K Number
K153703

Validate with FDA (Live)

Device Name
Guiding Catheter
Date Cleared
2016-05-12

(141 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Guiding Catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vasculature.

Device Description

The TXM Guiding Catheter is a guiding catheter, for application in intravascular vessels. It consists of a lubricous inner liner made from Teflon, a stainless steel braid over the inner liner. The outer shaft consists of durometer hardness and different lengths of colored Polyether block amide. A Polycarbonate hub is attached to the proximal end of the Guiding Catheter. A strain relief is placed at the proximal end of the Guiding Catheter and over the distal end of the device shafts will have an outer diameter of 5 French sizes (0.070") and 6 French sizes (0.082) respectively for each catheter. The 5 French sizes will have an inner diameter of 0.058" and the 6 French sizes will have an inner diameter of 0.070". The device is available in length of 100cm. The device is supplied sterile and is intended for single use.

AI/ML Overview

This document is a 510(k) premarket notification for a guiding catheter, indicating its substantial equivalence to a predicate device. It doesn't describe the evaluation of an AI/ML powered device, but rather a traditional medical device (catheter). Therefore, most of the requested information regarding AI/ML device evaluation (like sample sizes for test/training sets, expert ground truth, MRMC studies, standalone performance) is not applicable or cannot be extracted from this document.

However, I can extract the acceptance criteria and results for the bench testing and biocompatibility testing as presented for this traditional medical device.

1. Table of Acceptance Criteria and Reported Device Performance

For this medical device, the "acceptance criteria" are implied by meeting the specified standards or internal test methods, and the "reported device performance" is a simple "Pass" or "Fail" against those criteria.

Test NameApplicable Standard or Internal Test MethodAcceptance Criteria (Implied by standard and "Pass" result)Reported Device Performance
Air LeakISO 10555Meet ISO 10555 requirementsPass
In-Vitro Track ForceInternal Test MethodMeet internal method requirementsPass
StiffnessInternal Test MethodMeet internal method requirementsPass
Tensile StrengthISO 10555Meet ISO 10555 requirementsPass
Gravity Flow RateISO 10555Meet ISO 10555 requirementsPass
Liquid LeakISO 10555Meet ISO 10555 requirementsPass
Power Injector (Static Burst Pressure)ISO 10555Meet ISO 10555 requirementsPass
Tip Shape RetentionInternal Test MethodMeet internal method requirementsPass
Dimensional & Physical AttributesISO 10555Meet ISO 10555 requirementsPass
Kink TestingInternal Test MethodMeet internal method requirementsPass
Buckling TestInternal Test MethodMeet internal method requirementsPass
Corrosion ResistanceISO 10555Meet ISO 10555 requirementsPass
RadiopacityASTM-F640-12Meet ASTM-F640-12 requirementsPass
Torque ResponseInternal Test MethodMeet internal method requirementsPass
Torque to FailureInternal Test MethodMeet internal method requirementsPass
OvalizationInternal Test MethodMeet internal method requirementsPass
Simulated UseInternal Test MethodMeet internal method requirementsPass
Seal Strength (Pouch)ASTM F88/F88M-09Meet ASTM F88/F88M-09 requirementsPass
Dye Penetration TestASTM-F1929-12Meet ASTM-F1929-12 requirementsPass
Female Luer VerificationISO 594Meet ISO 594 requirementsPass
Shipping and TransportationISTA 3AMeet ISTA 3A requirementsPass
Accelerated AgingASTM F1980-07Meet ASTM F1980-07 requirementsPass
Biocompatibility Tests
CytotoxicityISO10993-5Meet ISO10993-5 requirementsPass
IrritationISO10993-10Meet ISO10993-10 requirementsPass
SensitizationISO10993-10Meet ISO10993-10 requirementsPass
Systemic ToxicityISO10993-11Meet ISO10993-11 requirementsPass
PyrogenicityUSP<151>, ISO10993-11, USP <85>Meet specified USP and ISO requirementsPass
HemocompatibilityISO10993-4Meet ISO10993-4 requirementsPass
Toxicological HazardISO10993-17Meet ISO10993-17 requirements (Risk Assessment)Pass
Sterilization Test
Ethylene Oxide GasRelevant validation standards (implicit)Sterility Assurance Level (SAL) of 10-6Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document refers to "non-aged devices (T=0)" and "devices subject to 3 years of accelerated aging (T=3)" for bench testing, but the specific number of devices tested for each test is not detailed. The data provenance is also not specified, as this is a device submission focused on bench and biocompatibility testing, not clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or expert evaluation for establishing ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or expert adjudication for test sets.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's evaluation is primarily based on meeting the specifications outlined in recognized international and national standards (e.g., ISO 10555, ASTM F640-12, ISO 10993 series, USP) and internal test methods. For example, for tensile strength, the ground truth is simply whether the device withstood the specified force according to the standard. For biocompatibility, it's whether the biological responses fall within acceptable limits defined by the ISO 10993 series.

