K Number
K153361

Validate with FDA (Live)

Date Cleared
2016-04-15

(147 days)

Product Code
Regulation Number
862.1825
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lumipulse G 25-0H Vitamin D For in vitro diagnostic use.

Lumipulse G 25-0H Vitamin D is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative determination of 25-hydroxyvitamin D (25-0H vitamin D) and other hydroxylated vitamin D metabolites in human serum and plasma (sodium heparin, lithium heparin, or dipotassium EDTA) on the LUMIPULSE G System.

Lumipulse G 25-0H Vitamin D is to be used as an aid in the assessment of vitamin D sufficiency. Assay results should be used in conjunction with other clinical or laboratory data to assist the clinician in making individual patient management decisions.

Lumipulse G 25-0H Vitamin D Calibrators Lumipulse G 25-0H Vitamin D Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G 25-0H Vitamin D on the LUMIPULSE G System.

Device Description

Lumipulse G 25-OH Vitamin D is an assay system, including a set of immunoassay reagents, for the quantitative measurement of 25-OH Vitamin D in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System. 25-OH Vitamin D Calibrator or specimen is initially auto-diluted with Specimen Diluent 1 in the system. 25-OH vitamin D in specimens is separated from vitamin D binding protein by the extraction agent and specifically bound to anti-25-OH vitamin D monoclonal antibody (sheep) on the particles, and antigen-antibody immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Alkaline phosphatase (ALP: calf)-labeled anti-(25-OH vitamin D/anti- 25-OH vitamin D monoclonal antibody immunocomplexes) recombinant chicken monoclonal antibody specifically binds to 25-OH vitamin D immunocomplexes on the particles, and additional immunocomplexes are formed The particles are washed and rinsed to remove unbound materials. Substrate Solution is added and mixed with the particles. AMPPD contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles. Luminescence (at a maximum wavelength of 477 nm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of 25-OH vitamin D.

AI/ML Overview

Here's an analysis of the acceptance criteria and study data for the Lumipulse G 25-OH Vitamin D device, based on the provided text:

Acceptance Criteria and Device Performance (Lumipulse G 25-OH Vitamin D)

| Acceptance Criteria Category | Specific Criteria | Reported Device Performance and Study Details |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analytical Performance | Precision/Reproducibility: Requires a total (within-laboratory) CV of ≤ 10%. | Precision: Lumipulse G 25-OH Vitamin D demonstrated precision ≤ 5.2% (total %CV), meeting the criteria. Study: A study was run according to CLSI guideline EP5-A3. Five human serum-based samples (specimen pools) and three controls were assayed in replicates of two at two separate times of the day for 20 days (total 80 replicates for each sample) using one LUMIPULSE G1200 System. |
| | Linearity/Assay Reportable Range: Needs to demonstrate linearity throughout the assay range. | Linearity: Linearity was found in the range of 6.9 to 150.0 ng/mL for serum and plasma. The device correlated with expected concentrations: Serum: y = 1.0382(x) + 1.251 (R-squared: 0.9978); Plasma: y = 1.0151(x) + 6.7669 (R-squared: 0.9936). Study: Consistent with CLSI Guideline EP6-A. Two human serum specimen pools and two K2EDTA plasma specimen pools with high 25-Hydroxyvitamin D2 and 25-Hydroxyvtiamin D3 levels were diluted with one human serum specimen pool and one dipotassium EDTA plasma specimen pool with low 25-Hydroxyvitamin D2 and 25-Hydroxyvtiamin D3 levels. |
| | High Dose Hook Effect: No high dose hook effect should be observed. | High Dose Hook: No high dose hook effect was observed for samples up to 3,000 ng/mL of 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3. |
| | Detection Limit: LoB, LoD, and LoQ of ≤ 4.0 ng/mL. | Detection Limit: LoB: 0.0 ng/mL LoD: 0.28 ng/mL LoQ: 3.33 ng/mL All are ≤ 4.0 ng/mL, meeting the criteria. Study: LoD and LoQ determined consistent with CLSI guideline EP17-A2. For LoD, seven low-level specimens were tested over 3 days using two LUMIPULSE G1200 Systems and two Lumipulse G 25-OH Vitamin D lots (120 determinations per panel). LoQ is the analyte level at which the CV is 10%. |
| | Analytical Specificity (Interferences): Average interference of ≤ 10% for each compound. | Interferences: Demonstrated average interference of ≤ 10% for a range of endogenous and therapeutic compounds (e.g., Bilirubin, Triglycerides, Hemoglobin, Acetaminophen, Ibuprofen, etc.). This meets the criteria. Study: Consistent with CLSI guideline EP7-A2. Human serum specimens with 25-OH vitamin D concentrations of approximately 20, 40, and 100 ng/mL were supplemented with potentially interfering compounds. |
| | Analytical Specificity (Cross-Reactivity): Cross-reactivity should be evaluated for similar substances. | Cross-Reactivity: Cross-reactivity was evaluated for various vitamin D metabolites and related compounds. Some compounds showed notable cross-reactivity (e.g., 1,25(OH)2 vitamin D2 at 143% at 10 ng/mL, 1,25(OH)2 vitamin D3 at 39% at 100 ng/mL, 24,25(OH)2 vitamin D3 at 21% at 100 ng/mL). The acceptance criteria for cross-reactivity are not explicitly stated as a percentage here, but the data is provided for evaluation against potential clinical impact.Study: Consistent with CLSI Protocol EP7-A2. Human serum specimens with 25-OH vitamin D concentrations of approximately 20, 40, and 100 ng/mL were supplemented with potentially cross-reacting compounds. |
| Method Comparison | Substantial Equivalence to Predicate Device: Acceptable correlation and bias compared to a legally marketed predicate device (LIAISON® 25 OH Vitamin D TOTAL Assay). | Comparison vs. Predicate (Liaison): n=137 Correlation Coefficient (r): 0.9476 Intercept: -3.073 (95% CI: -4.357 – -1.790) Slope: 1.09 (95% CI: 1.04 – 1.13) Average Bias: 0.423 ng/mL. This performance is presented to demonstrate substantial equivalence, implying it meets the necessary correlation and bias thresholds. Study: Consistent with CLSI guideline EP9-A3. Weighted Deming regression used. Range of samples: 4.0 to 107.3 ng/mL (Lumipulse G); 5.45 to 91.5 ng/mL (Liaison). |
| | Accuracy vs. Reference Method: Acceptable correlation and bias compared to a recognized reference measurement procedure (ID-HPLC-MS/MS). | Accuracy vs. ID-HPLC-MS/MS (CDC Reference Method): n=117 Correlation Coefficient (r): 0.9986 Intercept: -2.788 (95% CI: -3.225 – -2.351) Slope: 0.97 (95% CI: 0.96 – 0.99) Average Bias: -3.826 ng/mL. The high correlation and slope close to 1, with a small bias, indicate good accuracy. Study: Consistent with CLSI guideline EP9-A3. Weighted Deming regression used. Range of samples: 7.2 to 149.0 ng/mL (Lumipulse G); 8.65 to 153 ng/mL (ID-HPLC-MS/MS). |
| | Matrix Equivalency: Results should demonstrate equivalency between different sample matrices (serum, plasma types). | Matrix Equivalency: Demonstrated equivalency between serum (SST vs Red Top), K2EDTA plasma vs Red Top, Lithium Heparin vs Red Top, and Sodium Heparin vs Red Top. Correlation coefficients (r) ranged from 0.9964 to 0.9983, and slopes were close to 1. Study: Fifty (50) matched sets of serum (red top and serum separator tubes (SST)) and plasma (sodium heparin, lithium heparin, K2EDTA) samples were evaluated. |
| Expected/Reference Values | Establish expected values for apparently healthy adults. | Expected Values: Observed range for 25(OH) vitamin D concentrations in 287 apparently healthy adults was 7.6 to 47.6 ng/mL (2.5th to 97.5th percentile), with a mean of 22.9 ng/mL and median of 22.2 ng/mL. Study: Based on 322 samples from apparently healthy adults in the US (diverse regions, seasons, and skin types). Individuals with conditions or medications affecting vitamin D levels were excluded. Established according to CLSI Protocol EP28-A3c. |


