K Number
K161019

Validate with FDA (Live)

Device Name
Uroview FD
Date Cleared
2016-08-17

(127 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Uroview FD is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/ urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to: · Querying and retrieving patient information and /or images from other modalities. · X-ray examination of the urogenital area (e.g. cyctoscopy, kidney, bladder, ureter, urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow measurements. · Ultrasound examinations(in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum. · Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement , penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy). · Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy). · Laparoscopy (e.g. cholecystectomy, lymph node dissection, abdominal testis detection/correction, varicocele). - · Application of fistula (kidney/bladder). - · Simple procedures (e.g. urethra, testis, phimonis). - · Introcorporeal shock wave lithotripsy. - Uroflow/urodynamics. - · Pediatric radiological and therapeutic applications.

Device Description

The "Uroview FD" is a solid state detector fluoroscopic, X-ray system, primarily for urological applications (functional x-ray diagnostics, endourology and minimal invasive urology/surgery). The basic unit is a cantilevered, continuously adjustable, isocentric tilting patient table called the "Uroview FD Skeleton", which can be raised and tilted to provide convenient access for the patient as well as optimum and ergonomic operating conditions for the user. The Uroview FD X-ray system is a completely mounted system and is equipped with a digital imaging system with a dynamic flat detector, designed to replace traditional spot film devices using screen-film cassettes or "CR" plates and fluoroscopy with image intensifier CCD cameras. The system is also equipped with a generator and automatic, multilayer, square field collimation system intended for installation on stationary X-ray equipment. A rotating anode X-ray tube is mounted. The measuring chamber is placed between the patient and the detector in order to detect the actual dose value for the automatic exposure control (AEC), to provide consistent x-ray film appearance and to guarantee error-free images even at low kV values. Attenuation factor is low and X-ray scattering is reduced to minimum by using a moveable grid. The Uroview FD X-ray system includes the following major components: - . Uroview FD skeleton (urological table incl. tilting table) - . High Frequency RF X-ray generator - X-ray tube incl. housing - . Collimator - . Measuring chamber - . Grid - . Dynamic flat panel detector (Pixium RF4343 FL, originally cleared under K080859) - . Digital imaging workstation - Video monitors - . Accessories The Uroview FD X-ray System includes the following two software programs: HIRIS RF43 and Uroview FD software; these are explained briefly. The Hiris RF43 software controls the digital imaging system in fluoroscopy and radiography modes using the flat panel detector. The software is specifically designed for remote controlled fluoroscopy and radiography and emergency equipment and performs real-time X-ray diagnostics of the gastro-intestinal tract and the urogenital system. The Uroview FD software controls the patient table Uroview FD (Skeleton) and does not interact with the HIRIS RF43 software. It allows the user comprehensive control of the patient table via hand control and foot switch allowing operator to activate movement of the Uroview FD. The Uroview FD system is designed to meet the requirements in accordance with relevant sections of 21CFR 1020.30-1020.33.

AI/ML Overview

It appears there's a misunderstanding in the prompt's request for acceptance criteria and study details for an AI/Software as a Medical Device (SaMD). The provided text is a 510(k) summary for a physical medical device, the Uroview FD, which is an X-ray fluoroscopic system.

This document does not describe an AI/SaMD product or a clinical study evaluating its performance with respect to specific acceptance criteria in the way typically required for AI/SaMD. Instead, it focuses on demonstrating substantial equivalence of the Uroview FD X-ray system to a predicate physical device (UROSKOP Omnia) based on technological characteristics and non-clinical performance data.

Therefore, most of the requested information regarding AI/SaMD acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and standalone performance for an AI model cannot be extracted from this document, as it pertains to a different type of device and different regulatory pathway.

The document explicitly states:

  • "There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device." (Section 10. Clinical Performance Data)
  • The software mentioned (HIRIS RF43 and Uroview FD software) controls the digital imaging system and patient table, respectively, and is verified at a "Moderate Level of Concern" for functionality, compatibility, risk analysis, and user interface. This is standard software validation for a hardware device, not an AI algorithm performing diagnosis or detection.

