K Number
K160390

Validate with FDA (Live)

Date Cleared
2016-04-11

(60 days)

Product Code
Regulation Number
870.1220
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APT Map-It™ Diagnostic Mapping Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

Device Description

The Map-iTTM Diagnostic Mapping Catheters are biocompatible, flexible, radiopaque fixed and steerable catheters that are available in a variety of diameters, lengths, curve shapes, and electrode number and spacing configurations, including a guidewire tip configuration to aid the physician in accessing and mapping a patient's heart. They are available in fixed curve and steerable configurations. The fixed curve catheters are available in 4/3.3F (4F tapering to 3.3F), 4F, 5F and 6F sizes with a 0.026'' floppy guidewire tip configuration in the 4/3.3F and 4F sizes for access into smaller anatomies. The steerable catheters are available in 4F, 5F, 6F and 7F sizes. There is a tip electrode on all models except the guidewire tip and PV loop configuration: the band electrodes are set at predetermined distances along the catheter shaft. The catheters can be used for recording intracardiac signals in the atrial and ventricular regions of the heart and for cardiac stimulation during electrophysiology studies. The Map-iT™ Diagnostic Mapping Catheters are comprised of a Pebax catheter shaft, platinum-iridium electrodes and electrical connectors that are either shrouded pin for the bipolar configurations or industry standard Redel connectors. The distal end of the fixed catheters is preformed into industry standard shapes with a variety of electrodes spacing to reach specific areas in the targeted chamber of the heart. The steerable catheters are available in a variety of curves that are controlled to the precise location in the heart by pulling or pushing the plunger on the handle.

AI/ML Overview

The provided document is an FDA 510(k) premarket notification for the Map-it™ Diagnostic Mapping Catheters. The document states that the device is "substantially equivalent" to a legally marketed predicate device (St. Jude Medical Livewire™ Diagnostic Mapping Catheter). This means the FDA determined that the new device is as safe and effective as the predicate device, and thus does not require a new premarket approval (PMA).

Crucially, the document does not contain acceptance criteria or a study that proves the device meets specific acceptance criteria in the way one might expect for a new, de novo device or a clinical AI/ML device. Instead, the document focuses on demonstrating substantial equivalence to a predicate device.

Here's an breakdown based on your request, highlighting what is and isn't present in the document:


1. Table of Acceptance Criteria and Reported Device Performance

This information is not present in the document as specific performance metrics and acceptance thresholds. Instead, the performance is implicitly demonstrated by meeting existing standards and test methods as discussed in the "TESTING PERFORMED" section.

The document states:
"The Map-iT™ Diagnostic Mapping Catheters were thoroughly tested based on existing standards and test methods. These tests included visual, dimensional, electrical performance - continuity, leakage impedance, dielectric; mechanical performance - tensile, torque, deflection, shaft buckling, radio-detectability, corrosion; biocompatibility (short term <24 hours, blood contacting) and packaging. Testing demonstrates that the Map-iT™ Diagnostic Mapping Catheters met the specifications and are substantially equivalent to the predicate."

Therefore, the "acceptance criteria" here are implied to be adherence to the specifications of these standards and methods, and the "reported device performance" is that it "met the specifications."

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The testing described is pre-clinical (e.g., electrical, mechanical, biocompatibility) rather than a clinical study with a "test set" in the context of AI/ML or a new clinical performance claim.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as there is no mention of a "test set" requiring expert ground truth in the context of clinical performance evaluation for this device. The testing conducted is primarily engineering and materials-based.

4. Adjudication Method for the Test Set

This information is not applicable for the same reasons as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

An MRMC study was not done and is not applicable for this type of device submission. This is a diagnostic mapping catheter, not an AI/ML-assisted diagnostic software requiring human reader performance improvement assessment.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

A standalone performance study of an algorithm was not done and is not applicable. This is a physical medical device.

7. The Type of Ground Truth Used

For the engineering and materials performance, the "ground truth" implicitly used would be "specifications" derived from recognized standards and predicate device performance. For biocompatibility, it would be based on laboratory test results showing compliance with established biological safety standards. There is no mention of pathology, outcomes data, or expert consensus in relation to a diagnostic accuracy study for this device.

8. The Sample Size for the Training Set

This information is not applicable. This is not an AI/ML device, so there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set was Established

This information is not applicable for the same reasons as point 8.


Summary of Device Acceptance and Study:

The acceptance of the Map-iT™ Diagnostic Mapping Catheters by the FDA is based on demonstrating substantial equivalence to an existing predicate device (St. Jude Medical Livewire™ Diagnostic Mapping Catheters). The company performed a series of pre-clinical tests (visual, dimensional, electrical, mechanical, radio-detectability, corrosion, and biocompatibility) that confirmed the device "met the specifications" derived from existing standards and the characteristics of the predicate device. The document explicitly states: "There are no new questions raised regarding safety or efficacy of the Map-iT™ Diagnostic Mapping Catheters." This approach is common for 510(k) submissions where a new device is very similar in design, materials, and intended use to a legally marketed device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird-like shape.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2016

Access Point Technologies Ep, Inc. % Laurie Lewandowski Consultant Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, Minnesota 55123

Re: K160390

Trade/Device Name: Map-it™ Diagnostic Mapping Catheters Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: February 10, 2016 Received: February 11, 2016

Dear Laurie Lewandowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mitchell Stein

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160390

Device Name Map-ItTM Diagnostic Mapping Catheters

Indications for Use (Describe)

The APT Map-It™ Diagnostic Mapping Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Access Point Technologies. The logo is blue and features the letters "AP" connected together. Below the letters is the text "Access Point Technologies" in a blue rounded rectangle.

