K Number
K161017

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-05-09

(27 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ConMed True HD 3MOS Camera System is intended for use in a variety of endoscopic surgical procedures including but not limited to orthopedic, laparoscopic, urologic, sinuscopic, plastic and as an accessory for microscopic surgery.

Device Description

The ConMed True HD 3MOS Camera System consists of a camera control unit (CCU) and a camera head that is used in conjunction with endoscopes, light source, light guide cables, monitors and other ancillary equipment to allow for high definition visualization during minimally invasive surgical procedures.

AI/ML Overview

The provided text is a 510(k) summary for the ConMed True HD 3MOS Camera System. This document focuses on demonstrating substantial equivalence to a predicate device and does not contain detailed information about specific acceptance criteria or the study design with the granularity requested. Medical device 510(k) submissions typically summarize the testing performed, but do not often provide the full study details, especially for performance characteristics that are not directly tied to a clinical outcome or a specific diagnostic accuracy metric.

However, based on the limited information available in the provided text, here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The document broadly states: "The tests results demonstrated the safety and effectiveness of the devices in accordance with design specifications and applicable standards." and "Performance testing were conducted to demonstrate that the True HD 3MOS Camera System performs according to specifications and functions as intended."

However, specific quantitative acceptance criteria (e.g., minimum resolution, signal-to-noise ratio, color accuracy range) and their corresponding measured performance results are NOT provided in this summary.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Performance testing" and "System Testing" (Verification and Validation including Functional testing, Video Image, Software). It also lists various standards evaluations (Electromagnetic compatibility, Electrical Safety, Reprocessing, Risk analysis, Software validation).

However, specific sample sizes for these tests (e.g., number of camera units tested, number of video sequences analyzed) are NOT provided. Details regarding data provenance (e.g., country of origin, retrospective or prospective nature of data collection) are also NOT provided. These are typically engineering performance tests, not clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This product is an endoscopic camera system for visualization during surgery. The primary performance characteristic is image quality for surgical viewing. The document does not mention any use of experts to establish 'ground truth' for image quality or clinical outcomes in the context of a diagnostic or comparative study with human readers. The assessment of image quality in such devices is typically done objectively against technical specifications and subjectively by engineers or clinical specialists without a formal 'ground truth' establishment process as seen in diagnostic AI.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Since no experts were mentioned for establishing ground truth or evaluating diagnostic performance, there is no adjudication method described or applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or mentioned. This device is a camera system, not an AI software intended to assist human readers in interpretation. The goal is to provide high-definition visuals to the surgeon.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to an "algorithm only" performance. The ConMed True HD 3MOS Camera System is a hardware device (camera system) used for visualization. While it has software components, its primary function is not an "algorithm only" diagnostic performance that would typically be evaluated in this manner. The performance testing focuses on the camera's ability to produce images according to specifications.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For a camera system, "ground truth" would typically relate to objective image quality metrics (resolution, color accuracy, brightness, latency) compared against engineering specifications, or the ability to correctly visualize anatomical structures. The document states "The tests results demonstrated the safety and effectiveness of the devices in accordance with design specifications and applicable standards." This implies the "ground truth" for the device's performance is adherence to established technical specifications and standards (e.g., video resolution of 1920 x 1080 pixels). It does not mention ground truth based on expert consensus, pathology, or outcomes data in a clinical sense.

8. The sample size for the training set

This device does not appear to involve machine learning in a way that requires a 'training set' for a diagnostic algorithm. The software leveraged "key components from a camera base component procured from Panasonic (SOUP)." This implies it's an off-the-shelf software component, not a custom-trained AI model requiring a specific training dataset as understood in AI/ML performance evaluation. Therefore, there is no mention of a training set sample size.

9. How the ground truth for the training set was established

As there is no training set for a machine learning model, there is no ground truth establishment method described for a training set. The "ground truth" for this camera system would be based on its compliance with engineering and safety standards, as well as its ability to capture and display high-definition video during surgical procedures.

In summary: The provided 510(k) summary is for a hardware medical device (camera system) and focuses on demonstrating substantial equivalence through technical performance verification and validation, adherence to standards, and risk analysis. It does not contain the detailed information typically found in submissions for AI/ML-based diagnostic devices, which would involve concepts like training/test sets, ground truth establishment by experts, and comparative effectiveness studies with human readers.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three heads in profile, representing the department's focus on people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 9, 2016

ConMed Corporation Diair Cantisani Regulatory Affairs Specialist 525 French Road Utica, New York 13502

Re: K161017

Trade/Device Name: True HD 3MOS Camera System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: April 8, 2016 Received: April 12, 2016

Dear Diair Cantisani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161017

Device Name

ConMed True HD 3MOS Camera System

Indications for Use (Describe)

