K Number
K160595

Validate with FDA (Live)

Date Cleared
2016-11-18

(262 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PMS Steripack Tyvek Sterilization Pouches (TP) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method.

PMS Steripack Tyvek Sterilization Pouches (TP) with Chemical Indicator is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical devices and also maintain sterility of the enclosed devices until used.

The recommended (and validated) sterilization cycle parameters are:

The Steris V-PRO 1 PLUS LUMEN CYCLE is a single preprogramed standard sterilization cycle with NO cycle parameter variables allowed by the end user.

Chemical process indicator on the exterior of the Tyvek pouches and rolls indicates that the pouch has been exposed to V-PRO 1 Plus, Lumen Cycle sterilization process by changing color from red to blue or (lighter).

Devices with stainless steel lumens with the following internal diameter (ID) and maximum length can be processed in V-PRO 1 Plus, Lumen Cycle.

Pouch SizeMax. lumen lengthMin. internal diameterMax # Lumens
50 x200mm50 mm1.5 mm1
150 x300mm3 mm200 mm1
350 x600mm3 mm400 mm1
Device Description

PMSSteriback Tvvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are manufactured from tyvek material/laminated PET/PE film that are heat sealed on sides. Triple band seal provides three independent barriers to contamination, while reducing the risk of fibre tear.

PMSSteripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are divided into 2 main configuration groups as below.

Sterilization Pouches, Flat:

These pouches are manufactured from twek material/ laminated PET/PE plastic film that are heat sealed on 3 sides, the remaining open side is heat sealed by the end user.

Sterilization Rolls, Flat:

These rolls are manufactured from tyvek material and laminated PET/PE plastic film that are heat sealed on opposite two sides. The required size is cut from the roll and the remaining open sides are heat sealed by the end user.

AI/ML Overview

This document is a 510(k) summary for the PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator. It compares the submission device to a predicate device (STERIS Vis-U-All Low Temperature Sterilization Pouch/Tubing) to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and study information provided, focusing on the requested categories:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Design and Construction & Material Composition"Tyvek material and heat sealed laminated PET/PE plastic film. Externally printed H2O2 chemical process indicator ink. Pouches are manufactured from tyvek material/plastic film that are heat sealed on 3 sides. The remaining side is left opened and heat sealed by the end user. Rolls are manufactured from tyvek material/plastic film that are heat sealed on opposite two sides. The required size is cut from the roll and the remaining open two sides are heat sealed by the end user."
Indications for Use"PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method. PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical devices and also maintain sterility of the enclosed devices until used." "The recommended (and validated) sterilization cycle parameters are: The Steris V-PRO 1 PLUS LUMEN CYCLE is a single preprogramed standard sterilization cycle with NO cycle parameter variables allowed by the end user. Chemical process indicator on the exterior of the Tyvek pouches and rolls indicates that the pouch has been exposed to V-PRO 1 Plus, Lumen Cycle sterilization process by changing color from red to blue or (lighter). Devices with stainless steel lumens with the following internal diameter (ID) and maximum length can be processed in V-PRO 1 Plus, Lumen Cycle." (Specific lumen claims are listed in the document's Table 1.) "Worst Case Load: The worst case load for the 50 mm x 200 mm pouch is 0.058 lbs. of instruments. The worst case load for the 150 mm x 300 mm pouch is 0.630 lbs. of instruments. The worst case load for the 350 mm x 600 mm pouch is 3.054 lbs. of instruments."
Sterilization Properties"V-PRO 1 Plus, Lumen Cycle"
Principle of Operations"Medical device to be sterilized is put into pouch or roll and the open parts of the packages are closed by the end user with heat sealing. Sterilization Pouches then are subjected to sterilization operation in recommended (validated) VHP (Vaporized Hydrogen Peroxide) sterilization devices and the cycle as V-PRO 1 PLUS, Lumen Cycle. Sterilant penetration occurs through the tyvek material into the packages and microorganisms on the surfaces of the enclosed medical devices are destroyed via the sterilant."
Principles of Operation for Chemical Indicators"The Process Indicators Ink printed on the tyvek material will exhibit a color change after the pouch is exposed to Hydrogen Peroxide (H2O2) gas. Hydrogen Peroxide (H2O2) sterilization printed indicator bar changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor." The indicators are classified as Class 1 type process indicator according to ISO 11140-1 standard.
Shelf Life (Prior to Sterilization)3 years (Note: Predicate device is 2 years, but this is reported as "YES" for substantial equivalence)
Shelf Life (Post Sterilization)"Sterility was maintained for at least 30 days after processing in V-PRO 1 Plus, Lumen Cycle" (Note: Predicate device is 1 year, but this is reported as "YES" for substantial equivalence and meets the specified 30-day maintenance claim).
Configurations & Dimensions"Various sizes (width and height/length)"
Sterilant Penetration"The sterilization (VHP) was validated to a sterility assurance level (SAL) of 10-6"
Microbial Barrier Properties"Sterility was maintained for at least 30 days after processing in V-PRO 1 Plus, Lumen Cycle"
Material Compatibility"Suitability for use in V-PRO 1 Plus, Lumen Cycle sterilization process and cycle parameters."
Biocompatibility"Not direct patient-contacting devices; Materials are non-toxic and meet ISO 10993-1 requirements."
Package Integrity"Porous material providing a microbial barrier."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a separate "test set" sample size for evaluating the device's performance against the acceptance criteria. Instead, it refers to "non-clinical performance tests summaries" and mentions validation of sterilization to a "sterility assurance level (SAL) of 10-6".

