K Number
K160566

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-12-02

(277 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cerasorb Ortho Foam is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cerasorb Ortho Foam resorbs and is replaced with bone during the healing process.

Device Description

Cerasorb Ortho Foam is a highly porous composite material for bone augmentation and regeneration applications, consisting of porcine collagen and a resorbable ceramic. Through the use of pure phase ß-tricalcium phosphate with a regular interconnected porous structure, a degradation of the biomaterial is achieved simultaneously to natural bone regeneration. The pure phase B-tricalcium phosphate is made according to ASTM F 1088-04a. The ß-tricalcium phosphate component of Cerasorb Ortho Foam is a mixture of Cerasorb M Ortho and Cerasorb Ortho, which received previous 510(k) clearances (K040216 and K014156, respectively). The device has been tested to meet pyrogen limit specifications. Cerasorb Ortho Foam is available in five different sizes and two versions (mouldable foam and flexible foam strip) and serves to fill bone defects in the entire skeletal system.

AI/ML Overview

Here's an analysis based on the provided document regarding the acceptance criteria and study for the Cerasorb Ortho Foam device:

This document is a 510(k) premarket notification for a medical device called Cerasorb Ortho Foam. The purpose of a 510(k) submission is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device, rather than proving its absolute safety and effectiveness through extensive clinical trials as would be required for a PMA (Premarket Approval). Therefore, the "acceptance criteria" and "study" described here are focused on demonstrating substantial equivalence to a predicate, not necessarily meeting a specific performance metric in a standalone sense.

1. Table of Acceptance Criteria and Reported Device Performance:

Feature/CriteriaPredicate Device (Vitoss Scaffold Foam Bone Graft Material) PerformanceProposed Device (Cerasorb Ortho Foam) Performance (Claimed)Rationale for Equivalence
Intended UseBone void filler for skeletal system (extremities, spine, pelvis)Bone void filler for skeletal system (extremities, pelvis)Substantially equivalent (minor difference in explicitly listed anatomical locations, but overall 'bone void filler' function is same).
Target PopulationIndividual with bony defect resulting from surgery or traumaIndividual with bony defect resulting from surgery or traumaIdentical
Anatomical LocationsBony voids/gaps of extremities, spine, and pelvisBony voids/gaps of extremities and pelvisSubstantially equivalent (exclusion of 'spine' for proposed device not deemed significant, implies similar general application).
LabelingContains comparable intended use, warnings, and precautionsContains comparable intended use, warnings, and precautionsSubstantially equivalent
Chemical CompositionCalcium salt with Type I bovine collagenCalcium salt with Type I and Type III porcine collagenEquivalent, with rationale for porcine collagen for safety.
Mineral Phasesß-Tricalcium phosphate Ca3(PO4)2ß-Tricalcium phosphate Ca3(PO4)2Identical
Physical StructureTrabecular structure similar to cancellous boneTrabecular structure similar to cancellous boneIdentical
Pore Size (range)1-1000 µm0.1-500 µmWithin same general range, with statement "No significant functional implications are expected."
OsteoconductivityOsteoconductiveOsteoconductiveIdentical
Mechanical StrengthDoes not impart mechanical strength to surgical siteDoes not impart mechanical strength to surgical siteIdentical
SterilitySterilized by gamma irradiation, single use onlySterilized by gamma irradiation, single use onlyIdentical
BiocompatibilityEstablishedEstablishedEstablished, implying performance is comparable.
Dosage Form(s)Cylinders (Blocks), Strips, Packs, Flow, ShapesStrips, Packs (Mouldable Foam and Flexible Foam Strip)Different specific forms, but both offer forms for filling voids.
New Bone Formation (Animal Study)Not explicitly quantified, but implied successful performancePerformed at least as well as the predicate deviceDemonstrated non-inferiority to predicate in animal model.
Fibrosis and Inflammation (Animal Study)Not explicitly quantified, but implied successful performanceOccurrence evaluatedImplied comparable or acceptable levels.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set (Animal Study): The document states "A comparative study in sheep using the critical size defect model was performed." It does not specify the exact number of sheep used (sample size) or the number of defects created/treated.
  • Data Provenance: The study was "in sheep," which indicates animal data. The location of the study (country of origin) is not explicitly stated in this document but implied to be part of the manufacturer's (curasan AG, Germany) activities. It is a prospective animal study designed for comparison.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • The document does not mention the number of experts used or their qualifications for establishing ground truth in the animal study. The evaluation focused on "amount and quality of newly formed bone" and "occurrence of fibrosis and inflammation," which would typically involve histological analysis and potentially expert pathology assessment, but specifics are omitted.

