K Number
K160309

Validate with FDA (Live)

Device Name
Neurosoft TMS
Manufacturer
Date Cleared
2016-12-22

(322 days)

Product Code
Regulation Number
882.5805
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Neurosoft TMS is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

Device Description

The Neurosoft TMS is a repetitive transcranial magnetic stimulation (rTMS) system. This computerized medical device produces non-invasive, repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex for the treatment of Major Depressive Disorder.

AI/ML Overview

This document describes a 510(k) premarket notification for the Neurosoft TMS device. The core of this submission is to demonstrate substantial equivalence to predicate devices, namely the Tonica Electronik A/S MagVita TMS Therapy System and the Magstim Company Limited Rapid2 Therapy System. The document does not describe an acceptance criteria study with direct device performance measurement against these criteria, but instead relies on showing feature-by-feature similarity to already-approved predicate devices.

Therefore, the requested information elements cannot be fully populated as they would typically apply to a clinical trial or performance study against predefined metrics. However, an interpretation of "acceptance criteria" and "reported device performance" in this context would be the comparative features and specifications between the Neurosoft TMS and its predicate devices, as presented in Table 5-1.

Here's an analysis based on the provided text, addressing the points where information is available or can be inferred:

1. A table of acceptance criteria and the reported device performance

In the context of a 510(k) submission showing substantial equivalence, the "acceptance criteria" are implied by the characteristics of the predicate devices. The "reported device performance" is the Neurosoft TMS's specifications, which are presented as being equivalent, or not raising new safety/efficacy concerns, compared to the predicates.

Criteria (Implied "Acceptance Criteria" from Predicate Devices)Neurosoft TMS Reported Performance (Substantial Equivalence Claim)
Intended Use (Treatment of MDD in adult patients who failed prior antidepressant medication)Identical
Magnetic Field Intensity (120% of MT)120% of MT
Frequency (10 Hz)10 Hz
Train duration (4 sec)4 sec
Inter-train interval (26 sec)26 sec
Number of trains (75)75
Magnetic Pulses per Session (3000)3000
Treatment Session Duration (~37.0 - 37.5 min)~37.0 min
Sessions/week (5)5
Treatment Schedule (5 daily sessions for 6 weeks)5 daily sessions for 6 weeks
Area of brain to be stimulated (Frontal Cortex)Frontal Cortex
Coil Configuration (Figure-of-eight coil)Figure-of-eight coil
Coil Core material (Air core)Air core
Waveform (Biphasic sinusoid)Biphasic sinusoid
Active pulse width (μs) (~290-300 µs)280 µs (claimed to be substantially equivalent)
Max initial dB/dt (kT/s) near the coil surface (Various, e.g., 35-36 kT/s for predicates)FEC-02-100-C 25, AFEC-02-100-C 38, FEC-02-100 25, AFEC-02-100 32 (claimed to be substantially equivalent)
Coil temperature auto-disable (40-41 °C)41 °C (106 °F)
Electrical safety standards compliance (IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2)AAMI/ANSI ES 60601-1:2005/(R)2012 and IEC 60601-1-2
ISO Standards met (ISO 13485, ISO 10993-1, ISO 14971)DIN EN ISO 13485: 2012, ISO 10993-1: 2009, ISO 14971: 2007

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily describes non-clinical performance data for the Neurosoft TMS focusing on electrical safety, EMC, biocompatibility, software verification/validation, risk analysis, and magnetic field characteristics. There is no mention of a "test set" in the context of a clinical study, as the submission relies on substantial equivalence to predicate devices, which are already approved for clinical use.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This document does not describe a clinical study where ground truth would be established by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As no clinical test set is described, no adjudication method would be required.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a Repetitive Transcranial Magnetic Stimulation (rTMS) system, a treatment device, not an image analysis or diagnostic aid involving human readers or AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document describes non-clinical performance data for the Neurosoft TMS:

  • Electrical safety and electromagnetic compatibility testing (compliant with AAMI/ANSI ES 60601-1:2005/(R)2012 and IEC 60601-1-2: 2007).
  • Biocompatibility evaluation of coils (meet ISO 10993-1 (2009)).
  • Software verification and validation testing (performs as intended and in accordance with specifications, described in section 16, not provided here).
  • Risk analysis (compliance with ISO14971).
  • Magnetic field characteristics (plots and acoustic output measurements conducted as per FDA guidance).