Access Point Technologies 12560 Fletcher Lane Suite 300 Rogers, Minnesota 55374 USA Ph: 763-428-0010 Fax: 763-428-0012

==============================================================================================================================================================================

510(K) SUMMARY

Submitter:


Access Point Technologies EP (APT EP) dba APT Sean Carroll, President 12560 Fletcher Lane Suite 300 Rogers, MN 55374 763-428-0010 Telephone: Fax: 763-428-0012 Email: sean.carroll@apt-inc.us

Primary Contact:

Laurie Lewandowski
Consultant, Honkanen Consulting
738 Saddle Wood Drive
Eagan, MN 55123
Telephone:612-770-4038 (cell)
Fax:763-428-0012
Email:lalew207@gmail.com

DATE PREPARED:

April 9, 2016

NAME OF MEDICAL DEVICE:

Proprietary Name:Map-iT TM Diagnostic Mapping Catheters
Common/Usual Name:Diagnostic Electrophysiology Catheter
Classification Name:Electrode Recording Catheter

{4}------------------------------------------------

DEVICE CLASSIFICATION:

Classification Panel:Cardiovascular
Regulatory Class:II
Product Code:DRF
Regulation Number:21 CFR 870.1220

PREDICATE DEVICES:

Proprietary Name:LivewireTM Electrophysiology Catheter
Common/Usual Name:Diagnostic Mapping Catheter
Classification Name:Electrode Recording Catheter
510K Number:K022380

DEVICE DESCRIPTION:

The Map-iTTM Diagnostic Mapping Catheters are biocompatible, flexible, radiopaque fixed and steerable catheters that are available in a variety of diameters, lengths, curve shapes, and electrode number and spacing configurations, including a guidewire tip configuration to aid the physician in accessing and mapping a patient's heart. They are available in fixed curve and steerable configurations.

The fixed curve catheters are available in 4/3.3F (4F tapering to 3.3F), 4F, 5F and 6F sizes with a 0.026'' floppy guidewire tip configuration in the 4/3.3F and 4F sizes for access into smaller anatomies. The steerable catheters are available in 4F, 5F, 6F and 7F sizes. There is a tip electrode on all models except the guidewire tip and PV loop configuration: the band electrodes are set at predetermined distances along the catheter shaft. The catheters can be used for recording intracardiac signals in the atrial and ventricular regions of the heart and for cardiac stimulation during electrophysiology studies.

The Map-iT™ Diagnostic Mapping Catheters are comprised of a Pebax catheter shaft, platinum-iridium electrodes and electrical connectors that are either shrouded pin for the bipolar configurations or industry standard Redel connectors. The distal end of the fixed catheters is preformed into industry standard shapes with a variety of electrodes spacing to reach specific areas in the targeted chamber of the heart. The steerable catheters are available in a variety of curves that are controlled to the precise location in the heart by pulling or pushing the plunger on the handle. Table 5-1 lists the models and characteristics of the catheters.

{5}------------------------------------------------

INTENDED USE/INDICATION FOR USE:

The APT Map-It ™ Diagnostic Mapping Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:

Technologically the Map-iT™ Diagnostic Mapping Catheters are similar to St. Jude Medical Livewire™ Diagnostic Mapping Catheters in terms of indications for use, design, materials, technology and performance. Both families of catheters have shafts comprised of thermoplastic elastomer and platinum-iridium electrodes. Both come in multiple sizes, lengths, preformed curve shapes and configurations, including fixed and steerable, with varying number of electrodes and spacing's that are designed to reach specific areas in the targeted chambers of the heart. The 4/3.3F model functions the same as the predicate devices, with a taper for access. Similarly, the guidewire tip functions identically to a guidewire, providing easier access of the catheter to smaller anatomies.

TESTING PERFORMED

The Map-iT™ Diagnostic Mapping Catheters were thoroughly tested based on existing standards and test methods. These tests included visual, dimensional, electrical performance - continuity, leakage impedance, dielectric; mechanical performance - tensile, torque, deflection, shaft buckling, radio-detectability, corrosion; biocompatibility (short term <24 hours, blood contacting) and packaging. Testing demonstrates that the Map-iT™ Diagnostic Mapping Catheters met the specifications and are substantially equivalent to the predicate.

In addition, the Map-iT™ Diagnostic Mapping Catheters have been on the market in Europe since 2012.

There are no new questions raised regarding safety or efficacy of the Map-iTTM Diagnostic Mapping Catheters.

CONCLUSIONS

The Map-iTM Diagnostic Mapping Catheters are substantially equivalent in design, materials, sterilization, principles of operation, performance and indications for use to current commercially available catheters/cited predicates.

{6}------------------------------------------------

'able 5-1 Map-iT Diagnostic Mapping Catheter Models and Catheter Characteristics

Fixed Catheters

ModelFrench Size# of ElecElec. spacingRing Elec WidthTip Elec WidthCurveLength
2004/3.3F10Fixed Curve 4/3.3F Decapolar Catheters2-8-2-8-2-8-2-8-211.6S75
2014/3.3F102-10-2-10-2-10-2-10-211.6S75
2024/3.3F102-8-2-8-2-8-2-8-211.6S120
2034/3.3F102-10-2-10-2-10-2-10-211.6S120
2074/3.3F105-5-5-5-5-5-5-5-511.6S75
2094/3.3F105-5-5-5-5-5-5-5-511.6S120
2124/3.3F102-5-2-5-2-5-2-5-211.6S75
2134/3.3F102-5-2-5-2-5-2-5-211.6S120
Fixed Curve 4/3.3F Duo-decapolar Catheters
2504/3.3F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S75
2514/3.3F202-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-211.6S75
2524/3.3F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S120
2534/3.3F202-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-211.6S120
Fixed Curve 4F Bipolar Catheters
3004F21011.6MPA120
3014F21011.6MPB120
ModelFrenchSize# ofElecElec. spacingRingElecWidthTip ElecWidthCurveLength
9014704F410-10-10.11.6MPA120
9014714F410-10-10.11.6MPB120
9014724F410-10-10.11.6RVA120
9014734F410-10-10.11.6RVB120
9014744F410-10-10.11.6HIS120
9014754F410-10-10.11.6CS120
9014764F410-10-10.11.6CS65
9014774F45-5-5.11.6MPA120
9014784F45-5-5.11.6MPB120
9014794F45-5-5.11.6RVA120
9014804F45-5-5.11.6RVB120
9014814F45-5-5.11.6HIS120
9014824F42-5-2.11.6MPA120
9014834F42-5-2.11.6MPB65
9014844F42-5-2.11.6MPB120
9014854F42-2-2.11.6MPA120
9014864F42-2-2.11.6MPB120

