K Number
K160910

Validate with FDA (Live)

Device Name
Elecsys Syphilis
Manufacturer
Date Cleared
2016-07-28

(118 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro qualitative detection of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma. The test is intended as an aid in the diagnosis of syphilis infection with clinical signs and symptoms.

The Elecsys Syphilis immunoassay is not intended for use in screening blood or tissue donors. The effectiveness of this assay in testing blood or tissue donors has not been established.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 analyzer.

PreciControl Syphilis is intended for the quality control of the Elecsys Syphilis immunoassay on the cobas e 411 analyzer.

Device Description

The Elecsys Syphilis assay is a fully automated qualitative assay detecting IgG and IgM antibodies to Treponema pallidum, the causative agent of syphilis. Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of active or previous infection with Treponema pallidum in persons with signs and symptoms of syphilis, as well as in patients at risk for syphilis infection. This assay does not determine the stage of infection or associated disease.

PreciControl Syphilis is a lyophilized control based on human serum. It is used for monitoring the accuracy of the Elecsys Syphilis immunoassay.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Elecsys Syphilis assay, based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria for sensitivity and specificity. Instead, it presents the performance of the Elecsys Syphilis assay compared to a "final comparator algorithm" or "medically diagnosed individuals" and implies that "good agreement" and "substantial equivalence" are the overarching criteria.

Assuming the implicit acceptance criteria are high levels of positive percent agreement (PPA) and negative percent agreement (NPA) with the established comparator, here's a table summarizing the reported performance:

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Elecsys Syphilis)Study Context
PPA (Sensitivity)High agreement (e.g., >95%)*100% (228/228)Prospective Cohorts (Total)
95.00 to 99.83% (95% CI)Staged (Retrospective)
98.7% (155/157)Overall Retrospective Cohorts
100% (15/15)Pregnant (Retrospective)
NPA (Specificity)High agreement (e.g., >95%)*99.2% (2038/2054)Prospective Cohorts (Total)
99.40 to 99.96% (95% CI)Routine Syphilis (Prospective)
92.58 to 97.63% (95% CI)HIV (Prospective)
100% (301/301)Pregnant (Prospective)
100% (12/12)Staged (Retrospective)
100% (12/12)Overall Retrospective Cohorts

*Note: The document states "good agreement" and "substantially equivalent," which generally implies high sensitivity and specificity for diagnostic assays. The specific quantitative thresholds are not explicitly defined as "acceptance criteria" but are demonstrated by the reported results.


2. Sample Sizes and Data Provenance

  • Test Set Sample Sizes:

    • Prospective Cohorts: Total of 2282 samples.
      • Routine Syphilis testing: 1524 subjects
      • HIV positive subjects: 457 subjects
      • Pregnant women: 301 subjects
    • Retrospective Medically Diagnosed Individuals: Total of 169 specimens.
      • Pregnant positive women: 15
      • Subjects medically diagnosed with syphilis at different stages: 154
    • Apparently Healthy Individuals: 209 specimens.
    • Analytical Specificity: 266 human clinical samples from patients with medical conditions unrelated to Syphilis.
  • Data Provenance: The document generally refers to "human serum and plasma" samples.

    • Prospective Cohorts: "prospectively collected samples of the intended use population." A multi-center study was conducted in the EU and the US for cut-off validation, implying global data.
    • Retrospective Medically Diagnosed Individuals: "pre-selected retrospective cohort."
    • The "apparently healthy individuals" and "analytical specificity" samples are also clinical.
  • Retrospective or Prospective: Both retrospective and prospective data were used for clinical performance evaluation.


3. Number of Experts and Qualifications for Ground Truth

The document does not specify the "number of experts" or their "qualifications" used to establish the ground truth.


4. Adjudication Method for the Test Set

The ground truth for the clinical performance studies was established using a "composite algorithm" rather than expert adjudication.

  • Composite Algorithm: For the prospective cohorts, "all samples were tested according to a composite algorithm using FDA-cleared tests that included the predicate Syphilis immunoassay, the Rapid Plasma Reagin (RPR) non-treponemal specific assay and the Treponema pallidum Particle Agglutination, treponemal-specific assay."
  • For the retrospective medically diagnosed individuals, the ground truth was based on "medically diagnosed individuals" and the "final comparator results" from the composite algorithm.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not conducted. This device is an in-vitro diagnostic (IVD) immunoassay, not an AI-assisted diagnostic imaging or interpretation tool for human readers.


