K Number
K153767

Validate with FDA (Live)

Date Cleared
2016-03-29

(90 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

X-cera Zirconia Blanks are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.

Device Description

X-cera Zirconia Blanks are provided in either block, disc, cylinder shape and are derived from zirconia powder that has been processed via uni-axial die pressing, followed by iso-static pressing, to achieve final net shapes of uniform density and distribution. These Zirconia Blanks can be fabricated into various prosthetic dental device.

X-cera Zirconia Blanks are offered in ten(10) different product families of shapes and a multitude of different sizes as shown in Figure 1 and Table 1 and are capable of being machined into complex dental shapes using modern machining methods and are similar to the sizes and shapes of the predicate device.

AI/ML Overview

The document describes the submission of "X-cera Zirconia Blanks" for 510(k) premarket notification. The information provided outlines the device's technical specifications and a study conducted to demonstrate its substantial equivalence to a legally marketed predicate device, "Aidite Zirconia Dental Ceramics (K111291)".

Here's an analysis of the acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally based on compliance with international standards (ISO 6872, ISO 10993, ISO 7405) and internal design specifications. The reported device performance confirms compliance with these criteria through a series of non-clinical tests.

Acceptance Criteria (Standard / Design Requirement)Reported Device Performance (X-cera Zirconia Blanks)
Material Composition
ZrO₂88% ~ 96%
Y₂O₃4% ~10%
Al₂O₃<1%
HfO₂<0.05%
SiO₂<0.5%
All compositions comply with design specifications and manufacturer's values in IFU, and meet ISO 6872 Type II class 6 requirements.
Physical Properties (Post Sintering)
Density6.0~6.1 g/cm³ (Comply with values specified by manufacturer in IFU)
PorosityLess than 0.1% (Comply with values specified by manufacturer in IFU)
Shrinkage rate and Expand RateComply with acceptance criteria
Coefficient of linear thermal expansion (25 °C~500°C)$10.5±0.5×10^{-6} K^{-1}$
Mechanical Properties (Post Sintering)
Flexural Strength≥800MPa (comply with ISO 6872: 2008) AND ≥900MPa (comply with product design requirement)
Elastic modulus210±15 GPa
Fracture ToughnessComply with ISO 6872
Biocompatibility
ISO 7405 and ISO 10993 (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic, Genotoxicity)Passed all tests (Results comply with both ISO standards)
Applied Standards
ISO 6872Complied
ISO 10993-1Complied
ISO 7405Complied

Note: The document explicitly states "The X-cera Zirconia Blanks passed all testing stated above as shown by the acceptable results obtained." and "The device passed all the testing in accordance with national and international standards."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the specific sample sizes for each of the non-clinical tests (Flexural Strength, Density, Shrinkage, Biocompatibility, etc.). It only indicates that "testing has been performed" and the "test results comply."

The data provenance is from non-clinical performance data (laboratory testing) conducted to support the substantial equivalence of the device. The country of origin of the testing is not specified, but the manufacturer is Shenzhen XiangTong Photoelectricity Technology Co., Ltd. in China. The testing would be considered prospective in the context of demonstrating device performance against pre-defined criteria.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

There is no mention of experts being used to establish a "ground truth" in the traditional sense. This is because the study is a non-clinical performance study focusing on material properties and biocompatibility, not on interpretative outcomes like in medical imaging. The ground truth is established by the standards themselves (e.g., ISO 6872 sets the acceptable flexural strength) and the device's design specifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As this is a non-clinical material property and biocompatibility study, there is no human interpretation or adjudication involved to establish ground truth or assess performance in the way described by adjudication methods for clinical studies.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This type of study is irrelevant for a device like "X-cera Zirconia Blanks," which is a material for dental prosthetics and does not involve human readers or AI assistance in diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

No standalone algorithm performance study was done. This device is not an algorithm or an AI-powered system, so this type of study is not applicable. The performance evaluated is of the physical and chemical properties of the material itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the non-clinical performance tests is based on:

  1. International Standards: Specifically, ISO 6872 for dental ceramic materials, ISO 10993 for biological evaluation of medical devices, and ISO 7405 for preclinical evaluation of medical devices used in dentistry. These standards define acceptable ranges or minimum/maximum values for properties like flexural strength, density, and biocompatibility.
  2. Product Design Requirements: Internal specifications set by the manufacturer (e.g., flexural strength ≥ 900MPa).
  3. Manufacturer's Specified Values: Values stated for the device in the Instruction for Use (e.g., for density and porosity).

8. The sample size for the training set

Not applicable. This device is a material, not a machine learning algorithm, therefore there is no "training set."

