(126 days)
The SpineView X-Pac Expandable Lumbar Cage System is indicated for use as an intervertebral body fusion device in skeletally mature patients. The system is designed for use with autogenous bone graft material to facilitate spinal fusion. The system is intended for use in patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have at least 6 months of non-operative treatment prior to surgery. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via transforaminal approach using autogenous bone graft material. The system is intended for use with supplemental fixation systems cleared for use by the FDA for use in the thoracolumbar spine.
The Spine View X-Pac Expandable Lumbar Cage System is a lumbar interbody fusion system comprised of a cage implant and instruments. The implants are used to provide structural stability in skeletally mature individuals following discectomy and are placed via the transforaminal approach. The device is available in various sizes to accommodate varying anatomy. The primary difference between the previously cleared and the modified cage models is that the expansion of the upper plate is parallel to the bottom housing, whereas the previously cleared device expands at an angle. The implants are designed for use with autogenous bone graft material. Protrusions on the superior and inferior surfaces of each implant grip the endplates of the adjacent vertebrae to resist expulsion. The implants are manufactured from medical grade titanium alloy per ASTM F136 and ASTM F1295. The Spine View X-Pac Expandable Lumbar Cage System implants are single-use, provided non-sterile, and are intended to be cleaned and steam sterilized before use. The instruments are re-usable, provided nonsterile, and are intended to be cleaned and steam sterilized before use.
The provided text is a 510(k) premarket notification for the SpineView X-Pac Expandable Lumbar Cage System. It details the device's description, indications for use, and a comparison to a predicate device, focusing on mechanical testing performance data to demonstrate substantial equivalence.
However, the document does not contain any information about a study proving the device meets acceptance criteria related to an AI/machine learning model's performance, human reader studies (MRMC), or a standalone algorithm performance. The acceptance criteria discussed in the document are limited to the mechanical performance of the physical device (e.g., static and dynamic compression, subsidence, expulsion) in accordance with FDA guidance and ASTM standards for intervertebral fusion devices.
Therefore, I cannot provide the requested information regarding AI model performance, expert ground truth, adjudication methods, or MRMC studies because this document pertains to a physical medical device and not a software as a medical device (SaMD) or an AI-enabled diagnostic tool.
The information I can extract, pertaining to the physical device's acceptance criteria and performance, is as follows:
1. Table of Acceptance Criteria (Mechanical Performance) and Reported Device Performance:
| Acceptance Criteria (Mechanical Performance) | Performance Metric Mentioned | Reported Device Performance (Implied "Meets" or "Demonstrates SE") |
|---|---|---|
| Static Compression | Mechanical Testing | Conducted to demonstrate substantial equivalence to predicate. |
| Dynamic Compression | Mechanical Testing | Conducted to demonstrate substantial equivalence to predicate. |
| Static Compression-Shear | Mechanical Testing | Conducted to demonstrate substantial equivalence to predicate. |
| Dynamic Compression-Shear | Mechanical Testing | Conducted to demonstrate substantial equivalence to predicate. |
| Subsidence | Mechanical Testing | Conducted to demonstrate substantial equivalence to predicate. |
| Expulsion | Mechanical Testing | Conducted. |
Note: The document states that these tests were conducted "to demonstrate substantial equivalence to the predicate," implying that the performance met the thresholds necessary for this claim. Specific numerical performance values are not provided in this summary.
Regarding the other requested information (2-9), it is not applicable to this document as it does not describe an AI/ML-based device or a clinical study for software performance:
- 2. Sample size used for the test set and the data provenance: Not applicable to this type of device (physical implant). Mechanical testing utilizes test samples of the device itself, not patient data.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is based on engineered specifications and standardized test methods.
- 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for a physical implant, not an AI diagnostic tool.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For mechanical testing, the "ground truth" is defined by the performance requirements of the ASTM standards and FDA guidance for intervertebral fusion devices (e.g., specific load capacities, displacement limits).
- 8. The sample size for the training set: Not applicable. There is no AI model or training set described.
- 9. How the ground truth for the training set was established: Not applicable.
