K Number
K153596

Validate with FDA (Live)

Date Cleared
2016-08-09

(237 days)

Product Code
Regulation Number
862.3645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARK™ Oxcarbazepine Metabolite Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of Oxcarbazepine Metabolite in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of Oxcarbazepine Metabolite to help ensure appropriate therapy.

The ARK™ Oxcarbazepine Metabolite Calibrator is intended for use in calibration of the ARK Oxcarbazepine Metabolite Assay.

The ARK™ Oxcarbazepine Metabolite Control is an assayed quality control material intended for use in quality control of the ARK Oxcarcarbazepine Metabolite Assay.

For prescription use only. Caution: Federal Law restricts this device to sale by or on the order of a licensed practitioner.

Device Description

The ARK Oxcarbazepine Metabolite Assay is a homogeneous immunoassay based on competition between drug in the specimen and Oxcarbazepine Metabolite labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenyzme NAD functions only with the bacterial enzyme used in the assay.

The ARK Oxcarbazepine Metabolite Assay consists of reagents R1 anti-Oxcarbazepine Metabolite polyclonal antibody with substrate and R2 Oxcarbazepine Metabolite labeled with bacterial G6PDH enzyme. The ARK Oxcarbazepine Metabolite Calibrator consists of a six-level set to calibrate the assay, and the ARK Oxcarbazepine Metabolite Control consists of a three-level set used for quality control of the assay.

AI/ML Overview

The provided document describes the performance characteristics of the ARK™ Oxcarbazepine Metabolite Assay, Ark Oxcarbazepine Metabolite Calibrator, and Ark Oxcarbazepine Metabolite Control. This is a 510(k) premarket notification for a medical device (an in-vitro diagnostic assay), not an AI/ML powered device, therefore the information typically requested for AI/ML device studies (such as number of experts, adjudication methods, multi-reader multi-case studies, separate training/test sets with ground truth establishment methods for AI/ML models) are not applicable.

The acceptance criteria and performance data are detailed for several analytical validation studies.

Here's the breakdown of the requested information based on the provided document:


1. A table of acceptance criteria and the reported device performance

Performance CharacteristicAcceptance Criteria (Implicit from CLSI guidelines and successful submission)Reported Device Performance
Limit of Quantitation (LOQ)≤20% CV with ±15% recovery (according to CLSI EP17-A2)1.0 µg/mL
RecoveryDesired close agreement between theoretical and recovered concentrationsGenerally good, variations with S:R ratio. Example: For S:R 9:1, range 0.98 µg/mL (at 1.0 µg/mL theoretical) to 44.63 µg/mL (at 45.0 µg/mL theoretical).
LinearityPercent difference ±10% between 1st and 2nd order regressed values, or ≤ 0.20 µg/mL below 2.0 µg/mL (according to CLSI/NCCLS Protocol EP6-A)Linear relationship demonstrated between 1.0 and 50.0 µg/mL (y = 1.0388x -0.0693). All differences within acceptance criteria.
Assay RangeClinically relevant measurable range1.0 to 37.0 µg/mL
Method Comparison (vs. LC-MS/MS)Desired strong correlation (slope close to 1, y-intercept close to 0, high r²) (according to CLSI Protocol EP9-A3)Slope: 1.01 (0.98 to 1.04 95% CI) y-intercept: -0.38 (-0.84 to 0.12 95% CI) Correlation Coefficient (r²): 0.95 (0.94 to 0.97 95% CI)
Precision≤10% CV (Total CV)ARK Control: LOW: 5.7% CV MID: 4.8% CV HIGH: 5.1% CV Human Serum: LOW: 5.5% CV MID: 5.5% CV HIGH: 5.1% CV All results meet the ≤10% CV criterion.
Interfering Substances≤10% error in measurementAll tested substances (Human Albumin, Bilirubin, Cholesterol, Human IgG, Hemoglobin, Rheumatoid Factor, Triglycerides, Uric Acid) resulted in ≤10% error.
Stability (Serum Specimens)Defined stability period at various conditionsStable for at least 48 hours at room temperature (22 °C), 14 days refrigerated (2-8 °C), 3 months frozen (-20 °C), and after 3 freeze/thaw cycles.
Calibration Curve StabilityDefined stability period for stored calibrationEffective for at least 15 days.