Given this, I cannot provide the requested table and study details as they are framed for an AI/SaMD. I can, however, summarize the non-clinical performance acceptance criteria and proof of adherence as outlined in the document for the physical Uroview FD X-ray system:


Summary of Acceptance Criteria and Performance for the Uroview FD X-ray System (Non-AI/SaMD)

The Uroview FD is a physical X-ray fluoroscopic system, not an AI/SaMD. Its acceptance criteria and performance demonstration are based on non-clinical testing to prove substantial equivalence to a legally marketed predicate device (UROSKOP Omnia).

1. Table of Acceptance Criteria and Reported Device Performance

Criterion CategoryAcceptance Criterion (Standard or Internal Requirement)Reported Device Performance (Proof of Adherence)
BiocompatibilityISO 10993-1, 5, 10 (Cytotoxicity, Sensitization, Irritation reactivity)PASSED all testing
Electrical SafetyIEC 60601-1, IEC 60601-2-2PASSED required testing
Electromagnetic Compatibility (EMC)IEC 60601-1-2PASSED required testing
Electromagnetic Safety (Specific to X-ray)IEC 60601-1-6, IEC 60601-2-28 (X-ray tube assemblies), IEC 60601-2-43 (Interventional procedures), IEC 60601-2-54 (Radioscopy)PASSED required testing
Radiation Dose DocumentationIEC 61910-1PASSED required testing
Usability EngineeringIEC 62366PASSED required testing
Software Verification & ValidationIEC 62304 / FDA Guidance (for Moderate Level of Concern software, including system compatibility, risk analysis, user interface)PASSED required testing
Shelf Life TestingInternal requirement: 10 years based on 13,000 life cycles for unit components (tube arm support, cassette box, vertical system movement, table tilt, longitudinal movement, transversal table top movement)PASSED all testing
Packaging and Transport TestingInternal requirement for humidity, temperature, and vibration testing to maintain integrity through normal shipping and handlingPASSED all testing
Risk ManagementISO 14971All requirements met and risks reduced as far as possible
General Compliance21 CFR 1020.30-1020.33System designed to meet requirements

2. Sample Size Used for the Test Set and Data Provenance

Not applicable in the context of AI/SaMD performance evaluation. The data for these tests are derived from laboratory and engineering tests of the physical device and its components, not from patient data or clinical imaging test sets.

3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications

Not applicable. Ground truth as typically understood for AI models (e.g., diagnostic labels on medical images) is not relevant here. The "ground truth" for these tests refers to the engineering specifications and performance standards.

4. Adjudication Method for the Test Set

Not applicable. There is no adjudication process involving multiple human reviewers for these non-clinical engineering and electrical tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device." This study type is for evaluating the impact of AI on human reader performance, which isn't relevant for this physical X-ray system.

6. If a Standalone (Algorithm Only) Performance was done

No, a standalone algorithm performance study was not done. The software components are for controlling the hardware and imaging process, not for performing diagnostic tasks independently.

7. The Type of Ground Truth Used

The ground truth for this device's performance demonstration is based on engineering specifications, recognized national and international standards (e.g., IEC, ISO), and validated internal test procedures. This is typical for a hardware medical device seeking substantial equivalence.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device, so there is no "training set" for an algorithm.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, overlaid on three human profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17,2017

Pausch Medical GmbH % Oliver Eikenberg, Ph.D. Senior Consultant OA & RA EMERGO Global Consulting LLC 816 Congress Avenue, Suite 1400 AUSTIN TX 78701

Re: K161019

Trade/Device Name: Uroview FD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: July 04, 2016 Received: July 07, 2016

Dear Dr. Eikenberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161019

Device Name Uroview FD

Indications for Use (Describe)

The Uroview FD is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional xray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/ urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to: · Querying and retrieving patient information and /or images from other modalities.

· X-ray examination of the urogenital area (e.g. cyctoscopy, kidney, bladder, ureter, urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow

measurements. · Ultrasound examinations(in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum.

· Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement , penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy).

· Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy).

· Laparoscopy (e.g. cholecystectomy, lymph node dissection, abdominal testis detection/correction, varicocele).

