K Number
K153410

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-05-15

(537 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing can be used to cover and protect catheter sites and to secure devices to skin. Common applications include securing I.V. catheters, other intravascular catheters and percutaneous devices.

Tegaderm™ CHG I.V. Securement Dressing is intended to reduce vascular catheter colonization and catheter-related bloodstream infections (CRBSI) in patients with central venous or arterial catheters.

Device Description

3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing is used to cover and protect catheter sites and to secure devices to skin. It is available in a variety of shapes and sizes.

Tegaderm™ CHG I.V. Securement Dressing consists of a transparent adhesive dressing and an integrated gel pad containing 2% w/w Chlorhexidine Gluconate (CHG), a well-known antiseptic agent with broad spectrum antimicrobial and antifungal activity.

The transparent film provides an effective barrier against external contamination including fluids (waterproof), bacteria, and yeast, and protects the I.V. site.

In vitro testing (log reduction and barrier testing) demonstrates that the Tegaderm™ CHG gel pad in the Tegaderm™ CHG I.V. Securement Dressing has an antimicrobial effect against, and is a barrier to, a variety of gram-positive and gram-negative bacteria, and yeast in the dressing. The gel pad absorbs fluid.

Tegaderm™ CHG I.V. Securement dressing is transparent, allowing continual site observation, and is breathable, allowing good moisture vapor exchange.

The device is single-use, provided sterile and the sterilization method is ethylene oxide. There are no associated accessories.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes the performance of the 3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing in reducing Catheter-Related Bloodstream Infections (CRBSI). While explicit "acceptance criteria" (i.e., pre-defined thresholds the device must meet to be cleared) are not stated in a tabular format, the study results demonstrate the device's efficacy, which served as the basis for its expanded indication.

Implicit Acceptance Criteria (Based on Study Findings supporting Expanded Indication):

MetricAcceptance Criterion (Implicitly met by demonstrated performance)Reported Device Performance (Tegaderm™ CHG I.V. Securement Dressing)Control Group Performance (non-CHG dressing)P-value
CRBSI Reduction (Timsit Definition)Statistically significant reduction in CRBSI rate0.96% (9/938 patients)2.2% (21/941 patients)0.020
CRBSI Infections per 1000 Catheter Days (Timsit Definition)Statistically significant reduction in CRBSI rate per 1000 catheter-days0.521.29N/A (derived from patient reduction)
CRBSI Reduction (IDSA Definition - Sensitivity Analysis)Statistically significant reduction in CRBSI rate0.85% (8/938 patients)1.7% (16/941 patients)0.095
CRBSI Infections per 1000 Catheter Days (IDSA Definition)Statistically significant reduction in CRBSI rate per 1000 catheter-days0.460.94N/A (derived from patient reduction)
Percent Reduction in Patients with Infections (Timsit Definition)Significant percentage reduction57% reductionN/AN/A
Percent Reduction per 1000 Catheter Days (Timsit Definition)Significant percentage reduction60% reductionN/AN/A
Percent Reduction in Patients with Infections (IDSA Definition)Significant percentage reduction50% reductionN/AN/A
Percent Reduction per 1000 Catheter Days (IDSA Definition)Significant percentage reduction51% reductionN/AN/A

Additional Performance Data Mentioned (In Vitro):

  • Antimicrobial Effect: Log reduction and barrier testing demonstrate antimicrobial and antifungal effect against gram-positive and gram-negative bacteria, and yeast.
  • Suppression of Regrowth: Demonstrated suppression of regrowth over time.

2. Sample Size and Data Provenance

  • Sample Size for Test Set: The randomized controlled clinical trial included 1,879 subjects with 4,163 central venous and arterial catheters.
  • Data Provenance: The study was conducted at 11 hospitals (12 ICUs) in France. The data is prospective, as it was a randomized controlled trial.

3. Number of Experts and their Qualifications for Ground Truth

The document does not explicitly state the number of experts or their qualifications used to establish the ground truth for diagnosing CRBSI within the study. However, the study authors (Timsit JF et al.) likely followed established clinical protocols and expert consensus for CRBSI diagnosis within a critical care setting.

