K Number
K161182

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-08-15

(110 days)

Product Code
Regulation Number
866.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana® Trichomonas Assay is an in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection of Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swabs and female urine specimens obtained from symptomatic females to aid in the diagnosis of trichomoniasis. The Solana® Trichomonas Assay is intended for use only with the Solana® instrument.

Device Description

The Solana® Trichomonas Assay amplifies and detects Trichomonas vaginalis nucleic acids present in clinician-collected vaginal swabs and urine specimens from symptomatic and asymptomatic women. The assay targets a conserved multi-copy sequence of the T. vaqinalis DNA. The vaginal swab is eluted in a swab lysis tube or a urine specimen is added to a urine lysis tube, and the cells in either specimen type are lysed by simple heat treatment. After heat treatment, an aliquot of the lysed specimen is transferred into a dilution tube. An aliquot of the diluted sample is added to a reaction tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of T. vaginalis-specific target sequence. In Solana, the target sequence is amplified by T. vaginalis specific primers and detected by a T. vaginalis specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by T. vaginalis specific primers and detected by a PRC specific fluorescence probe. The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to T. vaqinalis or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and the results can be printed via a printer.

AI/ML Overview

Here's an analysis of the provided text, focusing on acceptance criteria and study details for the Solana® Trichomonas Assay:

Executive Summary:

The Solana® Trichomonas Assay is an in vitro diagnostic test for the qualitative detection of Trichomonas vaginalis using isothermal amplification technology (HDA). The device was evaluated in a multi-center clinical study comparing its performance against a composite reference method (Wet Mount and InPouch TV Culture) for both vaginal swab and urine specimens. The study included symptomatic and asymptomatic female patients.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" as distinct numerical targets that the FDA required the device to meet for approval. Instead, it presents the "Performance Characteristics" from the clinical study, which are then used to demonstrate substantial equivalence to a predicate device. The comparison table below uses the clinical study results as "reported device performance."

Note: The predicate device comparison table focuses on "Performance Characteristics" as opposed to explicit numerical acceptance criteria. The sensitivity and specificity values observed in the clinical trial would implicitly serve as the benchmark for demonstrating that the device performs as intended and is as safe and effective as the predicate device.

Performance MetricAcceptance Criteria (Implicit from Predicate & Guidance)Reported Device Performance (Solana® Trichomonas Assay)Specimen Type: Symptom Status
Vaginal Swabs
Sensitivity~98.6% - 100% (from Predicate)100% (92.9 to 100)Asymptomatic
98.6% (92.3 to 99.7)Symptomatic
Specificity~97.9% - 98.3% (from Predicate)98.9% (97.4 to 99.5)Asymptomatic
98.5% (97.0 to 99.3)Symptomatic
Female Urine Specimens(No predicate for urine data provided)
SensitivityN/A98.0% (89.5 to 99.6)Asymptomatic
92.9% (84.3 to 96.9)Symptomatic
SpecificityN/A98.4% (96.8 to 99.2)Asymptomatic
97.9% (96.2 to 98.8)Symptomatic

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set:
    • Vaginal Swabs: 1043 subjects (after re-testing of invalid results).
    • Female Urine Specimens: 1044 subjects (after re-testing of invalid results).
    • Reproducibility Study (Analytical Performance): 90 samples per category per workflow (swab/urine) across 3 sites (e.g., 90 Low Positive swab samples, 90 Negative swab samples, etc.). This totals to 540 swab samples and 540 urine samples for reproducibility.
  • Data Provenance:
    • Country of Origin: United States
    • Retrospective or Prospective: Prospective. The clinical study was performed from November 2015 through March 2016, and specimens were obtained from each subject after informed consent.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not specify the number of experts or their qualifications used to establish the ground truth beyond stating that the reference methods for the clinical study were "Wet Mount" and "InPouch TV Culture." These are laboratory-based diagnostic tests, typically performed by trained medical technologists or laboratory personnel, not necessarily "experts" in the sense of physicians or specialists establishing a consensus diagnosis.


4. Adjudication Method for the Test Set

  • Clinical Study: The adjudication method for the clinical study was based on a composite reference method. A specimen was considered positive if either the Wet Mount or the InPouch TV Culture was positive.
  • Discordant Analysis: For specimens where the Solana Assay results differed from the composite reference method, an FDA-cleared Trichomonas vaginalis molecular assay was used for further testing (referred to as "discordant testing").

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, a MRMC comparative effectiveness study was not explicitly mentioned or performed in the context of human readers improving with AI vs. without AI assistance. This device is a diagnostic assay (an in vitro diagnostic test) that provides qualitative results directly, not an AI-assisted interpretation by human readers. The clinical study evaluated the device's standalone performance compared to established clinical reference methods.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, a standalone performance evaluation was done. The Solana® Trichomonas Assay is an automated in vitro diagnostic test that uses "on-board method-specific algorithms" to measure and interpret fluorescent signals and report test results. The clinical study assessed the performance of this device as an algorithm-only system against the composite reference methods.