  • Sample Size for Test Set: Not explicitly stated as a single "test set" sample size in the provided text. The validation to SAL 10-6 typically involves a statistically significant number of biological indicator tests across multiple cycles and loads, but the exact number isn't quantified here. Individual tests, such as those for microbial barrier, material compatibility, and package integrity, would have their own sample sizes, but these are not precisely detailed.
  • Data Provenance: Not specified. The document states that PMS TIBBI CIHAZLAR TEKNOLOJISI SANA YI VE TICARET ANONIM ŞIRKETI is based in Turkey, suggesting the studies could have been conducted there or by a contracted lab. It doesn't specify if the data is retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This type of device (sterilization pouch with chemical indicator) does not typically involve human expert interpretation for "ground truth" in the same way clinical diagnostic imaging might. The "ground truth" for its performance is established through objective laboratory testing (e.g., microbiology for sterility, physical testing for barrier properties, chemical reaction for indicators). Therefore, information about experts for ground truth is not applicable in this context.

4. Adjudication Method for the Test Set

Not applicable for this type of device and testing. Performance is assessed through objective measurements against technical standards (e.g., ISO 11140-1, ISO 11607-1).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This device is a sterilization pouch and not an AI or diagnostic imaging system that would be evaluated for human reader improvement.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This is a physical medical device (sterilization packaging), not an AI algorithm. Its performance is inherent to its physical and chemical properties.

7. Type of Ground Truth Used

The "ground truth" for this device's performance is established through:

  • Sterility Assurance Level (SAL) of 10-6: Achieved through microbiological validation studies, which would involve biological indicators.
  • Microbial Barrier Properties: Determined through physical challenge tests (e.g., bacterial challenge tests, dye penetration tests).
  • Material Compatibility: Assessed through chemical and physical interaction studies with the sterilant and within the sterilization cycle.
  • Biocompatibility: Evaluated against ISO 10993-1 standards for materials that indirectly contact patients, typically through testing of extracts from the materials.
  • Chemical Indicator Efficacy: Verified by observing color change response to exposure/non-exposure to the sterilant, often against standards like ISO 11140-1.
  • Package Integrity: Assessed through tests like seal strength, visual inspection, and aging studies.
  • Outcomes data: While "patient outcomes" are not directly measured for the pouch itself, the ultimate outcome is the sterilization of medical devices, which is fundamental to patient safety.

