K Number
K160685

Validate with FDA (Live)

Date Cleared
2016-08-04

(146 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surveyor S4 Mobile Monitor is indicated for use:
• The Surveyor S4 Mobile Monitor is indicated for use to acquire, analyze and output multi-lead ECG and SpO2 signals.
The Mortara Surveyor S4 mobile monitor facilitates the monitoring of ECG and SpO2 parameters when used in conjunction with alarm annunciating devices such as the Surveyor Central Station for telemetric monitoring; however, the Surveyor S4 has no alarming capabilities of its own.
• The Surveyor S4 processes Heart Rate, ST segment deviation and arrhythmias such as, but not limited to, Ventricular Tachycardia, Ventricular Fibrillation, and Asystole by utilizing Mortara VERITAS™ ECG algorithm.
• The Surveyor S4 can deliver data from a 3/5-lead cable and diagnostic 12-lead ECG data to a receiving Surveyor Central station
• The Surveyor S4 Mobile Monitor is indicated for use as a mobile radio transceiver, allowing the patient to be ambulatory within the range of a Wi-Fi information infrastructure.
• The Surveyor S4 Mobile Monitor is indicated for use in adults, adolescents and children.
• The Surveyor S4 Mobile Monitor is a prescription device intended to be used by knowledgeable healthcare professionals within a healthcare or clinical setting. It is not intended as a sole means of diagnosis.

Device Description

The Surveyor™ S4 Mobile Monitor is a battery operated patient worn transceiver that communicates via IEEE 802.11 Wi-Fi technology to the Surveyor™ Central Station (K131929) which is the primary monitoring display and alarm source.
The Surveyor S4 transmitter has a touch sensitive, color display allowing for operational conveniences such as a battery level indicator, Wi-Fi Signal Strength and Surveyor Central Station slot information. The Surveyor S4 can display physiological waveforms, measured parameters, demographic data and other status information. The Surveyor S4 acquires, filters, and processes ECG or ECG and SpO2 parameters and sends filtered signals and parameters to the Surveyor Central Station where alarm processing and annunciation occurs.
The Surveyor Central is designed to communicate with the Surveyor S4 within a healthcare facilities Wi-Fi network infrastructure. The Surveyor Central provides primary monitoring of telemetry patients when used with the Surveyor S4. The central station displays numeric values, waveforms, alarm generation and management, data storage, individualized patient management, report generation and printing functionality.
The Surveyor S4 provides continuous data acquisition which can be visualized on the S4 itself for the following parameters:

  • o Electrocardiography (ECG) including heart rate and ECG waveforms
  • · Saturation percentage of peripheral oxygen (SpO2), including SpO2 plethysmogram waveform, pulse rate, and signal quality index.
  • o No alarm status or processing information of ECG or SpO2 is shown on the Surveyor S4 device.
  • o The Surveyor S4 processes Heart rate, ST segment deviation and arrhythmias, by utilizing Mortara VERITAS™ ECG algorithm and sends ECG data packets, including arrhythmias such as, but not limited to, Ventricular Tachycardia, Ventricular Fibrillation, and Asystole to the Surveyor Central Station for alarm processing and annunciation.
  • o The Surveyor S4 utilizes proprietary ECG lead cables available in AHA or IEC, snap or clip, 3-Lead, 5-Lead, and diagnostic 12 Lead configurations.
  • o The Surveyor S4 processes SpO2 percentage, pulse rate and signal quality through the Mortara SpO2 algorithm and sends data packets for SpO2 to the Surveyor Central Station for alarm processing and annunciation.
  • o Multiple SpO2 sensor styles allow connection of adults, adolescents, and children. SpO2 sensors are specified for use with the Surveyor S4 and available through Mortara Instrument.
AI/ML Overview

The device in question is the Mortara Surveyor S4 Mobile Monitor. The provided text primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study for acceptance criteria. However, it does mention one basic clinical study for SpO2 compliance.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly present a table of acceptance criteria with corresponding device performance for the entire device. It states that software verification and validation testing was conducted, and the device was designed and tested for compliance with specific standards (listed below). The only specific performance claim related to a study is for SpO2.