6. Standalone (Algorithm Only) Performance

Yes, the studies presented are for standalone (algorithm only) performance. The Elecsys Syphilis assay is an automated immunoassay that provides a qualitative result (reactive or non-reactive) without human interpretation in the loop for the primary result generation. Its performance is measured against established diagnostic algorithms and clinical diagnoses.


7. Type of Ground Truth Used

The primary ground truth used for evaluating clinical performance was a composite algorithm of FDA-cleared tests (predicate immunoassay, RPR, and TPPA). Additionally, a medical diagnosis was used for the retrospective cohort.


8. Sample Size for the Training Set

The document does not explicitly state the sample size for a "training set." This type of IVD immunoassay typically involves extensive assay development and optimization rather than a "training set" in the machine learning sense. The "cut-off determination" process likely utilized a set of samples, but these are not explicitly termed a "training set" with a specified size.

  • Cut-off Determination: "native human serum samples were measured with a prototype lot of the Elecsys Syphilis assay and a preliminary cut-off was set."
  • Final Validation: "Final validation of the cut-off was done in multi-center studies conducted in the EU and the US."

9. How the Ground Truth for the Training Set Was Established

As there's no explicitly defined "training set" in the machine learning context, the ground truth for cut-off determination would have been established through a combination of:

  • Known positive and negative samples: These would have been derived from individuals with confirmed syphilis infections (using reference methods or clinical diagnosis) and confirmed non-infected individuals.
  • Preliminary cut-off setting: Based on the distribution of results from these known samples.
  • Comparability with commercially available assays: Used to refine and validate the cut-off.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a row, with the first profile being the most prominent and the other two slightly behind it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Roche Diagnostics Angelo Pereira Regulatory Affairs Senior Program Manager 9115 Hague Road Indianapolis, IN, 46250

Re:K160910

Trade/Device Name: Elecsys Syphilis and PreciControl Syphilis Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LIP, JJX Dated: July 22, 2016 Received: July 25, 2016

Dear Mr. Pereira:

This letter corrects our substantially equivalent letter of July 28, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160910

Device Name Elecsys Syphilis PreciControl Syphilis

Indications for Use (Describe)

Immunoassay for the in vitro qualitative detection of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma. The test is intended as an aid in the diagnosis of syphilis infection with clinical signs and symptoms.

The Elecsys Syphilis immunoassay is not intended for use in screening blood or tissue donors. The effectiveness of this assay in testing blood or tissue donors has not been established.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 analyzer.

PreciControl Syphilis is intended for the quality control of the Elecsys Syphilis immunoassay on the cobas e 411 analyzer.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Elecsys Syphilis

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0416
ContactAngelo PereiraPhone: (317) 521-3544FAX: (317) 521-2324Email: angelo.pereira@roche.com
Date PreparedJuly 21, 2016
Proprietary NameElecsys Syphilis
Common NameSyphilis assay
Classification NameTreponema pallidum treponemal test reagent
Product CodesProduct Code: LIP; 21CFR866.3830Control Product Code: JJX
Predicate DevicesImmulite 2000 Syphilis Screen test system, K091361
Establishment RegistrationFor Elecsys Syphilis, the establishment registration number for RocheDiagnostics GmbH in Mannheim, Germany is 9610126, and for Penzberg,Germany, 9610529. The establishment registration number for RocheDiagnostics in the United States is 1823260.

DEVICE DESCRIPTION 1.

The Elecsys Syphilis assay is a fully automated qualitative assay detecting IgG and IgM antibodies to Treponema pallidum, the causative agent of syphilis. Assay results, in conjunction

{4}------------------------------------------------

with other laboratory results and clinical information, may be used to provide presumptive evidence of active or previous infection with Treponema pallidum in persons with signs and symptoms of syphilis, as well as in patients at risk for syphilis infection. This assay does not determine the stage of infection or associated disease.

PreciControl Syphilis is a lyophilized control based on human serum. It is used for monitoring the accuracy of the Elecsys Syphilis immunoassay.