{7}------------------------------------------------

Fixed Curve 4F Octapolar CathetersFixed Curve 4F Octapolar CathetersFixed Curve 4F Duo-decapolar Catheters
9012604F85-5-5-5-5-5-511.6MPA120
9012614F85-5-5-5-5-5-511.6MPB120
9012624F85-5-5-5-5-5-511.6HIS120
9012634F82-5-2-5-2-5-2-5-211.6MPA120
9012644F82-5-2-5-2-5-2-5-211.6MPB120
9012654F82-5-2-5-2-5-2-5-211.6MPB120
9012664F82-2-2-2-2-2-211.6MPA120
9012674F82-2-2-2-2-2-211.6MPB120
9012684F82-2-2-2-2-2-211.6HIS120
9012694F82-10-2-10-2-10-211.6CS120
9012704F82-10-2-10-2-10-211.6CS65
9012714F85-5-5-5-5-5-511.6CS120
9012724F85-5-5-5-5-5-511.6CS65
9012734F82-5-2-5-2-5-2-5-211.6CS120
9012744F82-5-2-5-2-5-2-5-211.6CS65
9017244F202-5-2-5-2-5-2-5-2-5-2-5-2-5-211.6S75
9017214F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S75
9017224F202-5-2-5-2-5-2-5-2-5-2-5-2-5-211.6S120
9011604F65-5-5-5-511.6 MPA 120
9011614F65-5-5-5-511.6 MPB 120
9011624F65-5-5-5-511.6 HIS 120
9011634F62-5-2-5-211.6 MPA 120
9011644F62-5-2-5-211.6 MPB 120
9011654F62-5-2-5-211.6 HIS 120
9011664F62-2-2-2-211.6 MPA 120
9011674F62-2-2-2-211.6 MPB 120
9011684F62-2-2-2-211.6 HIS 120
9011694F65-5-5-175-17511.6 MPA 120
9011704F65-5-5-175-17511.6 MPB 120
9011714F62-10-2-10-211.6 CS 120
9011724F62-10-2-10-211.6 CS 65
9011734F65-5-5-5-511.6 CS 120
9011744F65-5-5-5-511.6 CS 65
9011754F62-5-2-5-2-5-211.6 CS 120
9011764F62-5-2-5-2-5-211.6 CS 65
9015604F102-5-2-5-2-5-2-5-2-5-211.6 MPA 120
9015614F102-5-2-5-2-5-2-5-2-5-211.6 MPB 120

{8}------------------------------------------------

Fixed Curve 4F Decapolar Catheters (cont)Fixed Curve 4F Duo-decapolar Catheters (cont)
4F102-5-2-5-2-5-2-5-2-5-211.6HIS1209017234F202-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S120
4F102-8-2-8-2-8-2-8-211.6MPA120Fixed Curve 5F Bipolar Catheters
4F102-2-2-2-2-2-2-2-211.6MPB1209014005F21022MPA120
4F102-2-2-2-2-2-2-2-211.6HIS1209014015F21022MPB120
4F102-8-2-8-2-8-2-8-211.6CS120French# ofSize ElecElec. spacingRingElecWidthTip ElecWidthCurveLength
Fixed Curve 4F Decapolar CathetersFixed Curve 5F Quadripolar Catheters
4F102-8-2-8-2-8-2-8-211.6CS659014215F42-2-2.22MPA120
4F105-5-5-5-5-5-5-5-511.6CS1209014225F42-2-2.22MPB120
4F105-5-5-5-5-5-5-5-511.6CS659014235F410-10-10.22CS120
4F102-5-2-5-2-5-2-5-2-5-211.6CS1209014245F410-10-10.22CS65
4F102-5-2-5-2-5-2-5-2-5-211.6CS659014305F410-10-10.22MPB120
4F102-2-2-2-2-2-2-2-211.6CS1209014315F410-10-10.22MPA120
4F102-2-2-2-2-2-2-2-211.6CS659014325F410-10-10.22RVA120
4F102-8-2-8-2-8-2-8-211.6S759014335F410-10-10.22RVB120
4F102-10-2-10-2-10-2-10-211.6S759014345F410-10-10.22HIS120
4F102-8-2-8-2-8-2-8-211.6S1209014355F45-5-5.22MPB120
4F102-10-2-10-2-10-2-10-211.6S1209014365F45-5-5.22MPA120
Fixed Curve 4F Duo-decapolar Catheters9014375F45-5-5.22RVA120
4F202-5-2-5-2-5-2-5-2-5-2-5-2-5-211.6S759014385F45-5-5.22RVB120
4F202-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S759014395F45-5-5.22HIS120

