K Number
K160593

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-09-02

(185 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Indy OTW™ Vascular Retriever is intended to snare a foreign body and withdraw it to a peripheral vascular location.

Device Description

The Indy OTW™ Vascular Retriever consists of a 4-loop nitinol wire snare attached to an inner catheter (which is attached proximally to a metal cannula) and includes an outer Flexor sheath. The device is compatible with a 0.035 inch wire guide for over-the-wire introduction. The Flexor sheath has a radiopaque band incorporated within the sheath material to identify the location of the sheath's distal tip.

AI/ML Overview

The provided text describes the Cook Incorporated Indy OTW™ Vascular Retriever, a medical device for snaring and retrieving foreign bodies in the vascular system. However, it does not contain the specific information requested about the acceptance criteria or a study proving the device meets those criteria, particularly in the context of an AI/human-in-the-loop system.

The document is a 510(k) summary for a medical device cleared by the FDA. It focuses on demonstrating substantial equivalence to a predicate device through manufacturing and performance testing, not on the performance of a software algorithm or AI.

Therefore, I cannot fulfill your request for the specific points listed (AI performance, sample sizes for test/training sets, ground truth establishment, MRMC studies, standalone performance) based on the provided text.

The closest information related to "acceptance criteria" and "device performance" in this document is in Section VII, "PERFORMANCE DATA". It lists various types of engineering and materials testing performed, but these are for the physical device itself, not an AI component.

Here's what can be extracted from the provided text regarding "acceptance criteria" and "device performance" for the physical device, recognizing it does not apply to an AI system:

1. Table of Acceptance Criteria and Reported Device Performance (as much as can be inferred from the text):

Test TypeAcceptance Criteria (Inferred from text)Reported Device Performance (Summary from text)
Biocompatibility TestingAcceptable and demonstrate the device is biocompatible.All test results were acceptable.
Tensile Strength TestingPeak load values of joints between device components in accordance with predetermined acceptance criteria.Peak load values of joints between device components were in accordance with the predetermined acceptance criteria.
Simulated Use TestingCapable of snaring and removing wire guides and catheter fragments or snaring and repositioning wire guides and catheters within an anatomical model in accordance with the instructions for use.The predetermined acceptance criteria were met.
Liquid Leak Pressure TestingSheath hub, Touhy-Borst, and catheter hub did not leak water while pressurized.The predetermined acceptance criterion was met.
Structural Integrity TestingMaintained structural integrity under the conditions of the study.The predetermined acceptance criteria were met.
Radiopacity TestingCatheter tip, sheath tip, and wire loops are visible under standard fluoroscopy.The predetermined acceptance criterion was met.
Immersion Corrosion TestingNo evidence of corrosion was observed.No evidence of corrosion was observed. The predetermined acceptance criterion was met.

Regarding the other requested information (which is specific to AI/software performance):

  • 2. Sample sized used for the test set and the data provenance: Not applicable to the provided document, as it describes a physical medical device, not an AI system.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
  • 8. The sample size for the training set: Not applicable.
  • 9. How the ground truth for the training set was established: Not applicable.

The document details the evaluation of a physical vascular retriever based on mechanical, material, and simulated use tests designed to establish its safety and effectiveness and substantial equivalence to a predicate device. It does not involve any artificial intelligence or machine learning components.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged to suggest a sense of movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 2, 2016

Cook Incorporated Mr. Chad Schulenburg Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47404

Re: K160593

Trade/Device Name: Indy OTW Vascular Retriever Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: July 29, 2016 Received: August 1, 2016

Dear Mr. Schulenburg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160593

Device Name Indy OTW™ Vascular Retriever

Indications for Use (Describe)

The Indy OTW Vascular Retriever is intended to snare a foreign body and withdraw it to a peripheral vascular location.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The red background is a rectangle with a diagonal cut on the bottom right corner.

510(k) SUMMARY As required by 21 CFR §807.92 Date Prepared: July 29, 2016

I. SUBMITTER

Submission:Traditional 510(k) Premarket Notification
Applicant:Cook Incorporated
Contact:Chad Schulenburg
Applicant Address:Cook Incorporated
750 Daniels Way
Bloomington, IN 47404
Contact Phone Number:(812) 335-3575 ext. 104073
Contact Fax Number:(812) 332-0281

II. DEVICE

Trade Name: Common Name: Classification Name: Regulation/Class: Product Code

Indy OTWTM Vascular Retriever Percutaneous Retrieval Device Embolectomy Catheter 21 CFR §870.5150/Class II MMX

III. PREDICATE DEVICE

The device subject of this submission is considered substantially equivalent to the predicate device, the 4-Loop Vascular Retriever Snare (K112185).

