AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only.

The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only.

Diazyme Human Kappa/Lambda FLC Calibrator (DZ169A-CAL) is intended for use in establishing the calibration curve for the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.

Diazyme Human Kappa/Lambda FLC Control (DZ169A-CON) is intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.

Device Description

Diazyme Human Kappa (K) /Lambda (2) Free Light Chain Assay System consists of the following items:

  1. Diazyme Human Kappa (k) Free Light Chain Assay: containing reagent R1 and Reagent R2. Reagent 1 contains 100 mmol/L tris buffer, and 0.09% sodium azide Reagent 2 containing anti-human Kappa Free Light Chain Ig fraction coated on polystyrene beads and 0.09% sodium azide.
  2. Diazyme Human Lambda (λ) Free Light Chain Assay: containing reagent R1 and Reagent R2. Reagent 1 contains 100 mmol/L tris buffer, and 0.09% sodium azide Reagent 2 containing anti-human Lambda Free Light Chain Ig fraction coated on polystyrene beads and 0.09% sodium azide.
  3. Diazyme Kappa and Lambda FLC Calibrator Set: 5 levels, serum based, liquid stable format, intended for use in establishing the calibration curve for the Diazyme Human Kappa (k) and Lambda (λ) Free Light Chain Assay.
  4. Diazyme Kappa and Lambda FLC Control Set: 2 levels, serum based, liquid stable format, intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa (K) and Lambda (λ) Free Light Chain Assay.

Diazyme Human Kappa (k) /Lambda (2.) Free Light Chain Assays are run on clinical chemistry analyzers such as Roche Hitachi 917 analyzer (K023009). They never come into direct contact with patients. The patient's venous blood sample is first collected by a phlebotomist and then submitted for determination of Kappa and Lambda FLC concentration using Diazyme Human Kappa (k) and Lambda (λ) FLC Assay by trained professionals.

The Control Unit of the Roche Hitachi 917 analyzer uses a graphical interface to control all instrument functions. The computer, keyboard, and touchscreen monitor allows users to navigate through the software, enter assay, calibrator, and control information, and make test selections. The Diazyme Kappa and Lambda FLC Hitachi 917 application parameters provided are programmed into the Hitachi 917 analyzer. The reagents, calibrators and controls are loaded into the analyzer. The Roche Hitachi 917 stores Diazyme Human Kappa (k) /Lambda (λ) Free Light Chain Assay reagents in a refrigerated compartment. Reagent and sample pipettes automatically aspirate and dispense specified amounts of reagent or calibrators, controls, and sample into reaction cells. The change in absorbance is measured at specified wavelengths.

After a 6-point calibration, spline-fitting is used to smoothly fit polynomial functions through mean values of the response for calibrators of known concentrations. The Roche Hitachi 917 calculates the kappa and lambda FLC concentration of a patient sample by interpolation of the obtained signal to a stored 6-point calibration curve.

AI/ML Overview

The provided document describes the performance testing and acceptance criteria for the Diazyme Human Kappa (κ) and Lambda (λ) Free Light Chain Assays. The study aims to demonstrate substantial equivalence to predicate devices (Freelite® Human Kappa Free Kit and Freelite® Human Lambda Free Kit).

Here's an analysis of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Predicate Device)Reported Device Performance (Diazyme Assay)
Linear RangeKappa: 6.0-180 mg/L (standard); 150-4,500 mg/L & 3,000–90,000 mg/L (extended) Lambda: 8.0-270 mg/L (standard); 200-6750 mg/L & 4,000-135,000 mg/L (extended)Kappa FLC: 4.5-150 mg/L (standard); 90-3000 mg/L (extended) Lambda FLC: 6.1-200 mg/L (standard); 122-4000 mg/L (extended)
Precision (Within-run CV)Kappa FLC: 3.1% to 4.84% (13.56 mg/L to 51.06 mg/L) Lambda FLC: 5.2% to 8.4% (15.12 mg/L to 71.84 mg/L)Kappa FLC: 1.4% to 7.6% (6.0 mg/L to 2588.6 mg/L) (One lot, three analyzers) and 1.1% to 7.9% (5.81 to 2588.45 mg/L) (Three lots, one analyzer) Lambda FLC: 0.8% to 6.5% (8.3 mg/L to 3472.6 mg/L) (One lot, three analyzers) and 0.6% to 6.3% (8.16 to 3442.60 mg/L) (Three lots, one analyzer)
Precision (Between-run CV)Kappa FLC: 6.27% to 8.38% (12.02 mg/L to 54.66 mg/L) Lambda FLC: 4.72%-8.08% (18.43 mg/L to 71.71 mg/L)Kappa FLC: 0.6% to 4.9% (One lot, three analyzers) and 0.3% to 5.7% (Three lots, one analyzer) Lambda FLC: 0.4% to 4.5% (One lot, three analyzers) and 0.3% to 4.0% (Three lots, one analyzer)
Method Comparison (K-FLC)Linear regression yielded y = 0.9818x + 4.5206 and r = 0.9956Slope = 0.958; Correlation Coefficient (R²) = 0.977 95% CI Slope: 0.932 to 0.985; Intercept: -14.703 to 9.632 Agreement: PPA 96.67%, NPA 100%, TPA 97.62% (95% CI provided for each)
Method Comparison (L-FLC)Linear regression yielded y = 0.9266x + 9.5556 and r = 0.9752Slope = 1.054; Correlation Coefficient (R²) = 0.972 95% CI Slope: 1.021 to 1.086; Intercept: -20.746 to 22.018 Agreement: PPA 95.65%, NPA 94.74%, TPA 95.24% (95% CI provided for each)
Detection Limits (LOQ)N/A (Predicate states ranges, not specific LOQ)Kappa FLC: LOQ = 4.5 mg/L Lambda FLC: LOQ = 6.1 mg/L
InterferenceN/A (Predicate likely performs similar interference studies, but specific acceptance criteria for interferent levels are not stated)Less than 10% deviation for: Triglyceride: 1000 mg/dL Ascorbic Acid: 10 mM Bilirubin: 40 mg/dL Bilirubin Conjugated: 40 mg/dL Hemoglobin: 1000 mg/dL Rheumatoid Factor: 100 IU/mL
Hook EffectN/ANo high dose hook effect observed up to 100,000 mg/L for both Kappa and Lambda FLC.

