(122 days)
The BI IMAGE-X EVOLUTION device is an X-Ray generator designed for radiographic examination to assist with diagnosis of diseases of the teeth, jaw, and oral structure in combination with an intraoral image receptor.
The device is designed to be used by dental professionals for radiographic examinations to assist with diagnosis of diseases of the teeth, jaw, and oral structure by exposing an x-ray image receptor to ionizing radiation. The x-ray source, a tube, is located outside the mouth. All three conventional types of intraoral receptors can be used with this device: analog x-ray film, digital phosphorous plates, and digital x-ray sensors. The Bi Image-X Evolution is a wall mounted device or can also be a floor model on a stand.
This document is a 510(k) Pre-market Notification for the BI IMAGE-X EVOLUTION, an extraoral X-ray system. The submission aims to demonstrate substantial equivalence to a predicate device, the Dental X-Ray Z70 (K122582).
Here's an analysis of the provided information regarding acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly define "acceptance criteria" in the typical sense of quantitative performance thresholds for a diagnostic AI device (e.g., specific sensitivity, specificity, or AUC targets). Instead, it establishes substantial equivalence by comparing technological characteristics and adherence to safety standards.
Below is a table summarizing the technological comparison, which serves as the primary basis for demonstrating equivalence rather than meeting specific performance metrics for image analysis.
| Characteristic | Acceptance Criteria (Predicate Device K122582) | Reported Device Performance (Subject Device BI IMAGE-X EVOLUTION) |
|---|---|---|
| Intended Use | Extraoral X-ray system for dental radiographic examination and diagnosis of diseases of the teeth. | Extraoral X-ray generator that provides radiation for radiographic image formation for diagnosis. |
| Indications for Use | X-ray generator providing radiation for radiographic image formation for diagnosis. | Extraoral X-ray generator system for radiographic examinations to assist with diagnosis of diseases of the teeth, jaw, and oral structure in combination with an intraoral image receptor. |
| Electric Power Voltage | 120 VAC ± 10% or 230 VAC ± 10% | 108-132 VAC Monophase rectified |
| Rated Current | 9 A (120V) - 4A (230V) | 10A max (120V) |
| Type of Power Supply | 50/60 Hz | 50/60 Hz |
| kVp (kilovolt peak) | 70 kVp ± 10% | 70 kVp +/- 15% with 108-132 V |
| Exposure Time | 0.06 — 2.50 sec | 0.01-2.2 sec |
| Current to Tube | 7 mA ± 1.40 | 10 mA +/- 20% |
| Maximum Radiation Field | Ø 60 mm | Ø 60 mm |
| Focal Point | 0.8 mm x 0.8 mm (IEC 336) | 0.8mm (IEC366) |
| X-Ray Generator | AC - Alternate Current | AC - Alternate Current |
| Operator Exposure Control | Deadman Switch | Deadman Switch |
| Exposure Interval (Duty Cycle) | 1:60 | 1:32 |
| Minimum Distance SSD | 20 cm | 20 cm |
| Fuses | 10A 230V | 10A 250V |
| Operating Temperature Range | -20°C - + 55°C | -20°C - + 70°C |
| Fixed Arm | 27.56 inches (70 cm) | 27.56 inches (70 cm) |
| Standards of Conformity | EN 60601-1, EN 60601-2-28, EN 60601-2-54, EN 60601-1-3, EN 60601-1-2 | IEC 60601-1, IEC 60601-6, IEC 60601-3, IEC 60601-2-28, EN 60601-1-2 |
2. Sample Size Used for the Test Set and Data Provenance
This document describes an X-ray generator, not an AI/ML-driven diagnostic device that relies on a "test set" in the context of image analysis. Therefore, there is no mention of a test set sample size or data provenance related to image data. The "testing data reports" mentioned are for electrical, mechanical, and safety performance of the hardware.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. As this is not an AI/ML diagnostic device analyzing medical images, there's no "ground truth" established by experts in this context. The evaluation is based on technical specifications and compliance with standards.
4. Adjudication Method for the Test Set
Not applicable. There is no test set for image analysis requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. This document does not describe an MRMC study. The device is an X-ray generator aiming for substantial equivalence based on hardware and safety parameters, not on human reader performance with or without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
No. This is not an algorithm-only (AI) device. It is an X-ray generator.
7. The Type of Ground Truth Used
The "ground truth" in this context is adherence to recognized electrical and safety standards (IEC 60601 series, EN 60601 series, and 21CFR 1020.030, 1020.31). The testing involved physical measurements and evaluations against these objective standards by an independent laboratory.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/ML model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. No training set is involved.
Summary of the Study and Evidence Presented:
The study presented to prove the device meets acceptance criteria is primarily a technical comparison and compliance testing to demonstrate substantial equivalence to the predicate device (Dental X-Ray Z70, K122582) and adherence to relevant safety and performance standards.
-
Evidence Type:
- Technological Comparison: A detailed table (Table 5) comparing the subject device's specifications (e.g., kVp, mA, exposure time, focal point, power requirements) against those of the predicate device. The conclusion drawn is that the devices have "substantially similar technology characteristics."
