K Number
K153561

Validate with FDA (Live)

Date Cleared
2017-02-16

(430 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AG-607 blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, calf, or thigh. The AG-607 blood glucose monitoring system is intended to be used by a Single person and should not be shared.

The AG-607 blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-607 blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative Site Testing (AST) should be done only during steady-state times (when glucose levels is not changing rapidly).

The EGS-2003 test strips are for use with the AG-607 blood glucose meter to quantitatively measure glucose(sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, calf or thigh.

The AG-607Multi blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, calf, or thigh. The AG-607Multi blood glucose monitoring system is intended for testing outside the body(in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as and aid to monitor the effectiveness of diabetes control program. The system is only used with single-use auto-disabling lancing device.

The AG-607Multi blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative Site Testing (AST) should be done only during steady-state times (when glucose levels is not changing rapidly).

The EGS-2003Multi test strips are for use with the AG-607Multi blood glucose meter to quantitatively measure glucose(sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.

Device Description

The AG-607 and AG-607 Multi Glucose Monitoring System (BGMS) consist of blood glucose meter, single use test strips, sterile lancets, lancing device and the control solutions.

They are based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. It uses the same technological characteristics for testing with its predicate device.

AI/ML Overview

The document describes the AG-607 and AG-607 Multi Blood Glucose Monitoring Systems. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are based on ISO 15197 for in vitro diagnostic test systems for self-testing in managing diabetes mellitus. The document explicitly states that the device conforms to this standard. However, specific numerical performance criteria from ISO 15197 and the device's reported numerical performance against those criteria are not provided in detail in this document.

The document mentions a "Measurement Range" for the device. While this is a performance characteristic, it's not presented as an acceptance criterion against which specific test results are directly compared in a table.

Implied Performance (from device specifications):

Feature/SpecificationAG-607 Blood Glucose Monitoring System
Detection MethodAmperometry
EnzymeGlucose dehydrogenase
Sample SourceCapillary whole blood from AST and finger
Hematocrit Range20-60%
Operating Temperature Range$10\degree C~40\degree C (50\degree-104\degree F)$
Test Time5 second
Measurement Range20mg/dL-600mg/dL (1.1mmol/L~33.3mmol/L)
Sample VolumeMinimum 0.7 micro liter

2. Sample Size Used for the Test Set and Data Provenance

The document states "Non-clinical test and the clinical test are done according to the above standard [ISO 15197]". However, actual sample sizes for the test set (both non-clinical and clinical) are not specified in this document. The data provenance (e.g., country of origin of the data, retrospective or prospective) is also not explicitly mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not provided in the document. For a blood glucose monitoring system, the ground truth is typically established against a laboratory reference method, not by human experts.

4. Adjudication Method for the Test Set

This information is not provided in the document. Since blood glucose measurements are quantitative and typically compared against a reference method, an adjudication method for conflicting expert opinions would not typically apply.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This is not applicable as the device is a blood glucose monitoring system, not an AI-assisted diagnostic tool that involves human readers interpreting cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The device is a standalone blood glucose monitoring system. Its performance is evaluated independently of human interpretation of complex images or data. The document states "Non-clinical test and the clinical test are done according to the above standard [ISO 15197]", implying a standalone performance evaluation.

7. The Type of Ground Truth Used

While not explicitly stated, for blood glucose monitoring systems complying with ISO 15197, the ground truth is typically established using a laboratory reference method (e.g., a YSI analyzer) which is considered highly accurate for glucose concentration.

8. The Sample Size for the Training Set

This information is not provided in the document. Blood glucose measurement devices typically do not involve a "training set" in the machine learning sense. Instead, they are calibrated during manufacturing and their accuracy is verified through testing.

9. How the Ground Truth for the Training Set Was Established

This information is not provided in the document and is generally not applicable in the context of such devices. Calibration and verification processes ensure the device's accuracy against known glucose concentrations.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping to create a sense of depth.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

ANDON HEALTH CO., LTD YI LIU PRESIDENT NO. 3 JINPING STREET, YA AN ROAD, NANKAI DISTRICT TIANJIN 300190 CHINA

Re: K153561

Trade/Device Name: AG-607 Blood Glucose Monitoring System AG-607 Multi Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR Dated: January 20, 2017 Received: January 25, 2017

Dear Yi Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153561

Device Name

AG-607 Blood Glucose Monitoring System

Indications for Use (Describe)

The AG-607 blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, calf, or thigh. The AG-607 blood glucose monitoring system is intended to be used by a Single person and should not be shared.

The AG-607 blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-607 blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative Site Testing (AST) should be done only during steady-state times (when glucose levels is not changing rapidly).

The EGS-2003 test strips are for use with the AG-607 blood glucose meter to quantitatively measure glucose(sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, calf or thigh.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K153561

Device Name

AG-607 Multi Blood Glucose Monitoring System

Indications for Use (Describe)

The AG-607Multi blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, calf, or thigh. The AG-607Multi blood glucose monitoring system is intended for testing outside the body(in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as and aid to monitor the effectiveness of diabetes control program. The system is only used with single-use auto-disabling lancing device.