2. Sample Sizes Used for the Test Set and the Data Provenance

  • Method Comparison (Test Set):

    • Sample Size: 126 serum samples for both Kappa and Lambda FLC assays.
    • Data Provenance: The document states "clinical patient samples" but does not specify the country of origin. The samples included a mix of:
      • 39 Multiple Myeloma (MM)
      • 12 Monoclonal Gammopathy of Undetermined Significance (MGUS)
      • 34 Abnormal (due to other disease states)
      • 41 Normal (no disease)
    • Retrospective/Prospective: Not explicitly stated, but the sample categorization (MM, MGUS, Abnormal, Normal) suggests these were existing samples collected from patients with known diagnoses, implying a retrospective collection for the purpose of this method comparison study.
  • Precision Studies (Test Set):

    • Sample Size:
      • Kappa FLC: Eight levels of serum specimens + two levels of serum-based controls. Each sample/control was tested with 2 runs per day, duplicates, over 20 working days. The "n=240" for mean column likely refers to the total number of measurements for each ID (8 levels + 2 controls x 2 runs/day x 2 duplicates/run x 20 days = 800 measurements * n=240 is likely the total of measured values per ID across the entire study design, for example, 20 working days * 2 runs/day * 2 duplicates * total samples (8) = 640. No, it is 240, there are 8 levels of serum specimens, 2 levels of serum-based controls, 2 runs/day with duplicates over 20 working days. Therefore, n=240 possibly corresponds to (8+2) samples * 20 days * (2 runs/day * replicates). The precision studies involve 240 data points per ID item when using one lot of reagent across three analyzers (240 measurements for each S1-S8, Con1, Con2) and 240 for three lots of reagent on one analyzer.
      • Lambda FLC: Eight levels of serum specimens + two levels of serum-based controls. Same testing procedure as Kappa FLC, resulting in "n=240" for mean calculations.
    • Data Provenance: Not specified, but likely laboratory-prepared serum pools and controls.
    • Retrospective/Prospective: Prospective (laboratory conducted study).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This is an in-vitro diagnostic device measuring quantitative concentrations of analytes (Kappa and Lambda Free Light Chains). The "ground truth" for the test set in this context is the quantitative value determined by the predicate device, or in the case of analytical performance, the known concentration of the analyte in prepared control materials or spiked samples. This does not involve human expert interpretation of images or clinical cases that would require adjudication.

4. Adjudication Method for the Test Set

Not applicable. See point 3. The "ground truth" for method comparison is the measurement from the predicate device. For analytical performance studies (precision, linearity, detection limits, interference), the ground truth is either the known concentration in spiked samples/controls or derived through standard laboratory methodologies (e.g., gravimetric dilutions for linearity).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in-vitro diagnostic assay for quantitative measurement of analytes in serum, not an AI-based imaging or diagnostic tool requiring human reader interpretation or MRMC studies.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, the performance data presented (linearity, precision, method comparison, detection limits, interference, hook effect) are all standalone performance characteristics of the Diazyme assay system itself, without human interpretation of results as a primary outcome. The system quantitatively determines FLC concentration.

7. The Type of Ground Truth Used

  • Method Comparison: The "ground truth" for the test set was the quantitative measurement obtained from the predicate device (Freelite® Human Kappa Free Kit and Freelite® Human Lambda Free Kit).
  • Analytical Performance (Precision, Linearity, Detection Limit, Interference, Hook Effect): The "ground truth" was established using known concentrations of Kappa and Lambda FLC in serum specimens and controls, often prepared by spiking or dilution, and evaluated against established analytical methods (e.g., CLSI guidelines).

8. The Sample Size for the Training Set

Not directly applicable in the conventional sense of an AI/machine learning training set. This is a traditional immunoturbidimetric assay. However, developmental and optimization work for such assays would involve testing numerous samples and conditions to establish reagent formulations, reaction kinetics, and calibration curves. The document does not specify a distinct "training set" for the assay itself, but rather samples used in validation studies. The calibration curve is essentially "trained" using a 6-point calibration, for which the document states "calibrators of known concentrations" are used.