2. INTENDED USE

Elecsys Syphilis:

Immunoassay for the in vitro qualitative determination of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma. The test is intended as an aid in the diagnosis of syphilis infection in conjunction with clinical signs and symptoms.

The Elecsys Syphilis immunoassay is not intended for use in screening blood or tissue donors. The effectiveness of this assay in testing of blood and tissue donors has not been established.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 analyzer.

PreciControl Syphilis:

PreciControl Syphilis is intended for the quality control of the Elecsys Syphilis immunoassay on cobas e 411 analyzer.

{5}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS 3.

The Elecsys Syphilis assay is a qualitative assay that uses the established double antigen sandwich format. Biotinylated T. pallidum-specific recombinant antigens (TpN15, TpN17 and TpN47) and T. pallidum-specific recombinant antigens (TpN15, TpN17 and TpN47) labeled with a ruthenium complex react with anti T. pallidum-specific antibodies to form a sandwich complex. The assay uses monomeric and polymeric antigens, ensuring the detection of both IgG and IgM anti-treponemal antibodies. After addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically bound onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.

The predicate device is a solid phase, one-step chemiluminescent enzyme immunoassay. The solid phase (bead) is coated with purified recombinant T. pallidum 17 (Tp17) antigen. The liguid phase consists of alkaline phosphatase (bovine calf intestine) conjugated to purified recombinant T. pallidum 17 (Tp17) antigen.

Patient sample and the reagent are incubated together with the coated bead. During this time, total antibody to T. pallidum in the sample forms the antigen sandwich complex with purified recombinant T. pallidum 17 (Tp17) antigen on the bead and enzyme conjugated purified recombinant T. pallidum 17 (Tp17) antigen in the reagent. Unbound patient sample and enzyme conjugate are then removed by centrifugal washes. Finally, chemiluminescent substrate is added to the reaction tube containing the bead and the signal is generated in proportion to the bound enzyme.

Both the predicate device and Elecsys Syphilis provide prepackaged reagents, calibrators and controls for use on automated test systems. A comparison of the important similarities and differences of these assays is provided in the following table:

{6}------------------------------------------------

Table of similarities and differences

ParameterNew DevicePredicate
Intended UseImmunoassay for the in vitroqualitative determination of totalantibodies (IgG and IgM) toTreponema pallidum in human serumand plasma. The test is intended as anaid in the diagnosis of syphilis infectionin conjunction with clinical signs andsymptoms.The Elecsys Syphilis assay is notintended for use in screening blood orplasma donors.The electrochemiluminescenceimmunoassay "ECLIA" is intended foruse on cobas e 411 analyzerThe IMMULITE 2000 Syphilis Screentest is a treponemal testing procedurefor the qualitative detection ofantibodies to Treponema pallidum inhuman serum or heparinized plasmaon the IMMULITE 2000 analyzer asan aid I the diagnosis of syphilis.The IMMULITE 2000 Syphilis Screentest is not intended for use inscreening blood or plasma donors.
Analytes measuredAntibodies to Treponema pallidumsame
InstrumentCobas e 411IMMULITE 2000 Systems
MeasurementQualitativesame
Antigens usedRecombinant antigens TpN17,TpN15 andTpN47Recombinant antigen Tp17
Cut-off<1.00 Non-reactive≥1.00 Reactive<0.9 Non-reactive≥0.9 to <1.1 Indeterminate≥1.1 Reactive
Sample typeSerum, K2 EDTA, K3 EDTA, CPDA,NaCitrate and Li HeparinSerum, heparinized plasma
Assay typeDouble antigen sandwichelectrochemiluminescenceimmunoassayEnzyme labeled, one-stepchemiluminescent immunoassay

4. NON-CLINICAL PERFORMANCE EVALUATION

Precision

Precision was evaluated with serum samples on a single cobas e 411 Immunoassay Analyzer according to CLSI guideline EP5-A2. One reagent lot was evaluated. The protocol was performed by testing 2 controls (PreciControl 1 and PreciControl 2) and 6 human sera (HS) (negative, near cut-off, positive) in duplicate per run, 2 runs per day for 21 days. The samples were run in randomized order on the analyzer.

