K Number
K160317

Validate with FDA (Live)

Date Cleared
2016-02-29

(24 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

syngo.via MI Workflows are medical diagnostic applications for viewing, manipulation, 3Dvisualization and comparison of medical images from multiple imaging modalities and/or multiple time-points. The application supports functional data, such as PET or SPECT as well as anatomical datasets, such as CT or MR.

syngo.via MI Workflows enable visualization of information that would otherwise have to be visually compared disjointedly. syngo.via MI Workflows provide analytical tools to help the user assess, and document changes in morphological or functional activity at diagnostic and therapy follow-up examinations. syngo.via MI Workflows can perform harmonization of SUV (PET) across different PET systems or different reconstruction methods.

syngo via MI workflows support the interpretation and evaluations and follow up documentation of findings within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments.

Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. syngo.via MI Workflows are a complement to these standard procedures.

Device Description

The Total Lesion Glycolysis (TLG) feature resides within synqo.via MI Workflows and is a software only medical device which will be delivered on CD-ROM / DVD to be installed onto the commercially available Siemens syngo.via software platform by trained service personnel.

syngo.via MI Workflows is a medical diagnostic application for viewing, manipulation, 3Dvisualization and comparison of medical images from multiple imaqinq modalities and/or multiple time-points. The application supports functional data, such as PET or SPECT as well as anatomical datasets, such as CT or MR. The images can be viewed in a number of output formats including MIP and volume rendering.

synqo.via MI Workflows enable visualization of information that would otherwise have to be visually compared disjointedly. synqo.via MI Workflows provide analytical tools to help the user assess, and document changes in morphological or functional activity at diagnostic and therapy follow-up examinations. They additionally support the interpretation and evaluation of examinations and follow up documentation of findings within healthcare institutions, for example, in Radiology (Oncology), Nuclear Medicine and Cardiology environments.

The modifications to the syngo.via MI Workflows (K151192) include the addition of the Total Lesion Glycolysis (TLG) feature within the MM Oncology workflow. This feature will provide the ability for the user to calculate and display the TLG for a given volume of interest (VOI) in addition to calculating and trending the summed Molecular Tumor Volume (MTV) for all VOIs defined for the given PET series.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for a medical device called "syngo.via MI Workflows – Total Lesion Glycolysis (TLG) feature." This document describes the device, its indications for use, and claims substantial equivalence to a previously cleared device (K151192).

However, the document does not contain the detailed information required to describe the acceptance criteria and the specific study that proves the device meets those criteria, as typically found in a clinical study report. The document states that "Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values." This is a general statement and does not provide specific acceptance criteria or the study details.

Therefore, I cannot fully answer your request based on the provided text. I can, however, extract what little information is present about performance and testing.

Here's an analysis of the provided text, highlighting what can and cannot be answered:


Acceptance Criteria and Study for syngo.via MI Workflows – Total Lesion Glycolysis (TLG) feature

Based on the provided FDA 510(k) summary, specific acceptance criteria and detailed study information are not explicitly stated in a quantifiable manner (e.g., sensitivity, specificity, accuracy thresholds with corresponding confidence intervals). The document primarily focuses on demonstrating substantial equivalence to a predicate device and outlining general software validation processes.

1. Table of Acceptance Criteria and Reported Device Performance

Cannot be fully answered: The document does not provide a table of quantifiable acceptance criteria or reported device performance metrics such as sensitivity, specificity, or accuracy for the TLG feature. It only broadly states: "All testing has met the predetermined acceptance values."

2. Sample Size Used for the Test Set and Data Provenance

Cannot be answered: The document does not specify the sample size used for any test set or the provenance of any data (e.g., country of origin, retrospective or prospective) used for performance evaluation of the TLG feature.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

Cannot be answered: The document does not mention the number of experts or their qualifications used to establish ground truth for any testing of the TLG feature.

4. Adjudication Method for the Test Set

Cannot be answered: The document does not describe any adjudication method (e.g., 2+1, 3+1) used for establishing ground truth or evaluating the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Cannot be answered: The document does not indicate that an MRMC comparative effectiveness study was performed. There is no mention of human readers improving with or without AI assistance, nor any effect size.

