(85 days)
Radiance V3 is a software system intented for treatment planning and analysis of radiation therapy administered with devices suitable for intraoperative radiotherapy.
The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
The intended users of Radiance V3 shall be clinically qualified radiation therapy staff trained in using the system.
Radiance V3 is a treatment planning system, that is, a software program for planning and analysis of radiation therapy plans. Typically, a treatment plan is created by importing patient images obtained from a CT scanner, defining regions of interest either manually or semi-automatically, deciding on a treatment setup and objectives, optimizing the treatment parameters, comparing alternative plans to find the best compromise, computing the clinical dose distribution, approving the plan and exporting it.
Radiance V3 improvements over Radiance V2 (K133655) are:
- A Hybrid Monte Carlo dose computation algorithm for photons for the ● INTRABEAM.
- . A beam modeling tool which models and verifies the treatment unit model with measurements for the INTRABEAM.
- . An improved DICOM interface including PACS query&retrieve functionality. Storage SCP and DICOM.RT Structures and Dose exportation.
The list of compatible IOERT/IORT devices are:
- Intrabeam (a Carl Zeiss product) ●
- . NOVAC7 and NOVAC11 (a SIT product)
- . LIAC10 and LIAC12 (a SIT product)
- MOBETRON (an IntraOp Medical product)
- . Conventional LINACs with adapted cylindrical IOERT applicators (telescopic or fixed ones).
Radiance V3 has been tested with Elekta/Precise and Varian/21EX LINACs with particular IOERT cylindrical telescopic applicators.
Characteristics of radiance include:
-
- Image manipulation and visualization
-
- IORT applicator simulation
-
- Contouring manual and interpolation tools
-
- Dose calculation algorithms, including:
- a. For Intrabeam, Dose Painting for a fast (a few seconds) interpolation of PDD around the applicator or Hybrid Monte Carlo for a good combination of computation time (a few minutes) and accuracy.
- b. For IOERT, pencil beam for a fast (less than one minute) calculation of the dose or Monte Carlo for a good combination of computation time (between 1-10 minutes in most of the cases) and accuracy.
-
- Reporting.
-
- DICOM & DICOM.RT compatibility
Here's an analysis of the acceptance criteria and study information for the Radiance V3 device, based on the provided text:
Important Note: The provided document is a 510(k) summary, which often focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive de novo clinical trial data. Therefore, detailed performance metrics and statistical analyses typically found in full clinical study reports are not explicitly detailed here. The information below is extracted and inferred from the available text.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria in terms of performance metrics (e.g., specific accuracy thresholds for dose calculation). Instead, it focuses on verifying that the device meets its predefined product requirements and relevant industry standards. The performance is reported as meeting these requirements and passing various tests.
| Acceptance Criterion (Inferred from the document) | Reported Device Performance |
|---|---|
| New functionality requirements (Hybrid Monte Carlo for INTRABEAM) | Verification tests were written and executed, and the system passed these tests. The Hybrid Monte Carlo algorithm corrects the dose according to tissue density, providing a more accurate simulation of the dose. |
| Risk mitigation functions | Tests were executed to ensure risk mitigation functions as intended. The system passed these tests. |
| Continued safety and effectiveness of existing functionality (Regression Testing) | Regression tests were conducted to ensure continued safety and effectiveness of existing functionality. The system passed these tests. |
| Compliance with predefined product requirements | Validation and Verification Testing indicated that Radiance V3 meets its predefined product requirements. |
| Compliance with product standards | Validation and Verification Testing indicated compliance with: • IEC 61217 Radiotherapy equipment - Coordinates, movements and scales • IEC 62083 Medical electrical equipment - Requirements for the safety of radiotherapy treatment planning systems • IEC 62366 Medical devices - Application of usability engineering to medical devices |
| Accuracy of dose calculation functions (simulated clinical setup) | Validation testing involved algorithm testing which validated the accuracy of dose calculation functions using a simulated clinical setup. No specific quantitative accuracy metrics (e.g., % difference from a reference) are provided, but the product was "deemed fit for clinical use." The new Hybrid Monte Carlo algorithm is explicitly stated to provide "a more accurate simulation of the dose received to the tissue" compared to the predicate's 'Dose Painting' method. Algorithms were confirmed for a wide variety of field geometries, treatment units, setups, and patient positions. |
| Overall Safety and Effectiveness | Radiance V3 passed testing and was deemed safe and effective for its intended use, demonstrating substantial equivalence to the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not specify a "test set" in terms of number of cases or patients. Instead, it refers to "Over 150 tests" that were executed, including verification tests for new functionality, risk mitigation, and regression tests. These tests involved simulated clinical workflows and algorithm testing with simulated setups. It's not clear if these 150+ tests refer to individual "cases" or different functional aspects.