7. The Type of Ground Truth Used

  • Clinical Study: The ground truth for the clinical study was established using a composite reference method consisting of:
    • Wet Mount
    • InPouch TV Culture
      A specimen was considered positive if either of these reference tests was positive.
  • Analytical Sensitivity (LoD) and Inclusivity Studies: Ground truth was based on quantified strains of T. vaginalis (trophozoite/mL) at known concentrations.

8. The Sample Size for the Training Set

  • The document does not specify a separate "training set" sample size for the Solana® Trichomonas Assay. This is typical for in vitro diagnostic assays of this nature, where the "training" (development and optimization) of the assay's reagents and algorithms is usually an iterative laboratory process, not based on a distinct clinical training dataset to the same extent as, for example, a machine learning algorithm for image analysis. The clinical evaluation primarily serves as a validation set.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, a distinct "training set" with established ground truth in a clinical context is not explicitly described.
  • For analytical studies (e.g., LoD, inclusivity): Ground truth was established by using quantified strains of T. vaginalis at known concentrations (e.g., trophozoite/mL) diluted in negative clinical matrix. These known concentrations serve as the "ground truth" for evaluating analytical performance parameters. Development and optimization of the assay would rely on such controlled laboratory experiments.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol representing the human form, depicted as three overlapping profiles facing to the right.

Public Health Service

August 15, 2016

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Quidel Corporation Ronald H. Lollar Senior Director, Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, OH 45701

Re: K161182

Trade/Device Name: Solana® Trichomonas Assay Regulation Number: 21 CFR 866.3860 Regulation Name: Trichomonas vaginalis nucleic acid assay Regulatory Class: II Product Code: OUY Dated: July 14, 2016 Received: July 15, 2016

Dear Mr. Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

Page 2 - Mr. Lollar

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Steven R. Gitterman -S

for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161182

Device Name Solana® Trichomonas Assay

Indications for Use (Describe)

The Solana® Trichomonas Assay is an in vitro diagnostic test, using isothermal amplification technology (helicasedependent amplification, HDA), for the qualitative detection of Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swabs and female urine specimens obtained from symptomatic females to aid in the diagnosis of trichomoniasis. The Solana® Trichomonas Assay is intended for use only with the Solana® instrument.

Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 - Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

April 26, 2016

A. 510(k) Number:

K161182

B. Purpose for Submission:

To obtain substantial equivalence for the Solana Trichomonas Assay when performed on the Solana

C. Measurand:

Repeated DNA fragment located in T. vaginalis genome

D. Type of Test:

Helicase-dependent amplification (HDA)

{4}------------------------------------------------

E. Applicant:

Quidel Corporation

F. Proprietary and Established Names:

Solana Trichomonas Assay

G. Regulatory Information:

Product CodeClassificationRegulation SectionPanel
OUY –Trichomonas vaginalis nucleicacid amplificationtest systemClass II21 CFR 866.3860 Trichomonas vaginalis nucleic acid assayMicrobiology (83)

H. Intended Use:

1. Intended Use(s):

The Solana® Trichomonas Assay is an in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection of Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swabs and female urine specimens obtained from symptomatic or asymptomatic females to aid in the diagnosis of trichomoniasis. The Solana® Trichomonas Assay is intended for use only with the Solana® Instrument.

2. Indication(s) for Use:

Same as Intended Use.

3. Special conditions for use statement(s):

  • For in vitro diagnostic use only
  • For prescription use only
    1. Special instrument requirements:

Solana® Instrument

{5}------------------------------------------------

l. Device Description:

The Solana® Trichomonas Assay amplifies and detects Trichomonas vaginalis nucleic acids present in clinician-collected vaginal swabs and urine specimens from symptomatic and asymptomatic women. The assay targets a conserved multi-copy sequence of the T. vaqinalis DNA.

The vaginal swab is eluted in a swab lysis tube or a urine specimen is added to a urine lysis tube, and the cells in either specimen type are lysed by simple heat treatment. After heat treatment, an aliquot of the lysed specimen is transferred into a dilution tube. An aliquot of the diluted sample is added to a reaction tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of T. vaginalis-specific target sequence. In Solana, the target sequence is amplified by T. vaginalis specific primers and detected by a T. vaginalis specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by T. vaginalis specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to T. vaqinalis or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and the results can be printed via a printer.

Materials Provided:

Solana® Trichomonas Assay Kit: M304.S for swab testing or M304.U for urine testing

Solana® Trichomonas Assay Kit – M304.S
ComponentQuantityStorage
Lysis Buffer Tubes48 tubes/kit, 1.0 mL2° to 8°C
Dilution Tubes48 tubes/kit, 1.5 mL2° to 8°C
Reaction Tubes48 tubes/kit2° to 8°C
Solana® Trichomonas Assay Kit - M304.U
ComponentQuantityStorage
Lysis Buffer Tubes48 tubes/kit, 0.2 mL2° to 8°C
Dilution Tubes48 tubes/kit, 1.5 mL2° to 8°C
Reaction Tubes48 tubes/kit2° to 8°C

48 Tests per kit

{6}------------------------------------------------

Materials required but not provided:

