bacteria were tested at ≥10° CFU/mL or the highest available concentration, and viruses were tested at ≥10° TCID50/mL or the highest available concentration. The potential interfering organisms were spiked into simulated nasal matrix (SNM) containing either representative strains of influenza A, influenza B, RSV A, or RSV B near the LoD concentration. All target strain + cross reacting organism samples were tested in triplicate (n=3) on the ARIES® System.

{18}------------------------------------------------

Influenza A was correctly detected in all replicates. Influenza B was correctly detected in the presence of 29 cross reacting organisms, with 3 cross reacting organisms (Parainfluenza type 2, Coronavirus OC43, Corynebacterium diphtheriae) requiring additional testing of 5 replicates. For all repeated cross reacting organism + influenza B combinations, all 5 replicates correctly detected influenza B. Parainfluenza type 2, Coronavirus OC43, and Corynebacterium diphtheriae are considered non-reactive since 7 of 8 tested replicates generated correct results (initial testing of 3 replicates resulted in 2 out of 3 positive, and the repeat testing of 5 replicates resulted in 5 out of 5 positive results, denoted as 7 of 8).

RSV A was correctly detected in the presence of 28 cross reacting organisms, with 4 cross reacting organisms (Bordetella pertussis, Cytomegalovirus, Parainfluenza type 2, Epstein Barr virus) requiring additional testing of 5 replicates. For all repeated cross reacting organism + RSV A combinations, all 5 replicates correctly detected RSV A. Bordetella pertussis, Cytomegalovirus, Parainfluenza type 2, and Epstein Barr virus are considered non-reactive since 7 of 8 tested replicates generated correct results. RSV B was correctly detected in the presence of 28 cross reacting organisms, with 4 cross reacting organisms (Parainfluenza type 2, Rhinovirus type 1A, Enterovirus 71, Neisseria meningtidis) requiring additional testing of 5 replicates. For all repeated cross reacting organism + RSV B combinations excluding Enterovirus 71, all 5 replicates correctly detected RSV A. Parainfluenza type 2, Rhinovirus type 1A, and Neisseria meninqtidis are considered non-reactive since 7 of 8 tested replicates generated correct results. RSV B had an overall positivity of 6/8 when tested in the presence of Enterovirus 71. Additional testing of RSV B + Enterovirus 71 produced a final positivity of 3/3 using freshly prepared target material. Based on the final test result of 100% positivity (3/3) for RSV B + Enterovirus 71, Enterovirus 71 is considered non-reactive.

Carry-Over/Cross-Contamination:

Carry-over and Cross Contamination for the ARIES® Flu A/B and RSV Assay were evaluated by testing thirty (30) high concentration influenza A positive samples in series alternating with thirty (30) influenza A negative samples (UTM). The high positive samples were run adjacent to negative samples across ten (10) consecutive runs on one ARIES® System. No carry-over or cross contamination was observed, and the overall percent agreement was 100% for positive and negative samples.

Competitive Interference/Co-infection:

{19}------------------------------------------------

Image /page/19/Picture/1 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". A small registration mark is present to the right of the "x".

A study was designed to evaluate the ability of the ARIES® Flu A/B and RSV assay to detect influenza A, influenza B, RSV A, and RSV B in the presence of a co-infection. Competitive interference can occur when one analyte is near the LoD and an additional analyte is present at high concentration. Analytes were tested at high (> 1x10* TCID50/mL) and low concentrations (1.5X LoD) using 3 replicates in various combinations, excluding RSV A + RSV B due to the assay's inability to distinguish between RSV subtypes. Low concentration analytes were detected for 3 of the 4 assay targets with 100% positivity in all high concentration combinations. RSV A in presence of high concentration influenza A required a 3X LoD dilution to achieve 100% positivity (Table 11-10).

