(77 days)
The PASS OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radios, and degenerative disease of the facets with instability. The PASS OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the PASS OCT Spinal System may be connected to the PASS LP Spinal System rods with the dual diameter rods or dominos. Refer to the PASS LP Spinal System package insert for a list of the PASS LP Spinal System indications of use.
The PASS OCT Spinal System is a posterior system. which consists of a variety of shapes and sizes of rods, hooks, polyaxial screws, occipital plates, occipital bone screws, and connection components, which can be rigidly locked to the rod in a variety of configurations. See package insert of the system for labeling limitations.
The implants are manufactured in titanium alloy Ti-6AI-4V ELI conforming to ISO 5832-3 specifications and ASTM F136 specifications, in PEEK OPTIMA LT1 conforming to ASTM F2026 specifications and in cobalt-chromium molybdenum alloy Co-Cr28Mo6 that conforms to ISO 5832-12 and ASTM F1537 specifications, and also CP Titanium according to ASTM F67 and ISO 5832-2 specifications.
The purpose of this 510(k) submission is to add offset connectors to the PASS OCT Spinal System.
This FDA 510(k) summary describes a spinal implant system, not an AI/ML powered device. Therefore, many of the requested elements for describing the acceptance criteria and study proving device performance are not applicable.
However, I can extract the relevant information regarding the non-clinical testing performed for this medical device.
Device Name: PASS OCT Spinal System (specifically, additional components including offset connectors)
Device Type: Spinal implant system, specifically a posterior system for immobilizing spinal segments as an adjunct to fusion.
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state quantitative acceptance criteria in a table format with corresponding reported performance for the additional components (offset connectors). However, it refers to the standard used for mechanical evaluation.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Mechanical performance meeting requirements of ASTM F1798 (Implied: adequate strength and fatigue resistance for spinal implants) | The subject components (offset connectors) of the PASS OCT spinal system were mechanically evaluated in axial and torsional grip and dynamic flexion-extension per ASTM F1798. (This indicates the tests were performed and presumably met the standard, otherwise it would likely not be cleared.) |
| Biocompatibility requirements met (Implied: non-pyrogenic) | Pyrogenicity testing was conducted in support of PASS OCT component substantial equivalence. (This indicates the test was performed and presumably met the standard.) |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated in terms of number of devices tested. Mechanical testing typically involves a statistically relevant number of samples, but the exact count is not provided here.
- Data Provenance: Not applicable in the context of human data. The testing is non-clinical (bench testing), performed on the device components themselves.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. This is a non-clinical, bench testing study. "Ground truth" does not apply in the context of expert consensus for interpretations as it would for imaging or diagnostic AI. The "ground truth" for mechanical testing is established by the test methods and acceptance limits defined in the ASTM F1798 standard.
4. Adjudication Method for the Test Set:
Not applicable. Adjudication methods are relevant for human interpretation tasks, not for mechanical bench testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No. This is a non-clinical bench study of a spinal implant, not an AI/ML diagnostic or assistive device for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used:
For the non-clinical mechanical testing, the "ground truth" is defined by the specifications and requirements of the ASTM F1798 standard (Standard Practice for Testing Impact-Resistant Spinal Fracture Reduction Systems components in axial and torsional grip and dynamic flexion-extension). The pyrogenicity testing would adhere to established standards for biocompatibility.
8. The Sample Size for the Training Set:
Not applicable. There is no training set as this is a physical medical device, not an AI/ML algorithm.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. There is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three interconnected faces in profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 11, 2016
Medicrea International S.A. Mr. David Rvan VP Product Development and Marketing 14 Porte du Grand Lyon Neyron 01700 FRANCE
Re: K161149
Trade/Device Name: PASS OCT Spinal System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: June 23, 2016 Received: June 27, 2016
Dear Mr. Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K161149
Device Name PASS OCT Spinal System
Indications for Use (Describe)
The PASS OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radios, and degenerative disease of the facets with instability. The PASS OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the PASS OCT Spinal System may be connected to the PASS LP Spinal System rods with the dual diameter rods or dominos. Refer to the PASS LP Spinal System package insert for a list of the PASS LP Spinal System indications of use.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
MEDICREA INTERNATIONAL's PASS OCT additional components
In accordance with 21 CFR 807.92 of the Federal Code of Requlations the following 510(k) summary is submitted for the PASS OCT Spinal System- Additional Components:
Date Prepared: 20 June 2016
1. Submitter:
| Submitter: | Contact Person: |
|---|---|
| MEDICREA INTERNATIONAL S.A. | David RYAN |
| 14 Porte du Grand Lyon | MEDICREA INTERNATIONAL S.A. |
| NEYRON 01700 | 14 Porte du Grand Lyon |
| FRANCE | NEYRON 01700 |
| Phone Number : +33 4 72 01 87 87 | FRANCE |
-
- Trade name: PASS OCT Spinal System
Regulatory Identification/ Classification
Orthosis, Cervical pedicle screw spinal fixation Product Code: NKG Unclassified, Pre-Amendment
Spinal Interlaminal Fixation Orthosis Regulation Number: 21CFR 888.3050 Product Code: KWP Class II
-
- Predicate or legally marketed devices which are substantially equivalent:
Primary predicate:
- PASS OCT Spinal System, (MEDICREA INTERNATIONAL, K150918)
Additional predicate:
- Vertex Reconstruction System, (MEDTRONIC SOFAMOR DANEK USA; K143471) ●
No reference devices were used in this submission.