IV. DEVICE DESCRIPTION

The Indy OTW™ Vascular Retriever consists of a 4-loop nitinol wire snare attached to an inner catheter (which is attached proximally to a metal cannula) and includes an outer Flexor sheath. The device is compatible with a 0.035 inch wire guide for over-the-wire introduction. The Flexor sheath has a radiopaque band incorporated within the sheath material to identify the location of the sheath's distal tip.

V. INDICATIONS FOR USE

The Indy OTWIM Vascular Retriever is intended to snare a foreign body and withdraw it to a peripheral vascular location.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol to the right of the word. Below the word "COOK" is the word "MEDICAL" in white. The background is red.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Indy OTWTM Vascular Retriever and the predicate device, the 4-Loop Vascular Retriever Snare (K112185), are substantially equivalent in that these devices have the same intended use and similar technological characteristics.

PREDICATE DEVICESUBJECT DEVICE
4-Loop Vascular Retriever Snare(K112185)Indy OTW Vascular Retriever(Subject of this Submission)
Regulation Number870.5150Identical
Product CodeMMXIdentical
Classification NameEmbolectomy CatheterIdentical
ClassIIIdentical
Intended UseIntended for use in thecardiovascular system to manipulateand retrieve foreign objects,including, but not limited to, wireguides, coils, balloons, catheters,and filters.Intended to snare a foreign body andwithdraw it to a peripheral vascularlocation.
Model NumberPrefixVRSINDY
Snare Sub-Assembly5 Nitinol strands /2 Tantalum strands5 Nitinol strands /2 Platinum strands
Snare Loops4Identical
Inner CatheterRadiopaqueTetrafluoroethylene (TFE)Nylon radiopaque tubing
CatheterLength (cm)9055, 100
SheathNylonFlexor sheath:Nylon outer layer,Embedded steel coil,PTFE inner liner
Outer SheathSize (Fr)10.08.0
Connector CapNylonIdentical

Table 1: Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif, uppercase letters on a darker red background. The logo is simple and modern.

PREDICATE DEVICESUBJECT DEVICE
4-Loop Vascular Retriever Snare(K112185)Indy OTW Vascular Retriever(Subject of this Submission)
Side Arm AssemblyTuohy-Borst adapter material:Polycarbonate
Male luer cap with strap material:Identical
High density polyethylene/polyvinylchloride
Wire GuideCompatibility(diameter in inches)N/A0.035
PackagingPETG trayPET-PE / Tyvek pouchWire guide holderPET-PE / Tyvek pouch
Sterilization MethodEtOIdentical
SAL10-6Identical
Shelf Life3 yearsIdentical

The minor differences in dimensions, materials, and technology as compared to the predicate device were appropriately assessed and do not present any new questions of safety and/or effectiveness. The following devices were referenced to provide additional support: Flexor® Introducers (K142829), SeQure™ Snare System (K102484), Roadrunner® Uniglide® Hydrophilic Wire Guide (K110009), Approach® Pro ST Micro Wire Guides (K110163) and the Coda® Low Profile Balloon Catheter (K150970). Based on the comparison of the design, intended use, materials, fundamental technology, and principle of operation, the subject device is considered to be substantially equivalent to the currently marketed predicate device.

VII. PERFORMANCE DATA

Biocompatibility testing and performance testing were conducted to ensure that the Indy OTWTM Vascular Retriever will not elicit an adverse biological response in patients and will reliably perform as intended.

  • . Biocompatibility Testing - Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systematic Toxicity, Hemocompatibility, and Pyrogenicity tests were performed. All test results were acceptable and demonstrate that the device is biocompatible.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Cook Medical logo. The logo is white text on a red background. The word "COOK" is in a large, bold font, and the word "MEDICAL" is in a smaller font below it.

  • 트 Tensile Strength Testing - Peak load values of joints between device components were in accordance with the predetermined acceptance criteria demonstrating that the joints have sufficient strength to remain intact during use.
  • Simulated Use Testing The Indy OTW™ Vascular Retriever was capable of snaring and removing wire guides and catheter fragments or snaring and repositioning wire guides and catheters within an anatomical model in accordance with the instructions for use. The predetermined acceptance criteria were met.
  • . Liquid Leak Pressure Testing –The sheath hub, Touhy-Borst, and catheter hub did not leak water while pressurized. The predetermined acceptance criterion was met.
  • Structural Integrity Testing –The device maintained structural integrity under the conditions of the study. The predetermined acceptance criteria were met.
  • 그 Radiopacity Testing - The catheter tip, sheath tip, and wire loops are visible under standard fluoroscopy. The predetermined acceptance criterion was met.
  • Immersion Corrosion Testing – No evidence of corrosion was observed. The predetermined acceptance criterion was met.

VIII. CONCLUSIONS

The results of the testing provide reasonable assurance that the subject device has been designed such that it will function as intended to snare a foreign body and withdraw it to a peripheral vascular location. The subject device does not raise new questions of safety or effectiveness as compared to the predicate device. This supports a determination of substantial equivalence of the subject device to the predicate device.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).