K Number
K160882

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-10-28

(211 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PaX-i3D Green Premium (Model: PCT-90LH) is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the maxillofacial areas for a dental and ENT treatment in adult and pediatric dentistry. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by physicians, dentists, ENT specialist and x-ray technicians.

Device Description

PaX-i3D Green Premium (Model: PCT-90LH), a digital radiographic imaging system, acquires and processes multi FOV diagnostic images for a dentist and an ENT specifically designed for dental and ENT radiography, PaX-i3D Green Premium (Model: PCT-90LH) is a complete digital X-ray system equipped with imaging viewers, Xray generator and a dedicated SSXI detector. The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of head, neck and craniofacial anatomy for ENT, oral surgery, implant and orthodontic treatment. PaX-i3D Green Premium (Model: PCT-90LH) can also acquire 2D diagnostic image data in panoramic and cephalometric mode.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for the PaX-i3D Green Premium (Model: PCT-90LH) digital X-ray system, comparing it to a predicate device, PaX-i3D Ortho (TON-95LH).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with specific numerical targets. Instead, it describes acceptance as "performed similarly or better" than the predicate device and that "all test results were satisfactory." The key performance indicators evaluated are:

Acceptance Criterion (Based on Predicate Performance)Reported Device Performance (PaX-i3D Green Premium)
Modulation Transfer Function (MTF)Performed similarly or better than predicate device (Xmaru2430CF Master Plus)
Detective Quantum Efficiency (DQE)Performed similarly or better than predicate device (Xmaru2430CF Master Plus)
Normalized Noise Power Spectrum (NNPS)Similar to predicate device (Xmaru2430CF Master Plus)
Contrast (CT images)Equivalent or better than predicate device
Noise (CT images)Equivalent or better than predicate device
Contrast-to-Noise Ratio (CNR) (CT images)Equivalent or better than predicate device
MTF (CT images)Equivalent or better than predicate device
Noise levels (CT images with iterative reconstruction)Less than or equal to images produced by standard filtered back projection reconstruction
Conformance to 21 CFR Part 1020.33 and IEC 61223-3-5Satisfactory
Conformance to IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63 (electrical, mechanical, environmental safety/performance)Satisfactory
Conformance to IEC 60601-1-2 (EMC)Satisfactory
Conformance to NEMA PS 3.1-3.18 (DICOM)Satisfactory

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in the context of clinical images or specific patient data. The performance evaluations were conducted through non-clinical bench testing:

  • Sample Size: Not explicitly stated as a number of images or cases. The tests were performed on the device's components (detector) and the finished X-ray equipment.
  • Data Provenance: This was retrospective in the sense that the reference point for performance was the previously cleared predicate device. The tests themselves were conducted in a laboratory setting. The country of origin of the data is implied to be from the manufacturer (VATECH Co., Ltd., based in Korea) as it states "The sponsor tested the subject device in a laboratory and provided a non-clinical performance report."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

No explicit information is provided regarding the number or qualifications of experts used to establish ground truth for this non-clinical performance evaluation. The "ground truth" for the device's technical performance metrics (MTF, DQE, NNPS, Contrast, Noise, CNR) was established through objective physical measurements using established international standards (e.g., IEC 61223-3-5).

4. Adjudication Method for the Test Set

Not applicable. The performance evaluation was primarily objective, based on physical measurements of device characteristics against established standards and comparison to a predicate device's measured performance. There was no mention of human adjudication for the technical performance metrics.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This submission focuses on the substantial equivalence of an X-ray imaging system, not on an AI-driven image analysis or diagnostic aid that would involve human reader performance improvement.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

Yes, aspects of a "standalone" performance were evaluated, particularly for the iterative reconstruction algorithm. The text states: "The safety and effectiveness of iterative reconstruction algorithm have been demonstrated and confirmed by the previous 510k submission of the reference device, PaX-i3D Smart (PHT-30LFO) (K152106)." This implies an evaluation of the algorithm's output (image quality parameters such as noise reduction) independent of a human reader's diagnostic performance. The document explicitly focuses on the "algorithm only" performance for noise reduction in CT images.

7. The Type of Ground Truth Used

The ground truth used for the technical performance metrics (MTF, DQE, NNPS, Contrast, Noise, CNR) was objective physical measurement based on established international standards (e.g., IEC 61223-3-5). For the iterative reconstruction algorithm, the "ground truth" for its effectiveness was implicitly defined by its ability to produce images with "noise levels less than or equal to images produced by standard filtered back projection reconstruction."