9. How the Ground Truth for the Training Set Was Established

Again, not applicable for a "training set" in the AI sense. For the calibration curve of the assay, the "ground truth" (known concentrations) is established using:

  • Diazyme Human Kappa/Lambda FLC Calibrator Set: 5 levels, serum based, liquid stable format, containing purified Kappa Free Light Chains. The concentrations of these calibrators are presumably assigned through a validated process against reference methods or materials, though the specific method for assigning these values is not detailed in this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2016

DIAZYME LABORATORIES ABHIJIT DATTA DIRECTOR OF TECHNICAL OPERATIONS 12889 GREGG COURT POWAY CA 92064

Re: K153394

Trade/Device Name: Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay, Diazyme Human Kappa/lamda Free Light Chain Calibrator, Diazyme Human Kappa/lamda Free Light Chain Calibrator Regulation Number: 21 CFR 866.5550

Regulation Name: Immunoglobulin (light chain specific) immunological test system Regulatory Class: II Product Code: DFH, DEH Dated: October 28, 2016 Received: November 1, 2016

Dear Dr. Datta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kelly Oliner -S

FOR

Leonthena Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
510(k) Number (if known)153394
Device NameThe Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Kappa/Lambda FLC Calibrator Set, Diazyme Human Kappa/Lambda FLC Control Set.
Indications for Use (Describe)The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only. Diazyme Human Kappa/Lambda FLC Calibrator (DZ169A-CAL) is intended for use in establishing the calibration curve for the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only. Diazyme Human Kappa/Lambda FLC Control (DZ169A-CON) is intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)Page 1 of 1
-----------------------------------

PSC Publishing Services (301) 443-6740

EF

{3}------------------------------------------------

Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
510(k) Number (if known)153394
Device NameThe Diazyme Human Lambda Free Light Chain Assay, Diazyme Human Kappa/Lambda FLC Calibrator Set, Diazyme Human Kappa/Lambda FLC Control Set.
Indications for Use (Describe)The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only.Diazyme Human Kappa/Lambda FLC Calibrator (DZ169A-CAL) is intended for use in establishing the calibration curve for the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.Diazyme Human Kappa/Lambda FLC Control (DZ169A-CON) is intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's name:Diazyme Laboratories
Submitter's address:12889 Gregg CourtPoway, CA 92064USA
Name of Contact Person:Dr. Abhijit DattaDiazyme Laboratories12889 Gregg CourtPoway, CA 92064Phone: 858-455-4762Fax: 858-455-2120
Date the Summary was Prepared:October 28, 2016
Name of the DeviceDiazyme Human Kappa (κ) Free Light Chain AssayDiazyme Human Lambda (λ) Free Light Chain AssayDiazyme Human Kappa and Lambda FLC CalibratorSetDiazyme Human Kappa and Lambda FLC Control Set
Trade Name:Diazyme Human Kappa (κ) Free Light Chain AssayDiazyme Human Lambda (λ) Free Light Chain AssayDiazyme Human Kappa and Lambda FLC CalibratorSetDiazyme Human Kappa and Lambda FLC Control Set
Common/Usual NameDiazyme Human Kappa (κ) Free Light Chain AssayDiazyme Human Lambda (λ) Free Light Chain Assay
Device Classification NameKappa, antigen, antiserum, controlLambda, antigen, antiserum, control
Product code:DFH, Kappa, antigen, antiserum, controlDEH, Lambda, antigen, antiserum, control
Panel:Immunology (82)

{5}------------------------------------------------

Submission TypeTraditional 510k
Regulation Number21CFR 866.5550 Immunoglobulin (light chain specific)immunological test system
Device ClassClass II
Predicate Device:Freelite® Human Kappa Free Kit for use on the Hitachi911/912/917/Modular PFreelite® Human Lambda Free Kit for use on theHitachi 911/912/917/Modular P
Manufacturing AddressDiazyme Laboratories12889 Gregg CourtPoway, CA 92064USA
Establishment Registration2032900

DESCRIPTION OF THE DEVICE

Clinical Significance

Normal immunoglobulins are composed of smaller units called heavy chains and light chains. There are five types of heavy chains (alpha, delta, epsilon, gamma, and mu), and two types of light chains (kappa and lambda). The heavy and light chains are produced separately within plasma cells and are assembled to form a whole ("intact") immunoglobulin. When the light chains are attached to the heavy chains they are called bound, and when they are not attached they are called free. Kappa Free Light Chain (FLC) is a 22kD protein while Lambda FLC is usually a dimer of 44kD as presented in the serum. Elevated levels of Kappa or Lambda FLC are associated with plasma cell disorders such as multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinemia. AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus.

Assay Principle

Human Kappa (K) Free Light Chain Assay is based on a latex enhanced immunoturbidimetric assay. Kappa FLC in the sample binds to specific anti-Kappa FLC antibody, which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically and is proportional to the amount of Kappa FLC in the sample. The instrument calculates the Kappa FLC concentration by interpolation of obtained signal of a 6point calibration curve prepared from calibrators of known concentrations.