{9}------------------------------------------------

Fixed Curve 4F Duo-decapolar Catheters (cont)Fixed Curve 5F Quadripolar Catheters (cont)
4F202-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-211.6S1209014405F42-5-2.22MPB120
4F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S1209014415F42-5-2.22MPA120
Fixed Curve 5F Bipolar Catheters9014425F42-5-2.22RVA120
5F21022MPA1209014435F42-5-2.22RVB120
5F21022MPB1209014445F42-5-2.22HIS120
Fixed Curve 5F Hexapolar CathetersFixed Curve 5F Octapolar Catheters
5F65-5-5-5-522MPA1209012005F85-5-5-5-5-5-512MPA120
5F65-5-5-5-522MPB1209012015F85-5-5-5-5-5-512MPB120
5F65-5-5-5-522HIS1209012025F85-5-5-5-5-5-512HIS120
5F62-5-2-5-222MPA1209012035F82-5-2-5-2-5-212MPA120
5F62-5-2-5-222MPB1209012045F82-5-2-5-2-5-212MPB120
5F62-5-2-5-222HIS1209012055F82-5-2-5-2-5-212HIS120
5F62-2-2-2-212MPA1209012065F82-2-2-2-2-2-212MPA120
5F62-2-2-2-212MPB1209012075F82-2-2-2-2-2-212MPB120
5F62-2-2-2-212HIS1209012085F82-2-2-2-2-2-212HIS120
5F65-5-5-175-17522MPA1209012095F82-10-2-10-2-10-222CS120
5F65-5-5-175-17522MPB1209012105F82-10-2-10-2-10-222CS65
5F62-10-2-10-222CS1209012115F85-5-5-5-5-5-522CS120
5F62-10-2-10-222CS659012125F85-5-5-5-5-5-522CS65
5F65-5-5-5-522CS1209012135F82-5-2-5-2-5-222CS120
5F65-5-5-5-522CS659012145F82-5-2-5-2-5-222CS65
5F62-5-2-5-222CS120Fixed Curve 5F Duo-decapolar Catheters
9031505F82-5-2-5-2-5-2-5-2-51.6S120

{10}------------------------------------------------

Fixed Curve 5F Decapolar CathetersFixed Curve 5F Duo-decapolar Catheters (cont)Fixed Curve 6F Bipolar CathetersFixed Curve 6F Quadripolar Catheters
9015185F102-5-2-5-2-5-2-5-212MPA1209034515F202-8-2-8-2-8-2-8-2-8-2-8-211.6S759035006F21022MPA1209014456F410-10-10.22MPB120
9015195F102-5-2-5-2-5-2-5-212MPB1209034525F202-5-2-5-2-5-2-5-2-5-2-5-211.6S1209035016F21022MPB1209014466F410-10-10.22MPA120
9015205F102-5-2-5-2-5-2-5-212HIS1209034535F202-8-2-8-2-8-2-8-2-8-2-8-211.6S120Fixed Curve 6F Quadripolar Catheters9014476F410-10-10.22RVA120
9034535F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-211.6S120Fixed Curve 6F Bipolar Catheters9014486F410-10-10.22RVB120
9015225F102-2-2-2-2-2-2-2-212MPB1209014496F410-10-10.22HIS120
9015235F102-2-2-2-2-2-2-2-212HIS1209014506F45-5-5.22MPB120
9015245F105-5-5-5-5-5-5-5-512CS1209014516F45-5-5.22MPA120
9015255F105-5-5-5-5-5-5-5-512CS659014526F45-5-5.22RVA120
9015265F102-5-2-5-2-5-2-5-212CS1209014536F45-5-5.22RVB120
9015275F102-5-2-5-2-5-2-5-212CS659014546F45-5-5.22HIS120
9015285F102-2-2-2-2-2-2-2-212CS1209014556F42-5-2.22MPB120
9015295F102-2-2-2-2-2-2-2-212CS659014566F42-5-2.22MPA120
9015305F102-8-2-8-2-8-2-8-212CS1209014576F42-5-2.22RVA120
9015315F102-8-2-8-2-8-2-8-212CS659014586F42-5-2.22RVB120
9015785F104-4-4-29-3-3-3-3-312MPA1209014596F42-5-2.22HIS120
9015795F104-4-4-23-4-4-4-4-412MPA1209014606F42-2-2.22MPB120
9015805F104-4-4-29-4-4-4-4-412MPA1209014616F42-2-2.22MPA120
9015815F104-4-4-23-5-5-5-5-512MPA1209014626F410-10-10.22CS120
9015825F104-4-4-29-5-5-5-5-512MPA1209014636F410-10-10.22CS65
9015835F105-5-5-20-3-3-3-3-312MPA120
9015845F105-5-5-24-3-3-3-3-312MPA120
9015855F105-5-5-20-4-4-4-4-412MPA120
9015865F105-5-5-24-4-4-4-4-412MPA120
9015875F105-5-5-20-5-5-5-5-512MPA120
9015885F105-5-5-24-5-5-5-5-512MPA120
9015895F104-4-4-29-3-3-3-3-312MPA120

{11}------------------------------------------------

Fixed Curve 5F Decapolar Catheters (cont)Fixed Curve 6F Decapolar Catheters
9015905F104-4-4-23-4-4-4-4-412MPB1209015326F102-8-2-8-2-8-2-8-212CS120
9015915F104-4-4-29-4-4-4-4-412MPB1209015336F102-8-2-8-2-8-2-8-212CS65
9015925F104-4-4-23-5-5-5-5-512MPB1209015346F102-5-2-5-2-5-2-5-212MPA120
9015935F104-4-4-29-5-5-5-5-512MPB1209015356F102-5-2-5-2-5-2-5-212MPB120
9015945F105-5-5-20-3-3-3-3-312MPB1209015366F102-5-2-5-2-5-2-5-212HIS120
9015955F105-5-5-24-3-3-3-3-312MPB1209015376F102-2-2-2-2-2-2-2-212MPA120
9015965F105-5-5-20-4-4-4-4-412MPB1209015386F102-2-2-2-2-2-2-2-212MPB120
9015975F105-5-5-24-4-4-4-4-412MPB1209015396F102-8-2-8-2-8-2-8-212HIS120
9015985F105-5-5-20-5-5-5-5-512MPB1209015406F105-5-5-5-5-5-5-5-512CS120
9015995F105-5-5-24-5-5-5-5-512MPB1209015416F105-5-5-5-5-5-5-5-512CS65
Fixed Curve 6F Hexapolar Catheters
9011306F65-5-5-5-522MPA1209015426F102-5-2-5-2-5-2-5-212CS120
9011316F65-5-5-5-522MPB1209015436F102-5-2-5-2-5-2-5-212CS65
9011326F65-5-5-5-522HIS1209015446F102-2-2-2-2-2-2-2-212CS120
9011336F62-5-2-5-222MPA1209015456F102-2-2-2-2-2-2-2-212CS65
9011346F62-5-2-5-222MPB1209018226F104-4-4-29-3-3-3-3-312MPA120
9011356F62-5-2-5-222HIS1209018236F104-4-4-23-4-4-4-4-412MPA120
9018246F104-4-4-29-4-4-4-4-412MPA120

