(106 days)
The Gibralt® Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Gibralt® Spine System may be connected to the Gibralt® Occipital Spine System with rod-to-rod connectors. The Gibralt® Spine System may also be connected to the Exactech® Proliant® System, Exactech® Silverbolt® and Mainframe® Spinal Screw Systems, or Exactech® Hydralok® Exactech, using rod-to-rod connectors and transitional rods. Refer to the specific system package inserts for a list of their indications for use.
When used with the Gibralt® Spine System, the Gibralt® Occipital Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7), and the thoracic spine from T1 - T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Occipital Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Gibralt® Occipital Spine System may be connected to the Gibralt® Spine System, using rod-to-rod connectors. Refer to the Gibralt® Spine System package insert for a list of the Gibralt® Spine System indications of use.
The Occipital Bone Screws are limited to occipital fixation only.
The line extension and device modifications for the Gibralt® Spine System are intended to provide immobilization and stabilization of spinal segments in the upper thoracic, cervical, and occipital spine. The system components are manufactured from titanium and cobalt chromium. The modifications included in the scope of this submission are: expanded indications for use for both systems, incorporation of design changes, and addition of new components.
This document is a 510(k) Pre-Market Notification from the FDA regarding the "Gibralt® Spine System and Gibralt® Occipital Spine System." It is primarily a regulatory approval document and does not contain the detailed study information typically found in a clinical trial report or a comprehensive technical performance study.
Therefore, the specific information requested about acceptance criteria, detailed study design, sample sizes, expert qualifications, ground truth establishment, MRMC studies, or standalone performance of an AI/algorithm-based device cannot be fully provided from this document alone. This document describes a medical device (spinal fixation system) which is a physical implant, not an AI or algorithm-based diagnostic/prognostic device.
However, based on the provided text, I can extract information related to the device's performance data which serves as the "study that proves the device meets the acceptance criteria" in the context of this 510(k) submission, and the "acceptance criteria" can be inferred from the nature of the tests and the conclusion of substantial equivalence.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Inferred from testing standards and regulatory conclusion) | Reported Device Performance (from "PERFORMANCE DATA" section) |
|---|---|
| Device strength sufficient for intended use, substantially equivalent to predicate devices. | Non-clinical testing showed: |
| Consistency with ASTM F1717-14 standards for static and dynamic axial compression and compression bending. | - Static axial compression per ASTM F1717-14 |
| - Static axial compression bending per ASTM F1717-14 | |
| - Dynamic axial compression per ASTM F1717-14 | |
| - Dynamic axial compression bending per ASTM F1717-14 | |
| Conclusion: "The results of this non-clinical testing show that the strength of the Gibralt® Spine System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices." |
2. Sample size used for the test set and the data provenance
- Sample size: Not specified. The document mentions "the Gibralt® Spine System" and "the proposed line extensions and modified Gibralt® Spine System" were tested, implying multiple units or configurations were tested, but an exact number of samples (e.g., number of constructs tested) is not provided.
- Data provenance: Not explicitly stated, but the testing would have been conducted by or for Exactech, Inc. or Empirical Testing Corp. in a laboratory setting. Since it's device testing, it is prospective in the sense that the tests were specifically performed for this submission according to established ASTM standards. It is not clinical data from patients/countries.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This document pertains to mechanical testing of a spinal implant, not the assessment of medical images or patient data that would require expert consensus for ground truth. The "ground truth" here is the physical performance of the device under mechanical load, measured by objective instruments according to specified ASTM standards.
4. Adjudication method for the test set
- Not applicable. See point 3. Mechanical tests follow predefined protocols and objective measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a physical spinal implant system, not a diagnostic or prognostic AI/algorithm requiring human reader evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For this type of device, the "ground truth" for the mechanical performance testing is the objective measurement of physical properties and strength when subjected to various loads and cycles, as defined by the ASTM F1717-14 standard. There is no expert consensus, pathology, or outcomes data involved in generating the mechanical performance data in this context.
8. The sample size for the training set
- Not applicable. This document describes the mechanical testing of a physical medical device. There is no "training set" in the context of an AI/algorithm. The device design might be informed by previous designs and engineering principles, but this is not analogous to an AI training set.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 28, 2016
Exactech, Incorporated % Mr. Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K160697
Trade/Device Name: Gibralt® Spine System and Gibralt® Occipital Spine System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: May 31, 2016 Received: June 1, 2016
Dear Mr. Kenneth C. Maxwell II:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 | ||
|---|---|---|---|
| Food and Drug Administration | Expiration Date: January 31, 2017 | ||
| Indications for Use | See PRA Statement on last page. | ||
| 510(k) Number (if known) | K160697 | ||
| Device Name | Gibralt® Spine System | ||
| Indications for Use (Describe) | The Gibralt® Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The GibraltⓇ Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. | ||
| In order to achieve additional levels of fixation, the Gibralt® Spine System may be connected to the Gibralt® Occipital Spine System with rod-to-rod connectors. The Gibralt® Spine System may also be connected to the Exactech® Proliant® System, Exactech® Silverbolt® and Mainframe® Spinal Screw Systems, or Exactech® Hydralok® Exactech, using rod-to-rod connectors and transitional rods. Refer to the specific system package inserts for a list of their indications for use. | |||
| Type of Use (Select one or both, as applicable) | [X] Prescription Use (Part 21 CFR 801 Subpart D)[ ] Over-The-Counter Use (21 CFR 801 Subpart C) | [X] Prescription Use (Part 21 CFR 801 Subpart D) | [ ] Over-The-Counter Use (21 CFR 801 Subpart C) |
| [X] Prescription Use (Part 21 CFR 801 Subpart D) | [ ] Over-The-Counter Use (21 CFR 801 Subpart C) | ||
| PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. | |||
| FOR FDA USE ONLY | |||
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) |
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug Administration | Form Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement on last page. |
|---|---|
| ------------------------------------------------------------------------- | ---------------------------------------------------------------------------------------------------------- |
Indications for Use
| 510(k) Number (if known) | K160697 |
|---|---|
| Device Name | Gibralt® Occipital Spine System |
| Indications for Use (Describe) |
When used with the Gibralt® Spine System, the Gibralt® Occipital Spine System is intended
to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the
following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-
C7), and the thoracic spine from T1 - T3: traumatic spinal fractures and/or traumatic
dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors
involving the cervical/thoracic spine; and degenerative disease, including intractable
radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by
radiographic studies, and degenerative disease of the facets with instability. The GibraltⓇ
Occipital Spine System is also intended to restore the integrity of the spinal column even in the
absence of fusion for a limited time period in patients with advanced stage tumors involving
the cervical spine in whom life expectancy is of insufficient duration to permit achievement of
fusion.