The Diazyme Human Lambda (λ) Free Light Chain Assay is based on a latex enhanced immunoturbidimetric assay. Lambda FLC in the sample binds to specific anti-Lambda FLC antibody, which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically and is proportional to the amount of Lambda FLC in the sample. The instrument calculates the Kappa FLC concentration by interpolation of obtained signal of a 6-point calibration curve prepared from calibrators of known concentrations.

{6}------------------------------------------------

Detailed Device Description

Diazyme Human Kappa (K) /Lambda (2) Free Light Chain Assay System consists of the following items:

    1. Diazyme Human Kappa (k) Free Light Chain Assay: containing reagent R1 and Reagent R2. Reagent 1 contains 100 mmol/L tris buffer, and 0.09% sodium azide Reagent 2 containing anti-human Kappa Free Light Chain Ig fraction coated on polystyrene beads and 0.09% sodium azide.
    1. Diazyme Human Lambda (λ) Free Light Chain Assay: containing reagent R1 and Reagent R2. Reagent 1 contains 100 mmol/L tris buffer, and 0.09% sodium azide Reagent 2 containing anti-human Lambda Free Light Chain Ig fraction coated on polystyrene beads and 0.09% sodium azide.
    1. Diazyme Kappa and Lambda FLC Calibrator Set: 5 levels, serum based, liquid stable format, intended for use in establishing the calibration curve for the Diazyme Human Kappa (k) and Lambda (λ) Free Light Chain Assay.
    1. Diazyme Kappa and Lambda FLC Control Set: 2 levels, serum based, liquid stable format, intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa (K) and Lambda (λ) Free Light Chain Assay.

Diazyme Human Kappa (k) /Lambda (2.) Free Light Chain Assays are run on clinical chemistry analyzers such as Roche Hitachi 917 analyzer (K023009). They never come into direct contact with patients. The patient's venous blood sample is first collected by a phlebotomist and then submitted for determination of Kappa and Lambda FLC concentration using Diazyme Human Kappa (k) and Lambda (λ) FLC Assay by trained professionals.

The Control Unit of the Roche Hitachi 917 analyzer uses a graphical interface to control all instrument functions. The computer, keyboard, and touchscreen monitor allows users to navigate through the software, enter assay, calibrator, and control information, and make test selections. The Diazyme Kappa and Lambda FLC Hitachi 917 application parameters provided are programmed into the Hitachi 917 analyzer. The reagents, calibrators and controls are loaded into the analyzer. The Roche Hitachi 917 stores Diazyme Human Kappa (k) /Lambda (λ) Free Light Chain Assay reagents in a refrigerated compartment. Reagent and sample pipettes automatically aspirate and dispense specified amounts of reagent or calibrators, controls, and sample into reaction cells. The change in absorbance is measured at specified wavelengths.

See Reaction Scheme.

Image /page/6/Figure/9 description: The image shows two diagrams, one for Kappa FLC and one for Lambda FLC. The Kappa FLC diagram shows the following steps: R1 at 150 uL and a sample at 4.0 uL are added at time 0, then the mixture is heated to 37 degrees Celsius, R2 at 50 uL is added at 4.8 minutes, and the absorbance is measured at 570 nm after 10 minutes. The Lambda FLC diagram shows similar steps, but with different volumes and times.

Image /page/6/Figure/10 description: This image shows a timeline of a chemical reaction. At time zero, 120 microliters of R1 and 20 microliters of sample are added. The reaction is then incubated at 37 degrees Celsius, and at 4.8 minutes, 40 microliters of R2 are added. At 6 minutes, A1 is measured at 570nm, and at 10 minutes, A2 is measured at 570nm.

{7}------------------------------------------------

After a 6-point calibration, spline-fitting is used to smoothly fit polynomial functions through mean values of the response for calibrators of known concentrations. The Roche Hitachi 917 calculates the kappa and lambda FLC concentration of a patient sample by interpolation of the obtained signal to a stored 6-point calibration curve.

Indications for Use:

The Diazyme Human Kappa (K) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only.

The Diazyme Human Lambda (2) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only.

Diazyme Human Kappa/Lambda FLC Calibrator (DZ169A-CAL) is intended for use in establishing the calibration curve for the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.

Diazyme Human Kappa/Lambda FLC Control (DZ169A-CON) is intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.

Predicate Kappa FLC AssayDiazyme Human Kappa (κ) FLC Assay
Reagent 1: Supplementary Reagentcontaining 0.099% Azide aspreservative.Reagent 1100 mM Tris, 0.09% azide as preservative, ready touse
Reagent 2: Consisting of polyclonalmonospecific antibody coated ontopolystyrene latex.Reagent 2Latex particles coated with polyclonal rabbit and goatanti-Kappa FLC antibodies
Calibrator:Human sera that contain polyclonalkappa free light chain withpreservatives.CalibratorsLiquid stable calibrators prepared with serum,purified Kappa Free Light Chains with preservatives.
Calibrator SetCalibrator Set
2 x 1.5 mL Single-Level Calibratordiluted on board to 5 level calibrator1 x 1.0 mL Calibrator 11 x 1.0 mL Calibrator 21 x 1.0 mL Calibrator 3