  • · Application of fistula (kidney/bladder).
  • · Simple procedures (e.g. urethra, testis, phimonis).
  • · Introcorporeal shock wave lithotripsy.
  • Uroflow/urodynamics.
  • · Pediatric radiological and therapeutic applications.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Uroview FD

K161019

1. Submission Sponsor

Pausch Medical GmbH Graf-Zeppelin-Str. 1 Erlangen Bavaria, D-91056 GERMANY Office Phone number: +49 (0)9131 / 99 92 - 0 Contact: Christian Stoian Title: CEO

2. Submission Correspondent

Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 Office Phone: (512) 327.9997 Contact: Oliver Eikenberg, PhD, Senior Consultant, QA / RA Email: project.management@emergogroup.com

3. Date Prepared

15 August 2016

4. Device Identification

Trade/Proprietary Name:Uroview FD
Common/Usual Name:X-ray System
Classification Name:Image-intensified fluoroscopic x-ray system
Regulation Number:892.1650
Product Code:JAA
Device Class:Class II
Classification Panel:Radiology
Guidance:FORM FDA 3626 (1/14), A Guide for the Submission of Initial Reports onDiagnostic X-Ray Systems and Their Major Components

5. Legally Marketed Predicate Device

UROSKOP Omnia (K101491, 06/30/2010) from Siemens Medical Solutions USA Inc.

{4}------------------------------------------------

6. Device Description

The "Uroview FD" is a solid state detector fluoroscopic, X-ray system, primarily for urological applications (functional x-ray diagnostics, endourology and minimal invasive urology/surgery). The basic unit is a cantilevered, continuously adjustable, isocentric tilting patient table called the "Uroview FD Skeleton", which can be raised and tilted to provide convenient access for the patient as well as optimum and ergonomic operating conditions for the user. The Uroview FD X-ray system is a completely mounted system and is equipped with a digital imaging system with a dynamic flat detector, designed to replace traditional spot film devices using screen-film cassettes or "CR" plates and fluoroscopy with image intensifier CCD cameras. The system is also equipped with a generator and automatic, multilayer, square field collimation system intended for installation on stationary X-ray equipment. A rotating anode X-ray tube is mounted. The measuring chamber is placed between the patient and the detector in order to detect the actual dose value for the automatic exposure control (AEC), to provide consistent x-ray film appearance and to guarantee error-free images even at low kV values. Attenuation factor is low and X-ray scattering is reduced to minimum by using a moveable grid.

The Uroview FD X-ray system includes the following major components:

  • . Uroview FD skeleton (urological table incl. tilting table)
  • . High Frequency RF X-ray generator
  • X-ray tube incl. housing
  • . Collimator
  • . Measuring chamber
  • . Grid
  • . Dynamic flat panel detector (Pixium RF4343 FL, originally cleared under K080859)
  • . Digital imaging workstation
  • Video monitors
  • . Accessories

The Uroview FD X-ray System includes the following two software programs: HIRIS RF43 and Uroview FD software; these are explained briefly.

The Hiris RF43 software controls the digital imaging system in fluoroscopy and radiography modes using the flat panel detector. The software is specifically designed for remote controlled fluoroscopy and radiography and emergency equipment and performs real-time X-ray diagnostics of the gastro-intestinal tract and the urogenital system.

The Uroview FD software controls the patient table Uroview FD (Skeleton) and does not interact with the HIRIS RF43 software. It allows the user comprehensive control of the patient table via hand control and foot switch allowing operator to activate movement of the Uroview FD.

The Uroview FD system is designed to meet the requirements in accordance with relevant sections of 21CFR 1020.30-1020.33.

{5}------------------------------------------------

7. Indication for Use Statement

The "Uroview FD" is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/urologic treatment table, may be used for urological treatment, planning and diagnostic procedures, including but not limited to:

  • . Querying and retrieving patient information and/or images from other modalities
  • . X-ray examinations of the urogenital area (e.g. cystoscopy, kidney, bladder, ureter, and urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow measurements.
  • . Ultrasound examinations (in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum.
  • . Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement, penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy).
  • . Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy)
  • . Laparoscopy (e.g. cholecystectomy, nephrectomy, lymph node dissection, abdominal testis detection/correction, varicocele)
  • . Application of fistula (kidney/bladder)
  • Simple procedures (e.g. urethra, testis, phimosis)
  • Intracorporeal shock wave lithotripsy
  • . Uroflow/urodynamics
  • Paediatric radiological and therapeutic applications

8. Substantial Equivalence Discussion

The following table compares the Uroview FD to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The Uroview FD shares similar X-ray components (basic table unit, generator, X-ray tube and housing, beamlimiting device flat detector, digital imaging processing) to that of the predicate UROSKOP Omnia. X-ray generation and control equipped within the Uroview FD system is also similar to specifications used in the UROSKOP Omnia. Many of the components used in the Uroview FD system are commercially available and FDA-listed or FDA-cleared X-ray components. Detailed instructions for use including safety features are included within the device labeling and design and enable the user to operate the device in a safe and effective manner. The operators are trained health care professionals familiar with X-ray examinations to be performed. The following comparison Table 5A of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.