2. Sample size used for the test set and the data provenance

  • LOQ: Not explicitly stated how many unique samples, but "mean of six (6) replicate measurements" for recovery at different enantiomer ratios.
  • Recovery: Not explicitly stated how many unique samples, but "mean of six (6) replicate measurements" of Oxcarbazepine Metabolite was tabulated as a function of the enantiomer ratio.
  • Linearity: Not explicitly stated how many unique samples other than a "60.0 µg/mL serum sample was prepared and dilutions were made proportionally."
  • Method Comparison: 190 samples.
  • Precision: 3 levels of ARK Control (N=160 each) and 3 human serum pooled specimens (N=160 each). This means for each of the 6 material types, 160 measurements were taken (quadruplicate twice a day for 20 days).
  • Interfering Substances: Not explicitly stated the number of unique human serum samples, but substances were tested "in serum with known levels of Oxcarbazepine Metabolite (approximately 3 and 30 µg/mL)."
  • Specificity & Drug Interference: Not explicitly stated how many unique human serum samples, but tested with spiked compounds into normal human serum with known Oxcarbazepine Metabolite levels.
  • Sample Stability & Calibration Curve Stability: "supporting data" cited, but specific sample sizes are not provided within this document.

Data Provenance: The document does not specify the country of origin of the human serum samples. The studies are analytical validations performed retrospectively in a laboratory setting (e.g., Beckman Coulter AU480® automated clinical chemistry analyzer). It's not a prospective clinical trial with patient data.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is an analytical chemistry assay validation, not a diagnostic imaging AI/ML model. Therefore, "experts" in the sense of physicians establishing ground truth for patient cases are not applicable. The "ground truth" for the test set (e.g., concentration of Oxcarbazepine Metabolite) is established by highly accurate reference methods such as LC-MS/MS (for method comparison) or by precise gravimetric/volumetric preparation of controls and calibrators using certified reference materials. The qualifications of the personnel performing these analytical tests are implicitly assumed to be those typical for a laboratory setting conducting such validations.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is an analytical validation of an in-vitro diagnostic assay measuring a chemical concentration, not a study involving human readers or subjective interpretations requiring adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an analytical validation of an in-vitro diagnostic assay, not an AI-assisted diagnostic device for human readers.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is an "algorithm only" in the sense that it is an automated chemical assay system. Its performance (quantification of Oxcarbazepine Metabolite) is assessed independently through the various analytical studies (e.g., LOQ, linearity, precision, method comparison against LC-MS/MS). Human "human-in-the-loop" performance is not a direct component of the assay's function, though human operators are involved in running the assay and interpreting the results.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the analytical performance studies is primarily:

  • Reference Method: For method comparison, LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) is used as the reference "ground truth" method. LC-MS/MS is a highly accurate and precise analytical technique for quantifying specific compounds in complex mixtures.
  • Gravimetric/Volumetric Preparation: For calibrators and controls, the "ground truth" concentrations are established by precise gravimetric addition of certified Oxcarbazepine Metabolite powder to solvents. Purity is determined by NMR and elemental analysis.

8. The sample size for the training set

This is an immunoassay, not a machine learning or AI algorithm in the traditional sense that requires a "training set" for model development. The "training" of the assay refers to its calibration. The calibrators are prepared and value-assigned as described (e.g., "Two calibrated runs are performed using the Master Calibrator. In each run, five replicates of Master Lot (reference) and Test Lot are tested as matched pairs for each calibrator level.").


