K Number
K160345

Validate with FDA (Live)

Date Cleared
2017-03-03

(389 days)

Product Code
Regulation Number
872.3275
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Self-curing glass ionomer based radiopaque filling cement, for dental use only. Recommended Indications:
(1) Restoration of primary teeth.
(2) Restorations of Class III, V and limited Class I cavities.
(3) Base linings under composite and amalgam.
(4) Core build-up.

Device Description

Glass Ionomer Cement (Filling) has two types: Filling II. It is a device composed of various materials including Silica, Alumina, Strontium Fluoride in the powder composition and Poly Carboxylic Acid in the liquid composition which intended to serve as a permanent or temporary tooth filling or as base linings. It's a powder liquid system. The basic operation principle of Glass Ionomer Cement (Filling) is acid-base reaction.

AI/ML Overview

The provided document describes the acceptance criteria and the study (physical and chemical properties tests) for the "Glass Ionomer Cement (Filling)" device (K160345).

Here's the breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

ItemAcceptance Criteria (Standard Requirement - ISO 9917-1:2007)Reported Device Performance (Conclusion)
ComponentsLiquid: Free from deposits or filaments, no visible signs of gelation. Powder: Free from extraneous material, uniformly dispersed pigment (if colored). Unset cement: Homogeneous and smooth consistency.Within spec set by standard
Net setting timeFor restoration: 1.5~6 minWithin spec set by standard
Compressive strengthFor restoration: ≥100 MPaWithin spec set by standard
Acid erosion≤0.17 mmWithin spec set by standard
Acid-soluble Lead content≤100 mg/kgWithin spec set by standard
Radio-opacityAt least equivalent to that for the same thickness of aluminum.Within spec set by standard

2. Sample size used for the test set and the data provenance

The document does not specify the sample size used for each physical and chemical properties test. It also does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. Given that it refers to ISO standards and a submission by a Chinese company, it's highly probable the testing was conducted in a laboratory in China following international standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the study described is a series of laboratory-based physical and chemical properties tests against predefined ISO standards, not a clinical study involving human assessment or interpretation requiring expert ground truth establishment.

4. Adjudication method for the test set

This section is not applicable as the study described is a series of laboratory-based physical and chemical properties tests against predefined ISO standards. There is no mention of adjudication for these types of tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The document describes a traditional device clearance for a dental material (Glass Ionomer Cement), not an AI-powered diagnostic or assistive device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable as the device is a physical dental material, not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance criteria in this context is established by the ISO 9917-1:2007 standard for Dentistry — Water-based cements - Part 1: Powder/liquid acid-base cements. The reported performance is a direct measurement against these internationally recognized physical and chemical property specifications.

8. The sample size for the training set

This section is not applicable as the device is a physical dental material and not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

This section is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 3, 2017

Shandong Huge Dental Material Corporation Steven Song General Manager No. 68 Shanhai Road, Donggang District Rizhao City, 276800 CN

Re: K160345

Trade/Device Name: Glass Ionomer Cement Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: February 24, 2017 Received: March 1, 2017

Dear Steven Song:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160345

Device Name

Glass Ionomer Cement (Filling)

Indications for Use (Describe)

Self-curing glass ionomer based radiopaque filling cement, for dental use only. Recommended Indications:

(1) Restoration of primary teeth.

(2) Restorations of Class III, V and limited Class I cavities.

  • (3) Base linings under composite and amalgam.
    (4) Core build-up.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for a company called HUGE. The logo is set on an orange oval background. The word "HUGE" is written in large, white, sans-serif letters. To the right of the word "HUGE" are two Chinese characters.