{6}------------------------------------------------

Table 5A – Comparison of Characteristics for Uroview FD versus UROSKOP Omnia

Proposed DeviceUroview FDPredicate DeviceUROSKOP OmniaSimilarities /Differences
ManufacturerPausch Medical GmbHSiemens Medical Solutions USA, Inc.N/A
Product/Trade Name:Uroview FDUROSKOP OmniaN/A
510(k):PendingK101491N/A
EstablishmentRegistration No.96109032240869N/A
Product Code:JAAJAASame
Regulation Number892.1650892.1650Same
Device ClassificationNameImage-intensified fluoroscopic x-raysystemImage-intensified fluoroscopic x-raysystemSame
Class:IIIISame
Indications for useThe Uroview FD is a solid state detectorfluoroscopic X-ray system, primarily forurological applications (functional x-ray-diagnostics, endourology and minimal invasiveurology/surgery). The system, which includesa radiologic/ urologic treatment table, may beused for urological, treatment, planning anddiagnostic procedures including but notlimited to:Querying and retrieving patient informationand /or images from other modalities. X-ray examination of the urogenital area(e.g. cyctoscopy, kidney, bladder, ureter,urethra) including KUB, IVP, vasovesiculo-graphy, reflux-cystogram,cystourethrogram, and micturationcystourethrogram combined with uroflowmeasurements. Ultrasound examinations (in conjunctionwith a stand-alone ultrasound system) ofthe kidney, bladder, prostate, scrotum. Endourological interventions (e.g. of theurethra, prostate, bladder, sphincter, ostium,kidney and ureter, catheter placement , penileimplant placement, transurethral resection ofprostate or bladder, alternative treatment ofthe BPH and brachytherapy). Percutaneous interventions (e.g. PCNnephrolithotomy, resection, percutaneousnephrostomy). Laparoscopy (e.g. cholecystectomy,nephrectomy, lymph node dissection,abdominal testis detection/correction,varicocele). Application of fistula (kidney/bladder). Simple procedures (e.g. urethra, testis,phimosis). Intracorporeal shock wave lithotripsy. Uroflow/urodynamics.The UROSKOP Omnia is a solid state detectorfluoroscopic X-ray system, primarily forurological applications (functional x-ray-diagnostics, endourology and minimal invasiveurology/surgery). The system, which includesa radiologic/urologic treatment table, may beused for urological, gastroenterological andgynecological treatment, planning anddiagnostic procedures including but notlimited to:Querying and retrieving patient historyinformation and /or previous diagnosis andimages from other modalities. X-ray examination of the urogenital area(e.g. cyctoscopy, kidney, bladder, ureter,urethra) including KUB, IVP, vasovesiculo-graphy, reflux-cystogram,cystourethrogram, and micturationcystourethrogram combined with uroflowmeasurements. Ultrasound examinations (in conjunctionwith