Additional Information on the Studies:

  1. Sample sizes used for the test set and the data provenance:

    • Precision/Reproducibility: Five human serum-based samples and three controls, assayed in replicates of two at two separate times of the day for 20 days. This means 80 data points for each sample/control tested to derive the precision data. No specific country of origin is mentioned for these samples, but the general context is a US FDA submission. The nature of the study (analyzing manufactured controls and pooled human serum) suggests a prospective, controlled laboratory setting.
    • Linearity/Reportable Range: Two human serum specimen pools and two K2EDTA plasma specimen pools (high 25-OH vitamin D levels), diluted with one human serum specimen pool and one dipotassium EDTA plasma specimen pool (low 25-OH vitamin D levels). No specific number of individual samples is given, but it involves various dilutions of these pools.
    • Equimolar Recovery (D2/D3): Serum sample pool with low endogenous 25-OH vitamin D, supplemented to create 7 different ratios of D2 to D3. Each ratio was measured in n=3 replicates. Also, known concentrations of 25-OH vitamin D (D2 and D3) added to human serum and K2EDTA plasma samples (low endogenous levels). Measured in n=3 replicates for each spiked concentration across multiple serum/plasma samples.
    • Detection Limit (LoD): Seven low-level specimens, tested over 3 days using two LUMIPULSE G1200 Systems and two Lumipulse G 25-OH Vitamin D lots, resulting in 120 determinations for each panel.
    • Analytical Specificity (Interferences and Cross-Reactivity): Human serum specimens with 25-OH vitamin D concentrations of approximately 20, 40, and 100 ng/mL, supplemented with potentially interfering/cross-reacting compounds. The number of samples for each compound isn't specified but implies multiple measurements per compound at different 25-OHD levels.
    • Method Comparison (vs. Predicate): n=137 samples. Data provenance is not explicitly stated as retrospective or prospective, but clinical method comparison studies typically use prospectively collected patient samples or archived clinical samples. No country of origin is specified.
    • Method Comparison (vs. ID-HPLC-MS/MS): n=117 samples. Similar to the predicate comparison, data provenance is not explicitly stated, but it's a comparison to a CDC reference method.
    • Matrix Comparison: Fifty (50) matched sets of serum (red top and SST) and plasma (sodium heparin, lithium heparin, K2EDTA) samples.
    • Expected Values/Reference Range: 322 apparently healthy adults. These subjects were from two geographically diverse regions of the US (North: Northern California and New Jersey; South: Southern California and North Carolina) and sampled during spring/summer and fall/winter seasons. Specific exclusion criteria were applied. This data is prospectively collected for this purpose.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device is an in-vitro diagnostic assay for quantitative measurement, not an AI or imaging device requiring expert interpretation of results. Therefore, the concept of "experts establishing ground truth for a test set" with qualifications like "radiologist with 10 years of experience" is not directly applicable in the same way.
    • For the Method Comparison vs. ID-HPLC-MS/MS, the "ground truth" or reference measurement was established by the CDC reference method (ID-HPLC-MS/MS). The experts involved would be the analytical chemists and laboratory personnel operating and validating this highly accurate reference method. Their qualifications would involve extensive training and experience in mass spectrometry and laboratory diagnostics.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication methods like "2+1" or "3+1" are typically used in studies involving subjective interpretation (e.g., radiology reads) where multiple human readers agree or a tie-breaker is needed.
    • For this in-vitro diagnostic device, which produces quantitative numerical results, such an adjudication method is not relevant. The device output is a concentration value directly. Any discrepancies in laboratory measurements would be resolved through standard laboratory quality control procedures, repeat testing, or by comparison to a reference method (as seen with the ID-HPLC-MS/MS comparison).
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is an automated, quantitative immunoassay, not an AI-assisted diagnostic tool that aids human readers in interpreting images or other complex data. It provides a direct measurement of a biomarker.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance presented for the Lumipulse G 25-OH Vitamin D assay reflects its standalone performance as an automated, quantitative diagnostic system. It operates without human interpretive input for the final result. Human interaction is limited to sample loading, instrument operation, calibration, and review of results (which is standard for any lab test). The "algorithm" here refers to the immunoassay chemistry and the LUMIPULSE G System's processing of the chemiluminescent signal.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Method Comparison vs. ID-HPLC-MS/MS: The ground truth was based on a validated isotope dilution high performance liquid chromatography tandem mass spectrometry (ID-HPLC-MS/MS) 25-OH Vitamin D method performed as a CDC reference method. This is considered a highly accurate and precise analytical reference method.
    • For Linearity and Equimolar Recovery: Ground truth was established by gravimetric preparation and known concentrations of analytes, confirmed by UV spectrophotometric analysis.
    • For Precision, Detection Limit, Analytical Specificity, Matrix Comparison, and Expected Values: The ground truth for these studies involves the intrinsic properties of the samples themselves (e.g., true concentration, absence/presence of interferents, matrix type) as measured by the device and compared against expected statistical behavior or comparative methods.
  7. The sample size for the training set:

    • The document does not explicitly mention a "training set" in the context of machine learning. This device is an immunoassay, and its development involves optimizing reagents, assay conditions, and calibration curves through laboratory experimentation, rather than training an algorithm on a large dataset in the way AI models are trained.
    • However, the calibrator set itself can be considered analogous to "training data" for the instrument's internal curve fitting. The calibrator kit contains 6 concentrations (0, 10, 20, 50, 100, 150 ng/mL).
  8. How the ground truth for the training set was established:

    • As interpreted from the context of an immunoassay (analogous to the "training set" being the calibrators), the ground truth for the calibrators was established by:
      • Gravimetric preparation: The calibrators were prepared by carefully weighing and diluting the pure analyte to achieve specific concentrations.
      • Traceability: They are traceable to internal reference calibrator concentrations determined by UV spectrophotometric analysis.
      • Verification: Further verified by a Reference Method Procedure (University of Ghent). This ensures high accuracy and reliability of the assigned values for the calibrators.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 15, 2016

FUJIREBIO DIAGNOSTICS, INC. STACEY DOLAN MANAGER, REGULATORY AFFAIRS 201 GREAT VALLEY PARKWAY MALVERN, PA 19355

Re: K153361

Trade/Device Name: Lumipulse G 25-OH Vitamin D, Lumipulse G 25-OH Vitamin D Calibrators Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG, JIT Dated: March 17, 2016 Received: March 18, 2016

Dear Stacey Dolan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Katherine Serrano -S

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153361

Device Name Lumipulse G 25-0H Vitamin D Lumipulse G 25-0H Vitamin D Calibrators

Indications for Use (Describe) Lumipulse G 25-0H Vitamin D For in vitro diagnostic use.

Lumipulse G 25-0H Vitamin D is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative determination of 25-hydroxyvitamin D (25-0H vitamin D) and other hydroxylated vitamin D metabolites in human serum and plasma (sodium heparin, lithium heparin, or dipotassium EDTA) on the LUMIPULSE G System.

Lumipulse G 25-0H Vitamin D is to be used as an aid in the assessment of vitamin D sufficiency. Assay results should be used in conjunction with other clinical or laboratory data to assist the clinician in making individual patient management decisions.

Lumipulse G 25-0H Vitamin D Calibrators Lumipulse G 25-0H Vitamin D Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G 25-0H Vitamin D on the LUMIPULSE G System.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a combination of blue and green shapes, with a white teardrop shape in the center. The company name is written in blue, with "Diagnostics, Inc." written in a smaller, black font below.

Section 5 510(k) SUMMARY

Lumipulse G 25-OH Vitamin D K153361

Date: April 13, 2016

  • Submitter: Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355
  • Contact Person: Stacey Dolan Manager, Requlatory Affairs Phone: (610) 240-3843 Fax: (610) 240-3803 Email: dolans@fdi.com
  • Device Name: Lumipulse G 25-OH Vitamin D Classification: Class II MRG 75, Clinical Chemistry 21 CFR § 862.1825
  • Device Name: Lumipulse G 25-OH Vitamin D Calibrators Classification: Class II JIT 75. Clinical Chemistry 21 CFR § 862.1150
  • Predicate Device: K071480 DIASORIN. INC. LIAISON® 25 OH Vitamin D TOTAL Assay

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Device Description:

Lumipulse G 25-OH Vitamin D is an assay system, including a set of immunoassay reagents, for the quantitative measurement of 25-OH Vitamin D in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System. 25-OH Vitamin D Calibrator or specimen is initially auto-diluted with Specimen Diluent 1 in the system. 25-OH vitamin D in specimens is separated from vitamin D binding protein by the extraction agent and specifically bound to anti-25-OH vitamin D monoclonal antibody (sheep) on the particles, and antigen-antibody immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Alkaline phosphatase (ALP: calf)-labeled anti-(25-OH vitamin D/anti- 25-OH vitamin D monoclonal antibody immunocomplexes) recombinant chicken monoclonal antibody specifically binds to 25-OH vitamin D immunocomplexes on the particles,

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue abstract shape on the left, followed by the company name in blue, with "Diagnostics, Inc." in a smaller font size below the company name. The logo is clean and professional, with a focus on the company name.

and additional immunocomplexes are formed The particles are washed and rinsed to remove unbound materials. Substrate Solution is added and mixed with the particles. AMPPD contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles. Luminescence (at a maximum wavelength of 477 nm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of 25-OH vitamin D.