These are performed on the device itself and its components, representing "standalone" performance in terms of its technical and safety specifications.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical tests:

  • Electrical safety/EMC: Ground truth is defined by the requirements of the cited standards (e.g., AAMI/ANSI ES 60601-1, IEC 60601-1-2).
  • Biocompatibility: Ground truth is defined by the requirements of ISO 10993-1 (2009).
  • Software verification/validation: Ground truth is the defined software specifications and requirements.
  • Magnetic field characteristics: Ground truth is derived from physical measurements and comparison to established parameters of predicate devices and FDA guidance.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI-based diagnostic device that requires a training set of data.

9. How the ground truth for the training set was established

Not applicable. No training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to create a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 22, 2016

TeleEMG, LLC % Barry Ashar President MakroMed, Inc. 88 Stiles Road Salem, New Hampshire 03079

Re: K160309

Trade/Device Name: Neurosoft TMS Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: January 26, 2016 Received: February 4, 2016

Dear Barry Ashar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

William J.

Heetderks -A

Digitally signed by William J. Heetderks -A
DN: c=US, o=U.S. Government, ou=HHS,
ou=NIH, ou=People,
0.9.2342.19200300.100.1.1=0010149848,
cn=William J. Heetderks -A
Date: 2016.12.22 14:55:04 -05'00'

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160309

Device Name Neurosoft TMS

Indications for Use (Describe)

The Neurosoft TMS is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for the TeleEMG, LLC Neurosoft TMS

(per 21 CFR 807.92 and http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm)

1. SUBMITTER/510(K) HOLDER

TeleEMG. LLC 27 Arlington Rd., Building 2, Unit 1 Woburn, MA 01801, USA

Contact Person: Barry V. Ashar, Makromed, Inc. Telephone: (603) 890-3311 Date Prepared: November 16, 2016

2. DEVICE NAME

Proprietary Name:Neurosoft TMS
Regulation Name:Repetitive transcranial magnetic stimulation system
Regulation Number:21 CFR §882.5805
Classification Name:Transcranial Magnetic Stimulator
Device Class:Class II
Product Code:OBP

(Note: During its development Neurosoft TMS was also known as Neuro-MS/D. As a result, many of its documents may be titled with Neuro-MS/D name. It is our intention to commercialize it as Neurosoft TMS. All product labeling now refer to this product as Neurosoft TMS. Some of the support documents included in this submission may still refer to Neuro-MS/D. The two names are fully interchangeable as they represent one and the same product)

3. PREDICATE DEVICES

  • Tonica Electronik A/S, MagVita TMS Therapy System, K150641
  • Magstim Company Limited, Rapid2 Therapy System, K143531

4. DEVICE DESCRIPTION

The Neurosoft TMS is a repetitive transcranial magnetic stimulation (rTMS) system. This computerized medical device produces non-invasive, repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex for the treatment of Major Depressive Disorder.

{4}------------------------------------------------

The Neurosoft TMS principle of operation is based on the discharge of high voltage capacitor (1.8 kV) through stimulation coil; the pulsed magnetic field generated by the discharge current (up to 10 kA) penetrates through neuromuscular tissues nearby to induce electrical currents in cortical neurons.

The Neurosoft TMS consists of the following main components:

  • . Main unit of the magnetic stimulator
  • Cooling unit ●
  • Extra power supply unit ●
  • Coils
    • Cooled figure-of-eight coil FEC-02-100-C O
    • O Cooled figure-of-eight coil AFEC-02-100-C
    • Figure-of-eight coil FEC-02-100 (optional) o
    • o Figure-of-eight coil AFEC-02-100 (optional)
  • . K8 coil holder
  • K3 flexible arm for coil positioning
  • Trolley with casters ●

5. INTENDED USE

The Neurosoft TMS system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

6. STANDARDS

The Neurosoft TMS system has been tested and complies with the following standards:

  • DIN EN ISO 13485: 2012 .
  • ISO 10993-1: 2009 .
  • ISO 14971: 2007 .
  • AAMI/ANSI ES 60601-1:2005/(R)2012 ●
  • IEC 60601-1-2: 2007 .

7. NON-CLINICAL PERFORMANCE DATA

Electrical safety and electromagnetic compatibility testing demonstrate that the Neurosoft TMS is compliant with AAMI/ANSI ES 60601-1:2005/(R)2012 and IEC 60601-1-2: 2007.

Only the coils in the delivery set of the Neurosoft TMS have direct contact with the patients. The biocompatibility evaluation demonstrates that the coils meet ISO 10993-1 (2009) standards.

