K Number
K160987

Validate with FDA (Live)

Device Name
ReWalk(TM)
Date Cleared
2016-07-22

(105 days)

Product Code
Regulation Number
890.3480
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ReWalk™ orthotically fits to the lower limbs and part of the upper body and is intended to enable individuals with spinal cord injury at levels T7 to L5 to perform ambulatory function of a specially trained companion in accordance with the user assessment and training certification program. The device is also intended to enable individuals with spinal cord injury at levels T4 to perform ambulatory functions in rehabilitations in accordance with the user assessment and training certification program. The ReWalk™ is not intended for sports or stair climbing.

Device Description

The ReWalk™ orthotically fits to the lower limbs and part of the upper body.

AI/ML Overview

This document is an FDA 510(k) clearance letter for the ReWalk™ Powered Lower Extremity Exoskeleton. It does not contain information about acceptance criteria or a study proving that the device meets those criteria, as typically found in a clinical study report or a summary of safety and effectiveness (SSED).

The letter primarily confirms that the FDA has reviewed the 510(k) premarket notification and determined the device is substantially equivalent to legally marketed predicate devices for its stated indications for use. It lists the regulation number, product code, and regulatory class, and outlines general controls provisions of the Act that apply to the device.

Therefore, I cannot provide the requested information based on the provided text. The document is an administrative regulatory clearance, not a technical report detailing performance and validation studies.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, which is a symbol often associated with healthcare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 22, 2016

ReWalk Robotics Inc. John Hamilton VP, Regulatory Affairs/ Clinical 200 Donald Lynch Boulevard, Suite 204 Marlboro, MA 01752

Re: K160987

Trade/Device Name: ReWalk™M Regulation Number: 21 CFR 890.3480 Regulation Name: Powered Lower Extremity Exoskeleton Regulatory Class: Class II Product Code: PHL Dated: June 23, 2016 Received: June 24, 2016

Dear Mr. Hamilton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael J. Hoffmann -A

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160987

Device Name ReWalk™M

Indications for Use (Describe)

The ReWalk™ orthotically fits to the lower limbs and part of the upper body and is intended to enable individuals with spinal cord injury at levels T7 to L5 to perform ambulatory function of a specially trained companion in accordance with the user assessment and training certification program. The device is also intended to enable individuals with spinal cord injury at levels T4 to perform ambulatory functions in rehabilitations in accordance with the user assessment and training certification program. The ReWalk™ is not intended for sports or stair climbing.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 890.3480 Powered lower extremity exoskeleton.

(a)
Identification. A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
i.e., cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
e.g., indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
e.g., EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
e.g., level of supervision or assistance, and environment of use (e.g., indoors and/or outdoors) including obstacles and terrain).