K Number
K160081

Validate with FDA (Live)

Device Name
TENS Electrodes
Date Cleared
2016-10-19

(278 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The proposed device is intended for use as the disposable, conductive adhesive interface between the patient's skin and the electrical stimulator to apply electrical stimulation current, and is intended to be used with marketed electrical stimulators. i.e. TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). It's for OTC use.

Device Description

The proposed TENS Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application: The electrodes are designed for single-patient application use. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin. The proposed device mainly consists of electrode pad and wire. The electrode pad is available in rectangular shape, round shape, and butterfly shape, etc.

AI/ML Overview

The Cathay Manufacturing Corp. TENS Electrodes (K160081) were compared against a primary predicate device (K080276) and a reference device (K132588) to establish substantial equivalence. The study involved bench testing (non-clinical) to assess performance characteristics.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Test)Acceptance Criteria (Threshold)Reported Device Performance (K160081)
Performance StandardConforms to IEC 60601-1 and IEC 60601-2-10:2012Conforms to IEC 60601-1 and IEC 60601-2-10:2012
Electrical impedanceIndividual pad impedance below 200 ohms @ 60 kHz (for K160081). For predicates, "Pass if the needle of multi-meter moves" for x1k ohm range.Met: Individual pad impedance below 200 ohms @ 60 kHz. (Also passed predicate test method).
Adhesive performance on steel ball>= No.14 steel ball (Diameter: 11.112 mm) following AS07-003 / GB/T4852-2002Met: >= No.14 steel ball (Diameter: 11.112 mm)
Adhesive holding strengthA sample pad with 25mm length and 25 mm width; Hang a weight of 200g without detach within 30sMet: Successfully held 200g weight for at least 30s.
Tensile StrengthElectrode is held up with a weight of 500g hung on the wire, and can hold on at least 1 minuteMet: Successfully held 500g weight for at least 1 minute.
ConformabilityNo more than 10% of the adhesive area of the device shall have separated from the skin surface at 1 h after application.Met: No more than 10% separation after 1 hour.
Impedance Distribution uniformity1) No significant deviation between resistance values measured for each combination mode.2) Maximum and minimum values within ±10% of average value.Met: No significant deviation; max/min values within ±10% of average.
Stability and shelf life2 year2 year (demonstrated to be equivalent)

2. Sample Sizes Used for the Test Set and Data Provenance:

The document does not explicitly state the specific sample sizes (number of electrodes) used for each individual performance test (e.g., electrical impedance, tensile strength) on the subject device. It mentions "Largest model: CM100180FC (Rectangle shape: 100 x 180 mm); Smallest model: CM2222FC (square shape: 22 x 22 mm); Irregular representative: CM2542YC(crescent )" were used for testing electrode types, shapes, and sizes. This suggests at least one of each representative size and shape were included.

The data provenance is not specified beyond being "bench testing (non-clinical)." There is no information regarding the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth:

This section is not applicable as the study described is a non-clinical bench testing study of TENS electrodes, not a study involving human-in-the-loop performance or expert interpretation of diagnostic images/data. Therefore, there was no ground truth requiring expert establishment.

4. Adjudication Method:

This section is not applicable for the same reasons as above. There was no expert review or adjudication of results required for this type of non-clinical bench testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks, which is not applicable to TENS electrodes' physical and electrical performance evaluation.

6. Standalone Performance Study:

Yes, a standalone performance study (i.e., algorithm only without human-in-the-loop performance) was done, but not for an AI algorithm. The "algorithm" in this context refers to the device itself and its intrinsic performance as measured by various physical and electrical tests. The bench testing evaluated the device's performance directly against the defined acceptance criteria without human interaction influencing the measurement results.

7. Type of Ground Truth Used:

The "ground truth" for this study was based on objective, quantitative measurements and established industry standards (e.g., IEC 60601-1, IEC 60601-2-10:2012, AS07-003 / GB/T4852-2002). The results of the tests (e.g., resistance values, weight held, adhesion percentage) served as the direct evidence of performance against these pre-defined benchmarks.

8. Sample Size for the Training Set:

This section is not applicable. The device in question is a physical medical device (TENS electrodes), not an AI algorithm that requires a training set. The "design" and "validation" mentioned relate to engineering and manufacturing processes, not machine learning model training.

