K Number
K153371

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-03-28

(126 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sonicision Cordless Ultrasonic Dissection Device is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The device can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures. The Sonicision Cordless Ultrasonic Dissection Device can be used to coagulate isolated vessels up to 5 mm in diameter.

The Sonicision 13 cm device is also indicated for use in otorhinolaryngologic (ENT) procedures.

Device Description

The Sonicision Ultrasonic Dissection Device that is the subject of this 510(k) is a hand-held batterypowered device used to dissect through tissues and to coagulate vessels up to 5 mm in diameter. An assembled Sonicision device includes three components: (1) The Sonicision Dissector, 13 cm; (2) the Sonicision Battery; and (3) the Sonicision Generator. The dissector is a single-use item that is provided to the user in sterile form. The generator and battery are reusable components that are sterilized by the user. The clinical intended use is achieved by the surgeon when pressure is applied to tissue placed between the clamping jaw and the exposed portion of the probe while activating ultrasonic energy with a button on the handpiece.

AI/ML Overview

The provided text does not contain information about the acceptance criteria and the study that proves the device meets them in the format of a clinical study or performance goals with specific metrics for a new AI/software device. Instead, the document is a 510(k) summary for a physical medical device – the Sonicision Cordless Ultrasonic Dissection Device – seeking clearance based on substantial equivalence to a predicate device.

The "performance" section describes comparative ex vivo and in vivo testing to demonstrate that the new device performs similarly to predicate devices, particularly for its proposed use in Otorhinolaryngology (ENT) procedures.

Here's a breakdown of the available information, adapted to the requested format where possible, but highlighting that this is not a study proving an AI/software device meets acceptance criteria:

Study Type: Substantial Equivalence Evaluation for a Physical Surgical Device

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implicit - based on Predicate)Reported Device Performance (Sonicision)
Vessel Burst Strength (Ex vivo)Comparable to FOCUS+ deviceBlood vessels coagulated by Sonicision had comparable burst strength to those coagulated by FOCUS+.
Max Active Blade Temperature (Ex vivo)Comparable to FOCUS+ active bladeMaximum temperature of Sonicision active blade was comparable to FOCUS+ active blade after multiple activations on mesentery.
Max Distal Shaft Temperature (Ex vivo)Comparable to Thunderbeat distal shaftMaximum temperature of Sonicision distal shaft was comparable to Thunderbeat distal shaft after multiple activations on mesentery in ultrasonic mode.
Hemostasis Rates (Acute In vivo)Comparable to FOCUS+ deviceSonicision device and FOCUS+ device achieved comparable rates of hemostasis.
Clinical Acceptability (Usability/Human Factors)Clinically acceptable for ENT surgical procedures (subjective)Practicing surgeons found the 13 cm Sonicision device to be clinically acceptable for use in ENT surgical procedures.

2. Sample size used for the test set and the data provenance

  • Vessel Burst Testing: Not specified, but involved "blood vessels."
  • Ex vivo Tissue Testing (Temperatures): Not specified, but involved "mesentery."
  • Acute In vivo Testing (Hemostasis): Not specified.
  • Usability/Human Factors Validation: Not specified, but used a "human cadaver model."

The data provenance is from ex vivo (tissue from outside a living organism, likely animal or human cadaveric) and in vivo (within a living organism, likely animal) studies, along with human cadaver model studies. The country of origin is not specified but implicitly within the context of FDA submission, suggesting studies conducted under applicable US regulations or international standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Usability/Human Factors Validation: Involved "practicing surgeons" who were "predominantly users of the FOCUS+ device for ENT surgeries." The exact number is not specified, nor are specific years of experience. For the other performance tests (burst strength, temperatures, hemostasis), the ground truth is objective measurement, not expert consensus.

4. Adjudication method for the test set

  • Not applicable as the performance tests primarily involved objective measurements rather than subjective assessments requiring adjudication. For the usability study, the "ground truth" was the surgeon's subjective assessment of clinical acceptability, not a formally adjudicated outcome.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done. This document pertains to a physical surgical device, not an AI/software device. The usability study involved human operators using the device, but it was not an MRMC study comparing human reader performance with and without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This relates to a physical surgical device, not an algorithm.