{5}------------------------------------------------

Software verification and validation testing is described section 16 "Software" following the corresponding FDA software guidelines. It demonstrates that the software performs as intended and in accordance with specifications. The potential risks of the Neurosoft TMS have been identified and evaluated in compliance with ISO14971, and the risks were determined to be acceptable, or have been addressed with risk control measures.

Additionally, the non-clinical testing with the Neurosoft TMS included testing of the magnetic field characteristics of the system, as required by FDA's guidance document "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems". The magnetic field plots and acoustic output measurements were conducted to demonstrate safety and performance.

8. SUBSTANTIAL EQUIVALENCE

The Neurosoft TMS and the predicate devices have identical intended use /indication for use (see table 5-1), and the technological characteristics are very similar such that they in our view can be considered equivalent.

The design of the Neurosoft TMS is similar to that of the MagVita TMS Therapy System and the Rapid2 Therapy System, as all systems are based on applying transcranial magnetic stimulation by means of repetitive pulse trains at a predetermined frequency. All systems use the same mechanism of action, i.e., an electromechanical instrument that produces and delivers brief duration, rapidly alternating (pulsed) magnetic fields to induce electrical currents in localized regions of the prefrontal cortex.

The basic software capabilities related to treatment administration in the Neurosoft TMS are the same as these in the predicate devices.

The Neurosoft TMS and the predicate devices have the same components consisting of TMS stimulator with software, electromagnetic coil and a flexible arm for positioning of the treatment coil. The basic operational procedures including system setup, patient preparations, motor threshold determination, coil positioning and treatment with predefined treatment stimulation parameters are essentially the same. A thorough comparison among the Neurosoft TMS and the predicate devices is shown in a tabular form (see table 5-1) below:

{6}------------------------------------------------

CriteriaNeurosoft TMSMagVita TMS Therapy SystemRapid² Therapy System
Intended UseThe Neurosoft TMS is indicatedfor the treatment of MajorDepressive Disorder in adultpatients who have failed toreceive satisfactory improvementfrom prior antidepressantmedication in the currentepisode.The MagVita TMSTherapy System isindicated for the treatmentof Major DepressiveDisorder in adult patientswho have failed to receivesatisfactory improvementfrom prior antidepressantmedication in the currentepisode.The Rapid² TherapySystem is indicated forthe treatment of MajorDepressive Disorder inadult patients who havefailed to achievesatisfactory improvementfrom prior antidepressantmedication in the currentepisode.
Recommended standard treatment
Magnetic FieldIntensity120% of the MT120% of the MT120% of the MT
Frequency10 Hz10 Hz10 Hz
Train duration4 sec4 sec4 sec
Inter-train interval26 sec26 sec26 sec
Number of trains757575
Magnetic Pulsesper Session300030003000
Treatment SessionDuration~ 37.0 min37.0 min37.5 min
Sessions/week555
TreatmentSchedule5 daily sessions for 6 weeks5 daily sessions for 6weeks5 daily sessions for 6weeks
Area of brain to bestimulatedFrontal CortexFrontal CortexFrontal Cortex
Coils
Coils (includingoptionalaccessories)FEC-02-100-C, AFEC-02-100-CFEC-02-100 (optional), AFEC-02-100 (optional)C-B60C-B65Magstim Double 70mmAir Film Coil
ConfigurationFigure-of-eight coilFigure-of-eight coilFigure-of-eight coil
Core materialAir coreAir coreAir core
CoolingFEC-02-100-C & AFEC-02-100-C: Liquid coolingC-B65: Liquid coolingForced Air
FEC-02-100 & AFEC-02-100(optional accessories): NoneC-B60: None
Coil parametersFEC-02-100-CInner diameter - 47x50 mm1Outer diameter - 97x100 mm1Coil Cool-B65Inner diameter - 35 mmOuter diameter - 75 mmArea = 33000 mm²Average Inductance 12 μH
CriteriaNeurosoft TMSMagVita TMS Therapy SystemRapid2 Therapy System
Area = 184725 mm²Average Inductance 10 μHFlat spiral windingN = 16 (4 turns/wing x 2 wings x 2 layers)AFEC-02-100-CInner diameter - 36x51 mm1Outer diameter - 84x106 mm1Area = 184725 mm²Average Inductance 10 μHFlat spiral windingN = 16 (4 turns/wing x 2 wings x 2 layers)Winding height - 12 mmN = 2x (2 x 5)Flat spiral windingN = 3x19 turns/wing x 2wings
FEC-02-100Inner diameter - 47x50 mm1Outer diameter - 97x100 mm1Area = 91560 mm²Average Inductance 10 μHFlat spiral windingN = 16 (4 turns/wing x 2 wings x 2 layers)AFEC-02-100Inner diameter - 36x51 mm1Outer diameter - 84x106 mm1Area = 87200 mm²Average Inductance 10 μHFlat spiral windingN = 16 (4 turns/wing x 2 wings x 2 layers)Coil C-B60Inner diameter - 35 mmOuter diameter - 75 mmWinding height - 12 mmN = 2x (2 x 5)
Machine Output Parameters
Amplitude inStandard MotorThreshold (SMT)unitsFEC-02-100-C 0 - 1.89AFEC-02-100-C 0 - 2.38FEC-02-100 0 - 1.92AFEC-02-100 0 - 2.330 - 1.70.28 - 1.9
WaveformBiphasic sinusoidBiphasic sinusoidBiphasic sinusoid
Active pulse width(μs)280290300
Max initial dB/dt(kT/s) near the coilsurfaceFEC-02-100-C 25AFEC-02-100-C 38FEC-02-100 25AFEC-02-100 32C-B65: 36C-B60: 35N/A
The system willautomatically bedisabled when thecoil temperatureexceeds:41 °C (106 °F)41 °C (106 °F)40 °C (104 °F)
Frequency range0.1 - 30 (Stand-alone)0.1 - 30 or 0.1 - 100,0.1 - 30
CriteriaNeurosoft TMSMagVita TMS TherapySystemRapid2 Therapy System
(Hz)0.1 - 100 (with PC)depending on model
Pulse trainduration range (s)0.5 - 100Rep Rate: 0.1 ...100HzPulses in Train: 1,2,3,4 ...1000Train duration = Pulses inTrain / Rep Rate1 - 20
Inter-train intervalrange (s)0 - 3001 - 12010 - 60
Maximum trainsper session4800 = 2400 s [max session] /(0.5 s [min train]+ 0 s [minpause])500~140
Maximum numberof pulses persession72,000(Stand-alone) = 2400 s[max session] *30 Hz240,000(with PC) = 2400 s [maxsession] *100 Hz500,000 = 1,000 (pulsesmax per train) x 500 (trainsmax per session)65,000
Standards
ElectricalsafetyComplies with AAMI/ANSI ES60601-1:2005/(R)2012 and IEC60601-1-2Complies withIEC 60601-1, IEC 60601-1-1 and IEC 60601-1-2.Complies withEN 60601-1 andEN 60601-1-2
ISO StandardsmetCompany complies with DIN ENISO 13485: 2012ISO 10993-1: 2009ISO 14971: 2007Company complies withISO 13485:2012.Company complies withISO 13485: 2003ISO 10993-1: 2009ISO 14971