  • External controls for Trichomonas vaginalis (e.g. Quidel Molecular Trichomonas Assay Control Kit: M119), which contains positive and negative controls. This positive control contains intact non-viable, trophozoites and has been titered to be near the limit of detection for the assay. This negative control is the same matrix as the positive control, but is trophozoite-free. These controls serve as an external processing and extraction control
  • Sterile DNAse-free filter-blocked or positive displacement micropipettor tips
  • Micropipettor
  • BD BBL™ CultureSwab™ collection and transport device ●
  • . Stopwatch or timer
  • Scissors or a blade
  • Micro tube tray
  • Heat block capable of 95° C ± 2° C temperature ●
  • Thermometer
  • Solana Instrument ●

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      AmpliVue Trichomonas Assay
    1. Predicate 510(k) number(s): K143329

3. Comparison with predicate:

Similarities
ItemSolana® Trichomonas Assay(K161182)AmpliVue® Trichomonas Assay(K143329)
Intended UseThe Solana® Trichomonas Assay isan in vitro diagnostic test, usingisothermal amplificationtechnology (helicase-dependentamplification, HDA), for thequalitative detection ofTrichomonas vaginalis nucleic acidsThe AmpliVue® TrichomonasAssay is an in vitro diagnostictest, uses isothermalamplification technology(helicase-dependentamplification, HDA) for thequalitative detection of

{7}------------------------------------------------

Solana® Trichomonas Assay 7/26/2016 Page 5 of 21

510(k) Summary

Similarities
ItemSolana® Trichomonas Assay(K161182)AmpliVue® Trichomonas Assay(K143329)
isolated from clinician-collected vaginal swabs and female urine specimens obtained from symptomatic or asymptomatic females to aid in the diagnosis of trichomoniasis. The Solana® Trichomonas Assay is intended for use only with the Solana® instrument.Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swab specimens obtained from symptomatic or asymptomatic females to aid in the diagnosis of trichomoniasis.
TargetSequenceDetectedRepeated DNA fragment located in T. vaginalis genomeSame
AmplificationTechnologyHelicase-dependent amplification (HDA)Same
Differences
ItemSolana® Trichomonas Assay(K161182)AmpliVue® TrichomonasAssay (K143329)
Sample TypesClinician-collected VaginalSwabs, Female UrineClinician-collected VaginalSwabs
Self-ContainedSystem Assay aftersample preparationYesNo
Detection TechniqueAutomatedManual
InstrumentSolanaNone
PerformanceClinician-collected VaginalSpecimens: AsymptomaticClinician-collected VaginalSpecimens: Asymptomatic

{8}------------------------------------------------

Differences
ItemSolana® Trichomonas Assay(K161182)AmpliVue® TrichomonasAssay (K143329)
CharacteristicsSensitivity 100% (95% CI 92.9 –100)Specificity 98.9% (95% CI 97.4 –99.5)SymptomaticSensitivity 98.6% (95% CI 92.3 –99.7)Specificity 98.5% (95% CI 97.0 –99.3)Female Urine Specimens:AsymptomaticSensitivity 98% (95% CI 89.5 –99.6)Specificity 98.4% (95% CI 96.8 –99.2)SymptomaticSensitivity 92.9% (95% CI 84.3 –96.9)Specificity 97.9% (95% CI 96.2 –98.8)Sensitivity 100% (95% CI94.1 – 100)Specificity 98.3% (95% CI96.9 – 99.1)SymptomaticSensitivity 100% (95% CI93.9 – 100)Specificity 97.9% (95% CI95.5 – 99.0)

K. Standard/Guidance Document Referenced (if applicable):

Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection of Trichomonas vaginalis

L. Test Principle:

The vaginal swab is eluted in a swab lysis tube or a urine specimen is added to a urine lysis tube, and the cells in either specimen type are lysed by simple heat treatment. After heat treatment, an aliquot of the lysed specimen is transferred into a dilution tube. An aliquot of the diluted sample is added to a reaction tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of T. vaginalis-specific target sequence. In Solana, the target sequence is amplified by T. vaginalis specific primers

{9}------------------------------------------------

and detected by a T. vaqinalis specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by T. vaginalis specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to T. vaqinalis or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and the results can be printed via an attached printer.

M. Performance Characteristics:

1. Analytical performance:

a. Precision/Reproducibility:

Reproducibility

In order to confirm the reproducibility of the Solana Trichomonas Assay a blinded and randomized study consisting of the following four-member panel containing Trichomonas vaginalis positive and negative samples was performed: swab workflow - moderate positive (3x LoD), low positive (1x LoD), high negative (1/54x LoD), and a negative sample; urine workflow - moderate positive (3x LoD), low positive (1x LoD), high negative (1/27x LoD), and a negative sample. The testing was performed using appropriate swab or urine workflows at three (3) test sites (one inhouse laboratory and two (2) clinical sites). Each site tested a reproducibility panel and Assay Controls for five (5) days in triplicate. Testing was done by two (2) operators at each site. Each operator ran the panel once a day using one lot of Solana Trichomonas Assay. The testing was performed using both the swab and urine workflows.