Table 1 Summary of Assay Kit Components

{8}------------------------------------------------

1 x 1.0 mL Calibrator 41 x 1.0 mL Calibrator 5
Control Set (serum based)Control Set (serum based)
1 x 1.0mL Control 11 x 3.0mL Control 1
1 x 1.0mL Control 21 x 3.0mL Control 2
Predicate Lambda FLC AssayDiazyme Human Lamda (λ) FLC Assay
Reagent 1: Supplementary Reagentcontaining 0.099% Azide aspreservative.Reagent 1100 mM Tris, 0.09% azide as preservative, ready to use
Reagent 2: Consisting of polyclonalmonospecific antibody coated ontopolystyrene latex.Reagent 2Latex particles coated with polyclonal rabbit anti-Lambda FLC antibodies
Calibrator:Human sera that contain polyclonalkappa free light chain withpreservatives.CalibratorsLiquid stable calibrators prepared with serum,purified Kappa Free Light Chains with preservatives.
Calibrator SetCalibrator Set
2 x 1.5 mL Single-Level Calibratordiluted on board to 5 level calibrator1 x 1.0 mL Calibrator 11 x 1.0 mL Calibrator 21 x 1.0 mL Calibrator 31 x 1.0 mL Calibrator 41 x 1.0 mL Calibrator 5
Control Set (serum based)Control Set (serum based)
1 x 1.0mL Control 11 x 1.0mL Control 21 x 3.0mL Control 11 x 3.0mL Control 2

Substantial Equivalence Comparison

Indications for Use

Freelite® Human Kappa Free Kit for use(K023009) Freelite® Human Lambda FreeKit for use on the Hitachi 911/912/917/Modular P (K023131), (K040009)Diazyme Human Kappa (κ) andLambda (λ) Free Light Chain Assay.Equivalency
Kappa: This kit is intended for thequantitation of kappa free light chainsin serum and urine on the RocheHitachi 911 and Hitachi 912.Measurement of the various amountsThe Diazyme Human Kappa (κ) FreeLight Chain Assay is intended as a latexparticle enhanced immunoturbidimetricassay for the quantitative determinationof Kappa Free Light Chain (FLC)Similar

{9}------------------------------------------------

of the different types of light chainsaids in the diagnosis and monitoring ofmultiple myelomas, lymphocyticneoplasms, Waldenstrom'smacroglobulinemia and connectivetissue diseases, such as systemic lupusconcentration in serum on Hitachi 917analyzers. The measurement of KappaFLC in conjunction with Lambda FLCaids in the diagnosis of multiplemyeloma in conjunction with otherlaboratory findings. For in-vitro
erythematosus.diagnostic use only.
Lambda: This kit is intended for thequantitation of lambda free light chainsin serum and urine on the RocheHitachi 911 and Hitachi 912.Measurement of the various amountsof the different types of light chainsaids in the diagnosis and monitoring ofmultiple myelomas, lymphocyticneoplasms, Waldenstrom'sThe Diazyme Human Lambda (λ) FreeLight Chain Assay is intended as a latexparticle enhanced immunoturbidimetricassay for the quantitative determinationof Lambda Free Light Chain (FLC)concentration in serum on Hitachi 917analyzers. The measurement of LambdaFLC in conjunction with Kappa FLCaids in the diagnosis of multiple
macroglobulinemia and connectivemyeloma in conjunction with other
tissue diseases, such as systemic lupuslaboratory findings. For in-vitro
erythematosus.diagnostic use only

Principle

PrincipleEquivalency
Freelite® Human Kappa Free Kit for use(K023009) Freelite® Human Lambda FreeKit for use on the Hitachi 911/912/917/Modular P (K023131), (K040009)Diazyme Human Kappa (κ) and Lambda (λ) Free Light Chain Assay.
Evaluating the concentration of asoluble antigen by turbidimetryinvolves the addition of the test sampleto a solution containing the appropriateantibody in a reaction vessel or cuvette.A beam of light is passed through thecuvette and as the antigen-antibodyreaction proceeds, the light passingthrough the cuvette is scatteredincreasingly as insoluble immunecomplexes are formed. Light scatter ismonitored by measuring decrease inintensity of the incident beam of light.The antibody in the cuvette is in excessso the amount of immune complexformed is proportional to the antigenconcentration. A series of calibrators ofknown antigen concentration areassayed initially to produce acalibration curve of measured lightHuman Kappa (κ) and Lambda (λ) FreeLight Chain Assay is based on a latexenhanced immunoturbidimetric assay.Kappa FLC in the sample binds tospecific anti-Kappa FLC antibody, whichis coated on latex particles, and causesagglutination. The degree of the turbiditycaused by agglutination can be measuredoptically and is proportional to theamount of Kappa FLC in the sample.Same

{10}------------------------------------------------

scatter versus antigen concentration.Samples of unknown antigenconcentration can then be assayed andthe results read from the calibrationcurve.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------

Test Objective

Freelite® Human Kappa Free Kit for use(K023009) Freelite® Human LambdaFree Kit for use on the Hitachi911/912/917/ Modular P(K023131),(K040009)Diazyme Human Kappa (κ) andLambda (λ) Free Light Chain Assay.Equivalency
The quantitative determination ofKappa(κ) /Lambda(λ) Free Light ChainconcentrationFor the quantitative determination ofKappa and Lambda Free Light Chain(FLC) concentrationSame