9. How the ground truth for the training set was established

As described above, for an immunoassay, the "training set" is the calibrator set. The ground truth for the calibrators is established through:

  • Traceability to certified powder: The calibrators are traceable to certified Oxcarbazepine Metabolite powder. "The purity of Oxcarbazepine Metabolite in the certified raw material is determined by NMR and elemental analysis as performed by the supplier of the certified powder."
  • Gravimetric Addition: "Bulk solutions of the ARK Oxcarbazepine Metabolite Calibrator are prepared volumetrically using a stock solution prepared by gravimetric addition of powder to solvent."
  • Value Assignment: "Testing is performed with the ARK Oxcarbazepine Metabolite Assay on the Beckman Coulter AU480® automated analyzer. Two calibrated runs are performed using the Master Calibrator. In each run, five replicates of Master Lot (reference) and Test Lot are tested as matched pairs for each calibrator level. Mean values for ten replicates are calculated." This process ensures consistency and accuracy against a master reference.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 9, 2016

ARK DIAGNOSTICS, INC. KENNETH KASPER EXECUTIVE DIRECTOR, QUALITY AND REGULATORY AFFAIRS 48089 FREMONT BLVD. FREMONT CA 94538

Re: K153596

Trade/Device Name: Ark Oxcarbazepine Metabolite Assay, Ark Oxcarbazepine Metabolite Calibrator, Ark Oxcarbazepine Metabolite Control Regulation Number: 21 CFR 862.3645 Regulation Name: Neuroleptic drugs radioreceptor assay test system Regulatory Class: II Product Code: POX, DLJ, LAS Dated: July 6, 2016 Received: July 8, 2016

Dear Kenneth Kasper:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153596

Device Name ARKTM Oxcarbazepine Metabolite Assay ARK™M Oxcarbazepine Metabolite Calibrator ARK™ Oxcarbazepine Metabolite Control

Indications for Use (Describe)

The ARK™ Oxcarbazepine Metabolite Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of Oxcarbazepine Metabolite in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of Oxcarbazeoine Metabolite to help ensure appropriate therapy.

The ARK™ Oxcarbazepine Metabolite Calibration of the ARK Oxcarbazepine Metabolite Assay.

The ARK™ Oxcarbazepine Metabolite Control is an assayed quality control material intended for use in quality control of the ARK Oxcarbazepine Metabolite Assay.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is K153596.

807.92 (a)(1): Name:ARK Diagnostics, Inc.
Address:48089 Fremont BlvdFremont, CA 94538
Owner Operator Number:10027663
Establishment Registration:3005755244
Phone:(510) 270-6270
FAX:(510) 270-6298
Contact:Kenneth C. Kasper, PhD – (510) 270-6280Executive Director of Quality and Regulatory Affairs

Date prepared: August 9, 2016

807.92 (a)(2): Device name- trade name and common name, and classification

Trade name:ARK™ Oxcarbazepine Metabolite AssayARK™ Oxcarbazepine Metabolite CalibratorARK™ Oxcarbazepine Metabolite Control
Common Name:Homogeneous Enzyme Immunoassay
Classification:21 CFR 862.3645 POX Neuroleptic drugs radioreceptor assay testsystem; Class II; 21 CFR 862.3200 DLJ; 21 CFR 862.3280 LAS

807.92 (a)(3): Identification of the legally marketed predicate device

Emit® 2000 Carbamazepine AssayK010814
ARK™ Topiramate CalibratorK083799
ARK™ Topiramate ControlK083799

{4}------------------------------------------------

807.92 (a)(4): Device Description

The ARK Oxcarbazepine Metabolite Assay is a homogeneous immunoassay based on competition between drug in the specimen and Oxcarbazepine Metabolite labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenyzme NAD functions only with the bacterial enzyme used in the assay.

The ARK Oxcarbazepine Metabolite Assay consists of reagents R1 anti-Oxcarbazepine Metabolite polyclonal antibody with substrate and R2 Oxcarbazepine Metabolite labeled with bacterial G6PDH enzyme. The ARK Oxcarbazepine Metabolite Calibrator consists of a six-level set to calibrate the assay, and the ARK Oxcarbazepine Metabolite Control consists of a threelevel set used for quality control of the assay.

807.92 (a)(5): Intended Use / Indications for Use

The ARKIM Oxcarbazepine Metabolite Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of Oxcarbazepine Metabolite in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of Oxcarbazepine Metabolite to help ensure appropriate therapy.

The ARKIM Oxcarbazepine Metabolite Calibrator is intended for use in calibration of the ARK Oxcarbazepine Metabolite Assay.