The AG-607Multi blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative Site Testing (AST) should be done only during steady-state times (when glucose levels is not changing rapidly).

The EGS-2003 test strips are for use with the AG-607Multi blood glucose meter to quantitatively measure glucose(sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 Submitter's information

Name:Andon Health Co., Ltd.
Address:No.3 Jinping Street,Ya'an Road, Nankai District, Tianjin,P.R. China
Phone number:86-22-6052 6161
Fax number:86-22-6052 6162
Contact:Yi Liu
Date of Preparation:11/17/2015

2.0 Device information

Trade name:AG-607 Blood Glucose Monitoring System
AG-607 Multi Blood Glucose Monitoring System
Common name:Blood Glucose Monitoring System
Classification name:Blood Glucose Monitoring System

3.0 Classification

Product codeClassificationRegulation SectionPanel
NBWII21 CFR 862.1345Chemistry (75)
LFRII21 CFR 862.1345Chemistry (75)

4.0 Predicate device information

Manufacturer:Andon Medical Co., Ltd.
Device:AG-608N Single Blood Glucose Monitoring System
510(k) number:K110017

{5}------------------------------------------------

5.0 Device description

The AG-607 and AG-607 Multi Glucose Monitoring System (BGMS) consist of blood glucose meter, single use test strips, sterile lancets, lancing device and the control solutions.

They are based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. It uses the same technological characteristics for testing with its predicate device.

6.0 Intended use

The AG-607 blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf, or thigh. The AG-607 blood glucose monitoring system is intended to be used by a Single person and should not be shared.

The AG-607 blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-607 blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative Site Testing (AST) should be done only during steady-state times (when glucose levels is not changing rapidly).

The EGS-2003 test strips are for use with the AG-607 blood glucose meter to quantitatively measure glucose(sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.

The AG-607Multi blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf, or thigh. The AG-607Multi blood glucose monitoring system is intended for testing outside the body(in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as and aid to monitor the effectiveness of diabetes control program. The system is only used with single-use auto-disabling lancing device.

{6}------------------------------------------------

The AG-607Multi blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative Site Testing (AST) should be done only during steady-state times (when glucose levels is not changing rapidly).

The EGS-2003Multi test strips are for use with the AG-607Multi blood glucose meter to quantitatively measure glucose(sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.

CHARACTERISTICSNEW DEVICE:AG-607 Blood GlucoseMonitoring SystemPREDICATE:AG-608N Single BloodGlucose Monitoring System(K110017)
Detection MethodAmperometryAmperometry
EnzymeGlucose dehydrogenaseGlucose Oxidase
Type of MeterBiosensor (Electrode)Biosensor (Electrode)
Sample SourceCapillary whole blood fromAST(Alternative site testing)and fingerCapillary whole blood fromAST(Alternative site testing)and finger
Sample ApplicationBlood sample is placeddirectly to the test strip afterfinger or AST is lanced.Blood sample is placeddirectly to the test strip afterfinger or AST is lanced.
Hematocrit Range20-60%20-60%
OperatingTemperature Range$10℃~40℃(50°-104°F)$$10℃~40℃(50°-104°F)$
Dimensions110mm x 52mm x 20.5mm87mm×53mm ×9.9mm
DisplayLCDLCD
Result Presentationmg/dL or mmol/Lmg/dL or mmol/L
Memory Capabilities500 times500 times
Test StartAutomaticAutomatic
Test Time5 second5 second
Power SourceDC3.0V (2×LR03)DC 3.0V (CR2032)
Battery Life1000 times500 times
Measurement Range20mg/dL-600mg/dL(1.1mmol/L~33.3mmol/L)20mg/dL-600mg/dL(1.1mmol/L~33.3mmol/L)
Qualified Test StripEGS-2003 Test StripAGS-1000N Test Strip

7.0 Summary comparing technological characteristics with predicate device

{7}------------------------------------------------

Sample VolumeMinimum 0.7 micro literMinimum 0.7 micro liter
Connect MethodNAUSB
OutlookSee Picture 1See Picture 2

Picture 1: AG-607

Image /page/7/Picture/3 description: The image shows a white Andon blood glucose meter displaying a reading of 94 mg/dL. The time displayed on the meter is 8:01 AM. The meter has a digital display and a simple design with the brand name "andon" printed at the top.

Picture 2 AG-608N

Image /page/7/Picture/5 description: The image shows a white, rectangular device with a screen at the top and two triangular buttons labeled "M" and "S" below the screen. The screen is dark and appears to be off. The device has a simple, minimalist design and is likely an electronic device such as a timer or a remote control.

8.0 Performance summary

The AG-607 and AG-607 Multi Blood Glucose Monitoring System (BGMS) conform to the following standards:

  • · ISO 15197: In vitro diagnostic test systems- Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus .
    Non-clinical test and the clinical test are done according to the above standard.

9.0 Comparison to the predict device and the conclusion

The proposed device AG-607 is similar with the predicate device AG-608N, they are both for single patient use, and can test the blood glucose at the alternative site. The hematocrit range, the use function are all the same. The appearance, internal power source and the qualified test strip of the two device is different. Moreover, the new device AG-607 meter does not have the USB function.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.