6. Standalone (Algorithm Only) Performance Study

Partially Answered: The device is described as "software only" and provides "analytical tools to help the user assess, and document changes." This implies that the TLG feature itself performs calculations and provides numerical results. The general statement "Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met" suggests that the standalone performance of the algorithm for calculating TLG was evaluated to ensure its functionality and accuracy against specified requirements. However, no specific metrics or study details for this standalone performance are provided.

7. Type of Ground Truth Used

Cannot be answered: The document does not specify the type of ground truth used for any testing (e.g., expert consensus, pathology, outcomes data). In the context of a calculation feature like TLG, ground truth would typically refer to a validated reference method for calculating TLG, but this is not detailed.

8. Sample Size for the Training Set

Cannot be answered: The document does not mention a training set or its sample size. This is a software feature update to an existing platform, and typically training sets are more relevant for machine learning algorithms which are not explicitly described as being part of this specific TLG feature's core functionality beyond calculation.

9. How the Ground Truth for the Training Set Was Established

Cannot be answered: Since no training set is mentioned, how its ground truth was established cannot be answered.


Summary of Available Information from the Document:

The 510(k) summary for syngo.via MI Workflows – Total Lesion Glycolysis (TLG) feature focuses on demonstrating substantial equivalence to a predicate device (K151192) and confirming general software validation practices. It explicitly states:

  • "The features introduced into syngo.via VB10B had no impact on the technological characteristics already present in the commercially available predicate system."
  • "There are no differences in the Indications for Use or Fundamental Technological Characteristics of the synqo.via MI Workflows as compared to the currently commercially available software (K151192)."
  • "Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values."

This document serves as an FDA clearance declaration based on substantial equivalence, not a detailed technical report of a new device's performance validation study.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines that resemble ribbons or waves. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 29, 2016

Siemens Medical Solutions USA, Inc. % Ms. Veronica Padharia Regulatory Affairs Specialist 2501 N. Barrington Road HOFFMAN ESTATES IL 60192

Re: K160317

Trade/Device Name: syngo.via MI Workflows Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 3, 2016 Received: February 5, 2016

Dear Ms. Padharia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K160317

Device Name syngo.via MI Workflows

Indications for Use (Describe)

syngo.via MI Workflows are medical diagnostic applications for viewing, manipulation, 3D- visualization and comparison of medical images from multiple imaging modalities and/or multiple time-points. The application supports functional data, such as PET or SPECT as well as anatomical datasets, such as CT or MR.

syngo.via MI Workflows enable visualization of information that would otherwise have to be visually compared disjointedly. syngo.via MI Workflows provide analytical tools to help the user assess, and document changes in morphological or functional activity at diagnostic and therapy follow-up examinations. syngo.via MI Workflows can perform harmonization of SUV (PET) across different PET systems or different reconstruction methods.

syngo via MI workflows support the interpretation and evaluations and follow up documentation of findings within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments.

Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. syngo.via MI Workflows are a complement to these standard procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

as required by 21 CFR Part 807.87(h)

Identification of the Submitter

Primary Contact:Alternate Contact:
Submitter:Veronica PadhariaRegulatory Affairs SpecialistSiemens Medical SolutionsUSA, Inc.Molecular Imaging2501 Barrington RoadHoffman Estates, IL 60192Cynthia BuschRegulatory Technical SpecialistSiemens Medical SolutionsUSA, Inc.Molecular Imaging2501 Barrington RoadHoffman Estates, IL 60192
Telephone Number:(630) 877 – 5761(847) 304 - 7095
Fax Number:(847) 304 – 6023(847) 304 - 6023
Name / Address ofManufacturerSiemens Medical Solutions USA, IncMolecular Imaging2501 N. Barrington RoadHoffman Estates, IL 60192USA
Date of Submission:February 3, 2016
Identification of the product
Device Proprietary Name:syngo.via MI Workflows – Total Lesion Glycolysis (TLG) feature
Common Name:Image Processing Software
Classification Name:Picture Archiving and Communication System per 21 CFR892.2050
Product Code:LLZ
Classification Panel:Radiology
Device Class:Class II