The ARKIM Oxcarbazepine Metabolite Control is an assayed quality control material intended for use in quality control of the ARK Oxcarbazepine Metabolite Assay.

For prescription use only. Caution: Federal Law restricts this device to sale by or on the order of a licensed practitioner.

Summary and Explanation

Oxcarbazepine [10, 11-dihydro-10-oxo-5H-dibenzo[b,f]azepine-5-carboxamide] and eslicarbazapine [{{}-10-acetoxy-10,11-dihydro-5H-dibenz|b,f|azepine- 5-carboxamide] are prodrugs that are metabolized to an active metabolite (10,11-dihydro-10-hydroxy-5Hdibenz[b.flazepine-5-carboxamide). Oxcarbazepine Metabolite is often called 10-monohydroxy derivative (MHD) or referred to as licarbazepine (Trileptal, Novartis) is metabolized to two enantiomers (S)-MHD and (R)-MHD at a metabolite ratio of approximately 4:1, respectively.

Eslicarbazepine acetate (Antion Pharmaceuticals) is prescribed as adjunctive therapy for partial-onset seizures associated with epilepsy in adults. Eslicarbazepine acetate is [(S)-10-Acetoxy-10,11-dihydro-5Hdibenz[b,f]azepine-5-carboxamide] and metabolism to (S)-MHD is favored such that the metabolite ratio of (S)-MHD to (R)-MHD is approximately 19:1.

{5}------------------------------------------------

ARK Oxcarbazepine Metabolite Calibrator

The ARK Oxcarbazepine Metabolite Calibrator is traceable to certified Oxcarbazepine Metabolite powder. The purity of Oxcarbazepine Metabolite in the certified raw material is determined by NMR and elemental analysis as performed by the supplier of the certified powder.

Bulk solutions of the ARK Oxcarbazepine Metabolite Calibrator are prepared volumetrically using a stock solution prepared by gravimetric addition of powder to solvent. The concentration of Oxcarbazepine Metabolite in the respective bulk solution must agree within 5% of its corresponding master calibrator.

Value Assignment: Testing is performed with the ARK Oxcarbazepine Metabolite Assay on the Beckman Coulter AU480® automated analyzer. Two calibrated runs are performed using the Master Calibrator. In each run, five replicates of Master Lot (reference) and Test Lot are tested as matched pairs for each calibrator level. Mean values for ten replicates are calculated. Test lot mean values are expected to match the Master lot mean values within 5% allowance.

REFProduct DescriptionQuantity/Volume
5032-0002-00ARK Oxcarbazepine Metabolite Calibrators*Dropper vials
Oxcarbazepine Metabolite, buffer, bovineserum albumin, and sodium azide
A0.0 µg/mL1 X 4 mL
B2.0 µg/mL1 X 2 mL
C5.0 µg/mL1 X 2 mL
D12.0 µg/mL1 X 2 mL
E25.0 µg/mL1 X 2 mL
F50.0 µg/mL1 X 2 mL

*To convert results from µg/mL Oxcarbazepine Metabolite, multiply µg/mL by 3.933. Oxcarbazepine Metabolite levels become 7.9, 19.7, 47.2, 98.3, and 196.6 µmol/L for Calibrators B to F respectively.

{6}------------------------------------------------

ARK Oxcarbazepine Metabolite Control

The ARK Oxcarbazepine Metabolite Control is traceable to certified Oxcarbazepine Metabolite powder. The purity of Oxcarbazepine Metabolite in the certified raw material is determined by NMR and elemental analysis as performed by the supplier of the certified powder.

Bulk solutions of the ARK Oxcarbazepine Metabolite Control are prepared volumetrically using a stock solution prepared by gravimetric addition of powder to solvent. ARK manufactures the controls to contain Oxcarbazepine Metabolite within 10% of the target levels.

Value Assignment: Testing is performed with the ARK Oxcarbazepine Metabolite Assay on the Beckman Coulter AU480® automated analyzer, calibrated with the master calibrator lot. Three calibrated runs are performed using four replicates of each level per run. The expected control ranges are set according to mean values. Each laboratory should establish the mean value for each control level and its own ranges for each new lot of controls.