- Data Provenance: The testing involved "simulated clinical workflows" and "algorithm testing using a simulated clinical setup." This implies that the data was not derived from real patient studies but rather from synthetic or phantom data to simulate various treatment scenarios. Therefore, there is no country of origin for real patient data, and it is a simulated/retrospective (in the sense of being pre-designed scenarios) evaluation rather than prospective clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not provide information on the number of experts or their qualifications used to establish ground truth for the test set. Given that the testing involved "simulated clinical workflows" and "algorithm testing using a simulated clinical setup," the "ground truth" likely refers to established physical models, reference dose calculations (e.g., from more precise scientific methods or established phantoms), or mathematically derived correct outputs for the simulated scenarios.
4. Adjudication Method for the Test Set
The document does not specify an adjudication method. Since the ground truth for the test set seems to be derived from physical/mathematical models or established reference calculations in simulated environments, a human adjudication process by multiple experts is not directly applicable in the same way it would be for interpreting medical images. The evaluation of results against the expected simulated outcomes would be done internally by the engineering and testing teams.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. The document explicitly states: "Clinical trials were not performed as part of the development of this product." Therefore, there is no information on the effect size of human readers improving with or without AI assistance, as human reader performance was not evaluated.
6. Standalone (Algorithm Only) Performance Study
Yes, a standalone performance evaluation was done. The entire "Performance Data" section (Section 7) describes verification and validation testing performed on the Radiance V3 system itself. This included:
- Validation and Verification Testing against predefined product requirements and industry standards.
- Algorithm testing to validate the accuracy of dose calculation functions using a simulated clinical setup.
- Over 150 tests executed for new functionality, risk mitigation, and regression.
This testing addresses the performance of the algorithm and software in a standalone capacity, without direct human-in-the-loop performance measurement.
7. Type of Ground Truth Used
The ground truth used was primarily based on:
- Predefined product requirements: The system's output was compared against its own specified functionalities and performance targets.
- Industry standards (IEC 61217, 62083, 62366): Compliance with these standards served as a form of ground truth for safety, design, and usability.
- Simulated clinical setups and reference dose calculations: For dose calculation accuracy, the "ground truth" would have been established by known physical principles, validated models, or highly accurate computational methods for the simulated scenarios, confirming whether the algorithm's output matched the expected physical reality.
- Predicate device comparison: The substantial equivalence argument relies on the Radiance V3 performing comparably or better than the predicate (Radiance V2), particularly with the new Hybrid Monte Carlo algorithm offering "a more accurate simulation."
8. Sample Size for the Training Set
The document does not provide information on the sample size for the training set. Treatment planning software like Radiance V3, while complex, typically relies on established physics models and algorithms rather than machine learning models that require explicit "training sets" in the sense of labeled data for supervised learning. The algorithms are built upon scientific principles and validated against known physical phenomena, not "trained" on a dataset of cases. Therefore, a training set in the typical AI/ML context is not applicable here.
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" and associated "ground truth" in the machine learning sense is not applicable to this type of physics-based treatment planning software. The algorithms are based on physics equations and models validated through established scientific methods and comparisons to physical measurements (e.g., beam modeling and output factors for INTRABEAM measurements mentioned in the "Beam modeling tool" section).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 16, 2016
GMV Soluciones Globales Internet S.A.U. % Ms. Patsy Trisler Regulatory Consultant Oserve Group US Inc. P.O. Box 940 CHARLESTOWN NH 03603
Re: K153368
Trade/Device Name: Radiance V3 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: November 19, 2015 Received: November 23, 2015
Dear Ms. Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K153368
Device Name Radiance V3
Indications for Use (Describe)
Radiance V3 is a software system intented for treatment planning and analysis of radiation therapy administered with devices suitable for intraoperative radiotherapy.