{12}------------------------------------------------

Fixed Curve 6F Hexapolar Catheters (cont)
9011366F62-2-2-2-222MPA120
9011376F62-2-2-2-222MPB120
9011386F62-2-2-2-222HIS120
9011396F65-5-5-175-17522MPA120
9011406F65-5-5-175-17522MPB120
9011416F62-10-2-10-222CS120
9011426F62-10-2-10-222CS65
9011436F65-5-5-5-522CS120
9011446F65-5-5-5-522CS65
9011456F62-5-2-5-222CS120
9011466F62-5-2-5-222CS65
Fixed Curve 6F Octapolar Catheters
9012306F85-5-5-5-5-5-512MPA120
9012316F85-5-5-5-5-5-512MPB120
9012326F85-5-5-5-5-5-512HIS120
9012336F82-5-2-5-2-5-212MPA120
9012346F82-5-2-5-2-5-212MPB120
Fixed Curve 6F Decapolar Catheters (cont)
9018256F104-4-4-23-5-5-5-5-512MPA120
9018266F104-4-4-29-5-5-5-5-512MPA120
9018276F105-5-5-20-3-3-3-3-312MPA120
9018286F105-5-5-24-3-3-3-3-312MPA120
9018296F105-5-5-20-4-4-4-4-412MPA120
9018306F105-5-5-24-4-4-4-4-412MPA120
9018316F105-5-5-20-5-5-5-5-512MPA120
9018326F105-5-5-24-5-5-5-5-512MPA120
9018336F104-4-4-29-3-3-3-3-312MPB120
9018346F104-4-4-23-4-4-4-4-412MPB120
9018356F104-4-4-29-4-4-4-4-412MPB120
9018366F104-4-4-23-5-5-5-5-512MPB120
9018376F104-4-4-29-5-5-5-5-512MPB120
9018386F105-5-5-20-3-3-3-3-312MPB120
9018396F105-5-5-24-3-3-3-3-312MPB120
9018406F105-5-5-20-4-4-4-4-412MPB120
9018416F105-5-5-24-4-4-4-4-412MPB120
9018426F105-5-5-20-5-5-5-5-512MPB120
Fixed Curve 6F Decapolar Catheters (cont)9018436F105-5-5-24-5-5-5-5-512MPB120

{13}------------------------------------------------

65751201207512075120
SSSSSSS
GuidewireGuidewireGuidewireGuidewireGuidewireGuidewireGuidewire
Fixed Curve 4F Decapolar Catheters with Guidewire Tip2-8-2-8-2-8-2-8-22-8-2-8-2-8-2-8-22-10-2-10-2-10-2-10-22-10-2-10-2-10-2-10-2Fixed Curve 4F Duo-decapolar Catheters with Guidewire Tip2-8-2-8-2-8-2-8-2-8-2-8-22-8-2-8-2-8-2-8-2-8-2-8-2
101010102020
4F4F4F4F4F4F
903307903308903313903314903351903353
Fixed Curve 6F Octapolar Catheters (cont)
9012356F82-5-2-5-2-5-212HIS120
9012366F82-2-2-2-2-2-212MPA120
9012376F82-2-2-2-2-2-212MPB120
9012386F82-2-2-2-2-2-212HIS120
9012396F82-10-2-10-2-10-212CS120
9012406F82-10-2-10-2-10-212CS65
9012416F85-5-5-5-5-5-512CS120
9012426F85-5-5-5-5-5-512CS65
9012436F82-5-2-5-2-5-212CS120
9012446F82-5-2-5-2-5-212CS65
FIXED CURVE CATHETERS WITH GUIDEWIRE TIP
Fixed Curve 4/3.3F Decapolar Catheters with Guidewire Tip
9032144/3.3F102-8-2-8-2-8-2-8-21GuidewireS75
9032154/3.3F102-8-2-8-2-8-2-8-21GuidewireS120
9032174/3.3F105-5-5-5-5-5-5-5-51GuidewireS120
9032184/3.3F102-5-2-5-2-5-2-5-21GuidewireS75
9032194/3.3F102-5-2-5-2-5-2-5-21GuidewireS120
9032204/3.3F102-10-2-10-2-10-2-10-21GuidewireS75
9032214/3.3F102-10-2-10-2-10-2-10-21GuidewireS120

{14}------------------------------------------------

Fixed Curve 4/3.3F Duo-decapolar Catheters with Guidewire Tip
9032544/3.3F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-21GuidewireS75
9032554/3.3F202-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-21GuidewireS75
9032564/3.3F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-21GuidewireS120
9032574/3.3F202-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-21GuidewireS120
9032704/3.3F205-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-51GuidewireS120
9032714/3.3F205-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-51GuidewireS120
9032724/3.3F202-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-21GuidewireS75
9032734/3.3F202-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-21GuidewireS120