4. Adjudication Method for the Test Set

The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). However, the definition of CRBSI used in the Timsit study (and subsequent re-analysis using IDSA guidelines) implies a structured diagnostic process based on clinical criteria and laboratory results. This often involves review by infectious disease specialists or hospital epidemiologists, though the specific adjudication process is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted for this device. The study was a randomized controlled clinical trial comparing a medical device (dressing with CHG) to a control (dressing without CHG) in a real-world clinical setting, not assessing human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

No, a standalone (algorithm only) performance study was not done. The study evaluated the clinical effectiveness of a medical device (a dressing containing CHG) in a human-in-the-loop scenario (patients with catheters). There is no "algorithm" in the context of an antimicrobial dressing.

7. Type of Ground Truth Used

The ground truth used for determining CRBSI in the clinical study was based on clinical definitions of infection and laboratory confirmation. Two definitions were applied:

  • Timsit et al. (2012) Definition: A specific combination of positive peripheral blood cultures, positive quantitative catheter-tip culture, and absence of other infectious foci. For coagulase-negative staphylococci, matching pulse-field gel electrophoresis patterns were required.
  • Infectious Diseases Society of America (IDSA) 2009 Guidelines: A re-analysis incorporated IDSA's definition, which includes bacteremia/fungemia with an intravascular device, clinical manifestations, absence of other sources, and a combination of positive semiquantitative/quantitative catheter cultures or differential time to positivity.

8. Sample Size for the Training Set

The concept of a "training set" is not applicable here, as this is a medical device (dressing) and not an AI/ML algorithm. The study was a clinical trial evaluating the effectiveness of the dressing.

9. How the Ground Truth for the Training Set Was Established

As mentioned in point 8, there is no "training set" for this type of medical device. The clinical trial directly assessed the device's impact on CRBSI rates in patients.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 15, 2017

3M Company Melissa Forth Regulatory Affairs Specialist 3M Center 2510 Conway Ave Bldg 275-5W-06 St. Paul, Minnesota 55144

Re: K153410

Trade/Device Name: 3M Tegaderm CHG Chlorhexidine Gluconate I.V. Securement Dressing Regulatory Class: Unclassified Product Code: FRO Dated: February 23, 2017 Received: February 24, 2017

Dear Melissa Forth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153410

Device Name

3MTM Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing

Indications for Use (Describe)

3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing can be used to cover and protect catheter sites and to secure devices to skin. Common applications include securing I.V. catheters, other intravascular catheters and percutaneous devices.

Tegaderm™ CHG I.V. Securement Dressing is intended to reduce vascular catheter colonization and catheter-related bloodstream infections (CRBSI) in patients with central venous or arterial catheters.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

3M Center 2510 Conway Ave, Bldg. 275-5W-06 St. Paul, MN 55144 U.S.A. 651 733 1110

Image /page/3/Picture/2 description: The image shows the 3M logo. The logo is red and consists of the number 3 followed by the letter M. The logo is simple and recognizable.

510(k) Summary

510(k) Summary

This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.92.

I. SUBMITTER

Applicant Name:3M Health Care3M Center2510 Conway Ave, Bldg. 275-5W-06St. Paul, MN 55144-1000
Contact Person:Melissa J. ForthRegulatory Affairs Specialist
Phone:(651) 737-5016
Fax:(651) 737-5320
Email:mjforth@mmm.com
Date Prepared:May 10, 2017

II. DEVICE

Trade Names:3M™ Tegaderm™ CHG Chlorhexidine Gluconate
I.V. Securement Dressing
Common or Usual Name:Antimicrobial I.V. Securement Dressing
Classification:Unclassified
Product Code:FRO
Classification Name:Dressing, wound, drug

{4}------------------------------------------------

3M™ Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

III. PREDICATE DEVICE(S)

3MTM Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing (K063458. K080620)

BIOPATCH Antimicrobial Dressing (K003229)

No reference devices are used in this submission.

IV. DEVICE DESCRIPTION

3MTM Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing

3MTM Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing is used to cover and protect catheter sites and to secure devices to skin. It is available in a variety of shapes and sizes.

Tegaderm™ CHG I.V. Securement Dressing consists of a transparent adhesive dressing and an integrated gel pad containing 2% w/w Chlorhexidine Gluconate (CHG), a well-known antiseptic agent with broad spectrum antimicrobial and antifungal activity.

The transparent film provides an effective barrier against external contamination including fluids (waterproof), bacteria, and yeast, and protects the I.V. site.