Lumipulse G 25-OH Vitamin D is an assay system, including a set of immunoassay reagents, for the quantitative measurement of 25-OH Vitamin D in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System.

Lumipulse G 25-OH Vitamin D Immunoreaction Cartridges |REF|235089

The Lumipulse G 25-OH Vitamin D Immunoreaction Cartridges consists of 3 x 14 tests. Each kit contains the following:

1)Antibody-Coated Particle Solution (Liquid when used, 250 µL/Immunoreaction Cartridge) Contains 200 µg/mL anti-25-OH vitamin D monoclonal antibody (sheep)-coated particles, protein stabilizers (bovine and sheep) and chemical stabilizers in 0.15 M sodium chloride/Tris buffer. This solution contains gelatin and turns into gel at 15 ℃ or lower. Preservative: sodium azide.

2)Enzyme-Labeled Antibody Solution

(Liquid. 320 uL/Immunoreaction Cartridge)

Contains 3 µg/mL alkaline phosphatase (ALP: calf)-labeled anti-(25-0H vitamin D / anti-25-OH vitamin D monoclonal antibody immunocomplexes) recombinant chicken monoclonal antibody, protein stabilizers (bovine and calf) and chemical stabilizers in 0.15 M sodium chloride/MES buffer. Preservative: sodium azide.

Lumipulse G 25-OH Vitamin D Calibrators CAL 235096, Liquid 1x6 concentrations Each calibrator kit contains one bottle each of Calibrators 1 – 6. The calibrator kit is packaged separately.

CAL 1 0 ng/mL (0 nmol/L) 25-OH Vitamin D calibrator (1 x 1.5 mL) CAL 2 10 ng/mL (25 nmol/L) 25-OH Vitamin D calibrator (1 x 1.5 mL) CAL 3 20 ng/mL (50 nmol/L) 25-OH Vitamin D calibrator (1 x 1.5 mL) CAL 4 50 ng/mL (125 nmol/L) 25-OH Vitamin D calibrator (1 x 1.5 mL) CAL 5 100 ng/mL (250 nmol/L) 25-OH Vitamin D calibrator (1 x 1.5 mL) CAL 6 150 ng/mL (375 nmol/L) 25-OH Vitamin D calibrator (1 × 1.5 mL) Contains calcifediol in 0.1 M sodium chloride in HEPES buffer with protein stabilizer (bovine). Preservative: sodium azide

Device Intended Use:

Lumipulse G 25-OH Vitamin D For in vitro diagnostic use.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a combination of blue and green shapes, with a white space in the center. The company name is written in blue, with the word "FUJIREBIO" on top and "Diagnostics, Inc." below it in a smaller font.

Lumipulse G 25-OH Vitamin D is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative determination of 25-hydroxyvitamin D (25-OH vitamin D) and other hydroxylated vitamin D metabolites in human serum and plasma (sodium heparin, lithium heparin, or dipotassium EDTA) on the LUMIPULSE G System.

Lumipulse G 25-OH Vitamin D is to be used as an aid in the assessment of vitamin D sufficiency. Assay results should be used in conjunction with other clinical or laboratory data to assist the clinician in making individual patient management decisions.

Lumipulse G 25-OH Vitamin D Calibrators

Lumipulse G 25-OH Vitamin D Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G 25-OH Vitamin D on the LUMIPULSE G System.

Device Indications for Use:

Same as Intended Use

Substantial Equivalence:

Comparison between the Lumipulse G 25-OH Vitamin D Assay and LIAISON® 25 OH Vitamin D TOTAL Assay

Similarities and Differences
ParameterLumipulse G25-OH Vitamin D(Proposed Device)LIAISON® 25 OH VitaminD TOTAL Assay(Predicate Device)K071480
Device TypeClassificationIn vitro diagnosticClass IISame
Regulation Number21CFR § 862.1825 Vitamin D testsystem.Same
Principle of OperationAutomated QuantitativeChemiluminscent ImmunoassaySame
Specimen Collection MethodRoutine Phlebotomy TechniquesSame
Assay Range6.9 – 150 ng/mL4.0 – 150 ng/mL
Intended UseLumipulse G 25-OH Vitamin D is aChemiluminescent EnzymeImmunoassay (CLEIA) for thequantitative determination of 25-hydroxyvitamin D (25-OH vitaminD) and other hydroxylated vitaminD metabolites in human serum andplasma (sodium heparin, lithiumheparin, or dipotassium EDTA) onthe LUMIPULSE G System.Lumipulse G 25-OH Vitamin D is tobe used as an aid in theassessment of vitamin Dsufficiency.The LIAISON 25 OHVitamin D TOTAL Assayuses chemiluminescentimmunoassay (CLIA)technology for thequantitative determinationof 25-hydroxyvitamin D andother hydroxylated vitaminD metabolites in humanserum, EDTA-plasma orlithium heparin plasma tobe used in the assessmentof vitamin D sufficiency.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic with blue and green elements on the left. To the right of the graphic, the text "FUJIREBIO" is displayed in a bold, blue font. Below "FUJIREBIO", the text "Diagnostics, Inc." is written in a smaller, gray font.