- Performance Data: The submission includes "testing data reports" from an independent laboratory (Met Laboratories, Inc.) for:
- EMC and Electrical Safety: Conformance to IEC 60601-1 3rd Edition (including IEC 60601-1-6, 60601-1-3, 60601-2-28) and EN 650601-1-2.
- Radiation Performance Standards: Conformance to 21CFR 1020.030 and 1020.31 (presumably related to radiation output and limits for diagnostic X-ray systems).
- Software Verification and Validation: Performed by Radiotech, the manufacturer, and included in the submission. This indicates testing of the device's control software to ensure it functions as intended and safely.
- Risk Analysis: A risk assessment was performed, including recommendations for radiation exposure mitigation for users.
-
Conclusion of the Study: "The subject and the predicate device have the same intended use and the same technological features. The BI IMAGE – X and the Dental X-Ray Z70 share the same principles of operation and use similar imaging firmware. The conclusion is that the subject device is as safe and effective as the predicate."
In essence, the "acceptance criteria" are implicitly met by demonstrating that the device is technically comparable to a legally marketed predicate and complies with established international and national safety and performance standards for X-ray equipment. No clinical studies or AI performance metrics are
mentioned because the device itself is an X-ray generator, not an imaging analysis tool.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2016
Radiotecnologia Industrial S.A. De C.V. % Claude Berthoin, President Denterprise International, Inc./ 510k FDA Consulting 100 East Granada Blvd Suite 219 Ormond Beach, FL 32176
Re: K160458
Trade/Device Name: BI IMAGE-X EVOLUTION Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: EHD Dated: May 6, 2016 Received: May 10, 2016
Dear Claude Berthoin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oaks
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K160458
Device Name
BI IMAGE-X EVOLUTION
Indications for Use (Describe)
The BI IMAGE-X EVOLUTION device is an X-Ray generator designed for radiographic examination to assist with diagnosis of diseases of the teeth, jaw, and oral structure in combination with an intraoral image receptor.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a black and white drawing of an eagle. The eagle is facing to the right, and its wings are spread out. The eagle's head is turned to the side, and its beak is open. The eagle's feathers are detailed, and the overall image is well-drawn.
510k FDA Consulting
Medical Device Clearance
100 East Granada Blvd., Suite 219
Ormond Beach, FL 32176
386-506-8711
510(k) Summary
Submitter/Applicant
Radiotecnologia Industrial S.A. de C.V. Calle Puebla Sur, Manzana 4 Lote 5 Colonia Jardin Industria Ixtapaluca, Estado de Mexico, Mexico 56535
Contact: Marcelo Mojica, Engineering Manager Telephone: (01) 800-681-05-38 Date Prepared: February 15, 2016
Preparer/Consultant
Denterprise International, Inc. 100 East Granada Blvd., Suite 219 Ormond Beach, FL 32176
Phone: 386-506-8711 Fax: 855-235-7902
| Primary Contact: | Joyce St. Germain, Regulatory Executive |
|---|---|
| Joyce@510kFDA.com | |
| Secondary Contract: | Claude Berthoin, President |
| Claude@510kFDA.com |
Device Classification
| Trade Name: |
|---|
| Common Name: |
| Classification Name: |
| Regulation Number: |
| Product Code: |
| Regulatory Class: |
| 510k Review Panel: |
BI IMAGE-X EVOLUTION Extraoral Source X-ray System Unit, X-Ray, Extraoral with Timer 21 CFR 872.1800 EHD 2 Dental
{4}------------------------------------------------
Predicate Device
The subject device claims equivalence to the following legally marketed predicate:
| 510(k) Number: | K122582 |
|---|---|
| Date Cleared | December 7, 2012 |
| Trade Name: | Dental X-Ray Z70 |
| Common Name: | Extraoral Source X-ray System |
| Classification Name: | Unit, X-Ray, Extraoral with Timer |
| Regulation Number: | 21 CFR 872.1800 |
| Product Code: | EHD |
| Regulatory Class: | 2 |
| 510k Review Panel: | Dental |
Indication for Use
The BI IMAGE-X EVOLUTION is an extraoral x-ray generator system is designed to be used by dental professionals for radiographic examinations to assist with diagnosis of diseases of the teeth, jaw, and oral structure in combination with an intraoral image receptor.
Intended Use
The BI IMAGE -X EVOLUTION device is an extroral x-ray generator that provides the radiation necessary for the formation of a radiographic image from which a radiographic diagnosis may be made.
Device Description
The device is designed to be used by dental professionals for radiographic examinations to assist with diagnosis of diseases of the teeth, jaw, and oral structure by exposing an x-ray image receptor to ionizing radiation. The x-ray source, a tube, is located outside the mouth. All three conventional types of intraoral receptors can be used with this device: analog x-ray film, digital phosphorous plates, and digital x-ray sensors. The Bi Image-X Evolution is a wall mounted device or can also be a floor model on a stand. Choice of model is a matter of functional utility in the dental operatory and personal preference by the medical professional. The type of model used does not change the performance of the device.