Acceptance Criteria (Standard/Guideline)Reported Device Performance
IEC 60601-1:2005 Ed:3"The Surveyor S4 Mobile Monitor was designed and tested for compliance with the applicable clauses of the following standards..." (Implies compliance, but no specific performance metrics are given for this general safety standard).
IEC 60601-2-49: 2011"The Surveyor S4 Mobile Monitor was designed and tested for compliance with the applicable clauses of the following standards..." (Implies compliance, but no specific performance metrics are given for this multi-function patient monitoring standard).
IEC 60601-1-2:2007"The Surveyor S4 Mobile Monitor was designed and tested for compliance with the applicable clauses of the following standards..." (Implies compliance, but no specific performance metrics are given for this EMC standard).
ISO 80601-2-61:2011 Clause 201.12.1"A basic pulse oximeter clinical study was conducted to demonstrate the Surveyor S4 Mobile Monitor SpO2 compliance to IEC 80601-2-61. The study met the requirements of ISO 80601-2-61-2011 Clause 201.12.1." (This indicates the study successfully demonstrated compliance, but specific performance values like accuracy, bias, or RMSE are not provided in this summary). The standard itself would define the acceptable limits for these.
IEC 60601-2-25:2011-10 Edition 2.0"The Surveyor S4 Mobile Monitor was designed and tested for compliance with the applicable clauses of the following standards..." (Implies compliance, but no specific performance metrics are given for this electrocardiographs standard). Relevant to ECG accuracy, heart rate calculation, etc.
ANSI/AAMI EC57:2012"The Surveyor S4 Mobile Monitor was designed and tested for compliance with the applicable clauses of the following standards..." (Implies compliance, but no specific performance metrics are given for cardiac rhythm, ST segment measurement, and ventricular fibrillation algorithms. This standard specifies performance requirements and test methods for these algorithms, so compliance would mean meeting its defined accuracy, sensitivity, and specificity limits, among others).
Mortara VERITAS™ ECG algorithmThe device "processes Heart rate, ST segment deviation and arrhythmias such as, but not limited to, Ventricular Tachycardia, Ventricular Fibrillation, and Asystole by utilizing Mortara VERITAS™ ECG algorithm." No specific performance metrics (sensitivity, specificity, accuracy) are explicitly stated for this algorithm in the provided text, but compliance with ANSI/AAMI EC57 suggests it meets the performance requirements of that standard.
Mortara SpO2 algorithmThe device "processes SpO2 percentage, pulse rate and signal quality through the Mortara SpO2 algorithm." Compliance to ISO 80601-2-61 implies meeting the performance requirements for SpO2 and pulse rate accuracy.

Note: The FDA 510(k) summary typically establishes substantial equivalence, often by demonstrating adherence to recognized standards and comparing to predicate devices, rather than presenting full clinical trial results and specific acceptance criterion tables for every feature. The statement "The study met the requirements of ISO 80601-2-61-2011 Clause 201.12.1" is the most direct mention of acceptance criteria being met.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "A basic pulse oximeter clinical study was conducted to demonstrate the Surveyor S4 Mobile Monitor SpO2 compliance to IEC 80601-2-61."

  • Sample Size for Test Set: Not specified for the SpO2 clinical study.
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided for the SpO2 clinical study or any other part of the evaluation. For a pulse oximeter study conforming to ISO 80601-2-61, ground truth is typically established by comparing the device's SpO2 readings against co-oximeter readings from arterial blood samples, with interpretation often overseen by a qualified physician or clinical team, but details are absent here.

4. Adjudication Method for the Test Set

This information is not provided.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The document describes a "basic pulse oximeter clinical study" for SpO2 compliance and otherwise focuses on engineering testing against standards and substantial equivalence to predicate devices, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

For the SpO2 functionality, a standalone clinical study was done as "A basic pulse oximeter clinical study was conducted to demonstrate the Surveyor S4 Mobile Monitor SpO2 compliance to IEC 80601-2-61." This evaluates the algorithm's performance directly against established ground truth.