Type of Test

Freelite® Human Kappa Free Kit for useon the Hitachi 911/912/917/Modular P(K023009) Freelite® Human Lambda FreeKit for use on the Hitachi 911/912/917/Modular P (K023131), (K040009)Diazyme Human Kappa (κ) andLambda (λ) Free Light Chain Assay.Equivalency
QuantitativeQuantitativeSame

Specimen Type

Freelite® Human Kappa Free Kit for useon the Hitachi 911/912/917/Modular P(K023009)Freelite® Human Lambda Free Kit for useon the Hitachi 911/912/917/Modular P(K023131), (K040009)Diazyme Human Kappa (κ) andLambda (λ) Free Light Chain Assay.Equivalency.
Human serum.Human serum.Same
Human UrineHuman serum onlyN/A

Product Type

Freelite® Human Kappa Free Kit for useon the Hitachi 911/912/917/Modular P(K023009)Diazyme Human Kappa ( $κ$ ) andLambda ( $λ$ ) Free Light Chain Assay.Equivalency
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{11}------------------------------------------------

Freelite® Human Lambda Free Kit for useon the Hitachi 911/912/917/Modular P(K023131)(K040009)
Assay reagent kit, calibrator kit, qualitycontrol kit.Assay reagent kit, calibrator kit, qualitycontrol kit.Same

Performance

Freelite® Human Kappa Free Kit for useon the Hitachi 911/912/917/Modular P (K023009)Freelite® Human Lambda Free Kit for useon the Hitachi 911/912/917/Modular P(K023131),(K040009)Diazyme Human Kappa (κ) and Lambda (λ)Free Light Chain Assay.
Linear Range:Kappa: 6.0-180 mg/L150-4,500 mg/L and 3,000–90,000 mg/L (extended)Lambda: 8.0-270 mg/L200-6750 mg/L and 4,000-135,000 mg/L (extended)Linear Range:Kappa FLC 4.5-150 mg/LLambda FLC 6.1-200 mg/L
Precision:Kappa FLC Within-runCV 3.1% to 4.84%; 13.56 mg/L to 51.06 mg/LKappa FLC Between-runCV 6.27% to 8.38%; 12.02 mg/L to 54.66 mg/LLambda FLC Within-runCV 5.2% to 8.4%; 15.12 mg/L to 71.84 mg/LLambda FLC Between-runCV 4.72%-8.08%; 18.43 mg/L to 71.71 mg/LExtended Linear Range:Kappa FLC 90-3000 mg/L;Lambda FLC 122-4000 mg/L;Precision:Kappa FLC Within-runCV 1.4% to 7.6%; 6.0 mg/L to 2588.6 mg/LKappa FLC Between-runCV 1.2% to 7.3%; 6.0 mg/L to 2588.6 mg/LLambda FLC Within-runCV 0.8% to 6.5%; 8.3 mg/L to 3472.6 mg/LLambda FLC Between-runCV 0.9%-6.4%; 8.3 mg/L to 3472.6 mg/L
Method comparison:Kappa FLCLinear regression analysis yielded y = 0.9818x +4.5206 and the correlation coefficient (r) was0.9956 ;Lambda FLCLinear regression yielded y = 0.9266x + 9.5556and the correlation coefficient (r) was 0.9752Method comparison:Kappa FLCSlope = 0.958Correlation Coefficient (R²) = 0.977Lambda FLCSlope = 1.054Correlation Coefficient (R²) = 0.972

{12}------------------------------------------------

Calibrator Comparison

Freelite® Human Kappa Free Kit for useon the Hitachi 911/912/917/Modular P (K023009)Freelite® Human Lambda Free Kit for useon the Hitachi 911/912/917/Modular P(K023131), (K040009)Diazyme Human Kappa (κ) andLambda (λ) Free Light ChainAssay.Equivalency
Human sera based solution containing Kappa andLambda free light chainHuman sera based solutioncontaining Kappa and Lambdafree light chainSame
Liquid Stable One level calibrator for on-boardautomatic dilutionLiquid stable five levelcalibrator set is provided readyto useSimilar

Control Comparison

Freelite® Human Kappa Free Kit for useon the Hitachi 911/912/917/Modular P (K023009)Freelite® Human Lambda Free Kit for useon the Hitachi 911/912/917/Modular P(K023131), (K040009)Diazyme Human Kappa (κ)and Lambda (λ) Free LightChain Assay.Equivalency
Human sera based solution containing Kappa andLambda free light chainHuman sera based solutioncontaining Kappa and Lambdafree light chainSame
Separately packaged quality control kit designedfor specific assay.Separately packaged qualitycontrol kit designed forspecific assay.Same

PERFORMANCE TESTING SUMMARIES

Precision

Precision for Diazyme Human Kappa (к) FLC Assay

The precision of the Assay was evaluated according to CLSI EP5-A guideline. In the study, eight levels of serum specimens containing kappa FLC spanning AMR and two levels of serum based kappa FLC controls were tested with 2 runs per day with duplicates over 20 working days using multiple lots of the reagents on multiple analyzers. The precision data was analyzed according to three-way nested ANOVA and the results of mean (mg/L) and CV% are summarized below: One lot of the reagent on three analyzers