8. Sample Size for the Training Set

Not applicable. This is not a machine learning or AI device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2016

PMS TIBBI CIHAZLAR TEKNOLOJISI SANA YI VE TICARET ANONIM ŞIRKETI Derya Dikici, Ph.D. Business Development Manager Karaduvar Mahallesi, Serbest Bolge 11. Cadde No:46 Mersin Serbest Bolgesi, 33020, Akdeniz/MERSIN/TURKEY

Re: K160595

Trade/Device Name: PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: October 10, 2016 Received: October 19, 2016

Dear Derya Dikici:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160595

Device Name

PMSSteripack Tyvek Sterilization Pouch(TP) and Roll(TY) with Chemical Indicator

Indications for Use (Describe)

PMS Steripack Tyvek Sterilization Pouches (TP) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method.

PMS Steripack Tyvek Sterilization Pouches (TP) with Chemical Indicator is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical devices and also maintain sterility of the enclosed devices until used.

The recommended (and validated) sterilization cycle parameters are:

The Steris V-PRO 1 PLUS LUMEN CYCLE is a single preprogramed standard sterilization cycle with NO cycle parameter variables allowed by the end user.

Chemical process indicator on the exterior of the Tyvek pouches and rolls indicates that the pouch has been exposed to V-PRO 1 Plus, Lumen Cycle sterilization process by changing color from red to blue or (lighter).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

EF PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

Devices with stainless steel lumens with the following internal diameter (ID) and maximum length can be processed in V-PRO 1 Plus, Lumen Cycle.

Pouch SizeMax. lumen lengthMin. internal diameterMax # Lumens
50 x200mm50 mm1.5 mm1
150 x300mm3 mm200 mm1
350 x600mm3 mm400 mm1

Table 1. Lumen Claims

Worst Case Load:

The worst case load for the 50 mm x 200 mm pouch is 0.058 lbs. of instruments. The worst case load for the 150 mm x 300 mm pouch is 0.630 lbs. of instruments. The worst case load for the 350 mm x 600 mm pouch is 3.054 lbs. of instruments.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for PMS. The logo consists of a red and gray graphic on the left, followed by the letters "PMS" in gray. The graphic is a stylized shape with a red upper portion and a gray lower portion.

K160595

510 (k) Summary For PMSSteripack Tyvek Sterilization Pouch(TP) and Roll (TY) with Chemical Indicator

1.0. Name. address and contact

PMS TIBBI CIHAZLAR TEKNOLOJISI SANAYI VE TICARET ANONIM ŞIRKETI Karaduvar Mahallesi, Serbest Bolge 11. Cadde No:46 Mersin Serbest Bolgesi, 33020, Akdeniz/MERSIN/TURKEY

Phone90 324 2387042 – 90 542 648 6312
Fax90 324 2386549
Prepared by :Taner ErsenQuality Management Representative
Contact :Derya DikiciBusiness Development Manager
Phone :90 324 2387042- 90 542 648 6312
Fax :90 324 2386549
E-mail :derya.dikici@pmsmedikal.com
Date Prepared:November 17, 2016

{5}------------------------------------------------

2. Device Name

Trade Name: PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator

Common/Usual Name: Tyvek Sterilization Pouch and Roll

Device Classification Name: Wrap, Sterilization - Indicator, Physical/Chemical Sterilization Process

Product Code: FRG

Product Classification: Class II, 21 CFR 880.6850, General Hospital (FRG)

3. Predicate Device

K140487, STERIS Vis-U-All Low Temperature Sterilization Pouch/Tubing

4.Indications for use

PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method.

PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical devices and also maintain sterility of the enclosed devices until used.

The recommended (and validated) sterilization cvcle parameters are:

The Steris V-PRO 1 PLUS LUMEN CYCLE is a single preprogramed standard sterilization cycle with NO cycle parameter variables allowed by the end user.

Chemical process indicator on the exterior of the Tyvek pouches and rolls indicates that the pouch has been exposed to V-PRO 1 Plus, Lumen Cycle sterilization process by changing color from red to blue or (lighter).

Devices with stainless steel lumens with the following internal diameter (ID) and maximum length can be processed in V-PRO 1 Plus, Lumen Cycle.