9. How the Ground Truth for the Training Set was Established:

This section is not applicable for the same reason as above. There was no training set for an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 19, 2016

Cathay Manufacturing Corp. % Mr. Fan, Tao Consultant Shanghai YunTao Management Consultant Room 202, No. 14, Lane 986, Gaosi Road Pudong District Shanghai, 201203 China

Re: K160081

Trade/Device Name: TENS Electrodes Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: September 3, 2016 Received: September 19, 2016

Dear Mr. Fan. Tao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807): labeling (21 CFR Part 801 and Part 809): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michaelリ. Hoffmann -A

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160081

Device Name TENS Electrodes

Indications for Use (Describe)

The proposed device is intended for use as the disposable, conductive interface between the patient's skin and the electrical stimulator to apply electrical stimulation current, and is intended to be used with marketed electrical stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). It's for OTC use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter:CATHAY MANUFACTURING CORP.No. 328, Xishe Road, Maogang Town, Songjiang AreaShanghai, CN 201607
Contact:Mr. Fan, TaoPhone: (86)18616806984Fax: (86)-21-57876338fant01@uwellmed.com
Date Prepared:Oct. 18, 2016
Device Trade Name:TENS Electrodes
Device Common Name:Neurostimulation Electrodes
Classification name:Cutaneous Electrodes
Class:II
Regulation:882.1320
Product Code:GXY

Primary Predicate Device:

DeviceCompanyProduct Code510(k) Number
SELF-ADHESIVEELECTRODESCATHAY HealthcareEquipment Manufacturing Inc.GXYK080276

A Reference Device Chosen:

Top-Rank AdhesiveTop-Rank Health Care Co.,GXYK132588
ElectrodesLtd

{4}------------------------------------------------

1. Indication for Use:

The proposed device is intended for use as the disposable, conductive adhesive interface between the patient's skin and the electrical stimulator to apply electrical stimulation current, and is intended to be used with marketed electrical stimulators. i.e. TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation).

It's for OTC use.

2. Product Description:

The proposed TENS Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application:

The electrodes are designed for single-patient application use. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin.

The proposed device mainly consists of electrode pad and wire. The electrode pad is available in rectangular shape, round shape, and butterfly shape, etc.

ShapeProduct ModelShapeProduct Model
Image: Shape☐ CM2040FC ☐ CM5570FCImage: Shape☐ CM2222FC ☐ CM6060FC
☐ CM3050FC ☐ CM5575FC☐ CM3232FC ☐ CM6565FC
☐ CM3060FC ☐ CM5590FC☐ CM4040FC ☐ CM7070FC
☐ CM4050FC ☐ CM6070FC☐ CM4545FC ☐ CM7575FC
☐ CM4060FC ☐ CM6080FC☐ CM4848FC ☐ CM8080FC
☐ CM4070FC ☐ CM6090FC☐ CM5050FC ☐ CM8585FC
☐ CM4080FC ☐ CM60100FC☐ CM5555FC ☐ CM9090RC
☐ CM4090FC ☐ CM7080FCImage: Shape☐ CM22RC ☐ CM60RC
☐ CM40100FC ☐ CM70100FC☐ CM25RC ☐ CM65RC
☐ CM4580FC ☐ CM70120FC☐ CM32RC ☐ CM70RC
☐ CM4590FC ☐ CM75125FC☐ CM40RC ☐ CM75RC
☐ CM45100FC ☐ CM80100FC☐ CM45RC ☐ CM85RC
☐ CM5060FC ☐ CM80110FC☐ CM50RC ☐ CM90RC
☐ CM5070FC ☐ CM80120FCImage: Shape☐ CM55RC
☐ CM5090FC ☐ CM80130FC☐ CM4050TC ☐ CM5580SC
☐ CM50100FC ☐ CM90120FCImage: Shape☐ CM4560TC ☐ CM7090TC
☐ CM50130FC ☐ CM90130FC☐ CM4570SC
☐ CM5565FC ☐ CM100180FC☐ CM2542YC