REFProduct DescriptionQuality Control
5032-0003-00ARK Oxcarbazepine Metabolite Control* (4 mL)Expected Range
Oxcarbazepine Metabolite, buffer, bovine serumalbumin, and sodium azide (target level)(Mean µg/mL)
LOW (3.0 µg/mL)2.4 – 3.6
MID (10.0 µg/mL)8.5 – 11.5
HIGH (30.0 µg/mL)25.0 – 35.0

*To convert results from ug/mL Oxcarbazepine Metabolite, multiply µg/mL by 3.933. Oxcarbazepine Metabolite levels become 11.8, 39.3 and 118.0 µmol/L for LOW, MID and HIGH respectively.

{7}------------------------------------------------

807.92 (a)(6): Technological Similarities and Differences to the Predicate

SUBSTANTIAL EQUIVALENCE COMPARATIVE CHART

CharacteristicDevicePredicate
ARKTM Oxcarbazepine Metabolite AssayEmit® 2000 Carbamazepine Assay K010814
Intended UseThe ARKTM OxcarbazepineMetaboliteAssay is intended for the quantitativedetermination of Oxcarbazepine Metabolitein human serum on automated clinicalchemistry analyzers.The Emit® 2000 Carbamazepine Assay is ahomogeneous enzyme immunoassayintended for use in the quantitative analysisof carbamazepine in human serum or plasma.
Indications for UseThe measurements obtained are used inmonitoring levelsof OxcarbazepineMetaboliteto help ensure appropriatetherapy.The results obtained helps physiciansindividualize dosage regimens.
SampleSerumSerum or plasma
MethodologyHomogenous enzyme immunoassay (EIA)Homogenous enzyme immunoassay (EIA)
ReagentComponentsTwo (2) reagent system:Anti- Oxcarbazepine MetaboliteAntibody/Substrate Reagent (R1) containingrabbit polyclonal antibodies toOxcarbazepine Metabolite, glucose-6-phosphate, nicotinamide adeninedinucleotide, bovine serum albumin, sodiumazide, and stabilizersEnzyme Reagent (R2) containingOxcarbazepine Metabolite labeled withbacterial G6PDH, buffer, bovine serumalbumin, sodium azide, and stabilizersTwo (2) reagent system:Antibody/Substrate Reagent (R1) containingmouse monoclonal antibodies tocarbamazepine, glucose-6-phosphate,nicotinamide adenine dinucleotide.Enzyme Reagent (R2) containingcarbamazepine labeled with bacterialG6PDH, buffer.Sodium azide, buffer, preservatives, andstabilizers
Platform requiredAutomated clinical chemistry analyzerAutomated clinical chemistry analyzer
TestingenvironmentRoutine clinical laboratoryRoutine clinical laboratory
Reagent conditionand storageLiquid, 2-8° CLiquid, 2-8° C

{8}------------------------------------------------

Calibrators: Similarities and Differences
ItemCandidate Device (ARKOxcarbazepineCalibrator)Predicate Device (ARKTopiramate Calibrator;K083799)
Intended Use/Indicationsfor UseARK OxcarbazepineMetabolite Calibrator isintended for use incalibration of the ARKOxcarbazepine MetaboliteAssay.The ARK TopiramateCalibrator is intended foruse in calibration of theARK Topiramate Assay.
MatrixSynthetic protein matrix(buffer, bovine serumalbumin, preservatives)Same
Levels6Same
Controls: Similarities and Differences
ItemCandidate Device (ARK Oxcarbazepine Control)Predicate Device (ARK Topiramate Control; K083799)
Intended Use/Indications for UseThe ARK Oxcarbazepine Metabolite Control is an assayed quality control material intended for use in quality control of the ARK Oxcarbazepine Metabolite Assay.The ARK Topiramate Control is intended for use in quality control of the ARK Topiramate Assay.
MatrixSynthetic protein matrix (buffer, bovine serum albumin, preservatives)Same
Levels3 (LOW, MID, HIGH)Same

{9}------------------------------------------------

807.92 (b)(1) and 807.92 (b)(2): Brief Description of Nonclinical and Clinical Data

The following performance characteristics were obtained on the Beckman Coulter AU480® automated clinical chemistry analyzer. Unless otherwise stated, a metabolite S:R ratio of 9:1 was used to evaluate performance.

