K Number
K153466

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-01-14

(44 days)

Product Code
Regulation Number
876.1330
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PillCam COLON 2 capsule endoscopy system is intended to provide visualization of the colon. It may be used for detection of colon polyps in patients after an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible, In addition, it is intended for detection of colon polyps in patients with evidence of gastrointestinal bleeding of lower Gl origin. This applies only to patients with major risks for colonoscopy or moderate sedation, but who could tolerate colonoscopy and moderate sedation in the event a clinically significant colon abnormality was identified on capsule endoscopy.

Device Description

The Given PillCam Endoscopy system is comprised of four main subsystems; (1) the ingestible PillCam capsule, (2) the DR 3 PillCam Recorder, (3) the RAPID software, and (4) the Given Workstation and Accessories.

  1. Ingestible PillCam Capsule The disposable, ingestible PillCam Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via a RF communication channel to the PillCam Recorder located outside the body.
  2. DR 3 PillCam Recorder Recorder is an external The DR 3 PillCam receiving/recording unit that receives and stores the acquired images from the capsule. The Sensor Array/belt receives data from the PillCam capsule and transfers the data to the PillCam Recorder.
  3. RAPID Software The RAPID Software is a software application that is utilized to process, analyze, store, and view the acquired images collected from the PillCam Recorder to create a RAPID video of the images. The software also includes a reporting function to create detailed clinical reports, inservice training videos, and patient instruction forms.
  4. Given Workstation and Accessories The Workstation is a modified standard personal computer that is the operational platform for the RAPID software. Other accessories include a flat panel LCD monitor, a highcapacity mass storage device, and a high-capacity USB portable storage device.
AI/ML Overview

This document is a 510(k) summary for the PillCam COLON 2 Capsule Endoscopy System. It requests an expanded indication for use for detection of colon polyps.

Based on the provided text, the device itself (PillCam COLON 2 Capsule Endoscopy System) is identical to a previously cleared device (DEN 120023). No new clinical testing was performed or relied upon for this specific 510(k) submission for the expanded indication. The submission relies on the substantial equivalence of the existing device and its performance against "special controls for colon capsule imaging systems, as described in 21CFR 876.1330."

Therefore, the input does not provide information about a new study directly demonstrating the device meets specific acceptance criteria for this expanded indication. Instead, it states that the device's technological characteristics and optical features are the same as the predicate device (DEN 120023).

In summary, the provided text does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria, because the submission claims substantial equivalence without new clinical testing.

However, given the constraint to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, I will extract what is explicitly stated in the document about general performance data and and the basis for the approval.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of acceptance criteria and reported device performance for this expanded indication. It only states that the "device meets the special controls for colon capsule imaging systems, as described in 21CFR 876.1330."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The document explicitly states: "No clinical testing was performed or relied on for a determination of substantial equivalence."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no new clinical testing was performed for this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no new clinical testing was performed for this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as no new clinical testing was performed for this submission. This device is not an AI-assisted device in the context of this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This would not apply to a capsule endoscopy system. The system provides visualization, and a human (physician) interprets the images.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no new clinical testing was performed for this submission.

8. The sample size for the training set

Not applicable, as no new clinical testing was performed for this submission, and the device is not described as involving a machine learning algorithm that undergoes a "training set" in the context of this approval.

9. How the ground truth for the training set was established

Not applicable.

Conclusion based on the provided text:

The submission for the expanded indication of the PillCam COLON 2 Capsule Endoscopy System relies on the substantial equivalence to a predicate device (DEN 120023). The document explicitly states: "No clinical testing was performed or relied on for a determination of substantial equivalence." The acceptance criteria are broadly described as meeting "the special controls for colon capsule imaging systems, as described in 21CFR 876.1330," rather than specific performance metrics from a new clinical study.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 14, 2016

Given Imaging, Ltd. Hilla Debby Director, Clinical & Regulatory 2 Hacarmel St. New Industrial Park POB 258 Yoqneam, 20962 Israel

Re: K153466 Trade/Device Name: PillCam COLON 2 Capsule Endoscopy System Regulation Number: 21 CFR§ 876.1330 Regulation Name: Colon capsule imaging system Regulatory Class: II Product Code: PGD Dated: November 30, 2015 Received: December 1, 2015

Dear Hilla Debby,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K153466

Device Name

PillCam COLON 2 Capsule Endoscopy Sysytem

Indications for Use (Describe)

The PillCam COLON 2 capsule endoscopy system is intended to provide visualization of the colon. It may be used for detection of colon polyps in patients after an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible, In addition, it is intended for detection of colon polyps in patients with evidence of gastrointestinal bleeding of lower Gl origin. This applies only to patients with major risks for colonoscopy or moderate sedation, but who could tolerate colonoscopy and moderate sedation in the event a clinically significant colon abnormality was identified on capsule endoscopy.