Table 11-10: Co-infection Results Summary

Low Target AnalyteHigh Target AnalyteConditionPercent
A/Hong Kong/8/68B/Florida/04/06Influenza A + Influenza B100%
A/Hong Kong/8/68RSV A2Influenza A + RSV A100%
A/Hong Kong/8/68RSV B WV/14617/85Influenza A + RSV B100%
B/Florida/04/06A/Hong Kong/8/68Influenza B + Influenza A100%
B/Florida/04/06RSV A2Influenza B + RSV A100%
B/Florida/04/06RSV B WV/14617/85Influenza B + RSV B100%
RSV A2B/Florida/04/06RSV A + Influenza B100%
RSV A2 (3X LoD)A/Hong Kong/8/68RSV A + Influenza A100%
RSV B WV/14617/85A/Hong Kong/8/68RSV B + Influenza A100%
RSV B WV/14617/85B/Florida/04/06RSV B + Influenza B100%
  • g. Assay cut-off:
    Not applicable.

    1. Comparison Studies:
    • Method comparison with predicate device: a.

Not applicable.

  • b. Matrix comparison:
    A matrix equivalency study was performed to establish equivalency of a simulated matrix (50:50 dilution of pooled negative clinical specimens in Copan Universal Transport Media (UTM)) with a natural matrix (pooled negative clinical specimens)

{20}------------------------------------------------

when used in the ARIES® Flu A/B and RSV Assay. Study samples were prepared as a 4 point, 5-fold dilution series of representative Influenza A, Influenza B, and Respiratory Syncytial Virus (RSV) viral strain targets in both the simulated matrix and the pooled negative clinical matrix with the lowest dilution point targeting 3-5x the determined limit of detection. Each concentration level for each viral target strain was tested in 6 replicates for each matrix type for a total of 48 tests per viral strain. Additionally, 6 replicates of the simulated matrix and 6 replicates of the pooled negative clinical matrix were tested without the introduction of viral strain material. Equivalency was demonstrated based on the compared mean target Ct and SPC Ct values falling within 1 log (3.3 cycles) across the matrix types, and the accuracy of the clinical call in the simulated matrix compared with that of the natural matrix. The results of the study are shown in Tables 11-11 to 11-14. These results demonstrate the analytical performance of the ARIES® Flu A/B and RSV Assay is equivalent between the simulated matrix and the natural matrix.

Matrix typeDilution 1Dilution 2Dilution 3Dilution 4
Mean Ct ± SD (cycles)PositivityDelta CtMean Ct ± SD (cycles)PositivityDelta CtMean Ct ± SD (cycles)PositivityDelta CtMean Ct ± SD (cycles)PositivityDelta Ct
Simulated matrix26.2 ± 0.7100%28.4 ± 0.4100%32.1 ± 0.8100%33.8 ± 1.3100%
Negative Clinical Specimen matrix26.3 ± 1.1100%-0.129.4 ± 1.0100%-1.030.8 ± 0.4100%1.334.0 ± 1.083%a-0.2

a Positivity is within the 95% Cl of expected results of 35.9% - 99.6%

Table 11-12: ARIES® Flu A/B and RSV Assay Matrix Equivalency Results for Flu B

Influenza B/Florida/04/06
Matrix typeDilution 1Dilution 2Dilution 3Dilution 4
Mean Ct ± SD (cycles)PositivityDelta CtMean Ct ± SD (cycles)PositivityDelta CtMean Ct ± SD (cycles)PositivityDelta CtMean Ct ± SD (cycles)PositivityDelta Ct

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the Luminex logo. The logo is in black and white and features the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in Luminex. The logo is simple and modern.

Simulated matrix$28.0 \pm 0.5$100%0.1$30.2 \pm 0.7$100%-1.2$33.0 \pm 0.6$100%0.2$35.4 \pm 0.8$100%1.0
Negative ClinicalSpecimen matrix$28.1 \pm 0.3$100%$31.4 \pm 0.7$100%$32.9 \pm 0.8$100%$34.4 \pm 0.9$100%

Table 11-13: ARIES® Flu A/B and RSV Assay Matrix Equivalency Results for RSV

MatrixtypeRSV A2
Dilution 1Dilution 2Dilution 3Dilution 4
Mean Ct ±SD (cycles)PositivityDelta CtMean Ct ±SD (cycles)PositivityDelta CtMean Ct ±SD (cycles)PositivityDelta CtMean Ct ±SD (cycles)PositivityDelta Ct
Simulatedmatrix30.9 ±0.9100%32.5 ±0.6100%34.9 ±0.7100%35.3 ±1.4100%
NegativeClinicalSpecimenmatrix30.8 ±0.6100%0.233.1 ±0.6100%-0.635.6 ±1.0100%-0.735.6 ±0.8100%-0.4

Table 11-14: ARIES® Flu A/B and RSV Assay Matrix Equivalency Results for SPC

Matrix typeSPC
Mean Ct ± SD (cycles)PositivityDelta Ct
Simulated matrix27.6 ± 0.7100%
Negative ClinicalSpecimen matrix27.6 ± 0.8100%0.0

Swab comparison: C.