Limit of Quantitation (LOQ)

The LOQ of the ARK Oxcarbazepine Metabolite Assay was determined according to CLSI EP17-A2 and is defined as the lowest concentration for which acceptable inter-assay precision and recovery is observed (≤20% CV with ±15% recovery). The LOQ was determined to be 1.0 ug/mL, and may depend on analyzer-specific performance.

Recovery

Analytical recovery throughout the measurement range was assessed by adding concentrated Oxcarbazepine Metabolite into human serum negative for Oxcarbazepine Metabolite. The S:R ratio of each enantiomer was varied. The mean of six (6) replicate measurements of Oxcarbazepine Metabolite was tabulated as a function of the enantiomer ratio.

Mean Recovered Concentration (µg/mL)
TheoreticalConcentration(µg/mL)S:R1:1S:R4:1S:R9:1S:R19:1
1.00.770.930.980.95
4.03.783.923.943.86
8.07.478.188.167.82
15.014.1015.8014.9115.42
20.019.0321.6919.8121.02
35.033.7434.7133.5236.16
45.042.8946.8844.6349.46

{10}------------------------------------------------

Linearity

Linearity studies were performed as suggested in CLSI/NCCLS Protocol EP6-A. A 60.0 µg/mL serum sample was prepared and dilutions were made proportionally with human serum negative for Oxcarbazepine Metabolite. Oxcarbazepine Metabolite concentrations ranged from 1.0 to 50.0 ug/mL. Linearity at specific dilutions was considered acceptable if the percent difference was ±10% between the predicted 1st and 200 order regressed values or ≤ 0.20 µg/mL below 2.0 ug/mL. A linear relationship was demonstrated between 1.0 and 50.0 ug/mL (y = 1.0388x -0.0693). Results are shown below.

EstimatedValue(µg/mL)Results(µg/mL)1st OrderPredictedResults2nd OrderPredictedResultsDifference
1.001.000.971.110.14 µg/mL
3.003.193.053.112.2 %
5.005.145.125.120.0 %
10.0010.2610.3210.18-1.3 %
20.0021.0120.7120.41-1.4 %
30.0029.8831.0930.80-0.9 %
40.0041.9241.4841.36-0.3 %
50.0052.1351.8752.070.4 %

Assay Range

The measurement range of the ARK Oxcarbazepine Metabolite Assay is 1.0 to 37.0 ug/mL based on clinical concentrations tested. Report results below this range as <1.0 µg/mL or below the analyzer-specific lower LOQ established in your laboratory. Report results above this range as >37.0 ug/mL or test a diluted specimen having a concentration within the measurement range.

Specimens testing initially above the measurement range may be diluted in Calibrator A and retested. Multiply the assay result by the dilution factor to obtain the concentration of Oxcarbazepine Metabolite in the undiluted specimen.

{11}------------------------------------------------

Method Comparison

Correlation studies were performed using CLSI Protocol EP9-A3. Results from the ARK Oxcarbazepine Metabolite Assay were compared with results from LC-MS/MS. The Oxcarbazepine Metabolite concentrations ranged from 1.7 µg/mL to 36.4 µg/mL. Results of the Passing-Bablok regression analysis for the study are shown below (with 95% confidence limits).

Slope1.01(0.98 to 1.04)
y-intercept-0.38(-0.84 to 0.12)
Correlation Coefficient ( $r^2$ )0.95(0.94 to 0.97)
Number of Samples190

Image /page/11/Figure/3 description: This image is a scatter plot that compares two different methods of measuring Oxcarbazepine Metabolite. The x-axis represents the LC-MS/MS method in ug/mL, while the y-axis represents the ARK Oxcarbazepine Metabolite Assay method in ug/mL. The data points are clustered around a blue line, indicating a positive correlation between the two methods. The values on both axes range from 0.0 to 40.0 ug/mL.