PouchSizeMax.lumenlengthMin.InternaldiameterMax #Lumens
50 x200mm50 mm1.5 mm1
150 x300mm3 mm200 mm1
350 x600mm3 mm400 mm1

Table 1. Lumen Claims

Worst Case Load:

The worst case load for the 50 mm x 200 mm pouch is 0.058 lbs. of instruments. The worst case load for the 150 mm x 300 mm pouch is 0.630 lbs. of instruments. The worst case load for the 350 mm x 600 mm pouch is 3.054 lbs. of instruments.

{6}------------------------------------------------

5. Device Description

PMSSteriback Tvvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are manufactured from tyvek material/laminated PET/PE film that are heat sealed on sides. Triple band seal provides three independent barriers to contamination, while reducing the risk of fibre tear.

PMSSteripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are divided into 2 main configuration groups as below.

Sterilization Pouches, Flat:

These pouches are manufactured from twek material/ laminated PET/PE plastic film that are heat sealed on 3 sides, the remaining open side is heat sealed by the end user.

Sterilization Rolls, Flat:

These rolls are manufactured from tyvek material and laminated PET/PE plastic film that are heat sealed on opposite two sides. The required size is cut from the roll and the remaining open sides are heat sealed by the end user.

Available Sizes /Configurations for TP&TY:

Table 2. Product List of "PMSSteripack Tyvek Sterilization Pouch (TP) with Chemical Indicator "

PRODUCT LIST
(A)(B)(A)(B)(L)(V)Single(P)Distance(S)(T)
ProductTotalTotalTotalTotalSealThicknes
NameWidthHeightWidthHeightWidthSealBetweenDiameter (µm)
(mm)(mm)(in.)(in.)(mm)Width2 Seals(mm)
(mm)(mm)
TP05205020028102210223±18
TP052550250210102210223±18
TP053050300212102210223±18
TP053250320212¾102210223±18
TP053550350214102210223±18
TP75157515036102210223±18
TP75187518037102210223±18
TP75207520038102210223±18
TP75237523039102210223±18
TP752575250310102210223±18
TP752875280311102210223±18
TP753075300312102210223±18
TP753575350314102210223±18
TP754075400316102210223±18
TP754575450318102210223±18
TP0815801506102210223±18
TP081780170102210223±18
TP0818801807102210223±18
TP0820802008102210223±18
TP08278027010¾102210223±18
TP08608060024102210223±18
TP09569056022102210223±18
TP09609060024102210223±18
TP09619061024102210223±18
TP101510015046102210223±18
TP101810018047102210223±18
TP10191001904102210223±18
TP102010020048102210223±18
TP102310023049102210223±18
TP1025100250410102210223±18
TP1026100260410¾102210223±18
TP1027100270410¾102210223±18
TP1028100280411102210223±18
TP1030100300412102210223±18
TP1033100330413102210223±18
TP1035100350414102210223±18
TP1036100360414102210223±18
TP1040100400416102210223±18
TP1045100450418102210223±18
TP1050100500420102210223±18
TP1056100560422102210223±18
TP1057100570422102210223±18
TP1060100600424102210223±18
TP1061100610424102210223±18
TP112811028011102210223±18
TP122512025010102210223±18
TP123012030012102210223±18
TP124012040016102210223±18
TP125012050020102210223±18
TP1325130250510102210223±18
TP1338130380515102210223±18
TP152015020068102210223±18
TP1525150250610102210223±18
TP1527150270610%102210223±18
TP1528150280611102210223±18
TP1530150300612102210223±18
TP1532150320612½102210223±18
TP1535150350614102210223±18
TP1538150380615102210223±18
TP1540150400616102210223±18
TP1543150430617102210223±18
TP1545150450618102210223±18
TP1550150500620102210223±18
TP1552150520620%102210223±18
TP1555150550622102210223±18
TP1556150560622102210223±18
TP164016040016102210223±18
TP165816058023102210223±18
TP1722170220102210223±18
TP173517035014102210223±18
TP1830180300712102210223±18
TP1832180320712½102210223±18
TP1846180460718102210223±18
TP193319033013102210223±18
TP202020020088102210223±18
TP2025200250810102210223±18
TP2027200270810¾102210223±18
TP2030200300812102210223±18
TP2033200330813102210223±18
TP2035200350814102210223±18
TP2040200400816102210223±18
TP2041200410816102210223±18
TP2045200450818102210223±18
TP2050200500820102210223±18
TP2060200600824102210223±18
TP212821028011102210223±18
TP213521035014102210223±18
TP214021040016102210223±18
TP214221042016¾102210223±18
TP222822028011102210223±18
TP223522035014102210223±18
TP224222042016¾102210223±18
TP25252502501010102210223±18
TP25302503001012102210223±18
TP25352503501014102210223±18
TP25382503801015102210223±18
TP25402504001016102210223±18
TP25452504501018102210223±18
TP25502505001020102210223±18
TP273527035010¾14102210223±18
TP274527045010¾18102210223±18
TP28352803501114102210223±18
TP28402804001116102210223±18
TP28452804501118102210223±18
TP28502805001120102210223±18
TP28572805701122102210223±18
TP30383003801215102210223±18
TP30403004001216102210223±18
TP30423004201216¾102210223±18
TP30453004501218102210223±18
TP30463004601218102210223±18
TP30503005001220102210223±18
TP30553005501222102210223±18
TP30603006001224102210223±18
TP324532045012¾18102210223±18
TP325032050012¾20102210223±18
TP325532055012¾22102210223±18
TP33463304601318102210223±18
TP35403504001416102210223±18
TP35453504501418102210223±18
TP35503505001420102210223±18
TP35523505201420%102210223±18
TP35553505501422102210223±18
TP35603506001424102210223±18

