K Number
K153778

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-05-19

(140 days)

Product Code
Regulation Number
870.3300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nester® and Tornado® Embolization Coils are intended for arterial and venous embolization in the peripheral vasculature.

Device Description

The Nester® and Tornado® Embolization Coils are constructed from coiled platinum wire and synthetic fibers. The wire forms primary coil diameters that can be delivered through catheters with end hole diameters of 0.018, 0.035, and 0.038 inch. For the Nester® Embolization Coils, the extended embolus lengths range from 3 to 20 cm, and upon deployment the coiled embolus diameters range from 2 to 20 mm. For the Tornado Embolization Coils, the extended embolus lengths range from 2 to 14.2 cm, and upon deployment the coiled embolus tapering diameters range from 3/2 to 10/5 mm.

AI/ML Overview

This document describes K153778, an FDA 510(k) premarket notification for the Nester® and Tornado® Embolization Coils. It focuses on demonstrating substantial equivalence to a predicate device. The information provided is primarily related to benchtop performance, biocompatibility, and animal testing rather than AI/algorithm performance. Therefore, many of the requested criteria are not applicable.

Here's an analysis of the provided text based on your request:

1. A table of acceptance criteria and the reported device performance:

TestAcceptance CriteriaReported Device Performance
Biocompatibility Testing (Implantable Embolization Coils)All tests results met the acceptance criteria or demonstrated that the device is biocompatible.Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Subchronic Toxicity, Genotoxicity, Implantation, Hemocompatibility, and Pyrogenicity tests were performed. All tests results met the acceptance criteria, where applicable, or demonstrated that the device is biocompatible.
Biocompatibility Testing (Loading Stylet & Cartridge)All tests results met the acceptance criteria or demonstrated that the device is biocompatible.Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, and Hemocompatibility tests were performed. All tests results met the acceptance criteria, where applicable, or demonstrated that the device is biocompatible.
Wire Tensile TestingPeak load value ≥ minimum tensile strength requirements.Testing shows the raw wire has a peak load value greater than or equal to the minimum tensile strength requirements. The predetermined acceptance criteria were met.
Coil Tensile Testing(Not explicitly stated, but implied to characterize strength)Testing characterized the embolization coils' uniaxial tensile strength. (No specific numerical acceptance criteria mentioned, but assumed to be adequate for safe use).
Rotations to Failure Testing(Not explicitly stated, but implied to characterize strength)Testing characterized the embolization coils' torque strength. (No specific numerical acceptance criteria mentioned, but assumed to be adequate for safe use).
Delivery Friction TestingEmbolization coils fully deploy into the portion of the target artery in an anatomical model.Testing shows that the embolization coils fully deploy into the portion of the target artery in an anatomical model. The predetermined acceptance criteria were met.
Delivery Fatigue TestingNo visual defect after delivery.Testing shows that the embolization coils do not have any visual defect after delivery. The predetermined acceptance criteria were met.
Fiber Security TestingAn entire fiber is not released from the coils during simulated delivery conditions.Testing shows that an entire fiber is not released from the coils during simulated delivery conditions. The predetermined acceptance criteria were met.
MRI TestingCompatible within specified magnetic field interactions, artifact, and RF-induced heating.MRI compatibility was assessed by evaluating magnetic field interactions (displacement force and torque), artifact, and RF-induced heating in accordance with FDA guidance. (Stated as MR Conditional).
Animal Testing (Safety)Target arteries do not indicate a substantially adverse biological response.One-month, three-month, and six-month safety evaluations in swine were performed on platinum embolization coils. Testing shows that the target arteries do not indicate a substantially adverse biological response. The predetermined acceptance criteria were met.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size for Test Set: This document does not specify exact sample sizes for each bench test. For animal testing, it mentions "One-month, three-month, and six-month safety evaluations in swine," implying multiple animals, but the exact number is not stated.
  • Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective. Given it's a 510(k) submission for a medical device by Cook Incorporated (based in Bloomington, Indiana, USA), the testing would typically be conducted by or for the manufacturer, likely within the US or by certified labs globally. The nature of bench and animal testing is prospective in design for the purpose of demonstrating device performance and safety.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. This document describes a medical device (embolization coils) and its performance through bench and animal testing. It does not involve interpretation of medical images or data by experts to establish a "ground truth" in the way an AI algorithm might require. The "ground truth" here is the physical and biological performance as measured by the predetermined acceptance criteria in the various tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. As mentioned above, this is not an AI/diagnostic algorithm study requiring human adjudication of ground truth. Performance is determined by meeting objective engineering and biological criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI-assisted diagnostic study. It's a submission for a physical medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is not an AI/algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this device's performance is established through:
    • Objective engineering specifications and measurements: For tests like wire tensile strength, coil tensile strength, rotations to failure, delivery friction, delivery fatigue, and fiber security.
    • Established biological and toxicological standards: For biocompatibility testing (e.g., ISO 10993 series).
    • Pathological assessment (implied for animal studies): For the animal safety evaluations, although not explicitly stated, the assessment of "substantially adverse biological response" would typically involve histological examination (pathology) of the target arteries by qualified veterinary pathologists.
    • Regulatory guidance: MRI compatibility assessment followed FDA guidance.

8. The sample size for the training set:

  • Not Applicable. This is not an AI/machine learning device. Therefore, there is no "training set."