{12}------------------------------------------------

Precision

Precision was determined as described in CLSI Protocol EP5-A3. Tri-level controls and three human serum pooled specimens containing Oxcarbazepine Metabolite were used in the study. Each level was assayed in quadruplicate twice a day for 20 days. Each of the runs per day was separated by at least two hours. The within-run, between-day, total SD, and percent CVs were calculated. Acceptance criterion: ≤10% CV.

Within RunBetween DayTotal
SampleNMean(µg/mL)SDCV(%)SDCV (%)SDCV (%)
ARK Control
LOW1603.00.124.00.124.10.175.7
MID16010.10.373.60.333.20.484.8
HIGH16030.20.993.31.193.91.545.1
Human Serum
LOW1603.10.123.90.124.00.175.5
MID16010.10.383.80.363.60.555.5
HIGH16030.41.103.61.113.71.555.1

Interfering Substances

Interference studies were conducted using CLSI Protocol EP7-A2 as a guideline. Clinically high concentrations of the following potentially interfering substances in serum with known levels of Oxcarbazepine Metabolite (approximately 3 and 30 µg/mL) were evaluated. Each sample was assayed using the ARK Oxcarbazepine Metabolite Assay, along with a serum control of Oxcarbazepine Metabolite. Measurement of Oxcarbazepine Metabolite resulted in ≤10% error in the presence of interfering substances at the levels tested.

Percentage Recovery
InterferingSubstanceInterferentConcentration3 µg/mLOxcarbazepineMetabolite30 µg/mLOxcarbazepineMetabolite
Human Albumin12 g/dL102.295.1
Bilirubin - conjugated70 mg/dL108.6100.2
Bilirubin - unconjugated70 mg/dL102.792.4
Cholesterol602 mg/dL96.5103.5
Human IgG12 g/dL93.193.1
Hemoglobin1000 mg/dL105.7100.7
Rheumatoid Factor1000 IU/mL101.0103.9
Triglycerides1000 mg/dL96.694.3
Uric Acid30 mg/dL107.595.5

{13}------------------------------------------------

Specificity

MHD-Glucuronide and Dihydroxy-carbamazepine (synonymous with dihydroxy-derivative of oxcarbazepine or DHD) are secondary metabolites of Oxcarbazepine Metabolite (MHD). Oxcarbazepine and Eslicarbazepine acetate are parent drugs for MHD. Carbamazepine and its metabolites (Dihydrocarbamazepine and Carbamazepine-epoxide) are compounds structurally similar to MHD. All were tested for crossreactivity at the concentrations listed in the presence of MHD (20 ug/mL) in serum. MHD-Glucuronide levels may appear in serum greater than MHD in cases of renal impairment". MHD-Glucuronide and DHD levels are not crossreactive.

The parent drug oxcarbazepine crossreacted 22.2% (as did Eslicarbazepine Acetate), although neither Oxcarbazepine nor Eslicarbazepine Acetate are expected to be present with MHD at a significant level due to rapid renal clearance. Carbamazepine and its metabolites also crossreacted in the assay; the possibility of co-therapy or transition of therapy should be considered.

MetaboliteLevel Tested(µg/mL)Percent Cross-ReactivityPercentInterference
MHD Glucuronide201.61.6
MHD Glucuronide400.0-0.1
MHD Glucuronide1001.57.4
MHD Glucuronide2001.010.5
(DHD) Dihydro-dihydroxycarbamazepine5.0-11.3-2.9
Oxcarbazepine20.022.222.6
Eslicarbazepineacetate20.022.122.4
Carbamazepine20.020.420.7
Dihydro –Carbamazepine5.06.01.5
Carbamazepine-epoxide10.013.66.9

Drug Interference

Other anti-epileptic or coadministered drugs tested did not crossreact with Oxcarbazepine Metabolite-selective antibody. A high concentration of each compound was spiked into normal human serum with known levels of Oxcarbazepine Metabolite (approximately 3 and 30 ug/mL) and assayed along with a serum control of Oxcarbazepine Metabolite. Measurement of Oxcarbazepine Metabolite resulted in <10% error in the presence of drug compounds at the levels tested.