a stand-alone ultrasound system) ofthe kidney, bladder, prostate, scrotum. Endourological interventions (e.g. of theurethra, prostate, bladder, sphincter,ostium, kidney and ureter, catheterplacement , penile implant placement,transurethral resection of prostate orbladder, alternative treatment of the BPH,brachytherapy, as well as gynecologicalprocedures requiring radiological support). Percutaneous interventions (e.g. PCNnephrolithotomy, resection, percutaneousnephrostomy). Laparoscopy (e.g. cholecystectomy,nephrectomy, lymph node dissection,abdominal testis detection/correction,varicocele). Application of fistula (kidney/bladder). Simple procedures (e.g. urethra, testis,phimosis). Intracorporeal shock wave lithotripsy. Uroflow/urodynamics.Similar, onlyreduced forgastroentero-logical andgynecologicaltreatmentapplications
Proposed DeviceUroview FDPredicate DeviceUROSKOP OmniaSimilarities /Differences
Major ComponentsBasic Unit1. Basic Unit (urological table)1. Basic Unit (urological table)Same
2. Detector , 3. Collimator2. Detector , 3. Collimator
4. X-ray Generator4. X-ray Generator
5. X-ray tube and housing5. X-ray tube and housing
6. Measuring chamber6. Measuring chamber
7. Digital image workstation7. Digital image workstation
Uroview FD Basic Unit (urological table)
Basic Unit SystemOptional right-handed or lefthanded versionOptional right-handed or left handedversionSame
Tube Unit/FlatdetectorSynchronized longitudinal travel:20 cm (8")Synchronized longitudinal travel:15 cm (5.9")Similar, otherdimensions
Tilt rangeMotorized tilt: +/- 88 °lsocentric tilt: +/- 20 °Motorized tilt: +/- 90 °lsocentric tilt: +/- 15 °Similar, otherdimensions
Tube Assemblypark positionManual travel : 36.1 cm (14.2")Motorized travel : 32 cm (12.6")Similar, otherdimensions
Source-detectordistance115 cm (45.3 ")116 cm (45.7")Similar, otherdimensions
Tabletop-detectordistance6.4 cm (2.5 ")7 cm (2.8")Similar, otherdimensions
Powerrequirements:1/N 115/200/208/230/240 VACat 50 or 60 Hz3/PE~ 380/400/440/480 VAC (± 10%)at 50 or 60 HzSimilar, otherdimensions
Temperaturerange:+10°C to +40°C+15°C to +35°CSimilar, highertemperatureworking range
Relative humidity:20 % to 80%20 % to 75%Similar, higherhumidity workingrange
Barometricpressure:700 hPa to 1100 hPa700 hPa to 1060 hPaSimilar, higherpressure workingrange
Sterile:N/AN/AN/A
Battery OperatedN/AN/AN/A
Complies withApplicableVoluntaryStandardsIEC 60601-1IEC 60601-1-2IEC 60601-1-3IEC 60601-1-6IEC 60601-2-28IEC 60601-2-43IEC 60601-2-54IEC 62304IEC 62366ISO 14971IEC 60601-1IEC 60601-1-2IEC 60601-1-3IEC 60601-1-6IEC 60601-2-28IEC 60601-2-43IEC 60601-2-54IEC 62304IEC 62366ISO 14971Same
Complies withISO 10993-1 forBiocompatibilityyesyesSame
Electrical SafetyTesting PassedYesyesSame
Usable lifetimeTen yearsunknownn/a