Different models neither affect the device's indications for use nor raise differing issues of safety or effectiveness.
Comparison of Technological data on next page
{5}------------------------------------------------
Characteristics with Predicate
The following table compares technological and other characteristics of the subject and predicate device.
Table 5 -- Technological Comparison
| Subject Device | Predicate DeviceK122582 | |
|---|---|---|
| Device | BI IMAGE-X | Dental X-Ray Z70 |
| 510(k) Owner | RadiotecnologiaIndustrial S.A. de C.V.(Mexico) | XZeal Technologies, Inc.(Kissimmee, FL) |
| Classification & ProductCode | 872.72.1800; EHD | 872.1800; EHD |
| Device Description | Extroral x-ray systemused for dentalradiographicexaminations anddiagnosising. | Extroral x-ray systemused for dentalradiographicexaminations anddiagnosising. |
| Common Name | Extraoral Source X-raySystem | Extraoral Source X-raySystem |
| Models | Wall Mount, Mobileand Column Units | Wall Mount, Mobileand Column Units |
| Indication for Use | The BI IMAGE-XEVOLUTION is anextraoral x-raygenerator system isdesigned to be used bydental professionals forradiographicexaminations to assistwith diagnosis ofdiseases of the teeth,jaw, and oral structurein combination with anintraoral imagereceptor. | TheDental X-ray Z70device is an X-Raygenerator thatprovides the radiationnecessary for theformation of aradiographic imagefrom which aradiographic diagnosismay be made. |
| Intended Use | The BI IMAGE -XEVOLUTION device is anextraoral x-raygenerator that providesthe radiation necessaryfor the formation of aradiographic imagefrom which a | Extraoral source X-raysystem for dentalradiographicexamination anddiagnosis of diseasesof the teeth. |
| radiographic diagnosismay be made. | ||
| Electric Power Voltage | 108-132 VAC Monophaserectified | 120 VAC ± 10% or230 VAC ± 10% |
| Rated Current | 10A max (120V) | 9 A (120V) - 4A (230V) |
| Type of Power Supply | 50/60 Hz | 50/60 Hz |
| kVp (kilovolt peak) | 70 kVp +/- 15% with 108-132 V | 70 kVp ± 10% |
| Exposure Time | 0.01-2.2 sec | 0.06 — 2.50 sec |
| Current to Tube | 10 mA +/- 20% | 7 mA ± 1.40 |
| Maximum Radiation Field | Ø 60 mm | Ø 60 mm |
| Focal Point | 0.8mm (IEC366) | 0.8 mm x 0.8 mm(IEC 336) |
| X-Ray Generator | AC - Alternate Current | AC - Alternate Current |
| Operator Exposure Control | Deadman Switch | Deadman Switch |
| Exposure Interval(Duty Cycle) | 1:32 | 1:60 |
| Minimum Distancebetween the source andthe skin - SSD | 20 cm | 20 cm |
| Fuses | 10A 250V | 10A 230V |
| Operatiang TemperatureRange | -20°C - + 70°C | -20°C - + 55°C |
| Fixed Arm | 27.56 inches (70 cm) | 27.56 inches (70 cm) |
| Standards of Conformity | IEC 60601-1IEC 60601-6IEC 60601-3IEC 60601-2-28EN 60601-1-2 | EN 60601-1EN 60601-2-28EN 60601-2-54EN 60601-1-3EN 60601-1-2 |
{6}------------------------------------------------
The above comparison shows the subject and predicate devices have substantially similar technology characteristics.
{7}------------------------------------------------
Performance Data
The following performance data were provided in support of the substantial equivalence determination.
EMC and Electrical Safety – IEC 60601-1 3rd Edition (which include IEC test specifications: 60601-1-6, 60601-1-3, 60601-2-28) and EN 650601-1-2. The testing data reports for the subject device are provided in this petition. Tests were conducted by independent laboratory (Met Laboratories, Inc. 33439 Western Avenue, Union city, CA 94587.
This device conforms to all applicable performance standards in 21CFR 1020.030 and 1020.31.
Software - Software verification and validation testing and risk analysis assessment were performed by Radiotech and is a part of this submission.
Risk Analysis Information
Risk analysis includes particular recommendations to address radiation exposure to the user under mobile operating conditions. Operators should always read the user manuals for medical devices and take the necessary precautions before using the device. Methods to reduce exposure is recommended safety precautions such as wearing personnel monitoring and protective equipment.
Conclusion
The subject and the predicate device have the same intended use and the same technological features. The BI IMAGE – X and the Dental X-Ray Z70 share the same principles of operation and use similar imaging firmware. The conclusion is that the subject device is as safe and effective as the predicate.
The BI IMAGE - X warrants a finding of substantial equivalence to the legally marketed Dental X-Ray Z70 and thus clearance for premarket activities in the United States.
§ 872.1800 Extraoral source x-ray system.
(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.