{14}------------------------------------------------

#CompoundConcentration(µg/mL)Percentage Recovery
Oxcarbazepinemetabolite(3 µg/mL)Oxcarbazepinemetabolite(30 µg/mL)
1Acetaminophen20095.697.1
2Acetazolamide10099.990.3
3Acetylsalicylic acid100095.196.0
4Amikacin10091.792.0
5Amitriptyline10105.1101.1
6Amoxapine1099.398.0
7Amphotericin B10093.693.2
8Ampicillin10096.5100.2
9Ascorbic acid10092.891.1
10Baclofen10091.193.5
11Bupropion10109.698.8
12Caffeine10098.391.7
13Chloramphenicol25093.790.3
14Chlorpromazine1098.399.7
15Citalopram10102.999.3
16Clobazam10098.3103.2
17Clonazepam10104.699.2
18Cyclosporin A4091.290.2
19Diazepam20103.1100.3
20Digoxin1097.397.0
21Doxepin10107.4102.9
22Erythromycin20094.594.7
23Ethanol4000 (0.4%)91.6100.7
24Ethotoin10098.496.2
25Ethosuximide250103.2105.1
26Felbamate25093.093.8
27Fluoxetine2094.999.2
28Furosemide10095.292.8
29Gabapentin20092.2104.3
30Gentamicin10095.891.2
31Haloperidol10101.297.4
32Ibuprofen500103.391.6
33Imipramine10109.4100.4
35Kanamycin A20093.8109.0
35Lamotrigine40091.597.9
36Levetiracetam40097.794.7
37Lidocaine10096.897.7
38Lincomycin100090.7100.4
39Mephenytoin100100.797.3
40Mesoridazine1097.899.4
41Methicillin25093.596.2
#CompoundConcentration(µg/mL)Percentage Recovery
Oxcarbazepinemetabolite(3 µg/mL)Oxcarbazepinemetabolite(30 µg/mL)
42Naproxen600102.295.7
43Neomycin100095.6102.9
44Niacin10093.093.9
45Nitrazepam20106.398.5
46Nortriptyline10104.4102.0
47Olanzapine10105.8100.5
48Paroxetine1096.798.3
492-phenyl-2-ethyl-malonamide (PEMA)100094.693.9
50Penicillin V10095.493.8
51Perphenazine50104.9100.9
52Phenobarbital20090.294.7
53Phenytoin200100.199.6
54Pregabalin20091.590.2
55Primidone10095.092.4
56Procainamide10093.392.4
57Prochlorperazine10105.2101.6
58Ranitidine100102.1100.6
59Rifampin10093.392.7
60Risperidone10100.697.7
61Sertraline10098.993.4
62Spectinomycin10097.297.9
63Stiripentol10093.899.7
64Sulfamethoxazole400100.597.5
65Theophylline200100.5100.8
66Thioridazine10103.998.0
67Tobramycin10094.5101.3
68Tiagabine20091.693.5
69Topiramate25092.891.7
70Trimethoprim40101.293.6
71Valproic Acid60092.793.0
72Vancomycin250101.392.6
73Vigabatrin150103.296.9
74Zonisamide40092.191.4

ARK Diagnostics, Inc. – 510(k) Summary Rev 08
ARK Oxcarbazepine Metabolite Assay

{15}------------------------------------------------

{16}------------------------------------------------

Sample Stability

Serum specimens were shown to be stable for at least forty-eight (48) hours at room temperature (22 °C), fourteen (14) days when refrigerated (2-8 °C), frozen (-20 °C) for at least 3 months and after three (3) successive freeze/thaw cycles based on supporting data.

Calibration Curve Stability

A stored calibration was effective for at least 15 days based on supporting data.

807.92 (b)(3): Conclusions from Nonclinical Testing

As summarized above, the ARK Oxcarbazepine Metabolite Assay, the ARK Oxcarbazepine Metabolite Calibrator and the ARK Oxcarbazepine Metabolite Control are substantially equivalent to the ARK™ Topiramate Assay system. The ARK Oxcarbazepine Metabolite Assay system was shown to be safe and effective for its intended use based on performance studies.

§ 862.3645 Neuroleptic drugs radioreceptor assay test system.

(a)
Identification. A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.(b)
Classification. Class II.