{7}------------------------------------------------

Table 5A – Comparison of Characteristics for Uroview FD versus UROSKOP Omnia - Continued for Technological Characteristics

{8}------------------------------------------------

Table 5A – Comparison of Characteristics for Uroview FD versus UROSKOP Omnia

- Continued for Technological Characteristics
-------------------------------------------------
Proposed DeviceUroview FDPredicate DeviceUROSKOP OmniaSimilarities /Differences
Uroview FD Patient table and table top
Table topRadiolucent carbon-fibertabletop with foam mattress,allows controlled drainage offluids:120 cm (46.8 ") x 75.9 cm (29.9 ")Radiolucent carbon-fiber tabletopwith foam mattress, allowscontrolled drainage of fluids:115 cm (45.3 ") x 76 cm (29.9")Similar, otherdimensions
Tabletop extensions78 cm (30.7 ") x 75.9 cm (29.9 ")95 cm (37.4 ")x 76 cm (29.9 ")Similar, otherdimensions
Height adjustabletable topContinuously adjustable bymotor drive: 64 cm to 116 cm(25.2" to 45.7") , floatingContinuously adjustable bymotor drive: 72 cm to 122 cm(28.3" to 48 ") floatingSimilar, otherdimensions
LongitudinalmovementMotorized, continuous travel:+/- 18.8 cm (7.9")Motorized, continuous travel:+/-20 cm (7.9")Similar, otherdimensions
TransversemovementMotorized, continuous travel:+/- 7.1 cm (2.8")Motorized, continuous travel:+/-12.5 cm (4.9")Similar, otherdimensions
Table loadMax. 285 kg (628 lbs)Max. 272 kg (600 lbs)Similar
Display support armSpring-articulated arm mountedon the basic unit with 2 x 19"TFT color displaysSpring-articulated arm mountedon the basic unit with 2 x 19"TFT color displaysSame
TFT color displaySize 24"Image Matrix 1900 x 1200Maximum brightness 600 cd/m²Size 19"Image Matrix 1280 x 1024Maximum brightness 280 cd/m²Similar
Flat detector
43cm x 43cmHigh-resolution 2840 x 2880matrix with 148 μm pixel sizeand 16-bit digitization depthHigh-resolution 2840 x 2880matrix with 148 μm pixel sizeand 16-bit digitization depthSame
Input fields42 x 42 cm / 30 x 30 cm /20 x 20 cm / 15 x 15 cm42 x 42 cm / 30 x 30 cm/22 x 22 cm/15 x 15 cmSimilar
MaterialaSI with Csl scintillatoraSI with Csl scintillatorSame
Pixel size148 μm148 μmSame
Generator
DescriptionMicroprocessor-controlledhigh-frequency generator forradiography and fluoroscopyMicroprocessor-controlledhigh-frequency generator forradiography and fluoroscopySame
Power65 kW (650 mA at 100 kV)65 kW (650 mA at 100 kV)Same
Exposure voltage40 kV to 150 kV40 kV to 150 kVSame
X-ray tube
Nominal voltage150 kV150 kVSame
Nominal output40/80 kW40/80 kWSame
Focal spot nominalvalue0.6/1.20.6/1.0Similar
X-ray tube assemblyDual focusDual focusSame
collimatorAutomaticAutomaticSame
Digital Imaging System
Digital ImagingSystemFluoroscopy and radiographymodes (remote control) - using alarge flat panel detectorFluoroscopy and radiographymodes (remote control) - using alarge flat panel detectorSimilar

{9}------------------------------------------------

9. Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of the Uroview FD and in showing substantial equivalence to the predicate device that is subject to this 510(k) submission, Pausch Medical GmbH completed a number of non-clinical performance tests. The Uroview FD meets all the requirements for overall design, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.

The Uroview FD passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:

  • . Biocompatibility testing for the Uroview FD contacting materials including cytotoxicity, sensitization and irritation reactivity per ISO 10993-1, 5, 10: PASSED all testing
  • . Electrical safety testing per IEC 60601-1: PASSED required testing
  • . Electrical safety testing per IEC 60601-2-2: PASSED required testing
  • Electromagnetic Compatibility testing per IEC 60601-1-2: PASSED required testing
  • . Electromagnetic Safety testing per IEC 60601-1-6: PASSED required testing
  • . Electromagnetic Safety testing for X-ray tube assemblies for medical diagnosis per IEC 60601-2-28: PASSED required testing
  • . Electromagnetic Safety testing for X-ray equipment for interventional procedures per IEC 60601-2-43: PASSED required testing
  • . Electromagnetic Safety testing for X-ray equipment for radioscopy per IEC 60601-2-54: PASSED required testing
  • . Radiation dose documentation per IEC 61910-1: PASSED required testing
  • . Usability engineering testings per IEC 62366: PASSED required testing
  • . Software verification and validation testing has been completed on a functional level for a Moderate Level of Concern software including system compatibility testing, risk analysis and user interface testing per IEC 62304/FDA Guidance: PASSED required testing
  • . Shelf Life Testing for a period of ten (10) years based upon a 13000 life cycles for the Uroview FD unit including testings for tube arm support, cassette box, vertical system movement, table tilt, longitudinal movement, transversal table top movement: PASSED all testing
  • . Packaging and Transport Testing including humidity, temperature and vibration testing for Uroview FD to maintain integrity through normal shipping and handling: PASSED all testing
  • Risk Management per ISO 14971: all requirements were met and risks reduced as far as possible.

{10}------------------------------------------------

10. Clinical Performance Data

There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device(s).

It has been shown in this 510(k) submission that the difference between the Uroview FD and the predicate device UROSKOP Omnia does not raise any questions regarding its safety and effectiveness. Technological product characteristics, performance testing and compliance with voluntary standards, demonstrate that the Uroview FD device is substantially equivalent to the relevant aspects of the predicate device in terms of design, components, materials, principals of operation, performance characteristics, and intended use.

The Uroview FD system, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device(s).

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.