K Number
K153351

Validate with FDA (Live)

Device Name
Quantib Brain 1
Manufacturer
Date Cleared
2016-06-17

(210 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.

Device Description

Quantib™ Brain is post-processing analysis software for the GE Advantage (AW 4.7) or AW Server (AWS 3.2) platforms using Volume Viewer Apps. 12.3 Ext 6. It is intended for automatic labeling, visualization, and volumetric quantification of identifiable brain structures from magnetic resonance images (a 3D T1-weighted MR image, with an additional T2-weighted FLAIR MR image for white matter hyperintensities (WMH) segmentation). The segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments. Quantib™ Brain provides quantitative information on both the absolute and relative volume of the segmented regions. The automatic WMH segmentation is to be reviewed and if necessary, edited by the user before validation of the segmentation, after which volumetric information is accessible. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the radiology specialist in quantitative reporting.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for Quantib™ Brain 1, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" but rather presents a performance study comparing the device's output to manual segmentations. Therefore, the "acceptance criteria" are inferred from the reported performance values that were deemed acceptable for market clearance.

Metric (Inferred Acceptance Criterion)Reported Device Performance (Mean ± Standard Deviation)
Brain Tissue Segmentation (GM, WM, CSF)
Dice Index for CSF0.78 ± 0.05
Dice Index for GM0.83 ± 0.02
Dice Index for WM0.86 ± 0.02
Absolute Difference of Relative CSF Volume (pp)1.6 ± 1.0
Absolute Difference of Relative GM Volume (pp)2.8 ± 1.9
Absolute Difference of Relative WM Volume (pp)2.6 ± 1.6
ICV Segmentation
Dice Index for ICV0.97 ± 0.01
White Matter Hyperintensity (WMH) Segmentation
Dice Index for WMH0.61 ± 0.13
Absolute Difference of Relative WMH Volume (pp)0.6 ± 0.7

Note: The document states that the performance data "shows that Quantib™ Brain is as safe and effective as the predicate device," implying these performance metrics were sufficient to demonstrate substantial equivalence.


2. Sample Size Used for the Test Set and Data Provenance

  • Brain Tissue Segmentation (GM, WM, CSF, ICV):

    • Sample Size: 33 T1w MR images with 6 selected slices per scan for comparison.
    • Data Provenance: The set was "carefully selected to include data from multiple vendors and a series of representable scan settings." The document does not specify the country of origin or whether the data was retrospective or prospective.
  • White Matter Hyperintensity (WMH) Segmentation:

    • Sample Size: 30 3D T1w images with corresponding T2w FLAIR images.
    • Data Provenance: This set also "represented various scan settings." The document does not specify the country of origin or whether the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. It simply refers to "manual segmentations" for brain tissues and "manually segmented" WMHs.


4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1) for establishing the ground truth or resolving discrepancies in manual segmentations. It merely states "manual segmentations."


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, the document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The study focuses on comparing the algorithm's performance against manual segmentations, not on how human readers improve with or without AI assistance.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The reported "Algorithm performance" (Section VII.2) directly compares the Quantib™ Brain's automatic segmentations and volume measurements to manual segmentations for the same scans, without human intervention in the device's output during the test. The "Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs" statement in the Indications for Use refers to the intended use case, not the performance validation study's methodology. The study itself assesses the raw algorithmic output.


7. The Type of Ground Truth Used

The ground truth used was expert manual segmentation.

  • For brain tissue volumetry (GM, WM, CSF, ICV), the device's relative brain tissue volumes were compared to "relative volumes derived from manual segmentations."
  • For WMH, "WMHs were manually segmented on the T2w FLAR images" and compared to the device's automatic segmentation.

8. The Sample Size for the Training Set

The document does not specify the sample size for the training set. It mentions that "The segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments," but does not give a number for these atlases or the data used to create them or train the system.


9. How the Ground Truth for the Training Set Was Established

The document does not explicitly state how the ground truth for the training set was established. It mentions that the system "relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments." It can be inferred that these atlases contain expert-defined segmentations (label maps), but the method of their creation (e.g., by how many experts, what qualifications, adjudication) is not detailed.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2016

Quantib B.V. Floor van Leeuwen Quality & Regulatory Manager Westblaak 106 3012 KM Rotterdam NETHERLANDS

Re: K153351 Trade/Device Name: QuantibTM Brain 1 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: May 19, 2016 Received: May 23, 2016

Dear Floor van Leeuwen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K-153351

Device Name Quantib™ Brain 1

Indications for Use (Describe)

Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Figure/0 description: The image shows the logo for Quantib, a company focused on image-informed medicine. The logo features a stylized graphic to the left, composed of stacked, offset rectangles in shades of blue, orange, and brown, creating a visual effect reminiscent of data or imaging slices. To the right of the graphic is the company name "Quantib" in a teal sans-serif font, with the tagline "Image Informed Medicine" in a smaller font size underneath.