SITE
CategorySwabWorkflowSite #1Site #2Site #3Overall PercentAgreement95%ConfidenceInterval
#expectedresults/#tested%Agreement95%ConfidenceInterval#expectedresults/#tested%Agreement95%ConfidenceInterval#expectedresults/#tested%Agreement95%ConfidenceInterval
High Negative(1.89trophozoites/mL)25/3083%66.4% to92.7%22/3073%55.6% to85.8%15/3050%33.2%to66.8%62/9069%58.7% to77.5%

{10}------------------------------------------------

Solana® Trichomonas Assay 7/26/2016 Page 8 of 21

510(k) Summary

SITESITESITEOverall PercentAgreement95%ConfidenceInterval
CategorySwabWorkflow#expectedresults/#tested%Agreement95%ConfidenceInterval#expectedresults/#tested%Agreement95%ConfidenceInterval#expectedresults/#tested%Agreement95%ConfidenceInterval
Low Positive(102trophozoites/mL)30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6%to100%90/9095.9% to100%
ModeratePositive(306trophozoites/mL)30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6%to100%90/9095.9% to100%
Negative30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6%to100%90/9095.9% to100%
PositiveControl30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6%to100%90/9095.9% to100%
NegativeControl30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6%to100%90/9095.9% to100%
SITE
CategoryUrineWorkflow#expectedresults/#tested%Agreement95%ConfidenceInterval#expectedresults/#tested%Agreement95%ConfidenceInterval#expectedresults/#tested%Agreement95%ConfidenceIntervalOverall PercentAgreement95%ConfidenceInterval
High Negative(0.2trophozoites/mL)20/3067%48.8% to80.8%19/3063%45.5% to78.1%22/3073%55.6% to85.8%61/9068%57.6% to75.5%
Low Positive(4 trophozoites/mL)30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6% to100%90/90100%95.9% to100%
ModeratePositive(12trophozoites/mL)30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6% to100%90/90100%95.9% to100%
Negative30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6% to100%90/90100%95.9% to100%
PositiveControl30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6% to100%90/90100%95.9% to100%
NegativeControl30/30100%88.6% to100%30/30100%88.6% to100%30/30100%88.6% to100%90/90100%95.9% to100%

{11}------------------------------------------------

The results suggest that there are no significant differences between different users and different sites on different days using either the swab or the urine workflow. Reproducibility studies are acceptable.

  • b. Linearity/assay reportable range:
    Not applicable - This assay is qualitative.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Traceability:

Not applicable. This assay is qualitative.

Specimen Stability:

A study was performed to determine the stability of swab and urine samples when stored at 2 to 8°C or 30°C ± 2°C. Contrived positive swab and urine samples (2x LoD) were prepared. The testing time points for the specimens stored at 2° to 8°C were 0, 24, 48, 72, 73 hours, 5, 7, 8 days. The testing time points for the specimens stored at 30°C ± 2°C were 0, 2.5, 6.5, 24, 25, 27, 48, and 49 hours.

Based on the data generated by this study, swab specimens may be stored between 2° to 8°C for up to 7 days prior to testing. Swab specimens may be stored at up to 48-hours at 30°C ± 2°C or below prior to testing.

Based on the data generated by this study, urine specimens may be stored between 2° to 8°C for up to 7 days prior to testing. Urine specimens may be stored at up to 24-hours at 30°C ± 2°C or below prior to testing.

Controls:

Controls (Quidel Molecular Trichomonas Control Set #M119, which contains positive and negative controls, serves as an external processing and extraction control) were run on the Solana * Trichomonas Assay each day of testing. These controls are described as follows:

  • a. The internal control is used to detect HDA inhibitory specimens and to confirm the integrity of assay reagents and the operation of the Solana instrument. The internal control is included in the lysis tube.

{12}------------------------------------------------

  • b. External assay positive control (e.g. Quidel Molecular Trichomonas Control Set (M119)) serves as the assay positive control. Transfer 50 µL of positive control into a labeled lysis buffer tube and proceed with processing. The external assay positive control is intended to monitor substantial reagent and instrument failure.
  • External assay negative control (e.g. Quidel Molecular Trichomonas Control ﻥ Set (M119)) serves as the assay negative control. Transfer 50 µL of negative control into a labeled lysis buffer tube and proceed with processing. The external assay negative control is intended to detect reagent or environment contamination or carry-over by either T. vaginalis DNA or amplicons.
  • d. Detection limit:

The analytical sensitivity (limit of detection or LoD) of the Solana Trichomonas Assay was determined using quantified (trophozoite/mL) strains of two (2) T. vaginalis strains, one metronidazole-susceptible G3 and one metronidazole-resistant CDC888 serially diluted in negative clinical matrix. The LOD is defined as the lowest concentration at which 95% of all replicates tested positive.

The strains were freshly grown and quantified using a hemocytometer. The cells were serially diluted in negative clinical matrix to three (3) concentrations at 3×, 1×, and 1/3× in the preliminary LOD determination study.

LoD was confirmed by testing each reference strain with 20 replicates on three reagent lots in the negative vaginal swab and negative urine matrixes.

Trichomonas vaginalisreference strainSwab Workflow LoDUrine Workflow LoD
G3102 trophozoite /mL4 trophozoite /mL
CDC888306 trophozoite /mL108 trophozoite /mL
  • e. Analytical specificity:

Cross Reactivity:

A study was performed to evaluate the cross-reactivity of the Solana® Trichomonas Assay in the presence of forty-seven (47) microorganisms (37 bacteria, 4 yeast, 4 viruses, 2 parasite) potentially found in specimens collected to test for Trichomonas

{13}------------------------------------------------

vaginalis infection. Each microorganism was diluted in either swab negative matrix or negative urine matrix to the desired concentration (10° or higher CFU/mL or copies/mL for bacteria, yeast or DNA/RNA and 105 or higher pfu/mL or TCID50/mL for viruses). The strains included in the cross-reactive study are shown in the table below.

