K Number
K160621

Validate with FDA (Live)

Device Name
SIGNAPioneer
Manufacturer
Date Cleared
2016-03-29

(25 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.

Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Device Description

The SIGNA Pioneer features a 3.0T superconducting magnet with a 70cm bore size.

The RF receiver is equipped with up to 97 RF channels.

The system uses a combination of time-varying magnetic fields (gradients) and RF transmissions to obtain information regarding the density and position of nuclei exhibiting magnetic resonance.

The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences and reconstruction algorithms.

The SIGNA Pioneer uses multi-drive RF transmit for imaging of the head and body regions.

The SIGNA Pioneer is designed to conform to NEMA DICOM standards.

AI/ML Overview

The provided text is a 510(k) summary for the GE Healthcare SIGNA Pioneer Magnetic Resonance Diagnostic Device. It states that clinical studies were not required to support substantial equivalence for this submission (K160621). Therefore, the document does not contain information regarding detailed acceptance criteria, specific device performance metrics, sample sizes, provenance of data, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance for a study proving the device meets acceptance criteria.

The information provided only discusses the device's intended use, technological specifications, and compliance with voluntary standards, comparing it to a predicate device (SIGNA Pioneer, K143345). The substantial equivalence determination is based on non-clinical tests, design controls, and the fact that the indications for use are identical to the predicate device.

Therefore, it is not possible to provide the requested table and details as the document explicitly states that clinical studies were not used to support this 510(k) clearance.

The relevant statements from the document are:

  • "Summary of Clinical Tests: The subject of this premarket submission, SIGNA Pioneer, did not require clinical studies to support substantial equivalence." (Page 4)
  • "Scanning of human subjects on the SIGNA Pioneer system has been conducted at GE Healthcare facilities as part of design validation activities in order to ensure that the modified device meets user requirements." (Page 4)

This indicates that while some human subject scanning has occurred for design validation, it was not structured as a clinical study for demonstrating substantial equivalence against specific performance criteria for regulatory purposes.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of human profiles facing to the right, with three distinct head shapes stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 29, 2016

GE Healthcare % Mr. Glen Sabin Regulatory Affairs Director 3200 N. Grandview Blvd. WAUKESHA WI 53188

Re: K160621

Trade/Device Name: SIGNA Pioneer Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 3, 2016 Received: March 4, 2016

Dear Mr. Sabin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours

Michael D.'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K160621

Device Name SIGNA Pioneer

Indications for Use (Describe)

The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.

Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo of General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling lines, resembling water or air currents, which add a dynamic element to the design. The overall impression is one of a classic and recognizable corporate symbol.

Section 5: 510(k) Summary

SIGNA Pioneer

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a light blue hue, and the background is white.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:March 25, 2016
Submitter:GE Healthcare, (GE Healthcare Japan Corporation)7-127, Asahigaoka 4-chome, Hino-shi,Tokyo 191-8503 JAPAN
Primary Contact Person:Glen SabinRegulatory Affairs DirectorGE Healthcare, (GE Medical Systems, LLC)Phone: 262-521-6848
Secondary ContactPerson:Steve KachelmeyerSenior Regulatory Affairs DirectorGE Healthcare, (GE Medical Systems, LLC)Phone: 262-548-2432
Device Trade Name:SIGNA Pioneer
Common/Usual Name:Magnetic Resonance Diagnostic Device
Classification Names:Magnetic Resonance Diagnostic Device per 21 CFR 892.1000
Product Code:LNH
Predicate Device:SIGNA Pioneer (K143345)
Device Description:The SIGNA Pioneer features a 3.0T superconducting magnetwith a 70cm bore sizeThe RF receiver is equipped with up to 97 RF channels.The system uses a combination of time-varying magneticfields (gradients) and RF transmissions to obtain informationregarding the density and position of nuclei exhibitingmagnetic resonance.The system can image in the sagittal, coronal, axial, oblique,and double oblique planes, using various pulse sequences andreconstruction algorithms.The SIGNA Pioneer uses multi-drive RF transmit for imagingof the head and body regions.The SIGNA Pioneer is designed to conform to NEMADICOM standards.
Intended Use:The SIGNA Pioneer is a whole body magnetic resonancescanner designed to support high resolution, high signal-to-noise ratio, and short scan times.It is indicated for use as a diagnostic imaging device toproduce axial, sagittal, coronal, and oblique images,spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of the entirebody, including, but not limited to, head, neck, TMJ, spine,breast, heart, abdomen, pelvis, joints, prostate, blood vessels,and musculoskeletal regions of the body.Depending on the region of interest being imaged, contrastagents may be used.The images produced by the SIGNA Pioneer reflect the spatialdistribution or molecular environment of nuclei exhibitingmagnetic resonance. These images and/or spectra wheninterpreted by a trained physician yield information that mayassist in diagnosis.
Technology:The SIGNA Pioneer employs the same fundamental scientifictechnology as its predicate device.The SIGNA Pioneer system has been modified by theintroduction of a new 65 receive channel configuration, inaddition to the previously cleared 97 channel configuration.The modified SIGNA Pioneer system also includesenhancements to the software that controls SpecificAbsorption Rate (SAR).
Determination ofSubstantial Equivalence:Summary of Non-Clinical Tests:
Like the predicate device, the SIGNA Pioneer complies withthe following voluntary standards:IEC 60601-1 IEC 60601-1-2 IEC 60601-2-33 ISO 10993-1
In addition, the SIGNA Pioneer complies with the applicableNEMA MS standards for MRI and NEMA PS3 standard forDICOM, as does the predicate device.
The following quality assurance measures were applied to thedevelopment of the system, as they were for the predicate: Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation)
Summary of Clinical Tests:
The subject of this premarket submission, SIGNA Pioneer, didnot require clinical studies to support substantial equivalence.
Scanning of human subjects on the SIGNA Pioneer system hasbeen conducted at GE Healthcare facilities as part of designvalidation activities in order to ensure that the modified devicemeets user requirements.
Conclusion:The Indications for Use of the SIGNA Pioneer are identical tothe predicate device. The modifications to the SIGNA Pioneersystem do not change the fundamental scientific technology ofthe device. The results of design controls activitiesdemonstrate that the SIGNA Pioneer is substantiallyequivalent to the predicate with regards to safety and efficacy.GE Healthcare considers the SIGNA Pioneer to be as safe, aseffective, and performance is substantially equivalent to thepredicate device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three curved lines emanating from the circle, one on the top left, one on the top right, and one on the bottom left. The logo is simple and recognizable.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has a white border and a swirling design around the letters, giving it a dynamic and recognizable appearance.

GE Healthcare 510(k) Premarket Notification Submission

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.