(66 days)
The CX30N(CX30PQX) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. The display is not intended for mammography.
CX30N(CX30PQX) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for color image display. The WIDE controller board is installed into the PACS workstation computer system to display PACS medical images.
The provided text describes the 510(k) summary for the CX30N(CX30PQX) LCD Monitor System, which is a medical image display device. It is important to note that this document is for a medical monitor, not an AI/ML algorithm. Therefore, many of the typical acceptance criteria and study aspects related to AI/ML device performance (like MRMC studies, ground truth establishment for training/test sets, or expert consensus for labeling) are not applicable here.
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the technical performance specifications of the monitor and the tests conducted to demonstrate that the monitor meets these specifications and relevant safety standards for its intended use as a medical display.
Here's an analysis based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for a medical display like the CX30N(CX30PQX) are typically based on industry standards and performance specifications for medical imaging. The document highlights a comparison with a predicate device, which implicitly defines common performance expectations.
| Parameter | Acceptance Criteria (Implied / Predicate Performance) | Reported Device Performance (CX30N(CX30PQX)) |
|---|---|---|
| Intended Use | Display and view digital medical images for review and analysis by trained medical practitioners. (Not for mammography) | Display and view digital medical images for review and analysis by trained medical practitioners. (Not for mammography) |
| LCD Panel Size | 21.3" | 21.3" |
| Resolution | 2048 x 1536 | 2048 x 1536 |
| Pixel Pitch | 0.21mm x 0.21mm | 0.2109mm x 0.2109mm |
| Brightness | 1000 cd/m² (Predicate) | 900 cd/m² |
| Contrast Ratio | 1500:1 (Predicate) | 1400:1 |
| Input Signal | DVI-D, DisplayPort (Predicate) | DVI-I, DisplayPort |
| Safety | Compliance with IEC 60601-1 | Complies with IEC 60601-1 |
| EMC | Compliance with IEC 60601-1-2 | Complies with IEC 60601-1-2 |
| Effectiveness | Meeting standards for resolution, luminance, contrast, and noise. | Tests met acceptance criteria |
2. Sample Size Used for the Test Set and Data Provenance
For a medical monitor, the "test set" does not refer to clinical image data in the same way it would for an AI/ML algorithm. Instead, the testing involves the device itself.
- Sample Size: The document implies that testing was conducted on the CX30N(CX30PQX) device as a unit. It doesn't specify a "sample size" in terms of multiple devices, typically performance testing like resolution, luminance, contrast, and noise are done on prototypes or production samples to ensure they meet specifications.
- Data Provenance: Not applicable in the context of clinical image data. The "data" here are the performance measurements of the monitor itself. The manufacturer is WIDE Corporation, located in the Republic of Korea. The testing would have been conducted by the manufacturer or accredited testing labs.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. For a medical monitor, "ground truth" is established through physical measurement standards and calibration, not expert interpretation of images. The performance of the display is objectively measured against technical specifications and international standards (e.g., luminance, resolution). Human experts are the users of the display, not the ones establishing its ground truth.
4. Adjudication Method for the Test Set
Not applicable. This concept applies to the review of image data (e.g., for disease diagnosis) where there might be inter-reader variability. For a display device, performance measurements like brightness or resolution are objectively quantifiable and don't require adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. An MRMC study is a clinical study design used to evaluate the diagnostic performance of a medical imaging system or a CAD (Computer-Aided Detection)/AI device by assessing how human readers' diagnostic accuracy changes with and without the assistance of the device, across multiple cases. This is not applicable to a digital medical display itself, which is a display hardware, not a diagnostic aid.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. The CX30N(CX30PQX) is a display device, not an algorithm. Its "performance" is its ability to accurately display images according to specifications, not to interpret or diagnose.
7. The Type of Ground Truth Used
The ground truth for this device (a medical monitor) is based on:
- Technical Specifications: Defined parameters like resolution (e.g., 2048x1536 pixels), pixel pitch, brightness (cd/m²), contrast ratio, and input signal compatibility.
- International Standards: Compliance with recognized safety and performance standards such as IEC 60601-1 (general safety) and IEC 60601-1-2 (electromagnetic compatibility).
- Measured Performance: Calibration and testing results (e.g., resolution, luminance, contrast, noise) demonstrating that the device meets these specifications and standards.
8. The Sample Size for the Training Set
Not applicable. This device is a hardware display and does not involve AI/ML algorithms that require a "training set" of data.
9. How the Ground Truth for the Training Set was Established
Not applicable. As explained above, there is no AI/ML training set for this device.