MicroorganismStock ConcentrationMicroorganismStock Concentration
Acinetobacter Iwoffi1.0×106 CFU/mLHerpes simplex virus I1.0×105 TCID50/mL
Actinomyces israelii1.0×106 CFU/mLHerpes simplex virus II1.0×105 TCID50/mL
Atopobium vaginae1.0×106 CFU/mLKlebsiella oxytoca1.0×106 CFU/mL
Bacteroides fragilis1.0×106 CFU/mLLactobacillus acidophilus1.0×106 CFU/mL
Bifidobacterium adolescentis1.0×106 CFU/mLLactobacillus jensenii1.0×106 CFU/mL
Campylobacter jejuni1.0×106 CFU/mLLactobacillus vaginalis1.0×106 CFU/mL
Candida albicans1.0×106 CFU/mLListeria monocytogenes1.0×106 CFU/mL
Candida glabrata1.0×106 CFU/mLMobiluncus curtisii1.0×106 CFU/mL
Candida parapsilosis1.0×106 CFU/mLMycoplasma hominis1.0×106 CFU/mL
Candida tropicalis1.0×106 CFU/mLNeisseria gonorrhoeae1.0×106 CFU/mL
Chlamydia trachomatis1.0×106 CFU/mLPentatrichomonas hominis1.0×106 CFU/mL
Clostridium difficile1.0×106 CFU/mLPrevotella bivia1.0×106 CFU/mL
Clostridium perfingens1.0×106 CFU/mLPropionibacterium acnes1.0×106 CFU/mL
Corynebacterium genitalium1.0×106 CFU/mLProteus mirabilis1.0×106 CFU/mL
Cryptococcus neoformans1.0×106 CFU/mLPseudomonas aeruginosa1.0×106 CFU/mL
Enterobacter aerogenes1.0×106 CFU/mLStaphylococcus aureus (MRSA)1.0×106 CFU/mL
Enterococcus faecalis1.0×106 CFU/mLStaphylococcus epidermidis1.0×106 CFU/mL
Escherichia coli1.0×106 CFU/mLStreptococcus pyogenes1.0×106 CFU/mL
Fusobacterium nucleatum1.0×106 CFU/mLStreptococcus agalactiae1.0×106 CFU/mL
Gardnerella vaginalis1.0×106 CFU/mLTrichomonas tenax1.0×106 CFU/mL
MicroorganismStockConcentrationMicroorganismStockConcentration
Haemophilus ducreyi1.0×106 copies/mLEnterobacter cloacae1.0×106 CFU/mL
HIV-1 Subtype B RNA1.0×105 RNAcopies/mLHPV 16 (SiHa)1.0×105 copies/mL
Peptostreptococcus anaerobius1.0×106 copies/mLUreaplasma urealyticum DNA1.23 x108 cp/mL
Synthetic Mycoplasma genitalium DNA*1.0×106 copies/mL

{14}------------------------------------------------

  • Preparation includes fragments from the 16S gene, mgpA, and gap.

No cross-reactivity was seen with the Solana "Trichomonas Assay with any of fortyseven (47) microorganisms (37 bacteria, 4 yeast, 4 viruses, 2 parasite) tested.

Interference:

Vaginal Swab Specimens

A study was conducted to determine if the Solana "Trichomonas assay is inhibited in the presence of a panel of fourteen (14) substances potentially present in vaginal swab specimens collected to test for Trichomonas vaginalis infection. Each of the potential interfering substances was tested in three (3) replicates in the presence and absence of near LOD (2x) levels of two (2) strains of Trichomonas vaginalis in the Solana Trichomonas Assay. Substances were introduced into the assay at concentrations which were medically relevant.

ClassSubstancesConcentration Tested
BloodWhole blood with EDTA10% (v/v)
Seminal fluidSeminal fluid1% (v/v)
MucusMucin from Porcine Stomach1% (w/v)
Over the counter(OTC) vaginalproducts andcontraceptivesK-Y Personal Lubricant Jelly1% (w/v)
Ortho Options Gynol II Extra Strength Vaginal Contraceptive Jelly1% (w/v)
Summer's Eve Ultra Extra Strength Feminine Deodorant Spray1% (w/v)
Vagisil Creme Maximum Strength1% (w/v)

{15}------------------------------------------------

ClassSubstancesConcentration Tested
CVS Vinegar & Water Extra CleansingDisposable Douche (Glacial acetic acid)1% (v/v)
Summer's Eve Douche, Medicated1% (v/v)
IntravaginalHormonesEstradiol1% (w/v)
Hemerrhoidal CreamPreparation H1% (w/v)
LeukocytesLeukocytes106 cells/mL
Prescription vaginaltreatmentsAcyclovir (Acycloguanosine)0.05% (w/v) (1% of activeingredient of Zoviraxcream with Acyclovir at5%)
Metronidazole0.0075% (w/v) (1% ofactive ingredient ofVandazole gel withMetronidazole at 0.75%)

None of the substances tested interfered with the detection of either strain of 2x LoD Trichomonas vaginalis, or the detection of the internal control in negative specimens.