8. The Sample Size for the Training Set

Not applicable. This document describes the performance evaluation of an X-ray imaging system, not a machine learning or AI model that would require a dedicated training set. The "iterative reconstruction algorithm" mentioned is a signal processing algorithm, not a learning-based AI algorithm in the typical sense that would necessitate a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there was no training set in the context of machine learning.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, resembling a family or group of people.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2016

VATECH Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K160882

Trade/Device Name: PaX-i3D Green Premium (Model: PCT-90LH) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: September 22, 2016 Received: September 29, 2016

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160882

Device Name PaX-i3D Green Premium (Model: PCT-90LH)

Indications for Use (Describe)

PaX-i3D Green Premium (Model: PCT-90LH) is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the maxillofacial areas for a dental and ENT treatment in adult and pediatric dentistry. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by physicians, dentists, ENT specialist and x-ray technicians.

Type of Use (Select one or both, as applicable)
( Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 7 – 510(k) Summary

1. 510(k) Summary

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.

2. Date 510K Summary prepared: September 22, 2016

3. Administrative Information

Official Correspondent:Dave Kim / Mtech Group
8310 Buffalo Speedway, Houston, TX 77025
Tel: +713-467-2607
Fax: +713-583-8988
Contact person: Mr. Dave Kim (davekim@mtech-inc.net)
510(k) Submitter:VATECH Co., Ltd.
Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, Korea
Tel: +82-31-379-9492
Fax: +82-31-379-9400
Contact person: Daniel Kim / Manager (daniel.kim@vatech.co.kr)

4. Device Information

Type of 510(k) Submission:Traditional
Trade or Proprietary Name:PaX-i3D Green Premium (Model: PCT-90LH)
Common or Usual Name:Dental Computed Tomography X-ray System
Regulation Classification:Computed tomography x-ray system (21 CFR 892.1750)
Product Code:OAS
Class of Device:Class II
Panel:Radiology

5. Predicate Device Information

Manufacturer:VATECH Co., Ltd.
Predicate device:PaX-i3D Ortho (TON-95LH) / K132983
Common or Usual Name:X-Ray, Tomography, Computed, Dental
Regulation Classification:Computed tomography x-ray system (21 CFR 892.1750)
Product Code:OAS
Class of Device:Class II
Panel:Radiology

※ This predicate has not been subject to a design-related recall.

{4}------------------------------------------------

6. Reference Device Information

Manufacturer:VATECH Co., Ltd.
Reference device:PaX-i3D Smart (PHT-30LFO) / K152106
Common or Usual Name:X-Ray, Tomography, Computed, Dental
Regulation Classification:Computed tomography x-ray system (21 CFR 892.1750)
Product Code:OAS
Class of Device:Class II
Panel:Radiology

※ This predicate has not been subject to a design-related recall.

7. Device Description

PaX-i3D Green Premium (Model: PCT-90LH), a digital radiographic imaging system, acquires and processes multi FOV diagnostic images for a dentist and an ENT specifically designed for dental and ENT radiography, PaX-i3D Green Premium (Model: PCT-90LH) is a complete digital X-ray system equipped with imaging viewers, Xray generator and a dedicated SSXI detector.

The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of head, neck and craniofacial anatomy for ENT, oral surgery, implant and orthodontic treatment. PaX-i3D Green Premium (Model: PCT-90LH) can also acquire 2D diagnostic image data in panoramic and cephalometric mode.

Key components of the device

    1. PaX-i3D Green Premium (Model: PCT-90LH) digital x-ray equipment
    1. SSXI detector: Xmaru3104CF
ItemDescription
ModelXmaru3104CF
Detector TypeCMOS photodiode array
Pixel Size99 µm @ 2X2 binning : PANO
198 µm @ 4X4 binning : Dental CT, ENT CT, i-CEPH
Active Area310.4 x 36.4 mm
Frame Rate107 fps (2x2 Binning) : PANO
308 fps (4x4 Binning) : Dental CT, ENT CT, i-CEPH
Analogue-Digital Conversion14 bits
ConverterCsl:Ti
Energy Range60 ~ 120 kV
Readout TypeCharge amplifier array
Video OutputOptic

3) X-ray generator

- Option 1

ItemDescription
High VoltageGeneratorModelDG-08C22S1
Rated output power1,200 W
TypeInverter
Normal/Pulse kV60 ~ 120 kV

