K Number
K160305

Validate with FDA (Live)

Date Cleared
2016-05-27

(113 days)

Product Code
Regulation Number
880.5700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The neoBLUE® LED Phototherapy System is indicated for the treatment of hyperbilirubinemia for neonates and infants in a hospital environment, and administered by trained, professional medical staff, on the order of a licensed medical practitioner. The light can be used with a bassinet, incubator, open bed, or radiant warmer.

Device Description

The neoBLUE® Phototherapy System consists of two products – the neoBLUE LED Phototherapy light source (light) and the neoBLUE LED Phototherapy roll stand.

The neoBLUE LED Phototherapy System is a floor-standing, mobile phototherapy light that delivers a narrow band of high-intensity blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia. The light can be used independently of the roll stand, and can be placed directly on an incubator with a flattopped surface.

AI/ML Overview

This document describes the Natus Medical Incorporated neoBLUE® LED Phototherapy System (K160305).

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly present acceptance criteria in a tabular format with corresponding reported device performance metrics in the way one might see for an AI algorithm's diagnostic accuracy. Instead, the document focuses on compliance with established medical device standards and functional specifications for a phototherapy unit.

However, based on the non-clinical testing section, the implicit acceptance criteria are compliance with relevant safety and performance standards for medical electrical equipment and phototherapy devices.

Acceptance CriterionReported Device Performance
Safety Standards Compliance
IEC 60601-2-50:2009-03 Edition 2 (Infant Phototherapy Equipment)Passed all testing and complies.
IEC 60601-1:2005/(R)2012 and C1:2009/(R)2012 (General Requirements for Basic Safety)Passed all testing and complies.
IEC 60601-1-2: 2007/(R)2012 (Electromagnetic Compatibility)Passed all testing and complies.
IEC 60601-1-6:2013-10 Edition 3.1 (Usability)Passed all testing and complies.
Performance Specifications
Intensity Setting (High)35 µW/cm²/nm at 12 inches (30.5 cm)
Intensity Setting (Low)15 µW/cm²/nm at 12 inches (30.5 cm)
Light Emission SpectrumBlue LEDs emit 400-550 nm (peak 450-475 nm)
Non-significant UV/IR emissionReduce risk of skin damage and excessive warming
Equivalence to Predicate Device
Therapeutic OutputSame as predicate (K022196)
Functional Design ConsiderationsSeparation of light source from roll stand for independent use; addition of timer for LED working hours

2. Sample Size Used for the Test Set and Data Provenance

This is not applicable as the device is a phototherapy system and not a diagnostic AI algorithm that relies on a "test set" of patient data for performance evaluation in the typical sense. The testing performed was for compliance with electrical, safety, and performance standards, and for demonstrating substantial equivalence to a predicate device through engineering and design validation. The provenance in this context would refer to internal validation procedures and results.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This is not applicable. The device's performance is not evaluated against a "ground truth" derived from expert consensus on patient data. Instead, it is evaluated against established engineering specifications, safety standards, and functional requirements.

4. Adjudication Method

This is not applicable. Adjudication methods are typically used in studies involving human interpretation or subjective assessment where discrepancies need to be resolved (e.g., in clinical trials or diagnostic accuracy studies involving multiple readers). This document describes the regulatory clearance of a medical device based on compliance with standards and functional testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable. The neoBLUE® LED Phototherapy System is a treatment device, not a diagnostic AI tool intended to assist human readers in interpreting medical images or data. Therefore, an MRMC study related to reader performance is irrelevant to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is not applicable. The device is a standalone phototherapy unit; it is not an algorithm, and its function does not involve human interpretation of its output in the way an AI diagnostic tool would.

7. The Type of Ground Truth Used

The "ground truth" for this device's acceptance is based on:

  • Compliance with recognized international and national standards for medical electrical equipment and infant phototherapy.
  • Engineering specifications for light intensity, wavelength, and safety.
  • Demonstration of substantial equivalence to a legally marketed predicate device (K022196 Natus Blue Light Phototherapy Unit) in terms of intended use, operating principles, and therapeutic output.

8. The Sample Size for the Training Set

This is not applicable. The device is not an AI algorithm that undergoes a "training" phase with a dataset.

