K Number
K160710

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-05-13

(59 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Oasis Medical Lacrimal Intubation Set is intended for use for nasolacrimal intubation in patients of age and older. Indications for nasolacrimal intubation performed with the Oasis Medical Lacrimal Intubation device are:

  • · Canalicular pathologies (e.g., congenital or acquired stenosis, lacerations)
  • · During dacryocystorhinostomy
  • · Congenital lacrimal duct obstruction
Device Description

The Oasis Lacrimal Intubation Set with Retrieval Device (Model Summary: 6004) is a sterile (via gamma sterilization-VDmax method-25kGy), single-use, hand held, and ophthalmic surgical device. The device consists of a twelve inch medical grade silicone tube that is secured at each end to two 4 ½", 23 gauge stainless steel, olivetipped probes along with a 3″ plastic handled retriever with a stainless steel hook. The Oasis Lacrimal Intubation Set with Retrieval Device is used by a physician for nasolacrimal intubation.

AI/ML Overview

This document describes the Oasis Lacrimal Intubation Set with Retrieval Device and its substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria or device performance metrics in a pass/fail format. Instead, it focuses on demonstrating equivalence to a predicate device across several characteristics. The "performance" is implicitly deemed acceptable if it's equivalent to the legally marketed predicate.

However, we can infer the areas of assessment and the reported findings for each:

Acceptance Criteria AreaReported Device Performance (Comparison to Predicate Device)
Indications for UseEquivalent: Intended for nasolacrimal intubation in patients 12 months+ for canalicular pathologies (congenital/acquired stenosis, lacerations), during dacryocystorhinostomy, and congenital lacrimal duct obstruction.
Technological CharacteristicsDesign: Oasis Lacrimal Intubation Design (probes and silicone) is equivalent. Retrieval Device Design differs (single-use plastic handled vs. reusable stainless steel handled predicate, slightly angled/ergonomic hook on Oasis).
Mode of Action: Equivalent (intubation of lacrimal ducts enabling drainage of tears by capillarity).
Instructions for Use: Equivalent (insertion and removal).
Materials: Equivalent (Stainless steel and silicone).
SterilityRetrieval Device Sterility differs: Oasis device is provided sterile; predicate hook is not provided sterile. Mode of Sterilization for Oasis Lacrimal Intubation differs: Oasis is gamma sterilized (VDmax method-25kGy); predicate is EtO sterilized.
BiocompatibilityTested for Cytotoxicity, Sensitization, and Irritation. Concluded no adverse reaction. Silicone tubing (tissue contacting portion) considered to exceed USP Class VI due to manufacturing process not introducing novel impurities.
Durability/StrengthPerformed equivalently to predicate device in bench tests assessing dimensional attributes and device durability via strength testing (known pull force on junctions). Concluded similar safety, effectiveness, and performance due to equivalent overall design and strength.
Overall Safety & EffectivenessConcluded to be as safe and effective as the predicate device, based on independent laboratory and internal comparative bench testing, despite stated differences. Differences "do not affect the Oasis Medical Lacrimal Intubation Set with Retrieval Device in terms of relevant functionality and safety."

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical bench testing. It does not specify a numerical "sample size" in terms of patients or independent test articles for each specific test performed. However, it indicates:

  • Bench tests: A "series of bench tests" was conducted.
  • Biocompatibility tests: These were conducted for Cytotoxicity, Sensitization, and Irritation.
  • Provenance: The tests were conducted by "independent laboratories and internal comparative bench testing." No country of origin for the data is specified, nor is whether the testing was retrospective or prospective (it would inherently be prospective for product development).

3. Number of Experts Used to Establish Ground Truth and Qualifications

This document describes a medical device (intubation set), not a diagnostic AI device requiring expert ground truth for interpretation. Therefore, these categories are not applicable. The "truth" here is established by engineering specifications, material properties, and functionality comparison against a predicate device.

4. Adjudication Method for the Test Set

Not applicable, as this is bench testing of a physical device, not a diagnostic assessment requiring expert adjudication of findings.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No. This is a 510(k) submission for a physical medical device, not a diagnostic AI system requiring MRMC studies.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

No. This question applies to AI/software as a medical device. This is a physical device, so this is not applicable.

7. The Type of Ground Truth Used

For the non-clinical tests described:

  • Engineering specifications and standards: For dimensional attributes, material properties (e.g., USP Class VI for silicone).
  • Functional comparison to a predicate device: For durability and overall performance (e.g., "performed equivalently to the predicate device").
  • Standardized biocompatibility test protocols: For cytotoxicity, sensitization, and irritation.

8. The Sample Size for the Training Set

No. This question applies to AI/software as a medical device. This is a physical device, so this is not applicable.

