K Number
K160940

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-11-23

(233 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Masimo Disposable Transflectance Forehead Sensors are indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Masimo Disposable Transflectance Forehead Sensors are intended for use with adult and pediatric patients, weighing greater than 10 kg, who are well or poorly perfused in healthcare environments.

Device Description

The subject device, Masimo Disposable Transflectance Forehead Sensor (Forehead Sensor), is a single-use sensor and is supplied non-sterile. The sensor is available in two configurations, consisting of the LNCS TFA-1 with the DB9 sensor connector and the M-LNCS TFA-1 with the M15 sensor connector. The different sensor connectors allow for sensor connection to instruments with corresponding mating DB9 or M15 connectors. The subject device shall also be available with RD sensor connector. The subject device is intended to be used with instruments that include Masimo SET technology for the monitoring of:

  • Functional Oxygen Saturation of Arterial Hemoglobin (SpO2): The amount of oxyhemoglobin expressed a percentage of the hemoglobin that is available
  • Pulse Rate (PR): Measured in beats per minute (BPM) is based on the optical detection of peripheral flow pulse
  • Perfusion Index (PI): The ratio of the pulsatile blood flow to the non-pulsatile or static blood in peripheral tissue
  • Pleth Variability Index (PVI): Measure of dynamic changes in perfusion index that occur during the respiratory cycle.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Masimo Disposable Transflectance Forehead Sensor based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance Criteria (Predicate Device K032551)Reported Device Performance (Subject Device)
SpO2, no motion70-100 ± 2 % ARMS*70-100 ± 2 % ARMS*
SpO2, low perfusion70-100 ± 2 % ARMS*70-100 ± 2 % ARMS*
Pulse Rate, no motion25-240 ± 3 bpm ARMS*25-240 ± 3 bpm ARMS*
Pulse Rate, low perfusion25-240 ± 3 bpm ARMS*25-240 ± 3 bpm ARMS*
Patient population>30 kg (Predicate)>10 kg (Subject Device)
Type of useReusable (Predicate)Single-use (Subject Device)
SterilitySupplied non-sterileSupplied non-sterile
BiocompatibilityISO-10993 compliant materialsISO-10993 compliant materials
Electrical SafetyIEC60601-1 compliantIEC60601-1 compliant
EMC TestingIEC60601-1-2 compliantIEC60601-1-2 compliant

*ARMS (accuracy root mean square) per ISO-80601-2-61.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a specific number of subjects for the clinical accuracy study. The document mentions "healthy adult volunteers with light to dark skin pigmentation." However, it does state that the "variation in the accuracy specifications equals plus or minus one standard deviation, which encompassed approximately 68% of the tested population." This implies a sufficiently large sample to calculate a standard deviation for clinical accuracy, but the exact number isn't provided.
  • Data Provenance: The study was clinical and conducted by the manufacturer, Masimo Corporation. It involved "healthy adult volunteers." The location of these volunteers (e.g., country of origin) is not specified. The study appears to be prospective as it involved applying the sensors and taking measurements from volunteers.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. The ground truth for SpO2 measurements was established using a laboratory CO-oximeter for arterial blood samples, not human experts.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was based on objective laboratory measurements (CO-oximeter), not expert opinions requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This device is a sensor for pulse oximetry, and its performance is evaluated against objective physiological measurements, not through human reader interpretation of images or data.

6. If a Standalone Study Was Done

Yes, a standalone study was done. The document describes clinical performance testing where "The Forehead Sensors, connected to MX technology boards with Masimo SET technology, were applied to healthy adult volunteers...SpO2 measurements were compared against arterial blood samples analyzed by a laboratory CO-oximeter, in the range of 70 to 100%, to determine accuracy specifications." This describes the algorithm's (Masimo SET technology) performance in conjunction with the sensor independently.

7. The Type of Ground Truth Used

The ground truth used for SpO2 accuracy was objective laboratory measurements from a CO-oximeter analyzing arterial blood samples.