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Table 3. Product List of "PMSSteripack Tyvek Sterilization Rolls (TY) with Chemical Indicator"

PRODUCT LIST
(A)(B)Total (A)(B) Total(L) Seal(V) Single(P) Distance(T)
ProductTotalLengthTotalLengthWidthSealBetweenThickness
NameWidth(m)Width(ft)(mm)Width2 Seals(μm)
(cm)(in)(mm)
TY 05235232751022223±18
TY 75237.5233751022223±18
TY 102310234751022223±18
TY 1252312.5235751022223±18
TY 152315236751022223±18
TY 202320238751022223±18
TY 232323239751022223±18
TY 2523252310751022223±18
TY 3023302312751022223±18
TY 3523352314751022223±18
TY 3823382315751022223±18
TY 4023402316751022223±18
TY 4523452318751022223±18
TY 5023502320751022223±18
TY 053053021001022223±18
TY 75307.53031001022223±18
TY 1030103041001022223±18
TY 1253012.53051001022223±18
TY 1530153061001022223±18
TY 2030203081001022223±18
TY 2330233091001022223±18
TY 25302530101001022223±18
TY 30303030121001022223±18
TY 35303530141001022223±18
TY 38303830151001022223±18
TY 40304030161001022223±18
TY 45304530181001022223±18
TY 50305030201001022223±18
TY 057057022301022223±18
TY 75707.57032301022223±18
TY 1070107042301022223±18
TY 1270127052301022223±18
TY 1570157062301022223±18
TY 2070207082301022223±18
TY 25702570102301022223±18
TY 30703070122301022223±18
TY 35703570142301022223±18
TY 40704070162301022223±18
TY 45704570182301022223±18
TY 50705070202301022223±18
TY 05100510023301022223±18
TY 751007.510033301022223±18
TY 101001010043301022223±18
TY 12510012.510053301022223±18
TY 151001510063301022223±18
TY 201002010083301022223±18
TY 231002310093301022223±18
TY 2510025100103301022223±18
TY 3010030100123301022223±18
TY 3510035100143301022223±18
TY 3810038100153301022223±18
TY 4010040100163301022223±18
TY 4510045100183301022223±18
TY 5010050100203301022223±18

{11}------------------------------------------------

{12}------------------------------------------------

5.1. Representative Engineering Drawing of "PMSSteripack Tyvek Sterilization Pouch (TP) with Chemical Indicator"

Image /page/12/Figure/1 description: The image shows a technical drawing of a part with several dimensions labeled. The dimension 'B' is shown vertically on the right side of the part. At the bottom, 'A' indicates the width, and there are also dimensions labeled 'P=2', 'V=2', and 'L=10' indicating other measurements of the part.