{5}------------------------------------------------

ItemDescription
mA4 ~ 10 mA
CoolingAir (Optional fan cooling, ≥ 60 ℃ )
Total filtrationMin. 2.5 mm Al
Added filtration2.0 mm Al
X-ray TubeManufacturerSuperior
ModelSXR-130-15-0.5 (Stationary Anode type)
Focal spot size0.5 mm
Target Angle15 degree
Inherent FiltrationAt least 1.1 mm Al equivalent at 80 kV
Anode Heat Content21 kJ
Duty Cycle1:20 or more (Exposure time : Interval time)

- Option 2

ItemDescription
High VoltageGeneratorModelDG-08C22S2
Rated output power1,200 W
TypeInverter
Normal/PulsekVmA60 ~ 120 kV4 ~ 10 mA
CoolingAir (Optional fan cooling, ≥ 60 °C)
Total filtrationMin. 2.5 mm Al
Added filtration2.0 mm Al
X-ray TubeManufacturerC.E.I
ModelOX/115-05 (Stationary Anode type)
Focal spot size0.5 mm
Target Angle15 degree
Inherent Filtration0.5mm Al
Anode Heat Content30 kJ
Duty Cycle1:60 or more (Exposure time : Interval time)

4) PC system

ItemDescription
Operating SystemWindows 8.1 Professional 64-Bit OS
CPUIntel Xeon E5-1620v3 3.5GHz 2133 10MB cache CPU
RAM32GB DDR4-2133 ECC RAM
HDD1TB SATA 1st HDD
Graphics boardNVIDIA GEFORCE GTX980 Ti OC D5 6GB or greater
Ethernet interfaceBroadcom 5761 Gigabit PCIe NIC
Serial Port (RS232)HP Serial Port Adapter Kit
Power Supply≥ 700 Watts (90% Efficiency)
Slots1 PCI Express Gen3 x 8 Slot2 PCI Express Gen3 x 16 slot1 PCI Express Gen2 x 8 Slot1 PCI Express Gen2 x 4 Slot

5) Imaging software

  • ① EzDent-i: 2D viewer and patient management software
    |

  • ・ Manufacturer: EWOO SOFT Co., Ltd.

{6}------------------------------------------------

  • · 510(k) Number: K150747
  • · Classification Number: 21CFR892.2050
  • Product code: LLZ
  • ② Ez3D-i: 3D image viewer
    • Manufacturer: EWOO SOFT Co., Ltd.
    • · 510(k) Number: K150761
    • · Classification Number: 21CFR892.2050
    • · Product code: LLZ

8. Indication for use

PaX-i3D Green Premium (Model: PCT-90LH) is intended to produce panoramic, cephalometric or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, ENT, sinus and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.