Table 5-1. Side-by-Side Comparison of the Proposed Device with Cited Predicate Devices

{7}------------------------------------------------

{8}------------------------------------------------

1 see Fig. 1

Image /page/8/Figure/2 description: The image shows a diagram with several measurements labeled. The diagram appears to be a technical drawing of a mechanical component. The measurements are labeled as 'A', 'B', 'a', and 'b'.

Image /page/8/Figure/3 description: The image shows the text "Fig.1 Coil for magnetic stimulator". The text is in black font and is centered in the image. The text is likely a figure caption for a scientific paper or presentation.

9. CONCLUSION

The Neurosoft TMS and the predicate devices have identical intended use /indication for use, target population, treatment procedure, treatment position and all recommended standard

{9}------------------------------------------------

treatment protocol parameters (intensity, frequency, number of pulses in a train, number of trains in a session, number of treatment sessions).

All coils compared in Table 5.1 share the same transducer design (figure-of-eight). The tested magnetic properties of the Neurosoft TMS and the predicate devices are substantial equivalent for the coils.

The reliability of the positioning method used by the Neurosoft TMS is based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, as is the method used in the predicate devices. The method for identifying the correct treatment position in the Neurosoft TMS is at least as effective as the method employed by the predicate devices.

The Neurosoft TMS does not introduce any new safety considerations in comparison to the predicate device. All other identified differences between the two systems are minor and without any known impact on safety or efficacy.

Based on the information and supporting documentation provided in the premarket notification, the Neurosoft TMS is substantially equivalent to the cited predicate devices. Testing demonstrates that the Neurosoft TMS fulfills prospectively defined design and performance specifications.

§ 882.5805 Repetitive transcranial magnetic stimulation system.

(a)
Identification. A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.(b)
Classification. Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.