The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
The intended users of Radiance V3 shall be clinically qualified radiation therapy staff trained in using the system.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
1. SUBMITTER
510(k) Summarv
| Submitter Name: | GMV Soluciones Globales Internet S.A.U. | |
|---|---|---|
| Submitter Address: | Isaac Newton, 11; PTM Tres CantosMadrid 28760Spain | |
| Phone Number: | 011 34 91 807 22 70 | |
| Fax Number: | 011 34 91 807 21 99 | |
| Contact Person: | Carlos Illana Alejandro | |
| Date Prepared: | 30 October 2015 | |
| 2. DEVICE | ||
| Device Trade Name: | Radiance V3 | |
| Common Name: | Radiation Treatment Planning Software |
Classification Name, Medical charged-particle radiation therapy system 21 CFR 892.5050 Number & Product Code: MUJ
3. PREDICATE DEVICE
K133655 Predicate Devices: Radiance V2 (cleared 01/31/2015) This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
4. DEVICE DESCRIPTION
Radiance V3 is a treatment planning system, that is, a software program for planning and analysis of radiation therapy plans. Typically, a treatment plan is created by importing patient images obtained from a CT scanner, defining regions of interest either manually or semi-automatically, deciding on a treatment setup and objectives, optimizing the treatment parameters, comparing alternative plans to find the best compromise, computing the clinical dose distribution, approving the plan and exporting it.
Radiance V3 improvements over Radiance V2 (K133655) are:
- A Hybrid Monte Carlo dose computation algorithm for photons for the ● INTRABEAM.
- . A beam modeling tool which models and verifies the treatment unit model with measurements for the INTRABEAM.
- . An improved DICOM interface including PACS query&retrieve functionality. Storage SCP and DICOM.RT Structures and Dose exportation.
The list of compatible IOERT/IORT devices are:
- Intrabeam (a Carl Zeiss product) ●
- . NOVAC7 and NOVAC11 (a SIT product)
- . LIAC10 and LIAC12 (a SIT product)
- MOBETRON (an IntraOp Medical product)
- . Conventional LINACs with adapted cylindrical IOERT applicators (telescopic or fixed ones).
Radiance V3 has been tested with Elekta/Precise and Varian/21EX LINACs with particular IOERT cylindrical telescopic applicators.
{4}------------------------------------------------
Characteristics of radiance include:
-
- Image manipulation and visualization
-
- IORT applicator simulation
-
- Contouring manual and interpolation tools
-
- Dose calculation algorithms, including:
- a. For Intrabeam, Dose Painting for a fast (a few seconds) interpolation of PDD around the applicator or Hybrid Monte Carlo for a good combination of computation time (a few minutes) and accuracy.
- b. For IOERT, pencil beam for a fast (less than one minute) calculation of the dose or Monte Carlo for a good combination of computation time (between 1-10 minutes in most of the cases) and accuracy.
-
- Reporting.
-
- DICOM & DICOM.RT compatibility
5. INDICATIONS FOR USE
Radiance V3 is a software system intended for treatment planning and analysis of radiation therapy administered with devices suitable for intraoperative radiotherapy.
The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
The system functionality can be configured based on user needs.
The intended users of Radiance V3 shall be clinically qualified radiation therapy staff trained in using the system.
6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
| Summary ofTechnologicalCharacteristics | The technological characteristics are essentially the same asthose of the predicate.Both devices produce treatment plans with corresponding dosedistributions computed using a three dimensional dosimetryengine. Both devices have a function of electronic approval oftreatment plans by trained and authorized staff, and export inDICOM format for commencing treatment or archiving. |
|---|---|
| SubstantialEquivalence | From the standpoint of both functionality and workflow theRadiance V3 device is substantially equivalent to the identifiedpredicate as follows:• Within Radiance V3 and its predicate Radiance V2, the usercan adjust parameters to achieve a predicted outcome, ratherthan make a decision intra-operatively.• Radiance V3 and its predicate Radiance V2 are designed toanalyze and plan radiation treatments in three dimensions forthe purpose of treating patients with malignancies. |
{5}------------------------------------------------
- Radiance V3 and its predicate Radiance V2 provide treatment plans with estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by qualified medical personnel.
- · Radiance V3 and its predicate Radiance V2 use externally acquired medical images and user input to achieve the result.
The Radiance V3 dose distribution computation algorithms are identical to those implemented in Radiance V2 except for Hybrid Monte Carlo algorithm for INTRABEAM that has been added. Radiance V2 dose calculation algorithm is what we call 'Dose Painting' which is an interpolation of the PDD around the applicator (assuming isotropy). Hybrid Monte Carlo algorithm corrects the dose according to the tissue density which provides a more accurate simulation of the dose received to the tissue. Verification tests were written and executed to ensure that the system is working as designed. Over 150 tests were executed, including tests to verify requirements for new product functionality, tests to ensure the risks mitigation functions as intended, and regression tests to ensure continued safety and effectiveness of existing functionality. Radiance V3 passed testing and was deemed safe and effective for its intended use.