9. How the ground truth for the training set was established

Not applicable. As there is no training set for a machine learning algorithm, there is no ground truth established for it.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three interconnected faces in profile.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 29, 2016

Shenzhen XiangTong Photoelectricity Technology Co., Ltd. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313

Re: K153767

Trade/Device Name: X-cera Zirconia Blanks Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: March 9, 2016 Received: March 15, 2016

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153767

Device Name X-cera Zirconia Blanks

Indications for Use (Describe)

X-cera Zirconia Blanks are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

for

X-cera Zirconia Blanks

1. 510(k) Number

K153767

2. Submission Sponsor

Shenzhen XiangTong Photoelectricity Technology Co., Ltd. 2~4th floor, 1st Building, South Industrial park of Honghualing, Nanshan District, Shenzhen, China Phone: 0086-755-86001804 Fax: 0086-755-86001486 Contact: Yongqing Chen, Vice-President

3. Submission Correspondent

Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: +86-21-22815850 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Lee Fu, RAC, Senior Consultant, Regulatory Email: project.management@emergogroup.com

4. Date Prepared

03/29/2016

5. Device Identification

Trade/Proprietary Name:X-cera Zirconia Blanks
Common/Usual Name:Zirconia Blocks
Classification Name:Porcelain powder for clinical use
Classification Regulation:872.6660
Product Code:EIH
Device Class:Class II
Classification Panel:Dental Devices

6. Legally Marketed Predicate Device(s)

Aidite Zirconia Dental Ceramics (K111291)

{4}------------------------------------------------

7. Device Description

X-cera Zirconia Blanks are provided in either block, disc, cylinder shape and are derived from zirconia powder that has been processed via uni-axial die pressing, followed by iso-static pressing, to achieve final net shapes of uniform density and distribution. These Zirconia Blanks can be fabricated into various prosthetic dental device.

X-cera Zirconia Blanks are offered in ten(10) different product families of shapes and a multitude of different sizes as shown in Figure 1 and Table 1 and are capable of being machined into complex dental shapes using modern machining methods and are similar to the sizes and shapes of the predicate device.

Any difference in the net shapes and dimensions between X-cera Zirconia Blanks and those offered by the manufacturer of the predicate devices are incidental to the performance of the final dental prosthetic devices that will be fabricated from them.

Image /page/4/Figure/4 description: The image shows a variety of Zirconia Blanks. There are 10 different types of Zirconia Blanks, including XS series, XQ series, XW series, XZ series, XC series, XZZ series, XO series, XK series, XA series, and XM series. The Zirconia Blanks are all white, but they come in different shapes and sizes.

Figure 1 X-cera Zirconia Blanks

Table 2X-cera Zirconia Blanks Product Family(Model)
Family(Model)Shapes & Sizes available
XQ seriesCylinder: Φ22*25(mm)Disc: Φ98×8(mm)
XW seriesDiscs: Φ100×10 to Φ100×25(mm) (with 3mm protruding step)
XO seriesDiscs: Φ98×10 to 98×25(mm) (with 2mm protruding step)
XA seriesDiscs: Φ100×10 to Φ100×25(mm) (with Φ112×6mm Outer plastic ring)
XA seriesDiscs: Φ98×10 to Φ98×25(mm) (with Ø 115×10mm Outer ring)
XA seriesDiscs: Φ71×10 to Φ71×25(mm)
XA seriesBlocks: 40×20×16 to 65×30×20(mm)
XA seriesBlocks: 90×72×10 to 90×72×25(mm)(outer ring)
XK seriesDiscs: Φ98×10 to Φ98×25(mm)
XK seriesBlocks: 42×16×16 to 60×25×20(mm)
XK seriesCylinders: Φ16×16 to Φ20×20(mm)
XZ seriesDiscs: Φ95×10 to Φ95×25(mm)
XZ seriesBlocks: 43×25×16 to 93×75×25(mm)

{5}------------------------------------------------

Cylinders: $Φ$ 23×16 to $Φ$ 23×22(mm)
XM seriesBlocks: 43×25×16 to 75×42×22(mm)
XC seriesCylinders $Φ$ 25×26 to $Φ$ 25×60(mm) (with Outer plastic ring)
Blocks: 26×25×20 to 72×40×25(mm) (with Outer plastic ring)
XS seriesBlocks: 20×14×15 to 85×40×22(mm)
XZZ seriesDiscs: $Φ$ 95×10 to $Φ$ 95×25(mm)

Note: The form of size above has two forms, one is shown in D×H and the other is L×W× H, of which, "D"means Diameter, "H"means height, "L" means length, "W" means width.

8. Indication for Use Statement

X-cera Zirconia Blanks are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.