Urine Specimens

A study was conducted to determine if the Solana Trichomonas assay is inhibited in the presence of a panel of seventeen (17) substances potentially present in urine specimens collected to test for Trichomonas vaginalis infection. Each of the potential interfering substances was tested in three (3) replicates in the presence and absence of near LoD (2x) levels of two (2) strains of Trichomonas vaginalis in the Solana "Trichomonas Assay. Substances were introduced into the assay at concentrations which were medically relevant.

ClassSubstancesConcentration Tested
BloodWhole blood with EDTA1% (v/v)
Seminal fluidSeminal fluid5% (v/v)
MucusMucin from Porcine Stomach1% (w/v)
AZO Standard Urinary Relief Tablets(Phenazopyridine Hydrochloride)1.0 mg/mL

{16}------------------------------------------------

ClassSubstancesConcentration Tested
Analgesics &AntibioticsAcetylsalicylic Acid8 mg/mL
Acetaminophen3.2 mg/mL
Azithromycin1.0 mg/mL
Doxycycline0.5 mg/mL
Over the counterdeodorant spray andpowderSummer's Eve Feminine DeodorantPowder1% (w/v)
Summer's Eve Feminine Deodorant Spray1% (w/v)
AlbuminHuman Albumin10 mg/ml
GlucoseGlucose10 mg/ml
BilirubinBilirubin1 mg/ml
Acidic Urine (pH 4.0)Urine + N-Acetyl-L-CysteinepH 4.0
Alkaline Urine (pH 9.0)Urine + Ammonium Citrate & SodiumhydroxidepH 9.0
LeukocytesLeukocytes$10^6$ cells/mL
IntravaginalHormonesEstradiol1% (w/v)

None of the substances tested interfered with the detection of either strain of 2x LOD Trichomonas vaginalis, or the detection of the internal control in negative specimens.

Analytical Reactivity (Inclusivity):

A study was performed to verify the in silico inclusivity results with functional testing of the Solana Trichomonas Assay using twenty (20) additional strains of Trichomonas vaginalis tested in triplicate at concentrations at or near the 1x LoD levels of the assay in both the swab and urine workflows.

Bacterial StrainStrain Detected(Yes/No)Strain Detected(Yes/No)
CDC899YesYes
CDC938YesYes
CDC963YesYes
CDC1031YesYes

{17}------------------------------------------------

Bacterial StrainStrain Detected(Yes/No)Strain Detected(Yes/No)
CDC1256YesYes
PMGH25YesYes
BUSH20YesYes
CDC911YesYes
MOR31YesYes
CDC1080YesYes
B7708/1839YesYes
F1623YesYes
CDC1095YesYes
SD1YesYes
SA-384YesYes
CDC948YesYes
SD10YesYes
SA-A53YesYes
CDC1230YesYes
SA-A19YesYes

The twenty (20) additional strains of Trichomonas vaginalis tested in triplicate at concentrations near the LoD levels of the assay were detected in both the swab and urine workflows of the Solana Trichomonas assay.

  • f. Assay cut-off:
    Not applicable.

    1. Comparison studies:
    • a. Method comparison with predicate device:

Not applicable

  • b. Matrix comparison:
    Not applicable

    1. Clinical studies:

{18}------------------------------------------------

Clinical Sensitivity: a.

A multi-center study was performed to evaluate Solana Trichomonas Assay using one thousand forty-four (1044) clinician-collected vaginal swab and urine specimens obtained from asymptomatic (n=501) or symptomatic (n=543) patients. The clinician categorized the patients as symptomatic or asymptomatic at the of specimen collection. The study was performed November 2015 through March 2016 at four (4) locations in the United States. Specimens were obtained from each subject after informed consent was obtained. The study was conducted in accord with the Health Insurance Portability and Accountability Act (HIPAA).

Vaginal Swab

For each subject, three (3) vaginal specimens were collected using polyester or rayon Swabs w/ liquid Stuart's, and one (1) vaginal specimen was collected with a collection swab from a FDA-cleared molecular device. The four (4) clinician collected vaginal swabs were used for reference and Solana testing. The first two (2) polyester/rayon swabs were randomized, one swab was tested for the Wet Mount (reference method) and the other swab was used for the InPouch TV Culture (reference method). The third swab was used for testing the Solana Trichomonas Assay. The FDA-cleared molecular device collection swab was used for discordant testing.

All sensitivity and specificity calculations were based on a composite reference method of Wet Mount and InPouch TV culture. A specimen was considered positive if either test was positive.

One thousand forty-four (1044) clinician-collected vaginal swab specimens obtained from asymptomatic (n=501) or symptomatic (n=543) patients were tested by the composite reference method and the Solana Trichomonas Assay. Ten (10) specimens generated invalid results upon initial testing with the Solana Trichomonas Assay (0.96%). These specimens were re-tested according to the instructions provided in the package insert. Nine (9) of these specimens generated valid results upon re-testing (6 negative and 3 positive results), and one (1) specimen generated a second invalid result (0.1%). The table below shows the sensitivity, specificity, PPV, and NPV of the Solana Trichomonas Assay and the prevalence of T. vaginalis (by asymptomatic, symptomatic clinician designations and combined) for the remaining one thousand forty-three (1043) subjects.