7. PERFORMANCE DATA
| Non Clinical Data | Validation and Verification Testing carried out on the Radiance V3 indicates that it meets its predefined products requirements and requirements from the following product standards:• IEC 61217 Radiotherapy equipment - Coordinates, movements and scales• IEC 62083 Medical electrical equipment - Requirements for the safety of radiotherapy treatment planning systems• IEC 62366 Medical devices - Application of usability engineering to medical devices |
|---|---|
| Clinical Data | Clinical trials were not performed as part of the development of this product. Clinical testing on patients is not advantageous in demonstrating substantial equivalence or safety and effectiveness of the device since testing can be performed such that no human subjects are exposed to risk. Validation testing involved simulated clinical workflows, and algorithm testing which validated the accuracy of dose calculation functions using a simulated clinical setup. The product was deemed fit for clinical use. |
{6}------------------------------------------------
Item 4b of Table I in the FDA Guidance document entitled, Software Verification "Guidance for the Content of Premarket Submissions for and Validation Software Contained in Medical Devices asks, "Does the Software Testing Device control the delivery of potentially harmful enerqy that could result in death or serious injury, such as radiation treatment systems..." Radiance V3 does not directly control the linear accelerator that delivers the radiation. Once completed, plans are reviewed and approved by qualified clinicians and may be subject to quality assurance practices before treatment actually takes place. There is no automatic link between the Radiance V3 software and the linear accelerator. However, should a flaw in the treatment plan escape the notice of the qualified professionals using the Radiance V3 system, serious injury or death could result. Therefore, we believe Radiance V3 to be of major level of concern.
8. CONCLUSION
The information discussed above demonstrates that the Radiance V3 device is substantially equivalent to the predicate device.
{7}------------------------------------------------
| Feature | New Device: Radiance V3 | Predicate: Radiance V2 |
|---|---|---|
| 510(k) Number | K153368 | K133655 |
| Manufacturer | GMV Soluciones GlobalesInternet S.A.U. | GMV Soluciones GlobalesInternet S.A.U. |
| Classification #& Product Code | 21 CFR 892.5050MUJ | 21 CFR 892.5050MUJ |
| Indication foruse | Radiance V3 is a softwaresystem intended for treatmentplanning and analysis ofintraoperative radiation therapyadministered with devicessuitable for intraoperativeradiotherapy. | Radiance V2 is a softwaresystem intended for treatmentplanning and analysis ofintraoperative radiation therapyadministered with devicessuitable for intraoperativeradiotherapy. |
| The treatment plans providetreatment unit set-up parametersand estimates of dosedistributions expected during theproposed treatment, and may beused to administer treatmentsafter review and approval by theintended user. | The treatment plans providetreatment unit set-upparameters and estimates ofdose distributions expectedduring the proposed treatment,and may be used to administertreatments after review andapproval by the intended user. | |
| The system functionality can beconfigured based on user needs. | The system functionality can beconfigured based on userneeds. | |
| The intended users of RadianceV3 shall be clinically qualifiedradiation therapy staff trained inusing the system. | The intended users ofRadiance V2 shall be clinicallyqualified radiation therapy stafftrained in using the system. | |
| System Design | Software only | Software only |
| Calculation forelectrons | Dose distributions computedusing a three dimensional doseengine. | Dose distributions computedusing a three dimensional doseengine. |
| Pencil Beam computation andMonte Carlo Computation forelectrons | Pencil Beam computation andMonte Carlo Computation forelectrons | |
| Calculation forphotons | Dose Painting (Planningcalculation interpolation of PDDmeasurements) andHybrid Monte Carlo computation | Dose Painting (Planningcalculation interpolation of PDDmeasurements) |
| Feature | New Device: Radiance V3 | Predicate: Radiance V2 |
| Input | Externally acquired patientmedical images and user input.Calibration files and beammodeling measurements forINTRABEAM. | Externally acquired patientmedical images and user input.Calibration files forINTRABEAM. |
| Output | Treatment plans withcorresponding dose distributions | Treatment plans withcorresponding dosedistributions |
| Plan review andapproval | Allows electronic approval oftreatment plans by trained andauthorized staff | Allows electronic approval oftreatment plans by trained andauthorized staff |
| Dose calculationalgorithmconfirmation | Algorithms confirmed for a widevariety of field geometries,treatment units, treatment setupsand patient positions, includingdifferent dose grid resolutionsettings. | Algorithms confirmed for a widevariety of field geometries,treatment units, treatmentsetups and patient positions,including different dose gridresolution settings. |
| Beam modelingtool | Beam modeling of the treatmentunit based on relativemeasurements and outputfactors. | Beam modeling of thetreatment unit based on relativemeasurements and outputfactors. |
Summary of Technical Characteristics and Comparison to Predicate
{8}------------------------------------------------
N/A