005_510 (k) Summary

Date Summary Prepared: Feb. 24, 2017

Submitter Information:

NameShandong Huge Dental Material Corporation
AddressNo. 68 Shanhai Road, Donggang District, Rizhao City,Shandong Province, 276800, P.R. China
Telephone+86 633 2277285
Fax+86 633 2277298
ContactMr. Steven Song
E-mailzhangyj@hugedent.com

Device Name

Table 1: Device name
Trade nameGlass Ionomer Cement (Filling)
Common nameGlass Ionomer Cement
Classification nameDental Cement (21 C.F.R. 872.3275, Product Code: EMA)

Substantially Equivalent Device

Table 2: Predicate device information
Company NameDevice Name510 (k) No.
PrimaryPredicate DeviceGC AMERICA, INC.FUJI IIK980682
ReferencePredicate DeviceSILMET, LimitedProGlass Cements (ProGlass One, ProGlassTwo, ProGlass Two LC, ProGlass Nine,ProGlass Plus, ProGlass Silver)K101869
Shandong Huge DentalMaterial CorporationGlass Ionomer Cement (Luting)K161851

Description of Device

Glass Ionomer Cement (Filling) has two types: Filling II. It is a device composed of various materials including Silica, Alumina, Strontium Fluoride in the powder composition and Poly Carboxylic Acid in the liquid composition which intended to serve as a permanent or

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for "HUGE". The word "HUGE" is written in large, white, bold letters on an orange oval background. To the right of the word "HUGE" are two small Chinese characters.

temporary tooth filling or as base linings. It's a powder liquid system. The basic operation principle of Glass Ionomer Cement (Filling) is acid-base reaction. The physical properties of Glass Ionomer Cement (Filling) see page 5-4.

Intended for Use / including the indications for use

Self-curing glass ionomer based radiopaque filling cement, for dental use only.

  • (1) Restoration of primary teeth.
  • (2) Restorations of Class III, V and limited Class I cavities.
  • (3) Base linings under composite and amalgam.
  • (4) Core build-up.

Summary of Physical and Chemical Properties Tests

List of standards used:

ISO 9917-1:2007 Dentistry-- Water-based cements - Part 1: Powder/liquid acid-base cements

Table 4: Summary of physical and chemical properties test
ItemStandardRequirementsConclusion
ComponentsISO 9917-1:20075.2 Components5.2.1 LiquidFree from deposits or filaments on theinside of its container.No visible signs of gelation.5.2.2 PowderFree from extraneous material. If thepowder is coloured, the pigment shall beuniformly dispersed throughout thepowder.5.3 Unset cementHomogeneous and of a smoothconsistency.Within specset bystandard
Net setting timeISO 9917-1:20078.1 Net setting timeFor restoration: 1.5~6minWithin specset bystandard
CompressivestrengthISO 9917-1:20078.3 Compressive strengthFor restoration: ≥100MPaWithin specset bystandard
Acid erosionISO 9917-1:20078.4 Acid erosion≤0.17mmWithin specset bystandard
Acid-solubleLead contentISO 9917-1:20078.6.2 Acid-soluble Leadcontent≤100mg/kgWithin specset bystandard
Radio-opacityISO 9917-1:20078.7 Radio-opacityThe radio-opacity shall be at leastequivalent to that for the same thickness ofaluminium.Within specset bystandard

{5}------------------------------------------------

HUGE®

Shandong Huge Dental Material Corporation
www.hugedental.com T/+86 633 2277285 F/+86 633 2277298

No.68 Shanhai Road, Donggang District, Rizhao City
Shandong Province, 276800, P.R.China

echnological Characteristics

The new device has similar function, scientific concept, design, main materials and chemical composition as the predicate device

Comparison ItemsNew DevicePrimary Predicate DeviceReference Predicate Device
1)Regulatory ClassificationsGlass Ionomer Cement (Filling)21 C.F.R. 872.3275Product Code: EMAClass IIK98068221 C.F.R. 872.3275Product Code: EMAClass IIK10186921 C.F.R. 872.3275Product Code: EMAClass IIK16185121 C.F.R. 872.3275Product Code: EMAClass II
2)Intended Use-including the indications for useSelf-curing glass ionomer based radiopaque filling cement, for dental use only.(1) Restoration of primary teeth.(2) Restorations of Class III, V and limited Class I cavities.(3) Base linings under composite and amalgam.(4) Core build-up.FUJI II is a self-cure glass ionomer material intended for the use as a dental restorative for the following typical applications:1. Restoration of primary teeth.2. Core build-up.3. Class III, V and limited Class I cavities.Refer to page 5-6Self-curing glass ionomer based radiopaque luting cement, for dental use only. Recommended indications:(1) Cementation of metal-based inlays, onlays, crowns and bridges;(2) Cementation of high strength (zirconia based) all ceramic crowns and bridges;(3) Cementation of posts and screws made of metal or high-strength ceramic;(4) Cementation of orthodontic bands;(5) Restoration of caries in unstressed area (Luting II only).
3)Contraindications /Side Effects1. Pulp capping;2. In rare cases the product may cause sensitivity in some people. If any such reactions are experienced, discontinue the use of the product and refer to a1. Pulp capping;2. In rare cases the product may cause sensitivity in some people. If any such reactions are experienced, discontinue the use of the product and1. Pulp capping;2. In rare cases the product may cause sensitivity in some people. If any such reactions are experienced, discontinue the use of the product