In summary, this 510(k) submission focuses on demonstrating the substantial equivalence of a physical medical device (SpineView X-Pac Expandable Lumbar Cage System) through mechanical performance testing, rather than the performance of an AI/machine learning algorithm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2016
Spine View, Incorporated % Ms. Nancy Lincé Regulatory Affairs Consultant Lincé Consulting, LLC 111 Deerwood Road, Suite 200 San Ramon, California 94583
Re: K160856
Trade/Device Name: SpineView X-Pac Expandable Lumbar Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 1, 2016 Received: July 5, 2016
Dear Ms. Lincé:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K160856
Device Name
SpineView X-Pac Expandable Lumbar Cage System
Indications for Use (Describe)
The SpineView X-Pac Expandable Lumbar Cage System is indicated for use as an intervertebral body fusion device in skeletally mature patients. The system is designed for use with autogenous bone graft material to facilitate spinal fusion. The system is intended for use in patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have at least 6 months of non-operative treatment prior to surgery. In addition, these patients may have up to Grade 1 spondylolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via transforaminal approach using autogenous bone graft material. The system is intended for use with supplemental fixation systems cleared for use by the FDA for use in the thoracolumbar spine.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
A. Name, Address, Phone, and Fax Number of Applicant
Spine View, Inc. 48810 Kato Road, Suite 100E Fremont, CA 94538 Phone: (510) 743-5090 Fax: (510) 490-1753
B. Contact Person
Nancy Lincé Regulatory Affairs Consultant Lincé Consulting, LLC
Date Prepared C.
March 25, 2016
D. Device Name and Classification
| Trade Name: | SpineView X-Pac Expandable Lumbar Cage System |
|---|---|
| Common Name: | Intervertebral Body Fusion Device |
| CFR Classification: | 21 CFR§888.3080 |
| Classification Name: | Intervertebral Body Fusion Device |
| Product Code: | MAX |
E. Predicate Device
The Spine View X-Pac Expandable Lumbar Cage System is substantially equivalent to the previously cleared Spine View X-Pac Expandable Lumbar Cage System, K152539.
F. Device Description
The Spine View X-Pac Expandable Lumbar Cage System is a lumbar interbody fusion system comprised of a cage implant and instruments. The implants are used to provide structural stability in skeletally mature
{4}------------------------------------------------
individuals following discectomy and are placed via the transforaminal approach. The device is available in various sizes to accommodate varying anatomy. The primary difference between the previously cleared and the modified cage models is that the expansion of the upper plate is parallel to the bottom housing, whereas the previously cleared device expands at an angle. The implants are designed for use with autogenous bone graft material. Protrusions on the superior and inferior surfaces of each implant grip the endplates of the adjacent vertebrae to resist expulsion. The implants are manufactured from medical grade titanium alloy per ASTM F136 and ASTM F1295.
The Spine View X-Pac Expandable Lumbar Cage System implants are single-use, provided non-sterile, and are intended to be cleaned and steam sterilized before use. The instruments are re-usable, provided nonsterile, and are intended to be cleaned and steam sterilized before use.
Indications for Use / Intended Use G.
The Spine View X-Pac Expandable Lumbar Cage System is indicated for use as an intervertebral body fusion device in skeletally mature patients. The system is designed for use with autogenous bone graft material to facilitate spinal fusion.
The system is intended for use in patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have at least 6 months of non-operative treatment prior to surgery. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via transforaminal approach using autogenous bone graft material. The system is intended for use with supplemental fixation systems cleared for use by the FDA for use in the thoracolumbar spine.
{5}------------------------------------------------
H. Technological Comparison
The technological characteristics of the Spine View X-Pac Expandable Lumbar Cage System implants are similar to the predicate in terms of design, dimensions, intended use, materials, and performance characteristics.
l. Performance Data
Mechanical testing consisting of static and dynamic compression, static and dynamic compression-shear, and subsidence was conducted in accordance with "Class II Special Controls Guidance Document: Intervertebral Fusion Device", June 12, 2007, ASTM F2077, and ASTM F2267 to demonstrate substantial equivalence to the predicate. Expulsion testing was also conducted.
J. Basis for Substantial Equivalence
The Spine View X-Pac Expandable Lumbar Cage System is similar to the predicate with respect to technical characteristics, performance and intended use. The information provided supports the substantial equivalence of the X-Pac Expandable Lumbar Cage System to the legally marketed predicate.
N/A