{7}------------------------------------------------

RepeatabilityIntermediate precision
SampleMean [COI]SD[COl]CV [%](UCL* 95%)SD[COI]CV [%](UCL* 95%)n
HS negative 10.1030.0021.6 (2.0)0.0033.2 (4.1)84
HS negative 20.8210.01742.1 (2.7)0.0192.3 (2.7)84
HS positive 11.010.0282.8 (3.5)0.0333.2 (3.9)84
HS positive 21.120.0181.6 (2.1)0.0221.9 (2.3)84
HS positive 39.990.1711.7 (2.2)0.2622.6 (3.2)84
HS positive 450.20.9862.0 (2.5)1.242.5 (3.0)84
PreciControl 10.1060.0032.4 (3.1)0.0044.1 (5.1)84
PreciControl 24.950.1012.1 (2.6)0.1613.2 (4.1)84

Endogenous Interferences

Interference by endogenous substances was evaluated at three anti-T. pallidum antibody concentrations (negative, near cut-off and positive). Each serum pool was spiked with the interferent to the levels indicated in the table below.

Interfering substancemeasured up toNo interference seen uptoLabel Claim
Intralipid® (Lipemia)2000 mg/dL2000 mg/dL2000 mg/dL
Biotin70 ng/mL70 ng/mL60 ng/mL
Bilirubin66 mg/dL66 mg/dL66 mg/dL
Hemoglobin1000 mg/dL1000 mg/dL500 mg/dL
Rheumatoid Factor1500 IU/mL1500 IU/mL1500 IU/mL
human serum albumin10 g/dL10 g/dL10 g/dL
human IgG7.0 g/dL6.3 g/dL32 g/L
human IgM1.0 g/dL1.0 g/dL10 g/L
human IgA7.0 g/dL2.8 g/dL2.8 g/dL

Drug interferences

16 pharmaceutical compounds were spiked into human serum samples at three anti-T. pallidum antibody concentrations (negative, low-positive, high positive) and tested with the Elecsys Syphilis assay on the cobas e 411 analyzer. The concentration of the spiked aliquots was determined in 3-fold determination and compared to the anti-T. pallidum antibody concentration

{8}------------------------------------------------

determined for the reference aliquot (also in 3-fold determination) on one cobas e 411 analyzer. No interference was observed at the levels shown in the table below.

Drug testedConc. of spiked drug tested (mg/L)
Acetylcystein150
Ampicillin-Na1000
Ascorbic acid300
Cyclosporine5
Cefoxitin2500
Heparin5000 U/L
Levodopa20
Methyldopa +1.520
Metronidazole200
Phenylbutazone400
Doxycyclin50
Acetylsalicylic Acid1000
Rifampicin60
Acetaminophen200
Ibuprofen500
Theophylline100

Serum Plasma Matrix comparison

The effect of anticoagulants on the Elecsys Syphilis Immunoassay was determined by comparing values obtained from native samples spiked with anti-T. pallidum antibody (single donors) drawn into serum and plasma tubes.

The data support the package insert claim that serum, serum separation tubes, Li-Heparin plasma tubes, K2-EDTA- and K3-EDTA-plasma, Na Citrate and CPDA plasma specimens are acceptable sample types for use with the Elecsys Syphilis assay.

Analytical specificity

The analytical specificity of the Elecsys Syphilis assay was determined using human serum samples (single donors, native) spiked with potential cross-reactant antibodies.

{9}------------------------------------------------

The samples were tested in duplicates on the cobas e 411 Immunoassay Analyzer. The analytical specificity of the assay was determined using 266 human clinical samples from patients with medical conditions unrelated to Syphilis. The 9 samples that were reactive with the Elecsys Syphilis assay were confirmed as co-infected and positive for anti-treponema antibodies with another syphilis total antibody test. No cross-reactivity was found.

PotentialinterferenceNumbertestedNumberpositivesamples
Borrelia500
EBV IgG211
Rubella IgG171
E.coli antibodies181
Hepatitis A251
Hepatitis B102
Hepatitis C130
ANA151
RF190
HSV 1/2121
CMV IgG200
HIV-1251
Toxo IgG210
Total2669

5. CUT-OFF DETERMINATION

In order to define the cut-off for the Elecsys Syphilis assay, native human serum samples were measured with a prototype lot of the Elecsys Syphilis assay and a preliminary cut-off was set.