{6}------------------------------------------------

Shandong Huge Dental Material Corp
www.hugedental.com T / +86 633 2277285 F / +86 633

HUGE®

No.68 Shanhai Road, Donggang District, Rizhao City,
Shandong Province, 276800, P.R.China

No.68 Shanhai Road, Donggang District, Rizhao City,
Shandong Province, 276800, P.R.China oration 3 2277298

Comparison ItemsNew DevicePrimary Predicate DeviceReference Predicate Device
Compositionof MaterialsGlass Ionomer Cement (Filling)K980682K101869K161851
Mainly composed of powder:Silica, Alumina, StrontiumFluoride; Poly Carboxylic AcidMainly composed of liquid:Poly Carboxylic Acid; DistilledWaterMainly composed of powder:Silica, Alumina, StrontiumFluoride; Poly CarboxylicAcidMainly composed of liquid:Poly Carboxylic Acid;Distilled WaterMainly composed ofpowder:Silica, Alumina,Strontium Fluoride;Poly Carboxylic AcidMainly composed ofliquid:Poly Carboxylic Acid;Distilled WaterMainly composed of powder:Silica, Alumina, Strontium Fluoride;Poly Carboxylic AcidMainly composed of liquid:Poly Carboxylic Acid; DistilledWater
Physical PropertiesFilling IFilling IIK980682K101869K161851
ItemsPowder /liquid (g/g):1.8~2.0/1.02.0~2.1/1.02.7/1.0
Mixing time (min., sec.):min. 1'00"min.1'00"min.1'00"
Working time (min., sec.):1'00"~2'00"1'00"~2'00"1'45"~2'00"Refer to page 5-7
Net setting time (min., sec.):1'30"~6'00"1'30"~6'00"1'30"~6'00"
Compressive strength (MPa):min.100min.100min.100
Standards MetISO 9917-1:1991(E)ISO 9917-1:2007;ISO 7405:2008;ISO 10993-1:2009/(R) 2013;ISO 10993-3:2014;ISO 10993-5:2009/(R) 2014;ISO 9917-1:2007;ISO 7405:2008;ISO 10993-1:2009/(R) 2013;ISO 10993-3:2014;ISO 10993-5:2009/(R) 2014;ISO 10993-10:2010;ISO 10993-11:2006/(R) 2010;ISO 14971:2007/(R) 2010;ISO 15223-1:2012;ISO 13485:2003ISO 9917-1:2007;ISO 7405:2008;ISO 10993-1:2009/(R) 2013;ISO 10993-3:2014;ISO 10993-5:2009/(R) 2014;ISO 10993-10:2010;ISO 10993-11:2006/(R) 2010;ISO 14971:2007/(R) 2010;ISO 15223-1:2012;ISO 13485:2003
ISO 9917-1:2003ISO 10993-10:2010;ISO 10993-11:2006/(R) 2010;ISO 14971:2007/(R) 2010;ISO 15223-1:2012;ISO 13485:2003

5-4

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for a company called HUGE. The word "HUGE" is written in large, white letters on an orange oval background. To the right of the word "HUGE" are two small, white Chinese characters. The logo is simple and eye-catching.

Summary of Biocompatibility

The new device. Glass Ionomer Cement (Filling), is substantially equivalent to the predicate devices that have been on the market for decades and with no clinical adverse events. The formulation of new device does not contain any new or non-conventional chemicals compared to the legally marketed predicate device.