8. The Sample Size for the Training Set

The document does not provide information about a training set size. This is a medical device, and the "Masimo SET technology" refers to the underlying algorithm for pulse oximetry measurements. The document outlines verification and validation of the sensor and the expected accuracy of the associated technology, but does not detail the development or "training" of a machine learning algorithm in the way that would typically be described for AI software.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a training set and its ground truth in the context of a machine learning algorithm are not described for this device. The "Masimo SET technology" is pre-existing and forms the basis of the oximeter's function, with the sensor being a component validated for use with this technology.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 23, 2016

Masimo Corporation Marguerite Thomlinson Senior Director, Regulatory Affairs 52 Discovery Irvine, California 92618

Re: K160940

Trade/Device Name: Masimo Disposable Transflectance Forehead Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: October 24, 2016 Received: October 27, 2016

Dear Marguerite Thomlinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image contains a signature and a name. The name is "Tina Kiang-S". The signature is to the left of the name and is a series of loops and curves.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160940

Device Name

Masimo Disposable Transflectance Forehead Sensors

Indications for Use (Describe)

The Masimo Disposable Transflectance Forehead Sensors are indicated for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Masimo Disposable Transflectance Forehead Sensors are intended for use with adult and pediatric patients, weighing greater than 10 kg, who are well or poorly perfused in healthcare environments.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.

Submitter and Address ofManufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7000FAX: (949) 297-7592
Date:November 21, 2016
Contact:Marguerite ThomlinsonSenior Director, Regulatory Affairs
Trade Name:Masimo Disposable Transflectance Forehead Sensors
Common Name:Oximeter
Classification Regulation/Product Code:21 CFR 870.2700, Class II/DQA
Establishment RegistrationNumber:2031172
Reason for PremarketNotification:New Device
Predicate Device:K032551, Masimo LNOP TF-1 Reusable TransflectanceSensor
Performance StandardsNo performance standards for the above device have beenpromulgated pursuant to Section 514.

Device Description

The subject device, Masimo Disposable Transflectance Forehead Sensor (Forehead Sensor), is a single-use sensor and is supplied non-sterile. The sensor is available in two configurations, consisting of the LNCS TFA-1 with the DB9 sensor connector and the M-LNCS TFA-1 with the M15 sensor connector. The different sensor connectors allow for sensor connection to instruments with corresponding mating DB9 or M15 connectors. The subject device shall also be available with RD sensor connector. The subject device is intended to be used with instruments that include Masimo SET technology for the monitoring of:

  • Functional Oxygen Saturation of Arterial Hemoglobin (SpO2): The amount of ● oxyhemoglobin expressed a percentage of the hemoglobin that is available

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

  • Pulse Rate (PR): Measured in beats per minute (BPM) is based on the optical detection of peripheral flow pulse
  • . Perfusion Index (PI): The ratio of the pulsatile blood flow to the non-pulsatile or static blood in peripheral tissue
  • Pleth Variability Index (PVI): Measure of dynamic changes in perfusion index that occur ● during the respiratory cycle.

The Masimo SET technology is provided by the Masimo SET or Masimo Rainbow SET technology board, which is included in Masimo pulse oximeter instruments, Masimo pulse CO-oximeter instruments and third-party multi-parameter instruments.

Forehead Sensor Specifications
FeaturesSpecifications
Patient populationAdults and pediatrics > 10 kg
Measurement siteForehead
Type of useSingle-use
SterilitySupplied non-sterile
Accuracy*
SpO2, no motion70-100 ± 2 %
SpO2, low perfusion70-100 ± 2 %
Pulse Rate, no motion25-240 ± 3 bpm
Pulse Rate, low perfusion25-240 ± 3 bpm
Mechanical
Sensor dimensions/weightApproximately 2.5 x 1.5 in/ 1.0 oz (without cable)
Sensor cable length36 in, nominal
Patient-contacting materialsISO-10993 compliant
Environmental
Storage temperature-40°C to +70°C
Operating temperature+5°C to +40°C
Humidity10-95% humidity, non-condensing

The specifications of the Forehead Sensor are listed in the table below.