8. The sample size for the training set

Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or training sets.

9. How the ground truth for the training set was established

Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or training sets.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The symbol is enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 12, 2016

Texas Medical Technologies, Inc. % EJ Smith Consultant Smith Associates 1468 Harwell Avenue Crofton. Marvland 21114

Re: K153703

Trade/Device Name: Guiding Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: April 12, 2016 Received: April 12, 2016

Dear EJ Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153703

Device Name Guiding Catheter

Indications for Use (Describe)

The Guiding Catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vasculature.

Type of Use (Select one or both, as applicable)
-------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.1 - Company Information & Contact Person

Company Name:Texas Medical Technologies Inc.
Company Address:9005 Montana Ave. Ste. AEl Paso, Texas 79925
Telephone:(915) 774-4321
Fax:(915) 774-4323
Contact Person:Cesar Rios, Quality Assurance & Regulatory Manager
Date Prepared:4/11/2016

5.2 - Device Name & Classification

Proprietary Name:Guiding Catheter
Common Name:Catheter
Classification Name:Percutaneous Catheter
Regulation Number:21 CFR 870.1250
Product Code:DQY
Device Class:II

5.3 - Predicate Device

Legally Marketed Substantially Equivalent Predicate Device

Proprietary Name:Vista Brite Tip
Company Name:Cordis Corporation
Common Name:Percutaneous Catheter
Classification Name:Percutaneous Catheter
Regulation Number:21 CFR 870.1250
Product Code:DQY
Device Class:II
510(k) NumberK972978

5.4 - Device Description

The TXM Guiding Catheter is a guiding catheter, for application in intravascular vessels. It consists of a lubricous inner liner made from Teflon, a stainless steel braid over the inner liner. The outer shaft consists of durometer hardness and different lengths of colored Polyether block amide. A Polycarbonate hub is attached to the proximal end of the Guiding Catheter. A strain relief is placed at the proximal end of the Guiding Catheter and over the distal end of the device shafts will have an outer diameter of 5 French sizes (0.070") and 6 French sizes (0.082) respectively for each catheter. The 5 French sizes will have an inner diameter of 0.058" and the 6

{4}------------------------------------------------

French sizes will have an inner diameter of 0.070". The device is available in length of 100cm. The device is supplied sterile and is intended for single use.

The device is supplied sterile and is intended for single use

The following table lists the models and sizes available for Guiding Catheter.