4. Adjudication Method for the Test Set:

  • The document does not provide details on an adjudication method. It simply states that bone formation, fibrosis, and inflammation were "evaluated." This suggests standard scientific evaluation by the researchers involved in the study, but no formal adjudication process with independent experts is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, an MRMC comparative effectiveness study was not done. This type of study typically involves human readers (e.g., radiologists) evaluating medical images or cases with and without AI assistance to measure improvement in diagnostic performance. The Cerasorb Ortho Foam is a physical bone void filler, not an AI software device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This question is not applicable. Cerasorb Ortho Foam is a physical medical device (bone void filler), not an algorithm or AI system.

7. The Type of Ground Truth Used:

  • For the animal study (test set): The ground truth was established through histological evaluation of the amount and quality of newly formed bone, and the occurrence of fibrosis and inflammation at the surgical site in the sheep. This is a form of direct biological/pathological assessment.

8. The Sample Size for the Training Set:

  • This question is not applicable. Cerasorb Ortho Foam is a physical medical device. It does not involve a "training set" in the context of machine learning or AI. The product's development would have involved various material science and biological compatibility tests, but not a dataset for training an algorithm.

9. How the Ground Truth for the Training Set Was Established:

  • This question is not applicable, as there is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 2, 2016

Curasan AG % Dr. Karin Lübbers Managing Director YES Medical Device Services GmbH Bahnstrasse 42-46 Friedrichsdorf, Hesse 61381 Germany

Re: K160566

Trade/Device Name: Cerasorb Ortho Foam Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: October 28, 2016 Received: October 31, 2016

Dear Dr. Lübbers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K160566

Device Name Cerasorb Ortho Foam

Indications for Use (Describe)

Cerasorb Ortho Foam is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cerasorb Ortho Foam resorbs and is replaced with bone during process.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Yes" in white letters on a gray background. The "Y" has a green triangle above it. The background is a gray rectangle with rounded corners.

510(k) Summary

1. Submitter

This 510(k) application was created in January 2016

The 510(k) owner for this application is:

Companycurasan AG
AddressLindigstr. 463801 KleinostheimGermany
Phone:+49-6027-40900-0
Fax:+49-6027-40900-29
Email:info@curasan.de
Establishment registration number:3003771570

Contact person:

Name:Dr. Wolf-Dietrich Hübner
Position:Medical Director

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Yes" in white letters with a green triangle above the "Y". The word is set against a gray background that appears to have a glossy or reflective surface. The font is bold and sans-serif, giving the word a modern and clean look.

510(k) Application

Image /page/4/Picture/3 description: The image contains the word "curasan" in red, lowercase letters. The font is sans-serif and the letters are slightly rounded. The word is presented horizontally and appears to be a logo or brand name.

2. Device Name and Classification

Proprietary / Trade Name: Common Name: Classification Name: Classification: Product Code Panel

Cerasorb Ortho Foam Bone Void Filler Filler, Bone Void, Calcium Compound Class II per 21 CFR § 888.3045 MQV Orthopedic

3. Predicate Device

This application claims that Cerasorb Ortho Foam is substantially equivalent to the following predicate device:

510(k) numberK 032288
Proprietary / Trade Name:Vitoss Scaffold Foam Bone Graft Material
Common Name:Bone Void Filler
Classification Name:Filler, Bone Void, Calcium Compound
Classification:Class II per 21 CFR § 888.3045
Product CodeMQV
PanelOrthopedic
ManufacturerOrthovita, Inc., 45 Great Valley Parkway, Malvern, PA 19355

4. Description

Cerasorb Ortho Foam is a highly porous composite material for bone augmentation and regeneration applications, consisting of porcine collagen and a resorbable ceramic. Through the use of pure phase ß-tricalcium phosphate with a regular interconnected porous structure, a degradation of the biomaterial is achieved simultaneously to natural bone regeneration. The pure phase B-tricalcium phosphate is made according to ASTM F 1088-04a. The ß-tricalcium phosphate component of Cerasorb Ortho Foam is a mixture of Cerasorb M Ortho and Cerasorb Ortho, which received previous 510(k) clearances (K040216 and K014156, respectively). The device has been tested to meet pyrogen limit specifications.

Cerasorb Ortho Foam is available in five different sizes and two versions (mouldable foam and flexible foam strip) and serves to fill bone defects in the entire skeletal system.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Yes" in white letters on a gray background. The "Y" has a green triangle above it. The letters are in a sans-serif font and are slightly italicized. The background is a gradient of gray, with the lighter shade at the top and the darker shade at the bottom.

Image /page/5/Picture/2 description: The image shows the word "curasan" in red, lowercase letters. The letters are slightly tilted to the right, giving the word a dynamic appearance. The font is bold and sans-serif, contributing to the modern and clean look of the logo. The background is plain white, which makes the red letters stand out.