9. How the Ground Truth for the Training Set Was Established

This is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles suggesting a sense of community or collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 27, 2016

Natus Medical Incorporated Ms. Judy Buckham Regulatory Affairs Specialist 5900 First Avenue South Seattle, Washington 98108

Re: K160305

Trade/Device Name: neoBLUE® LED Phototherapy System Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: Class II Product Code: LBI. Dated: April 14, 2016 Received: April 18, 2016

Dear Ms. Buckham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang
-s

for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K160305

Device Name

neoBLUE® LED Phototherapy System

Indications for Use (Describe)

Indications for Use: The neoBLUE® LED Phototherapy System is indicated for the treatment of hyperbilirubinemia for neonates and infants in a hospital environment, and administered by trained, professional medical staff, on the order of a licensed medical practitioner. The light can be used with a bassinet, incubator, open bed, or radiant warmer.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

K160305 510(K) SUMMARY

Manufacturer'sName:Natus Medical Incorporated5900 First Avenue SouthSeattle, WA 98108
CorrespondingOfficial:Judy BuckhamRegulatory Affairs Specialist
Telephone206 268 5187
Number:Fax Number:206 268 5104
Preparation Date:April 14, 2016
Trade Name:neoBLUE® LED Phototherapy System
Common or UsualName:Neonatal Phototherapy Unit
ClassificationName and Number:Unit, Neonatal Phototherapy21 CFR 880.5700Product Code: LBI
Predicate Device:K022196 Natus Blue Light Phototherapy Unit

Device Description

The neoBLUE® Phototherapy System consists of two products – the neoBLUE LED Phototherapy light source (light) and the neoBLUE LED Phototherapy roll stand.

The neoBLUE LED Phototherapy System is a floor-standing, mobile phototherapy light that delivers a narrow band of high-intensity blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia. The light can be used independently of the roll stand, and can be placed directly on an incubator with a flattopped surface.

Intended Use

The neoBLUE LED Phototherapy System is indicated for the treatment of hyperbilirubinemia for neonates and infants in a hospital environment, and administered by trained, professional medical staff, on the order of a licensed medical practitioner. The light can be used with a bassinet, incubator, open bed, or radiant warmer.

{4}------------------------------------------------

The indications for use of the predicate (K022196) reference the phototherapy system as emitting "blue light." The indications for use for the subject device (neoBLUE LED Phototherapy System) do not include this statement because the system emits blue and yellow light. Removal of this sentence does not raise new questions of safety or effectiveness.

Technological Characteristics

The neoBLUE light has intensity settings, high and low, which are factory callbrated to provide intensity of 35 µW/cm²/nm at the high setting and 15 µW/cm²/nm at the low setting at a distance of 12 inches (30.5 cm) from the light enclosure to the baby. The light output can also be adjusted to accommodate different distances.

Blue LEDs emit light in the range of 400 - 550 nm (peak wavelength 450-475 nm). This range corresponds to the spectral absorption of light by bilirubin, and is thus considered to be the most effective for the degradation of bilirubin. Blue LEDs do not emit significant energy in the ultraviolet (UV) region of the spectrum, reducing the potential risk of skin damage. In addition. blue LEDs do not emit significant energy in the infrared (IR) region of the spectrum, minimizing concern about excessive warming of the infant.

The blue LEDs in the device produce the same therapeutic output as in the predicate. Yellow LEDs were intermixed with the blue LEDs in the new device so that the light output visually appears less blue in color, the light source was separated from the roll stand for independent use, and a timer was added to record the cumulative number of working hours for the LED lights.

Clinical Tests

N/A

Nonclinical Tests

The neoBLUE LED Phototherapy System passed all testing and complies with the following referenced standards:

  • IEC 60601-2-50:2009-03 Edition 2; Medical Electrical Equipment Part 2-50: Particular . Requirements for the Basic Safety and Essential Performance of Infant Phototherapy Equipment
  • IEC 60601-1:2005/(R)2012 and C1:2009/(R)2012; Medical Electrical Equipment -Part 1: ● General Requirements for Basic Safety
  • IEC 60601-1-2: 2007/(R)2012; Medical Electrical Equipment Part 1-2: General ● Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
  • IEC 60601-1-6:2013-10 Edition 3.1: Medical Electrical Equipment Part 1-6: General . Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability

In addition to compliance with the relevant standards, design verification and validation testing was performed to ensure that the neoBLUE LED Phototherapy System meets its performance specifications and demonstrates equivalence to the predicate device (K022196).

{5}------------------------------------------------

Conclusions

The performance verification/validation testing data and risk analysis documentation provided in this 510(k) support the conclusion that the neoBLUE LED Phototherapy system performs as intended and is substantially equivalent to the predicate device.

The neoBLUE LED Phototherapy System is substantially equivalent to the Natus Blue Light Phototherapy Unit cleared under K022196.

§ 880.5700 Neonatal phototherapy unit.

(a)
Identification. A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.(b)
Classification. Class II (performance standards).