K Number
K153436

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-06-06

(192 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Lateral Spine Truss System (STS) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-L5. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).

Device Description

The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone growth and fusion. The device is available in a variety of sizes to accommodate the patient's anatomy. The implant is made from Ti6Al4V alloy (ASTM F-1108).

AI/ML Overview

This document refers to a 510(k) premarket notification for a medical device called the "LATERAL Spine Truss System." It is a regulatory submission, not a study report, and therefore does not contain details about specific acceptance criteria or a study that "proves" the device meets acceptance criteria in the way a clinical trial or performance study would.

Instead, the document details the regulatory process by which the device was found substantially equivalent to existing devices. Substantial equivalence in the FDA 510(k) pathway means proving that a new device is as safe and effective as a legally marketed predicate device. This is often demonstrated through comparative analysis of technological characteristics, materials, indications for use, and sometimes non-clinical performance testing.

Here's an attempt to answer your questions based on the provided text, while also highlighting what information isn't present:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria in the typical sense (e.g., target accuracy, sensitivity, specificity values) for a diagnostic AI device, nor does it present "reported device performance" in terms of clinical outcomes or diagnostic metrics. This is a spinal implant, and the "performance" discussed is in terms of mechanical testing and equivalence to predicates.

Here's what can be inferred for mechanical testing:

Acceptance Criteria (Implied)Reported Device Performance
Mechanical performance comparable to predicate devices under various loading scenarios (compression, combined compression and shear)."Validated FEA... was conducted to evaluate the mechanical performance of the devices under different loading scenarios, including pure compression and combined compression and shear."
Subsidence performance meeting ASTM F2267-04."Other mechanical tests included subsidence per ASTM F2267-04."
Expulsion performance meeting industry accepted methodology."and expulsion testing per an industry accepted methodology."
Overall safety and effectiveness comparable to predicate devices."These comparisons demonstrate substantial equivalence to the predicate devices."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document describes mechanical testing, not a clinical study involving a "test set" of patient data in the context of AI. The "test set" would typically refer to a dataset used to evaluate an AI model's performance on unseen data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. This document is for a spinal implant, and the assessment of its substantial equivalence relies on mechanical testing and comparison to predicate devices, not on expert ground truth labeling of medical images or diagnostic classification.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided. This concept is relevant for studies involving human readers or AI in diagnostic tasks, where disagreements in labels or interpretations might need an adjudication process. It does not apply to the mechanical testing described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not provided. MRMC studies are used to evaluate diagnostic systems, especially AI-assisted ones, and compare them against human performance. This document is for a physical orthopedic implant and does not involve AI assistance for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided. The device is a physical implant, not an algorithm. Therefore, "standalone" algorithm performance is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mechanical testing, the "ground truth" implicitly refers to established engineering standards and measurements (e.g., ASTM F2267-04) and the performance characteristics of the predicate devices. The goal is to show the new device performs equivalently or acceptably according to these engineering benchmarks, rather than clinical ground truth like pathology or outcomes data in the context of an AI study.

8. The sample size for the training set

This information is not provided. The concept of a "training set" applies to machine learning algorithms. This document is for a physical medical device.

9. How the ground truth for the training set was established

This information is not provided as it's not applicable. There is no "training set" in the context of this regulatory submission for an orthopedic implant.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 6, 2016

4WEB. Incorporated % Rich Jansen, Ph.D. Consultant Silver Pine Consulting, LLC. 11821 Bramble Cove Drive Ft. Myers, Florida 33905

Re: K153436

Trade/Device Name: LATERAL Spine Truss System Regulation Number: 21 CFR 888.3080 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MAX Dated: May 9, 2016 Received: May 10, 2016

Dear Dr. Jansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

K153436

Page 1 of 1

510(k) Number (if known) K153436

Device Name LATERAL Spine Truss System

Indications for Use (Describe)

The Lateral Spine Truss System (STS) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-L5. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared:June 2, 2016
Contact:Jessee Hunt, President4WEB, Inc.6170 Research Rd. Suite 219Frisco, TX 75033Phone: (800) 285-7090Fax: 972-488-1816
Regulatory Contact:Rich Jansen, Pharm. D.Silver Pine Consulting, LLCrichj@s-pineconsulting.com
Trade Name:LATERAL Spine Truss System
Product Class:Class II
Classification:21 CFR §888.3080
Common Name:Intervertebral Body Fusion Device
Product Codes:MAX
Panel Code:87

Indications for Use:

The Lateral Spine Truss System (STS) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-L5. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).

Device Description:

The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone growth and fusion. The device is available in a variety of sizes to accommodate the patient's anatomy. The implant is made from Ti6Al4V alloy (ASTM F-1108).

Predicate Device(s):

The 4WEB LATERAL STS is substantially equivalent to the primary predicate, the 4WEB ALIF STS (K112316). Additional predicated devices are the 4WEB PSTS (K143258), 4WEB ALIF STS (K083894), and Medtronic Clydesdale Spinal System (K132897).

Performance Standards:

Validated FEA, comparing the subject device to PLIF STS (K143258), was conducted to evaluate the mechanical performance of the devices under different loading scenarios, including pure

{4}------------------------------------------------

compression and combined compression and shear. Other mechanical tests included subsidence per ASTM F2267-04 and expulsion testing per an industry accepted methodology.

Technological Characteristics:

4Web, Inc. has compared the Lateral STS to the predicate devices in regards to indications for use, materials, function, sizes and mechanical test results. These comparisons demonstrate substantial equivalence to the predicate devices.

Conclusion:

4WEB, Inc. concludes that these LATERAL STS devices are substantially equivalent to the predicate devices and raise no new questions of safety or effectiveness.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.