  • A: Total width
  • B: Total height
  • L: Seal width
  • V: Single seal width
  • P: Distance between 2 seals
  • S: diameter

{13}------------------------------------------------

5.2. Representative Engineering Drawing of "PMSSteripack Tyvek Sterilization Roll (TY) with Chemical Indicator"

Image /page/13/Figure/1 description: The image shows a rectangular diagram with dimensions labeled. The length 'B' is along the horizontal axis, while the height 'A' is along the vertical axis. There are also labels indicating 'V=2', 'P=2', and 'L=10', which likely represent specific measurements or parameters within the diagram.

  • A: Total width
  • B: Total length
  • L: Seal width
  • V: Single seal width
  • P: Distance between 2 seals

6. Description of the Principle of Operation

PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method.

Medical device to be sterilized is put into pouch or roll and the open parts of the packages are closed by the end user with heat sealing. Sterilization Pouches then are subjected to sterilization operation in recommended (validated) VHP (Vaporized Hydrogen Peroxide) sterilization devices and cycle as V-PRO 1 PLUS, Lumen Cycle.

Sterilant penetration occurs through the tyvek material into the packages and microorganisms on the surfaces of the enclosed medical devices are destroyed via of the sterilant.

The process indicator on "PMSSteripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator" is intended to be used by a health care provider to distinquish between processed and unprocessed units by changing color. Chemical process indicators

{14}------------------------------------------------

on the pouch and/or roll indicate that the pouch or roll has been exposed to sterilization process by changing color.

Chemical process indicators are printed on the pouch and roll exterior (printed on tyvek material component) change color when exposed to sterilant vapor during processing. After the sterilization is completed, the sterility of the enclosed medical device is maintained for 30 days. Chemical indicator bars printed on to the medical grade paper component of the pouches are 100 mm² size, having dimensions of 5 mmx20 mm.

The indicators used on the pouches are classified as Class 1 type process indicator according to the ISO 11140-1 standard.

7. Comparison of Submission Device and Predicate Device and Substantial Equivalence

Technological characteristics of the submission device and the comparison with predicate device are given in Table 7.1 as a summary. Performance comparison of submission device and predicate device is presented in Table 7.2.

DEVICE NAMESUBMISSION DEVICEPREDICATEDEVICESUBSTANTIALLYEQUIVALENCE
CHARACTERISTICSPMSSteripack Tyvek SterilizationPouch (TP) and Roll (TY) withChemical IndicatorSTERIS Vis-U-AllLow TemperatureSterilizationPouch/Tubing(K140487)YES
DESIGN ANDCONSTRUCTION&MATERIALCOMPOSITIONTyvek material and heat sealedlaminated PET/PE plastic filmExternally printed H2O2 chemicaprocess indicator ink. Pouches aremanufactured from tyvek material/plastic film that are heat sealed on 3sides. The remaining side is lefopened and heat sealed by the enduser. Rolls are manufactured fromtyvek material/plastic film that are heasealed on opposite two sides. Therequired size is cut from the roll andthe remaining open two sides are heasealed by the end user.Pouches areconstructed fronTyvek and plasticfilms, with theEthylene OxideProcess ChemicaIndicator Printed onboth sides of TyvekYES
INDICATIONS FORUSEPMS Steripack Tyvek SterilizationPouches (TP) and Rolls (TY) withChemical Indicator; are intended toprovide healthcare workers with aneffective method to enclose medicaldevices intended for sterilization invaporized hydrogen peroxide (VHP)sterilization method.PMS Steripack Tyvek SterilizationPouches (TP) and Rolls (TY) withChemical Indicator is a device intendedto be used to enclose another medicaldevice that is to be sterilized by ahealthcare provider. It is intended toallow sterilization of the enclosedmedical devices and also maintainsterility of the enclosed devices untilused.The Vis-U-All LowTemperatureSterilizationPouches/Tubing isa sterilizationcontainment pouchfor use by healthcare providers toenclose medicaldevices to besterilized in Lumenand Non LumenCycles in V-PRO60 LowTemperatureSterilizationSystem.The Pouchmaintains thesterility of theenclosed devicesuntil used.YES