Performance Characteristics of the Solana Trichomonas Assay with Vaginal Swabs by Symptom Status compared to the Composite Reference Method
Site NumberSymptom StatusNTPFPTNFNPrev%Sensitivity%(95% CI)Specificity%(95% CI)PPV %(95% CI)NPV %(95% CI)

{19}------------------------------------------------

Solana Trichomonas Assay 7/26/2016 Page 17 of 21

510(k) Summary

Site NumberSymptom StatusNTPFPTNFNPrev%Sensitivity% (95% CI)Specificity% (95% CI)PPV % (95% CI)NPV % (95% CI)
CombinedAsymptomatic501505446010.0100 (92.9 to 100)98.9 (97.4 to 99.5)90.9 (80.4 to 96.1)100 (99.1 to 100)
Symptomatic542697465112.998.6 (92.3 to 99.7)98.5 (97.0 to 99.3)90.8 (82.2 to 95.5)99.8 (98.8 to 100)
All104311912*9111*11.599.2 (95.4 to 99.9)98.7 (97.7 to 99.3)90.8 (84.7 to 94.7)99.7 (99.4 to 100)
Site 1Asymptomatic77707009.1100 (64.6 to 100)100 (94.8 to 100)100 (64.6 to 100)100 (94.8 to 100)
Symptomatic27212407.4100 (34.2 to 100)96.0 (80.5 to 99.3)66.7 (20.8 to 93.9)100 (86.2 to 100)
All104919408.7100 (70.1 to 100)98.9 (94.3 to 99.8)90.0 (59.6 to 98.2)100 (96.1 to 100)
Site 2Asymptomatic10813095012.0100 (77.2 to 100)100 (96.1 to 100)100 (77.2 to 100)100 (96.1 to 100)
Symptomatic213372174017.4100 (90.6 to 100)98.9 (80.5 to 99.3)94.9 (83.1 to 98.6)100 (97.8 to 100)
All321502269015.6100 (92.9 to 100)99.3 (97.3 to 99.8)96.2 (87.0 to 98.9)100 (98.6 to 100)
Site 3Asymptomatic146191126013.0100 (83.2 to 100)99.2 (95.7 to 99.9)95.0 (76.4 to 99.1)100 (97.0 to 100)
Symptomatic679157013.4100 (70.1 to 100)98.3 (90.9 to 99.7)90.0 (59.6 to 98.2)100 (93.7 to 100)
All213282183013.1100 (87.9 to 100)98.9 (96.1 to 99.7)85.9 (76.0 to 92.2)100 (99.3 to 100)
Site 4Asymptomatic17011415506.5100 (74.1 to 100)97.5 (93.7 to 99.0)73.3 (48.0 to 89.1)100 (97.6 to 100)
Symptomatic23521321019.495.5 (78.2 to 99.2)98.6 (95.9 to 99.5)87.5 (69.0 to 95.7)99.5 (97.4 to 99.9)
All40532736518.197.0 (84.7 to 99.5)98.1 (96.2 to 99.1)82.1 (67.3 to 91.0)99.7 (98.5 to 100)

*Of the one thousand forty-three 1043 specimens evaluated a total of thirteen (13) specimens were discordant. Of the twelve (12) discordant (Solana Positive/Composite Reference Method Negative) specimens, four (4) were positive by a FDA-cleared Trichomonas vaginalis molecular assay. The one (1) discordant (Solana Negative/Composite Reference Method Positive) specimen was negative by a FDA-cleared Trichomonas vaginalis molecular assay.

Urine

One thousand forty-four (1044) first catch urine specimens obtained from asymptomatic (n=501) or symptomatic (n=543) patients were tested by the Solana Trichomonas Assay. Five (5) specimens generated invalid results upon initial testing with the Solana Trichomonas Assay (0.5%). These specimens were re-tested according to the instructions provided in the package insert. All five (5) of these specimens generated valid results upon re-testing (four (4) negative and one (1) positive result). The table below shows the sensitivity, specificity, PPV, and NPV of the Solana Trichomonas Assay and the prevalence of T. vaginalis (by asymptomatic, symptomatic clinician designations and combined) for the one thousand forty-four (1044) subjects when compared to the corresponding Wet Mount and the InPouch TV Culture results.