{15}------------------------------------------------

Catheters
Steerable
ModelFrench Size# of ElecElec. spacingRing Elec WidthTip Elec WidthCurveLength (cm)
Steerable 4F Quadripolar Catheters
9036004F45-5-5.12MPD-B110
9036014F45-5-5.12MPD-D110
9036024F45-5-5.12MPD-DL110
9036034F42-5-2.12MPD-B110
9036044F42-5-2.12MPD-D110
9036054F42-5-2.12MPD-DL110
9036064F42-5-2.12MPD-S110
9036074F45-5-5.12MPD-S110
9036084F42-5-2.12MPD-SL110
9036094F45-5-5.12MPD-SL110
9036104F42-5-2.12MPD-CS-S110
9036114F42-8-2.12MPD-CS-S110
9036124F42-5-2.12MPD-CS-S110
9036134F45-5-5.12MPD-CS-S110
9036144F42-5-2.12MPD-CS-M110
9036154F42-8-2.12MPD-CS-M110
9036164F42-2-2.12MPD-CS-M110
9036174F45-5-5.12MPD-CS-M110
9036184F42-5-2.12MPD-CS-M110
ofElecElec. spacingRingElecWidthTipElecWidthCurveLength(cm)
Steerable 4F Pentapolar Catheters
55-5-5-17512MPD-B110
55-5-5-17512MPD-D110
55-5-5-17512MPD-DL110
52-5-2-17512MPD-B110
52-5-2-17512MPD-D110
52-5-2-17512MPD-DL110
52-2-2-17522MPD-S110
55-5-5-17522MPD-S110
52-2-2-17512MPD-SL110
55-5-5-17512MPD-SL110

{16}------------------------------------------------

Steerable 5F Pentapolar Catheters
1109043015F55-5-5-17512MPD-B110
1109043025F55-5-5-17512MPD-D110
1109043035F55-5-5-17512MPD-DL110
1109043045F52-5-2-17512MPD-B110
1109043055F52-5-2-17512MPD-D110
1109043065F52-5-2-17512MPD-DL110
1109043075F52-2-2-17512MPD-S110
1109043085F55-5-5-17512MPD-S110
1109043095F52-2-2-17512MPD-SL110
1109043105F55-5-5-17512MPD-SL110
Steerable 5F Decapolar Catheters
1109016565F102-5-2-5-2-5-2-5-212MPD-S110
1109016575F102-8-2-8-2-8-2-8-212MPD-S110
1109018805F102-5-2-5-2-5-2-5-212MPD-S-M110
1109018815F102-5-2-5-2-5-2-5-212MPD-S-L110
1109018825F102-8-2-8-2-8-2-8-212MPD-S-M110
Steerable 4F Quadripolar Catheters
9036194F42-8-2.12MPD-CS-L110
9036204F42-2-2.12MPD-CS-L110
9036214F45-5-5.12MPD-CS-L110
Steerable 5F Quadripolar Catheters
9016005F45-5-5.12MPD-B110
9016015F45-5-5.12MPD-D110
9016025F42-5-2.12MPD-B110
9016035F42-5-2.12MPD-D110
9017005F45-5-5.12MPD-DL110
9017015F42-5-2.12MPD-DL110
9016985F42-2-2.12MPD-S110
9016045F45-5-5.12MPD-S110
9017025F42-2-2.12MPD-SL110
9017035F45-5-5.12MPD-SL110
9037005F42-5-2.12MPD-CS-S110
9037015F42-8-2.12MPD-CS-S110
9037025F42-2-2.12MPD-CS-S110
9037035F45-5-5.12MPD-CS-S110
9037045F42-5-2.12MPD-CS-M110
9037055F42-8-2.12MPD-CS-M110

{17}------------------------------------------------

Catheters (cont)Steerable 5F Decapolar Catheters (cont)
12MPD-CS-M1109018835F102-8-2-8-2-8-2-8-212MPD-S-L110
12MPD-CS-M1109018845F105-5-5-5-5-5-5-5-512MPD-S-M110
12MPD-CS-L1109018855F105-5-5-5-5-5-5-5-512MPD-S-L110
12MPD-CS-L1109018865F102-2-2-2-2-2-2-2-212MPD-S-M110
12MPD-CS-L1109018875F102-2-2-2-2-2-2-2-212MPD-S-L110
12MPD-CS-L1109016505F105-5-5-5-5-5-5-5-512MPD-D110
12MPD-S-M1109016515F102-5-2-5-2-5-2-5-212MPD-D110
12MPD-S-M1109016525F102-2-2-2-2-2-2-2-212MPD-D110
12MPD-S-M1109017045F105-5-5-5-5-5-5-5-512MPD-DL110
12MPD-S-L1109017055F102-5-2-5-2-5-2-5-212MPD-DL110
12MPD-S-L1109017065F102-2-2-2-2-2-2-2-212MPD-DL110
12MPD-S-L1109017075F102-5-2-5-2-5-2-5-212MPD-SL110
9017085F102-8-2-8-2-8-2-8-212MPD-SL110
12MPD-B1109037305F102-5-2-5-2-5-2-5-212MPD-CS-S110
12MPD-D1109037315F102-8-2-8-2-8-2-8-212MPD-CS-S110
12MPD-B1109037325F102-2-2-2-2-2-2-2-212MPD-CS-S110
12MPD-DI1109037335F105-5-5-5-5-5-5-5-512MPD-CS-S110
Steerable 5F Quadripolar Catheters (cont)
9037065F42-2-2.12MPD-CS-M110
9037075F45-5-5.12MPD-CS-M110
9037085F42-5-2.12MPD-CS-L110
9037095F42-8-2.12MPD-CS-L110
9037105F42-2-2.12MPD-CS-L110
9037115F45-5-5.12MPD-CS-L110
9017925F42-5-2.12MPD-S-M110
9017935F45-5-5.12MPD-S-M110
9017945F42-2-2.12MPD-S-M110
9017955F42-5-2.12MPD-S-L110
9017965F45-5-5.12MPD-S-L110
9017975F42-2-2.12MPD-S-L110
Steerable 6F Quadripolar Catheters
9016106F45-5-5.12MPD-B110
9016116F45-5-5.12MPD-D110
9016126F42-5-2.12MPD-B110
9017096F45-5-512MPD-110