{4}------------------------------------------------

Marketed Devices to which Equivalence is claimed

DeviceManufacturer510(k) Number
syngo .via MIWorkflows VB10ASiemens Medical Solutions USA, IncK151192 (June 2015)

Device Description

The Total Lesion Glycolysis (TLG) feature resides within synqo.via MI Workflows and is a software only medical device which will be delivered on CD-ROM / DVD to be installed onto the commercially available Siemens syngo.via software platform by trained service personnel.

syngo.via MI Workflows is a medical diagnostic application for viewing, manipulation, 3Dvisualization and comparison of medical images from multiple imaqinq modalities and/or multiple time-points. The application supports functional data, such as PET or SPECT as well as anatomical datasets, such as CT or MR. The images can be viewed in a number of output formats including MIP and volume rendering.

synqo.via MI Workflows enable visualization of information that would otherwise have to be visually compared disjointedly. synqo.via MI Workflows provide analytical tools to help the user assess, and document changes in morphological or functional activity at diagnostic and therapy follow-up examinations. They additionally support the interpretation and evaluation of examinations and follow up documentation of findings within healthcare institutions, for example, in Radiology (Oncology), Nuclear Medicine and Cardiology environments.

The modifications to the syngo.via MI Workflows (K151192) include the addition of the Total Lesion Glycolysis (TLG) feature within the MM Oncology workflow. This feature will provide the ability for the user to calculate and display the TLG for a given volume of interest (VOI) in addition to calculating and trending the summed Molecular Tumor Volume (MTV) for all VOIs defined for the given PET series.

Technological Characteristics

The syngo.via MI Workflows VB10B software modifications are based on the commercially available syngo.via MI Workflows VB10A software (K151192). The features introduced into syngo.via VB10B had no impact on the technological characteristics already present in the commercially available predicate system.

syngo.via MI Workflows is intended to be run on the Siemens syngo.via software platform (K150843) either alone or with other advanced commercially cleared applications.

Performance Testing / Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.

Risk Management has been ensured via risk analyses in compliance with ISO 14971:2012 to identify and provide mitigation to potential hazards beginning early in the design cycle and

{5}------------------------------------------------

continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recoqnized and established industry standards for development including EN ISO 13485 and IEC 62304.

Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.

The device is designed and manufactured in accordance with Quality System Requlations as outlined in 21 CFR 820. The FDA recognized standards are listed as follows:

  • Recognition Number 13-8: IEC 62304 First Edition 2006-05
  • Recognition Number 12-238: NEMA PS 3.1 3.20 (2011) ●
  • o Recognition Number 5-40: ISO 14971 Second Edition 2007-03-01
  • Recognition Number 5-95: IEC 62366-1 Edition 1.0 2015-02
  • o Recognition Number 5-90: ISO 15223-1 Second Edition 2012-07-01

Indications for Use

syngo.via MI Workflows are medical diagnostic applications for viewing, manipulation, 3Dvisualization and comparison of medical images from multiple imaging modalities and/or multiple time-points. The application supports functional data, such as PET or SPECT as well as anatomical datasets, such as CT or MR.

syngo.via MI Workflows enable visualization of information that would otherwise have to be visually compared disjointedly. synqo.via MI Workflows provide analytical tools to help the user assess, and document changes in morphological or functional activity at diagnostic and therapy follow-up examinations. syngo.via MI Workflows can perform harmonization of SUV (PET) across different PET systems or different reconstruction methods.

syngo.via MI Workflows support the interpretation and evaluation of examinations and follow up documentation of findings within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments.

Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unreqistered images. syngo.via MI Workflows are a complement to these standard procedures.

Statement Regarding Substantial Equivalence:

There are no differences in the Indications for Use or Fundamental Technological Characteristics of the synqo.via MI Workflows as compared to the currently commercially available software (K151192). Both devices are used for viewing, manipulation, 3Dvisualization and comparison of medical images from multiple imaging modalities and/or multiple time-points

{6}------------------------------------------------

Additionally, there have been no changes that raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens opinion that the syngo.via MI Workflows software with the modifications outlined in this application is substantially equivalent to the predicate device.

N/A