A Nasopharyngeal Swab Equivalency study was performed to evaluate the reproducibility of the ARIES® Flu A/B & RSV Assay using three representative nasopharyngeal swab (NPS) types (flocked, rayon, and polyester) and two different swab sizes (minitip and regular). All swab types and swab sizes were evaluated with influenza A, influenza B, and RSV viral cultures prepared as a three point five-fold dilution series, transferred via select swab type and size to universal transport media (UTM), and tested with the ARIES® Flu A/B & RSV assay. In addition to the viral

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, dark gray circle above the "i" in "Luminex". The word is slightly blurred, giving it a soft appearance. A registered trademark symbol is present to the right of the word.

cultures, a negative sample consisting of swab transferred UTM was also evaluated. Overall, all swab types and sizes performed as expected generating 100% positivity for all evaluated target concentrations and 100% negativity for negative replicates (Table 11-15).

Assay TargetFinalConcentration(TCID50/mL)Swab TypeRegular Size PositivityMinitip Size Positivity
1x104.5Flocked100%100%
A/Hong Kong/8/68LoD = 1x102.401x103.8100%100%
1x103.1100%100%
1x104.5Polyester100%100%
1x103.8100%100%
1x103.1100%100%
1x104.5Rayon100%100%
1x103.8100%100%
1x103.1100%100%
1x102.4Flocked100%100%
B/Florida/04/06LoD = 1x100.301x101.7100%100%
1x101.0100%100%
1x102.4Polyester100%100%
1x101.7100%100%
1x101.0100%100%
1x101.5Flocked100%100%
RSV A2LoD = 1x10-0.571x100.8100%100%
1x100.1100%100%
1x101.5Polyester100%100%
1x100.8100%100%
1x100.1100%100%
1x101.5Rayon100%100%
1x100.8100%100%
1x100.1100%100%

Table 11-15: ARIES® Flu A/B & RSV Assay NPS Equivalency Results

  1. Clinical Performance:

The clinical performance of the ARIES® Flu A/B & RSV Assay was evaluated using leftover, de-identified, nasopharyngeal swab (NPS) specimens prospectively collected from pediatric or adult patients suspected of having respiratory tract infection during the 2014/2015 and 2015/2016 flu seasons. In the first phase of the prospective study (2014/2015 flu season) specimens were collected at 3 clinical sites located in the United States and Canada from 18-January-2015 to 20-March-2015. Clinical specimens accrued during the second phase of the prospective study (2015/2016 flu season) were collected from 02-November-2015 to 29-February-2016 at 4 clinical sites located in the United States and Canada. The clinical specimen collection sites were chosen based on the

{23}------------------------------------------------

Image /page/23/Picture/12 description: The image shows the word "Luminex" in a bold, sans-serif font. The letters are black, and there is a gray dot above the "i" in "Luminex". A small circle with a "c" inside is located to the right of the "x".

types of patients usually referred to them, and the prevalence of respiratory pathogens in order to ensure a broad coverage of respiratory organisms, patient ages, and geographical regions.

A total of 2504 nasopharyngeal swab specimens from subjects suspected of having respiratory tract infection were collected in the prospective study. Twenty-five (25) specimens were excluded based on inclusion criteria violation or protocol deviation leaving a total of 2479 eligible unique specimens for subsequent data analysis. Of these, 1017 were collected during the 2014/2015 Flu season while the remaining 1462 specimens were enrolled during the 2015/2016 Flu season.

All 2479 eligible clinical specimens were tested by an FDA-cleared molecular comparator and the ARIES® Flu A/B & RSV Assav. The comparator assay was performed in accordance with manufacturer's instructions at a centralized testing facility. Clinical runs and re-runs using ARIES® Flu A/B & RSV Assay were carried out by trained operators at the testing sites on clinical specimens that were either stored frozen at -65°C to -80°C (N=1316; 53.1%) or kept refrigerated at 2°C to 8°C (N=1163; 46.9%) prior to testing.