The ConMed True HD 3MOS Camera System is intended for use in a variety of endoscopic surgical procedures including but not limited to orthopedic, laparoscopic, urologic, sinuscopic, plastic and as an accessory for microscopic surgery.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a stylized blue swoosh on the left, followed by the word "CONMED" in bold, black letters. Below the company name is the word "CORPORATION" in smaller, blue letters.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS True HD 3MOS Camera System

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21CFR 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) number ______________________________________________________________________________________________________________________________________________

I. SUBMITTER

ConMed Corporation 525 French Road Utica, NY 13502

Establishment Registration: 1320894

II. COMPANY CONTACT

Diair Cantisani, Regulatory Affairs Specialist Phone: 508-948-2084 Fax: 508-366-8858 Date Prepared: February 29, 2016

III. DEVICE

Device Trade Name:True HD 3MOS Camera System
Common Name:HD 3MOS Camera System
Classification Name:Laparoscope, General & Plastic Surgery
Regulatory Class:Class II, per 21 CFR 876.1500
Review Panel:General and Plastic Surgery
Product Codes:GCJ

IV. PREDICATE DEVICE

Device Name:ConMed 3-CCD High Definition Digital Camera System
Company Name:ConMed Corporation
510(k) #:K063457

V. DEVICE DESCRIPTION

Modified Device

The ConMed True HD 3MOS Camera System consists of a camera control unit (CCU) and a camera head that is used in conjunction with endoscopes, light source, light guide cables, monitors and other ancillary equipment to allow for high definition visualization during minimally invasive surgical procedures.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a blue square with a white, curved shape inside, followed by the word "CONMED" in black, sans-serif font. Below the word "CONMED" is the word "CORPORATION" in a smaller, sans-serif font, underlined by a thin blue line.

Like the predicate device, sterilization of the camera head and cable is required before use. The camera heads must be sterilized using steam sterilization.

Unmodified Device

The ConMed HD Digital Camera System consists of a camera control unit (CCU) and a camera head that is used in conjunction with an endoscope to allow for high definition visualization during minimally invasive surgical procedures. Sterilization of the camera head and cable is required before use. The camera heads must be sterilized using steam sterilization.

VI. INTENDED USE / INDICATIONS FOR USE

The True HD 3MOS Camera System is intended for use in a variety of endoscopic surgical procedures including but not limited to orthopedic, laparoscopic, urologic, sinuscopic, plastic and as an accessory for microscopic surgery.

VII. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS

The ConMed True HD 3MOS Camera System has the same technological characteristics and design as the predicate device except for the following features:

  • . The camera head incorporates a CMOS image sensor in oppose to the CCD image sensor featured in the predicate device.
  • The proposed device software system leverages key components from a camera base component procured from Panasonic (SOUP) versus the in-house developed software system utilized in the predicate device.
  • . Unlike the predicate device that outputs 1280 x 720 pixels and 1920 x 1080 pixels video resolution the proposed device True HD 3MOS Camera System outputs 1920 x 1080 pixels only.
  • The camera control unit introduces a full metal chassis in oppose to the metal and plastic bezel chassis in the predicate device.

VIII. PERFORMANCE TESTING

Testing has been completed to demonstrate that the device performs as intended given the modification, and is substantially equivalent to the predicate device.

Completed test data includes the following:

  • . Performance testing were conducted to demonstrate that the True HD 3MOS Camera System performs according to specifications and functions as intended.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a blue abstract shape resembling a stylized "C" on the left. To the right of the shape is the word "CONMED" in bold, black letters. Below "CONMED" is the word "CORPORATION" in smaller, black letters.

The tests results demonstrated the safety and effectiveness of the devices in accordance with design specifications and applicable standards.

  • System Testing ●
    • o Verification
      • Functional testing
      • Video Image ■
      • Software
    • o Validation
      • Functional testing
      • Video Image
      • Software
  • Standards evaluation
    • Electromagnetic compatibility, IEC60601-1-2 о
    • Electrical Safety, IEC60601-1 O
    • Particular Standard, safety and essential performance of endoscopic O equipment, IEC60601-2-18
    • o Reprocessing, AAMI TIR12
  • . Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
  • The software validation activities were performed in accordance with the . IEC62304.
  • The instructions for reprocessing of the camera heads of the subject devices include Cleaning and Sterilization. All cleaning and sterilization methods have been validated.

SUBSTANCIAL EQUIVALENCE IX.

The differences between the predicate devices and the proposed device do not raise any new risks of safety or efficacy. Supporting information in this submission confirms that the ConMed True HD 3MOS Camera System is safe and effective for its intended use and is substantially equivalent in design, intended use, principals of operation, and technical characteristics to the ConMed 3-CCD High Definition Digital Camera System (K063457).

X. CONCLUSION

Based upon the testing and analysis performed, the ConMed True HD 3MOS Camera System is as safe, as effective, and performs as well as the ConMed 3-CCD High Definition Digital Camera System (K063457) for image visualization and user interaction.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.