For ECG aspects (Heart rate, ST segment deviation, arrhythmias using VERITAS™ ECG algorithm), the document states the device "was designed and tested for compliance with the applicable clauses of... ANSI/AAMI EC57:2012 Testing and reporting performance results of cardiac rhythm, ST segment measurement algorithms, ventricular fibrillation algorithm, and irregular rhythm (atrial fibrillation) algorithm." This standard specifically outlines methods for standalone algorithm performance testing, implying such testing was conducted to demonstrate compliance. However, specific performance metrics or detailed study descriptions are not provided in this summary.

7. The Type of Ground Truth Used

For the SpO2 clinical study, the ground truth would typically be established using co-oximetry from arterial blood samples, as dictated by the ISO 80601-2-61 standard for pulse oximeter accuracy testing. This involves comparing the device's SpO2 readings with direct measurements of oxygen saturation in arterial blood. While not explicitly stated, adherence to the standard implies this methodology.

For the ECG algorithms (Heart Rate, ST-segment, arrhythmia detection), ground truth for compliance with ANSI/AAMI EC57:2012 would typically be established by expert consensus of cardiologists or other qualified medical professionals reviewing annotated ECG recordings. This could also involve comparison against known, validated datasets or simulated waveforms. Again, the specific details are not provided in this summary but are implied by the standard.

8. The Sample Size for the Training Set

The document does not provide information regarding the training set sample size for any of its algorithms (VERITAS™ ECG algorithm or Mortara SpO2 algorithm). This submission focuses on the performance of the final, developed device and algorithm.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set was established. As this is a 510(k) submission for substantial equivalence, such details about algorithm development and training data are typically not included unless specifically requested or if it's a novel AI device requiring detailed algorithm transparency.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 4, 2016

Mortara Instrument, Inc. Sarah Weber Senior Regulatory Affairs Manager 7865 North 86th Street Milwaukee. Wisconsin 53224

Re: K160685

Trade/Device Name: Surveyor S4 Mobile Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DQA, DPS Dated: June 28, 2016 Received: June 29, 2016

Dear Sarah Weber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Munde Jellm
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K160685

Device Name Surveyor S4 Mobile Monitor

Indications for Use (Describe)

The Surveyor S4 Mobile Monitor is indicated for use:

· The Surveyor S4 is indicated for use to acquire, analyze and output multi-lead ECG and SpO2 signals. The Mortara Surveyor S4 mobile monitor facilitates the monitoring of ECG and SpO2 parameters when used in conjunction with alarm annunciating devices such as the Surveyor Central Station for telemetric monitoring; however, the Surveyor S4 has no alarming capabilities of its own.

  • The Surveyor S4 processes Heart rate, ST segment deviation and arrhythmias such as, but not limited to, Ventricular Tachycardia, Ventricular Fibrillation, and Asystole by utilizing Mortara VERITAS™ ECG algorithm

  • The Surveyor S4 can deliver data from a 3/5 lead cable and diagnostic 12-lead ECG data to a receiving Surveyor Central station.
    • The Surveyor S4 mobile monitor is indicated for use as a mobile radio transceiver, allowing the patient to be ambulatory within the range of a Wi-Fi information infrastructure.

· The Surveyor S4 mobile monitor is indicated for use in adults, adolescents and children.

• The Surveyor S4 mobile monitor is a prescription device intended to be used by knowledgeable healthcare professionals within a healthcare or clinical setting. It is not intended as a sole means of diagnosis.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo above the text "Traditional 510(k) Notification". The logo is a blue shape that resembles a heart with a flat bottom. The text is in a simple, sans-serif font and is centered below the logo.