Table 5.4.1. Guiding Catheter Models and Sizes

CommercialModel NumberFrenchsizeShaftLength (cm)OuterDiameter(inches)InnerDiameter(inches)Marker BandMaterialTip Shape
GC-510-JL3551000.0700.058TungstenJL 3.5
GC-510-JL4051000.0700.058TungstenJL 4
GC-510-JL4551000.0700.058TungstenJL 4.5
GC-510-JL5051000.0700.058TungstenJL 5
GC-510-JL6051000.0700.058TungstenJL 6
GC-510-JR3551000.0700.058TungstenAL 1
GC-510-JR4051000.0700.058TungstenAL 2
GC-510-JR5051000.0700.058TungstenAL 3
GC-510-JR6051000.0700.058TungstenJR 3.5
GC-510-AL1051000.0700.058TungstenJR 4
GC-510-AL2051000.0700.058TungstenJR 5
GC-510-AL3051000.0700.058TungstenJR 6
GC-510-AR1051000.0700.058TungstenAR 1
GC-510-AR2051000.0700.058TungstenAR 2
GC-510-AR3051000.0700.058TungstenAR 3
GC-510-MPA151000.0700.058TungstenMPA 1
GC-510-MPA151000.0700.058TungstenMPA 2
GC-510-MPB151000.0700.058TungstenMPB 1
GC-510-MPB251000.0700.058TungstenMPB 2
GC-610-JL3561000.0820.070TungstenJL 3.5
GC-610-JL4061000.0820.070TungstenJL 4
GC-610-JL4561000.0820.070TungstenJL 4.5
GC-610-JL5061000.0820.070TungstenJL 5
GC-610-JL6061000.0820.070TungstenJL 6
GC-610-JR3561000.0820.070TungstenAL 1
GC-610-JR4061000.0820.070TungstenAL 2
GC-610-JR5061000.0820.070TungstenAL 3
GC-610-JR6061000.0820.070TungstenJR 3.5
GC-610-AL1061000.0820.070TungstenJR 4
GC-610-AL2061000.0820.070TungstenJR 5
GC-610-AL3061000.0820.070TungstenJR 6
GC-610-AR1061000.0820.070TungstenAR 1
GC-610-AR2061000.0820.070TungstenAR 2
GC-610-AR3061000.0820.070TungstenAR 3
GC-610-MPA161000.0820.070TungstenMPA 1
GC-610-MPA161000.0820.070TungstenMPA 2
GC-610-MPB161000.0820.070TungstenMPB 1
GC-610-MPB261000.0820.070TungstenMPB 2

{5}------------------------------------------------

5.5 - Indications for Use

The guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vasculature.

The Indications for Use are identical, the devices are both intended for intravascular use and the Indications for Use do not change the intended use of the TXM Guiding Catheter when compared to the predicate device.

5.6 - Summary of Technological Characteristics Comparison

Based on a comparison of the indications for use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging, it is determined that the Guiding Catheter is substantially equivalent to the predicate device. Table 5.6 below provides a comparison of the Guiding Catheter and the predicate.

TechnicalCharacteristics /Principle ofOperationTXM Guiding CatheterVista Brite GuidingCatheterSubstantiallyEquivalent?
Length100 cm90 cm, 100 cm, 110 cm, 125 cmYes
Outer Diameter0.070"-0.082"0.068"-0.082"Yes
Inner Diameter0.058"-0.070"0.056"-0.070"Yes
ShapeJudkins right, Judkins left,Amplatz right, Amplatz leftand Multipurpose.Judkins right, Judkins left,Amplatz right, Amplatz leftand Multipurpose and others.Yes
Inner LinerMaterialPolytetrafluoroethylene(PTFE)*Polytetrafluoroethylene(PTFE)Yes
BraidReinforcementMaterialStainless Steel*Stainless SteelYes
RadiopaqueMarkerTungsten Material*Radiopaque fillerYes
Outer ShaftMaterialNylon, Pebax,Polyurethane*NylonYes
Luer MaterialPolycarbonate*UnknownYes
Luer ConnectorFemale Luer Connector*Female Luer ConnectorYes
Anatomical SiteUsePeripheral, CoronaryPeripheral, CoronaryYes
Delivery to SiteOver-the-wireOver-the-wireYes
GuidewireCompatibilityMaximum 0.038"Maximum 0.038"Yes
PackagingTyvek PouchTyvek PouchYes
SterilizationEtO GasEtO GasYes

Table 5.6.1 Comparison of the Guiding Catheter and the Predicate Device.