In summary, for the CX30N(CX30PQX) LCD Monitor System:
The acceptance criteria are primarily technical specifications and compliance with international safety and performance standards for medical displays. The study proving these criteria are met involves non-clinical bench testing to measure parameters like resolution, luminance, contrast, and noise, and to verify compliance with electrical safety and electromagnetic compatibility standards (IEC 60601-1 and IEC 60601-1-2). The document explicitly states that "No clinical studies were considered necessary and performed," which is typical for a device like a monitor where performance is assessed through engineering and physical measurements rather than clinical diagnostic accuracy studies.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, resembling a family or group of people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2016
WIDE Corporation % Mr. YeoJin Yun RA Manager 12 Wongomae-ro, Giheung-gu Yongin-si, Gyeonggi-do 17086 REPUBLIC OF KOREA
Re: K160348
Trade/Device Name: CX30N(CX30POX) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: February 5, 2016 Received: March 14, 2016
Dear Mr. Yun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K160348
Device Name CX30N(CX30PQX)
Indications for Use (Describe)
The CX30N(CX30PQX) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. The display is not intended for mammography.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
1. Date Prepared [21 CFR 807.92(a) (1)]
02/05//2016
2. Submitter's Information [21 CFR 807.92(a) (1)]
| Name of Sponsor: | WIDE Corporation. |
|---|---|
| Address: | 12 Wongomae-Ro, Giheung-Gu, Yongin-Si, Gyeonggi-Do17086, Republic of Korea |
| Contact Name: | YeoJin Yun |
| Telephone #: +82-31-218-1675 | |
| Fax #: +82-31-218-7400 | |
| Email: yyjin@widecorp.com | |
| Registration Number: | 3004082357 |
| Name of Manufacturer: | Same as Sponsor |
3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| Model Name: | CX30N(CX30PQX) |
|---|---|
| Common Name: | TFT LCD Medical Monitor System |
| Classification Name: | Display, Diagnostic Radiology |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | PGY |
| Device Class: | 2 |
| Review Panel: | Radiology |
4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| 510(k) Number: | K153354 |
|---|---|
| Applicant: | EIZO CORPORATION |
| Model Name: | 3MP Color LCD Monitor, RadiForce RX350 |
| Common Name: | TFT LCD Medical Monitor System |
| Classification Name: | Display, Diagnostic Radiology |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | PGY |
| Device Class: | 2 |
{4}------------------------------------------------
5. Description of the Device [21 CFR 807.92(a) (4)]
CX30N(CX30PQX) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for color image display. The WIDE controller board is installed into the PACS workstation computer system to display PACS medical images.
6. Intended Use [21 CFR 807.92(a) (5)]
CX30N(CX30PQX) is intended to be used to display and view digital medical images for review and analysis by trained medical practitioners. The display is not intended for mammography.
7. Technological Characteristics [21 CFR 807.92(a) (6)]
The device is an image display system which consists of computer software and components. The device does not contact the patient, nor does it control any life sustaining devices. A physician or trained medical practitioner provides ample opportunity for competent human intervention to interpret images and information being displayed.
| Parameter | Subject Device | Predicate Device |
|---|---|---|
| 510(k) Number | Unknown | K153354 |
| Model Name | CX30N(CX30PQX) | 3MP Color LCD Monitor.RadiForce RX350 |
| Manufacturer | WIDE Corporation. | EIZO CORPORATION |
| Common Name | TFT LCD Medical Monitor System | |
| Classification Name | Display, Diagnostic Radiology | |
| Classification Panel | Radiology | |
| Classification Regulation | 21 CFR 892.2050 | |
| Product Code | PGY | |
| Device Class | Class II | |
| Intended Use | CX30N(CX30PQX) is intendedto be used in displaying andviewing digital medical imagesfor review and analysis by trainedmedical practitioners. The displayis not intended formammography. | This product is intended to beused in displaying and viewingdigital images for review,analysis and diagnosis by trainedmedical practitioners. It does notsupport the display ofmammography images fordiagnosis. |
| LCD Panel Size | 21.3" | 21.3" |
8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]
{5}------------------------------------------------
| Parameter | Subject Device | Predicate Device |
|---|---|---|
| Resolution | 2048×1536 | 2048×1536 |
| Pixel pitch | 0.2109mm x 0.2109mm | 0.21mm x 0.21mm |
| Brightness | 900cd/m² | 1000cd/m² |
| Contrast Ratio | 1400 : 1 | 1500:1 |
| Input Signal | DVI-I, DisplayPort | DVI-D, DisplayPort |
| Power Supply | 100~240 VAC, 50/60Hz | 100~240 VAC, 50/60Hz |
| Color/Monochrome | Color | Color |
When compared to the predicate devices (K155354), the CX30N(CX30PQX) presented in this submission has the same of the followings:
- Intended Use
- Technological characteristics
- LCD Panel Size
- Resolution
- Pixel pitch
The two devices share the similar performance as the following:
- Contrast Ratio
- · Brightness
- Input Signal
9. Summary of Non-Clinical Data
CX30N(CX30PQX) comply with the following international and FDA-recognized consensus standards:
| IEC 60601-1: | Medical Electrical Equipment -- Part 1: GeneralRequirements For Basic Safety And Essential Performance |
|---|---|
| IEC 60601-1-2: | Medical Electrical Equipment - Part 1-2: GeneralRequirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility -Requirements And Tests |
The tests conducted for resolution, luminance, contrast and noise all met the acceptance criteria specified in the standards.
10. Summary of Clinical Data
No clinical studies were considered necessary and performed.
{6}------------------------------------------------
11. Conclusion [21 CFR 807.92(b) (3)]
Subject Device is substantially equivalent to the currently marketed and predicate devices in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, the safety of the subject device was validated through tests including IEC60601-1 and IEC 60601-1-2. The effectiveness of the device was validated through tests for resolution, luminance, contrast and noise.
The results of these tests demonstrate that CX30N(CX30PQX) meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing demonstrates that the device is as safe and effective as the predicate device and performs as well as the predicate device.
N/A