Section 5

510(k) Summary Statement

1. Submitter

Mortara Instrument, Inc.7865 North 86th StreetPrimary Contact
Milwaukee, WI 53224Sarah Weber
Senior Regulatory Affairs
Manager
sarah.weber@mortara.com
Telephone 414-354-1600Fax 414-354-4760Secondary Contact
Margaret Mucha
Director of Global Regulatory
Affairs
meg.mucha@mortara.com

2. Product Names

Device Trade NameSurveyor S4 Mobile Monitor
Common/ Usual NameMonitor, Physiological, Patient(With Arrhythmia Detection orAlarms)OximeterElectrocardiograph
ClassificationMonitor, Physiological, Patient (WithArrhythmia Detection or Alarms)870.1025MHX
Oximeter870.2700DQA
Electrocardiograph870.2340DPS

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a blue heart shape. The heart is formed by two curved lines that meet at the top and bottom. The heart is solid blue, and the background is white. The heart is centered in the image.

Traditional 510(k) Notification Note: There are no previous submissions for this device

3. Predicate Device to which this is Substantially Equivalent

Surveyor S4 Mobile MonitorK141020
ApexPro Telemetry SystemK080251

The primary predicate device is Surveyor S4 Mobile Monitor, K141020.

The S4 predicate device has not been subject to a design-related recall. The ApexPro predicate device does has not been under recall since 2011, Z-1522-2011.

4. Device Description

The Surveyor™ S4 Mobile Monitor is a battery operated patient worn transceiver that communicates via IEEE 802.11 Wi-Fi technology to the Surveyor™ Central Station (K131929) which is the primary monitoring display and alarm source.

The Surveyor S4 transmitter has a touch sensitive, color display allowing for operational conveniences such as a battery level indicator, Wi-Fi Signal Strength and Surveyor Central Station slot information. The Surveyor S4 can display physiological waveforms, measured parameters, demographic data and other status information. The Surveyor S4 acquires, filters, and processes ECG or ECG and SpO2 parameters and sends filtered signals and parameters to the Surveyor Central Station where alarm processing and annunciation occurs.

The Surveyor Central is designed to communicate with the Surveyor S4 within a healthcare facilities Wi-Fi network infrastructure. The Surveyor Central provides primary monitoring of telemetry patients when used with the Surveyor S4. The central station displays numeric values, waveforms, alarm generation and management, data storage, individualized patient management, report generation and printing functionality.

The Surveyor S4 provides continuous data acquisition which can be visualized on the S4 itself for the following parameters:

  • o Electrocardiography (ECG) including heart rate and ECG waveforms
  • · Saturation percentage of peripheral oxygen (SpO2), including SpO2 plethysmogram waveform, pulse rate, and signal quality index.
  • o No alarm status or processing information of ECG or SpO2 is shown on the Surveyor S4 device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a white heart shape inside of a blue square. The heart shape is formed by two curved lines that meet at the bottom and form a point at the top. The blue square provides a background for the heart shape, making it stand out. The logo is simple and clean, with a focus on the heart shape.

Traditional 510(k) Notification

  • o The Surveyor S4 processes Heart rate, ST segment deviation and arrhythmias, by utilizing Mortara VERITAS™ ECG algorithm and sends ECG data packets, including arrhythmias such as, but not limited to, Ventricular Tachycardia, Ventricular Fibrillation, and Asystole to the Surveyor Central Station for alarm processing and annunciation.
  • o The Surveyor S4 utilizes proprietary ECG lead cables available in AHA or IEC, snap or clip, 3-Lead, 5-Lead, and diagnostic 12 Lead configurations.

SpO2

  • o The Surveyor S4 processes SpO2 percentage, pulse rate and signal quality through the Mortara SpO2 algorithm and sends data packets for SpO2 to the Surveyor Central Station for alarm processing and annunciation.
  • o Multiple SpO2 sensor styles allow connection of adults, adolescents, and children. SpO2 sensors are specified for use with the Surveyor S4 and available through Mortara Instrument.

5. Intended Use

The Surveyor S4 Mobile Monitor is indicated for use:

·The Surveyor S4 Mobile Monitor is indicated for use to acquire, analyze and output multi-lead ECG and SpO2 signals.

The Mortara Surveyor S4 mobile monitor facilitates the monitoring of ECG and SpO2 parameters when used in conjunction with alarm annunciating devices such as the Surveyor Central Station for telemetric monitoring; however, the Surveyor S4 has no alarming capabilities of its own.