Similarities and Differences
ParameterLumipulse G25-OH Vitamin D(Proposed Device)LIAISON® 25 OH VitaminD TOTAL Assay(Predicate Device)K071480
Assay results should be used inconjunction with other clinical orlaboratory data to assist theclinician in making individual patientmanagement decisions.used in conjunction withother clinical or laboratorydata to assist the clinicianin making individual patientmanagement decisions inan adult population.
Instrument SystemLUMIPULSE G SystemLIAISON
Assay TypeDirect sandwichimmunoassay based onchemiluminescenttechnologyCompetitiveimmunoassay based onchemiluminescenttechnology
Sample VolumeSample Volume:20 μLMinimum Volume:120 μL for sample cups and 270 μLfor sample tubesSample Volume:25 μLMinimum Volume:250 μL
Type of SpecimenHuman serum or plasma (sodiumheparin, lithium heparin, ordipotassium EDTA)Human serum or plasma(lithium heparin ordipotassium EDTA)
Analyte Detected25-OH vitamin D and otherhydroxylated vitamin D metabolitesSame

Calibrator Set

Similarities and Differences
ParameterLumipulse G25-OH Vitamin D(Proposed Device)LIAISON® 25 OH VitaminD TOTAL Assay(Predicate Device)K071480
Number of Calibrators6 calibrators, 1.5 mL each2 calibrators, 1.0 mL each
Standardization/TraceabilityThe calibrators for use withLumipulse G 25-OH Vitamin D areprepared gravimetrically and aretraceable to internal referencecalibrator concentrationsdetermined by UVspectrophotometric analysis andverified by Reference MethodProcedure (University of Ghent).Calibrators are traceable toconcentrations determinedby UV spectrophotometricanalysis.
Calibration CurveFull Calibration CurveMaster Calibration Curve
MatrixHEPES buffer with protein stabilizer(bovine)Human serum and BSA

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo consists of two parts: a graphic on the left and the company name on the right. The graphic is a stylized shape in blue and green, resembling a water droplet or a stylized leaf. The company name, "FUJIREBIO Diagnostics, Inc.," is written in a clear, sans-serif font, with "FUJIREBIO" in a larger, bolder font and "Diagnostics, Inc." in a smaller font below it.

Performance Characteristics

Data were generated using the LUMIPULSE G1200 System.

1. Analytical performance:

  • a. Precision/Reproducibility:
    Lumipulse G 25-OH Vitamin D is ≤ 10% total (within-laboratory) CV (coefficient of variation). Lumipulse G 25-OH Vitamin D demonstrated precision ≤ 5.2% (total %CV) in a study run according to the Clinical and Laboratory Standards Institute (CLSI) guideline EP5-A3. Five human serum-based samples (specimen pools), and three controls were assayed in replicates of two at two separate times of the day for 20 for each sample) using one LUMIPULSE G1200 System. Data from this study are presented below.
Within-RunBetween RunWithin DayBetween-DayTotal
SampleMean(ng/mL)SD%CVSD%CVSD%CVSD%CV
ControlLevel 110.30.232.2%0.242.3%0.323.1%0.464.5%
ControlLevel 234.00.361.1%0.481.4%0.541.6%0.812.4%
ControlLevel 370.60.681.0%0.721.0%1.071.5%1.462.1%
Panel 17.70.212.8%0.202.6%0.283.6%0.405.2%
Panel 220.70.331.6%0.241.2%0.321.6%0.522.5%
Panel 341.20.531.3%0.451.1%0.671.6%0.972.3%
Panel 461.00.570.9%0.050.1%0.771.3%0.961.6%
Panel 5120.61.361.1%0.000.0%1.321.1%1.891.6%

b. Linearity/assay reportable range:

Lumipulse G 25-OH Vitamin D demonstrated linearity in a study consistent with the guidelines in the CLSI Guideline EP6-A. Two human serum specimen pools and two K2EDTA plasma specimen pools with high 25-Hydroxyvitamin D2 and 25-Hydroxyvtiamin D3 levels were diluted with one human serum specimen pool and one dipotassium EDTA plasma specimen pool with low 25-Hydroxyvitamin D2 and 25-Hydroxyvtiamin D3 levels throughout the range of the assay. The linearity was found in the range of 6.9 to 150.0 ng/mL for serum and plasma. Lumipulse G 25-OH Vitamin D correlated with expected concentrations according to the linear regression formulas:

y= 1.0382 (x) + 1.251; R-squared: 0.9978 Serum: Plasma: y= 1.0151 (x) + 6.7669; R-squared: 0.9936

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract shape on the left, followed by the company name in blue, and the word "Diagnostics, Inc." in black below it. The logo is clean and modern, and the colors are bright and eye-catching.

Lumipulse G 25-OH Vitamin D has been verified to detect 25-OH vitamin D2 and vitamin D3 in equal parts to the true molar concentration. 25-0H vitamin D2 and 25-OH vitamin D3 were prepared in seven (7) different ratios targeting a final concentration of 50 ng/mL in a serum sample pool with a low endogenous 25-OH vitamin D level. The concentration of 25-OH vitamin D was determined using Lumipulse G 25-OH Vitamin D and the resulting percent recovery was calculated. Samples with varying amounts of 25-OH vitamin D2 and 25-OH vitamin D3 recover within 100 ± 10%, ranging from 101% to 103%. The data are summarized in the following table

Ratio of vitaminD2 to vitamin D3Expected Conc.(ng/mL)Mean MeasuredConc. (ng/mL)(n=3)% Recovery
1to153.154.8103
1to254.055.6103
2to152.353.5102
1to354.455.8103
3to151.952.9102
1to755.055.7101
7to151.452.7103

Lumipulse G 25-OH Vitamin D recovers known concentrations of supplemental equimolar 25-OH vitamin D2 and 25-OH vitamin D3. A study was performed where known concentrations of 25-OH vitamin D (25-OH vitamin D2 and 25-OH vitamin D3) were added to human serum and Ko EDTA plasma samples with low endogenous 25-OH vitamin D levels. The concentration of 25-0H vitamin D was determined using the Lumipulse G 25-OH Vitamin D and the resulting percent recovery was calculated. 25-OH vitamin D recovery is 100 ± 10%, ranging from 92% to 106%. The data are summarized in the following table.