T ABLE-7.1: Comparison of Submission Device and Predicate Device (Characteristics)

{15}------------------------------------------------

The recommended (and validated)sterilization cycle parameters are:The Steris V-PRO 1 PLUS LUMENCYCLE is a single preprogramed standardsterilization cycle with NO cycle parametervariables allowed by the end user.Chemical process indicator on the exteriorof the Tyvek pouches and rolls indicatesthat the pouch has been exposed to V-Pouchesarintended to containdevices in suchmanner as to leavea minimum of oneinch betweenthedevices and seal orall sides.Deviceswithstainlesssteellumensand the
PRO 1 Plus, Lumen Cycle sterilizationprocess by changing color from red to blueor (lighter).Devices with stainless steel lumens withthe following internal diameter (ID) andmaximum length can be processed inV-PRO 1 Plus, Lumen Cycle.following minimuminternaldiameter(ID) and maximumlengthcanbeprocessed in Vis-U-AllLowTemperatureSterilization Pouch/Tubing using the
Worst Case Load:The worst case load for the 50 mm x 200mm pouch is 0.058 lbs. of instruments.lumen cycle.Single or duallumen devices
The worst case load for the 150 mm x 300mm pouch is 0.630 lbs. of instruments.0.77 mmID and 410 mmin length
The worst case load for the 350 mm x 600mm pouch is 3.054 lbs. of instruments.Triple LumenDevices1.2 mmID and 275 mm
Table 1. Lumen Claimsin length1.8 mmID and 310 mmin lengthor2.8 mmID and 317 mmin length
PouchSizeMax.lumenlengthMin.internaldiameterMax #Lumens
50 x200mm50mm1.5 mm1
150 x300mm3 mm200 mm1
350 x600mm3 mm400 mm1

{16}------------------------------------------------

STERILIZATIONPROPERTIESV-PRO 1 Plus, Lumen CycleV-PRO 60 LowTemperatureSterilization SystemLumen and NonLumen CyclesYES
PRINCIPLE OFOPERATIONSPMS Steripack Tyvek SterilizationPouches (TP) and Rolls (TY) withChemical Indicator; are intended toprovide healthcare workers with aneffective method to enclose medicaldevices intended for sterilization invaporized hydrogen peroxide (VHP)sterilization method.Medical device to be sterilized is putinto pouch or roll and the open parts ofthe packages are closed by the enduser with heat sealing. SterilizationPouches then are subjected tosterilization operation in recommended(validated) VHP (Vaporized HydrogerPeroxide) sterilization devices and thecycle as V-PRO 1 PLUS, LumenCycle.Sterilant penetration occurs throughthe tyvek material into the packagesand microorganisms on the surfaces ofthe enclosed medical devices aredestroyed via of the sterilant.The process indicator or a"PMSSteripack Tyvek SterilizationPouches (TP) and Rolls (TY) withChemical Indicator" is intended to beused by a health care provider todistinguish between processed andunprocessed units by changing color.Chemical process indicators on thepouch and/or roll indicate that thepouch or roll has been exposed tosterilization process by changing color.Chemical process indicators areprinted on the pouch and roll exterior(printed on tyvek material component)change color when exposed tosterilant vapor during processing. Afterthe sterilization is completed, thesterility of the enclosed medical deviceis maintained for 30 days. Chemicalindicator bars printed on to the medicalgrade paper component of thepouches are 100 mm² size, havingdimensions of 5 mmx20 mm.The indicators used on the pouchesare classified as Class 1 type processindicator according to the ISO 11140-1standard.The Vis-U-All LowTemperatureSterilizationPouches/Tubing isa sterilizationcontainment pouchfor use by healthcare providers toenclose medicaldevices to besterilized in Lumenand Non LumenCycles in V-PRO 60Low TemperatureSterilization System.The Pouchmaintains thesterility of theenclosed devicesuntil used.Pouches areintended to containdevices in such amanner as to leavea minimum of oneinch between thedevices and seal orall sides.Devices withstainless steelfollowing minimuminternal diameter(ID) and maximumlength can beprocessed in Vis-U-AllLow TemperatureSterilization Pouch/Tubing using thelumen cycle.Single or duallumen devices 0.77 mm IDand 410 mm inlength Triple LumenDevices 1.2 mm IDand 275 mm inlength 1.8 mm IDand 310 mm inlengthor 2.8 mm IDand 317 mm inlengthYES