In vitro testing (log reduction and barrier testing) demonstrates that the Tegaderm™ CHG gel pad in the Tegaderm™ CHG I.V. Securement Dressing has an antimicrobial effect against, and is a barrier to, a variety of gram-positive and gram-negative bacteria, and yeast in the dressing. The gel pad absorbs fluid.

Tegaderm™ CHG I.V. Securement dressing is transparent, allowing continual site observation, and is breathable, allowing good moisture vapor exchange.

The device is single-use, provided sterile and the sterilization method is ethylene oxide. There are no associated accessories.

CatalogNumberDevice NameDressing SizeAverage Amount ofCHG per Dressing(mg based on gel pad size)
16573M™ Tegaderm™ CHGChlorhexidine Gluconate I.V.Securement Dressing8.5 cm x 11.5 cm(3-1/2 X 4-1/2 in)45 mg
16583M™ Tegaderm™ CHGChlorhexidine Gluconate I.V.Securement Dressing10 cm X 12 cm(4 X 4-3/4 in)45 mg
16593M™ Tegaderm™ CHGChlorhexidine Gluconate I.V.Securement Dressing10 cm x 15.5 cm(4 X 6-1/8 in)78 mg

{5}------------------------------------------------

3M™ Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

Catalog NumberDevice NameDressing SizeAverage Amount of CHG per Dressing (mg based on gel pad size)
16603M™ Tegaderm™ CHGChlorhexidine Gluconate I.V.Securement Dressing7 cm x 8.5 cm(2 3/4 in. x 3 3/8 in.)15 mg
1877*3M™ PICC / CVC SecurementDevice + Tegaderm™ CHGChlorhexidine Gluconate I.V.Securement Dressing8.5 cm x 11.5 cm(3 1/2 x 4 1/2 in)45 mg
1879*3M™ PICC / CVC SecurementDevice + Tegaderm™ CHGChlorhexidine Gluconate I.V.Securement Dressing10 cm x 15.5 cm(4 x 6 1/8 in)78 mg
  • Note: Catalog numbers 1877 and 1879 are convenience kits. The 3M™ PICC / CVC Securement Device portion of the kit is a US Class I device.

V. INDICATIONS FOR USE

3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing

3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing can be used to cover and protect catheter sites and to secure devices to skin. Common applications include securing and covering I.V. catheters, other intravascular catheters and percutaneous devices.

Tegaderm™ CHG I.V. Securement Dressing is intended to reduce vascular catheter colonization and catheter-related bloodstream infections (CRBSI) in patients with central venous or arterial catheters.

{6}------------------------------------------------

3M™ Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Intended Use and Indications for Use

The Intended Use of the subject and predicate devices is that they are antimicrobial dressings with CHG, indications for Use of the subject and the predicate devices are set forth in detail in the table directly below.

3M™ Tegaderm™ CHG ChlorhexidineGluconate I.V. Securement Dressing(Submission Subject, K153410)BIOPATCH Antimicrobial Dressing(K003229)3M™ Tegaderm™ CHGChlorhexidine GluconateI.V. Securement Dressing(K063458, K080620)
3M™ Tegaderm™ CHG ChlorhexidineGluconate I.V. Securement Dressing can beused to cover and protect catheter sites and tosecure devices to skin. Common applicationsinclude I.V. catheters, other intravascularcatheters and percutaneous devices.Tegaderm™ CHG Chlorhexidine GluconateI.V. Securement Dressing is intended to reducevascular catheter colonization and catheter-related bloodstream infections (CRBSI) inpatients with central venous or arterialcatheters.BIOPATCH containing Chlorhexidine Gluconate isintended for use as a hydrophilic wound dressingthat is used to absorb exudate and to cover a woundcaused by the use of vascular and non-vascularpercutaneous medical devices such as: IV catheters,central venous lines, arterial catheters, dialysiscatheters peripherally inserted coronary catheters,mid-line catheters, drains, chest tubes, externallyplaced orthopedic pins, and epidural catheters. It isalso intended to reduce local infections, catheter-related blood stream infections (CRBSI), and skincolonization of microorganisms commonly relatedto CRBSI, in patients with central venous or arterialcatheters.3M™ Tegaderm™ CHG Dressings(Chlorhexidine Gluconate I.V.Securement Dressing) can be used tocover and protect catheter sites andto secure devices to skin. Commonapplications include IV catheters,other intravascular catheters andpercutaneous devices.