9. How the Ground Truth for the Training Set was Established

No. This question applies to AI/software as a medical device. This is a physical device, so this is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized representation of three human profiles facing right. The profiles are stacked on top of each other, creating a sense of depth and unity. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 13, 2016

Anodyne Surgical Ms. Valerie Anderson Director Quality Assurance and Regulatory Affairs 804 Corporate Centre Drive O' Fallon, Missouri 63368

Re: K160710

Trade/Device Name: Oasis Lacrimal Intubation Set with Retrieval Device Regulation Name: Lacrimal Stents and Intubation Sets Regulatory Class: Unclassified Product Code: OKS Dated: March 14, 2016 Received: March 15, 2016

Dear Ms. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, and Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) Not known

Device Name

Oasis Lacrimal Intubation Set with Retrieval Device

Indications for Use (Describe)

The Oasis Medical Lacrimal Intubation Set is intended for use for nasolacrimal intubation in patients of age and older. Indications for nasolacrimal intubation performed with the Oasis Medical Lacrimal Intubation device are:

  • · Canalicular pathologies (e.g., congenital or acquired stenosis, lacerations)
  • · During dacryocystorhinostomy
  • · Congenital lacrimal duct obstruction

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Anodyne Surgical. The logo includes the text "PSI / EYE-KO, Inc. dba" in red font at the top right. Below this text is the word "ANODYNE" in large blue font, with the word "SURGICAL" in smaller font below it. There are also red curved lines at the bottom left of the logo.

Oasis Lacrimal Intubation Set with Retrieval Device

510(K) Summary

Submitted in accordance with the requirements of 21 CFR 807.92

Applicant's Name andAddress:Anodyne804 Corporate Centre DriveO'Fallon, MO 63368
Contact Person:Valerie AndersonDirector Quality and Regulatory Affairs(T) 636-447-1010(F) 636-447-1786
Date Prepared:May 10, 2016
Device Trade Name:Oasis Lacrimal Intubation Set with Retrieval Device
Common Name:Lacrimal Intubation Set with Retrieval Device
Regulation Number:Unclassified
Regulation Name:Unclassified
Regulatory Class:Unclassified
Product Code:OKS
FDA Panel:Ophthalmic
Predicate Device:Sold separately - FCI, Crawford Bicanaliculus Intubation(Unclassified - K121142 - Product Code OKS) and Crawford Hook(Class 1 - 510(K) Exempt - Product Code HNQ - Reg. Number886.4350)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logos for PSI / EYE-KO, Inc. dba and Anodyne Surgical. The PSI / EYE-KO, Inc. dba logo features a red eye with the letters "P.S.I." and a map of the United States inside. The Anodyne Surgical logo features the word "ANODYNE" in blue, with the word "SURGICAL" in smaller letters below it. There are also three red curved lines below the word "ANODYNE".

510(K) Submission Oasis Lacrimal Intubation Set with Retrieval Device 510(K) Summary Submitted in accordance with the requirements of 21 CFR 807.92

Device Description The Oasis Lacrimal Intubation Set with Retrieval Device (Model Summary: 6004) is a sterile (via gamma sterilization-VDmax method-25kGy), single-use, hand held, and ophthalmic surgical device. The device consists of a twelve inch medical grade silicone tube that is secured at each end to two 4 ½", 23 gauge stainless steel, olivetipped probes along with a 3″ plastic handled retriever with a stainless steel hook. The Oasis Lacrimal Intubation Set with Retrieval Device is used by a physician for nasolacrimal intubation. The insertion and removal for the Oasis Lacrimal Intubation Set with Retrieval Device is equivalent to the insertion and removal of the predicate device as evident in the IFU for each device. For insertion, each punctum should be dilated and the lacrimal system should be probed to open any blockages by using a standard lacrimal probe. The first probe of the Lacrimal Intubation Set is passed through the upper punctum and across the upper canaliculus and then oriented down through the lacrimal system into the nose approximately 4cm. Proper positioning of the probe can be confirmed by achieving metal-to-metal contact with the retrieval instrument inserted through the nostril. The probe end should be located lateral to the inferior turbinate in the inferior meatus of the nose. The retrieval instrument can be used to engage the probe at its olive tip and pull the probe out of the nose. The second probe is passed down the inferior canaliculus and out

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains two logos. The first logo is for "PSI / EYE-KO, Inc. dba" and features a red eye-shaped graphic with the letters "P.S.I." and a map of the United States inside. The second logo is for "ANODYNE SURGICAL" and features the word "ANODYNE" in large blue letters above the word "SURGICAL" in smaller red letters. There are also red curved lines below the words "ANODYNE SURGICAL".

510(K) Submission Oasis Lacrimal Intubation Set with Retrieval Device 510(K) Summary Submitted in accordance with the requirements of 21 CFR 807.92

of the nose in a similar fashion.

The two exposed probes are cut from the silicone tube and removed. The ends of the silicone tube are tied to each other and securely knotted. The silicone tubes may be further secured by a tiny suture. Tuck the knotted silicone tubes up into the nose.