Type of Use (Select one or both, as applicable)

Prescription Use (21 CFR 801 Subpart D)
Over The Counter Use (21 CFR 801 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 80

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Given Imaging. The logo is in blue and white, and it features the word "GIVEN" in large, white letters. Below the word "GIVEN" is the word "IMAGING" in smaller, white letters. To the right of the logo are several blue squares with rounded corners.

510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.

Submitter Nameand Address:Given Imaging Ltd. (GI Solutions, Medtronic)2 Hacarmel StreetNew Industrial ParkPO Box 258Yokneam 20692IsraelTel.: 011-972-4-9097774Fax: 011-972-73-2501533
Contact Person:Hila DebbyDirector, Clinical & RegulatoryEmail: Hila.Debby@covidien.com
Phone Number:972 (4) 909-7774
Fax Number:(972) 73-2501533
EstablishmentRegistrationNumber:9710107
Date Prepared:January 13, 2016
Device TradeName(s):PillCam® COLON 2 capsule endoscopy system
Device CommonName:Colon capsule imaging system
Classification:Regulation No: 876.1330, Class: IIPanel: Gastroenterology/UrologyPGD- Colon Capsule Imaging System
PredicateDevice(s):PillCam COLON 2 capsule endoscopy system (DEN 120023)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image features the logo for Given Imaging, a medical technology company. The logo is set against a blue background and includes the company name in white, with the "V" in "Given" stylized as a green triangle pointing downwards. To the right of the logo are several blue squares of varying shades, arranged in a pattern.

General Device Description:

The Given PillCam Endoscopy system is comprised of four main subsystems; (1) the ingestible PillCam capsule, (2) the DR 3 PillCam Recorder, (3) the RAPID software, and (4) the Given Workstation and Accessories.

    1. Ingestible PillCam Capsule The disposable, ingestible PillCam Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via a RF communication channel to the PillCam Recorder located outside the body.
      1. DR 3 PillCam Recorder

Recorder is an external The DR 3 PillCam receiving/recording unit that receives and stores the acquired images from the capsule. The Sensor Array/belt receives data from the PillCam capsule and transfers the data to the PillCam Recorder.

3. RAPID Software

The RAPID Software is a software application that is utilized to process, analyze, store, and view the acquired images collected from the PillCam Recorder to create a RAPID video of the images. The software also includes a reporting function to create detailed clinical reports, inservice training videos, and patient instruction forms.

4. Given Workstation and Accessories

The Workstation is a modified standard personal computer that is the operational platform for the RAPID software. Other accessories include a flat panel LCD monitor, a highcapacity mass storage device, and a high-capacity USB portable storage device.

Indications for use: The PillCam COLON 2 capsule endoscopy system is intended to provide visualization of the colon. It may be used for detection of colon polyps in patients after an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible. In addition, it is intended for detection of colon polyps in patients with evidence of gastrointestinal bleeding of lower Gl origin. This applies only to patients with major risks for colonoscopy or moderate sedation, but who could tolerate colonoscopy and

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image features the logo for Given Imaging, a medical technology company. The logo is set against a blue background and consists of the company name in white lettering, with a stylized image of a capsule or pill integrated into the letter 'V'. To the right of the logo are several rounded squares in varying shades of blue, creating a visual pattern.

moderate sedation in the event a clinically significant colon abnormality was identified on capsule endoscopy.

According to the expanded indication above, PilllCam COLON 2 capsule can be used for detection of polyps in a group of patients who may not be good candidates for colonoscopy. Pillcam COLON 2 capsule procedure would help the doctors decide whether or not they want to perform the scoping, when there is a strong clinical suspicion of a large polyp or cancer and the patient is at significant risk with colonoscopy or sedation but can tolerate colonoscopy if necessary.