{20}------------------------------------------------

Site NumberSymptom StatusNTPFPTNFNPrev%Sensitivity%(95% CI)Specificity%(95% CI)PPV %(95% CI)NPV %(95% CI)
CombinedAsymptomatic501497444110.098.0(89.5 to 99.6)98.4(96.8 to 99.2)87.5(76.4 to 93.8)99.8(98.7 to 100)
Symptomatic5436510463512.992.9(84.3 to 96.9)97.9(96.2 to 98.8)86.7(77.2 to 92.6)98.9(97.5 to 99.5)
All104411417907611.595.0(89.5 to 97.7)98.2(97.1 to 98.8)87.0(80.2 to 91.7)99.3(98.6 to 99.7)
Site 1Asymptomatic77607019.185.7(48.7 to 97.4)100(94.8 to 100)100(61.0 to 100)98.6(92.4 to 99.8)
Symptomatic27232207.4100(34.2 to 100)88.0(70.0 to 95.8)40.0(11.8 t 76.9)100(85.1 to 100)
All104839218.788.9(56.5 to 98.0)96.6(91.1 to 98.9)72.7(43.4 to 90.3)98.9(94.2 to99.8)
Site 2Asymptomatic10813095012.0100(77.2 to 100)100(96.1 to 100)100(77.2 to 100)100(96.1 to 100)
Symptomatic214354173217.394.6(82.3 to 98.5)97.7(94.3 to 99.1)89.7(76.4 to 95.9)98.9(95.9 to 99.7)
All322484268215.596.0(86.5 to 98.9)98.5(96.3 to 99.4)92.3(81.8 to 97.0)99.3(97.3 to 99.8)
Site 3Asymptomatic146191126013.0100(83.2 to 100)99.2(95.7 to 99.9)95.0(76.4 to 99.1)100(97.0 to 100)
Symptomatic679058013.4100(70.1 to 100)100(93.8 to 100)100(70.1 to 100)100(93.8 to 100)
All213281184013.1100(87.9 to 100)99.5(97.0 to 99.9)96.6(82.8 to 99.4)100(97.9 to 100)
Site 4Asymptomatic17011615306.5100(74.1 to 100)96.2(92.0 to 98.3)64.7(41.3 to 82.7)100(97.6 to 100)
Symptomatic23519321039.486.4(66.7 to 95.3)98.6(95.9 to 99.5)86.4(66.7 to 95.3)99.5(97.4 to 99.9)
All40530936338.190.9(76.4 to 96.9)97.6(95.5 to 98.7)76.9(61.7 to 87.4)99.2(97.6 to 99.7)

b. Clinical specificity:

See Section 3a.

Other clinical supportive data (when a. and b. are not applicable): C.

Not applicable

4. Clinical cut-off:

Not applicable

    1. Expected values:

{21}------------------------------------------------

The prevalence of T. vaginalis (by asymptomatic clinician designations and combined) detected by the Solana Trichomonas Assay in the multi-center study was calculated and is provided in the table below.

Study Prevalence
Swab specimens
Symptom StatusCombinedSite 1Site 2Site 3Site 4
Asymptomatic10.0%9.1%12.0%13.0%6.5%
Symptomatic12.9%7.4%17.4%13.4%9.4%
Combined11.5%8.7%15.6%13.1%8.1%
Urine specimens
Symptom StatusCombinedSite 1Site 2Site 3Site 4
Asymptomatic10.0%9.1%12.0%13.0%6.5%
Symptomatic12.9%7.4%17.3%13.4%9.4%
Combined11.5%8.7%15.5%13.1%8.1%

Positive and Negative Predictive Values

The estimated positive predictive value (PPV) and negative predictive value (NPV) of the Solana Trichomonas Assay across different hypothetical prevalence rates are shown in the table below. These calculations are based on the overall estimated sensitivity and specificity for clinician-collected vaginal swab specimens in the Solana Trichomonas Assay clinical study.

Hypothetical PPV and NPV of the Solana Trichomonas Assay with clinician-collected vaginal swab specimens
Prevalence %PPV (%)NPV (%)
143.5100
260.9100
580.1100
1089.599.9
1593.199.9
2095.099.8
2596.299.7

The estimated positive predictive value (PPV) and negative predictive value (NPV) of the Solana Trichomonas Assay across different hypothetical prevalence rates are shown in the table below. These calculations are based on the overall estimated sensitivity and specificity for urine specimens in the Solana Trichomonas Assay clinical study.

{22}------------------------------------------------

Hypothetical PPV and NPV of the Solana Trichomonas Assay with Urinespecimens
Prevalence %PPV (%)NPV (%)
138.0100
255.1100
576.699.9
1086.699.8
1591.799.6
2094.099.5
2595.499.3

N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Instrument: Solana Instrument

O. System Descriptions:

1. Modes of Operation:

The Solana instrument heats each reaction tube to 64°C. If present, the target sequence is amplified by T. vaginalis specific primers and detected by a T. vaginalis specific fluorescence probe included in the Reaction Tube. The target probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target probes carry a ribonucleic acid. Upon annealing to T. vaqinalis amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. The Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and the results can be printed via a printer.

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes No

P. Proposed Labeling:

{23}------------------------------------------------

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and the special controls.

§ 866.3860

Trichomonas vaginalis nucleic acid assay.(a)
Identification. ATrichomonas vaginalis nucleic acid assay is a device that consists of primers, probes, enzymes, and controls for the amplification and detection of trichomonas nucleic acids in endocervical swabs, vaginal swabs, and female urine specimens, from women symptomatic for vaginitis, cervicitis, or urethritis and/or to aid in the diagnosis of trichomoniasis in asymptomatic women. The detection of trichomonas nucleic acids, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of trichomoniasis caused byTrichomonas vaginalis .(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection ofTrichomonas vaginalis; Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.