5-16

{18}------------------------------------------------

Steerable 6F Quadripolar Catheters (cont)Steerable 5F Decapolar Catheters (cont)
9016136F42-5-2.12MPD-D1109037345F102-5-2-5-2-5-2-5-212MPD-CS-M110
9017106F42-5-2.12MPD-DL1109037355F102-8-2-8-2-8-2-8-212MPD-CS-M110
9016996F42-2-2.12MPD-S1109037365F102-2-2-2-2-2-2-2-212MPD-CS-M110
9016146F45-5-5.12MPD-S1109037375F105-5-5-5-5-5-5-5-512MPD-CS-M110
9017126F42-2-2.12MPD-SL1109037385F102-5-2-5-2-5-2-5-212MPD-CS-L110
9017136F45-5-5.12MPD-SL1109037395F102-8-2-8-2-8-2-8-212MPD-CS-L110
9038006F42-5-2.12MPD-CS-S1109037405F102-2-2-2-2-2-2-2-212MPD-CS-L110
9038016F42-8-2.12MPD-CS-S1109037415F105-5-5-5-5-5-5-5-512MPD-CS-L110
9038026F42-2-2.12MPD-CS-S1109016535F105-5-5-5-5-5-5-5-512MPD-G110
9038036F45-5-5.12MPD-CS-S1109016545F102-5-2-5-2-5-2-5-212MPD-G110
9038046F42-5-2.12MPD-CS-M1109016555F102-2-2-2-2-2-2-2-212MPD-G110
9038056F42-8-2.12MPD-CS-M1109017805F102-5-2-5-2-5-2-5-212MPD-CS-M2110
9038066F42-2-2.12MPD-CS-M1109017815F102-5-2-5-2-5-2-5-212MPD-CS-L2110
9038076F45-5-5.12MPD-CS-M1109017825F102-8-2-8-2-8-2-8-212MPD-CS-M2110
9038086F42-5-2.12MPD-CS-L1109017835F102-8-2-8-2-8-2-8-212MPD-CS-L2110
9038096F42-8-2.12MPD-CS-L1109017845F105-5-5-5-5-5-5-5-512MPD-CS-M2110
9038106F42-2-2.12MPD-CS-L1109017855F105-5-5-5-5-5-5-5-512MPD-CS-L2110

{19}------------------------------------------------

Steerable 6F Duo-decapolar CathetersSteerable 7F Quadripolar Catheters
6F20127F412
9016706F202-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-2-10-212MPD-H1009016207F42-5-2.12MPD-A110
9016716F202-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-2-5-212MPD-H1009016217F42-5-2.12MPD-B110
9016726F205-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-512MPD-H1009016227F42-5-2.12MPD-C110
9016736F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-212MPD-H1009016237F42-5-2.14MPD-D110
9016746F202-8-2-8-2-8-2-8-2-60-2-8-2-8-2-8-2-8-212MPD-H1009016267F42-5-2.12MPD-B85
9016756F202-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-2-8-212MPD-L1009016277F42-5-2.12MPD-C85
9017197F42-5-2.14MPD-DL110
9016247F42-5-2.14MPD-E110
9016257F42-5-2.14MPD-F110
Steerable 6F Quadripolar Catheters (cont)Steerable 6F Pentapolar Catheters
9038116F45-5-5.12MPD-CS-L9043516F55-5-5-17512MPD-B9043575F52-2-2-175110
9017986F42-5-2.12MPD-S-M9043526F55-5-5-17512MPD-D9043585F55-5-5-175110
9017996F45-5-5.12MPD-S-M9043536F55-5-5-17512MPD-DL9043596F52-2-2-175110
9018766F42-2-2.12MPD-S-M9043546F52-5-2-17512MPD-B9043606F55-5-5-175110
9018776F42-5-2.12MPD-S-L9043556F52-5-2-17512MPD-D
9018786F45-5-5.12MPD-S-L9043566F52-5-2-17512MPD-DL
9018796F42-2-2.12MPD-S-L

{20}------------------------------------------------

Steerable 7F Quadripolar Catheters (cont)
9017207F42-5-2.14MPD-DL85
9016297F42-5-2.14MPD-E85
9016307F42-5-2.14MPD-F85
Steerable 7F Decapolar Catheters
9039007F105-5-5-5-5-5-5-5-512MPD-D110
9039017F102-5-2-5-2-5-2-5-212MPD-D110
9039027F102-2-2-2-2-2-2-2-212MPD-D110
9039037F105-5-5-5-5-5-5-5-512MPD-DL110
9039047F102-5-2-5-2-5-2-5-212MPD-DL110
9039057F102-2-2-2-2-2-2-2-212MPD-DL110
9039067F105-5-5-5-5-5-5-5-512MPD-G110
9039077F102-5-2-5-2-5-2-5-212MPD-G110
9039087F102-2-2-2-2-2-2-2-212MPD-G110
9039097F102-5-2-5-2-5-2-5-212MPD-S110
9039107F102-8-2-8-2-8-2-8-212MPD-S110
9039117F102-5-2-5-2-5-2-5-212MPD-SL110
9039127F102-8-2-8-2-8-2-8-212MPD-SL110
9039137F102-5-2-5-2-5-2-5-212MPD-CS-S110
Steerable 6F Decapolar Catheters
9018886F102-5-2-5-2-5-2-5-212MPD-S-M110
9018896F102-5-2-5-2-5-2-5-212MPD-S-L110
9018906F102-8-2-8-2-8-2-8-212MPD-S-M110
9018916F102-8-2-8-2-8-2-8-212MPD-S-L110
9018926F105-5-5-5-5-5-5-5-512MPD-S-M110
9018936F105-5-5-5-5-5-5-5-512MPD-S-L110
9018946F102-2-2-2-2-2-2-2-212MPD-S-M110
9018956F102-2-2-2-2-2-2-2-212MPD-S-L110
9016606F105-5-5-5-5-5-5-5-512MPD-D110
9016616F102-5-2-5-2-5-2-5-212MPD-D110
9016626F102-2-2-2-2-2-2-2-212MPD-D110
9017146F105-5-5-5-5-5-5-5-512MPD-DL110
9017156F102-5-2-5-2-5-2-5-212MPD-DL110
9017166F102-2-2-2-2-2-2-2-212MPD-DL110
9017176F102-5-2-5-2-5-2-5-212MPD-SL110
9017186F102-8-2-8-2-8-2-8-212MPD-SL110
9038306F102-5-2-5-2-5-2-5-212MPD-CS-S110