We selected our Glass Ionomer Cement (Filling II, shade: A3) as the representative model in biocompatibility tests because it is the worst case scenario. In the meantime, the two types of our Glass Ionomer Cement (Filling) has the same chemical compositions, indications for use, function mechanism and manufacturing process.

Biocompatibility tests were performed to satisfy the ISO 10993 standards. The test items include Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Subchronic Toxicity and Genotoxicity.

Summary of Substantial Equivalence

As with the comparison shown in substantial equivalence discussion, these devices are same or similar in almost all aspects. The new device adds an indication for use compared to the primary predicate device (Base linings under composite and amalgam) which fall within the indication for use of the reference predicate device (K101869, "Base/liner"). The details of physical properties are slightly different, but these devices are in equivalent to other legally marketed devices of this type.

It can be seen that the minor differences between the new device and the predicate device are not of significance. Shandong Huge Dental Material Corporation concludes that Glass Jonomer Cement (Filling) is substantially equivalent to the predicate device.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "HUGE" in white letters on an orange oval background. To the right of the word "HUGE" are two Chinese characters, also in white. The overall design is simple and eye-catching, with the bright orange color providing a strong contrast to the white text.

Photo of the device

Image /page/8/Picture/4 description: The image shows a dental cement kit by HUGE. The kit includes a box labeled "Glass Ionomer Cement (Filling) 1-1 PKG. Powder 15g Liquid 12g (10ml)", a bottle labeled "Glass Ionomer Cement (F) Powder 15g", a smaller bottle, a mixing pad, and a spatula. The box has text in multiple languages, including English, French, and Arabic.

Indications for use of the reference predicate device (K101869)

Indications for use

ProductsIndications for Use
ProGlass OneCementation of all types of metal, porcelain fused to metal, resin crowns, inlays,onlays & bridgesCementation of orthodontic bandsCementation of stainless steel crowns or orthodontic appliances retained with stainlesssteel crownsBase/liner
ProGlass TwoClass III, V and limited class I cavitiesRestoration of primary teethCore Build Up
ProGlass Two LCClass III and V restorationsRestoration of Cervical erosions and root surface cariesCore Build UpBase/Liner
ProGlass NineClass I & II cavitiesDecidious teeth: final restorative for Class I, II and VLong term restorative in non-load bearing areas of Class I, II and VIntermediate restorative & sandwich material for heavy stress bearingCore build up material
ProGlass PlusMetal-based restorationsCeramic inlaysReinforced ceramic crowns and bridgesAll kinds of acrylic/resin crowns, inlays, onlays and bridges
ProGlass SilverClass I, limited Class II, temporary fillingsRestoration of primary teethCore Build UpBase/Liner

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image features the word "HUGE" in bold, white letters against an orange, oval-shaped background. To the right of the word "HUGE" are two small Chinese characters, also in white. The overall design is simple and eye-catching, with the contrast between the white text and orange background making the word "HUGE" stand out.

Physical characteristics of the reference predicate device (K101869)

ProGlassOneProGlassTwoProGlassTwo LCProGlassNineProGlassPlusProGlassSilver
Powder /liquid2.4 / 1.03.5 / 1.02.3 /1.04.1 / 1.01.5 / 1.04.0 / 1.0
Mixing time (sec)30"30"30"30"30"20"
Working time2' 30" - 3"1' 30" - 2'3'2' 30"3'1'40"
Setting Time (min.sec)3'10"-3' 10" - 3'3' -3' 30"3'4'
Light Cure (sec)20"

Physical characteristics of the reference predicate device (K161851)

ItemsLuting ILuting IILuting II(only as restorationmaterials)
Powder /liquid (g/g):1.6~1.8/1.01.5~1.7/1.02.1~2.2/1.0
Mixing time (min., sec.):45"45"45"
Working time (min., sec.):≤3'00"2'00"~3'00"1'30"~2'00"
Net setting time (min., sec.):1'30"~8'00"1'30"~8'00"1'30"~6'00"
Film thickness (μ m):≤25≤25N/A
Compressive strength (MPa):≥50≥50≥100

§ 872.3275 Dental cement.

(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.