  • Accuracy was determined in terms of Arms (accuracy root mean square) per ISO-80601-2-61.

Intended Use/Indications for Use

The Masimo Disposable Transflectance Forehead Sensors are indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Masimo Disposable Transflectance Forehead Sensors are intended for use with adult and pediatric patients, weighing greater than 10 kg, who are well or poorly perfused in healthcare environments.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a stylized "V" symbol to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

Technological Characteristics

Principle of Operation

The subject device, Forehead Sensor, is a pulse oximetry sensor which is governed by the following principles:

  • Oxyhemoglobin (oxygentated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
  • The amount of arterial blood in tissue changes with a person's pulse (photoplethysymography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.

Mechanism of Action for Achieving the Intended Effect

The Forehead Sensor is placed on the patient forehead and secured with a headband. The other end of the sensor is connected to a patient cable, which in turn is connected to a pulse oximeter. The instrument is turned on when use begins. Once use is complete, the user turns the instrument power "off" and removes the sensor from the patient.

Summary of Technological Characteristics of Subject Device Compared to Predicate Device

Similarities and Differences between Subject Device and Predicate Device (K032551)

The subject device, Forehead Sensor, and the predicate (K032551) have the following key similarities:

  • . Both devices have the same intended use of pulse oximetry measurements under no motion and low perfusion conditions;
  • Both devices have the same forehead measurement site;
  • Both devices are supplied non-sterile; and ●
  • Both devices have the same accuracy specifications ●

The subject and predicate devices are mainly different in that:

  • the subject device is single-use device whereas the predicate is reusable; and ●
  • the subject device is intended for patients > 10 kg whereas the predicate is intended for ● patients >30 kg.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" and the address "52 Discovery, Irvine, CA 92618".

Substantial Equivalent

Table 12.5 Substantial Equivalent Discussion
FeatureMasimo Disposable Forehead SensorSubject Device, PendingMasimo Reusable Forehead Sensors, LNOP TF-1Predicate, K032551
510(k) No.General
Indications for useThe Masimo Disposable TransflectanceForehead Sensors are indicated forcontinuous noninvasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rate. TheMasimo Disposable TransflectanceForehead Sensors are intended for usewith adult and pediatric patients,weighing greater than 10 kg, who are wellor poorly perfused in healthcareenvironments.The following additional Masimo Sensor isindicated for the continuous noninvasive functionaloxygen saturation of arterial hemoglobin (SpO2) andpulse rate:Reusable oximetry transflectance sensor intendedfor adults and pediatrics greater than 30 kg inhospitals, hospital-type facilities, mobile, and homeenvironments.
Classification regulation/product codeSame as K03255121 CFR 870.2700, Class II / DQA
Principle of operationSame as K032551Pulse oximetry is governed by the principles of a)Oxyhemoglobin (oxygenated blood) and b)deoxyhemoglobin (non-oxygenated blood) differ intheir absorption of red and infrared light(spectrophotometry). The amount of arterial blood intissue changes with your pulse(photoplethysymography). Therefore, the amount oflight absorbed by the varying quantities of arterialblood changes as well.
Mechanism of actions forachieving the intended useSame as K032551The Forehead Sensor is placed on the patientforehead and secured with a headband. The otherend of the sensor is connected to a patient cable,which in turn is connected to a pulse oximeter. Theinstrument is turned on when use begins. Once useis complete, the user turns the instrument power"off" and removes the sensor from the patient.
Patient populationAdults and pediatrics >10 kgAdults and pediatrics >30 kg
Measurement siteSame as K032551Forehead
Type of useSingle-useReusable
SterilitySame as K032551Supplied non-sterile
CompatibilityInstruments with Masimo SETtechnology provided by Masimo SET orMasimo Rainbow SET technologyboards, including the MS, MS-2K andMX series boardsInstruments with Masimo SET technology providedby Masimo SET technology boards, including theMS series boards
Perfusion index (PI)
PI FunctionSame as K032551PI is the measure of the ratio of the pulsatile bloodflow to the non-pulsatile or static blood in peripheraltissue
Table 12.5 Substantial Equivalent Discussion
FeatureMasimo Disposable Forehead SensorMasimo Reusable Forehead Sensors, LNOP TF-1
510(k) No.Subject Device, PendingPredicate, K032551
PI RangeSame as K0325510.02% to 20%
Pleth variability index (PVI)Same as K032551PVI is the measure of dynamic changes in perfusion index that occur during the respiratory cycle. PVI is available based on whether the feature is enabled in the instrument with Masimo SET technology
PVI RangeSame as K0325510% to 100%
Accuracy, ARMS
SpO2, no motion70-100 ± 2 %70-100 ± 2 %
SpO2, low perfusion70-100 ± 2 %70-100 ± 2 %
Pulse rate, no motion25-240 ± 3 bpm25-240 ± 3 bpm
Pulse rate, low perfusion25-240 ± 3 bpm25-240 ± 3 bpm
Biocompatibility
Patient-contacting materialsDisposable adhesive sensor withbiocompatible patient contact materials:Lens: optically clear material for lighttransmission, polycarbonateLight barrier: polyurethaneSensor face: acrylic adhesiveCable jacket: PVC medical gradeReusable sensor with biocompatible patient contactmaterials:Lens: optically clear material for light transmission,silicon rubberLight barrier: PVCSensor face: PVCCable Jacket: PVC medical grade

