K Number
K160763

Validate with FDA (Live)

Date Cleared
2016-06-29

(100 days)

Product Code
Regulation Number
884.5300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ally® Liquid is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Device Description

ally® Liquid Personal Lubricant is a non-sterile, clear, non-staining, non-greasy, high viscosity liquid gel used as a personal lubricant. This product is highly lubricous and may be used with or without a latex or polyisoprene condom during intimate sexual activity. This product is not a contraceptive or spermicide. The product is packaged in 2 to 8 ounce plastic bottles with tamper-proof seals.

INGREDIANTS: Purified water, propanediol, hydroxyethylcellulose, hydroxyacetophenone, lactic acid

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the FDA for a personal lubricant (ally® Liquid Personal Lubricant). It is not a study proving device acceptance. The document focuses on regulatory approval based on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and performance data for a new medical device technology like AI.

Therefore, I cannot provide the requested information as the document does not contain:

  • A table of acceptance criteria and reported device performance in the context of a new device's efficacy study.
  • Details on sample sizes, data provenance, ground truth establishment methods, or expert qualifications for a test set.
  • Information about multi-reader multi-case (MRMC) comparative effectiveness studies or standalone algorithm performance.

The document primarily states that the device was found substantially equivalent to a legally marketed predicate device (Astroglide® Natural Liquid Personal Lubricant, K141581) based on meeting recognized consensus standards for biocompatibility, microbiology, condom compatibility, and shelf-life testing. These are regulatory compliance aspects, not detailed performance studies in the way you've outlined for AI or other complex medical devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2016

Guy & O'Neill, Inc. % John Ziobro Principal Consultant SpectraMedEx, LLC 3215 Golf Road, #149 Delafield, WI 53018

Re: K160763

Trade/Device Name: ally Liquid Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: Mav 27, 2016 Received: May 31, 2016

Dear John Ziobro,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160763

Device Name ally® Liquid Personal Lubricant

Indications for Use (Describe)

ally® Liquid is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number,"

{3}------------------------------------------------

Traditional 510(k) Summary

1. Summary Date:June 27, 2016
2. Applicant Name& Address:Guy & O'Neill, Inc.200 Industrial DriveFredonia, WI 53021
3. SubmissionCorrespondent:John F. ZiobroPrincipal ConsultantSpectraMedEx, LLC3215 Golf Road, #149Delafield, WI 53018Phone: 262-719-8922
4. Trade Name:ally® Liquid Personal Lubricant
5. Common Name:Personal Lubricant
6. Regulation:21 CFR 884.5300
7. Class:II
8. Classification Name:Condom
9. Advisory Panel:Obstetrical and Gynecological Therapeutic Devices
10. Predicate Device(s):Astroglide® Natural Liquid Personal Lubricant (K141581)
  1. Indication for Use:

ally® Liquid is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

12. Device Description:

ally® Liquid Personal Lubricant is a non-sterile, clear, non-staining, non-greasy, high viscosity liquid gel used as a personal lubricant. This product is highly lubricous and may be used with or without a latex or polyisoprene condom during intimate sexual activity. This product is not a contraceptive or spermicide. The product is packaged in 2 to 8 ounce plastic bottles with tamper-proof seals.

INGREDIANTS: Purified water, propanediol, hydroxyethylcellulose, hydroxyacetophenone, lactic acid

{4}------------------------------------------------

    1. Technological Characteristics:
      This product is a water-based personal lubricant, and its ingredients and physical parameters are comparable to the predicate device.
    1. Summary of Performance Data:
      Biocompatibility, microbiology, condom compatibility and shipping testing were conducted on the device by outside laboratories, in compliance with the following recognized consensus standards:
  • ASTM D7661-10, Standard Test Method for Determining Compatibility of Personal . Lubricants with Natural Rubber Latex Condoms for compliance with latex and polyisoprene condoms.

  • ISO 10993-5 Biological Evaluation of Medical Devices-Part 5: Tests for in vitro cytotoxicity .

  • ISO 10993-10 Biological Evaluation of Medical Devices-Part 10: Tests for irritation and . skin sensitization

  • ISO 10993-11 Biological Evaluation of Medical Devices-Part 11: Tests for systemic toxicity ●

  • United States Pharmacopeia, National Formulary General Chapter <88>: Biological ● Reactivity Tests, In Vivo

  • USP <51> Antimicrobial Effectiveness Testing ●

  • USP <61> Microbiological Examination of Nonsterile Products: Microbial Enumeration . Tests

  • USP <62> Microbiological Examination of Nonsterile Products: Tests for Specified . Microorganisms

  • ISTA 3A: Packaged-Products for Parcel Delivery System Shipment 70 Kg (150 lb) or Less .

Shelf-life testing:

Both accelerated and real time stability studies were performed on the ally® Liquid per ICH Q1A (R2) Guidelines. The device will be launched with a 10-month shelf life based on accelerated aging and then re-labeled with 2- and 3-year shelf-life once the results of the realtime testing are completed.

15. Conclusions

The proposed ally "Liquid Personal Lubricant and the predicate (K141581) have the same intended use and comparable technological characteristics. The differences in specific ingredients do not raise any different types of questions. Performance data demonstrated that the subject device is substantially equivalent to the predicate device in terms of safety and effectiveness.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.