{21}------------------------------------------------

Steerable 7F Decapolar Catheters (cont)
MPD-CS-S1109039147F102-8-2-8-2-8-2-8-212MPD-CS-S110
MPD-CS-S1109039157F102-2-2-2-2-2-2-2-212MPD-CS-S110
MPD-CS-S1109039167F105-5-5-5-5-5-5-5-512MPD-CS-S110
MPD-CS-M1109039177F102-5-2-5-2-5-2-5-212MPD-CS-M110
MPD-CS-M1109039187F102-8-2-8-2-8-2-8-212MPD-CS-M110
MPD-CS-M1109039197F102-2-2-2-2-2-2-2-212MPD-CS-M110
MPD-CS-M1109039217F105-5-5-5-5-5-5-5-512MPD-CS-M110
MPD-CS-L1109039227F102-5-2-5-2-5-2-5-212MPD-CS-L110
MPD-CS-L1109039237F102-8-2-8-2-8-2-8-212MPD-CS-L110
MPD-CS-L1109039247F102-2-2-2-2-2-2-2-212MPD-CS-L110
MPD-CS-L1109039257F105-5-5-5-5-5-5-5-512MPD-CS-L110
MPD-G110Steerable 7F Decapolar PV Catheters
MPD-G1109016417F103-3-3-3-3-3-3-3-31PV12110
MPD-G1109016427F104-4-4-4-4-4-4-4-41PV15110
MPD-S1109016437F105-5-5-5-5-5-5-5-51PV18110
MPD-S1109016447F106-6-6-6-6-6-6-6-61PV21110
MPD-CS-M21109016457F107-7-7-7-7-7-7-7-71PV24110
9038316F102-8-2-8-2-8-2-8-212110MPD-CS-S
9038336F102-2-2-2-2-2-2-2-212110MPD-CS-S
9038346F105-5-5-5-5-5-5-5-512110MPD-CS-S
9038356F102-5-2-5-2-5-2-5-212110MPD-CS-M
9038366F102-8-2-8-2-8-2-8-212110MPD-CS-M
9038376F102-2-2-2-2-2-2-2-212110MPD-CS-M
9038386F105-5-5-5-5-5-5-5-512110MPD-CS-M
9038396F102-5-2-5-2-5-2-5-212110MPD-CS-L
9038406F102-8-2-8-2-8-2-8-212110MPD-CS-L
9038416F102-2-2-2-2-2-2-2-212110MPD-CS-L
9038426F105-5-5-5-5-5-5-5-512110MPD-CS-L
9016636F105-5-5-5-5-5-5-5-512110MPD-G
9016646F102-5-2-5-2-5-2-5-212110MPD-G
9016656F102-2-2-2-2-2-2-2-212110MPD-G
9016666F102-5-2-5-2-5-2-5-212110MPD-S
9016676F102-8-2-8-2-8-2-8-212110MPD-S
9017866F102-5-2-5-2-5-2-5-212110MPD-CS-M2
Steerable 6F Decapolar Catheters (cont)

{22}------------------------------------------------

Steerable 7F Duo-decapolar PV Catheters
9039507F203-3-3-3-3-3-3-3-3-3-3-3-3-3-3-3-3-30.6-PV12110
9039517F204-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-40.6-PV15110
9039527F205-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-5-50.6-PV18110
9039537F206-6-6-6-6-6-6-6-6-6-6-6-6-6-6-6-6-6-60.6-PV21110
9039547F207-7-7-7-7-7-7-7-7-7-7-7-7-7-7-7-7-7-70.6-PV24110
1101101101101101001 0000 I100100100
TIS Camות מנויות מנואנתנויMPD- CS-M2כתבים במשMPD-HMPD-HMPD-FMPD-HMPD-HMPD-
IIIIIIIIIII
teerable 6F Decapolar Catheters (con5-2-5-2-5-2-58-2-8-2-8-2-8-2-8-2-8-25-5-5-5-5-55-5-5-5-5-5-5eerable 7F Duo-decapolar Catheter10-2-10-2-10-2-10-2 0-2-10-2-10-2-10-2 0-2-10-2-10-2-10-25-2-5-2-5-2-5-2-5-2- 5-2-5-2-5-2-5-2- 2-5-2-5-25-5-5-5-5-5-5-5-5-5 5-5-5-5-5-5-5-5- 5-5-5-5-58-2-8-2-8-2-8-2-1 8-2-8-2-8-2-8-2-1 2-8-2-8-28-2-8-2-8-2-8-2- 60-2-8-2-8-2-8-2 2-8-2-8-28-2-8-2-8-2-8-2 8-2-8-2-8-2-8-2 2-8-2-8-2
0101010101070000070000
1919191919ELELELELLEEL
01787901788901789901790901790168(90168901682901683901684901685

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).