IDMeann=240Within-RunBetweenRunBetweendayBetweenInstrTotal
S19.387.6%4.9%3.5%5.1%11.0%
S235.302.1%1.5%1.3%2.0%3.5%
S3122.372.4%0.9%1.8%0.8%3.2%
S45.957.5%3.0%8.5%N/A11.7%
S515.762.9%1.6%2.1%0.2%3.9%

{13}------------------------------------------------

S625.921.7%1.3%0.5%0.4%2.2%
S7139.231.5%1.5%0.7%1.1%2.5%
S82588.591.4%0.6%2.6%1.5%3.3%
Con116.703.2%2.6%2.4%0.9%4.9%
Con229.102.1%1.2%2.1%5.9%6.7%

Three lots of the reagents on one analyzer

IDMeann=240Within-RunBetweenRunBetweendayBetweenlotTotal
S19.497.9%5.7%5.7%2.6%11.6%
S235.142.9%0.8%1.5%0.8%3.5%
S3121.532.6%1.5%2.3%N/A3.8%
S45.816.3%2.6%2.7%0.7%7.4%
S515.402.4%0.9%0.6%2.2%3.5%
S625.661.1%0.3%0.6%1.0%1.7%
S7138.281.1%N/A1.0%1.0%1.8%
S82588.451.3%1.0%2.5%1.5%3.3%
Con116.644.2%1.4%3.0%1.6%5.6%
Con227.823.2%N/A2.7%1.8%4.6%

Precision for Diazyme Human Lambda (λ) FLC Assay

The precision of the Assay was evaluated according to CLSI EP5-A guideline. In the study, eight levels of serum specimens containing low, medium, and high lambda FLC respectively and two levels of serum based lambda FLC controls were tested with 2 runs per day with duplicates over 20 working days using multiple lots of the reagents on multiple analyzers. The precision data was analyzed according to three-way nested ANOVA and the results of mean (mg/L) and CV% are summarized below:

IDMeann=240Within-RunBetween RunBetween dayBetween InstrTotal
S111.896.3%4.5%6.1%N/A9.9%
S248.652.1%0.7%1.3%0.8%2.7%
S3144.551.6%1.0%1.3%0.6%2.4%
S48.316.5%0.4%3.1%0.8%7.3%
S522.273.0%1.4%2.6%1.1%4.3%
S635.092.1%1.2%1.4%1.0%2.9%
S7181.130.8%0.7%1.3%1.1%2.0%
S83472.593.6%N/A2.1%0.6%4.2%
Con127.623.9%N/A0.9%0.1%4.0%
Con262.581.3%0.7%N/A2.9%3.2%

One lot of the reagent on three analyzers

{14}------------------------------------------------

IDMeann=240Within-RunBetweenRunBetweendayBetweenlotTotal
S111.785.9%4.0%7.4%4.9%11.4%
S248.951.3%N/A1.7%0.3%2.2%
S3142.631.1%0.8%0.5%0.8%1.6%
S48.166.3%2.3%N/A3.1%7.4%
S521.902.9%1.3%N/A0.6%3.2%
S634.792.0%0.4%0.3%N/A2.1%
S7183.020.6%0.3%0.8%0.3%1.1%
S83442.603.2%N/A2.0%N/A3.8%
Con128.134.5%N/A0.2%3.8%5.9%
Con265.311.9%0.6%N/A1.2%2.3%

Three lots of the reagents on one analyzer

Linearity for Diazyme Human Lambda (λ) FLC Assay

The linearity of the Assay was evaluated according to CLSI EP6-A guideline. Based on the linearity data, limit of quantitation (LOQ=6.10 mg/L), the Analytical Measuring Range (AMR) is claimed to be 6.1-200 mg/L in standard mode. AMR is 6.1-4000 mg/L including extended mode.

Method Comparison

Method comparison for Diazyme Human Kappa (k) FLC Assay

The method comparison of the assay was evaluated following CLSI EP9-A2 protocol. A total of 126 serum samples ranging from 4.63-2975.80 mg/L were tested in comparison with predicate assay. Among the 126 samples, 39 were Multiple Myeloma (MM), 12 were Monoclonal Gammopathy of Undetermined Significance (MGUS), 34 were Abnormal (due to other disease states) and 41 were Normal (no disease), the concordance between the subject and predicate Kappa FLC is 98%. The results of regression analysis are summarized in the following table:

ParameterLinear RegressionDeming Regression
n126126
Slope0.9580.969
95% CI0.932 to 0.9850.943 to 0.996
Intercept-2.536-4.813
95% CI-14.703 to 9.632-17.014 to 7.388
Corr Coeff (R2)0.9770.977
Sample range4.63-2975.804.63-2975.80

Using CLSI EP12-A2: Evaluation of Qualitative Test Performance; Approved Guideline, Second Edition as guideline, the positive percent agreement (PPA), negative percent agreement (NPA) and overall percent agreement (TPA) for Kappa FLC comparisons to predicate and associated 95% confidence intervals were calculated. The results are listed in the table below:

PPANPATPA
-----------------

{15}------------------------------------------------

Agreement%96.67%100%97.62%
95% CI90.65 to 98.86%90.36 to 100%93.23 to 99.19%

Method comparison for Diazyme Human Lambda (λ) FLC Assay

The method comparison of the assay was evaluated following CLSI EP9-A2 protocol. A total of 126 serum samples ranging from 6.19-3917.80 mg/L were tested in comparison with predicate assay. Among the 106 samples, in which 39 were Multiple Myeloma (MM), 12 were Monoclonal Gammopathy of Undetermined Significance (MGUS), 34 were Abnormal (due to other disease states) and 41 were Normal (no disease), the concordance between the subject and predicate Lambda FLC was 95%. The results of regression analysis are summarized in the following table:

ParameterLinear RegressionDeming Regression
n126126
Slope1.0541.070
95% CI1.021 to 1.0861.038 to 1.102
Intercept0.636-3.160
95% CI-20.746 to 22.018-24.630 to 18.310
Corr Coeff (R2)0.9720.972
Sample range6.19-3917.806.19-3917.80

Using CLSI EP12-A2: Evaluation of Qualitative Test Performance; Approved Guideline, Second Edition as guideline, the positive percent agreement (PPA), negative percent agreement (NPA) and overall percent agreement (TPA) for Lambda FLC comparisons to predicate and associated 95% confidence intervals were calculated. The results are listed in the table below:

PPANPATPA
Agreement%95.65%94.74%95.24%
95% CI87.98 to 98.51%85.63 to 98.19%90.00 to 98.19%

Detection limit

Detection limit for Diazyme Human Kappa (K) FLC Assay

The LOB, LOD, LOQ of the Assay was determined according to CLSI EP17-A2. The LOB was determined to be 1.2 mg/L; the LOD was determined to be 2.0 mg/L; the LOQ was determined to be 4.5 mg/L.

Detection limit for Diazyme Human Lambda (^) FLC Assay

The LOB, LOD, LOQ of the Assay was determined according to CLSI EP17-A2. The LOB was determined to be 1.7 mg/L; the LOD was determined to be 2.9 mg/L; the LOQ was determined to be 6.1 mg/L.

Interference

The following substances normally present in the blood produced less than 10% deviation when tested at levels equal to the concentrations listed below, according to CLSI EP7-A2.

InterferentKappa FLC ConcentrationToleranceLambda FLC ConcentrationTolerance
------------------------------------------------------------------------------------------

{16}------------------------------------------------

InterferentKappa FLC ConcentrationToleranceLambda FLC ConcentrationTolerance
Triglyceride1000 mg/dL1000 mg/dL
Ascorbic Acid10 mM10 mM
Bilirubin40 mg/dL40 mg/dL
Bilirubin Conjugated40 mg/dL40 mg/dL
Hemoglobin1000 mg/dL1000 mg/dL
Rheumatoid Factor100 IU/mL100 IU/mL

Hook Effect

No high dose hook effect for Diazyme Human Kappa (k) FLC Assay was observed up to 100,000 mg/L.

No high dose hook effect for Diazyme Human Lambda (λ) FLC Assay was observed up to 100,000 mg/L.

Rationale for Considering the Device Substantially Equivalent to Devices Approved for Inter-State Commerce

Freelite® Human Kappa Free Kit/ Freelite® Human Lambda Free Kit was selected for method comparison with Diazyme Human Kappa(x)/Lambda (2.) Free Light Chain Assay. The reagents used for the Diazyme Human Kappa (x) and Lambda (λ) Free Light Chain Assay are similar to the predicate and were used to develop the application for the Hitachi 917 analyzer. The similarities and differences between the predicate reagent and the Diazyme Human Kappa (x) and Lambda (λ) Free Light Chain Assay are given in the table above. Detailed performance characteristics and comparison analysis are given in the filing and demonstrate substantial equivalence to predicate device.

The performance characteristics of the Diazyme Human Kappa (K) and Lambda (λ) Free Light Chain Assay are substantially similar to that of the approved predicate test. Performance data and risk analysis indicates that differences should not affect the safety and effectiveness of the Diazyme Human Kappa (k) and Lambda (λ) Free Light Chain Assay and offers users an in vitro diagnostic device system to measure Human Kappa (k) and Lambda (λ) Free Light Chain in human serum.

Conclusion

Detailed method comparison analysis presented in this 510k submission, together with linearity, precision and interference studies, demonstrates that the Diazyme Human Kappa(k)/Lambda (λ.) Free Light Chain Assay performance is acceptable, safe and effective. There is no significant deviations between the results obtained by Diazyme Human Kappa (k) and Lambda (λ) Free Light Chain Assay and the legally marketed predicate device Freelite® Human Kappa Free Kit/ Freelite® Human Lambda Free Kit (K023009, K023131, K040009) when testing clinical patient samples and is thus substantially similar.

§ 866.5550 Immunoglobulin (light chain specific) immunological test system.

(a)
Identification. An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.(b)
Classification. Class II (performance standards).