The indications for use statement is different between the predicate device. Specifically, the subject device indications for use includes an indication expansion "to reduce vasular catheter-related bloodstream infections (CRBS) in patients with central venous or arterial catheters." This statement is not included in the predicate: 3M™ Tegadern I V. Securement Dressing. However, it performance characteristic associated with the same intended use of being an antimicrobial dressing with CHG. To support this portion of the indications, 3M has provided performance data to show that the T.V. Securement Dressings perform as claimed.

{7}------------------------------------------------

3M™ Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

In addition, the indications for use statement is different between the Tegadern™ CHG I.V. Securement Dressing and the BIOPATCH Antimicrobial Dressing predicate. The Tegaderm™ CHG I.V. Securement Dressing "to reduce vascular catheter colonization and catherer-related bloodstream infections (CRBS) in patients with catheters" is similar to BIOPATCH's indication "to reduce local infections, catheter-related blood stream infections (CRBS), and skin colonization commonly related to CRBSI, in patients with central venous or arterial catheres in the Indications do not affect the safety and effectiveness of the device, and again, as stated above, do not alter the intended therapeutic, drasmostic, or surgical use of the subject device.

Technology Comparison

The Tegadern™ CHG I.V. Securement Dressing and the BIOPATCH Antimicrobial Dressing have similar technological characteristics. Both devices contain the same antimicrobial, choconate (CHG). The minor differences in technological characteristics do not raise different questions of safety or effectiveness, as confirmed by the testing activities described in this submission.

Characteristic3M™ Tegaderm™ CHG I.VSecurement Dressing(Submission Subject, K153410)BIOPATCH AntimicrobialDressing(K003229)3M™ Tegaderm™ CHGChlorhexidine GluconateI.V. Securement Dressing(K063458, K080620)
AntimicrobialagentCHG Chlorhexidine Gluconate (CHG)Gel pad contains 2% w/w chlorhexidinegluconateCHG Chlorhexidine Gluconate(CHG) contains chlorhexidinegluconate(no subsequent 510k filings)CHG Chlorhexidine Gluconate (CHG)Gel pad contains 2% w/wchlorhexidine gluconate(unchanged from original 510kclearance)
DressingconfigurationThe Chlorhexidine Gluconate I.V.Securement Dressing consists of atransparent, semi-permeable adhesivedressing and an integrated hydrophilicgel pad containing 2% w/wChlorhexidine Gluconate (CHG).A hydrophilic polyurethaneabsorptive foam withChlorhexidine Gluconate (CHG).A separate unspecified dressingcover is needed.The Chlorhexidine Gluconate I.V.Securement Dressing consists of atransparent adhesive dressing and anintegrated hydrophilic gel padcontaining 2% w/w ChlorhexidineGluconate (CHG).
AbsorbentYesYesYes
TransparentYesN/A (opaque)Yes
BreathableYesYesYes

{8}------------------------------------------------

3M™ Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

Characteristic3MTM Tegaderm™ CHG I.VSecurement Dressing(Submission Subject, K153410)BIOPATCH AntimicrobialDressing(K003229)3MTM Tegaderm™ CHGChlorhexidine GluconateI.V. Securement Dressing(K063458, K080620)
Broad spectrum,continuous,antimicrobial /Broad spectrumantifungalYesYesYes

{9}------------------------------------------------

3MTM Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

VII. PERFORMANCE DATA

The following performance data are provided in support of the substantial equivalence determination for this indication expansion submission.

Performance Tests

The following performance testing was provided in this submission to support a determination of substantial equivalence to the predicate device.

  • Suppression of regrowth over time ●
  • In vitro testing (log reduction and barrier testing) demonstrates that the Tegaderm™ CHG gel pad in the Tegaderm™ CHG I.V. Securement Dressing has an antimicrobial effect against, and is a barrier to, a variety of gram-positive and gram-negative bacteria, and yeast in the dressing.

Clinical Studies -- Randomized Controlled Clinical Trial

The proposed broadened indication for use for 3M Tegaderm CHG Chlorhexidine Gluconate I.V. Securement Dressing is supported by a randomized, multi-arm, controlled clinical trial consisting of 1,879 subjects with 4.163 central venous and arterial catheters conducted at 11 hospitals in France (12 ICUs). This trial found that the addition of 3M Tegaderm CHG Chlorhexidine Gluconate I.V. Securement Dressing in institutions already following routine infection control techniques led to a reduction in catheter colonization and catheter-related infections. The full study details are available in Timsit JF et al. Randomized controlled trial of chlorhexidine dressing and highly adhesive dressing for preventing catheter-related infections in critically ill adults. Am J Crit Care Med. 2012;186(12): 1272-1278.