  • · The Surveyor S4 processes Heart Rate, ST segment deviation and arrhythmias such as, but not limited to, Ventricular Tachycardia, Ventricular Fibrillation, and Asystole by utilizing Mortara VERITAS™ ECG algorithm.
  • · The Surveyor S4 can deliver data from a 3/5-lead cable and_diagnostic 12-lead ECG data to a receiving Surveyor Central_station
  • The Surveyor S4 Mobile Monitor is indicated for use as a mobile radio transceiver, allowing the patient to be ambulatory within the range of a Wi-Fi information infrastructure.

ECG

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with a heart shape in the center, formed by two curved lines. The heart is white, and the background is blue. Below the logo, the text "Traditional 510(k) Notification" is written in a simple, sans-serif font. The text is black and appears to be centered beneath the logo.

· The Surveyor S4 Mobile Monitor is indicated for use in adults, adolescents and children.

· The Surveyor S4 Mobile Monitor is a prescription device intended to be used by knowledgeable healthcare professionals within a healthcare or clinical setting. It is not intended as a sole means of diagnosis.

6. Technological Characteristics

The Surveyor S4 Mobile Monitor employs the same functional scientific technology as its predicate device Surveyor S4 Mobile Monitor (K141020). It is a prescription device designed to acquire and transmit diagnostic quality ECG in a clinical setting while allowing the patient to be ambulatory. The Surveyor S4 Mobile Monitor is suitable for use as a telemetry solution within centralized ECG monitoring systems or other clinical settings for patients connected to telemetry transceivers

The Surveyor S4 Mobile Monitor is designed to work with ECG when used with a compatible multi-parameter Telemetry Central Station System such as the Mortara Surveyor Central Station (K131929) and other compatible receiving devices.

The Surveyor S4 Mobile Monitor was designed and manufactured by Mortara Instrument according to 21 CFR Part 820. The Surveyor S4 Mobile Monitor is substantially equivalent to The Surveyor S4 Mobile Monitor (Predicate K141020) with the following technological differences:

  • . Added 3/4/5 lead ECG
  • Added shielded 3/4/5- wire ECG patient cable ●
  • Added SpO2 functionality ●
  • Added SpO2 sensor with optional extender cable .
510(k)K080251K141020Present SubmissionChangeexplanation
BRANDApexProSurveyor S4 MobileMonitor (Primary)Surveyor S4 MobileMonitor
COMPANYGE MedicalSystemsInformationTechnologiesMortara Instrument, Inc.Mortara Instrument, Inc.
SoftwareVersionProprietaryV1.00V1.2.0
510 (K)NumberK080251K141020Present Application
Indications forUseThe ApexProTelemetrySystem isThe Surveyor S4 MobileMonitor is indicated foruse:The Surveyor S4 MobileMonitor is indicated for use:• The Surveyor S4
Equivalent.Indications revised
510(k)K080251K141020Present SubmissionChange explanation
BRANDApexProSurveyor S4 Mobile Monitor (Primary)Surveyor S4 Mobile Monitor
COMPANYGE Medical Systems Information TechnologiesMortara Instrument, Inc.Mortara Instrument, Inc.
Software VersionProprietaryV1.00V1.2.0
510 (K) NumberK080251K141020Present Application
intended for use under the direct supervision of a licensed healthcare practitioner. The system is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The system processes this physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations. The ApexPro Telemetry System is intended to be installed in the hospital or clinical environment in order to provide clinicians with patient physiological data, while allowing for patient mobility. These systems areThe Surveyor S4 Mobile Monitor is indicated for use in adult & pediatric patient populations. The Mortara Surveyor S4 Mobile Monitor facilitates the monitoring of ECG signals. The Surveyor S4 Mobile Monitor is a prescription device intended to be used by knowledgeable healthcare professionals within a healthcare facility or clinical pharmacology unit. The Surveyor S4 Mobile Monitor is indicated for use in a clinical setting by a physician, or by trained personnel acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The Surveyor S4 Mobile Monitor is indicated for use to acquire and output electrocardiographic data. The Surveyor S4 Mobile Monitor is indicated for use as aMobile Monitor is indicated for use to acquire, analyze and output multi-lead ECG, SpO2 and other physiological signals. The Mortara Surveyor S4 Mobile Monitor facilitates the real-time monitoring of clinical parameters when used in conjunction with alarm annunciating devices such as a Central Station for telemetric monitoring. The Surveyor S4 can deliver diagnostic 12-lead ECG data to a receiving station. The Surveyor S4 Mobile Monitor is indicated for use as a mobile radio transceiver, allowing the patient to be ambulatory within the range of a Wi-Fi information infrastructure. The Surveyor S4 Mobile Monitor is indicated for use in adults, adolescents and children. The Surveyor S4 Mobile Monitor is a prescription device intended tofor clarity and to provide patient population categories in accordance with FDA guidance "Providing Information about Pediatric Uses of Medical Devices - Guidance for Industry and Food and Drug Administration Staff" dated May 1, 2014. Added SpO2 functionality.
510(k)K080251K141020Present Submission
BRANDApexProSurveyor S4 MobileMonitor (Primary)Surveyor S4 MobileMonitor
COMPANYGE MedicalSystemsInformationTechnologiesMortara Instrument, Inc.Mortara Instrument, Inc.Changeexplanation
SoftwareVersionProprietaryV1.00V1.2.0
510 (K)NumberK080251K141020Present Application
typicallydeployed insub-acutecare areas inhospitals orclinical siteswhere patientmobility canenhance theeffectivenessof the medicalproceduresadministered.Thephysiologicalparametersmonitoredinclude ECG,noninvasivebloodpressure,non-invasivetemperatureand SpO2.The ApexProTelemetrySystem isintended toprovide ECGdatavia Ethernetto thecomputerplatform forprocessing.TheApexProTelemetrySystem isalso intendedto providephysiologicdata over theUnity Networkto clinicalinformationsystems fordisplay.radiofrequencyphysiologicalsignaltransceiver,receiving anddelivering real-time acquisitionand transmissionof simultaneouselectrocardiographic data, whileallowing thepatient to beambulatorywithin the rangeof the antennanetwork.The Mortara Li-IonBattery Charger isintended for chargingonly the MortaraRechargeable Li-Ionbattery pack.be used byknowledgeablehealthcareprofessionalswithin a healthcareor clinical setting.It is not intendedas a sole meansof diagnosis.
510(k)K080251K141020Present Submission
BRANDApexProSurveyor S4 MobileMonitor (Primary)Surveyor S4 MobileMonitor
COMPANYGE MedicalSystemsInformationTechnologiesMortara Instrument, Inc.Mortara Instrument, Inc.Changeexplanation
SoftwareVersionProprietaryV1.00V1.2.0
510 (K)NumberK080251K141020Present Application
OperatingSystemProprietaryLinux derivativeLinux derivativeEquivalent
NetworkInterfaceDigital/608-614 MhzDigital/Wireless networkinterface perIEEE 802.11g/nDigital/Wireless networkinterface per IEEE802.11g/nEquivalentBoth devicesuse a digital,wireless form ofcommunicationbut utilizedifferentfrequencyranges andsignalingprotocol.
TargetPopulationAdult,pediatricAdult, pediatricAdults, adolescents,childrenEquivalent.Subject deviceincludes asubset of theAdults toNeonatepopulation of thepredicatedevice.Revised patientpopulationcategories inaccordance withFDA quidance"ProvidingInformationabout PediatricUses of MedicalDevices -Guidance forIndustry andFood and DrugAdministrationStaff" datedMay 1, 2014.
ECG3, 5, or 6-leadECG12-lead diagnostic ECG3/4/5-lead, and12-leaddiagnostic ECGEquivalent
510(k)K080251K141020Present Submission
BRANDApexProSurveyor S4 MobileMonitor (Primary)Surveyor S4 MobileMonitor
COMPANYGE MedicalSystemsInformationTechnologiesMortara Instrument, Inc.Mortara Instrument, Inc.Changeexplanation
SoftwareVersionProprietaryV1.00V1.2.0
510 (K)NumberK080251K141020Present Application
ECG SamplingRate240samples/second500 Samples/second500 samples/secEquivalentThe subjectdevice isequivalent to theprevioussubmission ofthe samedevice.
SpO2YesNoYesNewfunctionalityadded
GraphicDisplayNoYesYesEquivalent
FrequencyTransmission608-614WMTS or1,395-1,400WMTS,dependent ontransmittermodelIEEE 802.11g/n, 2.4GHz bandIEEE 802.11g/n, 2.4GHz bandEquivalent
TransmissionTypeChannelizedDigital wireless inaccordance withIEEE 802.11g/nDigital wireless inaccordance withIEEE 802.11g/nEquivalent
Battery Type2 AA3 AA disposable, orrechargeable Li-Ion Pack3 AA disposable, orrechargeable Li-Ion PackEquivalent
Lead CheckYesYesYesEquivalent
InternalAntennaNoYesYesEquivalent
CompatibleSystemsCARESCAPECentral Station(K133882)Surveyor Central Station(K131929)Surveyor Central Station(K131929)Equivalent
Patient Cable3, 5, and 6-wireMulti-LinkLeadwiresLeadForm 10-wire ECGLeadForm 10-wire ECGShielded 3/4/5-wire ECGEquivalent
510(k)K080251K141020Present SubmissionChange explanation
BRANDApexProSurveyor S4 MobileMonitor (Primary)Surveyor S4 MobileMonitor
COMPANYGE MedicalSystemsInformationTechnologiesMortara Instrument, Inc.Mortara Instrument, Inc.
SoftwareVersionProprietaryV1.00V1.2.0
510 (K)NumberK080251K141020Present Application
SpO2 CableMasimouSpO2 pulseoximetryutilizingMasimo SETMeasure-throughMotion andlow perfusionpulseoximetrytechnologywithcompatibleLNCS stylesensors.N/ASpO2 sensors - disposable,clip, SoftTipTMw/ optional extender cable
EquivalentAdded SpO2 asa new feature tosubject device.