Next, the comparability in patient samples to commercially available assays was reviewed.

Final validation of the cut-off was done in multi-center studies conducted in the EU and the US.

The classification of samples is as follows:

Sample result < 1.0 COIsample is regarded as non-reactive
Sample result ≥ 1.0 COIsample is regarded as reactive

{10}------------------------------------------------

CLINICAL PERFORMANCE 6.

Reproducibility

A reproducibility study was conducted following CLSI EP5-A2 at three sites incorporating a 7 member panel consisting of 5 serum pools (high negative, low positive and moderately positive) and 2 controls that were assayed for 5 days, 2 runs per day, 3 replicates per run. Data from all three sites were combined to achieve SD and percent CV for repeatability (within-run), between run, between-day, between site and reproducibility. The overall reproducibility (imprecision) data are summarized in the following table:

MeanWithin RunRepeatabilityBetweenRunBetweenDayBetweenSiteReproducibility
SampleNCOISD95% CI% CVSD% CVSD% CVSD% CVSD95% CI% CV
HSP 06901.020.01(0.01, 0.01)1.160.022.280.000.000.011.290.03(0.02, 0.04)2.86
HSP 07901.130.02(0.02, 0.03)1.900.032.570.000.000.010.790.04(0.03, 0.05)3.29
HSP 08900.940.01(0.01, 0.02)1.580.022.550.000.000.010.750.03(0.02, 0.04)3.09
HSP 09900.850.02(0.01, 0.02)1.880.022.490.000.000.011.230.03(0.02, 0.04)3.36
HSP 10903.240.09(0.08, 0.11)2.780.082.610.000.000.000.000.12(0.11, 0.15)3.82
SYPHPC1900.120.00(0.00, 0.00)2.410.002.390.001.150.014.470.01(0.00, 0.02)5.73
SYPHPC2904.760.08(0.07, 0.10)1.650.142.890.000.000.030.700.16(0.13, 0.21)3.40

Clinical performance in prospective cohorts

A total of 2282 prospectively collected samples of the intended use population were tested at 3 sites using the Elecsys Syphilis assay, including 1524 subjects for routine syphilis testing, 457 HIV positive subjects and 301 pregnant women. In addition, all samples were tested according to a composite algorithm using FDA-cleared tests that included the predicate Syphilis immunoassay, the Rapid Plasma Reagin (RPR) non-treponemal specific assay and the Treponema pallidum Particle Agglutination, treponemal-specific assay.

The following table summarizes the performance of Elecsys Syphilis compared with the final comparator results for all prospective cohorts.

{11}------------------------------------------------

Final Comparator Results
Elecsys SyphilisPositive for SyphilisNegative for SyphilisTotal
Reactive22816244
Non reactive020382038
Total22820542282

Elecsys Syphilis compared with Final Comparator for all prospective cohorts

Percent agreement by category for prospective cohorts

CohortPositive PercentAgreementPPA95 % CINegative PercentAgreementNPA95 % CI
%Ratio%Ratio
RoutineSyphilis10066/6694.56 to 100.0099.81455/145899.40 to 99.96
HIV100162/16297.75 to 100.0095.6282/29592.58 to 97.63
PregnantN/A0/0N/A100301/30198.78 to 100.00
Total100228/22898.40 to 100.0099.22038/205498.74 to 99.55

Clinical performance in retrospective medically diagnosed individuals

Clinical performance in the pre-selected retrospective cohort included a total of 169 specimens, including 15 pregnant positive women and 154 subjects medically diagnosed with syphilis at different stages. The comparison between the Elecsys Syphilis results and the final comparator results is given in the following table.