9. How the ground truth for the training set was established:

  • Not Applicable. This is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 19, 2016

Cook Incorporated David Lehr, RAC Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404

Re: K153778

Trade/Device Name: Nester® and Tornado® Embolization Coils Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: May 11, 2016 Received: May 12, 2016

Dear David Lehr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153778

Device Name

Nester® and Tornado® Embolization Coils

Indications for Use (Describe)

The Nester® and Tornado® Embolization Coils are intended for arterial and venous embolization in the peripheral vasculature.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 812.339.2235 TOLLEREE: 800.457.4500 WWW.COOKMEDICAL.COM

Image /page/3/Picture/2 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, with a registered trademark symbol next to it. Below the word "COOK" is the word "MEDICAL" in white, set against a red background. The red background is a rectangle that extends slightly beyond the width of the word "COOK".

2.0 510(k) SUMMARY

As required by 21 CFR §807.92 Date Prepared: April 1, 2016

I. SUBMITTER

Submission:Traditional 510(k) Premarket Notification
Applicant:Cook Incorporated
Contact:David Lehr, RAC
Applicant Address:Cook Incorporated750 Daniels WayBloomington, IN 47404
Contact Phone Number:(812) 335-3575 ext. 102309
Contact Fax Number:(812) 332-0281

II. DEVICE

Trade Name:Nester® and Tornado® Embolization Coils
Common Name:Vascular Embolization Device
Classification Name:Device, Vascular, For Promoting Embolization
Regulation/Class:21 CFR §870.3300/Class II
Product CodeKRD

III. PREDICATE DEVICE

The device subject of this submission is considered substantially equivalent to the predicate device, the Cook Retracta® Detachable Embolization Coil (K123712/K151676). This predicate has never been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The Nester® and Tornado® Embolization Coils are constructed from coiled platinum wire and synthetic fibers. The wire forms primary coil diameters that can be delivered through catheters with end hole diameters of 0.018, 0.035, and 0.038 inch. For the Nester® Embolization Coils, the extended embolus lengths range from 3 to 20 cm, and upon deployment the coiled embolus diameters range from 2 to 20 mm. For the Tornado Embolization Coils, the extended embolus lengths range from 2 to 14.2 cm, and upon deployment the coiled embolus tapering diameters range from 3/2 to 10/5 mm.

V. INDICATIONS FOR USE

The Nester® and Tornado® Embolization Coils are intended for arterial and venous embolization in the peripheral vasculature.

{4}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject devices, the Nester® and Tornado® Embolization Coils, and the predicate device, the Retracta® Detachable Embolization Coil (K123712/151676), are substantially equivalent in that these devices have identical intended uses and similar technological characteristics. The predicate device is a combined embolization coil and delivery wire. Its embolization coil (made of platinum wire and nylon fibers) is delivered by a detachment mechanism. The subject devices are pushable embolization coils (also made of platinum wire and nylon fibers) delivered with a wire guide through a delivery catheter. The predicate coils (intended for delivery through catheters with an end hole size of 0.035 inch) have a helical shape, and are available in lengths of 7 cm or 14 cm and diameters ranging from 4 to 20 mm. The subject Nester coils (intended for delivery through catheters with end hole sizes of 0.018, 0.035, and 0.038 inch) also have a helical shape and are available in lengths ranging from 2 to 20 cm and in diameters ranging from 2 to 20 mm. The subject Tornado coils (also intended for delivery through catheters with end hole sizes of 0.018, 0.035, and 0.038 inch) have a tapered vortex (or tornado) shape and are available in lengths ranging from 2.0 to 14.2 cm and tapering diameters ranging from 3/2 to 10/5 mm. Additionally, the subject devices, like the predicate device, are labeled as MR Conditional. However, there are differences in the MR scanning conditions to reflect the testing performed on the subject devices. Based on the comparison of the design, intended use, materials, fundamental technology, and principle of operation, the subject devices are considered to be substantially equivalent to the currently marketed predicate device.

VII. PERFORMANCE DATA

The subject devices underwent the applicable testing listed below to ensure reliable design and performance under the testing parameters. Performance (bench, MRI, and animal) and biocompatibility testing was conducted in accordance with applicable FDA guidance documents to confirm the reliable performance of critical device characteristics.

  • . Biocompatibility Testing - Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Subchronic Toxicity, Genotoxicity, Implantation, Hemocompatibility, and Pyrogenicity tests were performed on the implantable embolization coils. Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, and Hemocompatibility tests were performed on the device's

{5}------------------------------------------------

loading stylet and loading cartridge. All tests results met the acceptance criteria, where applicable, or demonstrated that the device is biocompatible.

  • . Wire Tensile Testing - Testing shows the raw wire has a peak load value greater than or equal to the minimum tensile strength requirements. The predetermined acceptance criteria were met.
  • . Coil Tensile Testing - Testing characterized the embolization coils' uniaxial tensile strength.
  • . Rotations to Failure Testing - Testing characterized the embolization coils' torque strength.
  • . Delivery Friction Testing - Testing shows that the embolization coils fully deploy into the portion of the target artery in an anatomical model. The predetermined acceptance criteria were met.
  • . Delivery Fatigue Testing – Testing shows that the embolization coils do not have any visual defect after delivery. The predetermined acceptance criteria were met.
  • Fiber Security Testing Testing shows that an entire fiber is not released from the . coils during simulated delivery conditions. The predetermined acceptance criteria were met.
  • MRI Testing MRI compatibility was assessed by evaluating magnetic field . interactions (displacement force and torque), artifact, and RF-induced heating in accordance with FDA guidance titled, Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment.
  • . Animal Testing - One-month, three-month, and six-month safety evaluations in swine were performed on platinum embolization coils. Testing shows that the target arteries do not indicate a substantially adverse biological response. The predetermined acceptance criteria were met.

VIII. CONCLUSIONS

The data included in this submission indicate that the subject devices do not raise new questions of safety or effectiveness compared to the predicate device. This supports a determination of substantial equivalence.

§ 870.3300 Vascular embolization device.

(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).