9. Substantial Equivalence Discussion

The following table compares the X-cera Zirconia Blanks to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

ManufacturerShenzhen XiangTongPhotoelectricityTechnology Co., ltd.Qinhuangdao AiditeHigh-Technical CeramicsCo.SIGNIFICANTDIFFERENCES
Trade NameX-cera Zirconia BlanksAidite Zirconia DentalCeramics/
510(k) Number/K111291/
Product CodeEIHEIHSame
RegulationNumber21 CFR 872.666021 CFR 872.6660Same
Regulation NamePowder, PorcelainPowder, PorcelainSame
Indications for Useare indicated for theproduction of artificialteeth in fixed orremovable dentures, orfor jacket crowns,facings, and veneersare indicated for theproduction of artificialteeth in fixed orremovable dentures, orfor jacket crowns,facings, and veneersSame
Directions for UsePrescription UsePrescription UseSame
FundamentalScientificTechnologyIso static pressingIso static pressingSame
Main workingprincipleBe capable of beingmachined into complexdental shapes usingmodern machiningBe capable of beingmachined into complexdental shapes usingmodern machiningSame
ManufacturerShenzhen XiangTongPhotoelectricityTechnology Co., ltd.Qinhuangdao AiditeHigh-Technical CeramicsCo.SIGNIFICANTDIFFERENCES
Trade NameX-cera Zirconia BlanksAidite Zirconia DentalCeramics/
methodsmethods
Applied StandardsISO 6872ISO 10993-1ISO 7405ISO 6872ISO 10993-1ISO 7405Same
Main ChemicalComposition (inmass-%)ZrO₂ 88% ~ 96%Y₂O₃ 4% ~10%Al₂O₃ <1%HfO₂ <0.05%SiO₂ <0.5%ZrO₂ <96%Y₂O₃>4%Al₂O₃ <1%HfO₂>1%SiO₂ < 0.02%Analysis 1
CrystalMorphologyTetragonalTetragonalSame
ColorwhitewhiteSame
Sinteringtemperature1480°C1480°CSame
SterileNon-sterileNon-sterileSame
Density (postsintering)6.0~6.1 g/cm³6.05g/cm³Analysis 2
Shrinkage (postsintering)19%~22%22%Analysis 3
CategorizationAccording to ISO7405 and ISO10993-11) Categorization bynature of contact:External communicatingdevices2) Categorization byduration of contactPermanent contactdevices1) Categorization bynature of contact:Surface-contactingdevices2) Categorization byduration of contactPermanent contactdevicesSame
Flexural Strength(post sintering)Comply with ISO 6872Comply with ISO 6872Same
Elastic modulus(post sintering)210±15 GPa210 GPaSame
PorosityLess than 0.1%0%Same
Type and Classaccording to ISO6872Type II Class 6Type II Class 6Same
Fracture Toughness(after sintered)Comply with ISO 6872Comply with ISO 6872Same
Coefficient of linearthermal expansion$10.5±0.5×10^{-6} K^{-1} (25 °C~500°C)$$10.5±0.5×10^{-6} K^{-1} (25 °C~500°C)$Same

Table 3 – Comparison of Characteristics

{6}------------------------------------------------

Analysis 1

They are both conformance with the requirement ISO 6872 Type II class 6.

{7}------------------------------------------------

Analysis 2

The value of Density (post sintering) of Aidite Zirconia Dental Ceramics is a fixed value, and the range of the proposed cover the value of the predicate.

Analysis 3

Proposed device and predicate device are both conformance to the same mechanical and chemical property and biocompatibility standard (ISO 6872, ISO 10993-1, ISO 7405).

10. Non-Clinical Performance Data

The following testing has been performed to support substantial equivalence:

  • Flexural Strength Test - The test results of mean flexural strength (post sintering) comply with the ISO 6872: 2008 (≥800MPa) and product design requirement (≥900MPa).
  • . Density and Apparent Porosity Test - The test results of density and apparent porosity comply with the values specified by manufacturer in the Instruction for Use.
  • Shrinkage rate and Expand Rate Test - The test results of shrinkage rate and expand rate comply with the acceptance criteria.
  • . Physical and chemical test - The test results comply with the ISO 6872: 2008.
  • . Zirconia powder Composition Test - The test results comply with the design specifications and the values specified by manufacturer in the Instruction for Use.
  • . Biocompatibility Test – Results comply with ISO 7405 and ISO 10993 including In Vitro Cytotoxicity Test, Sensitization Test, Intracutaneous Reactivity Test, Acute Systemic Test, and Genotoxicity.

The X-cera Zirconia Blanks passed all testing stated above as shown by the acceptable results obtained.

The device passed all the testing in accordance with national and international standards.

11. Clinical Performance Data

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device.

{8}------------------------------------------------

12. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.

The X-cera Zirconia Blanks, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.