4. Description of the device:
The PASS OCT Spinal System is a posterior system. which consists of a variety of shapes and sizes of rods, hooks, polyaxial screws, occipital plates, occipital bone screws, and connection components, which can be rigidly locked to the rod in a variety of configurations. See package insert of the system for labeling limitations.
The implants are manufactured in titanium alloy Ti-6AI-4V ELI conforming to ISO 5832-3 specifications and ASTM F136 specifications, in PEEK OPTIMA LT1 conforming to ASTM F2026 specifications and in cobalt-chromium molybdenum alloy Co-Cr28Mo6 that conforms to ISO 5832-
{4}------------------------------------------------
12 and ASTM F1537 specifications, and also CP Titanium according to ASTM F67 and ISO 5832-2 specifications.
Never use stainless steel and titanium implant components in the same construct.
To achieve best results, do not use any of the PASS OCT Spinal System implant components with components from any other systems or manufacturer unless specifically labeled to do so in this or another MEDICREA® INTERNATIONAL document.
MATERIALS: Titanium Alloy (Ti-6Al-4V) according to the ASTM F136-11 & ISO 5832-30; CP Titanium according to the ASTM F67 and ISO 5832-2: PEEK OPTIMA LT1® according to the ASTM F2026-10 and cobalt-chromium molybdenum alloy Co-Cr28Mo6 according to ISO5832-12 and ASTM F1537.
Function: The PASS OCT spinal system was developed as an implant:
- · To provide immobilization and stabilization of posterior spinal segments
- to auqment the development of a solid spinal fusion
- · to provide stability to ease fusion
- to be mechanically resistant to allow the fusion of the operated level
The purpose of this 510(k) submission is to add offset connectors to the PASS OCT Spinal System.
5. Indication for Use
The PASS OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PASS OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the PASS OCT Spinal System may be connected to the PASS LP Spinal System rods with the dual diameter rods or dominos. Refer to the PASS LP Spinal System package insert for a list of the PASS LP Spinal System indications of use.
Substantial equivalence claimed to predicate devices 6.
The PASS OCT Spinal system components are technologically similar to the already cleared PASS OCT Spinal System Components and Vertex Reconstruction System Components in terms of intended use, materials used, mechanical safety and performances
- . PASS OCT Spinal System, (MEDICREA INTERNATIONAL, K150918)
- Vertex Reconstruction System, (MEDTRONIC SOFAMOR DANEK USA; K143471) .
{5}------------------------------------------------
The table below compares the features and characteristics of PASS OCT components to their predicate devices.
| Device | MEDICREAINTERNATIONALPASS OCT SpinalSystemNEW COMPONENT | MEDICREAINTERNATIONALPASS OCT SpinalSystem | MEDTRONIC SOFAMORDANEKVertex ReconstructionSystem |
|---|---|---|---|
| 510(k) number | Unknown | K150918 | K143471 |
| Intended use | |||
| Occipital | Yes | Yes | Yes |
| Cervical | Yes | Yes | Yes |
| Thoracic | Yes | Yes | Yes |
| Components | |||
| Anchoragemeans | -Polyaxial screws- Polyaxial hooks- Occipital bonescrews | -Polyaxial screws- Polyaxial hooks- Occipital bone screws | - Multi-axial screws- Hooks- Occipital bone screws |
| Offset Connector | Yes | No | Yes |
| Materials | |||
| Titanium Alloy (Ti-6Al-4V) according toASTM F136 & ISO5832-3- Grade Titaniumaccording to ASTMF67-13 & ISO 5832-2PEEK OPTIMA LT1conforming to ASTMF2026Co-Cr 28Mo6 alloy 1(following theASTMF1537 | Titanium Alloy (Ti-6Al-4V) according to ASTMF136 & ISO 5832-3- Grade Titaniumaccording to ASTM F67-13 & ISO 5832-2PEEK OPTIMA LT1conforming to ASTMF2026Co-Cr 28Mo6 alloy 1(following theASTMF1537 | Titanium Alloy (Ti-6Al-4V) according to ASTMF136 & ISO 5832-3- Grade Titaniumaccording to ASTM F67& ISO 5832-2- Shape Memory Alloy(Nitinol-NiTi) accordingto ASTM F2063 |
Non-clinical Test Summary: 7.
The subject components of the PASS OCT spinal were mechanically evaluated in axial and torsional grip and dynamic flexion-extension per ASTM F1798.
Pyrogenicity testing was conducted in support to PASS OCT component substantial equivalence.
8. Clinical Test Summary
No clinical data was provided.
9. Conclusions Non clinical and Clinical
The PASS OCT Spinal system components are substantially equivalent to legally marketed predicate device.
N/A