Comparison Matrix:

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

7. Determination of Substantial Equivalence – Non-clinical

Software verification and validation testing was conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content fo Premarket Submissions for Software Contained in Medical Devices" and "Off-the-Shelf Software Use in Medical Devices."

The Surveyor S4 Mobile Monitor was designed and tested for compliance with the applicable clauses of the following standards:

Image /page/11/Picture/4 description: The image shows a logo above the text "Traditional 510(k) Notification". The logo is a blue square with a white heart shape inside. The text is centered below the logo and is in a simple, sans-serif font.

  • . IEC 60601-1:2005 Ed:3 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • IEC 60601-2-49: 2011 Medical electrical equipment Part . 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
  • IEC 60601-1-2:2007 Medical Electrical Equipment Part . 1-2: General requirements for safety and essential performance - Collateral standard: Electromagnetic Compatibility - Requirements and tests

{12}------------------------------------------------

  • ISO 80601-2-61:2011 Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
  • IEC 60601-2-25:2011-10 Edition 2.0, Medical Electrical . Equipment. Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
  • ANSI/AAMI EC57:2012 Testing and reporting performance results of cardiac rhythm, ST segment measurement algorithms, ventricular fibrillation algorithm, and irregular rhythm (atrial fibrillation) algorithm

8. Determination of Substantial Equivalence – Clinical

A basic pulse oximeter clinical study was conducted to demonstrate the Surveyor S4 Mobile Monitor SpO2 compliance to IEC 80601-2-61. The study met the requirements of ISO 80601-2-61-2011 Clause 201.12.1.

9. Conclusion

Changes to the Surveyor S4 Mobile Monitor include:

  • · Added SpO2 functionality
  • . Added SpO2 extender cable
  • Added shielded 3/4/5-wire ECG .

Mortara Instrument, Inc. considers the Surveyor S4 Mobile Monitor to be as safe, as effective and performance is substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.