wire: _ 0 2. Omn Packaging : 22PCS 04PCS 6PCS 88PCS

{5}------------------------------------------------

ELEMENTOFCOMPARISONSUBJECT DEVICEPrimary Predicate DeviceA Reference Device chosen
K080276K132588
SELF-ADHESIVEELECTRODESTop-Rank Adhesive Electrodes
Indications forUseThe proposed device isintended for use as thedisposable, conductiveadhesive interface betweenthe patient's skin and theelectrical stimulator to applyelectrical stimulationcurrent, and is intended tobe used with marketedelectrical stimulators, i.e.TENS (TranscutaneousElectrical NerveStimulation) and EMS(Electrical MuscularStimulation).It's for OTC useThe electrotherapyelectrodes areintended to be used toapply electricalstimulation current tothe patient's skin orrecord physiologicalsignalsAdhesive electrodes areintended for use as thedisposable, conductiveadhesive interface betweenthe patient's skin and theelectrical stimulator. Top-RankAdhesive Electrodes areintended to be used withmarketed ElectricalStimulators, i.e. TENS(Transcutaneous ElectricalNerve Stimulation) andEMS(Electrical MuscularStimulation). It's for OTC use.
OTC orPrescriptionFor OTC UseFor Prescription useFor OTC use
SterilityNon-sterileNon-sterileNon-sterile
Design FeaturesSubstrate / Wire/ Hydro-GelSubstrate / Wire/ Hydro-GelSubstrate /Wire/Hydro-Gel
Scrim / Conductive FiberScrim / Conductive FiberScrim/Conductive Fiber
Carbon Conductive Film / LinerCarbon Conductive Film / LinerCarbon Conductive Film / Liner
Materials:Substrate;Wire:Hydro-gel;Scrim;Conductivefiber;Carbonconductivefilm;LinerNon-woven fabric+ adhesiveNon-woven fabric+ adhesiveNon-woven fabric+ adhesive
Wire & Terminal coated PVCWire & Terminal coated PVCWire & Terminal coated PVC
Hydro-gelHydro-gelHydro-gel
PET fabricPET fabricPET fabric
Carbon fiber + reinforcingfiberCarbon fiber + reinforcing fiberCarbon fiber + reinforcing fiber
Poly-isobutylene, CarbonBlackPoly-isobutylene, Carbon BlackPoly-isobutylene, Carbon Black
Transparent PET linercoated siliconTransparent PET liner coatedsiliconTransparent PET liner coated silicon
BiocompatibilityConforms toISO10993Conforms toISO10993Conforms toISO10993
Labelingconfirm to 21 CFR Part 801confirm to 21 CFR Part 801confirm to 21 CFR Part 801
PerformanceConforms to IEC 60601-1andIEC 60601-2-10:2012,Part 2 Particularrequirements for the basicsafety and essentialperformance of nerve andmuscle stimulators.Conforms to IEC 60601-1andIEC 60601-2-10:2012,Part 2 Particularrequirements for the basicsafety and essentialperformance of nerve andmuscle stimulators.Conforms to IEC 60601-1andIEC 60601-2-10:2012,Part 2 Particular requirements for thebasic safety and essentialperformance of nerve and musclestimulators.
ElectricalimpedanceIndividual pad impedancebelow 200 ohms @ 60kHzBy linking the connector togel using a multi-meter. Setthe range at x1k ohm.Pass if the needle of multi-meter movesBy linking the connector to gel usinga multi-meter. Set the range at x1kohm.Pass if the needle of multi-metermoves
ELEMENTOFCOMPARISONSUBJECT DEVICEPrimary Predicate DeviceA Reference Device chosen
K080276K132588
SELF-ADHESIVEELECTRODESTop-Rank Adhesive Electrodes
Electrode types,shapes, andsizes used fortestingLargest model:CM100180FC (Rectangleshape: 100 x 180 mm);Smallest model:CM2222FC(squareshape: 22 x 22 mm);Irregular representative:CM2542YC(crescent )Largest and irregular model:CM100200 (butterfly shape:100 x 200 mm)Smallest model:CM2222 (Square shape: 22x 22 mm)A most common used sample:AF5050WP(Square Shape: 50 x50mm)
ToAdhesiveperformance on>= No.14 steel ball(Diameter: 11.112 mm)following AS07-003 /GB/T4852-2002>= No.14 steel ball(Diameter: 11.112 mm)following AS07-003 /GB/T4852-2002>= No.14 steel ball (Diameter:11.112 mm) following AS07-003 /GB/T4852-2002
Adhesiveholding strengthperformanceA sample pad with 25mmlength and 25 mm width;Hang a weight of 200gwithout detach within 30sA sample pad with 25mmlength and 25 mm width;Hang a weight of 200gwithout detach within 30sA sample pad with 25mm length and25 mm width:Hang a weight of 200g withoutdetach within 30s
Tensile StrengthElectrode is held up with aweight of 500g hung onthe wire, and can hold onat least 1 minutesElectrode is held up with aweight of 500g hung on thewire, and can hold on at least1 minutesElectrode is held up with a weight of500g hung on the wire, and can holdon at least 1 minutes
TestConformabilityNo more than 10% of theadhesive area of thedevice shall haveseparated from the skinsurface at 1 h afterapplication.No more than 10% of theadhesive area of the deviceshall have separated fromthe skin surface at 1 h afterapplication.No more than 10% of the adhesivearea of the device shall haveseparated from the skin surface at 1h after application.
ImpedanceDistributionuniformity test1) No significant deviationbetween resistancevalues measured for eachcombination mode.2) The maximum valueand the minimum valueshould be within ±10% ofaverage value1) No significant deviationbetween resistance valuesmeasured for eachcombination mode.2) The maximum value andthe minimum value should bewithin ±10% of averagevalue1) No significant deviation betweenresistance values measured for eachcombination mode.2) The maximum value and theminimum value should be within±10% of average value
Stability andshelf life2 year2 year2 year
Re-usableFor single patientFor single patientFor single patient
IntendedpopulationNo specified populationNo specified populationNo specified population