Out of the 2479 clinical specimens included in the prospective analysis, 2458 (2458/2479; 99.2%) generated valid results with ARIES® Flu A/B & RSV Assay during initial testing. There were 21 specimens (21/2479, 0.8%) that were re-tested with ARIES® Flu A/B & RSV Assay because they yielded initial invalid results. All 21 specimens in question generated valid ARIES results upon repeat testing.

ARIES® Flu A/B & RSV Assay Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for Influenza A, Influenza B and RSV are summarized in the tables below. Discordant specimens where the ARIES® Flu A/B & RSV Assay results were different from the comparator assay result were further assessed by bi-directional sequencing using analytically validated primers that targeted genomic regions distinct from the ARIES® Flu A/B & RSV Assay. Results from discordant testing analysis were not included in the calculation of Positive Percent and Negative Percent Agreement for each target. These results are, however, included as footnotes in the performance evaluation tables for each analyte.

ARIES Flu A/B & RSV AssayReference
PositiveNegativeNo CallTOTAL
Positive2993421334
Negative131213112145
No Call0000
TOTAL312216522479
95% CI

Table 11-16: ARIES Flu A/B & RSV Assay Performance for Influenza A (N=2479)

Confidential & Restricted

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is slightly slanted to the right. There is a small circle with a "c" in the middle to the bottom right of the word.

Positive Percent Agreement95.8%93.0% - 97.8%
Negative Percent Agreement98.4%97.8% - 98.9%

් Seven (7) ARIES® Flu A/B & RSV Assay negative specimens that were positive by the comparator method (i.e. False Negative by bi-directional sequencing analytically validated primers that targeted genomic regions distinct from the ARES® Flu A/B & RSV Assav.

² Four (4) ARIES® Flu A/B & RSV Assay positive by the comparator method (i.e. False Positive) tested positive by bi-directional sequencing analytically validated primers that targeted genomic regions distinct from the ARIES® Flu A/B & RSV Assay.

Table 11-17: ARIES Flu A/B & RSV Assay Performance for Influenza B (N=2479)

ARIES Flu A/B & RSV AssayReference
PositiveNegativeNo CallTOTAL
Positive45142059
Negative31241702420
No Call0000
TOTAL48243102479
95% CI
Positive Percent Agreement93.8%82.8% - 98.7%
Negative Percent Agreement99.4%99.0% - 99.7%

² Two (2) ARIES® Flu A/B & RSV Assay negative specimens that were positive by the comparator method (i.e. False Negative by bi-directional sequencing analytically validated primers that targeted genomic regions distinct from the ARIES® Flu A/B & RSV Assay.

1 Three (3) ARIES® Flu A/B & RSV Assay positive speciment that were negative by the comparator method (i.e. False Positive by bi-directional sequencing analytically validated primers that targeted genomic regions distinct from the ARIES® Flu A/B & RSV Assay.

ARIES Flu A/B & RSV Assay
PositiveNegativeNo CallTOTAL
Positive2703620306
Negative81216502173
No Call0000
TOTAL278220102479
95% CI
Positive Percent Agreement97.1%94.4% - 98.7%
Negative PercentAgreement98.4%97.7% - 98.9%

Table 11-18: ARIES Flu A/B & RSV Assay Performance for RSV (N=2479)

2 ଠne (1) ARIES® Flu A/B & RSV Assay negative specitive by the comparator method (i.e. False Negative) tested negative by bidirectional sequencing analytically validated primers that targeted genomic regions distinct from the ARIES® Flu A/B & RSV Assay.

{25}------------------------------------------------

4 Thirty-two (32) ARIES® Flu A/B & RSV Assay positive specimens that were negative by the comparator method (i.e. False Positive) tested positive by bi-directional sequencing analytically validated primers that targeted genomic regions distinct from the ARES® Flu A/B & RSV Assay.

Due to the low prevalence of Influenza B observed in the prospective study, the sample set was supplemented with 40 banked (pre-selected) Influenza B positive specimens collected at a single clinical laboratory (collection site) in Canada. The presence of the expected target (Influenza B) in each of the pre-selected specimens was confirmed by the comparator assay. In order to minimize bias, the pre-selected positive specimens were tested along with 40 unique negative clinical specimens in a randomized, blinded fashion at 4 external testing sites. Of the 80 Influenza B positive and negative specimens included in this study, 78 (78/80; 97.5%) generated valid results with ARIES® Flu A/B & RSV Assay on the first attempt. Invalid results were generated for 2 specimens (2/80; 2.5%) tested. Both specimens in question yielded valid ARIES results upon repeat testing.