  • The technological characteristics of the system (Ingestible Technological Characteristics: PillCam Capsule, DR 3 PillCam Recorder, RAPID Software, Given Workstation and Accessories) are identical to the predicate device (DEN120023).
    Optical Features Same as the predicate device (DEN 120023)

Bench testing: Same as the predicate device (DEN 120023)

  • Clinical testing No clinical testing was performed or relied on for a determination of substantial equivalence.
  • Performance Data: The device meets the special controls for colon capsule imaging systems, as described in 21CFR 876.1330.
  • Conclusion: The device that is submitted for an expanded indication is identical to an already cleared device (DEN 120023).

§ 876.1330 Colon capsule endoscopy system.

(a)
Identification. A prescription, single-use ingestible capsule designed to acquire video images during natural propulsion through the digestive system. It is specifically designed to visualize the colon for the detection of polyps. It is intended for use only in patients who had an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The capsule must be demonstrated to be biocompatible.
(2) Non-clinical testing data must demonstrate the mechanical and functional integrity of the device under physically stressed conditions. The following performance characteristics must be tested and detailed protocols must be provided for each test:
(i) Bite test to ensure that the capsule can withstand extreme cases of biting.
(ii) pH resistance test to evaluate integrity of the capsule when exposed to a range of pH values.
(iii) Battery life test to demonstrate that the capsule's operating time is not constrained by the battery capacity.
(iv) Shelf-life testing to demonstrate that the device performs as intended at the proposed shelf-life date.
(v) Optical testing to evaluate fundamental image quality characteristics such as resolution, field of view, depth of field, distortion, signal-to-noise ratio, uniformity, and image artifacts. A test must be performed to evaluate the potential of scratches, caused by travelling through the gastrointestinal tract, on the transparent window of the capsule and their impact on the optical and color performance.
(vi) An optical safety analysis must be performed based on maximum (worst-case) light exposure to internal gastrointestinal mucosa, and covering ultraviolet, visible, and near-infrared ranges, as appropriate. A mitigation analysis must be provided.
(vii) A color performance test must be provided to compare the color differences between the input scene and output image.
(viii) The video viewer must clearly present the temporal or spatial relationship between any two frames as a real-time lapse or a travel distance. The video viewer must alert the user when the specific video interval is captured at a frame rate lower than the nominal one due to communication errors.
(ix) A performance test evaluating the latency caused by any adaptive algorithm such as adjustable frame rate must be provided.
(x) If the capsule includes a localization module, a localization performance test must be performed to verify the accuracy and precision of locating the capsule position within the colon.
(xi) A data transmission test must be performed to verify the robustness of the data transmission between the capsule and the recorder. Controlled signal attenuation should be included for simulating a non-ideal environment.
(xii) Software validation, verification, and hazards analysis must be provided.
(xiii) Electrical equipment safety, including thermal and mechanical safety and electromagnetic compatibility (EMC) testing must be performed. If the environments of intended use include locations outside of hospitals and clinics, appropriate higher immunity test levels must be used. Labeling must include appropriate EMC information.
(xiv) Information demonstrating immunity from wireless hazards.
(3) The clinical performance characteristics of the device for the detection of colon polyps must be established. Demonstration of the performance characteristics must include assessment of positive percent agreement and negative percent agreement compared to a clinically acceptable alternative structural imaging method.
(4) Clinician labeling must include:
(i) Specific instructions and the clinical and technical expertise needed for the safe use of the device.
(ii) A detailed summary of the clinical testing pertinent to use of the device, including the percentage of patients in which a polyp was correctly identified by capsule endoscopy, but also the percent of patients in which the capsule either missed or falsely identified a polyp with respect to the clinically acceptable alternative structural imaging method.
(iii) The colon cleansing procedure.
(iv) A detailed summary of the device technical parameters.
(v) A detailed summary of the device- and procedure-related complications pertinent to use of the device.
(vi) An expiration date/shelf life.
(5) Patient labeling must include:
(i) An explanation of the device and the mechanism of operation.
(ii) Patient preparation procedure.
(iii) A brief summary of the clinical study. The summary should not only include the percentage of patients in which a polyp was correctly identified by capsule endoscopy, but also the percent of patients in which the capsule either missed or falsely identified a polyp with respect to the clinically acceptable alternative structural imaging method.
(iv) A summary of the device- and procedure-related complications pertinent to use of the device.