SampleMean MeasuredConc.(ng/mL)(n=3)Expected Conc.(ng/mL)%Recovery
Serum 129.531.095
44.245.398
57.757.9100
89.989.8100
112.9112.4100
137.1135.7101
Serum 223.924.199
38.038.499
50.651.099
83.282.9100
105.0105.5100
136.3128.8106
Serum 329.329.0101
43.743.3101
56.155.9100

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes arranged around a white teardrop shape. To the right of the graphic, the word "FUJIREBIO" is written in a bold, blue sans-serif font. Below "FUJIREBIO", the words "Diagnostics, Inc." are written in a smaller, gray sans-serif font.

SampleMean MeasuredConc.(ng/mL)(n=3)Expected Conc.(ng/mL)%Recovery
88.587.8101
113.1110.4102
137.3133.7103
26.726.5101
38.140.594
52.156.093
Plasma 183.386.097
110.2114.896
137.9143.496
23.123.997
34.937.992
49.053.492
Plasma 279.983.496
106.9112.295
134.9140.896
20.721.895
33.435.893
48.351.394
Plasma 377.281.395
103.8110.194
130.5138.794

High dose hook is a phenomenon whereby very high level specimens may read within the dynamic range of the assay. For Lumipulse G 25-OH Vitamin D, no high dose hook effect was observed for samples containing up to 3,000 ng/mL of 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3.

C. Traceability, Stability, Expected values (controls, calibrators, or methods):

Lumipulse G 25-OH Vitamin D Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G 25-OH Vitamin D on the LUMIPULSE G System for the quantitative determination of 25-hydroxyvitamin D (25-OH Vitamin D) in human serum and plasma (sodium heparin, lithium heparin, or dipotassium EDTA).

6 Bottles (1.5 mL each) are supplied for the LUMIPULSE G 25-OH Vitamin D Calibrators. Preservative: Sodium azide. The calibrators are at the following concentrations:

CalibratorConcentration(ng/mL)Concentration(nmol/L)
CAL 100
CAL 21025
CAL 32050
CAL 450125

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and green shapes, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, black font.

CAL 5100250
CAL 6150375

The calibrators for use with Lumipulse G 25-OH Vitamin D are prepared gravimetrically and are traceable to internal reference calibrator concentrations determined by UV spectrophotometric analysis and verified by Reference Method Procedure (University of Ghent).

Shelf life

The shelf life for Lumipulse G 25-OH Vitamin D Immunoreaction Cartridges and the Lumipulse G 25-OH Vitamin D Calibrators is 12 months at 2-10°C.

On board the LUMIPULSE G1200

The Lumipulse G 25-OH Vitamin D Immunoreaction Cartridges are sealed unit dose stored at 2-10°C. To reduce risk for any misuse, the package insert states "The Lumipulse G 25-OH Vitamin D Immunoreaction Cartridges can be stored (refrigerated unit) on-board the LUMIPULSE G System for a maximum of 30 days".

The package insert recommends calibrator curve storage on the LUMIPULSE G1200 for a maximum of 30 days.

Transport Conditions

Lumipulse G 25-OH Vitamin D Immunoreaction Cartridges and the Lumipulse G 25-OH Vitamin D Calibrators are shipped at 2-10°C.

Materials will be shipped to the end user using an insulated container and a predetermined configuration of gel (cold and/or frozen) packs to maintain the product for up to 72 hours when stored at ambient temperature.

  • d. Detection limit:
    The Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) of Lumipulse G 25-OH Vitamin D on the LUMIPULSE G1200 is ≤ 4.0 ng/mL.

    1. LoB and LoD
      The LoB for Lumipulse G 25-OH Vitamin D was 0.0 ng/mL.

The LoD for Lumipulse G 25-OH Vitamin D on the LUMIPULSE G1200 was 0.28 ng/mL, determined consistent with the CLSI guideline EP17-A2. Seven low-level specimens were tested over 3 days using two LUMIPULSE G1200 Systems and two Lumipulse G 25-OH Vitamin D lots giving 120 determinations for each panel.

    1. LoQ
      The LOQ for Lumipulse G 25-OH Vitamin D on the LUMIPULSE G1200 was 3.33 ng/mL. determined consistent with the CLSI guideline EP17-A2. The LoQ is the analyte level at which the CV is 10%.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and green graphic to the left of the company name. The company name is written in blue, with the word "FUJIREBIO" on top and "Diagnostics, Inc." on the bottom in a smaller, black font.

e. Analytical specificity:

The Lumipulse G 25-OH Vitamin D demonstrated an average interference of ≤ 10% (for each compound) in a study consistent with the guidelines in the CLSI guideline EP7-A2. Human serum specimens with 25-OH vitamin D concentrations of approximately 20, 40 and 100 ng/mL were supplemented with potentially interfering compounds. The following compounds were tested and found not to interfere with the test.

EndogenousInterferencesTestConcentration
Free Bilirubin (unconjugated)60 mg/dL
Conjugated Bilirubin60 mg/dL
Triglycerides (Intralipid 20% Emulsion)1000 mg/dL
Hemoglobin500 mg/dL
Total Protein (Human Serum Albumin)11 g/dL
Immunoglobulin G (IgG)5 g/dL
Biotin19.7 mg/dL
Uric Acid24 mg/dL
Cholesterol500 mg/dL
L-Ascorbic Acid3 mg/dL
Human Vitamin D Binding Protein0.200 mg/mL
Human Anti-Mouse Antibodies (HAMA)1,000 ng/mL
Rheumatoid Factor (RF)1.000 IU/mL

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and green graphic to the left of the company name. The company name is written in blue, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.