{7}------------------------------------------------

Subject DevicePredicate DeviceReference Device
Device NamePaX-i3D Green Premium(Model: PCT-90LH)PaX-i3D Ortho (TON-95LH)PaX-i3D Smart (PHT-30LFO)
Applicant NameVATECH Co., Ltd.VATECH Co., Ltd.VATECH Co., Ltd.
510(k) NumberN/AK132983K152106
Device Classification NameX-Ray, Tomography, Computed,DentalX-Ray, Tomography, Computed,DentalX-Ray, Tomography, Computed,Dental
Classification Product CodeOASOASOAS
Regulation Number21 CFR 892.175021 CFR 892.175021 CFR 892.1750
Indications for UsePaX-i3D Green Premium (Model: PCT-90LH) is intended to producepanoramic, cephalometric or 3Ddigital x-ray images.It provides diagnostic details of thedento-maxillofacial, ENT, sinus andTMJ for adult and pediatric patients.The system also utilizes carpal imagesfor orthodontic treatment. The deviceis to be operated by healthcareprofessionals.PaX-i3D Ortho (TON-95LH) is acomputed tomography x-ray systemintended to produce cross-sectionalimages of the oral anatomy byreconstructing a three dimensionalradiographic images from the sameaxial plane taken at different angles.It provides diagnostic details of themaxillofacial areas for a dentaltreatment in adult and pediatricdentistry. The system also acquirescarpal images for orthodontictreatment. The device is operatedand used by physicians, dentists,and x-ray technicians.PHT-30LFO is a computed tomographyx-ray system intended to producepanoramic, cephalometric or cross-sectional images of the oralanatomy on a real time basis bycomputer reconstruction of x-rayimage data from the same axialplane taken at different angles. Itprovides diagnostic details of theanatomic structures by acquiring360° rotational image sequences oforal and maxillofacial area for aprecise treatment planning in adultand pediatric dentistry. The device isoperated and used by physicians,dentists, and x-ray technicians.
Performance SpecificationPanoramic, Cephalometric andcomputed tomographyComputed tomographyPanoramic, cephalometric andcomputed tomography
Input VoltageAC 100 - 240 VAC 100 - 240 VAC 100 - 240 V
X-Ray source(Option 1) SXR-130-15-0.5(Option 2) OX/115-05SXR-130-15-0.5D-052SB
Tube Voltage60 - 120 kV50 - 120 kV50 - 99 kV
Tube Current4 - 10 mA4 - 10 mA4 - 16 mA
Focal Spot Size0.5 mm0.5 mm
Exposure TimeMax. 18 sMax. 18 s
Slice WidthMin. 0.1 mmMin. 0.1 mmMin. 0.1 mm
Total Filtration(Option 1) Min. 2.5 mm AlMin. 2.8 mm Al2.8 mm Al
(Option 2) Min. 2.5 mm Al
MechanicalCompact designCompact designCompact design
ElectricalLDCP logic circuitLDCP logic circuitLDCP logic circuit
SoftwareDICOM 3.0 Format compatibleDICOM 3.0 Format compatibleDICOM 3.0 Format compatible
2D Image Viewing ProgramEzDent-iEzDent-iEasyDent
3D Image Viewing ProgramEz3D-iEz3D-i OrthoEz3D Plus
Anatomical SitesMaxillofacialMaxillofacialMaxillofacial
ImageReceptorComputedTomographyXmaru2430CF Master PlusXmaru1404CF
PanoramicXmaru3104CFReconstructed from 3D image data
CephalometricReconstructed from 3D image dataXmaru2301CF1210SGA910SGAXmaru2301CF-O
Size of Imaging VolumeMax. 210 x 190 mmMax. 240 x 210 mmXmaru1404CF: Max. 10 x 8.5 mm
PixelResolutionComputedTomography2.5 lp/mm -4x4 binning5 lp/mm -2x2 binning2.5 lp/mm -4x4 binningXmaru1404CF:- 5.0 lp/mm - 2x2 binning- 2.5 lp/mm - 4x4 binning
Panoramic5 lp/mm -2x2 binningN/AXmaru1404CF:- 5.0 lp/mm - 2x2 binning- 2.5 lp/mm - 4x4 binning
Cephalometric2.5 lp/mm -4x4 binningN/AXmaru2301CF: 5 lp/mm1210SGA: 3.9 lp/mm910SGA: 3.9 lp/mmXmaru2301CF-O: 5 lp/mm
Pixel SizeComputedTomography198 $\mu$ m -4x4 binning99 $\mu$ m-2x2 binning198 $\mu$ m-4x4 binningXmaru1404CF:- 99 $\mu$ m- 2x2 binning- 198 $\mu$ m- 4x4 binning
Panoramic99 $\mu$ m -2X2 binningN/AXmaru1404CF:- 99 $\mu$ m- 2x2 binning- 198 $\mu$ m- 4x4 binning
Cephalometric198 $\mu$ m -4x4 binningN/AXmaru2301CF: 100 x 100 $\mu$ m1210SGA: 127 x 127 $\mu$ m910SGA: 127 x 127 $\mu$ mXmaru2301CF-O: 100 x 100 $\mu$ m

9. Comparison of Technological characteristics with the predicate device



{8}------------------------------------------------

VATECH Co., Ltd. Premarket Notification 510(k)

{9}------------------------------------------------

10. Performance Data

Summary of Performance Testing

The PaX-i3D Green Premium (Model: PCT-90LH) digital X-ray system described in this 510(k) is similar to the predicate device, PaX-i3D Ortho (TON-95LH) in terms of indications for use, materials, safety characteristics, and X-ray source.

Furthermore, the following information further substantial equivalence between the subject device and predicate device:

The fundamental technological characteristics of the subject and predicate device are similar.

The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the predicated device for the purpose of comparing the performance between the predicate device. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.

The differences are as follows.

  • The subject device is equipped with Xmaru3104CF, a SSXI detector, which is used to capture PANO, i-। CEPH, CT (Dental / ENT) images.
  • । The subject device utilizes iterative reconstruction algorithm, a CBCT imaging processing algorithm.
  • -In addition to the Dental CT mode, the subject device imaging modes; PANO, i-CEPH, ENT CT and 3D Photo.
  • -The subject device includes imaging viewer programs; EzDent-i (K150747) and Ez3D-i (K150761).