The Timsit study defined a catheter-related bloodstream infection (CRBSI) as:

"CRBSI was a combination of (a) one or more positive peripheral blood cultures sampled immediately before or within 48 hours after catheter removal; (b) a positive quantitative catheter-tip culture positive (using the 1.000 CFU/ml threshold when vortexing technique was used or 100 CFU/ml threshold when sonication technique was used) for the same microorganisms (same species and same susceptibility pattern) or a blood-culture differential time-to-positivity of 2 hours or more; and (c) no other infectious focus explaining the positive blood cultures. In patients with blood cultures positive for coagulasenegative staphylococci, the same pulse-field gel electrophoresis patterns in the catheter tip and blood cultures was required for a diagnosis of CRBSI."

Utilizing the definition of CRBSI described in the Timsit et al. (2012) publication, the study authors found that a dressing containing CHG (i.e. Tegaderm CHG Securement Dressing), reduced the risk of CRBSI by 57% compared to dressings without CHG (control). The following results were obtained:

{10}------------------------------------------------

3MTM Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

Control Group(non CHG dressing)Test Group(Tegaderm™ CHG I.V.Securement Dressing)ComparisonP-value
PatientswithInfectionInfections per1000 CatheterDaysPatients withInfectionInfections per1000 CatheterDaysPercentReduction inPatients withInfectionsPercentReduction per1000 CatheterDays
2.2%(21/941)1.290.96%(9/938)0.5257%60%0.020

It should be noted that the Timsit definition differs from the CRBSI definition published by the Infectious Diseases Society of America (IDSA). See below for the IDSA definition. As a sensitivity analysis for the Timsit study results, the data was re-analyzed applying the Infectious Diseases Society of America (IDSA) 2009 guidelines, Clinical Practice Guidelines for the Diagnosis and Management of Intravascular Catheter-Related Infection: 2009 Update by the Infectious Diseases Society of America. In a re-analysis of all 31 CRBSI cases, each subject was reviewed to determine whether they fully satisfied criteria for CRBSI per the IDSA Guidelines.

The IDSA 2009 guidelines use the following definition for CRBSI:

"Catheter related Bacteremia or fungemia in a patient who has an intravascular device and >1 positive blood culture result obtained from the peripheral vein, clinical manifestations of infection (e.g., fever, chills, and/ or hypotension), and no apparent source for bloodstream infection (with the exception of the catheter). One of the following should be present: a positive result of semiquantitative (>15 cfu per catheter segment) or quantitative (>10° cfu per catheter segment) catheter culture, whereby the same organism (species) is isolated from a catheter segment and a peripheral blood culture; simultaneous quantitative cultures of blood with a ratio of >3:1 cfu/mL of blood (catheter vs. peripheral blood); differential time to positivity (growth in a culture of blood obtained through a catheter hub is detected by an automated blood culture system at least 2 h earlier than a culture of simultaneously drawn peripheral blood of equal volume)."

{11}------------------------------------------------

3MTM Tegaderm™ CHG Chlorhexidine Gluconate Dressings 510(k) Premarket Notification - Traditional

Control Group(non CHG dressing)Test Group(Tegaderm ™ CHG I.V.Securement Dressing)ComparisonP-value
Patients withInfectionInfections per1000 CatheterDaysPatients withInfectionInfections per1000 CatheterDaysPercentReduction inPatients withInfectionsPercentReduction per1000 CatheterDays
1.7%(16/941)0.940.85%(8/938)0.4650%51%0.095

The following results were obtained from the re-analysis with the IDSA guidelines:

VIII. CONCLUSIONS

3MTM Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing has been shown to reduce the incidence of catheter-related bloodstream infection (CRBSI) and catheter colonization in a large, randomized, controlled clinical trial. The Timsit 2012 study supports the expanded indication statement requested by 3M.

The 3M™ Tegaderm™ CHG Chlorhexidine Gluconate I.V. Securement Dressing remains substantially equivalent to 3M Company's currently cleared and marketed 3M™ Tegaderm™ CHG I.V. Securement Dressing, and is substantially equivalent to the BIOPATCH predicate. The differences proposed do not affect the safety and effectiveness of the device when used as labeled.

N/A