4. Summary of Technology Characteristics and Performance Testing (Comparison to the Primary Predicate Device and Reference Device)

{6}------------------------------------------------

{7}------------------------------------------------

ComparisonSummaryCompare to primary predicate device specified in K080276, our device and thepredicate device are same in Essential Components, raw materials, physicalfeatures, and same manufacturing processes. One difference is that our proposeddevice does not claim for recording physiological signals as the predicate devicedoes. Another difference is predicate device is intended for prescription use, and ourproposed product is for Over-the-Counter (OTC) use.Our product and the predicate device are the same in intended for use, essentialcomponents, physical and general safety features based on ISO internationalstandards. The hydro-gel used in K080276 is not different from what is used inCATHAY TENS Electrodes.Compare to both primary predicate device and reference device, the proposeddevice has different test method for electrical impedance. However, this test methodis a more sensitive & improved test method. Previously, proposed device testmethod for electrical impedance was exactly the same as predicate device andreference device, and all tests can pass previous test method also.
Compare to the primary predicate device and reference device, the subject deviceis constructed in the same manner,is constructed of same or similar materials,have same or similar indications for use, andhave similar performance characteristicsAny minor visual, dimensional or labeling differences between the primarypredicate device and reference device and the proposed electrodes do not pose riskto their performance and usage.

5. Bench Testing (non-clinical):

Compare to primary predicate product specified in K080276, our device and the predicate device are same in Essential Components, raw materials, physical features, and same manufacturing processes. The biocompatibility performance equivalence evidence of proposed electrode can be demonstrated.

The proposed TENS Electrodes evaluated through Electrode Performance testing (Electrical Impedance Performance, Adhesive Performance, and Tensile Strenght), Biocompatibility testing, Stability testing, and Reuse testing.

The comparison tests conclusively prove that the proposed electrode is similar / equivalence as performance efficient at electrical impedance, adhesive performances, tensile strength, conformability, and impedance distribution uniformity performance as the primary predicate device and reference device.

For lead wires & cables used in proposed device, the safety performances are demonstrated by the Third party, TUV Rheiland, through testing following IEC

{8}------------------------------------------------

60601-1:2005+CORR. 1(2006) + CORR.2(2007) + AM1(2012), Part 1General requirements for basic safety and essential performance, and IEC 60601-2-10:2012, Part 2 Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.

The results of bench testing & type testing provide reasonable assurance that the proposed device has been designed and validated to assure conformance to the requirements for its indication for use.

6. Conclusion:

Based on the indication for use, technology characteristics, and performance testing, the proposed product, TENS Electrodes, has been shown to be appropriate for its indication for use and is considered to be substantially equivalent to the primary predicate device and reference device.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).