In order to achieve additional levels of fixation, the Gibralt® Occipital Spine System may be
connected to the Gibralt® Spine System, using rod-to-rod connectors. Refer to the Gibralt®
Spine System package insert for a list of the Gibralt® Spine System indications of use.
The Occipital Bone Screws are limited to occipital fixation only.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
| FORM FDA 3881 (9/13)Page 1 of 2 | PSC Publishing Services (301) 443-6740 EF |
|---|---|
| ------------------------------------- | ------------------------------------------- |
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
5.510(K) SUMMARY
| Submitter's Name: | Exactech, Inc. |
|---|---|
| Submitter's Address: | 2320 NW 66th CourtGainesville, FL 32658 |
| Submitter's Telephone: | 352.377.1140 |
| Contact Person: | Kenneth C. Maxwell IIEmpirical Testing Corp.719.291.6874 |
| Date Summary was Prepared: | 23 June 2016 |
| Trade or Proprietary Name: | Gibralt® Spine SystemGibralt® Occipital Spine System |
| Common or Usual Name: | Orthosis, Cervical Spinal Pedicle FixationAppliance, Fixation, Spinal Interlaminal |
| Classification: | Unclassified, Class II per 21 CFR §888.3050 |
| Product Code: | NKG, KWP |
| Classification Panel: | Orthopedic Devices |
DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The line extension and device modifications for the Gibralt® Spine System are intended to provide immobilization and stabilization of spinal segments in the upper thoracic, cervical, and occipital spine. The system components are manufactured from titanium and cobalt chromium. The modifications included in the scope of this submission are: expanded indications for use for both systems, incorporation of design changes, and addition of new components.
INDICATIONS FOR USE
Gibralt® Spine System:
The Gibralt® Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Gibralt® Spine System may be connected to the Gibralt® Occipital Spine System with rod-to-rod connectors. The Gibralt®
{7}------------------------------------------------
Spine System may also be connected to the Exactech® Proliant® System. Exactech® Silverbolt® and Mainframe® Spinal Screw Systems, or Exactech® Hydralok® Exactech, using rod-to-rod connectors and transitional rods. Refer to the specific system package inserts for a list of their indications for use.
Gibralt® Occipital Spine System:
When used with the Gibralt® Spine System, the Gibralt® Occipital Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7), and the thoracic spine from T1 - T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Occipital Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Gibralt® Occipital Spine System may be connected to the Gibralt® Spine System, using rod-to-rod connectors. Refer to the Gibralt® Spine System package insert for a list of the Gibralt® Spine System indications of use.
The Occipital Bone Screws are limited to occipital fixation only.
The indications for use for the Gibralt® Spine System and the Gibralt® Occipital Spine System are similar to that of the predicate devices in Table 5-1.
TECHNOLOGICAL CHARACTERISTICS
Gibralt® Spine System components are made from materials that conform to titanium per ASTM F136 and cobalt chromium per ASTM F1537. The Gibralt® Occipital Spine System components are made from materials that conform to titanium per ASTM F136. The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates:
- Indications for Use
- Materials of Manufacture ●
{8}------------------------------------------------
| 510k Number | Trade or Proprietary orModel Name | Manufacturer | Type |
|---|---|---|---|
| K142838 | Synapse Occipital-Cervical-Thoracic System | Synthes USA | PrimaryPredicate |
| K083071 | VERTEXTM ReconstructionSystem | Medtronic | Reference |
| K050979 | S4 Spinal System | Aesculap®, Inc. | Reference |
Table 5-1 Predicate Devices
PERFORMANCE DATA
The proposed line extensions and modified Gibralt® Spine System have been tested in the following test modes:
- Static axial compression per ASTM F1717-14
- Static axial compression bending per ASTM F1717-14 ●
- Dynamic axial compression per ASTM F1717-14 ●
- o Dynamic axial compression bending per ASTM F1717-14
The results of this non-clinical testing show that the strength of the Gibralt® Spine System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the Gibralt® Spine System and Gibralt® Occipital Spine System are substantially equivalent to the predicate devices.
N/A