Therapeutic drugTest
InterferencesConcentration
Acetaminophen1455 µmol/L
Acetylsalicylic acid3.65 mmol/L
Alendronate350 mg/L
Ampicillin344 µmol/L
Ascorbic Acid375 µmol/L
Caffeine309 µmol/L
Chloramphenicol155 µmol/L
Digoxin8.7 µmol/L
EinsAlpha (Alfacalcidol)3633 µg/L
Hydrochlorothiazide22.2 µmol/L
Ibandronate52 mg/L
Ibuprofen2486 µmol/L
Indomethacin103 µmol/L
Lidocaine57.9 µmol/L
Lovastatin1932 µmol/L
Metoprolol18.7 µmol/L
Naproxen2247 µmol/L
Pamidron90 mg/L
Risedronate175 mg/L
Theophylline243 µmol/L
Warfarin37.5 µmol/L
Zometa (Zoledronic Acid)4 mg/L

Lumipulse G 25-OH Vitamin D on the LUMIPULSE G1200 System was evaluated for cross-reactivity of the assay with other substances that are similar in structure to 25(OH) Vitamin D in a study consistent with the guidelines in the CLSI Protocol EP7-A2. Human serum specimens with 25-OH vitamin D concentrations of approximately 20, 40 and 100 ng/mL were supplemented with potentially cross-reacting compounds. The compounds were tested at the concentrations listed below and found to have the following percent cross-reactivity.

SubstanceTest Concentration(ng/mL)% Cross-Reactivity
3-epi-25(OH) vitamin D31002
3-epi-25(OH) vitamin D21000
1,25(OH)2 vitamin D210143
1,25(OH)2 vitamin D210024
1,25(OH)2 vitamin D310039
24,25(OH)2 vitamin D310021
Vitamin D3 (Cholecalciferol)20,0000
Vitamin D2 (Ergocalciferol)20,0000
1αOH Vitamin D3 (alfacalcidol)8,0000
Paricalcitol (Zemplar)25-2

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes that resemble a water droplet. To the right of the graphic is the company name, "FUJIREBIO," in a bold, blue sans-serif font, with the registered trademark symbol next to it. Below the company name is the text "Diagnostics, Inc." in a smaller, black sans-serif font.

  • f. Assay cut-off:
    See Clinical Cutoff in 4. below.

    1. Comparison studies:
    • a. Method Comparison Liaison® 25 OH Vitamin D TOTAL

Lumipulse G 25-OH Vitamin D method comparison study was performed using specimens consistent with CLSI guideline EP9-A3. The weighted Deming regression method was used to compare Lumipulse G 25-OH Vitamin D to Liaison® 25 OH Vitamin D TOTAL. The data are summarized in the following table and presented graphically.

Range of samples: 4.0 to 107.3 ng/mL. (Lumipulse G 25-OH Vitamin D) 5.45 to 91.5 ng/mL (Liaison® 25 OH Vitamin D TOTAL)

Lumipulse G 25-OH Vitamin D vs. Liaison® 25 OH Vitamin D TOTAL
------------------------------------------------------------------
nCorrelationCoefficient(r)Intercept(95% CI)Slope(95% CI)Average Bias(ng/mL)
1370.9476-3.0731.090.423
-4.357 –-1.7901.04 – 1.13

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes forming a circular design with a white teardrop shape in the center. To the right of the graphic, the company name "FUJIREBIO" is written in large, bold, blue letters, with "Diagnostics, Inc." written in smaller, gray letters below it.

Image /page/14/Figure/2 description: The image is a scatter plot titled "Deming regression". The x-axis is labeled "LIAISON ng/mL" and ranges from 0 to 80. The y-axis is labeled "Lumipulse ng/mL" and ranges from 0 to 100. The plot shows a positive correlation between the two variables, with a regression line and confidence intervals displayed.

b. Method Comparison – ID- HPLC-MS/MS

To evaluate accuracy of Lumipulse G 25-OH Vitamin D, an additional method comparison study was performed using specimens consistent with CLSI guideline EP9-A3. The weighted Deming regression method was used to compare Lumipulse G 25-OH Vitamin D to reference Vitamin D values determined by the CDC reference method, a validated isotope dilution high performance liquid chromatography tandem mass spectrometry (ID- HPLC-MS/MS) 25-OH Vitamin D. The data are summarized in the following table and presented graphically.

Range of samples: 7.2 to 149.0 ng/mL (Lumipulse G 25-OH Vitamin D) 8.65 to 153 ng/mL (ID- HPLC-MS/MS)

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. On the left side of the logo is a blue and green abstract design. To the right of the design is the company name, "FUJIREBIO" in large, blue, sans-serif font. Below the company name is "Diagnostics, Inc." in a smaller, gray, sans-serif font.

Lumipulse G 25-OH Vitamin D vs. ID- HPLC-MS/MS
nCorrelationCoefficient(r)Intercept(95% CI)Slope(95% CI)Average Bias(ng/mL)
1170.9986-2.7880.97-3.826
-3.225 –-2.3510.96 – 0.99

Image /page/15/Figure/3 description: The image is a scatter plot showing the correlation between Lumipulse and ID-HPLC-MS/MS 25-OH Vitamin D measurements. The x-axis represents ID-HPLC-MS/MS 25-OH Vitamin D in ng/mL, ranging from approximately 0 to 150. The y-axis represents Lumipulse in ng/mL, also ranging from approximately 0 to 140. A Deming regression line is plotted through the data points, along with dashed lines indicating the confidence interval.

Matrix Comparison ﻥ

Lumipulse G 25-OH Vitamin D on the LUMIPULSE G1200 System was evaluated for matrix differences by performing a study using fifty (50) matched sets of serum (red top and serum separator tubes (SST)) and plasma (sodium heparin, lithium heparin, K2EDTA) samples. The results demonstrated equivalency between matrices and are presented as regression equations in the following table:

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. On the left is a blue and green abstract symbol. To the right of the symbol is the company name, "FUJIREBIO" in large, blue, bold letters. Below that is "Diagnostics, Inc." in smaller, black letters.