Cerasorb Ortho Foam

Available sizes:
Product name
Cerasorb Ortho Mouldable Foam
Cerasorb Ortho Mouldable Foam
Cerasorb Ortho Mouldable Foam
Cerasorb Ortho Flexible Foam Strip
Cerasorb Ortho Flexible Foam Strip

Dimensions LxWxH [mm] 25x25x4 25x50x4 25x100x4 25x50x4 25x100x4

5. Intended Use

Cerasorb Ortho Foam is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cerasorb Ortho Foam resorbs and is replaced with bone during the healing process.

6. Substantial Equivalence

Cerasorb Ortho Foam is substantially equivalent to the predicate device Vitoss Scaffold Foam Bone Graft Material. Please refer to the table below that discusses the substantial equivalence of the two devices in detail.

Device Comparison Table
FeatureProposed device:Cerasorb Ortho FoamPredicate deviceVitoss Scaffold Foam Bone GraftMaterial
Intended useCerasorb Ortho Foam is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cerasorb Ortho Foam resorbs and is replaced with bone during the healing process.Vitoss Scaffold Foam Bone Graft Material is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss Scaffold Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss Scaffold Foam should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously. Vitoss Scaffold Foam Bone Graft Material is intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
Abbreviated intended useBone void fillerBone void filler

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Yes" in a stylized font. The "Y" is white and has a green triangle above it, while the "es" is white and slightly blurred. The background is a gradient of gray, with the darker shade at the bottom and the lighter shade at the top. The overall design is simple and modern.

Image /page/6/Picture/3 description: The image shows the word "curasan" in red, lowercase letters. The font is sans-serif and the letters are slightly rounded. The word is displayed horizontally and is the only element in the image.

Device Comparison Table
FeatureProposed device:Cerasorb Ortho FoamPredicate deviceVitoss Scaffold Foam Bone GraftMaterial
Target PopulationIndividual with bony defect resulting from surgery or traumaIndividual with bony defect resulting from surgery or trauma
Anatomical locationsBony voids or gaps of the skeletal system, i. e., the extremities and pelvisBony voids or gaps of the skeletal system, i. e., the extremities, spine and pelvis
LabelingContains comparable intended use, warnings, and precautions as the labeling of the Vitoss Scaffold Foam Bone Graft Material. The wording is not identical, but substantially equivalent.-
Materials
Chemical compositionCalcium salt with Type I and Type III porcine collagenCalcium salt with Type I bovine collagen
Mineral phases$ β $ -Tricalcium phosphate Ca3(PO4)2$ β $ -Tricalcium phosphate Ca3(PO4)2
Design
Physical structureTrabecular structure similar to cancellos boneTrabecular structure similar to cancellos bone
Pore size (range)0.1-500 $ μ $ m1-1000 $ μ $ m
Performance
OsteoconductivityOsteoconductiveOsteoconductive
Mechanical strengthDoes not impart mechanical strength to surgical siteDoes not impart mechanical strength to surgical site
SterilitySterilized by gamma irradiation, single use onlySterilized by gamma irradiation, single use only
BiocompatibilityEstablishedEstablished
Dosage Form(s)Strips, PacksCylinders (Blocks), Strips, Packs, Flow, Shapes

One difference between Cerasorb Ortho Foam and the predicate device is that porcine collagen type I and type III was used instead of bovine collagen. This change was done for safety reasons as for porcine collagen the risk of Transmissible Spongiform Encephalopathy transmission is reduced. Moreover, the pore size of the integrated ß-tricalcium phosphate particles is on the average smaller for Cerasorb Ortho Foam, but within the same range. No significant functional implications are expected.

Based on the identified characteristics it is claimed that the device is substantially equivalent to the predicate device.

7. Performance Testing

A comparative study in sheep using the critical size defect model was performed. Cerasorb Ortho Flexible Foam Strip and Cerasorb Ortho Mouldable Foam were tested and compared to the predicate device Vitoss Scaffold Foam Bone Graft Material. The amount and quality

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Yes" in white letters on a gray background. The "Y" has a green triangle above it. The letters are in a bold, sans-serif font. The background is a gradient, with the top being lighter than the bottom.

Cerasorb Ortho Foam

Image /page/7/Picture/3 description: The image shows the word "curasan" in red, lowercase letters. The letters are slightly rounded and have a uniform thickness. The word is presented in a simple, sans-serif font. The background is plain white, which makes the red letters stand out.

of the newly formed bone was evaluated as well as the occurrence of fibrosis and inflammation at the surgical site. It was concluded that Cerasorb Ortho Flexible Foam Strip and Cerasorb Ortho Mouldable Foam performed at least as well as the predicate device Vitoss Scaffold Foam Bone Graft.

Based on the functional testing performed, substantial equivalence to the predicate device Vitoss Scaffold Foam Bone Graft Material is given.

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.