PaX-i3D Green Premium (Model: PCT-90LH), a digital radiographic imaging system is equipped with Xmaru3104CF, a SSXI detector, which is used to capture PANO, i-CEPH, Dental CT and ENT CT images. It has different active areas compared to Xmaru2430CF Master Plus of PaX-i3D Ortho (TON-95LH). Xmaru3104CF detector of the subject device offers different data output interface compared to the Xmaru2430CF Master Plus for PaX-i3D Ortho (TON-95LH). The added imaging modes of the subject device; PANO, i-CEPH and ENT CT require different radiograph exposure settings and field of view. However, the operating principles and indications for use of the radiograph imaging system remains similar as the predicate device. Both devices have the 3D photo mode which takes a 3D facial photograph of a patient and does not require X-ray exposure.

Non-clinical test was conducted for Xmaru3104CF of PaX-i3D Green Premium (Model: PCT-90LH). More specifically, the Modulation Transfer Function (MTF), Detective Quantum Efficiency (DQE) and Normalized Noise Power Spectrum (NNPS) performance were measured for Xmaru3104CF of the subject device and Xmaru2430CF Master Plus for the predicate device. (MTF and DQE results are objective performance evaluation parameters for a digital detector. MTF measures the resolution and expressed by the contrast difference in percentage. DQE represents the overall index for the detector's ability to converting incident X-ray into an image signal without much loss to the actual image object.) Based on the Non-Clinical Test results, the Xmaru3104CF detector has performed similarly or better than the predicate device, Xmaru2430CF Master Plus, in terms of the overall MTF and DQE performance.

The results of NNPS performance for Xmaru2430CF-Master Plus were similar. The change in the data output interface has not caused any changes in the performance or the level of artifacts. All performance parameters for both detectors have shown similar results.

The acceptance test was done according to the requirements of 21 CFR Part 1020.33 and IEC 61223-3-5, international performance standard for computed tomography X-ray system.

{10}------------------------------------------------

In addition, CT Image evaluation was performed on the finished X-ray equipment with the detector mounted. Contrast, Noise, CNR, and MTF, the representative indicators for 2D and 3D CT image quality were measured with iterative reconstruction algorithm. The results demonstrated that the general image quality of the subject device is equivalent or better than the predicate device.

The CBCT image reconstruction for PaX-i3D Green Premium (Model: PCT-90LH) applies iterative reconstruction algorithm for the reduction of noise in the CT image capture mode. Iterative reconstruction algorithm reconstructs raw data from the PaX-i3D Green Premium (Model: PCT-90LH) CBCT system to produce images containing noise levels less than or equal to images produced by standard filtered back projection reconstruction. The safety and effectiveness of iterative reconstruction algorithm have been demonstrated and confirmed by the previous 510k submission of the reference device, PaX-i3D Smart (PHT-30LFO) (K152106).

Clinical Data

The clinical considerations on the images in terms of quality and effectiveness sustained the diagnostic reliability and usefulness of images produced by the PaX-i3D Green Premium (Model: PCT-90LH) applications.

Software Verification and Validation Testing

Software verification and validation were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.

PaX-i3D Green Premium (Model: PCT-90LH) utilizes imaging viewer programs; EzDent-i (K150747), Ez3D-i (K150761), both with FDA clearance.

Safety, EMC and Performance Data

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1(Ed. 3, 2005), IEC 60601-1-3 (Ed. 2, 2008), IEC 60601-2-63 (Ed. 1, 2012) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2.

The manufacturing facility is in conformance with the relevant EPRC standards as specified in 21 CFR 1020.30, 31, and 33 and the records are available for review.

PaX-i3D Green Premium (Model: PCT-90LH) conforms to the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.

Non-clinical consideration report according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" was provided.

Bench testing according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 18, Performance Testing - Bench" were performed.

Acceptance test and CT image evaluation report according to IEC 61223-3-4 and IEC 61223-3-5 were also performed.

All test results were satisfactory.

11. Conclusions

The subject device and the predicate device have similar indications for use and demonstrated similar technical characteristics. As demonstrated in the performance test, the Xmaru3104CF performed similar or better than the predicate device in various performance parameters such as DQE, MTF and NNPS. In addition, 2D and 3D CT

{11}------------------------------------------------

image evaluation of Contrast, Noise, CNR, and MTF further demonstrated the performance equivalency between the subject and predicate device. Quality assurance procedures are adhered to, and the specifications and functional requirements were met as the test results indicated.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, VATECH Co., Ltd. concludes that PaX-i3D Green Premium (Model: PCT-90LH) is substantially equivalent to the predicate device as described herein.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.