7. The type of ground truth used

  • Vessel Burst Strength: Objective measurement (burst pressure).
  • Max Active Blade Temperature: Objective measurement (temperature readings).
  • Max Distal Shaft Temperature: Objective measurement (temperature readings).
  • Hemostasis Rates: Objective observation of bleeding control.
  • Clinical Acceptability (Usability/Human Factors): Expert subjective assessment/opinion from practicing surgeons.

8. The sample size for the training set

  • Not applicable. This is a physical device, and the described studies are for performance validation, not for training an algorithm.

9. How the ground truth for the training set was established

  • Not applicable. (See point 8)

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement. The profiles are arranged in a way that they also resemble a bird in flight. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 28, 2016

Covidien LLC Ms. Nancy Sauer Principal Regulatory Affairs Specialist 5920 Longbow Drive Boulder, Colorado 80301

Re: K153371

Trade/Device Name: Sonicision Cordless Ultrasonic Dissection Device Regulation Name: Instrument, Ultrasonic Surgical Regulatory Class: Unclassified Product Code: LFL Dated: February 24, 2016 Received: February 26, 2016

Dear Ms. Sauer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153371

Device Name

Sonicision™ Cordless Ultrasonic Dissection Device

Indications for Use (Describe)

The Sonicision Cordless Ultrasonic Dissection Device is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The device can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, urologic, exposure to orthopedic structures (such as spine and ioint space) and other open and endoscopic procedures. The Sonicision Cordless Ultrasonic Dissection Device can be used to coagulate isolated vessels up to 5 mm in diameter.

The Sonicision 13 cm device is also indicated for use in otorhinolaryngologic (ENT) procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

510(k) Number: K153371

Date summary prepared: 03/23/16

510(k) Submitter/Holder

Covidien 5920 Longbow Drive Boulder, CO 80301

Contact

Nancy Sauer Regulatory Affairs Product Manager Telephone: 303-581-6791 Fax: 303-530-6313 Email: nancy.k.sauer@medtronic.com

Regulatory Information

Trade Name:Sonicision™ Cordless Ultrasonic Dissection Device• Sonicision Dissector, 13 cm• Sonicision Generator• Sonicision Battery
Common Name:Ultrasonic Dissection Device
Regulation Number:NA
Regulation Name:Instrument, Ultrasonic Surgical
Regulatory Class:Unclassified
Product Code:LFL

Predicate Device

The primary predicate device for this submission is the Harmonic FOCUS®+ Shears with Adaptive Tissue Technology (FOCUS+), cleared by 510(k) K133314. The FOCUS+ device is an ultrasonic dissector indicated for use in otorhinolaryngology (ENT) procedures. The Olympus Thunderbeat 0510IC device serves as a reference predicate device (cleared under K132703). The Thunderbeat is a combination ultrasonic/radiofrequency device that has a shaft-based design, similar in size and shape to the Sonicision device assembled with the 13 cm dissector. The Sonicision generator and battery are the same as the designs cleared under 510(k) K101797 and the 13 cm dissector is the same as the design cleared under 510(k) K141371.

Device Description

The Sonicision Ultrasonic Dissection Device that is the subject of this 510(k) is a hand-held batterypowered device used to dissect through tissues and to coagulate vessels up to 5 mm in diameter. An assembled Sonicision device includes three components: (1) The Sonicision Dissector, 13 cm; (2) the Sonicision Battery; and (3) the Sonicision Generator. The dissector is a single-use item that is provided to the user in sterile form. The generator and battery are reusable components that are sterilized by the user. The clinical intended use is achieved by the surgeon when pressure is applied to tissue placed between the clamping jaw and the exposed portion of the probe while activating ultrasonic energy with a button on the handpiece.

{4}------------------------------------------------

Indications for Use

The Sonicision Cordless Ultrasonic Dissection Device is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The device can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures. The Sonicision Cordless Ultrasonic Dissection Device can be used to coagulate isolated vessels up to 5 mm in diameter.

The Sonicision 13 cm device is also indicated for use in otorhinolaryngologic (ENT) procedures.