  • Denotes a patient-contacting material.

{6}------------------------------------------------

5.7 - Testing Summary

The following bench tests were performed to evaluate the design elements and performance characteristics of the Guiding Catheter and to demonstrate substantial equivalence to the predicate device. The Guiding Catheter met the predetermined acceptance criteria. Testing was performed on non-aged devices (T=0) as well as on devices subject to 3 years of accelerated aging (T=3). Tests results show that the Guiding Catheter is substantially equivalent to the predicate device.

5.7.1- Bench Testing Table

Table 5.7.1 below provides a summary of the bench testing performed on the Guiding Catheter.

Test#Test NameApplicableStandardor InternalTestMethodTestResults(T=3)Test#Test NameApplicableStandardor InternalTestMethodTestResults(T=3)
1Air LeakISO 10555T=0:T=3Pass12CorrosionResistanceISO 10555T=0:T=3Pass
2In-VitroTrack ForceInternalTestMethodT=0;T=3Pass13RadiopacityASTM-F640-12T=0;T=3Pass
3StiffnessInternalTestMethodT=0:T=3Pass14TorqueResponseInternalTestMethodT=0:T=3Pass
4TensileStrengthISO 10555T=0;T=3Pass15Torque toFailureInternalTestMethodT=0;T=3Pass
5Gravity FlowRateISO 10555T=0:T=3Pass16OvalizationInternalTestMethodT=0:T=3Pass
6Liquid LeakISO 10555T=0:T=3Pass17Simulated UseInternalTestMethodT=0:T=3Pass
7PowerInjector(Static BurstPressure)ISO 10555T=0;T=3Pass18Seal Strength(Pouch)ASTMF88/F88M-09T=0;T=3Pass
8Tip ShapeRetentionInternalTestMethodT=0;T=3Pass19DyePenetrationTestASTM-F1929-12T=0;T=3Pass
9Dimensional& PhysicalAttributesISO 10555T=0:T=3Pass20Female LuerVerificationISO 594T=0:T=3Pass
10Kink TestingInternalTestMethodT=0;T=3Pass21Shipping andTransportationISTA 3AT=0:T=3Pass
11BucklingTestInternalTestMethodT=0:T=3Pass22AcceleratedAgingASTMF1980-07T=0:T=3Pass

Table 5.7.1. Bench Testing Performed on Guiding Catheter

{7}------------------------------------------------

5.7.2 - Biocompatibility

The Guiding Catheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hours). Biocompatibility testing was performed in accordance with ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process'' (2009). Table 5.7.2 below describes the testing performed to determine biocompatibility. All testing met the predetermined acceptance criteria.

Biological EffectTestCompliance Standard
CytotoxicityL929 Neutral Red Uptake (NRU) - ISOISO10993-5
IrritationIntracutaneous Injection - ISOISO10993-10
SensitizationKligman MaximizationMurine Local Lymph AssayISO10993-10
Systemic ToxicityISO Acute Systemic Toxicity TestISO10993-11
PyrogenicityPyrogen Test in RabbitLimulus Amebocyte Lysate$USP<151>$ISO10993-11$USP <85>$ 38, NF 33, 2015
HemocompatibilityHemolysis-Complete (Direct and Indirect)Complement ActivationIn-Vivo ThrombogenicityISO10993-4
Toxicological HazardRisk Assessment of Extractables and LeachablesISO10993-17

Table 5.7.2. Summary of Biocompatibility Testing for the Guiding Catheter

5.8 - Sterilization Testing Summary

Validation Sterilization ProcessSterility Assurance Level (SAL)Validation Result
Ethylene Oxide Gas10-6Pass

5.9 – Conclusion

The Guiding Catheter is substantially equivalent in intended use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).