{17}------------------------------------------------

PRINCIPLES OFOPERATION FORCHEMICALINDICATORSThe Process Indicators Ink printed onthe tyvek material will exhibit a colorchange after the pouch is exposed toHydrogen Peroxide (H2O2) gas.Hydrogen Peroxide (H2O2) sterilizationprinted indicator bar changes from redto blue (or lighter) when exposed tohydrogen peroxide vapor.The ProcessIndicator Ink printedon the tyvek willexhibit a colorchange after thepouch is exposedto sterilization.YES
SHELF LIFE(Prior to Sterilization)3 years2 yearsYES
SHELF LIFE(Post Sterilization)30 days1 yearYES
CONFIGURATIONS&DIMENSIONSVarious sizes (width and height/length)Various sizes(width andheight/length)YES

TABLE-7.2: Comparison Nonclinical Tests of Submission Device and Predicate Device (Performance)

DEVICE NAMEPROPOSED DEVICEPREDICATE DEVICESUBSTANTIALLYEQUIVALENCE
PERFROMANCE"PMSSteripack TyvekSterilization Pouch (TP) andRoll (TY) with ChemicalIndicator"STERIS Vis-U-All LowTemperature SterilizationPouch/Tubing (K140487)YES
Sterilant PenetrationThe sterilization (VHP) wasvalidated to a sterility assurancelevel (SAL) of 10-6Meets requirementsYES
Microbial BarrierPropertiesSterility was maintained for atleast 30 days after processing inV-PRO 1 Plus, Lumen CycleMeets requirementsYES
MaterialCompatibilitySuitability for use in V-PRO 1Plus, Lumen Cycle sterilizationprocess and cycle parameters.Meets requirementsYES
BiocompatibilityNot direct patient-contactingdevices; Materials are non-toxicand meet ISO 10993-1requirements.Meets requirementsYES
Shelf Life (poststerilization)Sterility was maintained for atleast 30 days after processing inV-PRO 1 Plus, Lumen CycleMeets requirementsYES
Package IntegrityPorous material providing amicrobial barrier.Meets requirementsYES

8. Conclusion

"PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator" and "Vis-U-All Low Temperature Sterilization Pouch/Tubing" (submission device) (predicate device) are both single use devices that are used to enclose another medical device to be sterilized in recommended (validated) sterilization cycles.

PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator and predicate device have many similar technological characteristics. Vis-U-All Low Temperature

{18}------------------------------------------------

Sterilization Pouch/Tubing are made from tyvek material and plastic film by heat sealing. On the other hand, PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator and Vis-U-All Low Temperature Sterilization Pouch/Tubing have similar design features and they all have various sizes.

PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator and predicate device have similar sterilization methods (VH2O2). Submission device sterilization system is; VPRO 1 PLUS, Lumen Cycle; Predicate device sterilization system is V-PRO 60 Low Temperature Sterilization System, Lumen and Non Lumen Cycles.

PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator and device have identical performance characteristics considering predicate Sterilant Penetration, Package Integity, Sterility Maintenance, Biocompatibility and Chemical Indicator Efficiacy.

Both the subject device "PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator" and the predicate device "Vis-U-All Low Temperature Sterilization Pouch/Tubing" meet the requirements of ANSI/ AAMI/ ISO 11140-1:2005 at 50 ℃ only and 11607-1:2006/(R)2010.

As indicated in Table 7.2. based on the results of non-clinical performance tests summaries of the subject and predicate device; any differences between subject and predicate device do not impact safety and effectiveness.

In conclusion, the subject device "PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator" is substantially equivalent to predicate device K140487 STERIS Vis-U-All Low Temperature Sterilization Pouch/Tubing. Based on the intended use, technological characteristics, and performance data, the subject PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator, is substantially equivalent and is as safe and as effective as the legally marketed predicate device.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).