The results from pre-selected specimens were analyzed separately from those of the prospective data set. ARIES® Flu A/B & RSV Assay accurately detected all 40 Influenza B positive specimens tested (100%; 95% confidence interval, 91.2% - 100%).

Based on the data collected from the clinical study, Positive Percent Agreement acceptance criteria of 90% with a lower bound 95% confidence interval of at least 80% were achieved for Influenza A, Influenza B and RSV. Negative Percent Agreement acceptance criteria of 90% with a lower bound 95% confidence interval of at least 90% were also achieved for all targets probed by the ARIES® Flu A/B & RSV Assay.

The results generated from this clinical study demonstrate that the diagnostic accuracy of the ARIES® Flu A/B & RSV Assay is acceptable for the detection of Influenza A, Influenza B and RSV in nasopharyngeal swabs collected from patients suspected of respiratory tract infection (RTI).

4. Expected values:

ARIES® Flu A/B & RSV Assay positive results (expected values) after allowable re-runs for each individual target are summarized for each of the enrollment periods, per age groups and per site in Tables 11-19 to 11-22 below.

Table 11-19: Expected Values (As determined by ARIES® Flu A/B & RSV) – Summary by Age Groups for the ARIES® Flu A/B & RSV Prospective Clinical Study (January 2015 – March 2015)

TargetOverall(n=1017)0-1 year(n=190)>1-5 years(n=92)>5-21 years(n=131)>21-65 years(n=307)>65 years(n=297)
-----------------------------------------------------------------------------------------------------------------------------------------------

{26}------------------------------------------------

(Analyte)No.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected Value
Influenza A17016.7%84.2%1314.1%2519.1%5317.3%7123.9%
Influenza B313.0%10.5%33.3%118.4%103.3%62.0%
RSV10410.2%5026.3%1415.2%43.1%154.9%217.1%

Table 11-20: Expected Values (As determined by ARIES® Flu A/B & RSV) – Summary by Age Groups for the ARIES® Flu A/B & RSV Prospective Clinical Study (November 2015 – February 2016)

Overall(n=1462)0-1 year(n=244)>1-5 years(n=131)>5-21 years(n=114)>21-65 years(n=538)>65 years(n=435)
Target(Analyte)No.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected Value
Influenza A16411.2%93.7%1713.0%1513.2%9818.2%255.7%
Influenza B281.9%31.2%21.5%97.9%91.7%51.1%
RSV20213.8%8334.0%3325.2%108.8%244.5%5212.0%

Table 11-21: Expected Values (As determined by ARIES® Flu A/B & RSV) – Summary by Site for the ARIES® Flu A/B & RSV Prospective Clinical Study (January 2015 – March 2015)

Target (Analyte)Overall (n=1017)Site 01 (n=238)Site 02 (n=225)Site 04 (n=554)
No.Expected ValueNo.Expected ValueNo.Expected ValueNo.Expected Value
Influenza A17016.7%4217.6%3917.3%8916.1%
Influenza B313.0%20.8%177.6%122.2%
RSV10410.2%125.0%208.9%7213.0%

{27}------------------------------------------------

Overall (n=1462)Site 01 (n=573)Site 02 (n=102)Site 03 (n=387)Site 04 (n=400)
Target(Analyte)No.ExpectedValueNo.ExpectedValueNo.ExpectedValueNo.ExpectedValueNo.ExpectedValue
Influenza A16411.2%7412.9%1211.8%164.1%6215.5%
Influenza B281.9%81.4%43.9%30.8%133.3%
RSV20213.8%5810.1%1211.8%5313.7%7919.8%

Table 11-22: Expected Values (As determined by ARIES® Flu A/B & RSV) – Summary by Site for the ARIES® Flu A/B & RSV Prospective Clinical Study (November 2015 – February 2016)

N. Proposed Labeling:

The labeling provided in the submission satisfies the requirements of 21 CFR 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.