510(k) Summary

I. SUBMITTER

Quantib B.V. Westblaak 106 3012 KM Rotterdam Phone: (+31) 108 41 17 49 Contact Person: Floor van Leeuwen Date Prepared: November 18th, 2015

II. DEVICE

Name of Device: Quantib™ Brain 1 Common or Usual Name: Quantib™ Brain Classification Name: System, image processing, radiology (892.2050) Regulatory Class: II Product Code: Picture archiving and communication system (LLZ)

III. PREDICATE DEVICE

Device: QBrain® Manufacturer: Medis medical imaging systems 510(k) Reg. No: K050703 This predicate has not been subject to a design-related recall. Regulatory Class: II Product Code: Picture archiving and communication system (LLZ)

IV. DEVICE DESCRIPTION

Quantib™ Brain is post-processing analysis software for the GE Advantage (AW 4.7) or AW Server (AWS 3.2) platforms using Volume Viewer Apps. 12.3 Ext 6. It is intended for automatic labeling, visualization, and volumetric quantification of identifiable brain structures from magnetic resonance images (a 3D T1weighted MR image, with an additional T2-weighted FLAIR MR image for white matter hyperintensities (WMH) segmentation). The segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments. Quantib™ Brain provides quantitative information on both the absolute and relative volume of the segmented regions. The automatic WMH segmentation is to be reviewed and if necessary, edited by the user before validation of the segmentation, after which volumetric information is accessible. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the radiology specialist in quantitative reporting.

{4}------------------------------------------------

V. INTENDED USE

Intended use Quantib™ Brain

Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.

Intended us predicate device QBrain®

Intended use

The QBrain® software has been developed for the objective and reproducible analysis of MR images of the brain. It performs quantitative analyses on MR brain image based on automatic segmentation. More specifically, it quantifies the volumes of intracranial cavities, areas that contain cerebrospinal fluid (CSF), and white matter hyperintensities (lesions).

Indications for use

QBrain® is able to read DCIOM MR image from all major MRI vendors. Mask data, generated by automatic segmentation and/or manual editing, and quantitative results ca be saved in separate files enabling the comparison of results from different users and easy export to standard spreadsheet software.

Neur-(radio)logists in hospitals and specialists in core labs use the QBrain stand-alone analytical software package in image post-processing. The provided objective and quantitative values support the diagnostic decision process or are used in the evaluation of follow-up studies about on and/or therapy response.

Intended use comparison

Both device and predicate device are designed to assist trained medical professionals, i.e. radiologists, in the evaluation and assessment of MR brain images with a set of tools for visualizing the location and quantification of brain structures. Both are intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on MR images.

{5}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Both device and predicate device are software for the automatic labeling, visualization, and quantification (segmentation) of the volume of specific areas of the brain. Atlas-based segmentation is the technological principle for both devices.

Both devices require a 3D T1w MR image for tissue-type segment White Matter Hyperintensities (WMH), Quantib™ Brain requires an additional T2w FLAIR scan. Histogram analysis and trained classifiers are used in combination with atlas-based tissue segmentation to improve the accuracy and specificity of WMH segmentation.

Quantib™ Brain reports the following measurements: absolute volume and relative to ICV) of grey matter (GM), white matter (WM), cerebrospinal fluid (CSF), and intracranial volume (ICV); brain volume (=GM+WM); total volume of WM hyperintensities (WMH); relative WMH volume (relative to WM); total number of WMH.

Quantib™ BrainK050703 QBrain®
Brain tissue and WMH segmentationBrain tissue segmentation (intracranialtissue volume GM+WM, CSF volume, lobevolumes, cerebellum volume). WMHsegmentation (total WMH volume).
Absolute and relative volumesAbsolute volumes
Software plugs into the AW and AW serverplatformsStand-alone application

Following are the differences between Quantib™ Brain and the predicate device:

Table 5-1 Differences with predicate device

VII. PERFORMANCE DATA:

1. Quality and safety

Quantib™ Brain was designed in compliance with the following process standards:

  • ISO 14971 – Medical devices - Application of risk management to medical devices
  • . IEC 62304 – Medical device software – Software life cycle processes

The following quality assurance measures were applied to Quantib™ Brain development:

  • Risk and hazard analysis
  • Design reviews
  • Unit level testing
  • . Integration testing
  • System testing
  • Performance testing
  • . Usability engineering

2. Algorithm performance

{6}------------------------------------------------

To validate the quality of Quantib™ Brain volume measurements and segmentations, we compared the relative brain tissue volumes to relative volumes derived from manual segmentations for the same scan. We performed this analysis for GM, WM, CSF, ICV, and WMHs.

For the brain volumetry protocol (segmentation and measures of GM, WM, CSF, and ICV) the test set included 33 T1w MR images. The set was carefully selected to include data from multiple vendors and a series of representable scan settings. For each scan we selected six (6) slices for comparison. The results are summarized in Table 5-2.

Dice indexAbsolute difference of therelative volumes [pp]
CSF0.78 ± 0.051.6 ± 1.0
GM0.83 ± 0.022.8 ± 1.9
WM0.86 ± 0.022.6 ± 1.6
ICV0.97 ± 0.01

Table 5-2 Results of comparison between manual and automatic brain tissue seqmentation. Reported values are averages ± std. dev., computed over 6 segmented slices of 33 scans. The Dice index provides a measure for overlap of manual and automatic segmentations (1 = perfect overlap). The absolute differences of the relative volumes (of the brain tissues) are averages ± std. dev. in percentage points.

The test set for the WMH protocol included 30 3D T1w images with corresponding T2w FLAIR images. This set also represented various scan settings. WMHs were manually segmented on the T2w FLAR images and compared to Quantib™ Brain automatic segmentation output. The average Dice overlap between the manual segmentations and Quantib™ Brain segmentations was 0.61 ± 0.13. The absolute difference of the relative volumes (for WMHs) was 0.6 ± 0.7 percentage points.

VIII. CONCLUSIONS

By virtue of its intended use and physical and technological characteristics, Quantib™ Brain is substantially equivalent to a device that has been approved for marketing in the United States. The performance data shows that Quantib™ Brain is as safe and effective as the predicate device.

N/A