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the word "Masimo" are the words "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".

The difference in the indications for use (IFU) between the subject and predicate devices does not raise different questions of safety and effectiveness because both devices have the same intended use as forehead sensors for pulse oximetry, same principle of operation and same accuracy specifications. Although the patient-contacting materials for the subject device and predicate are slightly different, the materials are the same in that they are compliant to ISO-10993-1. Additionally, since the patient population is the same for the subject device and the predicate, there is no expected difference in the subject device's performance with the slight extended weight range for the subject device. Please note the reference device in K051271, where the same patient population of greater than 10 kg does not require additional testing. Furthermore, the subject device has been validated in regards to human factors and usability.

In summary, the accompanied verification and validation data for the subject device, including testing for any differences, demonstrated that the subject device is as safe and effective as the predicate and therefore substantially equivalent to the predicate.

Non-clinical Testing

The subject devices were subjected to bench testing. The following non-clinical testing, as applicable, was performed in accordance with Masimo design control requirements and quality

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a stylized red checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".

system to demonstrate substantial equivalence of the subject device with its predicates:

  • Electrical safety testing per IEC60601-1
  • . EMC testing per IEC60601-1-2
  • Pulse oximetry testing, including simulator testing of pulse rate and low perfusion, per ISO 80601-2-61,
  • Biocompatibility testing per ISO-10993-1, ISO-10993-5 and ISO-10993-10 ●
  • Usability testing per FDA Human Factors and Usability Guidance
  • Software verification per FDA Software Guidance
  • Mechanical and environmental testing

Clinical Testing

Clinical performance testing was completed in accordance with ISO 80601-2-61 to evaluate the accuracy of functional oxygen saturation (SpO2) measurements. The Forehead Sensors, connected to MX technology boards with Masimo SET technology, were applied to healthy adult volunteers with light to dark skin pigmentation. SpO2 measurements were compared against arterial blood samples analyzed by a laboratory CO-oximeter, in the range of 70 to 100%, to determine accuracy specifications in terms of accuracy root mean square (Arms). The variation in the accuracy specifications equals plus or minus one standard deviation, which encompassed approximately 68% of the tested population. Arms calculation incorporates error due to bias and imprecision (standard deviation).

Conclusion

The non-clinical and clinical testing provided in this 510(k) submission demonstrates that the subject devices. Forehead Sensors are as safe and effective as the predicate and therefore substantially equivalent to the predicate in K032551.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).