Technological Characteristics

Sonicision 13 cm dissector

This is a sterile single-use device that includes the following features:

  • . Active blade that vibrates at ultrasonic frequency and delivers the energy that provides the tissue effect.
  • Clamping jaw that the surgeon uses to provide pressure to vessels, tissues, or vascular bundles as needed to deliver the desired tissue effect.
  • Controls for activating the delivery of ultrasonic energy and for opening and closing the clamping ● iaw.
  • Features that interface with the Sonicision generator and the Sonicision battery ●

Sonicision Generator

This is a reusable component that is sterilized by the user facility using a low-temperature hydrogen peroxide plasma process. It has the following features

  • Transducer that converts electrical energy into ultrasonic energy (resonant frequency of approximately 55.5 kHz)
  • . Indicator LEDs that notify the user of the state of the assembled device
  • Control software

Sonicision Battery

This is a reusable component that is sterilized by the user facility using a low-temperature hydrogen peroxide plasma process. It has the following features

  • Li-ion chemistry
  • 14.8 VDC, rechargeable

Comparison to the Predicate Device

Similarities

  • . Same intended use and indicated for the same types of surgeries
  • Same contraindications ●
  • Same operating principle a metal waveguide vibrates at ultrasonic frequencies to create thermal . and mechanical effects that disrupt cells and tissues. The user can also apply pressure with the passive jaw to achieve specific tissue effects (e.g., coagulating a vessel).
  • Same resonant frequency, approx. 55.5 kHz
  • Both have two modes of operation: a minimum power mode that is used for vessel coagulation ● and a maximum power mode that is used for dissection.

{5}------------------------------------------------

Differences

The table below outlines the major differences between the two devices.

CharacteristicSonicisionFOCUS+
Power source14.8 VDC Lithium-ion batteryElectrical current delivered throughEthicon Gen11 generator
Instrument form factorShaft-based designShears-style design
Jaw shapeStraightCurved
Footswitching optionNoYes
Audio and Visual IndicatorsVisual and audio indicators producedby the hand-held deviceAudio and visual indicators produced bythe generator.

The impact of these differences was evaluated through:

(1) Comparative studies described below showing equivalent performance of the Sonicision and FOCUS+ devices

(2) Comparison to a reference predicate, the Thunderbeat 0510IC device, a shaft-based device also indicated for use in ENT surgery, which has a similar size and shape to the Sonicision device.

(3) Human factors evaluation of the 13 cm Sonicision Cordless Ultrasonic Dissection Device for use in ENT surgeries, also described below

Performance

Because the Sonicision Cordless Ultrasonic Dissection Device has been previously cleared for use in other types of surgery, the performance testing in this 510(k) focused only on the testing needed to demonstrate suitability of the Sonicision system for otorhinolaryngology (ENT) surgical procedures and substantial equivalence to the predicate device for characteristics deemed relevant for ENT surgical procedures.

  • . Ex vivo burst testing showed that blood vessels coagulated by the Sonicision device had comparable burst strength to the same type of blood vessels coagulated by the FOCUS+ device.
  • . Ex vivo tissue testing showed that the maximum temperature of the Sonicision active blade was comparable to the maximum temperature of the FOCUS+ active blade after multiple activations on mesentery.
  • . Ex vivo tissue testing showed that the maximum temperature of the Sonicision distal shaft was comparable to the maximum temperature of the Thunderbeat distal shaft after multiple activations on mesentery in the ultrasonic mode.
  • Acute in vivo testing showed that the Sonicision device and the FOCUS+ device achieved comparable rates of hemostasis.
  • . Usability/ human factors validation in a human cadaver model demonstrated that practicing surgeons could successfully complete representative ENT procedures using the 13 cm Sonicision device. The surgeons, who were predominantly users of the FOCUS+ device for ENT surgeries, found that the 13 cm Sonicision device is clinically acceptable for use in ENT surgical procedures.

{6}------------------------------------------------

Conclusion

The comparison of device characteristics and the review of the performance data support the conclusion that the 13 cm Sonicision Cordless Ultrasonic Dissection Device is substantially equivalent to the Harmonic FOCUS+ Shears with Advanced Tissue Technology device with regard to a general indication for use in otorhinolaryngology (ENT) surgery.

N/A