For removal, the silicone tube is located in the eye between the upper and lower puncta and pulling the tube upward. The silicone tube is then cut and pulled out completely from the ducts.

The knot used to tie the silicone tubes together is too big to pass through the canaliculus and the silicone tubes must be removed through the nose. In older patients it may be possible to expel the knot by having the patient blow their nose. In small children, location of the knot and silicone tube removal may need to be performed under general anesthesia.

Duration of intubation (patient contact of the silicone tube portion) is 24 hours to 30 days.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains two logos. The first logo is for "PSI / EYE-KO, Inc. dba" and features a stylized eye with the letters "P.S.I." inside, along with an outline of the United States. The second logo is for "ANODYNE SURGICAL" and is written in a bold, sans-serif font, with the word "ANODYNE" in a larger font size than "SURGICAL". There are also red curved lines underneath the word "ANODYNE".

Oasis Lacrimal Intubation Set with Retrieval Device

510(K) Summary

Submitted in accordance with the requirements of 21 CFR 807.92

Indications for Use: The Oasis Medical Lacrimal Intubation Set with Retrieval Device is intended for use for nasolacrimal intubation in patients 12 months of age and older. Indications for nasolacrimal intubation performed with the Oasis Medical Lacrimal Intubation device are:

  • Canalicular pathologies (e.g., congenital or acquired stenosis, lacerations)
  • During dacryocystorhinostomy
  • Congenital lacrimal duct obstruction

The Indications for Use statement for the Oasis Medical Lacrimal Intubation Set with Retrieval Device is equivalent to the predicate device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Anodyne Surgical. The logo includes the text "PSI / EYE-KO, Inc. dba" in red font at the top right. Below that is the word "ANODYNE" in large blue font, with the word "SURGICAL" in smaller red font below it. There are also red curved lines at the bottom left of the logo.

Oasis Lacrimal Intubation Set with Retrieval Device

510(K) Summary

Submitted in accordance with the requirements of 21 CFR 807.92

Summary ofTechnologicalcharacteristics:The Oasis Lacrimal Intubation with Retrieval Device is equivalent tothe predicate device in terms of:
Oasis Lacrimal Intubation Design – probes and silicone
Mode of Action – intubation of the lacrimal ducts enablingdrainage of tears by capillarity
Instructions for Use – insertion and removal
Oasis Lacrimal Intubation Materials - Stainless steel andsilicone
Differences include:
Retrieval Device Design - The predicate hook is reusablewhereas the Oasis Medical Lacrimal Intubation Set Retrieval

Device is single-use and disposable (plastic handle). The reusable predicate hook is more substantial (stainless steel handle) for multiple uses. Slightly angled, ergonomic hook on the Oasis Medical Lacrimal Intubation Set Retrieval Device.

Retrieval Device Sterility - The predicate device hook is not provided sterile whereas the Oasis Medical Lacrimal Intubation Set with Retrieval Device is provided sterile.

Mode of Sterilization for Oasis Lacrimal Intubation - The predicate intubation device is EtO sterilized whereas the Oasis Medical Lacrimal Intubation Set with Retrieval Device is gamma sterilized via VDmax method (25kGy).

The device differences do not affect the Oasis Medical Lacrimal Intubation Set with Retrieval Device in terms of relevant functionality and safety.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logos for PSI / EYE-KO, Inc. dba and Anodyne Surgical. The PSI logo features an eye with the letters P.S.I. and an outline of the United States inside. The Anodyne Surgical logo is in blue and red, with the word "ANODYNE" in large, bold letters and the word "SURGICAL" underneath in smaller letters.

Oasis Lacrimal Intubation Set with Retrieval Device

510(K) Summary

Submitted in accordance with the requirements of 21 CFR 807.92

Biocompatibility:The biocompatibility evaluation for the Oasis Lacrimal Intubationwith Retrieval Device included the following tests:CytotoxicitySensitizationIrritationTest results concluded this device did not elicit an adverse reactionduring the above testing. The silicone tubing is considered thetissue contacting portion. The duration of contact is 24 hours to30 days. In consideration of the Oasis Lacrimal Intubation withRetrieval Device manufacturing process, no novel impurities areintroduced or added to the silicone and therefore the siliconematerial biocompatibility profile exceeds USP Class VI.
Summary of Non-clinicaltests:The subject device performed equivalently to the predicate devicein a series of bench tests which included an assessment ofdimensional attributes and device durability via strength testing bymeans of applying a known pull force on the device junctions. Itwas concluded in this side-by-side comparison, the subject deviceand the predicate device have similar safety, effectiveness, andperformance profiles because their overall design and strength areequivalent.
Substantial EquivalenceBasis:The conclusions performed by independent laboratories andinternal comparative bench testing provide objective evidence tosubstantiate the Oasis Lacrimal Intubation Set with RetrievalDevice is as safe and effective as the predicate device that iscurrently marketed for the same intended use.

N/A