Elecsys Syphilis compared with Final Comparator for retrospective cohorts
Final Comparator Results
Elecsys SyphilisPositive for SyphilisNegative for SyphilisTotal
Reactive1550155
Non-reactive21214
Total15712169

{12}------------------------------------------------

CohortPositive PercentAgreementPPA95 % CINegative PercentAgreementNPA95 % CI
%Ratio%Ratio
Pregnant(Retrospective)10015/1578.20 to100.00N/A0/0N/A
Staged98.6140/14295.00 to 99.8310012/1273.54 to100.00
Overall(Retrospective)98.7155/15795.47 to 99.8510012/1273.54 to 99.55

Percent agreement by category for retrospective cohorts

Clinical performance in pregnant women

A total of 316 pregnant female samples were tested in the study. Of these, 301 were prospectively collected and 15 were retrospectively collected. The percent agreement between the Elecsys Syphilis results and the final comparator results is shown below, stratified by pregnancy trimesters compared with the Composite Algorithm for pregnant women:

Percent agreement of Elecsys Syphilis with final comparator results for Pregnant Women
CohortPositive PercentAgreementPPA95 % CINegative PercentAgreementNPA95 % CI
%%
Pregnant(Prospective)N/AN/A10098.78 to100.00
1st TrimesterN/AN/A10096.38 to100.00
2nd TrimesterN/AN/A10097.09 to100.00
3rd TrimesterN/AN/A10095.26 to100.00
Pregnant(Retrospective)10078.20 to100.00N/AN/A
1st Trimester10059.04 to100.00N/AN/A
3rd Trimester10063.06 to100.00N/AN/A
RatioRatio
Pregnant(Prospective)0/0N/A301/301
1st Trimester0/0N/A100/100
2nd Trimester0/0N/A125/125
3rd Trimester0/0N/A76/76
Pregnant(Retrospective)15/150/0N/A
1st Trimester7/70/0N/A
3rd Trimester8/80/0N/A

{13}------------------------------------------------

CohortPositive PercentAgreementPPA95 % CINegative PercentAgreementNPA95 % CI
%Ratio%Ratio
100.00

Clinical performance in Medically Diagnosed Individuals

Samples were collected from 154 individuals diagnosed with primary, secondary or latent syphilis. They included 10 females and 144 males. Results of the Elecsys Syphilis assay for this cohort are summarized below.

Syphilis stageTreatment statusNReactiveNon-reactive
Elecsys Syphilis Results
PrimaryTreated291613*
PrimaryUntreated25250
SecondaryTreated25241
SecondaryUntreated25250
LatentTreated25250
LatentUntreated25250

Reactivity of Elecsys Syphilis in subjects medically diagnosed with syphilis

  • 12 of these samples also tested negative for Syphilis with the composite testing algorithm

Performance in apparently healthy individuals

Specimens were collected from 209 apparently healthy individuals. Of these, 80 were female and 129 male. The results of the Elecsys Syphilis assay for this cohort are shown below

Elecsys Syphilis Results
ReactiveNon-reactive
NN
Female9 (11.3 %)71 (88.7 %)80
Male11 (8.5 %)118 (91.5 %)129
Total20 (9.6 %)189 (90.4 %)209

{14}------------------------------------------------

Expected values

In this clinical study there were 2282 prospectively collected specimens for the intended use population that were tested with the Elecsys Syphilis assay. There were 244 reactive samples for a 10.7% prevalence of T. pallidum antibodies in the study population. The distribution of the Elecsys Syphilis reactive and non-reactive results is summarized below by age and gender.

Age Range(years)Elecsys Syphilis Results
GenderReactiveNon-reactiveTotal
18 21Female0 (0.00 %)247 (100 %)247
Male6 (3.5 %)165 (96.5 %)171
22 29Female0 (0.00 %)358 (100 %)358
Male19 (10.7 %)159 (89.3 %)178
30 39Female10 (3.7 %)261 (96.3 %)271
Male13 (12.3 %)93 (87.7 %)106
40 49Female29 (12.1 %)210 (87.9 %)239
Male29 (20.0 %)116 (80.0 %)145
50 59Female47 (20.8 %)179 (79.2 %)226
Male55 (27.1 %)148 (72.9 %)203
60 69Female11 (23.9 %)35 (76.1 %)46
Male23 (29.5 %)55 (70.5 %)78
70 79Female0 (0.00 %)5 (100 %)5
Male2 (22.2 %)7 (77.8 %)9
CombinedTotal244 (10.7 %)2038 (89.3 %)2282

{15}------------------------------------------------

CONCLUSIONS 7.

There was good agreement between the Elecsys Syphilis assay and the results from the final comparator algorithm.

A comparison of the device features, intended use, non-clinical and clinical data supports a decision that Elecsys Syphilis is substantially equivalent to the predicate assay.

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).