Matrix ComparisonSampleRange(ng/mL)Regression Equation andCorrelation Coefficient
SST versus Red Top8.6 –113.7$y= 1.0076(x) – 0.2178;$$r=0.9981$
K₂EDTA versus Red Top8.6 –111.2$y= 0.9673(x) + 0.0008;$$r=0.9982$
Lithium Heparin versus Red Top8.8 –119.2$y= 1.019(x) – 0.5969;$$r=0.9964$
Sodium Heparin versus Red Top8.8 –116.3$y= 1.0103(x) – 0.4566;$$r=0.9983$

3. Clinical studies:

  • a. Clinical sensitivity:
    Not applicable

  • b. Clinical specificity:
    Not applicable

  • c. Other clinical supportive data (when a. and b. are not applicable):
    Not applicable

4. Clinical cut-off:

See Expected Values below.

5. Expected values/Reference range:

It is recommended that each laboratory establish its own range, which may be unique to the population it serves depending upon geographical, patient, and environmental factors.

There is considerable discussion of the serum concentrations of 25-OH Vitamin D associated with deficiency, adequacy for bone health and optimal overall health. In the past, vitamin D deficiency was defined as serum 25-0H vitamin D below 10 ng/mL. The World Health Organization (WHO) defined vitamin D insufficiency as serum 25-OH vitamin D below 20 ng/mL. However, others recently started to define vitamin D deficiency as serum 25-OH vitamin D level below 20 ng/mL and vitamin D insufficiency as less than 30 ng/mL. A review of the available literature suggests the following recommendations for 25-0H vitamin D levels:

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and green graphic to the left of the company name. The company name is written in blue, with "Diagnostics, Inc." written in a smaller font size below "FUJIREBIO".

LevelRanges according to WorldHealth OrganizationRanges according toEndocrine Society ClinicalPractice Guideline
Deficient< 10 ng/mL< 20 ng/mL
Insufficient10 – 19 ng/mL20 – 29 ng/mL
Sufficient20 – 100 ng/mL30 – 100 ng/mL
Potential Toxicity>100 ng/mL>100 ng/mL

Results from a total of 322 samples were used to determine the distribution of Lumipulse G 25-OH Vitamin D results in apparently healthy adults.

Excluded from the study were individuals who had history of current use of dietary supplements containing > 2,000 IU per day of Vitamin D, history of Vitamin D deficiency, any disease considered chronic, history of seizures, bariatric surgery, parathyroid and thyroid disease, were pregnant or lactating, had an active malignancy or were diagnosed with cancer other than basal/squamous cell skin cancer within 5 years, were receiving chemotherapy or radiation treatment, had a family history of parathyroid or calcium regulatory disease, were on medicines known to affect absorption or increase catabolism, or had abnormal serum levels for calcium PTH or TSH.

The subjects ranged in age from 18 to 74 years (median age of 36) and were representative of the overall US population in terms of sex (46.9% male and 53.1% female) and ethnicity/race (60.0% White, 7.5% Black, 19.3% Hispanic, 8.1% Asian/Pacific Islander and 5.3% Other). The majority of the population was < 50 years old (231/322, 71.7%) with a median age of 28.

To represent a broad spectrum of UV light exposure, the study population included adult subjects from 2 geographically diverse regions of the US (North: Northern California and New Jersey and South: Southern California and North Carolina) that were sampled during spring/summer (April through September) and fall/winter (October through March) seasons. The majority of patients was from Northern sites (247/322, 76.7%) and sampled during the fall/winter season (241/322, 74.8%).

To represent a broad spectrum of skin types, the assessment of skin type was done using the Fitzpatrick Skin Test questionnaire, which was completed by each study participant. Subjects were classified into six categories of the Fitzpatrick scale. There were 265 (82.3%) lighter skinned subjects (Fitzpatrick scale Types I, III) and 57 (17.7%) darker skinned subjects (Fitzpatrick scale Types IV. V. The majority of subjects (198/322, 61.5%) were Type III. Type I and Type VI had the smallest populations (0.6% and 0.3%, respectively).

The observed range of 25(OH) vitamin D concentrations, established according to Clinical and Laboratory Standards Institute (CLSI) Protocol EP28-A3c is summarized in the table below.

nMean Conc.MedianConc.Observed Range(2.5th to 97.5th percentile)
28722.9ng/mL22.2ng/mL7.6 to 47.6ng/mL

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a blue abstract design on the left, resembling a stylized water droplet or a stylized flower. To the right of the design, the text "FUJIREBIO" is written in a bold, blue sans-serif font. Below "FUJIREBIO", the text "Diagnostics, Inc." is written in a smaller, black sans-serif font.

It is recommended that each laboratory establish its own reference value for the population of interest for Lumipulse G 25-OH Vitamin D on the LUMIPULSE G1200 System.

Standard/Guidance Document Referenced:

  • . ISO 17511:2003 Measurement of Quantities in Biological Samples - Metrological Traceability of Values Assigned to Calibrator and Control Materials
  • . CLSI EP5-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
  • . CLSI EP7-A2 - Interference Testing in Clinical Chemistry: Approved Guideline-Second Edition
  • CLSI EP28-A3c - Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline-Third Edition
  • . CLSI EP17-A2 - Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline Second Edition
  • . CLSI EP6-A - Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
  • . CLSI EP9-A3 - Measurement Procedure Comparison and Bias Estimation Using Patient Samples; approved Guideline - Third Edition
  • CLSI EP25-A - Evaluation of Stability of In Vitro Diagnostic Reagents: Approved Guideline
  • . Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
  • . Test Principle: Guidance for Industry and Food and Drug Administration Staff - eCopy Program for Medical Device Submissions (December 31, 2012)
  • . Guidance for Industry and Food and Drug Administration Staff - Refuse to Accept Policy for 510(k)s (August 4, 2015)

Conclusion

The results of these analytical (nonclinical) and clinical studies demonstrate that the Lumipulse G 25-OH Vitamin D assay is substantially equivalent to the performance of the Liaison® 25 OH Vitamin D TOTAL assay.

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.